|
Class 2 Device Recall Alinity c Hemoglobin Reagent Kit |
|
Date Initiated by Firm |
June 09, 2022 |
Create Date |
July 22, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1455-2022 |
Recall Event ID |
90499 |
510(K)Number |
K130255
|
Product Classification |
Hemoglobin A1c test system - Product Code PDJ
|
Product |
REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL |
Code Information |
UDI/DI: 00380740135591/ Lots: 59561UQ04, 59779UQ05, 60229UQ07, 60540UQ09, 61011UQ10, 61349UQ01, 61658UQ01
|
Recalling Firm/ Manufacturer |
Abbott Laboratories 1915 Hurd Dr Irving TX 75038
|
For Additional Information Contact |
Suchin Song 224-668-0343
|
Manufacturer Reason for Recall |
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
On June 9, 2022, Abbott issued a "Urgent Medical Device Correction" to affected consignees via FedEx.
1. Immediately install the updated assay file versions listed below (as applicable for your laboratory), and manually configure the Low-Linearity value. The assay files can be obtained from www.corelaboratory.abbott
2. For the whole blood application, manually configure the THbWB Low-Linearity to 53.8278. For the hemolysate application, manually configure the THbH Low-Linearity to 1250.
For detailed information on editing assay parameters, refer to Changing assay configuration settings, Change a linearity range in the ARCHITECT System Operations Manual, Section 2.
3. Please review this letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for reviewing previously reported patient results.
4. Complete and return the Customer Reply Form.
5. Please retain this letter for your laboratory records.
6. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. |
Quantity in Commerce |
7,391 units |
Distribution |
US: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND NE NJ NV NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV
OUS: ALBANIA ALGERIA ANDORRA ANGOLA ARGENTINA ARMENIA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BANGLADESH BELARUS BELGIUM BERMUDA BOSNIA AND HERZEGOVI BRAZIL BRUNEI BULGARIA BURKINA FASO CANADA CHILE CHINA COLOMBIA COSTA RICA Country CROATIA CYPRUS CZECH REPUBLIC ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GABON GERMANY GREECE GUATEMALA HONG KONG HUNGARY INDIA INDONESIA IRAQ IRELAND ISRAEL ITALY JAMAICA JORDAN KAZAKHSTAN KENYA KUWAIT KYRGYZSTAN LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MALDIVES MALI MEXICO MOLDOVA MOROCCO MOZAMBIQUE NAMIBIA NETHERLANDS NIGERIA NORWAY OMAN PAKISTAN PALESTINIAN TERRITOR PANAMA PARAGUAY PERU PHILIPPINES POLAND PORTUGAL QATAR REPUBLIC OF KOSOVO REPUBLIC OF YEMEN ROMANIA RUSSIA RWANDA SAUDI ARABIA SENEGAL SERBIA SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SPAIN ST. LUCIA ST. VINCENT SWEDEN SWITZERLAND TANZANIA THAILAND THE GAMBIA TRINIDAD & TOBAGO TUNISIA TURKEY TURKS & CAICOS UAE UGANDA UNITED KINGDOM URUGUAY UZBEKISTAN VIETNAM ZIMBABWE
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = PDJ and Original Applicant = ABBOTT LABORATORIES
|
|
|
|