• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Drager Infinity CentralStation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Drager Infinity CentralStationsee related information
Date Initiated by FirmMarch 04, 2022
Create DateApril 27, 2022
Recall Status1 Open3, Classified
Recall NumberZ-1012-2022
Recall Event ID 89877
510(K)NumberK130711 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductDrager Infinity CentralStation Software versions VG2.1.2 and VG2.1.2 SU12. For patient monitoring with Infinity M300 wireless telemetry devices.
Code Information Unique Device Identifier (UDI) for US distribution: (01)04049098001878(11)200626(21)1OQ84F-01013(93)MS26800-40, (01)04049098001878(11)190516(21)1OQ80D-00211(93)MS26800-36, (01)04049098001878(11)191218(21)1OQ84F-00362(93)MS26800-37, (01)04049098001878(11)200323(21)1OQ84F-00617(93)MS26800-38, (01)04049098001878(11)190903(21)1OQ84F-00103(93)MS26800-37, (01)04049098001878(11)190903(21)1OQ84F-00105(93)MS26800-37, (01)04049098001878(11)190903(21)1OQ84F-00104(93)MS26800-37, (01)04049098001878(11)190903(21)1OQ84F-00099(93)MS26800-37, (01)04049098001878(11)190807(21)1OQ84F-00065(93)MS26800-36, (01)04049098001878(11)190807(21)1OQ84F-00081(93)MS26800-36, (01)04049098001878(11)190807(21)1OQ84F-00120(93)MS26800-36, (01)04049098001878(11)190807(21)1OQ84F-00080(93)MS26800-36, (01)04049098001878(11)190807(21)1OQ84F-00070(93)MS26800-36, (01)04049098001878(11)200630(21)1OQ84F-01005(93)MS26800-40, (01)04049098001878(11)190730(21)1OQ80D-00059(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ84F-00043(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ7XL-00479(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ7XL-00427(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ80D-00010(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ7XL-00792(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ84F-00088(93)MS26800-36, (01)04049098001878(11)190730(21)1OQ7XL-00067(93)MS26800-36, (01)04049098001878(11)200323(21)1OQ84F-00606(93)MS26800-38, (01)04049098001878(11)190830(21)1OQ84F-00110(93)MS26800-37, (01)04049098001878(11)190902(21)1OQ84F-00098(93)MS26800-36, (01)04049098001878(11)200911(21)1OQ84F-01196(93)MS26800-41, (01)04049098001878(11)200123(21)1OQ84F-00496(93)MS26800-37, (01)04049098001878(11)200123(21)1OQ84F-00483(93)MS26800-37, (01)04049098001878(11)200123(21)1OQ84F-00507(93)MS26800-37, (01)04049098001878(11)200123(21)1OQ84F-00480(93)MS26800-37, (01)04049098001878(11)191217(21)1OQ84F-00364(93)MS26800-38, (01)04049098001878(11)200304(21)1OQ84F-00559(93)MS26800-38, (01)04049098001878(11)200228(21)1OQ84F-00548(93)MS26800-38, (01)04049098001878(11)200313(21)1OQ84F-00575(93)MS26800-38, (01)04049098001878(11)200720(21)1OQ84F-01030(93)MS26800-40, (01)04049098001878(11)200121(21)1OQ84F-00485(93)MS26800-38, (01)04049098001878(11)200220(21)1OQ84F-00516(93)MS26800-38, (01)04049098001878(11)190506(21)1OQ80D-00193(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00211(93)MS26800-36, (01)04049098001878(11)200124(21)1OQ84F-00481(93)MS26800-37, (01)04049098001878(11)200124(21)1OQ84F-00500(93)MS26800-37, (01)04049098001878(11)200420(21)1OQ84F-00828(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00826(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00803(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00830(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00796(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00832(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00824(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00806(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00829(93)MS26800-40, (01)04049098001878(11)200420(21)1OQ84F-00804(93)MS26800-40, (01)04049098001878(11)200911(21)1OQ84F-01243(93)MS26800-41, (01)04049098001878(11)200911(21)1OQ84F-01212(93)MS26800-41, (01)04049098001878(11)201117(21)1OQ84F-01461(93)MS26800-43, (01)04049098001878(11)201127(21)1OQ84F-01520(93)MS26800-43, (01)04049098001878(11)201127(21)1OQ84F-01456(93)MS26800-43, (01)04049098001878(11)201127(21)1OQ84F-01469(93)MS26800-43, (01)04049098001878(11)201209(21)1OQ84F-01375(93)MS26800-43, (01)04049098001878(11)201123(21)1OQ84F-01429(93)MS26800-43, (01)04049098001878(11)190702(21)1OQ80D-00338(93)MS26800-36, (01)04049098001878(11)190703(21)1OQ80D-00368(93)MS26800-36, (01)04049098001878(11)201113(21)1OQ84F-01428(93)MS26800-43, (01)04049098001878(11)201207(21)1OQ84F-01252(93)MS26800-43, (01)04049098001878(11)210115(21)1OQ84F-01598(93)MS26800-43, (01)04049098001878(11)200922(21)1OQ84F-01257(93)MS26800-41, (01)04049098001878(11)201214(21)1OQ84F-01378(93)MS26800-43, (01)04049098001878(11)200819(21)1OQ84F-01073(93)MS26800-40, (01)04049098001878(11)200123(21)1OQ84F-00482(93)MS26800-38, (01)04049098001878(11)191209(21)1OQ84F-00394(93)MS26800-37, (01)04049098001878(11)200605(21)1OQ84F-00932(93)MS26800-40, (01)04049098001878(11)200323(21)1OQ84F-00594(93)MS26800-38, (01)04049098001878(11)200323(21)1OQ84F-00602(93)MS26800-38, (01)04049098001878(11)190308(21)1OQ80D-00156(93)MS26800-36, (01)04049098001878(11)200806(21)1OQ84F-01118(93)MS26800-40, (01)04049098001878(11)200806(21)1OQ84F-01115(93)MS26800-40, (01)04049098001878(11)200806(21)1OQ84F-01130(93)MS26800-40, (01)04049098001878(11)190212(21)1OQ80D-00057(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00062(93)MS26800-36, (01)04049098001878(11)200115(21)1OQ84F-00424(93)MS26800-38, (01)04049098001878(11)190703(21)1OQ80D-00361(93)MS26800-36, (01)04049098001878(11)190703(21)1OQ80D-00366(93)MS26800-36, (01)04049098001878(11)190830(21)1OQ84F-00107(93)MS26800-36, (01)04049098001878(11)190830(21)1OQ84F-00121(93)MS26800-36, (01)04049098001878(11)190830(21)1OQ84F-00094(93)MS26800-36, (01)04049098001878(11)200515(21)1OQ84F-00849(93)MS26800-40, (01)04049098001878(11)200515(21)1OQ84F-00897(93)MS26800-40, (01)04049098001878(11)200702(21)1OQ84F-01017(93)MS26800-40, (01)04049098001878(11)190830(21)1OQ84F-00100(93)MS26800-36, (01)04049098001878(11)191217(21)1OQ84F-00365(93)MS26800-38, (01)04049098001878(11)190213(21)1OQ80D-00056(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ7XL-00407(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00045(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00040(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00043(93)MS26800-36, (01)04049098001878(11)190926(21)1OQ84F-00179(93)MS26800-37, (01)04049098001878(11)200619(21)1OQ84F-00913(93)MS26800-40, (01)04049098001878(11)200619(21)1OQ84F-00950(93)MS26800-40, (01)04049098001878(11)201104(21)1OQ84F-01508(93)MS26800-43, (01)04049098001878(11)190708(21)1OQ80D-00369(93)MS26800-36, (01)04049098001878(11)190621(21)1OQ80D-00321(93)MS26800-36, (01)04049098001878(11)190621(21)1OQ80D-00316(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00035(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00036(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00038(93)MS26800-36, (01)04049098001878(11)190213(21)1OQ80D-00049(93)MS26800-36, (01)04049098001878(11)190425(21)1OQ80D-00172(93)MS26800-36, (01)04049098001878(11)190510(21)1OQ80D-00187(93)MS26800-36, (01)04049098001878(11)200525(21)1OQ84F-00952(93)MS26800-40, (01)04049098001878(11)200827(21)1OQ84F-01164(93)MS26800-40, (01)04049098001878(11)200828(21)1OQ84F-01169(93)MS26800-40, (01)04049098001878(11)200828(21)1OQ84F-01158(93)MS26800-40, (01)04049098001878(11)200828(21)1OQ84F-01160(93)MS26800-40, (01)04049098001878(11)200814(21)1OQ84F-01170(93)MS26800-40, (01)04049098001878(11)190214(21)1OQ80D-00039(93)MS26800-36, (01)04049098001878(11)190826(21)1OQ84F-00086(93)MS26800-37, (01)04049098001878(11)191016(21)1OQ84F-00199(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00245(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00191(93)MS26800-36, (01)04049098001878(11)190816(21)1OQ84F-00091(93)MS26800-36, (01)04049098001878(11)190816(21)1OQ84F-00055(93)MS26800-36, (01)04049098001878(11)190816(21)1OQ84F-00056(93)MS26800-36, (01)04049098001878(11)190816(21)1OQ84F-00059(93)MS26800-36, (01)04049098001878(11)190816(21)1OQ84F-00054(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00251(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00198(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00180(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00221(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00189(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00215(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00234(93)MS26800-36, (01)04049098001878(11)191016(21)1OQ84F-00218(93)MS26800-36, (01)04049098001878(11)190515(21)1OQ80D-00205(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00264(93)MS26800-36. (01)04049098001878(11)190527(21)1OQ80D-00262(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00263(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00273(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00265(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00258(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00261(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00276(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00281(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00282(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00268(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00284(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00269(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00287(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00286(93)MS26800-36, (01)04049098001878(11)190527(21)1OQ80D-00278(93)MS26800-36, (01)04049098001878(11)190902(21)1OQ84F-00096(93)MS26800-36, (01)04049098001878(11)190902(21)1OQ84F-00111(93)MS26800-36, (01)04049098001878(11)190902(21)1OQ84F-00093(93)MS26800-36, (01)04049098001878(11)191126(21)1OQ84F-00328(93)MS26800-38, (01)04049098001878(11)191126(21)1OQ84F-00332(93)MS26800-38, (01)04049098001878(11)191213(21)1OQ84F-00368(93)MS26800-37, (01)04049098001878(11)191213(21)1OQ84F-00367(93)MS26800-37, (01)04049098001878(11)191219(21)1OQ84F-00429(93)MS26800-38, (01)04049098001878(11)191219(21)1OQ84F-00378(93)MS26800-38, (01)04049098001878(11)200701(21)1OQ84F-01006(93)MS26800-40, (01)04049098001878(11)191219(21)1OQ84F-00389(93)MS26800-37, (01)04049098001878(11)190816(21)1OQ84F-00060(93)MS26800-36, (01)04049098001878(11)190705(21)1OQ80D-00362(93)MS26800-37, (01)04049098001878(11)190705(21)1OQ80D-00364(93)MS26800-37, (01)04049098001878(11)190705(21)1OQ80D-00384(93)MS26800-37, (01)04049098001878(11)200828(21)1OQ84F-01177(93)MS26800-40, (01)04049098001878(11)210115(21)1OQ84F-01514(93)MS26800-43, (01)04049098001878(11)190318(21)1OQ80D-00130(93)MS26800-36, (01)04049098001878(11)200728(21)1OQ84F-01088(93)MS26800-40, (01)04049098001878(11)200728(21)1OQ84F-01092(93)MS26800-40, (01)04049098001878(11)200414(21)1OQ84F-00783(93)MS26800-40, (01)04049098001878(11)200414(21)1OQ84F-00759(93)MS26800-40, (01)04049098001878(11)200414(21)1OQ84F-00782(93)MS26800-40, (01)04049098001878(11)200113(21)1OQ84F-00382(93)MS26800-38, (01)04049098001878(11)191212(21)1OQ84F-00347(93)MS26800-37, (01)04049098001878(11)200228(21)1OQ84F-00471(93)MS26800-38, (01)04049098001878(11)200720(21)1OQ84F-01074(93)MS26800-40, (01)04049098001878(11)190705(21)1OQ80D-00358(93)MS26800-37, (01)04049098001878(11)200622(21)1OQ84F-00905(93)MS26800-40, (01)04049098001878(11)190213(21)1OQ80D-00042(93)MS26800-36, (01)04049098001878(11)190919(21)1OQ84F-00160(93)MS26800-37, (01)04049098001878(11)190625(21)1OQ80D-00299(93)MS26800-36, (01)04049098001878(11)190814(21)1OQ84F-00073(93)MS26800-36, (01)04049098001878(11)200626(21)1OQ84F-00996(93)MS26800-40, (01)04049098001878(11)191021(21)1OQ84F-00269(93)MS26800-37, (01)04049098001878(11)190814(21)1OQ84F-00116(93)MS26800-37, (01)04049098001878(11)190726(21)1OQ84F-00109(93)MS26800-37, (01)04049098001878(11)200909(21)1OQ84F-01233(93)MS26800-41, (01)04049098001878(11)200820(21)1OQ84F-01163(93)MS26800-40, (01)04049098001878(11)200220(21)1OQ84F-00513(93)MS26800-38, (01)04049098001878(11)190508(21)1OQ80D-00185(93)MS26800-36, (01)04049098001878(11)190523(21)1OQ80D-00240(93)MS26800-36, (01)04049098001878(11)200324(21)1OQ84F-00598(93)MS26800-38, (01)04049098001878(11)200324(21)1OQ84F-00596(93)MS26800-38, (01)04049098001878(11)200324(21)1OQ84F-00618(93)MS26800-38, (01)04049098001878(11)200324(21)1OQ84F-00622(93)MS26800-38, (01)04049098001878(11)200325(21)1OQ84F-00592(93)MS26800-38, (01)04049098001878(11)190313(21)1OQ80D-00090(93)MS26800-36, (01)04049098001878(11)200716(21)1OQ84F-01031(93)MS26800-40, (01)04049098001878(11)190214(21)1OQ80D-00041(93)MS26800-36, (01)04049098001878(11)190513(21)1OQ80D-00214(93)MS26800-36, (01)04049098001878(11)200402(21)1OQ84F-00716(93)MS26800-38, (01)04049098001878(11)190313(21)1OQ80D-00089(93)MS26800-36, (01)04049098001878(11)190816(21)1OQ84F-00075(93)MS26800-36, (01)04049098001878(11)190528(21)1OQ80D-00244(93)MS26800-36, (01)04049098001878(11)190909(21)1OQ84F-00122(93)MS26800-37, (01)04049098001878(11)200203(21)1OQ84F-00452(93)MS26800-38, (01)04049098001878(11)190916(21)1OQ84F-00144(93)MS26800-36, (01)04049098001878(11)190916(21)1OQ84F-00145(93)MS26800-36, (01)04049098001878(11)190916(21)1OQ84F-00124(93)MS26800-36, (01)04049098001878(11)190515(21)1OQ80D-00207(93)MS26800-36, (01)04049098001878(11)190515(21)1OQ80D-00224(93)MS26800-36, (01)04049098001878(11)190515(21)1OQ80D-00210(93)MS26800-36, (01)04049098001878(11)200828(21)1OQ84F-01157(93)MS26800-40, (01)04049098001878(11)200324(21)1OQ84F-00600(93)MS26800-38, (01)04049098001878(11)200324(21)1OQ84F-00614(93)MS26800-38, (01)04049098001878(11)200325(21)1OQ84F-00612(93)MS26800-38, (01)04049098001878(11)200310(21)1OQ84F-00565(93)MS26800-38, (01)04049098001878(11)200107(21)1OQ84F-00411(93)MS26800-38, (01)04049098001878(11)200220(21)1OQ84F-00512(93)MS26800-38, (01)04049098001878(11)200123(21)1OQ84F-00502(93)MS26800-37, (01)04049098001878(11)190926(21)1OQ84F-00163(93)MS26800-37, (01)04049098001878(11)190926(21)1OQ84F-00175(93)MS26800-37, (01)04049098001878(11)200203(21)1OQ84F-00450(93)MS26800-38, (01)04049098001878(11)190313(21)1OQ80D-00088(93)MS26800-36, (01)04049098001878(11)200228(21)1OQ84F-00550(93)MS26800-38, (01)04049098001878(11)190621(21)1OQ80D-00349(93)MS26800-36, (01)04049098001878(11)190621(21)1OQ80D-00331(93)MS26800-36, (01)04049098001878(11)190621(21)1OQ80D-00351(93)MS26800-36, (01)04049098001878(11)200728(21)1OQ84F-01040(93)MS26800-40, (01)04049098001878(11)190705(21)1OQ80D-00373(93)MS26800-36, (01)04049098001878(11)190520(21)1OQ80D-00227(93)MS26800-36, (01)04049098001878(11)200128(21)1OQ84F-00456(93)MS26800-38, (01)04049098001878(11)190606(21)1OQ80D-00310(93)MS26800-36, (01)04049098001878(11)190708(21)1OQ80D-00370(93)MS26800-36, (01)04049098001878(11)190708(21)1OQ80D-00371(93)MS26800-36, (01)04049098001878(11)190814(21)1OQ7XL-00447(93)MS26800-37, (01)04049098001878(11)190705(21)1OQ80D-00379(93)MS26800-36, (01)04049098001878(11)191030(21)1OQ84F-00261(93)MS26800-36, (01)04049098001878(11)191030(21)1OQ84F-00263(93)MS26800-37, (01)04049098001878(11)200214(21)1OQ84F-00529(93)MS26800-37, (01)04049098001878(11)190709(21)1OQ84F-00008(93)MS26800-36, (01)04049098001878(11)190709(21)1OQ80D-00383(93)MS26800-36, (01)04049098001878(11)190321(21)1OQ80D-00140(93)MS26800-36, (01)04049098001878(11)190705(21)1OQ80D-00385(93)MS26800-36, (01)04049098001878(11)190705(21)1OQ80D-00363(93)MS26800-36, (01)04049098001878(11)200918(21)1OQ84F-01209(93)MS26800-41, (01)04049098001878(11)190814(21)1OQ84F-00047(93)MS26800-37, (01)04049098001878(11)200918(21)1OQ84F-01198(93)MS26800-41, (01)04049098001878(11)190814(21)1OQ84F-00076(93)MS26800-37, (01)04049098001878(11)200122(21)1OQ84F-00505(93)MS26800-38, (01)04049098001878(11)190814(21)1OQ7XL-00365(93)MS26800-37, (01)04049098001878(11)190318(21)1OQ80D-00138(93)MS26800-36, (01)04049098001878(11)200318(21)1OQ84F-00580(93)MS26800-38, (01)04049098001878(11)200928(21)1OQ84F-01315(93)MS26800-43, (01)04049098001878(11)200928(21)1OQ84F-01254(93)MS26800-43, (01)04049098001878(11)190319(21)1OQ80D-00131(93)MS26800-36, (01)04049098001878(11)190808(21)1OQ84F-00052(93)MS26800-36, (01)04049098001878(11)191213(21)1OQ84F-00376(93)MS26800-36, (01)04049098001878(11)200330(21)1OQ84F-00609(93)MS26800-40, (01)04049098001878(11)200330(21)1OQ84F-00610(93)MS26800-40, (01)04049098001878(11)200330(21)1OQ84F-00593(93)MS26800-40, (01)04049098001878(11)200330(21)1OQ84F-00625(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00627(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00615(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00630(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00683(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00623(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00624(93)MS26800-40, (01)04049098001878(11)200331(21)1OQ84F-00631(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00703(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00697(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00707(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00712(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00710(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00692(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00689(93)MS26800-40, (01)04049098001878(11)200401(21)1OQ84F-00674(93)MS26800-40, (01)04049098001878(11)200427(21)1OQ84F-00853(93)MS26800-40, (01)04049098001878(11)200427(21)1OQ84F-00845(93)MS26800-40, (01)04049098001878(11)201210(21)1OQ84F-01477(93)MS26800-43, (01)04049098001878(11)191105(21)1OQ84F-00203(93)MS26800-37, (01)04049098001878(11)191105(21)1OQ84F-00224(93)MS26800-37, (01)04049098001878(11)170214(21)12946M00555(93)MS26800-27, (01)04049098001878(11)171113(21)1OQ7WR-00418(93)MS26800-30, (01)04049098001878(11)181026(21)1OQ7XL-00668(93)MS26800-33, (01)04049098001878(11)181113(21)1OQ7XL-00747(93)MS26800-33, (01)04049098001878(11)181113(21)1OQ7XL-00708(93)MS26800-33, (01)04049098001878(11)181113(21)1OQ7XL-00720(93)MS26800-33, (01)04049098001878(11)181113(21)1OQ7XL-00722(93)MS26800-33, (01)04049098001878(11)181113(21)1OQ7XL-00715(93)MS26800-33, (01)04049098001878(11)181211(21)1OQ80D-00011(93)MS26800-33, (01)04049098001878(11)180903(21)1OQ7XL-00503(93)MS26800-30, (01)04049098001878(11)180903(21)1OQ7XL-00534(93)MS26800-30, (01)04049098001878(11)180627(21)1OQ7XL-00409(93)MS26800-32, (01)04049098001878(11)171005(21)1OQ7WR-00389(93)MS26800-30, (01)04049098001878(11)181126(21)1OQ7XL-00766(93)MS26800-33, (01)04049098001878(11)181002(21)1OQ7XL-00617(93)MS26800-33, (01)04049098001878(11)181106(21)1OQ7XL-00741(93)MS26800-33, (01)04049098001878(11)190114(21)1OQ80D-00054(93)MS26800-33, (01)04049098001878(11)180208(21)1OQ7XL-00064(93)MS26800-31, (01)04049098001878(11)161006(21)12946M00311(93)MS26800-27, (01)04049098001878(11)181207(21)1OQ80D-00007(93)MS26800-33, (01)04049098001878(11)181207(21)1OQ80D-00005(93)MS26800-33, (01)04049098001878(11)181207(21)1OQ80D-00004(93)MS26800-33, (01)04049098001878(11)190117(21)1OQ80D-00061(93)MS26800-33, (01)04049098001878(11)161202(21)12946M00448(93)MS26800-27, (01)04049098001878(11)161202(21)12946M00450(93)MS26800-27, (01)04049098001878(11)180614(21)1OQ7XL-00352(93)MS26800-33, (01)04049098001878(11)180614(21)1OQ7XL-00324(93)MS26800-33, (01)04049098001878(11)180614(21)1OQ7XL-00383(93)MS26800-33, (01)04049098001878(11)180831(21)1OQ7XL-00520(93)MS26800-30, (01)04049098001878(11)180903(21)1OQ7XL-00519(93)MS26800-30, (01)04049098001878(11)180831(21)1OQ7XL-00495(93)MS26800-30, (01)04049098001878(11)170420(21)1OQ7TC-00088(93)MS26800-28, (01)04049098001878(11)170420(21)1OQ7TC-00111(93)MS26800-28, (01)04049098001878(11)180704(21)1OQ7XL-00452(93)MS26800-32, (01)04049098001878(11)181120(21)1OQ7XL-00768(93)MS26800-33, (01)04049098001878(11)170726(21)1OQ7WR-00181(93)MS26800-30, (01)04049098001878(11)170726(21)1OQ7WR-00178(93)MS26800-30, (01)04049098001878(11)180607(21)1OQ7XL-00325(93)MS26800-30, (01)04049098001878(11)180607(21)1OQ7XL-00296(93)MS26800-30, (01)04049098001878(11)180208(21)1OQ7XL-00030(93)MS26800-31, (01)04049098001878(11)170731(21)1OQ7WR-00216(93)MS26800-30, (01)04049098001878(11)181211(21)1OQ80D-00009(93)MS26800-30, (01)04049098001878(11)190103(21)1OQ7XL-00790(93)MS26800-33, (01)04049098001878(11)180904(21)1OQ7XL-00528(93)MS26800-30, (01)04049098001878(11)190117(21)1OQ80D-00060(93)MS26800-33, (01)04049098001878(11)181126(21)1OQ7XL-00757(93)MS26800-33, (01)04049098001878(11)181017(21)1OQ7XL-00638(93)MS26800-33, (01)04049098001878(11)180720(21)1OQ7XL-00366(93)MS26800-30, (01)04049098001878(11)181211(21)1OQ80D-00016(93)MS26800-30, (01)04049098001878(11)160622(21)12738M00594(93)MS26800-24, (01)04049098001878(11)160229(21)12738M00337(93)MS26800-20, (01)04049098001878(11)151110(21)12738M00125(93)MS26800-18, (01)04049098001878(11)151130(21)12738M00201(93)MS26800-18, (01)04049098001878(11)160122(21)12738M00293(93)MS26800-20. Additional US Serial Numbers (w/o UDI): 12946M00316, 12946M00314, 12946M00269, 11553M00212, 11516M00186, 12051M00467, 11516M00155, 11516M00359, 11553M00646, 11097M00482, 12381M00904, 12381M00952, 11097M00313, 11407M00119, 11097M00661, 12051M00327, 11731M00246, 11731M00230, 11731M00245, 11516M00314, 11802M00360, 11553M00190, 11407M00061, 12381M00513, 12381M00557, 11187M00009, 12051M00412, 12381M00528, 12051M00381, 12381M00551, 12381M00566, 12381M00478, 12381M00505, 12381M00500, 11553M00883, 12381M00473, 12381M00464, 12381M00531, 11097M00451.
FEI Number 3005783425
Recalling Firm/
Manufacturer
Draegar Medical Systems, Inc.
6 Tech Dr
Andover MA 01810-2434
For Additional Information ContactMike Kelhart
215-660-2349
Manufacturer Reason
for Recall
Software issue resulting in temporary loss of central monitoring functionality.
FDA Determined
Cause 2
Software design
ActionA notification signed 03/08/2022, issued via certified mail on 03/11/2022 and 03/24/2022, notified customers of the impacted software versions, that a software update will be deployed to correct the issue, and provided interim instructions regarding use of USB devices. Customers are to notify all users and return the acknowledgement form. For questions please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Quantity in Commerce1950 (1574 OUS + 376 US)
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IA, ID, IL, IN, LA, MN, MS, NC, NE, NH, NY, OH, PA, SD, TN, TX, WA, WI, and 58 additional foreign countries.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MHX
-
-