|
Class 2 Device Recall Incisive CT |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
March 20, 2023 |
Date Posted |
May 12, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1585-2023 |
Recall Event ID |
92117 |
510(K)Number |
K212441 K180015
|
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product |
Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72814
|
Code Information |
UDI-DI: (01)00884838105508(21) + Serial number
US Serial Numbers:
550243, 34111, 34092, 34066, 34064, 34047, 34042, 34025, 34031, 34033
OUS Serial Numbers:
System Serial Number
550263
550249
550239
550215
550211
550091
550188
550187
550186
550174
550172
552041
550164
34107
550140
34062
550099
550082
550071
550068
550059
34020
550038
550040
550021
550036
550018
550014
550011
550010
550009
550008
550007
550240
550126
550241
|
Recalling Firm/ Manufacturer |
Philips North America 222 Jacobs St Cambridge MA 02141-2296
|
For Additional Information Contact |
Philips Customer Care Solutions Center 800-722-9377
|
Manufacturer Reason for Recall |
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Philips Medical issued Urgent Medical Device Correction Letter Multi-Function Switch Unload Pedal Foot Entrapment on 3/20/23 to US consignees via Certified United States Postal Service (USPS) with delivery confirmation. International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take:
Immediately stop using the Unload Pedal of the Foot Switch until Philips installs a solution on
your system.
" Instead of using the Unload Pedal, please use the Unload Patient function on the
Gantry Control Panels, or use other Table In/Out, Up/Down functions on the Gantry
Control Panels, CT Scan Control Box, or Hardware lnterventional Controls to fulfill the
normal patient unloading function, as instructed in the IFU.
" You may continue to use Load Pedal, and Free-Float Pedal of Foot Switch, as they are not
affected by this Urgent Medical Device Correction Letter.
" Please complete and return the attached response form to Philips promptly and no later than
30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device
Correction Letter, understanding of the issue, and required actions to be taken.
Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed.
Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve this issue (refer to FCO72800794).
If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377). |
Quantity in Commerce |
10 units US; 36 units OUS |
Distribution |
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, MI, MO, NY, TN,
TX, VT and the countries of Australia, Chile, Costa Rica, Czech Republic, Dominican, Republic Germany, India, Israel, Japan, Lithuania, Philippines, Poland, Sint Maarten (Dutch part), South Africa, Thailand, United Kingdom, Utd. Arab. Emir. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = Philips Healthcare (Suzhou) Co., Ltd.
|
|
|
|