Date Initiated by Firm | May 25, 2022 |
Date Posted | September 26, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1793-2022 |
Recall Event ID |
90731 |
510(K)Number | K070461 K152630 K171504 K173626 K854902 |
Product Classification |
Indicator, biological sterilization process - Product Code FRC
|
Product | a. STERIS CORP Celerity Celerity 20 Sterilization Indicator Challenge Pack Steam 3 Inch Length Model Number: LCB053.
b. STERIS CORP VERIFY Verify SixCess Sterilization Flash Indicator Strip Steam 2-3/4 Inch Model Number: PCC006.
c. STERIS CORP VERIFY Verify Sterilization Biological Indicator Pack Steam Model Number: S3069.
d. STERIS CORP Celerity Celerity Sterilization Biological Indicator Vial Vaporized Hydrogen Peroxide Model Number:LCB044.
e. STERIS CORP Celerity Celerity 20 Sterilization Indicator Challenge Pack Steam 3 Inch Length Model Number: LCB049.
f. STERIS CORP VERIFY Sterilization Indicator Challenge Pack Steam Catalog #LCB035. |
Code Information |
GTIN:
a. 00724995161569; 10724995161566.
b. 00724995084653; 10724995084650.
c. 00724995023867; 10724995023864.
d. 00724995154233; 10724995154230.
e. 00724995161552; 10724995161559.
f. 00724995002558. |
Recalling Firm/ Manufacturer |
Mckesson Medical-Surgical Inc. Corporate Office 9954 Maryland Drive Deep Run Iii Ste. 4000 Richmond VA 23233
|
For Additional Information Contact | 804-553-2281 |
Manufacturer Reason for Recall | Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. |
FDA Determined Cause 2 | Environmental control |
Action | Mckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 sedgwick@sedgwick.com. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent possible patient harm. Please complete and return the enclosed response form as soon as possible even if you do not have any product on hand. Customers should contact Sedgwick directly by phone at 1-888-943-2401 to return the recalled products or if they have additional questions regarding this notification. Representatives are available Monday Friday, 8:00 am 5:00 pm PT |
Quantity in Commerce | a. 10 bx, b. 20 bx, c. 7 bx, d. 6 bx, e. 18 bx, f. 1 |
Distribution | US Nationwide distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = FRC 510(K)s with Product Code = FRC
|