• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medicina

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medicinasee related information
Date Initiated by FirmJuly 04, 2022
Date PostedOctober 20, 2022
Recall Status1 Open3, Classified
Recall NumberZ-0114-2023
Recall Event ID 90849
Product Classification Container, liquid medication, graduated - Product Code KYW
ProductMedicina Reusable Oral syringe 2.5ml Code: OTH25
Code Information UDI-DI: (01)05056115402876 Batch number: HM18042, HM21019, HM21054
Recalling Firm/
Manufacturer
Medicina Uk Ltd
Unit 2
Rivington View Business Park
Bolton United Kingdom
Manufacturer Reason
for Recall
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMedicina Ltd issued Field Safety Corrective Action notice to the US Distributor 07/04/2022 via email. The US Distributor is conducting recall to the user level letter issued the week of July 25th via United States Postal Service. . Letter states reason for recall, health risk and action to take: Action to be taken - Please carefully read this Field Safety Notice and return the Acknowledgement Form - Please do not use any of the batches of product listed on this notice - If you have product to return, please complete the information on the Acknowledgement Form and our team will arrange the return for you - Please distribute throughout your organisations and to relevant personnel - If you have any questions or queries in relation to this notice, please contact us using the information below Contact information regarding FSN Shaista Patel QA/RA Administrator Unit 2, Rivington View Business Park, Station Road, Blackrod, Bolton BL6 5BN Email: quality.administrator@medicina.co.uk Tel: 01204 695050
Quantity in Commerce20 boxes (2000 syringes)
DistributionNationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-