• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CardioMEMSTM HF System Patient Electronics System (PES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CardioMEMSTM HF System Patient Electronics System (PES)see related information
Date Initiated by FirmFebruary 07, 2023
Date PostedMarch 03, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1219-2023
Recall Event ID 91610
PMA NumberP100045 
Product Classification System, hemodynamic, implantable - Product Code MOM
ProductCardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.
Code Information UDI/DI 05414734509176, Serial Numbers: M140300439, M140300440, M140300441, M140300445, M140300446, M140300447, M140300449, M140300450, M140300451, M140300454, M140300458, M140300459, M140300460, M140300463, M140500480, M140500486, M140800987, M140800988, M140800992, M140800995, M140801002, M140801003, M140801004, M140801007, M140801010, M140801011, M141001528, M141001529, M141001530, M141001531, M141001536, M141001539, M141001541, M141001544, M141001545, M141001546, M141101678, M141101681, M141101682, M141101683, M141101684, M141101685, M141101686, M141101693, M141101696, M141101697, M141101700, M141101701, M141101703, M141101705, M141101706, M150102038, M150102039, M150102049, M150102052, M150102054, M150102055, M150102059, M150102061, M150102062, M150102064, M150102066, M150102399, M150102402, M150102404, M150102409, M150102410, M150102414, M150102416, M150202537, M150202538, M150202541, M150202542, M150202544, M150202545, M150302658, M150302660, M150302662, M150302664, M150302666, M150302667, M150302669, M150302673, M150302680, M150302682, M150302683, M150302687, M150302689, M150302690, M150302692, M150302694, M150302695, M150302702, M150302703, M150302707, M150302711, M150302712, M150302713, M150302716, M150302721, M150302723, M150302724, M150302725, M150302729, M150302737, M150302739, M150302742, M150302747, M150302748, M150302751, M150302755, M150302757, M150302758, M150302759, M150302762, M150302769, M150302773, M150302777, M150302778, M150302779, M150302780, M150302782, M150302783, M150302786, M150403048, M150403050, M150403051, M150403057, M150403066, M150805139, M150805143, M150805151, M150805154, M150805155, M150805156, M150805158, M150805160, M150805161, M150805162, M150805166, M150905677, M150905680, M150905682, M150905683, M150905685, M150905688, M150905689, M150905691, M150905692, M150905697, M150905698, M150905699, M150905702, M150905703, M150905704, M150905706, M150905707, M150905708, M150905709, M150905711, M150905714, M151206628, M151206631, M151206633, M151206636, M151206637, M151206641, M151206644, M151206645, M151206647, M151206648, M151206651, M151206655, M151206659, M151206662, M151206665, M160206849, M160206853, M160206854, M160206857, M160206861, M160206867, M160206869, M160206872, M160206876, M160206879, M160206880, M160206882, M160206885, M160206888, M160206896, M160206899, M160206900, M160206902, M160206903, M160206906, M160807699, M160807700, M160807701, M160807772.
FEI Number 3004936110
Recalling Firm/
Manufacturer
St. Jude Medical
387 Technology Cir Nw Ste 500
Atlanta GA 30313-2424
For Additional Information ContactMr. Justin Paquette
651-756-6293
Manufacturer Reason
for Recall
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
FDA Determined
Cause 2
Device Design
ActionAbbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to continue to follow CardioMEMS Instructions for Use (IFU). Abbott will notify patients of the emissions issue. If you suspect a CardioMEMS Electronics System has interfered with another medical device during use, report the event to Abbott Remote Care Technical Support at 1-844-MYCMEMS (692-6367) (US). Customers were also instructed to complete and return the acknowledgement and receipt form. Abbott notified patients on about 02/07/2023 via mailed letter. Customers were instructed that they could continue use of their CardioMEMS System and that their doctor has been notified of the issue as well. If there is suspicion that interference may be occurring during a sensor reading or any other quality issues, please report it to: Abbott Technical Support at 1-844-MYCMEMS (692-6367) (US); Monday-Friday: 8am-8pm EST
Quantity in Commerce200 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = MOM
-
-