• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BD Pyxis" MedBank System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BD Pyxis" MedBank Systemsee related information
Date Initiated by FirmDecember 12, 2022
Date PostedJanuary 10, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0928-2023
Recall Event ID 91099
Product Classification Cabinet, table and tray, anesthesia - Product Code BRY
ProductBD Pyxis MedBank System - Product Label/labeling pending
Code Information BD Pyxis" MedBank Tower Main UDI Code: 10885403512568 Catalog Numbers: 169-23 to 169-81, 169-100, 169-132, 169-138, 169-139 BD Pyxis" MedBank Mini Main UDI Code: 10885403512544 Catalog Numbers: 169-101 - 169-113, 169-137 BD Pyxis" MEDFLEX BD Pyxis" MEDFLEX 2.0 UDI Codes:10885403512704 Catalog Numbers: 1119-00 1139-00 MedBank 200/500 UDI Code: N/A Catalog Numbers: 1137-00 BD Pyxis MEDPASS v1.0 or prior UDI Code: 10885403513817 Catalog Number: Not Provided BD Pyxis" MedBank Cabinet Software v2.2 or prior UDI Codes: 10885403513794 Catalog Number: N/A Affected Serial Numbers: 12940473 12971999 12972001 12972004 12972006 12972007 12982486 12982487 12982488 12982489 12982490 12982491 12982492 12982493 13517006 13523350 13598718 13629024 13727050 13739164 13754724 13766710 13766711 13767953 13769979 13770265 13770291 13789575 13792769 13801727 13802400 13817034 13819501 13826187 13829362 13829363 13829364 13832440 13843251 13849951 13853916 13853918 13856246 13873024 13875126 13875127 13877499 13877712 13877713 13877714 13877715 13879543 13879544 13880558 13883357 13884040 13884382 13884848 13884851 13885104 13885635 13889874 13889875 13889876 13889877 13896746 13905178 13908407 13909109 13909805 13909807 13909811 13909812 13920301 13920929 13921844 13922339 13922383 13923418 13923419 13923420 13926990 13929455 13929458 13931363 13931627 13933837 13934472 13939322 13939327 13939451 13941235 13941263 13943281 13944200 13944201 13945364 13945365 13945495 13945496 13945498 13946708 13947011 13947037 13947040 13947516 13947517 13947572 13947846 13948724 13949142 13950437 13954824 13956160 13961110 13961111 13961120 13963180 13963181 13963182 13964429 13965052 13966639 13973538 13977588 13985115 14005805 14006853 14007599 14008329 14011497 14011500 14017694 14032802 14036852 14037891 14045577 14054190 14054191 14062430 14075911 14083989 14085448 14087128 14095393 14095395 14101518 14101521 14101910 14101916 14112317 14112319 14112706 14112707 14125212 14128591 14133084 14133085 14133099 14133103 14133113 14136249 14138680 14138681 14138682 14149625 14152490 14153379 14154304 14156545 14156546 14158187 14159916 14160164 14160165 14160333 14160508 14167572 14167574 14167868 14168580 14169355 14169356 14172397 14172832 14173161 14174161 14175161 14175162 14175169 14175359 14175362 14177119 14177208 14177214 14177215 14177244 14178224 14178226 14182716 14183152 14183625 14185685 14185687 14190642 14194700 14194702 14194703 14196851 14197815 14197820 14202337 14202339 14208007 14208008 14208010 14208012 14208013 14208014 14211059 14211060 14211518 14212643 14212645 14214950 14216279 14216280 14220872 14222746 14223835 14225991 14228149 14228150 14228154 14229977 14233034 14233036 14242419 14242423 14244101 14248200 14248332 14248333 14251098 14254812 14254813 14254815 14254816 14258664 14265195 14265655 14269897 14269900 14270255 14275112 14275274 14278095 14279544 14279552 14279557 14297227 14298866 14298868 14304052 14304058 14304062 14304064 14304065 14304066 14306555 14308767 14309974 14314757 14317945 14317946 14324386 14325479 14332058 14333231 14334176 14336726 14336727 14337290 14338618 14338825 14338826 14347747 14347748 14348708 14348709 14349106 14351350 14351351 14352459 14364542 14364547 14365785 14365786 14374963 14374964 14374965 14377075 14377077 14377078 14377079 14377908 14378351 14378428 14392749 14395640 14395945 14403134 14403135 14403527 14403531 14404307 14406629 14406871 14407523 14407524 14408997 14408998 14410129 14410130 14410133 14410137 14414459 14414503 14414504 14416475 14419505 14419506 14419507 14419508 14419509 14419510 14420205 14420206 14420207 14420208 14420211 14420212 14420213 14420221 14420222 14420223 14427167 14429784 14440591 14441382 14442352 14442359 14442585 14442586 14442684 14442685 14442686 14442800 14443184 14443824 14443825 14443826 14443827 14443828 14443829 14443830 14443831 14443832 14443834 14443835 14443836 14443838 14443839 14443841 14443842 14443843 14443844 14443845 14443849 14445911 14448922 14452824 14453838 14454437 14454438 14456757 14456782 14456783 14458256 14458259 14458260 14460703 14460704 14460705 14460706 14460707 14461204 14461824 14461825 14461826 14461827 14461828 14461829 14461830 14461831 14461834 14461835 14461836 14461837 14461838 14461839 14461840 14461841 14461842 14461843 14461844 14461845 14461846 14461847 14461848 14462298 14469771 14470092 14472073 14472338 14472490 14472491 14472961 14472962 14472964 14472966 14472967 14472968 14472969 14472970 14472971 14472972 14472973 14472974 14472975 14472977 14472979 14472980 14472981 14472982 14472983 14472984 14472985 14472987 14472988 14472989 14472990 14473285 14473356 14473357 14473358 14473359 14473360 14473361 14473362 14473364 14473365 14473366 14473367 14473368 14473369 14473370 14473371 14473372 14473373 14473374 14473375 14473376 14473377 14473379 14473380 14474845 14475827 14475828 14475829 14475830 14475831 14475832 14475834 14475835 14475836 14475837 14475839 14478424 14478425 14478427 14478428 14478429 14478430 14478431 14478432 14478433 14478434 14478435 14485431 14486821 14491507 14491510 14491511 14491512 14491513 14491514 14491516 14491517 14491518 14491519 14491520 14491522 14491523 14491524 14491525 14491526 14491527 14491529 14491530 14491533 14491534 14491535 14491536 14491537 14492272 14509113 14509114 14509115 14509119 14509121 14509122 14509123 14509124 14509125 14509126 14509128 14509129 14509130 14509131 14509133 14509134 14509135 14509136 14509137 14509138 14509139 14509141 14513976 14518086 14531591 14532282 14532284 14532285 14532286 14532287 14532288 14532289 14532290 14532291 14532292 14532293 14532294 14532295 14532296 14532297 14532298 14532299 14532300 14532301 14532302 14532303 14532304 14532305 14532831 14533080 14533082 14533087 14533088 14533089 14533090 14533092 14533095 14533097 14533098 14535388 14535390 14535391 14535392 14537046 14540083 14540084 14540085 14540087 14540088 14540090 14540091 14540092 14540095 14540096 14540097 14540099 14540101 14540103 14540107 14540108 14540109 14540110 14540111 14540113 14540117 14540120 14540123 14540879 14545019 14549534 14549535 14549537 14549542 14549543 14549546 14549910 14549912 14549913 14549918 14549923 14549924 14549926 14549927 14549928 14549929 14549930 14549932 14549933 14549934 14549935 14549936 14568024 14568025 14568026 14568027 14568028 14568029 14568030 14568031 14568032 14568034 14568035 14568036 14568037 14568038 14568041 14568042 14568043 14568045 14568046 14568048 14568050 14568051 14568054 14568057 14568061 14568064 14579394 14586208 14602043 14602044 14602048 14602049 14602050 14602051 14603139 14603141 14603142 14603143 14603144 14603145 14603146 14603147 14603150 14603151 14603152 14603153 14603154 14603155 14603156 14603157 14603159 14603160 14603161 14603162 14603163 14603164 14603165 14603166 14603167 14603168 14603169 14603170 14603171 14603172 14603173 14603174 14603175 14603177 14603179 14603180 14603181 14603182 14604911 14607360 14607361 14607363 14607364 14607365 14607366 14607367 14607368 14607369 14607370 14618948 14624517 14630724 14634261 14635117 14635118 14635119 14635120 14635121 14635122 14648248 14649202 14649204 14649226 14649227 14649229 14649230 14649232 14649235 14649237 14649239 14649240 14649245 14649246 14649247 14649249 14649250 14649376 14651893 14656794 14657501 14657505 14657506 14657508 14657509 14671989 14671992 14671993 14671996 14671997 14671998 14672001 14672003 14672004 14672005 14672006 14672007 14672009 14672318 14672320 14678240 14678246 14678251 14678256 14678258 14678259 14678261 14678262 14678266 14686584 14689357 14695731 14695732 14695735 14695737 14697048 14697049 14697051 14697340 14697341 14697342 14697343 14699031 14699033 14699034 14699396 14699397 14699401 14699403 14699405 14699408 14699415 14699416 14709254 14709300 14711347 14711353 14714409 14715885 14719829 14719830 14720526 14721960 14721962 14721963 14721964 14721965 14721966 14721967 14721968 14723237 14723238 14723239 14723240 14723241 14723242 14723243 14723245 14723247 14723248 14723845 14723850 14723851 14723853 14723855 14723857 14723864 14723866 14723867 14728127 14728129 14728138 14728140 14728142 14728143 14730440 14730441 14730442 14730443 14730444 14730445 14730446 14730447 14730448 14730449 14730450 14730451 14730452 14730456 14730458 14730460 14730461 14730462 14730464 14730465 14730470 14730471 14737820 14737821 14737822 14737823 14737824 14737825 14737826 14737827 14737828 14737829 14737830 14737891 14739282 14740165 14740987 14740989 14740990 14740991 14742541 14742544 14742550 14742553 14742555 14742644 14742646 14743460 14743886 14743920 14743921 14743922 14743923 14743925 14743929 14743931 14743932 14743933 14743934 14743935 14743937 14743938 14743939 14750082 14750083 14750085 14750086 14750087 14750088 14750089 14750090 14757641 14757649 14772052 14781068 14781336 14781337 14781643 14787919 14787922 14787923 14787924 14787932 14787933 14787934 14787935 14787936 14787938 14787939 14787940 14787942 14787943 14787944 14788043 14788044 14788045 14788046 14788047 14788048 14788780 14788782 14797285 14797286 14803084 14813912 14813913 14834314 14844339 14861726 14861727 14861745 14861746 14861747 14861748 14861749 14861750 14861751 14861753 14861756 14867006 14867007 14868948 14874159 14874160 14874161 14874162 14874163 14874165 14874166 14874167 14874168 14883876 14888627 14888639 14890183 14891492 14894199 14894200 14894201 14894202 14897577 14897578 14914161 14914167 14925744 14925745 14925746 14925749 14925753 14937418 14942507 14942508 14942509 14942514 14947855 14947856 14947857 14947858 14947859 14949191 14949192 14949193 14949194 14951338 14951344 14951354 14951360 14951377 14951386 14958134 14961873 14961874 14961875 14961876 14961877 14961879 14961880 14961881 14961882 14961883 14961884 14963124 14963125 14963126 14972153 14977179 14977993 14990472 14990476 14997626 15000252 15000254 15000454 15000459 15003164 15010575 15019500 15019502 15019503 15019504 15019691 15023135 15023136 15024804 15024805 15027467 15027468 15027469 15041225 15041227 15042014 15049510 15049511 15055753 15057618 15057619 15063109 15063111 15063119 15066163 15070085 15070091 15071140 15071249 15078908 15078909 15079605 15079607 15079608 15079609 15079610 15080053 15080054 15080055 15080056 15080372 15080374 15082441 15082442 15082443 15082444 15082447 15082448 15084076 15086026 15086028 15086034 15086035 15088025 15088028 15088545 15089507 15089508 15090717 15090718 15091677 15093012 15095959 15095960 15095961 15095962 15095963 15095964 15095967 15096552 15096795 15098243 15098244 15101740 15101744 15102966 15102967 15109527 15109587 15115950 15115951 15115955 15115956 15115957 15115959 15115960 15124351 15124352 15125772 15125773 15125774 15125776 15125777 15125778 15125779 15125781 15125782 15125784 15133017 15133219 15133220 15167059 15167060 15167061 15167062 15167063 15167471 15169037 15175656 15176580 15176582 15176767 15182137 15182585 15188733 15188738 15190476 15192525 15201877 15201878 15203211 15203212 15204115 15204117 15204118 15204120 15206267 15206269 15206270 15210088 15210089 15210090 15210091 15210094 15210095 15210104 15210106 15213239 15216160 15216161 15219130 15219131 15219133 15219134 15220389 15220390 15220391 15220392 15220393 15220394 15222490 15222619 15222620 15223560 15223562 15223568 15223570 15223571 15226385 15226387 15227522 15227523 15227524 15227525 15229107 15229108 15229109 15229115 15229116 15229118 15229119 15229138 15229142 15229143 15229144 15229145 15229146 15230957 15236627 15240793 15242046 15242047 15242596 15242599 15242602 15242603 15242605 15242606 15245774 15245775 15245776 15245777 15245863 15248755 15248756 15248757 15248758 15253442 15253443 15253736 15253737 15253738 15253739 15253741 15253744 15254772 15254774 15254775 15254776 15254778 15257622 15257953 15257954 15260920 15260921 15260923 15262193 15263414 15265921 15265924 15267500 15267501 15267502 15267544 15267546 15267549 15268831 15268835 15268836 15268838 15268840 15268850 15268851 15268856 15273301 15274576 15274578 15274579 15276086 15276088 15276090 15277012 15277013 15279442 15279443 15279444 15279450 15279452 15279453 15279454 15279455 15279458 15282117 15282121 15282134 15282137 15282139 15282141 15282144 15286817 15286829 15286832 15286943 15290616 15290618 15290619 15290668 15290669 15290673 15290675 15294474 15294475 15295764 15295765 15295766 15295767 15295768 15295769 15295771 15295772 15295773 15295774 15295775 15295776 15295777 15295778 15295779 15295780 15295781 15295782 15295783 15295784 15295785 15295786 15295787 15295788 15295789 15295790 15295791 15295792 15295793 15295794 15295795 15295796 15295797 15295799 15295800 15295802 15295804 15295805 15295807 15295809 15295811 15295812 15295813 15295814 15295815 15295949 15296040 15296041 15296048 15296050 15298294 15298295 15298296 15299618 15299619 15299620 15299621 15299622 15299938 15299940 15299942 15300909 15300911 15302367 15303222 15303223 15303224 15305089 15305090 15305242 15305249 15306296 15306297 15308353 15308354 15308357 15309653 15309654 15312343 15312344 15312346 15312347 15314017 15315526 15315572 15315573 15317097 15317099 15317100 15317101 15317102 15317104 15317120 15318393 15318395 15318399 15318400 15318633 15318635 15320783 15322145 15322148 15322152 15322316 15322317 15322318 15322319 15322320 15322321 15322322 15322323 15322324 15322325 15322326 15326804 15326806 15326869 15326870 15326872 15328486 15328487 15328488 15328489 15328491 15328492 15328493 15328494 15328495 15328496 15328498 15328499 15328500 15328501 15328502 15328503 15328504 15328505 15328507 15328509 15328510 15328511 15328512 15328514 15331136 15337472 15340238 15340239 15340240 15340241 15340242 15340243 15340245 15340251 15342790 15342841 15353860 15353868 15353870 15353872 15353874 15353876 15353878 15353879 15353881 15353883 15353884 15353886 15353888 15353889 15353890 15353891 15353892 15353893 15353894 15353895 15353896 15353897 15353898 15353899 15353901 15353902 15353903 15353904 15353905 15353906 15353915 15353916 15353917 15353918 15353919 15353920 15353925 15353926 15353927 15353928 15353929 15353930 15355354 15356965 15356968 15356973 15356977 15357385 15357386 15357387 15357388 15357389 15361553 15361554 15361555 15361558 15361559 15363281 15364475 15366304 15366309 15366310 15366311 15366313 15366314 15366315 15366316 15366317 15366318 15366319 15366320 15366321 15366322 15366323 15366324 15366325 15366326 15366327 15366328 15366329 15366947 15366948 15366949 15366950 15369939 15371597 15371598 15371599 15371600 15371601 15371602 15371603 15371604 15371606 15374173 15376146 15376276 15378111 15378112 15378113 15378114 15378115 15378116 15378118 15378119 15378120 15378121 15383468 15383469 15383470 15383471 15383472 15383473 15383474 15383475 15383476 15383477 15383478 15383479 15383480 15383481 15383483 15383484 15383485 15383486 15383487 15383488 15383489 15389013 15389116 15389559 15389561 15389562 15389563 15389564 15389565 15389566 15389567 15389568 15389570 15389673 15391311 15391406 15391407 15391644 15392443 15392447 15392449 15392451 15393520 15393558 15396726 15398897 15398899 15398901 15398903 15398905 15398907 15398909 15398911 15398913 15398915 15409635 15409636 15409637 15409638 15409639 15409640 15409641 15410881 15410882 15410883 15410884 15410885 15410886 15410887 15410889 15410890 15410891 15410892 15410893 15410894 15410895 15410896 15410897 15410898 15410900 15410902 15415129 15415130 15415131 15415132 15415133 15415134 15415135 15415136 15415137 15415138 15416741 15422127 15422128 15422129 15422130 15422131 15422132 15422133 15422134 15422135 15422136 15422137 15422138 15422139 15422140 15422141 15422142 15422143 15422145 15422146 15422147 15422148 15422149 15422151 15422152 15422153 15422154 15422155 15422156 15422157 15422158 15429584 15430944 15430946 15430947 15430948 15430949 15430950 15430951 15430952 15430954 15430955 15430956 15430958 15430959 15430960 15430961 15430962 15430963 15430964 15430965 15430966 15437809 15437811 15437812 15437814 15437815 15437816 15437817 15437818 15437819 15437820 15437821 15437823 15437824 15437825 15437826 15437827 15437828 15437830 15437831 15437833 15437840 15441355 15441356 15441357 15441358 15441360 15441361 15441362 15441364 15441366 15441367 15442768 15442769 15442770 15442771 15442772 15442773 15442774 15442776 15442778 15442780 15442781 15442783 15442784 15442785 15442786 15442787 15442788 15442789 15442790 15442791 15442792 15442793 15442794 15442795 15442800 15452295 15452296 15452358 15452359 15452360 15460705 15460706 15460707 15460996 15460997 15461858 15463095 15463096 15463097 15463216 15463217 15463218 15463219 15463220 15464898 15467518 15470033 15470034 15470035 15471183 15473692 15473693 15474551 15474596 15475980 15483513 15489236 15489237 15489238 15489239 15489240 15489241 15489242 15489243 15489244 15489245 15490718 15490719 15490720 15490721 15490722 15490723 15490724 15490725 15490726 15490727 15490728 15490729 15490730 15490731 15490732 15490733 15490734 15490735 15490736 15490737 15490739 15490740 15490741 15490742 15490743 15490744 15490745 15490746 15490747 15490748 15490749 15490751 15490752 15490753 15490754 15490755 15490756 15490757 15490758 15490760 15490761 15490762 15490763 15490764 15490765 15490766 15490767 15490768 15490769 15490770 15490886 15490887 15490888 15492211 15492248 15492257 15492258 15492259 15492260 15492261 15492262 15492263 15492264 15492265 15492266 15492267 15492268 15492270 15492271 15492272 15492273 15492274 15492275 15492276 15492278 15492649 15498908 15499040 15499041 15499042 15499043 15499044 15499045 15499046 15499047 15499048 15499049 15499050 15499051 15499052 15499053 15499054 15499055 15499056 15499057 15499058 15499059 15499060 15499061 15499062 15499063 15499064 15500684 15500685 15500687 15500689 15500691 15500693 15500695 15500696 15500697 15500698 15500700 15500702 15500704 15500706 15500854 15503509 15503510 15503511 15503512 15503513 15503514 15503515 15503516 15503517 15503518 15503519 15504819 15504820 15504822 15504823 15504825 15504827 15504829 15504830 15504832 15504833 15504834 15504835 15504836 15504837 15504838 15504839 15504840 15504841 15506570 15506571 15506572 15506573 15506574 15506575 15506577 15506578 15506579 15506580 15506581 15506582 15506583 15506585 15506586 15506587 15506588 15506589 15506590 15506591 15506592 15506593 15506594 15506595 15506596 15509359 15509361 15509364 15509365 15509366 15513825 15513826 15513830 15513831 15513833 15513834 15513835 15517161 15518256 15518257 15518258 15518259 15518260 15518261 15518262 15518263 15518264 15518265 15518266 15518267 15518268 15518269 15518270 15518271 15518272 15518273 15518274 15518275 15518276 15518277 15519700 15519704 15521199 15521200 15521201 15521203 15537380 15546742 15547813 15547816 15547817 15547818 15547819 15547820 15547821 15547822 15547823 15547824 15547825 15547826 15547827 15547828 15547830 15550828 15550829 15550830 15550831 15550832 15550834 15550835 15550838 15550839 15550840 15550842 15550843 15550844 15550845 15550846 15550847 15550848 15550850 15550852 15550853 15550855 15550857 15550858 15550860 15550861 15554707 15557264 15557588 15557611 15561250 15561251 15561252 15561253 15561254 15561255 15561256 15561257 15561258 15561259 15561260 15561261 15561262 15561263 15561264 15561265 15561266 15561268 15561269 15561270 15561271 15561272 15561273 15561274 15561275 15561697 15562706 15563875 15564775 15564776 15564777 15564778 15564780 15564781 15564782 15564783 15564784 15564785 15564786 15564787 15564788 15564789 15564790 15564792 15564793 15564795 15564796 15564798 15564893 15564894 15564895 15564896 15564897 15569604 15569605 15569606 15569607 15569608 15569609 15569610 15569611 15569612 15569613 15569614 15569615 15569616 15569617 15569618 15569619 15569620 15569621 15569622 15569623 15569624 15569625 15569626 15569627 15569628 15569629 15569630 15574744 15575589 15575972 15575973 15575974 15575975 15575976 15575977 15575978 15575979 15575980 15575981 15575982 15575983 15575984 15575985 15575986 15575987 15575988 15575989 15575990 15575991 15575992 15575993 15575994 15575995 15575996 15579794 15579795 15581964 15581965 15583141 15583988 15583990 15584199 15587194 15587198 15587203 15587204 15587205 15587220 15588927 15590119 15590120 15598083 15598086 15598088 15598090 15598091 15598092 15598093 15598094 15598095 15598096 15598097 15598098 15598099 15598100 15598101 15598102 15598103 15598104 15598105 15598106 15598108 15598109 15598110 15598111 15598113 15600756 15600757 15600761 15600762 15600763 15600765 15600766 15600767 15600768 15600769 15600770 15600771 15600772 15600773 15600774 15600775 15600776 15600777 15600778 15600780 15600781 15600782 15600783 15600784 15600786 15600787 15600788 15606143 15606144 15606145 15606146 15606147 15606148 15606149 15606150 15606151 15606152 15606153 15606154 15606155 15606156 15606157 15606159 15606160 15606161 15606163 15606164 15606165 15606167 15606168 15606169 15606170 15606171 15613096 15614585 15614789 15614790 15614792 15614794 15614796 15615141 15616394 15617717 15618003 15618005 15618006 15622611 15622613 15622635 15622984 15622985 15622986 15622987 15622988 15622989 15631092 15631093 15631094 15631095 15631097 15631098 15631099 15631100 15631101 15631103 15631104 15631106 15631107 15631108 15631109 15631112 15631114 15631116 15631117 15631119 15631120 15631122 15631124 15631125 15631126 15631138 15632965 15632966 15632967 15632968 15632969 15632970 15632971 15632972 15632973 15632974 15632975 15632976 15632978 15632979 15632980 15632982 15632983 15632984 15632985 15632986 15632987 15632988 15632989 15632990 15632991 15633002 15636281 15636282 15636283 15638495 15638496 15639664 15639665 15639666 15639667 15639668 15639669 15639670 15639671 15639672 15639673 15639674 15639675 15639676 15639677 15639678 15639679 15639681 15639683 15639684 15639685 15639686 15639687 15639688 15639689 15639690 15639691 15639693 15639694 15639695 15639696 15639697 15639698 15639699 15639700 15639702 15639703 15639704 15639705 15639706 15639707 15639708 15639709 15639710 15639711 15639712 15639713 15639714 15639715 15639716 15639717 15639718 15639720 15639721 15639722 15639723 15639724 15639725 15639726 15639729 15643124 15643125 15643128 15645825 15645826 15645828 15645830 15647364 15647367 15650666 15652059 15652138 15652143 15652145 15652147 15653516 15653518 15653519 15653520 15653523 15653524 15654695 15654697 15654698 15654700 15654701 15654702 15654703 15654704 15654705 15654706 15654707 15654708 15654709 15654710 15654711 15654712 15654713 15654714 15654716 15654717 15654718 15654719 15654720 15654721 15654722 15657863 15657864 15657865 15657866 15657867 15660912 15660914 15661681 15661682 15662803 15662804 15662805 15662807 15662808 15662811 15662812 15662813 15662814 15662816 15662817 15662818 15662819 15662820 15662821 15662822 15662823 15662824 15662825 15662827 15662829 15662830 15662831 15662832 15662833 15662834 15662835 15662837 15662838 15662839 15662840 15662841 15666662 15666663 15666664 15666665 15666666 15666759 15666760 15666761 15666762 15666763 15666764 15666765 15666767 15666769 15666771 15671101 15671102 15674963 15674964 15674965 15674966 15674967 15674968 15674969 15674971 15674972 15674973 15675069 15679237 15679243 15679245 15679247 15679254 15679263 15679269 15679417 15682324 15684798 15684799 15684800 15684801 15684802 15684803 15684804 15684805 15684806 15684807 15684808 15684809 15684810 15684811 15684812 15684813 15684814 15684815 15684816 15684817 15691543 15693248 15693279 15700819 15709537 15709538 15709540 15709841 15709842 15709844 15709845 15709847 15709848 15709849 15709851 15709852 15709853 15709854 15709855 15709856 15709857 15709858 15709860 15709861 15709862 15709865 15709867 15709869 15709871 15709873 15709875 15713948 15713949 15713950 15713951 15713952 15715735 15715737 15715738 15715739 15715741 15715742 15715743 15715744 15715745 15715746 15715747 15715748 15715750 15715751 15715753 15715756 15715757 15715758 15715759 15715760 15715761 15715762 15715763 15715764 15715766 15723790 15723791 15723792 15723793 15723794 15723795 15723796 15723797 15723798 15723799 15723800 15723801 15723802 15723803 15723804 15723806 15723808 15723809 15723810 15723811 15723812 15723813 15723814 15723815 15723816 15723817 15723818 15723819 15723820 15723821 15728638 15728639 15728640 15728641 15728642 15728643 15728644 15728646 15728648 15728649 15728651 15728652 15728653 15728654 15728655 15728656 15728657 15728659 15728660 15728661 15728662 15728665 15728666 15728667 15728668 15728672 15728674 15728675 15728676 15728677 15728678 15728680 15728681 15728682 15728685 15736420 15736421 15736422 15736423 15736424 15736425 15736426 15736427 15736428 15736429 15736430 15736431 15736432 15736433 15736434 15736435 15736436 15736437 15736438 15736439 15736440 15736441 15736442 15736443 15736444 15736823 15736824 15736832 15740633 15740634 15743366 15743368 15743369 15743370 15743371 15743372 15743373 15743374 15743375 15743376 15743377 15743378 15743379 15743380 15743381 15743382 15743383 15743384 15743385 15743386 15743387 15743388 15743389 15743391 15743392 15743400 15747055 15747056 15747057 15747059 15747061 15747062 15747064 15747065 15747066 15747067 15747069 15747070 15747071 15747072 15747073 15747074 15747075 15747076 15747078 15747079 15747080 15747082 15747083 15747086 15747087 15747089 15747090 15747091 15747093 15747094 15748541 15748543 15748544 15761370 15761371 15761372 15761373 15761374 15761375 15761376 15761377 15761378 15761379 15761380 15761381 15761382 15761383 15761384 15761385 15761386 15761387 15761388 15761389 15761390 15761391 15761392 15761393 15761394 15761395 15761397 15761398 15761399 15761400 15761401 15761402 15761403 15761404 15761405 15770172 15770175 15770177 15770178 15770179 15770181 15770183 15770184 15770186 15770187 15770189 15770191 15770194 15770197 15770200 15770202 15770205 15770207 15770210 15770212 15770215 15770217 15770219 15770221 15770223 15770227 15770228 15770232 15774188 15774799 15775726 15775727 15775728 15775732 15775733 15775749 15775750 15775751 15775753 15775755 15775756 15775759 15775760 15775761 15775763 15775764 15775765 15775766 15775767 15775769 15775770 15775772 15775774 15775775 15775776 15775777 15775778 15775779 15775780 15775781 15779749 15779750 15779751 15779891 15783778 15783779 15783780 15783781 15783783 15783785 15783786 15783787 15783788 15783789 15783790 15783791 15783793 15783794 15783795 15783796 15783798 15783800 15783801 15783802 15783804 15783805 15783806 15783807 15783808 15784846 15784847 15784848 15784849 15784850 15784851 15786145 15786146 15786147 15786148 15786149 15786150 15786151 15786152 15786154 15786155 15786156 15786157 15786158 15786159 15786160 15786161 15786162 15786163 15786164 15786166 15786167 15786168 15786169 15786171 15786172 15786173 15786174 15788868 15788870 15788872 15803598 15803599 15803600 15803601 15803602 15803603 15803604 15803605 15803606 15803607 15803608 15803609 15803610 15803611 15803612 15803613 15803614 15803615 15803616 15803617 15803618 15803619 15803620 15803621 15803622 15803627 15803628 15803629 15803630 15803631 15803632 15803633 15803634 15803635 15803636 15803637 15803799 15806681 15806683 15806684 15806687 15806689 15806691 15806692 15806694 15806695 15806696 15806698 15806699 15806701 15806702 15806703 15806706 15806707 15806708 15806709 15806710 15806711 15806713 15806714 15806716 15806717 15806718 15806719 15806720 15806721 15806723 15806724 15806726 15806727 15815815 15815816 15815817 15815818 15815820 15815821 15815822 15815823 15815824 15815825 15815826 15815827 15815828 15815829 15815830 15815831 15815832 15815833 15815834 15815835 15815836 15815837 15815838 15815839 15815840 15824205 15824206 15824207 15824208 15824209 15824210 15824211 15824212 15824213 15824215 15824216 15824219 15824220 15824221 15824222 15824223 15824224 15824225 15824227 15824229 15824230 15824231 15824233 15824235 15824236 15830892 15832649 15832650 15832651 15832652 15832653 15832654 15832655 15832656 15832657 15832659 15832660 15832661 15832662 15832663 15832664 15832665 15832666 15832667 15832668 15832670 15832671 15832673 15832674 15832675 15832676 15832678 15832679 15832680 15832682 15832683 15832684 15832685 15832686 15832687 15832688 15834992 15835058 15835059 15836188 15852348 15852349 15852900 15852901 15853040 15853041 15853042 15853043 15853044 15853045 15853046 15853047 15853048 15853049 15853050 15853051 15853052 15853053 15853054 15853055 15853056 15853057 15853058 15853059 15853295 15853296 15853297 15853298 15853299 15853300 15853301 15853302 15853303 15853304 15853794 15854191 15854192 15854193 15854194 15854195 15854196 15854197 15854198 15854200 15854201 15854202 15854203 15854204 15854205 15854206 15854214 15854215 15854216 15854217 15854218 15854592 15854727 15854728 15854729 15854730 15854731 15854732 15854733 15854734 15854735 15854736 15854737 15854738 15854739 15854740 15854741 15854742 15854743 15854744 15854745 15854746 15854748 15854749 15854750 15854989 15855080 15855081 15855085 15855198 15855487 15855488 15855489 15855491 15855492 15855493 15855494 15855495 15855496 15855497 15855649 15855650 15855651 15855652 15855653 15867335 15867336 15867337 15867338 15867340 15867341 15867342 15867343 15867344 15867345 15867346 15867347 15867348 15867349 15872411 15872412 15872414 15872415 15872417 15872418 15872419 15872421 15872422 15872424 15872426 15872427 15872429 15872430 15872431 15872432 15879976 15879981 15879982 15879984 15879985 15879986 15879987 15879988 15879989 15879990 15879991 15879992 15879994 15879995 15879996 15879997 15879998 15880008 15880009 15888945 15888946 15888947 15888948 15888949 15888951 15888952 15888953 15888954 15888955 15888957 15888958 15888959 15888960 15888961 15888962 15888963 15888964 15888965 15888966 15888967 15888968 15888969 15895413 15895414 15895416 15895417 15895419 15895420 15895422 15895423 15895424 15895425 15895426 15895427 15895428 15895429 15895430 15895435 15895496 15901342 15901343 15901349 15901350 15901351 15902408 15902409 15902410 15902411 15902412 15902413 15902414 15902415 15902416 15902417 15902418 15902420 15902421 15902423 15902424 15906045 15906046 15906047 15906048 15906049 15906051 15906052 15906053 15906054 15906055 15906056 15906057 15906058 15906059 15906060 15906062 15906065 15906066 15906068 15906069 15906071 15906073 15906074 15906075 15906076 15906077 15906078 15906079 15906080 15906081 15906082 15906084 15906085 15906086 15906087 15906088 15906089 15906091 15906093 15906182 15910167 15910168 15910169 15910170 15910807 15910808 15910828 15910829 15910830 15910831 15910832 15910833 15910834 15910835 15910836 15910837 15910838 15910840 15910842 15910844 15910845 15910846 15910847 15910848 15910849 15910850 15912620 15912621 15912622 15912623 15914113 15914114 15914115 15914116 15914117 15914118 15914119 15914120 15914121 15914122 15914123 15914124 15914125 15914126 15914127 15914128 15914129 15914130 15914131 15914132 15914133 15914134 15914135 15914136 15919981 15919982 15919983 15919997 15919998 15919999 15920000 15920001 15920002 15920003 15920004 15920005 15920006 15920007 15920008 15920009 15920010 15920011 15920012 15920013 15920015 15920016 15920017 15920018 15920019 15920020 15920021 15920022 15921430 15921431 15921434 15921436 15921439 15921441 15921443 15921445 15921447 15921449 15921450 15921452 15921454 15921456 15921457 15921459 15921461 15921464 15921466 15921468 15921469 15921471 15921473 15921475 15921476 15921740 15921741 15921742 15921743 15922804 15922805 15922806 15922807 15922808 15922809 15922810 15922811 15922812 15922813 15922814 15922815 15922817 15922818 15922819 15922821 15922822 15922823 15923716 15923962 15923963 15923965 15923966 15923967 15923968 15923969 15923970 15923971 15923972 15923973 15923974 15923975 15924528 15924529 15924530 15924532 15924533 15924534 15924535 15924536 15924537 15924538 15924539 15924540 15924541 15924542 15924566 15924567 15925586 15925587 15925588 15925589 15925590 15925591 15925592 15925593 15925594 15925595 15925596 15925597 15925598 15925599 15925600 15925601 15925602 15925603 15925604 15927971 15927972 15927973 15927974 15927975 15927979 15927980 15927981 15927982 15927983 15927984 15927985 15927986 15927987 15927988 15927989 15927990 15927991 15927992 15927993 15927994 15928168 15928169 15928762 15928763 15928764 15928765 15928766 15928768 15929552 15929553 15929554 15929555 15929556 15929557 15929558 15929559 15929560 15929561 15929562 15929563 15929564 15929989 15930652 15930653 15930654 15930655 15930656 15930743 15930745 15930746 15930748 15930750 15930752 15930754 15930756 15930758 15930783 15930784 15930785 15930786 15930787 15930788 15930789 15930790 15930791 15930792 15930793 15930794 15930795 15930796 15930797 15931005 15931006 15931007 15931008 15931009 15931010 15931011 15931012 15931013 15931014 15931015 15931016 15931017 15931018 15931019 15931343 15932821 15932822 15932823 15932824 15932826 15932827 15932828 15932829 15932831 15932832 15932833 15932834 15932835 15932836 15932837 15932838 15932839 15932840 15932841 15933056 15933057 15933706 15933707 15933708 15933709 15933710 15933711 15933712 15933713 15933714 15933715 15933716 15933717 15933719 15933721 15933723 15933724 15933725 15933726 15933727 15933728 15933729 15933730 15933731 15933732 15933733 15933734 15933735 15933736 15933737 15933738 15933739 15933740 15933741 15933742 15933743 15933744 15933746 15933748 15933750 15935583 15935584 15935585 15935586 15935587 15935588 15935589 15935590 15935591 15935592 15935593 15935594 15935595 15935596 15935597 15935598 15935599 15935600 15935601 15935602 15935611 15935612 15935613 15935614 15935615 15935616 15935617 15935618 15935619 15935620 15935621 15935622 15935623 15935624 15935625 15935626 15935627 15935628 15935629 15935631 15937569 15937570 15937571 15937573 15937574 15937575 15937576 15937577 15937578 15937579 15937580 15937581 15937582 15937583 15937584 15937585 15937586 15937587 15937588 15937957 15938054 15938055 15939077 15939078 15939079 15939080 15939081 15939082 15939083 15939084 15939085 15939254 15939255 15940048 15940050 15940051 15940052 15940053 15940054 15940083 15940084 15940085 15940086 15940088 15940090 15940092 15940094 15940096 15940097 15940099 15940100 15940101 15940102 15940103 15940104 15940105 15940106 15940107 15940108 15940109 15940110 15940111 15940112 15940113 15940114 15940115 15940116 15940117 15940118 15940119 15940120 15940121 15940122 15940123 15940124 15940125 15940126 15940127 15940128 15940129 15940130 15940131 15940132 15940133 15940134 15940135 15940136 15940137 15940138 15940139 15940141 15940142 15940143 15940144 15940145 15940146 15940147 15940148 15940149 15940150 15940368 15940369 15940370 15940371 15940372 15940373 15940374 15940375 15940376 15940377 15940378 15940380 15940382 15940384 15940385 15940387 15940389 15940391 15940393 15940394 15940396 15940398 15940400 15940401 15940403 15940405 15940407 15940408 15940410 15940412 15940593 15940620 15940923 15940924 15940925 15940926 15940927 15940928 15940929 15940930 15940931 15941307 15941311 15941654 15941655 15941657 15941658 15941659 15941660 15941661 15941662 15941663 15941664 15941666 15941668 15941669 15941671 15941673 15941675 15941677 15941679 15941681 15941682 15941685 15941687 15941690 15941692 15941694 15941696 15941698 15941700 15941701 15941703 15942157 15942158 15942159 15943470 15943742 15943743 15943744 15943745 15943746 15943747 15943748 15943749 15943750 15943751 15943752 15943753 15943754 15943755 15943756 15943757 15943758 15943759 15943760 15943761 15943762 15943763 15943764 15943765 15943766 15943767 15943768 15943769 15943770 15943771 15943772 15943773 15943774 15943775 15943776 15943777 15943778 15943781 15943782 15944068 15944070 15945024 15945025 15945026 15945372 15945373 15946892 15946893 15946894 15946895 15946896 15946897 15948854 15948855 15949545 15949546 15949550 15949551 15949552 15949553 15949555 15949556 15949557 15949558 15949559 15949560 15949561 15949562 15949563 15949564 15949565 15949566 15949567 15949568 15949569 15949570 15949571 15949572 15949573 15949575 15949576 15949577 15949578 15949579 15949580 15949594 15949595 15950395 15950396 15950397 15950398 15950399 15952351 15952352 15952353 15952354 15952355 15952356 15952357 15952358 15952359 15952360 15952361 15952362 15952363 15952364 15952365 15952366 15952367 15952368 15952369 15952370 15952371 15952372 15952373 15952374 15952375 15952376 15952377 15952378 15952379 15952380 15954738 15954739 15954740 15956999 15957002 15957596 15957597 15957598 15957599 15957600 15957601 15957602 15957603 15957605 15957606 15957607 15957608 15957609 15957610 15957611 15957612 15957613 15957614 15957615 15957616 15957617 15957618 15957620 15957621 15957622 15957624 15957625 15957626 15957627 15957628 15958407 15958408 15958409 15958410 15958411 15958412 15958413 15958414 15958415 15958416 15958417 15958418 15958419 15958420 15958421 15958422 15958423 15958424 15958426 15958427 15958428 15958429 15958430 15958431 15958432 15958433 15958434 15958435 15958436 15958437 15958438 15958440 15958442 15958443 15958444 15958445 15958446 15958447 15958448 15958449 15958450 15958451 15958452 15958453 15958454 15958456 15958457 15958458 15958459 15958460 15958461 15958462 15958463 15959537 15959538 15959539 15959540 15960095 15960096 15960097 15960098 15960099 15960104 15960105 15960106 15960107 15960108 15960109 15960110 15960111 15961960 15961961 15961962 15961968 15961979 15961980 15961981 15961982 15961984 15961985 15961986 15961987 15961989 15961990 15961991 15961992 15961993 15961994 15961995 15961996 15961997 15961998 15961999 15962000 15962001 15962002 15962003 15962004 15962005 15962006 15962007 15962008 15962009 15962010 15962011 15962012 15962013 15962014 15962015 15962016 15962018 15962019 15962020 15962021 15965841 15965842 15965843 15965844 15965845 15965846 15965847 15965848 15965849 15965850 15965851 15965852 15965853 15965854 15965855 15965856 15965857 15965858 15965859 15965860 15965861 15965862 15965863 15965864 15965865 15965867 15967360 15967361 15967362 15967374 15967375 15967376 15967377 15967378 15967379 15967380 15967381 15967382 15967383 15967384 15967385 15967386 15967387 15967388 15967389 15967390 15967391 15967392 15967393 15967394 15967396 15967398 15967399 15967401 15967402 15967403 15967404 15967405 15967406 15967407 15967408 15967409 15967410 15967411 15967412 15967413 15967414 15967415 15978026 15978027 15978028 15978029 15978030 15978031 15978032 15978033 15978034 15978035 15978036 15978037 15978038 15978556 15978557 15978563 15978564 15978565 15978567 15978568 15979042 15979043 15979044 15979045 15979046 15979047 15979048 15979049 15979050 15979051 15979052 15979053 15979054 15979055 15979487 15979488 15979489 15979490 15979508 15979509 15979510 15979511 15979512 15979513 15979514 15979515 15979516 15979517 15979518 15979519 15979520 15979521 15979522 15979523 15979524 15980588 15980589 15980590 15980591 15980592 15980593 15980594 15980595 15980596 15980597 15980598 15980599 15980600 15980601 15980602 15980603 15980604 15980605 15980606 15980607 15980608 15980609 15980610 15983615 15983616 15983617 15983618 15983619 15983620 15984458 15984459 15984460 15984461 15984462 15984463 15984464 15984465 15987751 15987755 15987756 15987757 15988573 15988574 15988575 15988576 15988577 15988578 15988579 15988580 15988581 15988582 15988583 15988584 15988585 15988586 15988587 15988588 15988589 15988590 15988591 15988592 15988639 15988883 15988898 15988899 15988900 15988901 15988902 15989186 15989187 15989188 15989189 15989190 15989191 15989192 15989193 15989194 15989195 15989196 15989197 15989198 15989199 15989200 15989201 15989202 15991648 15991649 15991650 15991651 15991724 15991725 15991726 15991727 15991728 15991729 15991730 15991731 15991732 15991733 15991734 15991735 15991736 15991737 15991738 15991739 15991741 15991742 15991744 15991749 15991783 15991784 15991785 15991786 15991787 15991788 15991789 15991790 15991791 15991792 15991793 15991794 15991795 15991796 15991797 15991798 15992162 15992164 15992165 15992167 15992168 15992169 15992170 15992171 15992172 15992173 15992174 15992175 15992176 15992177 15992178 15992179 15992180 15992181 15992182 15992183 15992184 15992185 15992186 15992187 15992188 15992740 15992741 15992743 15992744 15992745 15992746 15992747 15992748 15992749 15992750 15992751 15992752 15992753 15992754 15992755 15992756 15992757 15993197 15993198 15993199 15993200 15993201 15993202 15993203 15993204 15993205 15993207 15993209 15993957 15993958 15993959 15993960 15993961 15993962 15993963 15993964 15993965 15993966 15993967 15993968 15993969 15993970 15993971 15993972 15994032 15994033 15994364 15994365 15994366 15994367 15994368 15994369 15994370 15994371 15996316 15996317 15996318 15996319 15996320 15996321 15996322 15996323 15996655 15996656 15996657 15996658 15996659 15996660 15996661 15996662 15996663 15996664 15996665 15996667 15996668 15996672 15996676 15996677 15996678 15996679 15997470 15997471 15997472 15997473 15997474 15997476 15997477 15997479 15997480 15997481 15997482 15997483 15997484 15997485 15997486 15997487 15997488 15997489 15997808 15997809 15997810 15997811 15997812 15997813 15997815 15997816 15997817 15997818 15997819 15997820 15997821 15997822 15997823 15997824 15997825 15997826 15997827 15997828 15997858 15997859 15997860 15997861 15997862 15997863 15997864 15997865 15997866 15997867 15997868 15997869 15997870 15997871 15997872 15997873 15997874 15997875 15997876 15997877 15997878 15998595 15998596 15999405 15999407 15999408 15999410 15999411 15999412 15999413 15999414 15999415 15999416 15999418 15999419 15999420 15999421 15999422 15999423 15999424 15999425 15999426 15999427 15999428 15999429 15999440 15999441 15999442 15999443 15999444 15999791 15999792 15999793 15999794 15999796 15999797 15999799 15999800 15999942 15999943 15999944 15999945 15999946 15999947 15999948 16000233 16000234 16000235 16000236 16000237 16000481 16000745 16000746 16000747 16000748 16000749 16000750 16000751 16000752 16000753 16000754 16000755 16000756 16000758 16000759 16000760 16000761 16000762 16001038 16001039 16001040 16001041 16001042 16001043 16001044 16001045 16001046 16001047 16001048 16001049 16001050 16001051 16001054 16001055 16001067 16001069 16001070 16001071 16001173 16001227 16002030 16002034 16002036 16002037 16002107 16002108 16002109 16002110 16002111 16002112 16002120 16002121 16002122 16002123 16002124 16002125 16002712 16002713 16003482 16003484 16003486 16003487 16003489 16003491 16003493 16003495 16003497 16003499 16003501 16003503 16003505 16003507 16003508 16003510 16003512 16003514 16003516 16003518 16003931 16004153 16004425 16005132 16005133 16005134 16005135 16005136 16005137 16005138 16005139 16005141 16005142 16005144 16005391 16005392 16006027 16006028 16006029 16006030 16006031 16006032 16006033 16006034 16006035 16006036 16006037 16006038 16006039 16006040 16006041 16006042 16006043 16006044 16006045 16006046 16008004 16008006 16009044 16009046 16009048 16009049 16009050 16009051 16009052 16009053 16009054 16009055 16009056 16009057 16009059 16009060 16009061 16009062 16009063 16009064 16009065 16009066 16009067 16009068 16009069 16009070 16009071 16009072 16012164 16012165 16012166 16012167 16012168 16012169 16012170 16012171 16012172 16012173 16012174 16012175 16012176 16012177 16012178 16012179 16012180 16012181 16012182 16012183 16012184 16012338 16012339 16012340 16012341 16012342 16012343 16012344 16012345 16012346 16012347 16012348 16012349 16012350 16012351 16012352 16012353 16012354 16012355 16012356 16012357 16012358 16012359 16012360 16012361 16012362 16012363 16012364 16012685 16012860 16012863 16012864 16012865 16012866 16012867 16012869 16012871 16012872 16012873 16014447 16014448 16014451 16014896 16015766 16015767 16015768 16015769 16015770 16015771 16015772 16015773 16015774 16015775 16015776 16015777 16015778 16015779 16015780 16015781 16015782 16015783 16015784 16015785 16015786 16015787 16015788 16015789 16015790 16015791 16015792 16015793 16015794 16015795 16015796 16015797 16015798 16015799 16015800 16015801 16015802 16015803 16015804 16015805 16015806 16016146 16017209 16017210 16017211 16017212 16017213 16017214 16017216 16017217 16019470 16019472 16019479 16020097 16020099 16022057 16022732 16023030 16023031 16023032 16023033 16023034 16023035 16023040 16023041 16023042 16023043 16023044 16023045 16023046 16023047 16023048 16023049 16025386 16026280 16026281 16026282 16026283 16026284 16026285 16026286 16026287 16026288 16026289 16026290 16026291 16026292 16026293 16026294 16026295 16026296 16026297 16026298 16026299 16026300 16026885 16026886 16026888 16026889 16026890 16026891 16026892 16026893 16026894 16026895 16026897 16026898 16026899 16026900 16026901 16026902 16026903 16026904 16026905 16026906 16026907 16026908 16026909 16026910 16026911 16026912 16026913 16026914 16026915 16026916 16026917 16026918 16029669 16030464 16031886 16031887 16031888 16031889 16031890 16031891 16031892 16032397 16032398 16032399 16033122 16033123 16033124 16033125 16034341 16034342 16034343 16034344 16034345 16034791 16034792 16034793 16035155 16035156 16035157 16035948 16035949 16036183 16036389 16036390 16036872 16036875 16036880 16036884 16036885 16037349 16037350 16037351 16037352 16037627 16037628 16037630 16037631 16037632 16037633 16037634 16037635 16037980 16038218 16038219 16038618 16038620 16038621 16038622 16038623 16038624 16038626 16039315 16039316 16039317 16039553 16039555 16039556 16039561 16039565 16039729 16039764 16039766 16039767 16039768 16039769 16039770 16039771 16039772 16039774 16039776 16040063 16040064 16040065 16040066 16040067 16040068 16040154 16040193 16040194 16040231 16040233 16040234 16040728 16040729 16040730 16040731 16040732 16040911 16040913 16040914 16040915 16040917 16040918 16040919 16042046 16042047 16042123 16042131 16042993 16042994 16043221 16043223 16043226 16043229 16044053 16044063 16044065 16044067 16044069 16044071 16044073 16044075 16044081 16044082 16044084 16044086 16044087 16044088 16044090 16044091 16044093 16044094 16044095 16044096 16044097 16044098 16044100 16045065 16045107 16045109 16045112 16045113 16046162 16046163 16046164 16046165 16046167 16046168 16046169 16046170 16046297 16046298 16046299 16046300 16046301 16046303 16046304 16046305 16047561 16047562 16047564 16047565 16047566 16047926 16047927 16047931 16047932 16047933 16047934 16047936 16048634 16048636 16048637 16048764 16048765 16048767 16048768 16048769 16048771 16049076 16049077 16049572 16049573 16049574 16049575 16049576 16049577 16049580 16049592 16049596 16049602 16049606 16049607 16049609 16049610 16051731 16051732 16051733 16051734 16051736 16051737 16051738 16051739 16051740 16051741 16052084 16052473 16052474 16052475 16052476 16052477 16052478 16052480 16052482 16052484 16052519 16052520 16052521 16053361 16054660 16054661 16054663 16054916 16054917 16054918 16054919 16054920 16054921 16054922 16054923 16054983 16054985 16054986 16054987 16055121 16055122 16055123 16055124 16055213 16055229 16055230 16055231 16055232 16055233 16055234 16055243 16055302 16055907 16055908 16055909 16055910 16055911 16055912 16055913 16055967 16056243 16056394 16056395 16056396 16056397 16056398 16056399 16056400 16056401 16056463 16056464 16056465 16056466 16056467 16056469 16056470 16056471 16056658 16056659 16056660 16056661 16056662 16056663 16056664 16056665 16056920 16056921 16056922 16056923 16056924 16056926 16056927 16056928 16056929 16056930 16056931 16056933 16056934 16056935 16058105 16058109 16058110 16058111 16058114 16058116 16058205 16058207 16058634 16058635 16058636 16058637 16058638 16058639 16060437 16060438 16060439 16060440 16060441 16060442 16060443 16060444 16060445 16060446 16060447 16060448 16060449 16060450 16060451 16060452 16060453 16060454 16060455 16060456 16060457 16060458 16060459 16060460 16060461 16060462 16060463 16060464 16060465 16060466 16060467 16060468 16060469 16060470 16060471 16060472 16060473 16060474 16060475 16060476 16060477 16060478 16060479 16060480 16060481 16060482 16060483 16060484 16060485 16060486 16060487 16060488 16060489 16060490 16060491 16060492 16060493 16060494 16060495 16060496 16060497 16060498 16060499 16060508 16060509 16060510 16060511 16060788 16060789 16060790 16060791 16060792 16060793 16060794 16060795 16060796 16060797 16060798 16060799 16060910 16060913 16060914 16060915 16060918 16060919 16060920 16060921 16060922 16060923 16060924 16061390 16061391 16061392 16061393 16062430 16062431 16062432 16062433 16062514 16062515 16062516 16062517 16062518 16062519 16062520 16062521 16062522 16062523 16062524 16062525 16062526 16062527 16062528 16062529 16062530 16062531 16062532 16062533 16062942 16062943 16062944 16062945 16062946 16062947 16062948 16062949 16062950 16062951 16062952 16062953 16062954 16062955 16062956 16062957 16063139 16063140 16063141 16063142 16063143 16063144 16063244 16063250 16063702 16063703 16063704 16063705 16063706 16063707 16063826 16063911 16063924 16064699 16064700 16064701 16064702 16064703 16064705 16065031 16065032 16065033 16065034 16065035 16065246 16065250 16066049 16066050 16066051 16066055 16066058 16066060 16066061 16066062 16066740 16066745 16067151 16068039 16068264 16068549 16068550 16068551 16068552 16068553 16068554 16068555 16068556 16068557 16069111 16069112 16069113 16069114 16069115 16069116 16069117 16069118 16069119 16069120 16069121 16069122 16069123 16069124 16069125 16069126 16069142 16069143 16070529 16072260 16073418 16073419 16073420 16073423 16073425 16073427 16073430 16073432 16073435 16073436 16073437 16073440 16073443 16073448 16073449 16073454 16073456 16073458 16073462 16073464 16073466 16073467 16073468 16073477 16073481 16073486 16073488 16073490 16073491 16073494 16073514 16073515 16073516 16073518 16073520 16073522 16073523 16073525 16073526 16073528 16073529 16073530 16073532 16073534 16074261 16075071 16075670 16077584 16077585 16079832 16079833 16079834 16079835 16079836 16079837 16079838 16079839 16079840 16079841 16079842 16079843 16079844 16079845 16079846 16079847 16079848 16079849 16079850 16079851 16079852 16079853 16079854 16079855 16079856 16079858 16079860 16079862 16079863 16079865 16079867 16079868 16079869 16079870 16079871 16079872 16079873 16079874 16079875 16079876 16079877 16079878 16079879 16079880 16079881 16079882 16079883 16079884 16079885 16079886 16080412 16081052 16081312 16081478 16081522 16081524 16081525 16081624 16081878 16082187 16082189 16082213 16082214 16082216 16082217 16082239 16082788 16082797 16082798 16082799 16082800 16082802 16082950 16083483 16083489 16083490 16083491 16083492 16083493 16083494 16083495 16083496 16083497 16083498 16083499 16083500 16083501 16083695 16083747 16083748 16083749 16083751 16083752 16083753 16083754 16083755 16083756 16083757 16084020 16084022 16084023 16084024 16084025 16084026 16084028 16084089 16084090 16084091 16084092 16084093 16084094 16084095 16084096 16084097 16084098 16084099 16084100 16084101 16084102 16084103 16084104 16084105 16084106 16084107 16084108 16084109 16084110 16084111 16084112 16084113 16084114 16084115 16084116 16084117 16084118 16084119 16084120 16084121 16084122 16084123 16084124 16084125 16084126 16084127 16084128 16084129 16084130 16084131 16084132 16084133 16084134 16084135 16084136 16084137 16084138 16084139 16084140 16084141 16084142 16084143 16084144 16084145 16084146 16084147 16084148 16084149 16084150 16084151 16084152 16084153 16084154 16084155 16084156 16084157 16084158 16084159 16084160 16084161 16084162 16084163 16084164 16084165 16084166 16084167 16084168 16084169 16084170 16084171 16084463 16084464 16084465 16084466 16084467 16084469 16084470 16084471 16084492 16084502 16084954 16084955 16084959 16084960 16084961 16084962 16084963 16085412 16085440 16085441 16085442 16085443 16085444 16085446 16085447 16085632 16085674 16085675 16085676 16085677 16085678 16086055 16086058 16086059 16086061 16086062 16086063 16086065 16086066 16086067 16086068 16086069 16086070 16086071 16086072 16086073 16086074 16086075 16086076 16086077 16086078 16086079 16086485 16086486 16086487 16086488 16086489 16086998 16087001 16087002 16087003 16087004 16087005 16087006 16087007 16087008 16087009 16087687 16087688 16087689 16087690 16087691 16087692 16087693 16087694 16087695 16087696 16087808 16087809 16087811 16087812 16087813 16088361 16088362 16088363 16088364 16088365 16088366 16088367 16088368 16088369 16088370 16088371 16088905 16088909 16088910 16088911 16088912 16088913 16088914 16088915 16088917 16088919 16088920 16088921 16088999 16089000 16089001 16089019 16089021 16089022 16089561 16089562 16089912 16089956 16090202 16090203 16090204 16090308 16090325 16090393 16090399 16090401 16090406 16090687 16091227 16091228 16091258 16091818 16091819 16091820 16091885 16092773 16092775 16092800 16092801 16092802 16092803 16092804 16092805 16093533 16093639 16093640 16093716 16093717 16093719 16095332 16095333 16095334 16095335 16095336 16095337 16095338 16095339 16095340 16095341 16097230 16097406 16097407 16097437 16097441 16097442 16097985 16101319 16102550 16102551 16103625 16103626 16103627 16103628 16103629 16103630 16103631 16103632 16103633 16103634 16103635 16103636 16103637 16103638 16103639 16103640 16103641 16103642 16103643 16103644 16103645 16103646 16103647 16103648 16103649 16103650 16103651 16103662 16103663 16103664 16103665 16103666 16103667 16103668 16103669 16103670 16103671 16103672 16103673 16103674 16103675 16103676 16103677 16103678 16103679 16103680 16103681 16103682 16103683 16103684 16103685 16103686 16103687 16103688 16103689 16103690 16103691 16103692 16103693 16103694 16103695 16103696 16103697 16103698 16103699 16103700 16103701 16103702 16103703 16103704 16103705 16103706 16103707 16103708 16103709 16103710 16103711 16103712 16103713 16103714 16103715 16103716 16103717 16103718 16103719 16103720 16103721 16103722 16103723 16103724 16103725 16103726 16103727 16105080 16105081 16105083 16105084 16105085 16105086 16105087 16105088 16105090 16105091 16105092 16105398 16105399 16105400 16105401 16105402 16105403 16105404 16105405 16105406 16105407 16105415 16105989 16105990 16105991 16105992 16105993 16105995 16105996 16105997 16106201 16106202 16106203 16106204 16106205 16106206 16106587 16106589 16106591 16106592 16106594 16106595 16106631 16106636 16107014 16107015 16107016 16107017 16107018 16107019 16107020 16107021 16107022 16107083 16107084 16107085 16107091 16107092 16107286 16107287 16107288 16107289 16107290 16107291 16107567 16107568 16107569 16107660 16107663 16108132 16108133 16108134 16108526 16108527 16108528 16108529 16108530 16108531 16108532 16108533 16108534 16108535 16108536 16108537 16108539 16108541 16108543 16108544 16108545 16108824 16108827 16108828 16108830 16108831 16108832 16108833 16108913 16108914 16108917 16109066 16109067 16109068 16109069 16109070 16109071 16109072 16109073 16109075 16109076 16109077 16109224 16109225 16109227 16109228 16109317 16109318 16109319 16109320 16109322 16109323 16109695 16109696 16109697 16109698 16109699 16109700 16109701 16109702 16109703 16109704 16109705 16109706 16109707 16109708 16109709 16109710 16109711 16109712 16109713 16109714 16109715 16109716 16109717 16109718 16109720 16109721 16109722 16109723 16109724 16109725 16109726 16109727 16109728 16109729 16109730 16109731 16109732 16109733 16109734 16109735 16109736 16109737 16109738 16109739 16109740 16109741 16109742 16109743 16109745 16109746 16109747 16109748 16109749 16109750 16109751 16109752 16109753 16109754 16109755 16109756 16109757 16109758 16109759 16109760 16109761 16109762 16109763 16109764 16109765 16109766 16109767 16109768 16109769 16109771 16109772 16109773 16109774 16109775 16109776 16109777 16109778 16109779 16109780 16109781 16109782 16109783 16109784 16109785 16109786 16109787 16109788 16109789 16109790 16109791 16109792 16109793 16109795 16109796 16109885 16109886 16109888 16109889 16109890 16109891 16109892 16109984 16109988 16109989 16109991 16109992 16109993 16109994 16109995 16109996 16109997 16109998 16109999 16110001 16110002 16110003 16110135 16110138 16110143 16110144 16110145 16110147 16110257 16110258 16110511 16110532 16110888 16110889 16111183 16111184 16111185 16111186 16111187 16111188 16111189 16111232 16111233 16111234 16111668 16111669 16111670 16111671 16111672 16111846 16111847 16111848 16111849 16111850 16111852 16111853 16112040 16112042 16112043 16112044 16112045 16112046 16112047 16112048 16112171 16112173 16112174 16112175 16112176 16112177 16112178 16112179 16112184 16112185 16112186 16112187 16112376 16112377 16112378 16112379 16112380 16112381 16112382 16112383 16112384 16112385 16112386 16112388 16112389 16112390 16112391 16112392 16112393 16112396 16112397 16112398 16112756 16112757 16112758 16112759 16112760 16112761 16112762 16112763 16112764 16112765 16112766 16112767 16112768 16112769 16112770 16112771 16112772 16112773 16112774 16112775 16112776 16112777 16112778 16112779 16112780 16112781 16112782 16112783 16112808 16112812 16112853 16113239 16113240 16113241 16113307 16113461 16113462 16113480 16113481 16113482 16113483 16113484 16113485 16113486 16113487 16113488 16113489 16113490 16113491 16113492 16113493 16113494 16113495 16113496 16113498 16113500 16113501 16113502 16113503 16113504 16113505 16113506 16114228 16115895 16115957 16115958 16115959 16116015 16116027 16116718 16116719 16116720 16116721 16116722 16116723 16116724 16116725 16116726 16116727 16116729 16116730 16116731 16116732 16116733 16116734 16116735 16116736 16116737 16116738 16116739 16116740 16116742 16116743 16118352 16118353 16118364 16118366 16118367 16118368 16118369 16118370 16118371 16118372 16118373 16119404 16119405 16119406 16119407 16119408 16119409 16119877 16119929 16119930 16119931 16119932 16119933 16119934 16120070 16120443 16120704 16120707 16120708 16120931 16120934 16120938 16120948 16120950 16120952 16121796 16121797 16122575 16122576 16122742 16122744 16122746 16122748 16122750 16122752 16122753 16123677 16124076 16124077 16124078 16124079 16124080 16124081 16124082 16124083 16124084 16124085 16124086 16124087 16124093 16124122 16124123 16124124 16124125 16124126 16124127 16124128 16124129 16124130 16124131 16124132 16124133 16124134 16124135 16124136 16124137 16124138 16124139 16124140 16124141 16124142 16124143 16124144 16124145 16124146 16124603 16124604 16124605 16124904 16124905 16124906 16124907 16124908 16124909 16124910 16125092 16125674 16125676 16125678 16125679 16125697 16125698 16125699 16125701 16126158 16126159 16126160 16126161 16126673 16127651 16127652 16127653 16127654 16127655 16128394 16128843 16128844 16128845 16128847 16128848 16128849 16128850 16128851 16128853 16128854 16128855 16128856 16128859 16128861 16129009 16129012 16129013 16129014 16129015 16129016 16129017 16129507 16129770 16130129 16130295 16130296 16130308 16130309 16130310 16130311 16130312 16130314 16131741 16131742 16132248 16132438 16132658 16132831 16134107 16134108 16134109 16134110 16134111 16134691 16134692 16134693 16135184 16135186 16135187 16135188 16135189 16135190 16135191 16135192 16136374 16136375 16136376 16136377 16136378 16136379 16136380 16136418 16136419 16136421 16136422 16136726 16136727 16136728 16136729 16136730 16136731 16136732 16136733 16136734 16137617 16137618 16137673 16137674 16137675 16138035 16138036 16138037 16138038 16138039 16138040 16138041 16138042 16138043 16138044 16138825 16138827 16138828 16138829 16138830 16138831 16138835 16138836 16138837 16139153 16139154 16139672 16139673 16139674 16139675 16139676 16139677 16139678 16139679 16139684 16139686 16139797 16139836 16140468 16140524 16141542 16141543 16141545 16141547 16141549 16141550 16141551 16141585 16141588 16142010 16142772 16143254 16143255 16144318 16144319 16144320 16144321 16144322 16144323 16144324 16144325 16144326 16144327 16144328 16144329 16144331 16144332 16144333 16144335 16144336 16144339 16144807 16144810 16144811 16145473 16145921 16145922 16145923 16145924 16146405 16146406 16146407 16146408 16147169 16147248 16147319 16147499 16147608 16147609 16147788 16148945 16148946 16148959 16148978 16148979 16148982 16149364 16149476 16149477 16151040 16151041 16151042 16151043 16151378 16151379 16151380 16151381 16151382 16151468 16151583 16151584 16151585 16151587 16151699 16151700 16151701 16152697 16153564 16154249 16154250 16154252 16154253 16154254 16154518 16154519 16155517 16155518 16155519 16155520 16155521 16155522 16155812 16155813 16155814 16155815 16155816 16155817 16155818 16155819 16155820 16155821 16155822 16155823 16155824 16155825 16155826 16155827 16155828 16155829 16155830 16155831 16155832 16155833 16155834 16155835 16155836 16155837 16155838 16155839 16155840 16155841 16155842 16155843 16155844 16155846 16155847 16155848 16155849 16155850 16155851 16155852 16155853 16155854 16155855 16155856 16155857 16155858 16155859 16155860 16155861 16155862 16155863 16155864 16155865 16155866 16155867 16155868 16155869 16155870 16156626 16157016 16157018 16157021 16158463 16158464 16158465 16158512 16158513 16158514 16158515 16158516 16158517 16158518 16158519 16158752 16158753 16159137 16159138 16159556 16159757 16159758 16159759 16159780 16159781 16159783 16159784 16159786 16159789 16159987 16159988 16159989 16160108 16160246 16160247 16160248 16160274 16160275 16160283 16160284 16160285 16160286 16160287 16160288 16160289 16160290 16161290 16161325 16161326 16161327 16161328 16161329 16161330 16161331 16161332 16161333 16161335
FEI Number 2016493
Recalling Firm/
Manufacturer
CareFusion 303, Inc.
10020 Pacific Mesa Blvd
San Diego CA 92121-4386
For Additional Information ContactMr. Puneet Sarin
888-876-7128
Manufacturer Reason
for Recall
Due to customer complaints related to the MedBank software which may indicate that a different medication is loaded in a cubie (drawer) pocket than what is physically loaded and listed on the pocket label. As a result of this issue, a patient could receive the wrong medication or wrong dose of medication that could lead to patient harm.
FDA Determined
Cause 2
Software design
ActionOn 12/12/2022, the firm sent an "Urgent: Medical Device Correction" Letter via FedEx 2-Day mail to customers informing them of a software anomaly affecting BD Pyxis MedBank Cabinets with software version v.2.2 or prior. The firm was made aware that when using an atypical workflow the software would indicate that a different medication is loaded in a CUBIE pocket that what is physically loaded and listed on the CUBIE pocket label. This issue may occur when a new CUBIE is placed into an empty position that the software considers already occupied. Customer are instructed to: To mitigate and avoid this issue, only restock or destock CUBIE pockets when in the Restock / Destock workflow. If a CUBIE is removed atypically, and you do not have the privileges to activate the Restock/Destock workflow, call BD Support to assist with inserting the CUBIE correctly into the MedBank automated dispensing cabinet. Do not manually insert the CUBIE without support. Actions to be taken by Pharmacy: 1. Provide a copy of this customer letter to all facilities within their distribution network that have the affected products. 2. Promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that they understand the information provided in this letter. Actions to be taken by the Customer: 1. Post this Customer Notification on all affected devices until additional software updates have been released. 2. Promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 3. If customers experience a memory overwrite issue, reach out to BD Technical Support/Product Complaints at 1-866-930-9251. Actions to be taken by BD: 1. A software update to version 2.3.1.8 is currently in progress and will be available to all affected customers by December 2022. BD will contact customers to schedule a convenient time to update the software. BD will re
Quantity in Commerce6,796 systems
DistributionU.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-