| | Class 2 Device Recall ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES |  |
| Date Initiated by Firm | November 24, 2021 |
| Create Date | March 04, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0724-2022 |
| Recall Event ID |
89375 |
| 510(K)Number | K141979 K161563 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray:
Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039.
Allura Xper F010 {C+F). 722003
Allura Xper F010/10 722005
Allura Xper F020 722006
Allura Xper F020/10 and F020/20 722008
Allura Xper F010 722010
Altura Xper F010/10 722011
Allura Xper F020 722012
Allura Xper FD20 biplane 722013
Allura Xper FDlO OR Table 722014
Allura Xper FD20 OR Table 722015
Allura Xper FDl0/10 OR Table 722019
Allura Xper FD20 Biplane OR Table 722020
Allura Xper FD10 OR Table 722022
Allura Xper FD20 OR Table 722023
Allura Xper FD10/10 OR Table 722024
Altura Xper FD20 Biplane OR Table 722025
Allura Xper FD10 722026
Allura Xper FD10/10 722027
Altura Xper FD20/10 biplane 722029
Allura Xper FD10 OR Table 722033
Allura Xper FD10/10 OR Table 722034
Allura Xper FD20/10 OR Table 722036
Allure Xper F020/20 722038
Altura Xper FD20/20 biplane OR Table 722039 |
| Code Information |
Allura Xper FD10 722003/722026 (US) S/N: 24 40 71 162 191 266 283 294 319 347 406 485 825 959 965 967 985 990 1246 1353 1444 1944 1949 UDI: (01)00884838054189(21)825 (01)00884838054189(21)967 (01)00884838054189(21)1246 (01)00884838054189(21)965 (01)00884838054189(21)959 (01)00884838054189(21)990 (01)00884838054189(21)985 Allura Xper FD10/10 722005 S/N: 144 US OUS: 37 52 74 79 80 108 279 417 459 Allura Xper FD20 722006 S/N: US 44 72 234 246 263 268 327 333 429 469 471 498 514 535 550 554 588 604 718 754 766 861 1174 1274 1355 1400 1783 1831 1844 1878 1886 1931 2018 UDI: (01)00884838054202(21)893 OUS: 11 40 89 102 112 125 151 244 364 408 431 484 568 580 594 597 602 640 643 704 811 940 971 1262 1312 1402 1405 1424 1465 1481 1658 1691 1794 1796 1815 1852 1859 1879 1891 1948 1959 1974 1985 1988 2141 2311 2521 2554 Allura Xper FD20 Biplane 722008 S/N: 33, 82, 374 (US) OUS: 39 86 224 249 336 384 388 532 Allura Xper FD10 722010 S/N:OUS 163 340 1422 1429 1434 Allura Xper FD20 722012 S/N: OUS 395 463 927 1013 1518 2652 2854 2918 Allura Xper FD20 OR Table 722015 S/N: 30 (OUS) Allura Xper FD10 722026 S/N OUS: 315 701 748 884 888 943 948 960 971 975 978 984 989 995 Allura Xper FD10/10 722027 S/N: OUS 268 278 Allura Xper FD20/10 722029 S/N:OUS 9 Allura Xper FD10 722026 S/N : US 825 959 965 967 985 990 1246 OUS: 11 40 89 102 112 125 151 244 364 408 431 484 568 580 594 597 602 640 643 704 811 940 971 1262 1312 1402 1405 1424 1465 1481 1658 1691 1794 1796 1815 1852 1859 1879 1891 1948 1959 1974 1985 1988 2141 2311 2521 2554 Allura Xper FD10/10 266 722027 S/N: 266 Allura Xper FD20 722028 S/N: 893 UDI: (01)00884838054202(21)893 Allura Xper FD20/10 722029 S/N: 91, 94 US UDI: (01)00884838054219(21)94 OUS: S/N:6 |
| FEI Number |
3016618143
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
Manufacturer Reason for Recall | Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire. |
FDA Determined Cause 2 | Process control |
| Action | Philips Issued Medical Device Correction letter dated 11/24/21 to U.S.A, Customers via certified mailing from the United States Postal Service (USPS). Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk and action to take:
Philips will inspect the affected systems for leaks and repair as required. In addition, Philips will install an extended drip tray to prevent damage from any future leaks. This service will be provided free of charge for affected systems.
A Philips service representative will contact customers with affected devices to arrange for the service.
If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377 (reference FCC 72200484). |
| Quantity in Commerce | USA: 63 systems R.O.W.: 127 systems |
| Distribution | US Nationwide
Foreign:
Argentina
Austria
Bangladesh
Brazil
Chile
China
Denmark
Dominican Republic
France
Germany
Greece
India
Indonesia
Iran
Israel
Italy
Japan
Korea, Republic of
Malaysia
Mexico
Netherlands
Philippines
Russian Federation
Saudi Arabia
South Africa
Spain
Sri Lanka
Switzerland
Taiwan
Thailand
Turkey
United Kingdom
Virgin Islands (U.S.) |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB 510(K)s with Product Code = OWB
|
|
|
|