• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1726-2023
Recall Event ID 92328
PMA NumberP980016S713 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductICD COBALT DR MRI IS1 DF4, Model Number DDPB3D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00763000178390, Lot Serial Numbers: RSO600207S, RSO600934S, RSO602587S, RSO602860S, RSO603892S, RSO600679S, RSO612257S, RSO602636S, RSO604992S, RSO609567S, RSO609647S, RSO609795S, RSO609900S, RSO610931S, RSO607136S, RSO609278S, RSO601095S, RSO601096S, RSO601098S, RSO601472S, RSO601473S, RSO601474S, RSO601475S, RSO601476S, RSO601477S, RSO601478S, RSO601479S, RSO604357S, RSO604358S, RSO605928S, RSO606135S, RSO606155S, RSO603632S, RSO603681S, RSO607412S, RSO607605S, RSO610650S, RSO600968S, RSO601258S, RSO603258S, RSO603300S, RSO603339S, RSO603742S, RSO603773S, RSO612385S, RSO601910S, RSO601911S, RSO601913S, RSO602604S, RSO602611S, RSO603824S, RSO603826S, RSO603876S, RSO609857S, RSO610204S, RSO609865S, RSO610317S, RSO601050S, RSO601392S, RSO601568S, RSO601741S, RSO604464S, RSO600230S, RSO610679S, RSO612158S, RSO612159S, RSO608401S, RSO608413S, RSO613154S, RSO613155S, RSO609928S, RSO610881S, RSO611808S, RSO600081S, RSO600082S, RSO600083S, RSO600951S, RSO600986S, RSO601604S, RSO604972S, RSO605522S, RSO606607S, RSO607741S, RSO606041S, RSO601440S, RSO605663S, RSO607178S, RSO607206S, RSO609698S, RSO603723S, RSO602519S, RSO603022S, RSO603027S, RSO603552S, RSO603805S, RSO603832S, RSO603833S, RSO604179S, RSO604194S, RSO604476S, RSO604477S, RSO604521S, RSO604971S, RSO604986S, RSO605288S, RSO605313S, RSO606154S, RSO606633S, RSO606984S, RSO607027S, RSO607059S, RSO607060S, RSO607066S, RSO607258S, RSO609917S, RSO601016S, RSO602843S, RSO603638S, RSO603828S, RSO603945S, RSO604008S, RSO604027S, RSO604176S, RSO607364S, RSO607365S, RSO608201S, RSO600768S, RSO602527S, RSO601005S, RSO601737S, RSO610417S, RSO601500S, RSO610630S, RSO612087S, RSO612192S, RSO612270S, RSO612547S, RSO610576S, RSO611766S, RSO611771S, RSO611773S, RSO600905S, RSO601136S, RSO601571S, RSO602734S, RSO603672S, RSO605325S, RSO606464S, RSO606516S, RSO607087S, RSO607179S, RSO607202S, RSO607203S, RSO607602S, RSO602875S, RSO602876S, RSO602877S, RSO602878S, RSO602880S, RSO602881S, RSO602882S, RSO602883S, RSO602884S, RSO604398S, RSO604416S, RSO604417S, RSO604418S, RSO604419S, RSO604420S, RSO604421S, RSO604422S, RSO604423S, RSO604424S, RSO604425S, RSO604426S, RSO604428S, RSO600619S, RSO602618S, RSO603959S, RSO603960S, RSO603961S, RSO604005S, RSO611972S, RSO611973S, RSO601416S, RSO606602S, RSO602744S, RSO603285S, RSO603286S, RSO603946S, RSO604015S, RSO604499S, RSO604965S, RSO605434S, RSO605526S, RSO605827S, RSO605981S, RSO605996S, RSO606140S, RSO606182S, RSO606990S, RSO608017S, RSO605212S, RSO607379S, RSO600542S, RSO601565S, RSO604252S, RSO604674S, RSO608842S, RSO609738S, RSO609739S, RSO610464S, RSO610539S, RSO610961S, RSO610967S, RSO611676S, RSO612239S, RSO612640S, RSO600675S, RSO600875S, RSO600876S, RSO600889S, RSO601564S, RSO601672S, RSO601692S, RSO601745S, RSO601752S, RSO603092S, RSO603653S, RSO603692S, RSO604011S, RSO604122S, RSO604251S, RSO605012S, RSO605236S, RSO605513S, RSO605821S, RSO605893S, RSO605916S, RSO606459S, RSO606528S, RSO606540S, RSO606965S, RSO606966S, RSO607237S, RSO607994S, RSO608027S, RSO608841S, RSO609002S, RSO609003S, RSO609670S, RSO609759S, RSO610510S, RSO610637S, RSO610767S, RSO611793S, RSO612692S, RSO600826S, RSO601006S, RSO601007S, RSO601044S, RSO601778S, RSO602544S, RSO602645S, RSO602707S, RSO602789S, RSO602846S, RSO603570S, RSO603710S, RSO604003S, RSO604012S, RSO604158S, RSO605435S, RSO606448S, RSO607739S, RSO600997S, RSO601143S, RSO601144S, RSO602895S, RSO602925S, RSO603346S, RSO609665S, RSO611127S, RSO600710S, RSO605598S, RSO606058S, RSO610370S, RSO603034S, RSO603711S, RSO604060S, RSO611419S, RSO601756S, RSO601825S, RSO601828S, RSO601866S, RSO604240S, RSO604519S, RSO607211S, RSO607250S, RSO608553S, RSO612321S, RSO609837S, RSO610453S, RSO610561S, RSO610850S, RSO610851S, RSO612544S, RSO607226S, RSO612324S, RSO601331S, RSO601813S, RSO606333S, RSO608018S, RSO600400S, RSO600959S, RSO601134S, RSO601383S, RSO609594S, RSO612179S, RSO601175S, RSO601915S, RSO603815S, RSO605975S, RSO607858S, RSO608939S, RSO611580S, RSO611744S, RSO605075S, RSO605076S, RSO605083S, RSO605086S, RSO605094S, RSO605099S, RSO605103S, RSO605105S, RSO605110S, RSO610959S, RSO600716S, RSO600717S, RSO600924S, RSO601179S, RSO601244S, RSO601335S, RSO601395S, RSO601396S, RSO601523S, RSO601731S, RSO603166S, RSO603488S, RSO603608S, RSO603702S, RSO604495S, RSO604497S, RSO604557S, RSO604558S, RSO605031S, RSO605616S, RSO607022S, RSO607597S, RSO607601S, RSO607654S, RSO607995S, RSO608747S, RSO609006S, RSO604814S, RSO600698S, RSO601710S, RSO602323S, RSO602936S, RSO605037S, RSO601130S, RSO600845S, RSO603422S, RSO603798S, RSO604962S, RSO608238S, RSO608535S, RSO608792S, RSO602340S, RSO605028S, RSO605417S, RSO606314S, RSO610970S, RSO602377S, RSO602378S, RSO602407S, RSO602408S, RSO602413S, RSO602417S, RSO603643S, RSO603644S, RSO603897S, RSO604383S, RSO604384S, RSO604385S, RSO604387S, RSO604388S, RSO604389S, RSO604390S, RSO604391S, RSO604392S, RSO604393S, RSO604394S, RSO604395S, RSO604396S, RSO604397S, RSO604399S, RSO603426S, RSO609731S, RSO610709S, RSO610719S, RSO611384S, RSO612126S, RSO612242S, RSO612350S, RSO612519S, RSO612520S, RSO603687S, RSO603908S, RSO604002S, RSO605098S, RSO605753S, RSO605884S, RSO606130S, RSO606219S, RSO606470S, RSO606969S, RSO607563S, RSO607594S, RSO607874S, RSO607875S, RSO607926S, RSO608386S, RSO608595S, RSO608617S, RSO608997S, RSO601321S, RSO605761S, RSO609630S, RSO610819S, RSO611142S, RSO612129S, RSO612309S, RSO612311S, RSO604624S, RSO604625S, RSO604626S, RSO606560S, RSO606981S, RSO607350S, RSO607959S, RSO608256S, RSO609612S, RSO610710S, RSO611571S, RSO611577S, RSO611708S, RSO612010S, RSO612011S, RSO612168S, RSO612178S, RSO601158S, RSO604260S, RSO605063S, RSO605130S, RSO605494S, RSO605496S, RSO605732S, RSO606265S, RSO606339S, RSO606536S, RSO607823S, RSO607900S, RSO607934S, RSO608024S, RSO608615S, RSO608663S, RSO608694S, RSO609256S, RSO609296S, RSO609415S, RSO609583S, RSO609631S, RSO609842S, RSO609905S, RSO610347S, RSO610497S, RSO610836S, RSO610837S, RSO610839S, RSO611616S, RSO611617S, RSO612049S, RSO612050S, RSO612051S, RSO612267S, RSO612330S, RSO612331S, RSO612511S, RSO600862S, RSO600910S, RSO600912S, RSO600913S, RSO600914S, RSO600915S, RSO600916S, RSO601240S, RSO601243S, RSO601300S, RSO601361S, RSO601563S, RSO601690S, RSO601887S, RSO601929S, RSO602758S, RSO602889S, RSO602958S, RSO603071S, RSO603579S, RSO603590S, RSO603595S, RSO603821S, RSO603831S, RSO603890S, RSO603903S, RSO603910S, RSO603971S, RSO603984S, RSO604086S, RSO604588S, RSO604870S, RSO605005S, RSO605006S, RSO605101S, RSO605218S, RSO605253S, RSO605262S, RSO605361S, RSO605395S, RSO605412S, RSO605765S, RSO605883S, RSO605890S, RSO606120S, RSO606143S, RSO606268S, RSO606500S, RSO606557S, RSO606604S, RSO606662S, RSO606663S, RSO606777S, RSO607101S, RSO607199S, RSO607200S, RSO607256S, RSO607325S, RSO607960S, RSO608051S, RSO608053S, RSO608258S, RSO608305S, RSO608309S, RSO608473S, RSO608478S, RSO608566S, RSO608568S, RSO608622S, RSO608625S, RSO608716S, RSO608769S, RSO608771S, RSO608947S, RSO608948S, RSO609142S, RSO609199S, RSO611980S, RSO611981S, RSO611982S, RSO611983S, RSO609632S, RSO609637S, RSO609849S, RSO609932S, RSO610429S, RSO610568S, RSO610742S, RSO611356S, RSO611655S, RSO612100S, RSO600676S, RSO600748S, RSO600891S, RSO600898S, RSO601114S, RSO601620S, RSO604023S, RSO604735S, RSO605729S, RSO606124S, RSO606320S, RSO606476S, RSO606501S, RSO606568S, RSO606615S, RSO607002S, RSO607123S, RSO607305S, RSO607307S, RSO607533S, RSO607856S, RSO607864S, RSO607910S, RSO607938S, RSO607945S, RSO607947S, RSO608333S, RSO608685S, RSO608810S, RSO608909S, RSO612282S, RSO601847S, RSO603623S, RSO604303S, RSO605413S, RSO605661S, RSO609667S, RSO609668S, RSO609954S, RSO609956S, RSO610952S, RSO610965S, RSO603312S, RSO604206S, RSO604265S, RSO604445S, RSO605120S, RSO605374S, RSO605398S, RSO605845S, RSO605888S, RSO605910S, RSO606295S, RSO606988S, RSO607607S, RSO607623S, RSO607922S, RSO607958S, RSO608292S, RSO608746S, RSO608845S, RSO609912S, RSO610854S, RSO611622S, RSO606255S, RSO607072S, RSO608032S, RSO608619S, RSO608777S, RSO609440S, RSO612130S, RSO607564S, RSO609509S, RSO601201S, RSO611621S, RSO612582S, RSO605340S, RSO605956S, RSO606832S, RSO610401S, RSO611661S, RSO600992S, RSO609581S, RSO610560S, RSO612191S, RSO612315S, RSO612540S, RSO605922S, RSO606125S, RSO608803S, RSO606956S, RSO605352S, RSO607025S, RSO601056S, RSO610487S, RSO612128S, RSO612165S, RSO606290S, RSO600183S, RSO600645S, RSO600678S, RSO600752S, RSO600911S, RSO601088S, RSO601412S, RSO601815S, RSO601830S, RSO601853S, RSO602450S, RSO602746S, RSO602763S, RSO602932S, RSO603060S, RSO603260S, RSO603379S, RSO603539S, RSO603704S, RSO603848S, RSO609350S, RSO600784S, RSO601665S, RSO601860S, RSO603380S, RSO605029S, RSO606349S, RSO612306S, RSO612566S, RSO600167S, RSO600168S, RSO600685S, RSO600686S, RSO600818S, RSO600879S, RSO600880S, RSO600881S, RSO600892S, RSO600893S, RSO600894S, RSO600895S, RSO601123S, RSO602625S, RSO602628S, RSO603229S, RSO603478S, RSO603629S, RSO603630S, RSO603657S, RSO603709S, RSO603727S, RSO605102S, RSO605254S, RSO605711S, RSO610419S, RSO603576S, RSO609646S, RSO610502S, RSO612017S, RSO612057S, RSO601493S, RSO606259S, RSO606538S, RSO607174S, RSO607283S, RSO608560S, RSO608599S, RSO613123S, RSO610585S, RSO611639S, RSO601068S, RSO603452S, RSO604575S, RSO604761S, RSO605092S, RSO606546S, RSO606766S, RSO606962S, RSO608026S, RSO611718S, RSO606519S, RSO606852S, RSO600251S, RSO601170S, RSO603094S, RSO603240S, RSO603373S, RSO609551S, RSO610587S, RSO611731S, RSO600180S, RSO600481S, RSO600922S, RSO601430S, RSO601685S, RSO610794S, RSO611643S, RSO612225S, RSO612474S, RSO612541S, RSO612619S, RSO608456S, RSO612706S, RSO609749S, RSO609886S, RSO610558S, RSO610975S, RSO611569S, RSO612069S, RSO612135S, RSO612491S, RSO612560S, RSO601505S, RSO601728S, RSO602755S, RSO604245S, RSO604250S, RSO604448S, RSO604564S, RSO605197S, RSO605311S, RSO605622S, RSO605667S, RSO605828S, RSO605844S, RSO606275S, RSO606357S, RSO606409S, RSO606485S, RSO607041S, RSO607055S, RSO607092S, RSO607105S, RSO607109S, RSO607134S, RSO607149S, RSO607415S, RSO607580S, RSO607636S, RSO607637S, RSO607896S, RSO608451S, RSO608604S, RSO608661S, RSO610410S, RSO610960S, RSO610966S, RSO610968S, RSO610428S, RSO603994S, RSO603997S, RSO604374S, RSO605178S, RSO605637S, RSO606023S, RSO606260S, RSO606364S, RSO606559S, RSO606595S, RSO608644S, RSO609833S, RSO610351S, RSO610384S, RSO610408S, RSO611399S, RSO611704S, RSO612207S, RSO612241S, RSO612650S, RSO612652S, RSO600975S, RSO601353S, RSO601593S, RSO601920S, RSO602397S, RSO602723S, RSO603207S, RSO604299S, RSO604310S, RSO604872S, RSO605157S, RSO605172S, RSO605224S, RSO605250S, RSO605281S, RSO605823S, RSO606370S, RSO606373S, RSO607294S, RSO607436S, RSO607817S, RSO610702S, RSO605300S, RSO607037S, RSO601739S, RSO605210S, RSO609806S, RSO612196S, RSO601209S, RSO612637S, RSO612638S, RSO612639S, RSO612710S, RSO612711S, RSO612712S, RSO610372S, RSO610641S, RSO600731S, RSO600732S, RSO600733S, RSO600811S, RSO600812S, RSO600816S, RSO600902S, RSO601062S, RSO601084S, RSO601140S, RSO601183S, RSO601279S, RSO601282S, RSO601284S, RSO601303S, RSO601798S, RSO601871S, RSO601878S, RSO601895S, RSO602873S, RSO602917S, RSO603070S, RSO603085S, RSO603170S, RSO603227S, RSO604070S, RSO604487S, RSO605205S, RSO607445S, RSO608191S, RSO608254S, RSO608579S, RSO608580S, RSO608630S, RSO608631S, RSO608653S, RSO609860S, RSO600267S, RSO600770S, RSO601471S, RSO603375S, RSO605306S, RSO605353S, RSO607148S, RSO607151S, RSO607794S, RSO609007S, RSO613216S, RSO609629S, RSO609817S, RSO611293S, RSO611574S, RSO611992S, RSO612206S, RSO612240S, RSO600955S, RSO603847S, RSO605100S, RSO605373S, RSO608237S, RSO608765S, RSO609011S, RSO610872S, RSO611790S, RSO600213S, RSO600216S, RSO600243S, RSO600872S, RSO600933S, RSO601148S, RSO601373S, RSO601628S, RSO602548S, RSO602592S, RSO602753S, RSO603013S, RSO603019S, RSO603790S, RSO604004S, RSO604124S, RSO604136S, RSO604200S, RSO604223S, RSO604945S, RSO605416S, RSO606838S, RSO609423S, RSO610969S, RSO611772S, RSO611785S, RSO609633S, RSO609636S, RSO609688S, RSO609856S, RSO609936S, RSO611806S, RSO600864S, RSO600965S, RSO601039S, RSO601174S, RSO601178S, RSO601180S, RSO601253S, RSO601280S, RSO601324S, RSO601325S, RSO601609S, RSO601618S, RSO601640S, RSO601642S, RSO601643S, RSO601647S, RSO602309S, RSO602310S, RSO602481S, RSO602516S, RSO602552S, RSO602556S, RSO602558S, RSO602568S, RSO603058S, RSO603118S, RSO603165S, RSO603167S, RSO603213S, RSO603233S, RSO603267S, RSO603290S, RSO603331S, RSO603384S, RSO603409S, RSO603412S, RSO603476S, RSO603511S, RSO603571S, RSO603637S, RSO603808S, RSO603840S, RSO603873S, RSO603874S, RSO603894S, RSO603965S, RSO603969S, RSO603985S, RSO604033S, RSO604278S, RSO604484S, RSO604485S, RSO604493S, RSO604498S, RSO604563S, RSO605038S, RSO605133S, RSO605184S, RSO605403S, RSO605419S, RSO605422S, RSO605423S, RSO605592S, RSO605594S, RSO605878S, RSO605958S, RSO605960S, RSO605965S, RSO605966S, RSO606443S, RSO608542S, RSO608607S, RSO608646S, RSO608815S, RSO601394S, RSO601666S, RSO604222S, RSO605158S, RSO605625S, RSO606136S, RSO606236S, RSO609591S, RSO609915S, RSO608559S, RSO611819S, RSO611820S, RSO611821S, RSO611974S, RSO612691S, RSO612553S, RSO612554S, RSO612628S, RSO608379S, RSO613261S, RSO613272S, RSO609371S, RSO601754S, RSO602976S, RSO603507S, RSO609747S, RSO610624S, RSO612358S, RSO612499S, RSO603278S, RSO606975S, RSO613223S, RSO609726S, RSO611409S, RSO611597S, RSO601344S, RSO606799S, RSO601081S, RSO601286S, RSO601316S, RSO603020S, RSO603309S, RSO603377S, RSO603434S, RSO603462S, RSO603498S, RSO603517S, RSO603532S, RSO604053S, RSO604071S, RSO604078S, RSO604160S, RSO604214S, RSO604550S, RSO604999S, RSO605043S, RSO605046S, RSO605387S, RSO605518S, RSO605544S, RSO605852S, RSO605907S, RSO606127S, RSO606234S, RSO607111S, RSO607290S, RSO607423S, RSO608610S, RSO608744S, RSO608922S, RSO606354S, RSO612348S, RSO605867S, RSO611503S, RSO601548S, RSO601827S, RSO603217S, RSO605162S, RSO605367S, RSO606495S, RSO607232S, RSO607876S, RSO608593S, RSO600136S, RSO601377S, RSO604277S, RSO605362S, RSO606597S, RSO606874S, RSO608414S, RSO610577S, RSO605726S, RSO610849S, RSO608261S, RSO608262S, RSO609452S, RSO609701S, RSO611599S, RSO601333S, RSO601670S, RSO601804S, RSO604262S, RSO604263S, RSO605073S, RSO605913S, RSO606095S, RSO604006S, RSO600252S, RSO600515S, RSO600518S, RSO601085S, RSO602566S, RSO604117S, RSO605219S, RSO606170S, RSO606171S, RSO607341S, RSO609840S, RSO612677S, RSO601239S, RSO601454S, RSO601455S, RSO603247S, RSO605332S, RSO605718S, RSO605849S, RSO605879S, RSO606251S, RSO607165S, RSO607949S, RSO610067S, RSO610068S, RSO611898S, RSO612292S, RSO612293S, RSO612483S, RSO612484S, RSO609267S, RSO613161S, RSO613162S, RSO609735S, RSO609951S, RSO610463S, RSO610532S, RSO610535S, RSO610538S, RSO610609S, RSO600664S, RSO600669S, RSO601211S, RSO601824S, RSO602370S, RSO603445S, RSO603506S, RSO603567S, RSO603732S, RSO604080S, RSO604555S, RSO604966S, RSO605243S, RSO605382S, RSO605702S, RSO606011S, RSO606289S, RSO606381S, RSO607295S, RSO607352S, RSO607611S, RSO607681S, RSO608670S, RSO608745S, RSO608904S, RSO608906S, RSO608916S, RSO609270S, RSO601470S, RSO603463S, RSO605013S, RSO612076S, RSO606269S, RSO609839S, RSO609841S, RSO609909S, RSO610842S, RSO612175S, RSO612332S, RSO612571S, RSO612573S, RSO605199S, RSO608009S, RSO608194S, RSO608195S, RSO608713S, RSO608888S, RSO609281S, RSO609304S, RSO609466S, RSO605053S, RSO605085S, RSO605107S, RSO605301S, RSO605943S, RSO610760S, RSO609618S, RSO603303S, RSO604261S, RSO604264S, RSO605740S, RSO608785S, RSO609282S, RSO612188S, RSO612497S, RSO613255S, RSO607213S, RSO600145S, RSO600210S, RSO600237S, RSO600526S, RSO600801S, RSO600948S, RSO601370S, RSO601562S, RSO605936S, RSO606419S, RSO601426S, RSO607172S, RSO600648S, RSO600774S, RSO601543S, RSO601701S, RSO605507S, RSO607347S, RSO609940S, RSO605292S, RSO607869S, RSO609893S, RSO611991S, RSO612523S, RSO602896S, RSO603256S, RSO603647S, RSO603728S, RSO603729S, RSO604115S, RSO604318S, RSO604323S, RSO604593S, RSO604659S, RSO604893S, RSO605009S, RSO605349S, RSO605564S, RSO605602S, RSO605617S, RSO605741S, RSO606305S, RSO606411S, RSO606434S, RSO606517S, RSO606522S, RSO606570S, RSO606998S, RSO607021S, RSO607173S, RSO607603S, RSO607604S, RSO608886S, RSO613196S, RSO602800S, RSO603393S, RSO606549S, RSO610387S, RSO610492S, RSO612498S, RSO603073S, RSO604401S, RSO604402S, RSO604403S, RSO607685S, RSO608573S, RSO609287S, RSO609307S, RSO603597S, RSO606382S, RSO609734S, RSO601880S, RSO603518S, RSO604292S, RSO605326S, RSO606257S, RSO606368S, RSO609763S, RSO609925S, RSO609926S, RSO610542S, RSO610720S, RSO611518S, RSO612099S, RSO612383S, RSO601248S, RSO601249S, RSO601250S, RSO601252S, RSO601292S, RSO601375S, RSO601534S, RSO601544S, RSO601721S, RSO601767S, RSO601822S, RSO601883S, RSO602503S, RSO602739S, RSO602890S, RSO602914S, RSO602954S, RSO602974S, RSO603000S, RSO603127S, RSO603130S, RSO603168S, RSO603475S, RSO603686S, RSO603779S, RSO603872S, RSO604035S, RSO604177S, RSO604560S, RSO604561S, RSO604565S, RSO604815S, RSO605016S, RSO605017S, RSO605018S, RSO605020S, RSO605022S, RSO605062S, RSO605303S, RSO605304S, RSO605305S, RSO605308S, RSO605309S, RSO605359S, RSO605575S, RSO605759S, RSO605886S, RSO605999S, RSO606213S, RSO606233S, RSO606248S, RSO606277S, RSO606450S, RSO606465S, RSO606535S, RSO606555S, RSO607119S, RSO607405S, RSO607684S, RSO607972S, RSO608022S, RSO608023S, RSO608224S, RSO608257S, RSO608316S, RSO608702S, RSO608703S, RSO609596S, RSO609799S, RSO611098S, RSO600813S, RSO600814S, RSO603140S, RSO603408S, RSO603455S, RSO603634S, RSO603842S, RSO604125S, RSO604473S, RSO605175S, RSO605176S, RSO606408S, RSO606533S, RSO607939S, RSO607982S, RSO608677S, RSO611547S, RSO611549S, RSO612081S, RSO612269S, RSO600936S, RSO601054S, RSO603023S, RSO603639S, RSO604219S, RSO605647S, RSO606899S, RSO607164S, RSO607340S, RSO607673S, RSO608783S, RSO608952S, RSO609638S, RSO609708S, RSO609538S, RSO609570S, RSO609620S, RSO609622S, RSO609625S, RSO609626S, RSO609692S, RSO609810S, RSO609823S, RSO609826S, RSO609904S, RSO610375S, RSO610380S, RSO610392S, RSO610476S, RSO610581S, RSO610892S, RSO610905S, RSO611646S, RSO600140S, RSO600144S, RSO600595S, RSO601067S, RSO601083S, RSO601124S, RSO601132S, RSO601307S, RSO601408S, RSO601409S, RSO601488S, RSO601542S, RSO601553S, RSO601684S, RSO601695S, RSO601775S, RSO601832S, RSO601901S, RSO602720S, RSO602732S, RSO602814S, RSO602825S, RSO602935S, RSO602972S, RSO603018S, RSO603068S, RSO603106S, RSO603172S, RSO603204S, RSO603243S, RSO603392S, RSO603402S, RSO603444S, RSO603446S, RSO603605S, RSO603631S, RSO603695S, RSO603819S, RSO604021S, RSO604144S, RSO604151S, RSO604165S, RSO604202S, RSO604505S, RSO604559S, RSO604595S, RSO604939S, RSO605014S, RSO605164S, RSO605174S, RSO605179S, RSO605193S, RSO605464S, RSO605613S, RSO605688S, RSO605703S, RSO605767S, RSO605915S, RSO605937S, RSO606060S, RSO606105S, RSO606344S, RSO606345S, RSO606414S, RSO606513S, RSO606584S, RSO606588S, RSO606653S, RSO606987S, RSO607125S, RSO607147S, RSO607155S, RSO607208S, RSO607225S, RSO607279S, RSO607333S, RSO607334S, RSO607335S, RSO607348S, RSO607358S, RSO607376S, RSO607378S, RSO607705S, RSO607715S, RSO607718S, RSO607722S, RSO607732S, RSO607861S, RSO607901S, RSO607992S, RSO608004S, RSO608029S, RSO608209S, RSO608210S, RSO608211S, RSO608212S, RSO608214S, RSO608215S, RSO608245S, RSO608314S, RSO608317S, RSO608491S, RSO608492S, RSO608495S, RSO608496S, RSO608497S, RSO608503S, RSO608524S, RSO608633S, RSO608634S, RSO608635S, RSO608637S, RSO608671S, RSO608672S, RSO608675S, RSO608737S, RSO608826S, RSO608980S, RSO608981S, RSO609273S, RSO609274S, RSO609332S, RSO609360S, RSO601588S, RSO602903S, RSO604131S, RSO605365S, RSO605521S, RSO602771S, RSO602777S, RSO610605S, RSO611589S, RSO611782S, RSO611783S, RSO606331S, RSO606398S, RSO612250S, RSO600847S, RSO603341S, RSO603424S, RSO603525S, RSO604350S, RSO606026S, RSO609595S, RSO600863S, RSO600942S, RSO601141S, RSO601268S, RSO601278S, RSO601814S, RSO602551S, RSO602560S, RSO602579S, RSO602586S, RSO602820S, RSO603205S, RSO603474S, RSO603807S, RSO603835S, RSO603937S, RSO603939S, RSO603940S, RSO603975S, RSO604036S, RSO604532S, RSO604549S, RSO604989S, RSO605338S, RSO605492S, RSO605968S, RSO605976S, RSO606329S, RSO606634S, RSO606693S, RSO606879S, RSO607177S, RSO607261S, RSO607355S, RSO607751S, RSO607890S, RSO608513S, RSO608707S, RSO610146S, RSO610160S, RSO610162S, RSO612518S, RSO600627S, RSO600660S, RSO600741S, RSO600926S, RSO605291S, RSO605302S, RSO605418S, RSO605730S, RSO605989S, RSO606554S, RSO607756S, RSO609877S, RSO609878S, RSO609879S, RSO609880S, RSO609881S, RSO610039S, RSO610436S, RSO610439S, RSO610442S, RSO610443S, RSO610444S, RSO610801S, RSO611517S, RSO611521S, RSO611610S, RSO611815S, RSO611816S, RSO611899S, RSO611900S, RSO611901S, RSO612164S, RSO612166S, RSO612167S, RSO600776S, RSO602336S, RSO604492S, RSO604494S, RSO604496S, RSO604500S, RSO604501S, RSO604694S, RSO604697S, RSO604699S, RSO604700S, RSO604703S, RSO604705S, RSO604708S, RSO604714S, RSO604716S, RSO604927S, RSO604933S, RSO604937S, RSO604947S, RSO608548S, RSO608549S, RSO608550S, RSO608551S, RSO609017S, RSO609106S, RSO609107S, RSO609108S, RSO609109S, RSO609110S, RSO611074S, RSO612775S, RSO612776S, RSO612777S, RSO612778S, RSO612780S, RSO612781S, RSO612782S, RSO612783S, RSO612784S, RSO612785S, RSO612786S, RSO612788S, RSO612789S, RSO612790S, RSO612791S, RSO600702S, RSO600780S, RSO600781S, RSO600782S, RSO600783S, RSO600848S, RSO600853S, RSO601107S, RSO601429S, RSO603565S, RSO605605S, RSO605606S, RSO605731S, RSO605991S, RSO606193S, RSO607357S, RSO607424S, RSO607701S, RSO607702S, RSO607703S, RSO609755S, RSO609757S, RSO609758S, RSO610771S, RSO611800S, RSO612042S, RSO600785S, RSO600804S, RSO600833S, RSO601164S, RSO601236S, RSO601255S, RSO601256S, RSO601350S, RSO601499S, RSO601898S, RSO602603S, RSO602733S, RSO602852S, RSO603139S, RSO603242S, RSO603523S, RSO605393S, RSO605536S, RSO605585S, RSO605607S, RSO607366S, RSO607686S, RSO607687S, RSO607727S, RSO607730S, RSO608033S, RSO611147S, RSO612203S, RSO612228S, RSO600408S, RSO605674S, RSO606220S, RSO606426S, RSO608990S, RSO600275S, RSO600945S, RSO600979S, RSO605161S, RSO605557S, RSO605854S, RSO605863S, RSO606201S, RSO606209S, RSO611510S, RSO612002S, RSO612097S, RSO600178S, RSO601421S, RSO602805S, RSO602864S, RSO604616S, RSO606126S, RSO602299S, RSO611723S, RSO603468S, RSO604139S, RSO604140S, RSO604381S, RSO605638S, RSO605639S, RSO607387S, RSO607530S, RSO603160S, RSO604056S, RSO606993S, RSO609736S, RSO609953S, RSO610606S, RSO610955S, RSO610958S, RSO611145S, RSO611759S, RSO611762S, RSO611765S, RSO611779S, RSO611780S, RSO602440S, RSO603306S, RSO603482S, RSO603694S, RSO606107S, RSO606460S, RSO607212S, RSO608388S, RSO612936S, RSO612480S, RSO611627S, RSO600071S, RSO600115S, RSO600116S, RSO600219S, RSO600944S, RSO601181S, RSO601658S, RSO603556S, RSO603865S, RSO603955S, RSO605707S, RSO607011S, RSO600588S, RSO611807S, RSO612082S, RSO612327S, RSO600079S, RSO603884S, RSO603924S, RSO604351S, RSO604352S, RSO608321S, RSO608326S, RSO613250S, RSO600852S, RSO604256S, RSO607962S, RSO607963S, RSO601540S, RSO603785S, RSO603788S, RSO603825S, RSO603849S, RSO603850S, RSO603851S, RSO604312S, RSO604511S, RSO604512S, RSO604910S, RSO604925S, RSO605074S, RSO605698S, RSO605872S, RSO606304S, RSO606307S, RSO606319S, RSO606323S, RSO606326S, RSO607356S, RSO607386S, RSO607396S, RSO607578S, RSO607829S, RSO607830S, RSO607831S, RSO607832S, RSO609552S, RSO609650S, RSO610470S, RSO610549S, RSO611713S, RSO612202S, RSO600778S, RSO600906S, RSO601058S, RSO601541S, RSO603069S, RSO603584S, RSO604969S, RSO605252S, RSO605384S, RSO605712S, RSO606045S, RSO606652S, RSO607980S, RSO608498S, RSO609374S, RSO609562S, RSO609902S, RSO610475S, RSO611524S, RSO611550S, RSO611674S, RSO612345S, RSO600907S, RSO607198S, RSO610588S, RSO610936S, RSO612344S, RSO600323S, RSO600332S, RSO600788S, RSO600874S, RSO603335S, RSO603486S, RSO608988S, RSO611978S, RSO611979S, RSO612763S, RSO612764S, RSO609527S, RSO609528S, RSO609635S, RSO609689S, RSO609855S, RSO609935S, RSO610527S, RSO610887S, RSO610888S, RSO610942S, RSO610943S, RSO611638S, RSO611804S, RSO611805S, RSO611839S, RSO612083S, RSO612084S, RSO612085S, RSO612328S, RSO612329S, RSO600108S, RSO600984S, RSO601078S, RSO601103S, RSO601623S, RSO601639S, RSO601712S, RSO602334S, RSO602510S, RSO602555S, RSO602574S, RSO602584S, RSO603232S, RSO603235S, RSO603581S, RSO603854S, RSO603856S, RSO603860S, RSO603863S, RSO603866S, RSO603867S, RSO604488S, RSO604490S, RSO605322S, RSO605324S, RSO605836S, RSO605870S, RSO605873S, RSO605876S, RSO605933S, RSO606417S, RSO606565S, RSO606569S, RSO606574S, RSO606580S, RSO606585S, RSO606589S, RSO606590S, RSO607407S, RSO607410S, RSO607624S, RSO607626S, RSO607627S, RSO607628S, RSO607988S, RSO608543S, RSO611681S, RSO609768S, RSO600714S, RSO600739S, RSO600792S, RSO601434S, RSO601823S, RSO601843S, RSO602822S, RSO602902S, RSO603673S, RSO603905S, RSO604066S, RSO604162S, RSO604290S, RSO604305S, RSO604973S, RSO605108S, RSO605244S, RSO605248S, RSO605249S, RSO605714S, RSO606478S, RSO607167S, RSO607259S, RSO607313S, RSO607911S, RSO607912S, RSO600540S, RSO600639S, RSO607485S, RSO604472S, RSO609858S, RSO610897S, RSO611015S, RSO611642S, RSO612124S, RSO612646S, RSO607017S, RSO609514S, RSO609517S, RSO609613S, RSO609832S, RSO610356S, RSO610457S, RSO610474S, RSO610615S, RSO610707S, RSO610716S, RSO611573S, RSO611702S, RSO612110S, RSO612172S, RSO612209S, RSO612487S, RSO612649S, RSO612655S, RSO601575S, RSO601831S, RSO603176S, RSO603737S, RSO603896S, RSO604585S, RSO604802S, RSO604902S, RSO604903S, RSO605558S, RSO605737S, RSO606474S, RSO606561S, RSO607525S, RSO607591S, RSO607720S, RSO608236S, RSO608381S, RSO608387S, RSO608487S, RSO608616S, RSO608719S, RSO613203S, RSO613204S, RSO609564S, RSO609565S, RSO609771S, RSO609884S, RSO610385S, RSO600535S, RSO600652S, RSO600743S, RSO600744S, RSO600963S, RSO600971S, RSO600990S, RSO601018S, RSO601235S, RSO601275S, RSO601532S, RSO601567S, RSO601788S, RSO601797S, RSO601859S, RSO601891S, RSO602452S, RSO602553S, RSO602806S, RSO602957S, RSO603001S, RSO603230S, RSO603342S, RSO603369S, RSO603542S, RSO603554S, RSO603747S, RSO603999S, RSO604077S, RSO604858S, RSO604926S, RSO605061S, RSO605208S, RSO605261S, RSO605331S, RSO605333S, RSO605683S, RSO605709S, RSO605997S, RSO606347S, RSO606567S, RSO607122S, RSO607351S, RSO607662S, RSO608735S, RSO608995S, RSO609372S, RSO601587S, RSO601805S, RSO611603S, RSO603323S, RSO604286S, RSO609510S, RSO602803S, RSO609938S, RSO600599S, RSO600609S, RSO600877S, RSO601302S, RSO601533S, RSO602767S, RSO602810S, RSO603005S, RSO603371S, RSO603492S, RSO603703S, RSO605335S, RSO605701S, RSO606048S, RSO604042S, RSO608844S, RSO609746S, RSO612046S, RSO612047S, RSO612143S, RSO612274S, RSO612354S, RSO612583S, RSO612584S, RSO612690S, RSO604990S, RSO606644S, RSO607552S, RSO607976S, RSO607977S, RSO608177S, RSO608180S, RSO608942S, RSO609428S, RSO609546S, RSO605127S, RSO607760S, RSO609867S, RSO610205S, RSO610477S, RSO610634S, RSO610635S, RSO611750S, RSO612000S, RSO612094S, RSO612095S, RSO612276S, RSO612608S, RSO600151S, RSO600169S, RSO600170S, RSO600831S, RSO600935S, RSO601063S, RSO601086S, RSO601089S, RSO601156S, RSO601617S, RSO602316S, RSO603110S, RSO603502S, RSO603591S, RSO603809S, RSO603811S, RSO605268S, RSO605336S, RSO605463S, RSO606029S, RSO607987S, RSO608065S, RSO608760S, RSO600841S, RSO602731S, RSO603028S, RSO607322S, RSO608248S, RSO609542S, RSO609803S, RSO610365S, RSO610376S, RSO610880S, RSO600734S, RSO605289S, RSO606174S, RSO606429S, RSO607571S, RSO608318S, RSO608502S, RSO608927S, RSO601789S, RSO607140S, RSO607648S, RSO608192S, RSO604483S, RSO604489S, RSO602301S, RSO603202S, RSO606631S, RSO600854S, RSO600856S, RSO605457S, RSO605656S, RSO606888S, RSO609532S, RSO609557S, RSO609558S, RSO609677S, RSO609807S, RSO609847S, RSO609910S, RSO610945S, RSO600132S, RSO600133S, RSO600164S, RSO600166S, RSO600271S, RSO600295S, RSO600468S, RSO600662S, RSO600697S, RSO600819S, RSO600930S, RSO600941S, RSO601035S, RSO601196S, RSO601343S, RSO601365S, RSO601441S, RSO601660S, RSO601714S, RSO602298S, RSO602311S, RSO602315S, RSO602331S, RSO602942S, RSO602956S, RSO603038S, RSO603189S, RSO603310S, RSO603327S, RSO603354S, RSO603427S, RSO603976S, RSO603988S, RSO604026S, RSO604218S, RSO604462S, RSO605659S, RSO605898S, RSO605935S, RSO605993S, RSO606133S, RSO606469S, RSO606639S, RSO606758S, RSO606836S, RSO606847S, RSO606887S, RSO606995S, RSO607065S, RSO607088S, RSO607171S, RSO607242S, RSO607617S, RSO607621S, RSO607737S, RSO607749S, RSO607991S, RSO608353S, RSO608426S, RSO608816S, RSO608817S, RSO608932S, RSO609676S, RSO609725S, RSO601326S, RSO601327S, RSO603754S, RSO603953S, RSO604221S, RSO604241S, RSO605072S, RSO605410S, RSO607131S, RSO608868S, RSO604887S, RSO607112S, RSO608873S, RSO612726S, RSO612727S, RSO612728S, RSO612729S, RSO612730S, RSO612732S, RSO612733S, RSO612734S, RSO612735S, RSO612736S, RSO612737S, RSO612740S, RSO612742S, RSO612743S, RSO612744S, RSO612745S, RSO612747S, RSO612748S, RSO612749S, RSO612750S, RSO612751S, RSO612058S, RSO612144S, RSO612215S, RSO600953S, RSO601645S, RSO606297S, RSO608202S, RSO609605S, RSO609675S, RSO609680S, RSO609684S, RSO609752S, RSO609835S, RSO609916S, RSO610412S, RSO610430S, RSO610482S, RSO600233S, RSO600543S, RSO600544S, RSO600793S, RSO601121S, RSO601261S, RSO601574S, RSO602990S, RSO603216S, RSO603360S, RSO603367S, RSO603370S, RSO603382S, RSO603601S, RSO603880S, RSO603956S, RSO604138S, RSO604213S, RSO604874S, RSO604950S, RSO604980S, RSO605315S, RSO605488S, RSO605501S, RSO605697S, RSO605796S, RSO605832S, RSO605952S, RSO605957S, RSO606118S, RSO606121S, RSO606178S, RSO606605S, RSO606894S, RSO607096S, RSO607162S, RSO607236S, RSO607246S, RSO607249S, RSO607537S, RSO607726S, RSO607801S, RSO607811S, RSO607993S, RSO608727S, RSO608728S, RSO608729S, RSO608818S, RSO608969S, RSO608972S, RSO608973S, RSO609018S, RSO609402S, RSO601131S, RSO600119S, RSO600122S, RSO600361S, RSO600656S, RSO600873S, RSO601381S, RSO603540S, RSO603786S, RSO604321S, RSO605624S, RSO609491S, RSO609494S, RSO609495S, RSO609740S, RSO609862S, RSO610478S, RSO610491S, RSO610798S, RSO610804S, RSO611511S, RSO611749S, RSO612003S, RSO612156S, RSO612609S, RSO612701S, RSO601064S, RSO601627S, RSO602892S, RSO603404S, RSO603802S, RSO603803S, RSO604029S, RSO605140S, RSO605173S, RSO605453S, RSO605459S, RSO605466S, RSO605520S, RSO605676S, RSO605970S, RSO606097S, RSO606425S, RSO606880S, RSO606904S, RSO607782S, RSO607784S, RSO607929S, RSO608069S, RSO608071S, RSO608320S, RSO608761S, RSO600673S, RSO600674S, RSO603496S, RSO603683S, RSO603743S, RSO604266S, RSO605859S, RSO605860S, RSO606306S, RSO606992S, RSO607104S, RSO607526S, RSO607797S, RSO607798S, RSO607946S, RSO612757S, RSO609898S, RSO600989S, RSO601271S, RSO601310S, RSO601582S, RSO601774S, RSO608342S, RSO612096S, RSO604581S, RSO605641S, RSO602750S, RSO604737S, RSO608399S, RSO603654S, RSO604019S, RSO605358S, RSO606542S, RSO607128S, RSO608036S, RSO601664S, RSO600107S, RSO601008S, RSO601443S, RSO601624S, RSO602738S, RSO603122S, RSO603208S, RSO605835S, RSO605880S, RSO605881S, RSO607003S, RSO608598S, RSO609706S, RSO601102S, RSO601322S, RSO601329S, RSO601667S, RSO601889S, RSO602759S, RSO602984S, RSO603129S, RSO603311S, RSO603736S, RSO604967S, RSO605404S, RSO605682S, RSO606302S, RSO606545S, RSO607338S, RSO607710S, RSO607712S, RSO608044S, RSO608048S, RSO608965S, RSO605528S, RSO607121S, RSO600109S, RSO600113S, RSO600187S, RSO600264S, RSO600269S, RSO601004S, RSO601917S, RSO602304S, RSO602325S, RSO602449S, RSO602518S, RSO602561S, RSO602567S, RSO603816S, RSO603857S, RSO605784S, RSO605853S, RSO605953S, RSO605954S, RSO606841S, RSO609294S, RSO612534S, RSO603056S, RSO612251S, RSO612352S, RSO609277S, RSO607893S, RSO610684S, RSO610685S, RSO607740S, RSO609257S, RSO605228S, RSO610171S, RSO610590S, RSO609691S, RSO609883S, RSO609889S, RSO610418S, RSO610466S, RSO603458S, RSO605230S, RSO605290S, RSO605376S, RSO605415S, RSO605675S, RSO606433S, RSO606627S, RSO606681S, RSO607005S, RSO607210S, RSO607377S, RSO607566S, RSO607860S, RSO608636S, RSO608924S, RSO608984S, RSO609368S, RSO600722S, RSO600777S, RSO601223S, RSO601254S, RSO601337S, RSO601526S, RSO601529S, RSO601551S, RSO601661S, RSO601800S, RSO602724S, RSO603136S, RSO603245S, RSO603287S, RSO603288S, RSO603658S, RSO604327S, RSO604447S, RSO604856S, RSO605420S, RSO605506S, RSO606380S, RSO607050S, RSO607694S, RSO609262S, RSO600644S, RSO606756S, RSO605317S, RSO605079S, RSO604126S, RSO604427S, RSO610413S, RSO610414S, RSO612148S, RSO612149S, RSO612312S, RSO603062S, RSO606190S, RSO606266S, RSO606267S, RSO606435S, RSO606782S, RSO607557S, RSO608639S, RSO608733S, RSO611612S, RSO612486S, RSO612488S, RSO612653S, RSO612654S, RSO600659S, RSO603237S, RSO605643S, RSO603620S, RSO605429S, RSO607324S, RSO609996S, RSO610012S, RSO610024S, RSO610026S, RSO610203S, RSO610330S, RSO610433S, RSO610435S, RSO610445S, RSO610446S, RSO611011S, RSO611027S, RSO611318S, RSO611341S, RSO611833S, RSO611834S, RSO611835S, RSO612526S, RSO612527S, RSO604375S, RSO604507S, RSO605024S, RSO605040S, RSO605042S, RSO605934S, RSO606229S, RSO606328S, RSO606480S, RSO606492S, RSO606529S, RSO606893S, RSO606918S, RSO607024S, RSO607282S, RSO607417S, RSO607575S, RSO607618S, RSO607839S, RSO607854S, RSO607877S, RSO608359S, RSO608428S, RSO608432S, RSO608433S, RSO608459S, RSO608467S, RSO609057S, RSO609059S, RSO609061S, RSO609062S, RSO609394S, RSO609399S, RSO609413S, RSO612702S, RSO613068S, RSO613183S, RSO602858S, RSO603017S, RSO603095S, RSO603797S, RSO603981S, RSO609545S, RSO601120S, RSO601776S, RSO601779S, RSO601781S, RSO602977S, RSO602978S, RSO603093S, RSO603133S, RSO603503S, RSO603764S, RSO605106S, RSO605905S, RSO608015S, RSO608016S, RSO608145S, RSO609588S, RSO609743S, RSO609744S, RSO609838S, RSO609907S, RSO610346S, RSO610348S, RSO610564S, RSO610566S, RSO610825S, RSO610826S, RSO610845S, RSO610848S, RSO611618S, RSO611619S, RSO611739S, RSO611742S, RSO611748S, RSO611986S, RSO611987S, RSO611988S, RSO611989S, RSO612023S, RSO612333S, RSO612334S, RSO612568S, RSO600094S, RSO600100S, RSO600495S, RSO601199S, RSO601206S, RSO601246S, RSO601374S, RSO601681S, RSO601777S, RSO601858S, RSO602766S, RSO603011S, RSO603246S, RSO604404S, RSO605246S, RSO605487S, RSO605493S, RSO605519S, RSO607979S, RSO608623S, RSO608624S, RSO609302S, RSO609455S, RSO609580S, RSO609720S, RSO609721S, RSO609872S, RSO609873S, RSO609876S, RSO610556S, RSO610724S, RSO610729S, RSO610730S, RSO600141S, RSO600156S, RSO600235S, RSO600244S, RSO600260S, RSO600261S, RSO600338S, RSO600608S, RSO600684S, RSO600946S, RSO601020S, RSO601048S, RSO601139S, RSO601226S, RSO601338S, RSO601339S, RSO601346S, RSO601447S, RSO601644S, RSO601762S, RSO602302S, RSO602332S, RSO602348S, RSO602355S, RSO602525S, RSO602588S, RSO602741S, RSO602760S, RSO602812S, RSO603295S, RSO603869S, RSO603895S, RSO603904S, RSO603909S, RSO603954S, RSO603968S, RSO604161S, RSO604167S, RSO604280S, RSO604443S, RSO605021S, RSO605112S, RSO605467S, RSO605566S, RSO605700S, RSO605727S, RSO605762S, RSO605874S, RSO605951S, RSO605988S, RSO606069S, RSO606094S, RSO606145S, RSO606157S, RSO606312S, RSO606525S, RSO606624S, RSO606739S, RSO606750S, RSO606814S, RSO606820S, RSO606867S, RSO606877S, RSO606925S, RSO607217S, RSO607725S, RSO607735S, RSO607899S, RSO608014S, RSO608375S, RSO608879S, RSO608880S, RSO608881S, RSO601212S, RSO610636S, RSO601718S, RSO603844S, RSO603898S, RSO606273S, RSO608652S, RSO601025S, RSO601844S, RSO605225S, RSO600460S, RSO600463S, RSO600522S, RSO600553S, RSO600835S, RSO600859S, RSO601277S, RSO604135S, RSO609614S, RSO605772S, RSO605781S, RSO605804S, RSO601027S, RSO601082S, RSO601308S, RSO601761S, RSO602838S, RSO602985S, RSO603714S, RSO604288S, RSO605109S, RSO605264S, RSO605739S, RSO606035S, RSO606036S, RSO606250S, RSO607161S, RSO600754S, RSO601189S, RSO603568S, RSO604289S, RSO600672S, RSO601675S, RSO603134S, RSO603378S, RSO604259S, RSO611767S, RSO603055S, RSO603322S, RSO603566S, RSO603624S, RSO604195S, RSO604556S, RSO605391S, RSO606606S, RSO600533S, RSO600634S, RSO600635S, RSO600810S, RSO601385S, RSO601530S, RSO602541S, RSO602779S, RSO602963S, RSO603021S, RSO603405S, RSO603587S, RSO603640S, RSO604198S, RSO604255S, RSO604446S, RSO605096S, RSO605977S, RSO609634S, RSO609640S, RSO610357S, RSO610397S, RSO610399S, RSO610563S, RSO612145S, RSO612572S, RSO607541S, RSO608172S, RSO608303S, RSO608474S, RSO608658S, RSO600181S, RSO610313S, RSO612620S, RSO612703S, RSO602509S, RSO605146S, RSO605746S, RSO605985S, RSO606593S, RSO606839S, RSO607098S, RSO607551S, RSO607800S, RSO607867S, RSO608457S, RSO604568S, RSO604360S, RSO600860S, RSO601535S, RSO602565S, RSO604230S, RSO606146S, RSO607719S, RSO600572S, RSO600573S, RSO600589S, RSO600591S, RSO600709S, RSO601069S, RSO605239S, RSO608632S, RSO608955S, RSO602907S, RSO602908S, RSO602909S, RSO602910S, RSO602911S, RSO604225S, RSO604226S, RSO604228S, RSO604430S, RSO604431S, RSO604434S, RSO604435S, RSO604436S, RSO604437S, RSO604438S, RSO604439S, RSO604440S, RSO604441S, RSO604442S, RSO604444S, RSO603184S, RSO603186S, RSO603187S, RSO603188S, RSO603459S, RSO603460S, RSO603461S, RSO603472S, RSO603580S, RSO603713S, RSO600637S, RSO604353S, RSO605400S, RSO605733S, RSO606366S, RSO605653S, RSO600272S, RSO608606S, RSO611587S, RSO611777S, RSO612066S, RSO600454S, RSO600625S, RSO600657S, RSO601431S, RSO603252S, RSO605241S, RSO605266S, RSO606311S, RSO607871S, RSO605614S, RSO606641S, RSO608717S, RSO610608S, RSO602872S, RSO603778S, RSO604058S, RSO604254S, RSO609548S, RSO609641S, RSO609642S, RSO609836S, RSO609882S, RSO610321S, RSO610379S, RSO610383S, RSO610480S, RSO610896S, RSO610939S, RSO611675S, RSO611736S, RSO612346S, RSO600545S, RSO601581S, RSO601603S, RSO603541S, RSO604085S, RSO604414S, RSO605188S, RSO605190S, RSO605399S, RSO605465S, RSO605671S, RSO605684S, RSO605825S, RSO606027S, RSO606218S, RSO606334S, RSO606335S, RSO606367S, RSO606511S, RSO607049S, RSO607054S, RSO607124S, RSO607130S, RSO607689S, RSO607745S, RSO607968S, RSO607971S, RSO607983S, RSO607989S, RSO608193S, RSO608329S, RSO608330S, RSO608331S, RSO608929S, RSO608992S, RSO610895S, RSO609589S, RSO610841S, RSO610871S, RSO611751S, RSO611752S, RSO612019S, RSO612055S, RSO612213S, RSO612268S, RSO612531S, RSO612681S, RSO604687S, RSO605125S, RSO605343S, RSO605840S, RSO606964S, RSO606974S, RSO607251S, RSO607393S, RSO607589S, RSO607767S, RSO607768S, RSO608055S, RSO608056S, RSO608296S, RSO608299S, RSO608302S, RSO612510S, RSO603035S, RSO605411S, RSO605839S, RSO601015S, RSO606343S, RSO609600S, RSO609601S, RSO609602S, RSO609643S, RSO609644S, RSO609645S, RSO610782S, RSO611798S, RSO600447S, RSO600747S, RSO600923S, RSO601185S, RSO601191S, RSO601238S, RSO601273S, RSO601439S, RSO601578S, RSO601683S, RSO601711S, RSO601906S, RSO601922S, RSO602920S, RSO602988S, RSO603215S, RSO603289S, RSO603431S, RSO603497S, RSO603795S, RSO604082S, RSO604088S, RSO604246S, RSO604613S, RSO604829S, RSO605129S, RSO605196S, RSO605549S, RSO605632S, RSO605668S, RSO605704S, RSO605705S, RSO606044S, RSO606046S, RSO606192S, RSO606361S, RSO606385S, RSO606521S, RSO606967S, RSO607342S, RSO607343S, RSO607576S, RSO607581S, RSO607643S, RSO607644S, RSO607924S, RSO608253S, RSO608522S, RSO608523S, RSO608801S, RSO608802S, RSO608945S, RSO608946S, RSO609260S, RSO605153S, RSO608074S, RSO609499S, RSO610753S, RSO610754S, RSO610790S, RSO612005S, RSO612039S, RSO612222S, RSO604963S, RSO605026S, RSO605027S, RSO605167S, RSO605168S, RSO605170S, RSO605627S, RSO606056S, RSO606657S, RSO606864S, RSO606919S, RSO606920S, RSO606922S, RSO607103S, RSO607892S, RSO608067S, RSO608175S, RSO608176S, RSO608181S, RSO608189S, RSO612115S, RSO612163S, RSO612557S, RSO613150S, RSO610579S, RSO610580S, RSO610610S, RSO610611S, RSO610956S, RSO610957S, RSO611591S, RSO611594S, RSO611761S, RSO611886S, RSO600982S, RSO601449S, RSO602576S, RSO603508S, RSO603983S, RSO604297S, RSO604968S, RSO605990S, RSO606324S, RSO606337S, RSO606510S, RSO607159S, RSO607956S, RSO610337S, RSO604920S, RSO605752S, RSO608797S, RSO612771S, RSO609493S, RSO609572S, RSO609659S, RSO609663S, RSO609815S, RSO609864S, RSO609868S, RSO610316S, RSO610324S, RSO610327S, RSO610328S, RSO610454S, RSO610555S, RSO610632S, RSO610633S, RSO610698S, RSO610807S, RSO610808S, RSO610811S, RSO610815S, RSO610816S, RSO611509S, RSO611512S, RSO611578S, RSO611579S, RSO611581S, RSO611583S, RSO611745S, RSO611746S, RSO611996S, RSO611997S, RSO611999S, RSO612001S, RSO612004S, RSO612093S, RSO612098S, RSO612153S, RSO612226S, RSO612369S, RSO612543S, RSO612611S, RSO612612S, RSO612613S, RSO612614S, RSO612615S, RSO612707S, RSO612708S, RSO600087S, RSO600088S, RSO600089S, RSO600857S, RSO600974S, RSO600976S, RSO600977S, RSO601090S, RSO601314S, RSO601342S, RSO601621S, RSO601641S, RSO602314S, RSO602328S, RSO602547S, RSO602575S, RSO602785S, RSO603214S, RSO603272S, RSO603593S, RSO603671S, RSO603787S, RSO604057S, RSO604133S, RSO604533S, RSO605080S, RSO605148S, RSO605156S, RSO605203S, RSO605204S, RSO605272S, RSO605310S, RSO605327S, RSO605378S, RSO605580S, RSO605695S, RSO605923S, RSO605929S, RSO605967S, RSO606001S, RSO606049S, RSO606057S, RSO606226S, RSO606240S, RSO606629S, RSO606680S, RSO606917S, RSO607113S, RSO607245S, RSO607369S, RSO607630S, RSO607944S, RSO608030S, RSO608574S, RSO608575S, RSO608577S, RSO608578S, RSO608627S, RSO608731S, RSO608732S, RSO608763S, RSO608918S, RSO608957S, RSO608958S, RSO608959S, RSO608961S, RSO608962S, RSO612610S, RSO611720S, RSO600388S, RSO600450S, RSO600713S, RSO601059S, RSO601171S, RSO601448S, RSO603223S, RSO604471S, RSO604517S, RSO605375S, RSO601485S, RSO601606S, RSO602451S, RSO610783S, RSO611482S, RSO602926S, RSO603750S, RSO609703S, RSO609948S, RSO610604S, RSO610766S, RSO610946S, RSO610948S, RSO610949S, RSO610963S, RSO611398S, RSO611770S, RSO601508S, RSO603107S, RSO603114S, RSO603116S, RSO603484S, RSO604623S, RSO605918S, RSO606300S, RSO606444S, RSO607368S, RSO608288S, RSO608680S, RSO600428S, RSO601259S, RSO612589S, RSO606913S, RSO608890S, RSO612719S, RSO610541S, RSO600556S, RSO600596S, RSO600753S, RSO600961S, RSO601177S, RSO601386S, RSO601387S, RSO601391S, RSO602700S, RSO604257S, RSO604361S, RSO605769S, RSO607182S, RSO607483S, RSO608297S, RSO608454S, RSO600111S, RSO600112S, RSO600146S, RSO601049S, RSO602557S, RSO606122S, RSO606941S, RSO607663S, RSO607733S, RSO607750S, RSO607795S, RSO605664S, RSO609529S, RSO609530S, RSO609686S, RSO609761S, RSO610528S, RSO610573S, RSO610972S, RSO611330S, RSO611636S, RSO601407S, RSO601834S, RSO603993S, RSO604307S, RSO604308S, RSO604329S, RSO605149S, RSO605610S, RSO605631S, RSO606153S, RSO606338S, RSO606951S, RSO607169S, RSO607204S, RSO607998S, RSO607999S, RSO608313S, RSO608647S, RSO608700S, RSO608724S, RSO608725S, RSO608726S, RSO608814S, RSO608892S, RSO612545S, RSO600919S, RSO603557S, RSO606512S, RSO608743S, RSO605817S, RSO605940S, RSO607700S, RSO609592S, RSO609593S, RSO609648S, RSO609649S, RSO609653S, RSO610360S, RSO610406S, RSO610407S, RSO610569S, RSO610570S, RSO610621S, RSO610622S, RSO610623S, RSO610852S, RSO610855S, RSO610858S, RSO611489S, RSO611506S, RSO611623S, RSO611625S, RSO611755S, RSO611756S, RSO612014S, RSO612015S, RSO612044S, RSO612059S, RSO612060S, RSO612177S, RSO612258S, RSO612260S, RSO612529S, RSO612530S, RSO612585S, RSO612586S, RSO612683S, RSO612689S, RSO601262S, RSO601267S, RSO601403S, RSO603048S, RSO603313S, RSO603314S, RSO603531S, RSO603564S, RSO603577S, RSO603814S, RSO603879S, RSO604301S, RSO604302S, RSO604685S, RSO604904S, RSO605265S, RSO605267S, RSO605687S, RSO605689S, RSO605864S, RSO605939S, RSO606197S, RSO606242S, RSO606246S, RSO606254S, RSO606473S, RSO606551S, RSO606645S, RSO606648S, RSO606885S, RSO606959S, RSO606982S, RSO607000S, RSO607135S, RSO607370S, RSO607440S, RSO607536S, RSO607562S, RSO607674S, RSO607679S, RSO607789S, RSO607981S, RSO608059S, RSO608062S, RSO608178S, RSO608179S, RSO608200S, RSO608259S, RSO608479S, RSO608520S, RSO608570S, RSO608648S, RSO608708S, RSO608718S, RSO608800S, RSO608804S, RSO608805S, RSO608943S, RSO608944S, RSO609268S, RSO609283S, RSO609285S, RSO609286S, RSO609288S, RSO609305S, RSO609427S, RSO609430S, RSO609434S, RSO612579S, RSO609724S, RSO612603S, RSO600212S, RSO607667S, RSO609748S, RSO600211S, RSO600218S, RSO600258S, RSO600858S, RSO600962S, RSO600967S, RSO601105S, RSO601369S, RSO601632S, RSO601713S, RSO602581S, RSO602826S, RSO603403S, RSO603441S, RSO603676S, RSO605144S, RSO603088S, RSO610499S, RSO611389S, RSO611486S, RSO612054S, RSO612142S, RSO612577S, RSO604467S, RSO604468S, RSO604794S, RSO604805S, RSO605590S, RSO605948S, RSO605949S, RSO606227S, RSO606484S, RSO606614S, RSO606828S, RSO606849S, RSO606851S, RSO606868S, RSO606889S, RSO606976S, RSO607397S, RSO607550S, RSO607805S, RSO607807S, RSO606330S, RSO606556S, RSO612894S, RSO612895S, RSO609711S, RSO612043S, RSO608914S, RSO600527S, RSO600555S, RSO600878S, RSO601151S, RSO601297S, RSO601345S, RSO602562S, RSO602729S, RSO603200S, RSO603344S, RSO604132S, RSO604474S, RSO606187S, RSO607661S, RSO608521S, RSO612162S, RSO603395S, RSO612373S, RSO608233S, RSO608764S, RSO612036S, RSO610876S, RSO608082S, RSO609237S, RSO612337S, RSO612593S, RSO600124S, RSO605532S, RSO607073S, RSO609290S, RSO612029S, RSO602418S, RSO603401S, RSO603648S, RSO603868S, RSO605019S, RSO605736S, RSO605858S, RSO608197S, RSO600836S, RSO602303S, RSO605744S, RSO600176S, RSO600978S, RSO601363S, RSO601364S, RSO601648S, RSO601650S, RSO602904S, RSO603834S, RSO603935S, RSO605312S, RSO605612S, RSO605938S, RSO606160S, RSO606362S, RSO606778S, RSO606915S, RSO608072S, RSO608482S, RSO611710S, RSO611732S, RSO602849S, RSO603131S, RSO607057S, RSO601569S, RSO601686S, RSO603891S, RSO609533S, RSO609604S, RSO610431S, RSO610647S, RSO601552S, RSO603947S, RSO604048S, RSO604306S, RSO605974S, RSO607052S, RSO607058S, RSO610529S, RSO610910S, RSO610932S, RSO612377S, RSO611789S, RSO612700S, RSO604047S, RSO605756S, RSO606735S, RSO606957S, RSO612659S, RSO601659S, RSO608609S, RSO610505S, RSO611500S, RSO612218S, RSO601117S, RSO604204S, RSO608198S, RSO601113S, RSO607743S, RSO610540S, RSO600789S, RSO602970S, RSO603400S, RSO603499S, RSO606316S, RSO604051S, RSO610403S, RSO612053S, RSO612335S, RSO605775S, RSO600616S, RSO609863S, RSO610325S, RSO603174S, RSO607651S, RSO608576S, RSO608628S, RSO608234S, RSO609685S, RSO610374S, RSO610796S, RSO607666S, RSO603534S, RSO605143S, RSO600538S, RSO601669S, RSO605469S, RSO606022S, RSO607916S, RSO606110S, RSO606111S, RSO606116S, RSO608462S, RSO608463S, RSO608464S, RSO608465S, RSO609503S, RSO609574S, RSO609575S, RSO609672S, RSO609871S, RSO610526S, RSO601146S, RSO601848S, RSO602585S, RSO602938S, RSO604347S, RSO604368S, RSO604369S, RSO604371S, RSO604372S, RSO604373S, RSO604449S, RSO605141S, RSO605154S, RSO605277S, RSO605278S, RSO605320S, RSO605408S, RSO605449S, RSO605484S, RSO605497S, RSO605576S, RSO605577S, RSO605578S, RSO605660S, RSO605754S, RSO605868S, RSO605926S, RSO605978S, RSO606014S, RSO606015S, RSO606016S, RSO606018S, RSO606021S, RSO606249S, RSO606420S, RSO606421S, RSO606943S, RSO607020S, RSO607067S, RSO607114S, RSO607166S, RSO607186S, RSO607189S, RSO607193S, RSO607297S, RSO607303S, RSO607353S, RSO607375S, RSO607731S, RSO607770S, RSO607788S, RSO607857S, RSO607865S, RSO607866S, RSO607870S, RSO607915S, RSO607990S, RSO608066S, RSO608068S, RSO608231S, RSO608357S, RSO608485S, RSO608554S, RSO608555S, RSO608651S, RSO608722S, RSO608723S, RSO608756S, RSO608757S, RSO608964S, RSO609010S, RSO609263S, RSO606024S, RSO610907S, RSO611553S, RSO611648S, RSO611659S, RSO612231S, RSO612340S, RSO612658S, RSO604049S, RSO608325S, RSO608347S, RSO608739S, RSO600700S, RSO601215S, RSO601503S, RSO603154S, RSO603156S, RSO603158S, RSO604846S, RSO605068S, RSO605570S, RSO606090S, RSO606279S, RSO608912S, RSO610567S, RSO609751S, RSO600172S, RSO605609S, RSO605969S, RSO607019S, RSO607777S, RSO608608S, RSO609577S, RSO609621S, RSO609728S, RSO609892S, RSO609894S, RSO609895S, RSO610354S, RSO610715S, RSO610717S, RSO612183S, RSO612245S, RSO612495S, RSO612565S, RSO612667S, RSO612668S, RSO601154S, RSO602745S, RSO602965S, RSO603274S, RSO603283S, RSO603294S, RSO603298S, RSO603436S, RSO603527S, RSO603611S, RSO603622S, RSO603726S, RSO603748S, RSO603756S, RSO603782S, RSO603991S, RSO604014S, RSO604017S, RSO604123S, RSO604359S, RSO604450S, RSO605008S, RSO605227S, RSO607872S, RSO608538S, RSO608540S, RSO608666S, RSO608784S, RSO608889S, RSO609298S, RSO610421S, RSO610731S, RSO612227S, RSO607909S, RSO607948S, RSO611825S, RSO611826S, RSO612120S, RSO612117S, RSO612118S, RSO612119S, RSO612122S, RSO613248S, RSO613249S, RSO600795S, RSO602330S, RSO604016S, RSO605402S, RSO611576S, RSO612558S, RSO608488S, RSO604067S, RSO605745S, RSO605861S, RSO607761S, RSO610757S, RSO610906S, RSO611711S, RSO601841S, RSO603705S, RSO609762S, RSO610631S, RSO612086S, RSO612371S, RSO601203S, RSO601401S, RSO602571S, RSO602862S, RSO603052S, RSO603791S, RSO603852S, RSO605369S, RSO605649S, RSO606062S, RSO606562S, RSO609553S, RSO609559S, RSO609660S, RSO609661S, RSO609664S, RSO609700S, RSO610947S, RSO611600S, RSO611781S, RSO612389S, RSO600996S, RSO601334S, RSO601586S, RSO601894S, RSO602479S, RSO602630S, RSO603096S, RSO603249S, RSO603282S, RSO603626S, RSO603739S, RSO604247S, RSO604901S, RSO605201S, RSO605207S, RSO605510S, RSO605914S, RSO606054S, RSO606340S, RSO606428S, RSO606532S, RSO606566S, RSO606647S, RSO607046S, RSO607255S, RSO607371S, RSO607418S, RSO607606S, RSO607813S, RSO607880S, RSO607881S, RSO608005S, RSO608006S, RSO608679S, RSO608742S, RSO608843S, RSO604143S, RSO605535S, RSO611678S, RSO609225S, RSO609854S, RSO610547S, RSO600205S, RSO600530S, RSO600531S, RSO600549S, RSO600631S, RSO600668S, RSO600691S, RSO600762S, RSO600763S, RSO600765S, RSO600766S, RSO601352S, RSO601372S, RSO601389S, RSO601573S, RSO601751S, RSO601783S, RSO601809S, RSO603006S, RSO603316S, RSO603712S, RSO604069S, RSO604154S, RSO604869S, RSO605555S, RSO605706S, RSO606341S, RSO607698S, RSO607699S, RSO608856S, RSO602520S, RSO603864S, RSO600682S, RSO601229S, RSO601288S, RSO601833S, RSO605334S, RSO600171S, RSO600215S, RSO601100S, RSO601422S, RSO601531S, RSO601580S, RSO602526S, RSO602529S, RSO602701S, RSO602747S, RSO602808S, RSO602870S, RSO602945S, RSO603072S, RSO603074S, RSO603079S, RSO603089S, RSO603164S, RSO603201S, RSO603317S, RSO603343S, RSO603418S, RSO603545S, RSO603546S, RSO603635S, RSO603915S, RSO603919S, RSO603978S, RSO604216S, RSO604268S, RSO604269S, RSO604621S, RSO605003S, RSO605930S, RSO606214S, RSO606232S, RSO608205S, RSO603042S, RSO603043S, RSO603044S, RSO603045S, RSO603046S, RSO603049S, RSO603050S, RSO603053S, RSO603054S, RSO603543S, RSO603544S, RSO603547S, RSO603548S, RSO603549S, RSO603550S, RSO603551S, RSO604166S, RSO604168S, RSO604169S, RSO604170S, RSO604171S, RSO604173S, RSO604175S, RSO604178S, RSO604180S, RSO604182S, RSO604184S, RSO604185S, RSO604186S, RSO604187S, RSO604189S, RSO604190S, RSO604191S, RSO604192S, RSO607582S, RSO607583S, RSO607584S, RSO607585S, RSO607586S, RSO611716S, RSO611717S, RSO611990S, RSO601219S, RSO601220S, RSO601801S, RSO602339S, RSO602850S, RSO604529S, RSO605724S, RSO606386S, RSO606392S, RSO606393S, RSO606394S, RSO606497S, RSO606527S, RSO606547S, RSO606553S, RSO606960S, RSO606978S, RSO607095S, RSO607099S, RSO607289S, RSO607319S, RSO607321S, RSO607717S, RSO607923S, RSO611632S, RSO611634S, RSO612116S, RSO609671S, RSO610319S, RSO610341S, RSO610455S, RSO610638S, RSO610941S, RSO600091S, RSO600248S, RSO600253S, RSO601135S, RSO601625S, RSO601646S, RSO602358S, RSO602514S, RSO602781S, RSO603143S, RSO603179S, RSO603357S, RSO604052S, RSO604415S, RSO605537S, RSO605950S, RSO606104S, RSO606592S, RSO606999S, RSO607692S, RSO608882S, RSO609449S, RSO610722S, RSO602350S, RSO603015S, RSO603480S, RSO603481S, RSO600208S, RSO600246S, RSO601014S, RSO601336S, RSO602961S, RSO603211S, RSO603893S, RSO605307S, RSO608787S, RSO612926S, RSO611700S, RSO603817S, RSO606724S, RSO611567S, RSO611568S, RSO611701S, RSO612247S, RSO612248S, RSO612303S, RSO610574S, RSO609753S, RSO610882S, RSO610883S, RSO601313S, RSO606902S, RSO610088S, RSO610866S, RSO611667S, RSO600699S, RSO602839S, RSO604333S, RSO608956S, RSO601301S, RSO609628S, RSO609931S, RSO609933S, RSO609934S, RSO610543S, RSO610572S, RSO610652S, RSO610745S, RSO610746S, RSO610759S, RSO610985S, RSO611028S, RSO611380S, RSO611665S, RSO611796S, RSO612035S, RSO612090S, RSO612161S, RSO612182S, RSO612284S, RSO612347S, RSO612384S, RSO612563S, RSO612564S, RSO612643S, RSO600093S, RSO600153S, RSO600214S, RSO600611S, RSO600661S, RSO600708S, RSO600758S, RSO600846S, RSO601024S, RSO601026S, RSO601153S, RSO601355S, RSO601605S, RSO601619S, RSO601634S, RSO601638S, RSO601724S, RSO601869S, RSO602344S, RSO602905S, RSO602906S, RSO602933S, RSO602964S, RSO602975S, RSO603090S, RSO603141S, RSO603661S, RSO603862S, RSO603970S, RSO604149S, RSO605067S, RSO605180S, RSO605181S, RSO605182S, RSO605183S, RSO605293S, RSO605646S, RSO606786S, RSO606802S, RSO606900S, RSO607039S, RSO607168S, RSO607265S, RSO607647S, RSO608045S, RSO608046S, RSO608811S, RSO608855S, RSO608968S, RSO613192S, RSO611885S, RSO610655S, RSO605747S, RSO607532S, RSO608794S, RSO611812S, RSO611814S, RSO610329S, RSO612255S, RSO612256S, RSO602641S, RSO610481S, RSO601566S, RSO604063S, RSO607723S, RSO603641S, RSO606873S, RSO608000S, RSO608450S, RSO611607S, RSO602490S, RSO602722S, RSO602782S, RSO602874S, RSO603305S, RSO603979S, RSO604378S, RSO605088S, RSO605904S, RSO606499S, RSO606961S, RSO606991S, RSO607555S, RSO601283S, RSO603618S, RSO605561S, RSO605670S, RSO605803S, RSO605947S, RSO606173S, RSO606238S, RSO606537S, RSO606677S, RSO606994S, RSO607432S, RSO607638S, RSO608563S, RSO602893S, RSO610642S, RSO600647S, RSO606286S, RSO609828S, RSO601492S, RSO602927S, RSO602951S, RSO603768S, RSO607937S, RSO607332S, RSO600202S, RSO605553S, RSO606293S, RSO606610S, RSO607133S, RSO607859S, RSO612596S, RSO612599S, RSO612601S, RSO609737S, RSO610508S, RSO612113S, RSO612244S, RSO600013S, RSO600204S, RSO600268S, RSO600273S, RSO601197S, RSO601549S, RSO607894S, RSO608789S, RSO609297S, RSO604295S, RSO604531S, RSO604548S, RSO609489S, RSO610318S, RSO610520S, RSO611601S, RSO602550S, RSO605318S, RSO607758S, RSO608687S, RSO612077S, RSO612186S, RSO612357S, RSO612686S, RSO605565S, RSO607895S, RSO608532S, RSO608534S, RSO609526S, RSO609573S, RSO610332S, RSO610699S, RSO611280S, RSO612137S, RSO612212S, RSO600636S, RSO600966S, RSO601319S, RSO601453S, RSO601885S, RSO601886S, RSO601892S, RSO601893S, RSO604242S, RSO604379S, RSO605041S, RSO605185S, RSO605511S, RSO605515S, RSO605963S, RSO605971S, RSO606313S, RSO606467S, RSO606548S, RSO606972S, RSO607263S, RSO607264S, RSO607528S, RSO607529S, RSO607675S, RSO607690S, RSO607796S, RSO607873S, RSO607885S, RSO608034S, RSO608150S, RSO608235S, RSO608264S, RSO608770S, RSO608788S, RSO608967S, RSO601367S, RSO602432S, RSO604820S, RSO604827S, RSO609556S, RSO609560S, RSO611768S, RSO612629S, RSO612630S, RSO601518S, RSO601732S, RSO601733S, RSO601743S, RSO601744S, RSO601746S, RSO601747S, RSO603596S, RSO603693S, RSO603783S, RSO603843S, RSO603906S, RSO603907S, RSO604515S, RSO604544S, RSO604545S, RSO604602S, RSO604890S, RSO605396S, RSO606059S, RSO607091S, RSO607138S, RSO607791S, RSO607996S, RSO608871S, RSO608872S, RSO608905S, RSO600195S, RSO601111S, RSO601152S, RSO602981S, RSO603075S, RSO600438S, RSO600969S, RSO609702S, RSO600078S, RSO601311S, RSO609606S, RSO609607S, RSO609678S, RSO609911S, RSO600161S, RSO600318S, RSO601051S, RSO601281S, RSO601656S, RSO603789S, RSO605111S, RSO605114S, RSO605275S, RSO605533S, RSO605608S, RSO605620S, RSO605720S, RSO605882S, RSO607053S, RSO607780S, RSO607985S, RSO608348S, RSO608813S, RSO609326S, RSO601749S, RSO602629S, RSO602848S, RSO603030S, RSO603066S, RSO603171S, RSO603794S, RSO606550S, RSO606576S, RSO609146S, RSO607115S, RSO608049S, RSO612320S, RSO610339S, RSO610340S, RSO610639S, RSO610728S, RSO610734S, RSO600254S, RSO600259S, RSO601208S, RSO601418S, RSO601622S, RSO601673S, RSO601912S, RSO602710S, RSO603913S, RSO604030S, RSO604129S, RSO604293S, RSO605652S, RSO605776S, RSO605782S, RSO606075S, RSO606632S, RSO606835S, RSO606923S, RSO607086S, RSO607524S, RSO607882S, RSO607898S, RSO608013S, RSO608686S, RSO610747S, RSO603853S, RSO612238S, RSO603190S, RSO604547S, RSO608691S, RSO603128S, RSO605584S, RSO605871S, RSO606649S, RSO606921S, RSO608058S, RSO608060S, RSO608061S, RSO610388S, RSO603319S, RSO603752S, RSO606436S, RSO601720S, RSO602847S, RSO602578S, RSO605150S, RSO607613S, RSO610393S, RSO610479S, RSO611520S, RSO611714S, RSO611719S, RSO612279S, RSO612359S, RSO612665S, RSO601539S, RSO601694S, RSO603296S, RSO603411S, RSO603413S, RSO603504S, RSO605078S, RSO605407S, RSO606371S, RSO606422S, RSO606498S, RSO606564S, RSO606582S, RSO607634S, RSO607746S, RSO608530S, RSO608928S, RSO609369S, RSO612199S, RSO612317S, RSO612372S, RSO612503S, RSO607383S, RSO607384S, RSO608611S, RSO612507S, RSO612524S, RSO612618S, RSO600376S, RSO600809S, RSO601073S, RSO601237S, RSO602389S, RSO602947S, RSO603228S, RSO603589S, RSO604020S, RSO604248S, RSO605386S, RSO607345S, RSO608693S, RSO609920S, RSO612136S, RSO600508S, RSO600705S, RSO601099S, RSO601766S, RSO601863S, RSO602824S, RSO602983S, RSO603145S, RSO603195S, RSO603261S, RSO603958S, RSO604469S, RSO604470S, RSO604478S, RSO605541S, RSO605908S, RSO606504S, RSO607318S, RSO607460S, RSO608915S, RSO608935S, RSO609913S, RSO610404S, RSO610459S, RSO610460S, RSO610620S, RSO610853S, RSO610856S, RSO611754S, RSO612018S, RSO612056S, RSO612336S, RSO612501S, RSO601651S, RSO602593S, RSO603792S, RSO603951S, RSO604159S, RSO604272S, RSO605351S, RSO605710S, RSO607427S, RSO608618S, RSO608941S, RSO609438S, RSO609444S, RSO601877S, RSO611757S, RSO611758S, RSO611760S, RSO612171S, RSO600587S, RSO601537S, RSO601538S, RSO601559S, RSO601819S, RSO601820S, RSO602614S, RSO603209S, RSO603359S, RSO603372S, RSO603449S, RSO603615S, RSO603899S, RSO604525S, RSO604553S, RSO605054S, RSO605065S, RSO605077S, RSO605126S, RSO605134S, RSO605245S, RSO605279S, RSO605636S, RSO605648S, RSO605818S, RSO606165S, RSO606285S, RSO606317S, RSO606318S, RSO606342S, RSO606400S, RSO607889S, RSO608552S, RSO609005S, RSO609480S, RSO609587S, RSO609811S, RSO610349S, RSO610395S, RSO610847S, RSO612024S, RSO612025S, RSO612173S, RSO600601S, RSO600603S, RSO600802S, RSO600817S, RSO601376S, RSO601437S, RSO601458S, RSO601572S, RSO601607S, RSO601854S, RSO601876S, RSO602765S, RSO602922S, RSO603177S, RSO603239S, RSO603376S, RSO603430S, RSO603515S, RSO603636S, RSO603989S, RSO604074S, RSO605595S, RSO606456S, RSO607668S, RSO607714S, RSO607908S, RSO608052S, RSO608564S, RSO609741S, RSO608171S, RSO609742S, RSO610495S, RSO610616S, RSO610617S, RSO611740S, RSO611741S, RSO612020S, RSO612048S, RSO612147S, RSO600602S, RSO600604S, RSO601087S, RSO601147S, RSO601360S, RSO601521S, RSO601864S, RSO602778S, RSO602987S, RSO603115S, RSO603181S, RSO603516S, RSO603998S, RSO604064S, RSO604065S, RSO605095S, RSO605165S, RSO606050S, RSO606088S, RSO606591S, RSO606697S, RSO606989S, RSO607176S, RSO607421S, RSO607429S, RSO607527S, RSO607553S, RSO608050S, RSO608565S, RSO608602S, RSO608621S, RSO608715S, RSO608774S, RSO609301S, RSO609454S, RSO600850S, RSO605155S, RSO609733S, RSO609820S, RSO610344S, RSO600366S, RSO601040S, RSO603972S, RSO604522S, RSO604523S, RSO604530S, RSO605430S, RSO605574S, RSO605581S, RSO605582S, RSO606626S, RSO606898S, RSO607771S, RSO607878S, RSO607879S, RSO607928S, RSO607930S, RSO600102S, RSO600104S, RSO600118S, RSO600398S, RSO600800S, RSO600909S, RSO600972S, RSO601357S, RSO601450S, RSO601598S, RSO601616S, RSO601793S, RSO602454S, RSO602507S, RSO602582S, RSO602639S, RSO603491S, RSO603977S, RSO603992S, RSO604031S, RSO604045S, RSO604294S, RSO604317S, RSO604319S, RSO605151S, RSO605401S, RSO605942S, RSO605944S, RSO606968S, RSO600099S, RSO600101S, RSO600236S, RSO600681S, RSO600746S, RSO600947S, RSO600987S, RSO600998S, RSO601247S, RSO601328S, RSO601368S, RSO601371S, RSO601536S, RSO601611S, RSO601613S, RSO601629S, RSO601653S, RSO601654S, RSO601916S, RSO602300S, RSO602786S, RSO602827S, RSO603120S, RSO603345S, RSO603845S, RSO603882S, RSO604183S, RSO605447S, RSO605562S, RSO605618S, RSO605629S, RSO605755S, RSO605824S, RSO609523S, RSO609750S, RSO609824S, RSO609825S, RSO609885S, RSO609887S, RSO610345S, RSO610461S, RSO610509S, RSO610703S, RSO611508S, RSO611566S, RSO611679S, RSO611699S, RSO612034S, RSO612134S, RSO612210S, RSO612211S, RSO612559S, RSO601270S, RSO601873S, RSO603137S, RSO603349S, RSO603606S, RSO603607S, RSO604193S, RSO604313S, RSO604314S, RSO605299S, RSO605583S, RSO606131S, RSO606594S, RSO606973S, RSO607317S, RSO607920S, RSO608354S, RSO608392S, RSO608484S, RSO609299S, RSO613197S, RSO610878S, RTP602711S, RSO607219S, RSO607301S, RSO607309S, RSO607315S, RSO607326S, RSO607327S, RSO607329S, RSO607330S, RSO607573S, RSO607645S, RSO607658S, RSO607659S, RSO607850S, RSO608040S, RSO608092S, RSO608097S, RSO608103S, RSO608105S, RSO608106S, RSO608123S, RSO608227S, RSO608275S, RSO608276S, RSO608336S, RSO608337S, RSO608338S, RSO608339S, RSO608352S, RSO608356S, RSO608362S, RSO608367S, RSO608368S, RSO608369S, RSO608409S, RSO608440S, RSO608443S, RSO608444S, RSO608455S, RSO608458S, RSO608546S, RSO610069S, RSO610070S, RSO610071S, RSO610072S, RSO610073S, RSO610074S, RSO610075S, RSO610077S, RSO610078S, RSO610079S, RSO610578S, RSO611823S, RSO611824S, RSO612555S, RSO612556S, RSO612713S, RSO612714S, RSO612715S, RSO601460S, RSO601461S, RSO601462S, RSO601463S, RSO601464S, RSO601465S, RSO601466S, RSO601467S, RSO601468S, RSO601469S, RSO604300S, RSO604320S, RSO604322S, RSO604324S, RSO604366S, RSO604370S, RSO605452S, RSO605472S, RSO605474S, RSO605542S, RSO605543S, RSO605545S, RSO605547S, RSO605548S, RSO605552S, RSO605554S, RSO607388S, RSO607399S, RSO607400S, RSO607401S, RSO608315S, RSO608361S, RSO608363S, RSO608364S, RSO608365S, RSO608371S, RSO608376S, RSO608382S, RSO608385S, RSO608389S, RSO608585S, RSO608586S, RSO608587S, RSO608588S, RSO608589S, RSO608590S, RSO608591S, RSO608592S, RSO609064S, RSO609068S, RSO609070S, RSO609076S, RSO609080S, RSO609084S, RSO609085S, RSO612716S, RSO612717S, RSO612718S, RSO612724S, RSO612725S, RSO612738S, RSO612739S, RSO612741S, RSO612746S, RSO612755S, RSO612760S, RSO612761S, RSO612762S, RSO612772S, RSO612773S, RSO612779S, RSO612787S, RSO612846S, RSO612878S, RSO612880S, RSO612885S, RSO612886S, RSO612891S, RSO612893S, RSO612908S, RSO612935S, RSO612955S, RSO612970S, RSO613132S, RSO613133S, RSO613151S, RSO613152S, RSO613153S, RSO609808S, RSO609846S, RSO610648S, RSO610886S, RSO610889S, RSO610890S, RSO611668S, RSO611669S, RSO611809S, RSO612285S, RSO612326S, RSO612514S, RSO612515S, RSO600092S, RSO600238S, RSO600240S, RSO600241S, RSO600242S, RSO600305S, RSO600806S, RSO600938S, RSO601030S, RSO601384S, RSO601436S, RSO602329S, RSO603276S, RSO603830S, RSO604134S, RSO604199S, RSO604201S, RSO604211S, RSO604779S, RSO605479S, RSO605569S, RSO606031S, RSO606034S, RSO606148S, RSO606152S, RSO606789S, RSO606950S, RSO606996S, RSO607139S, RSO607243S, RSO607395S, RSO607398S, RSO607404S, RSO607742S, RSO607747S, RSO612720S, RSO612721S, RSO612722S, RSO612723S, RSO612752S, RSO612758S, RSO612903S, RSO612904S, RSO612909S, RSO612910S, RSO612911S, RSO612912S, RSO612914S, RSO612915S, RSO612918S, RSO612919S, RSO612920S, RSO612921S, RSO612922S, RSO612923S, RSO612924S, RSO612930S, RSO612931S, RSO612932S, RSO612933S, RSO612937S, RSO612938S, RSO612939S, RSO612940S, RSO612942S, RSO612943S, RSO612945S, RSO612948S, RSO612949S, RSO612953S, RSO612954S, RSO612956S, RSO612958S, RSO612965S, RSO612975S, RSO613081S, RSO613082S, RSO613083S, RSO613084S, RSO613085S, RSO613087S, RSO613089S, RSO613091S, RSO613104S, RSO613127S, RSO613129S, RSO613130S, RSO613135S, RSO613136S, RSO613139S, RSO613140S, RSO613145S, RSO613147S, RSO613157S, RSO613159S, RSO613160S, RSO613163S, RSO613176S, RSO613185S, RSO613190S, RSO613191S, RSO613193S, RSO613198S, RSO613200S, RSO613201S, RSO613205S, RSO613206S, RSO613208S, RSO613214S, RSO613137S, RSO613138S, RSO613262S, RSO613263S, RSO613264S, RSO610918S, RSO606507S, RSO600995S, RSO600999S, RSO601738S, RSO606530S, RSO608900S, RSO610678S, RSO611354S, RSO612063S, RSO612189S, RSO602921S, RSO607969S, RSO608020S, RSO608118S, RSO608659S, RSO611554S, RSO601515S, RSO601740S, RSO602971S, RSO613247S, RSO600138S, RSO603150S, RSO608247S, RSO609829S, RSO611552S, RSO612504S, RSO600834S, RSO601857S, RSO603026S, RSO608493S, RSO605117S, RSO607554S, RSO600134S, RSO600135S, RSO600869S, RSO601029S, RSO601110S, RSO610583S, RSO610827S, RSO610828S, RSO610833S, RSO611075S, RSO611138S, RSO611794S, RSO612065S, RSO612261S, RSO612263S, RSO612535S, RSO603495S, RSO606493S, RSO612696S, RSO606691S, RSO607970S, RSO609845S, RSO610371S, RSO610877S, RSO612072S, RSO612367S, RSO603191S, RSO603769S, RSO604991S, RSO608115S, RSO608692S, RSO609666S, RSO609906S, RSO610400S, RSO610843S, RSO612028S, RSO612512S, RSO612533S, RSO612570S, RSO600927S, RSO601704S, RSO603443S, RSO603621S, RSO603781S, RSO604346S, RSO605060S, RSO605069S, RSO605215S, RSO605257S, RSO606132S, RSO606396S, RSO607542S, RSO607765S, RSO607799S, RSO607951S, RSO608054S, RSO608170S, RSO608304S, RSO608516S, RSO608600S, RSO608714S, RSO608772S, RSO608949S, RSO608950S, RSO609303S, RSO610704S, RSO611507S, RSO612301S, RSO607306S, RSO608390S, RSO600730S, RSO602754S, RSO602950S, RSO606089S, RSO607535S, RSO607804S, RSO609554S, RSO605421S, RSO608287S, RSO609707S, RSO603292S, RSO603307S, RSO603410S, RSO603447S, RSO603533S, RSO603569S, RSO604007S, RSO604876S, RSO604993S, RSO605644S, RSO605855S, RSO606168S, RSO607069S, RSO605389S, RSO608007S, RSO600547S, RSO601584S, RSO603138S, RSO603801S, RSO609732S, RSO609830S, RSO609831S, RSO610473S, RSO610551S, RSO610708S, RSO610718S, RSO610974S, RSO611575S, RSO611706S, RSO611993S, RSO612109S, RSO612111S, RSO612127S, RSO612243S, RSO612485S, RSO612521S, RSO612656S, RSO601041S, RSO601590S, RSO605748S, RSO605808S, RSO606359S, RSO606447S, RSO606955S, RSO609919S, RSO610973S, RSO611635S, RSO612092S, RSO612193S, RSO603760S, RSO603883S, RSO604362S, RSO605137S, RSO612091S, RSO604776S, RSO604997S, RSO605339S, RSO605346S, RSO605431S, RSO600191S, RSO600618S, RSO603487S, RSO611615S, RSO612676S, RSO612680S, RSO600220S, RSO604465S, RSO600725S, RSO610516S, RSO612366S, RSO612699S, RSO600988S, RSO601846S, RSO602619S, RSO603076S, RSO603091S, RSO603146S, RSO603510S, RSO603708S, RSO604481S, RSO605116S, RSO605276S, RSO605355S, RSO605444S, RSO605512S, RSO605619S, RSO605875S, RSO606573S, RSO606637S, RSO606642S, RSO607284S, RSO607416S, RSO607711S, RSO608597S, RSO608720S, RSO609669S, RSO610488S, RSO610800S, RSO611677S, RSO612275S, RSO600983S, RSO601340S, RSO601717S, RSO603938S, RSO605256S, RSO605820S, RSO605984S, RSO606907S, RSO608232S, RSO608730S, RSO603538S, RSO609208S, RSO610900S, RSO611391S, RSO612375S, RSO600520S, RSO600646S, RSO600757S, RSO601583S, RSO601723S, RSO601811S, RSO602871S, RSO602986S, RSO603456S, RSO603650S, RSO604849S, RSO605350S, RSO605505S, RSO605811S, RSO606405S, RSO606871S, RSO607964S, RSO608246S, RSO602916S, RSO603196S, RSO605084S, RSO605316S, RSO606455S, RSO606700S, RSO606977S, RSO607716S, RSO602531S, RSO602955S, RSO609727S, RSO609821S, RSO610343S, RSO604975S, RSO604988S, RSO605468S, RSO607883S, RSO611649S, RSO610355S, RSO611513S, RSO600245S, RSO604681S, RSO605877S, RSO608991S, RSO603293S, RSO600174S, RSO612580S, RSO601576S, RSO602341S, RSO608472S, RSO600640S, RSO605119S, RSO605234S, RSO605834S, RSO605850S, RSO601192S, RSO605087S, RSO612494S, RSO610494S, RSO611640S, RSO612645S, RSO600155S, RSO600256S, RSO600257S, RSO600263S, RSO600395S, RSO600489S, RSO600490S, RSO600548S, RSO600597S, RSO600799S, RSO600822S, RSO600885S, RSO600901S, RSO600921S, RSO601194S, RSO601719S, RSO605409S, RSO605433S, RSO606239S, RSO607195S, RSO600224S, RSO600466S, RSO601184S, RSO603111S, RSO603183S, RSO603429S, RSO604766S, RSO604775S, RSO611998S, RSO609497S, RSO609571S, RSO609869S, RSO610323S, RSO610489S, RSO610490S, RSO610797S, RSO600077S, RSO600114S, RSO600143S, RSO600687S, RSO600970S, RSO601126S, RSO601127S, RSO601674S, RSO602501S, RSO602537S, RSO602580S, RSO603029S, RSO603153S, RSO603175S, RSO603900S, RSO605177S, RSO605269S, RSO605271S, RSO605273S, RSO605294S, RSO605485S, RSO605499S, RSO606100S, RSO606670S, RSO607016S, RSO608509S, RSO608510S, RSO600193S, RSO600194S, RSO600419S, RSO600494S, RSO600638S, RSO600740S, RSO600769S, RSO600890S, RSO601119S, RSO601231S, RSO601691S, RSO601826S, RSO601867S, RSO602549S, RSO602590S, RSO602740S, RSO602886S, RSO602968S, RSO603057S, RSO605128S, RSO605286S, RSO606321S, RSO603173S, RSO608351S, RSO611605S, RSO610930S, RSO612661S, RSO603064S, RSO603457S, RSO604282S, RSO607382S, RSO600701S, RSO603151S, RSO603152S, RSO603155S, RSO603432S, RSO603453S, RSO603616S, RSO603619S, RSO604594S, RSO605121S, RSO605255S, RSO606283S, RSO610333S, RSO604506S, RSO604509S, RSO604510S, RSO605901S, RSO609001S, RSO601423S, RSO612492S, RSO612522S, RSO613209S, RSO610589S, RSO610917S, RSO610927S, RSO611641S, RSO603036S, RSO603469S, RSO609272S, RSO609244S, RSO605342S, RSO606047S, RSO610472S, RSO610559S, RSO603674S, RSO609772S, RSO609699S, RSO609794S, RSO609834S, RSO610320S, RSO610378S, RSO610501S, RSO610915S, RSO610919S, RSO611519S, RSO611647S, RSO603224S, RSO604583S, RSO609358S, RSO600865S, RSO600937S, RSO600985S, RSO602801S, RSO603442S, RSO603980S, RSO603982S, RSO606000S, RSO606301S, RSO606558S, RSO605163S, RSO606243S, RSO606650S, RSO607196S, RSO607278S, RSO608025S, RSO605406S, RSO606183S, RSO609918S, RSO600159S, RSO600163S, RSO600165S, RSO600184S, RSO601138S, RSO601232S, RSO601380S, RSO602591S, RSO603364S, RSO611840S, RSO609284S, RSO609522S, RSO609610S, RSO609822S, RSO610420S, RSO610427S, RSO610462S, RSO610612S, RSO610653S, RSO610705S, RSO610706S, RSO611352S, RSO611387S, RSO612006S, RSO612008S, RSO612075S, RSO612561S, RSO601874S, RSO601875S, RSO602313S, RSO603121S, RSO603253S, RSO603479S, RSO603656S, RSO604231S, RSO604232S, RSO604985S, RSO605451S, RSO605455S, RSO605657S, RSO605681S, RSO605758S, RSO605911S, RSO606123S, RSO606482S, RSO606541S, RSO607014S, RSO607609S, RSO607693S, RSO607919S, RSO607950S, RSO608643S, RSO608667S, RSO608668S, RSO608698S, RSO608887S, RSO608893S, RSO608919S, RSO608999S, RSO609000S, RSO609073S, RSO609258S, RSO609547S, RSO609697S, RSO609801S, RSO610322S, RSO610929S, RSO611670S, RSO612379S, RSO600148S, RSO600650S, RSO601914S, RSO603203S, RSO603318S, RSO603578S, RSO604220S, RSO604998S, RSO605280S, RSO605282S, RSO605283S, RSO605368S, RSO606431S, RSO607230S, RSO607234S, RSO607331S, RSO607336S, RSO608494S, RSO608987S, RSO600622S, RSO600623S, RSO600624S, RSO602944S, RSO602946S, RSO604454S, RSO604455S, RSO604456S, RSO604457S, RSO604458S, RSO604459S, RSO604460S, RSO610603S, RSO600606S, RSO600607S, RSO600629S, RSO600796S, RSO600994S, RSO601052S, RSO601061S, RSO601070S, RSO601304S, RSO601414S, RSO601495S, RSO602564S, RSO603813S, RSO610389S, RSO602959S, RSO603397S, RSO601287S, RSO609939S, RSO611431S, RSO602811S, RSO601042S, RSO601347S, RSO601382S, RSO601676S, RSO602737S, RSO603911S, RSO605571S, RSO607614S, RSO607615S, RSO607616S, RSO609924S, RSO607268S, RSO608850S, RSO611588S, RSO611764S, RSO611801S, RSO606475S, RSO609246S, RSO600736S, RSO600737S, RSO601725S, RSO602776S, RSO603423S, RSO603536S, RSO606461S, RSO600761S, RSO603612S, RSO604137S, RSO608583S, RSO603099S, RSO611984S, RSO611985S, RSO610865S, RSO600137S, RSO601706S, RSO601709S, RSO602559S, RSO603082S, RSO603582S, RSO603599S, RSO603995S, RSO604793S, RSO605568S, RSO605591S, RSO606085S, RSO606952S, RSO607631S, RSO607653S, RSO607954S, RSO607955S, RSO604274S, RSO604275S, RSO604807S, RSO604825S, RSO604847S, RSO600706S, RSO600707S, RSO601028S, RSO605490S, RSO605563S, RSO600266S, RSO600594S, RSO600886S, RSO601009S, RSO601118S, RSO601417S, RSO601897S, RSO602887S, RSO602888S, RSO603065S, RSO603381S, RSO604212S, RSO605015S, RSO605039S, RSO606043S, RSO606506S, RSO608748S, RSO601608S, RSO602319S, RSO602321S, RSO603494S, RSO602317S, RSO610353S, RSO600203S, RSO600950S, RSO608779S, RSO610649S, RSO611528S, RSO611637S, RSO612195S, RSO603759S, RSO603761S, RSO604270S, RSO604367S, RSO605160S, RSO605354S, RSO605458S, RSO605760S, RSO605973S, RSO606247S, RSO607664S, RSO608894S, RSO609875S, RSO610336S, RSO611727S, RSO612012S, RSO612605S, RSO601487S, RSO601525S, RSO602919S, RSO603004S, RSO603009S, RSO606658S, RSO606660S, RSO606661S, RSO607380S, RSO608182S, RSO608512S, RSO608684S, RSO612927S, RSO609888S, RSO611715S, RSO612660S, RSO603197S, RSO603206S, RSO603304S, RSO603916S, RSO604062S, RSO604354S, RSO605481S, RSO606583S, RSO607235S, RSO612952S, RSO601715S, RSO601716S, RSO602995S, RSO602996S, RSO602997S, RSO603670S, RSO603698S, RSO603700S, RSO604365S, RSO604571S, RSO605931S, RSO605932S, RSO606704S, RSO606706S, RSO606708S, RSO606711S, RSO607783S, RSO607786S, RSO607787S, RSO607825S, RSO607826S, RSO612905S, RSO612907S, RSO612089S, RSO601501S, RSO601502S, RSO601547S, RSO605650S, RSO607827S, RSO609802S, RSO608721S, RSO601017S, RSO610735S, RSO601312S, RSO601410S, RSO601560S, RSO601561S, RSO603148S, RSO603603S, RSO604116S, RSO605223S, RSO606284S, RSO606308S, RSO606390S, RSO607863S, RSO610405S, RSO611499S, RSO612259S, RSO601242S, RSO602949S, RSO603519S, RSO603520S, RSO603573S, RSO603986S, RSO604797S, RSO605055S, RSO606616S, RSO606619S, RSO608199S, RSO608626S, RSO608705S, RSO610873S, RSO610875S, RSO611502S, RSO611564S, RSO612078S, RSO612181S, RSO612525S, RSO600158S, RSO600173S, RSO600201S, RSO600234S, RSO600282S, RSO600728S, RSO600964S, RSO601046S, RSO601652S, RSO601682S, RSO602308S, RSO602543S, RSO602563S, RSO602855S, RSO604970S, RSO605232S, RSO605298S, RSO605567S, RSO605588S, RSO606618S, RSO606805S, RSO606983S, RSO607093S, RSO607655S, RSO607704S, RSO608174S, RSO608196S, RSO608781S, RSO609421S, RSO613257S, RSO609504S, RSO608812S, RSO609249S, RSO613188S, RSO609549S, RSO610037S, RSO612103S, RSO612552S, RSO602635S, RSO603390S, RSO605090S, RSO607620S, RSO612300S, RSO606404S, RSO607228S, RSO607612S, RSO608242S, RSO608536S, RSO608662S, RSO609234S, RSO608581S, RSO605779S, RSO607107S, RSO607267S, RSO607793S, RSO607986S, RSO608001S, RSO608002S, RSO608003S, RSO608012S, RSO609289S, RSO610038S, RSO610422S, RSO611730S, RSO600821S, RSO601166S, RSO601405S, RSO601818S, RSO605498S, RSO605509S, RSO605539S, RSO605992S, RSO606071S, RSO610565S, RSO607588S, RSO609773S, RSO610654S, RSO611514S, RSO611707S, RSO606327S, RSO608208S, RSO608500S, RSO608736S, RSO608926S, RSO610364S, RSO610894S, RSO610928S, RSO611734S, RSO612343S, RSO612675S, RSO601687S, RSO601689S, RSO602931S, RSO603103S, RSO603942S, RSO603973S, RSO604839S, RSO605159S, RSO605462S, RSO606099S, RSO607775S, RSO608319S, RSO608979S, RSO609276S, RSO609346S, RSO603192S, RSO605357S, RSO609704S, RSO609705S, RSO610950S, RSO611775S, RSO611894S, RSO611896S, RSO611897S, RSO600226S, RSO600565S, RSO601077S, RSO601079S, RSO601092S, RSO601094S, RSO601169S, RSO601264S, RSO601317S, RSO601420S, RSO601428S, RSO601451S, RSO601513S, RSO601726S, RSO601799S, RSO601836S, RSO601862S, RSO603063S, RSO603104S, RSO603535S, RSO603646S, RSO603652S, RSO603749S, RSO603765S, RSO604145S, RSO604207S, RSO604208S, RSO604209S, RSO604578S, RSO605104S, RSO605235S, RSO605321S, RSO605323S, RSO605360S, RSO605679S, RSO605822S, RSO606271S, RSO606303S, RSO606622S, RSO607390S, RSO607391S, RSO607917S, RSO608291S, RSO610493S, RSO610657S, RSO600742S, RSO603003S, RSO610402S, RSO610546S, RSO610586S, RSO611664S, RSO611729S, RSO601163S, RSO601200S, RSO601398S, RSO601504S, RSO601780S, RSO601852S, RSO602809S, RSO603254S, RSO603347S, RSO603398S, RSO603696S, RSO604141S, RSO604628S, RSO605665S, RSO606963S, RSO608250S, RSO608327S, RSO600805S, RSO606914S, RSO610937S, RSO612669S, RSO612674S, RSO612636S, RSO602505S, RSO602969S, RSO603505S, RSO604325S, RSO605319S, RSO606463S, RSO611582S, RSO603284S, RSO604205S, RSO604838S, RSO606096S, RSO606336S, RSO607961S, RSO603562S, RSO604332S, RSO606870S, RSO607707S, RSO604326S, RSO604328S, RSO604897S, RSO605439S, RSO605478S, RSO605693S, RSO605757S, RSO606949S, RSO607023S, RSO604364S, RSO605725S, RSO609796S, RSO602784S, RSO606294S, RSO606544S, RSO601452S, RSO601514S, RSO601839S, RSO601850S, RSO602599S, RSO603784S, RSO603917S, RSO604001S, RSO604258S, RSO605132S, RSO605337S, RSO605372S, RSO605426S, RSO605672S, RSO605678S, RSO605829S, RSO606064S, RSO606980S, RSO607385S, RSO607389S, RSO607577S, RSO607695S, RSO607696S, RSO608341S, RSO608499S, RSO608505S, RSO608740S, RSO608839S, RSO608994S, RSO609279S, RSO603149S, RSO603553S, RSO603881S, RSO605425S, RSO606395S, RSO607691S, RSO606061S, RSO608469S, RSO608558S, RSO600096S, RSO600232S, RSO600952S, RSO602521S, RSO602939S, RSO603706S, RSO606596S, RSO609723S, RSO611609S, RSO611611S, RSO602324S, RSO605795S, RSO606147S, RSO609722S, RSO610515S, RSO610554S, RSO601359S, RSO602842S, RSO608490S, RSO609492S, RSO609505S, RSO609576S, RSO609870S, RSO610312S, RSO610315S, RSO610338S, RSO610363S, RSO610552S, RSO610553S, RSO610627S, RSO610629S, RSO610640S, RSO610744S, RSO610749S, RSO611721S, RSO612062S, RSO612150S, RSO612190S, RSO612278S, RSO612622S, RSO612704S, RSO601366S, RSO601397S, RSO601655S, RSO602577S, RSO602857S, RSO603108S, RSO603437S, RSO603454S, RSO603521S, RSO603806S, RSO603923S, RSO604000S, RSO604284S, RSO604479S, RSO605356S, RSO605491S, RSO605586S, RSO606004S, RSO606080S, RSO606228S, RSO606356S, RSO606384S, RSO606621S, RSO606933S, RSO607669S, RSO607676S, RSO607905S, RSO608073S, RSO608119S, RSO608378S, RSO608557S, RSO608755S, RSO608953S, RSO610366S, RSO612061S, RSO612277S, RSO603466S, RSO604075S, RSO601415S, RSO610913S, RSO610912S, RSO601362S, RSO606496S, RSO601043S, RSO603818S, RSO609852S, RSO600445S, RSO602346S, RSO603888S, RSO603987S, RSO603996S, RSO601122S, RSO601817S, RSO603722S, RSO606452S, RSO608866S, RSO606477S, RSO602865S, RSO612562S, RSO601358S, RSO612647S, RSO609804S, RSO610512S, RSO610584S, RSO611614S, RSO611799S, RSO604105S, RSO606322S, RSO607344S, RSO607346S, RSO609760S, RSO612380S, RSO600838S, RSO601787S, RSO602774S, RSO603087S, RSO603169S, RSO603385S, RSO603386S, RSO603388S, RSO603389S, RSO600787S, RSO600823S, RSO601066S, RSO605770S, RSO602703S, RSO604172S, RSO605857S, RSO611776S, RSO612232S, RSO612623S, RSO612624S, RSO612625S, RSO612626S, RSO612627S, RSO612792S, RSO612793S, RSO612794S, RSO612795S, RSO612796S, RSO612797S, RSO612798S, RSO612799S, RSO612800S, RSO612801S, RSO603667S, RSO613221S, RSO613269S, RSO613273S, RSO601803S, RSO603421S, RSO605830S, RSO609745S, RSO609914S, RSO610618S, RSO612016S, RSO600095S, RSO600190S, RSO600196S, RSO600704S, RSO600884S, RSO600943S, RSO601115S, RSO601133S, RSO601649S, RSO602991S, RSO603717S, RSO603837S, RSO604068S, RSO605032S, RSO605480S, RSO606667S, RSO607108S, RSO607142S, RSO608471S, RSO608514S, RSO608620S, RSO608649S, RSO600192S, RSO600225S, RSO603885S, RSO603886S, RSO607257S, RSO607808S, RSO600228S, RSO601399S, RSO612198S, RSO610701S, RSO612007S, RSO612133S, RSO612185S, RSO612302S, RSO612390S, RSO601438S, RSO601705S, RSO603526S, RSO604109S, RSO605998S, RSO606494S, RSO607748S, RSO607888S, RSO604486S, RSO609805S, RSO610513S, RSO610772S, RSO611146S, RSO611458S, RSO611613S, RSO612037S, RSO612038S, RSO612140S, RSO612197S, RSO612229S, RSO612304S, RSO612305S, RSO612363S, RSO612496S, RSO612537S, RSO612602S, RSO612695S, RSO612697S, RSO600721S, RSO603225S, RSO603259S, RSO604491S, RSO605010S, RSO605216S, RSO605440S, RSO605461S, RSO605483S, RSO605708S, RSO605917S, RSO606092S, RSO607143S, RSO608605S, RSO608690S, RSO606217S, RSO612071S, RSO610522S, RSO610787S, RSO611602S, RSO600084S, RSO600085S, RSO600086S, RSO600262S, RSO600377S, RSO602318S, RSO602597S, RSO603157S, RSO603887S, RSO605443S, RSO605446S, RSO605630S, RSO605651S, RSO605750S, RSO607262S, RSO608370S, RSO608688S, RSO611787S, RSO610331S, RSO600775S, RSO600824S, RSO600830S, RSO601697S, RSO604093S, RSO606113S, RSO609416S, RSO604682S, RSO607056S, RSO608796S, RSO600760S, RSO601489S, RSO601890S, RSO605441S, RSO608876S, RSO600791S, RSO610643S, RSO611666S, RSO612313S, RSO601845S, RSO602817S, RSO603161S, RSO603182S, RSO603645S, RSO603675S, RSO603766S, RSO603838S, RSO605135S, RSO605390S, RSO605450S, RSO605587S, RSO605833S, RSO606184S, RSO608366S, RSO608629S, RSO608806S, RSO608808S, RSO608809S, RSO613124S, RSO613125S, RSO609714S, RSO609715S, RSO609716S, RSO610142S, RSO610166S, RSO610167S, RSO610169S, RSO610170S, RSO610997S, RSO610999S, RSO602644S, RSO602651S, RSO602702S, RSO602705S, RSO602719S, RSO602898S, RSO602899S, RSO606270S, RSO606375S, RSO606376S, RSO606377S, RSO606378S, RSO607247S, RSO607252S, RSO607320S, RSO607328S, RSO607849S, RSO607851S, RSO608101S, RSO608102S, RSO608104S, RSO608107S, RSO608108S, RSO608110S, RSO608111S, RSO608112S, RSO609026S, RSO609027S, RSO609029S, RSO609163S, RSO609164S, RSO609167S, RSO609168S, RSO609170S, RSO609171S, RSO609172S, RSO609173S, RSO610006S, RSO610011S, RSO610139S, RSO610140S, RSO610143S, RSO610145S, RSO610149S, RSO610151S, RSO610152S, RSO610153S, RSO610154S, RSO610155S, RSO610156S, RSO610164S, RSO610165S, RSO610168S, RSO610692S, RSO610994S, RSO610995S, RSO613236S, RSO613237S, RSO613238S, RSO613239S, RSO613240S, RSO613241S, RSO613242S, RSO613243S, RSO613244S, RSO613245S, RSO610736S, RSO612184S, RSO612316S, RSO600797S, RSO601390S, RSO601902S, RSO602600S, RSO602642S, RSO602711S, RSO602718S, RSO602897S, RSO603651S, RSO604156S, RSO605122S, RSO605838S, RSO607141S, RSO607248S, RSO607254S, RSO607848S, RSO608870S, RSO600738S, RSO603435S, RSO604974S, RSO605251S, RSO606282S, RSO606481S, RSO603266S, RSO605347S, RSO605862S, RSO606445S, RSO607006S, RSO600554S, RSO603477S, RSO606298S, RSO606413S, RSO602802S, RSO602869S, RSO603193S, RSO604121S, RSO605716S, RSO607806S, RSO602930S, RSO603264S, RSO603265S, RSO609293S, RSO600255S, RSO611620S, RSO609597S, RSO609598S, RSO609599S, RSO601856S, RSO601881S, RSO604539S, RSO608752S, RSO608807S, RSO605717S, RSO601216S, RSO601217S, RSO601294S, RSO601356S, RSO601379S, RSO601496S, RSO601755S, RSO601763S, RSO603067S, RSO603588S, RSO606244S, RSO608422S, RSO600643S, RSO601456S, RSO601570S, RSO601577S, RSO601842S, RSO601861S, RSO605091S, RSO607559S, RSO607688S, RSO607721S, RSO609730S, RSO609922S, RSO609923S, RSO610409S, RSO610711S, RSO611572S, RSO611705S, RSO612125S, RSO612194S, RSO612296S, RSO612298S, RSO612299S, RSO612651S, RSO600117S, RSO600612S, RSO600613S, RSO600630S, RSO600663S, RSO600696S, RSO600719S, RSO600844S, RSO601011S, RSO601116S, RSO601172S, RSO601589S, RSO601592S, RSO601837S, RSO601870S, RSO601884S, RSO602496S, RSO602621S, RSO602998S, RSO603374S, RSO603490S, RSO603730S, RSO604061S, RSO604311S, RSO604979S, RSO605004S, RSO605007S, RSO605206S, RSO605213S, RSO605237S, RSO605238S, RSO605517S, RSO605692S, RSO605768S, RSO605946S, RSO606416S, RSO606505S, RSO606518S, RSO606598S, RSO606954S, RSO607100S, RSO607102S, RSO607420S, RSO607646S, RSO607809S, RSO607897S, RSO607940S, RSO608223S, RSO608486S, RSO608614S, RSO608696S, RSO608791S, RSO608793S, RSO608936S, RSO608998S, RSO600247S, RSO600476S, RSO601128S, RSO601274S, RSO601557S, RSO601795S, RSO601812S, RSO603315S, RSO603685S, RSO603912S, RSO604024S, RSO601010S, RSO601193S, RSO601227S, RSO601233S, RSO601446S, RSO601785S, RSO603350S, RSO605680S, RSO605831S, RSO606508S, RSO606587S, RSO606754S, RSO607240S, RSO608993S, RSO610335S, RSO610628S, RSO610677S, RSO610740S, RSO611722S, RSO601349S, RSO604894S, RSO605516S, RSO605559S, RSO605892S, RSO607523S, RSO608021S, RSO608070S, RSO608954S, RSO610625S, RSO610761S, RSO600590S, RSO610562S, RSO603741S, RSO604601S, RSO605774S, RSO606882S, RSO607042S, RSO608934S, RSO610879S, RSO602342S, RSO605330S, RSO605979S, RSO606179S, RSO607456S, RSO606441S, RSO606462S, RSO601671S, RSO601678S, RSO601757S, RSO600110S, RSO601188S, RSO601433S, RSO601546S, RSO601677S, RSO601784S, RSO602305S, RSO602924S, RSO603875S, RSO604032S, RSO604055S, RSO605685S, RSO606144S, RSO610773S, RSO612234S, RSO612698S, RSO603010S, RSO603738S, RSO603967S, RSO604562S, RSO608655S, RSO609280S, RSO600897S, RSO601245S, RSO603244S, RSO605640S, RSO606101S, RSO610456S, RSO610598S, RSO611688S, RSO600693S, RSO605723S, RSO609891S, RSO606741S, RSO612648S, RSO600097S, RSO600270S, RSO603340S, RSO603563S, RSO603757S, RSO603918S, RSO604309S, RSO605242S, RSO608674S, RSO600620S, RSO600621S, RSO604520S, RSO604526S, RSO604527S, RSO604528S, RSO604534S, RSO604536S, RSO604541S, RSO604546S, RSO604567S, RSO600188S, RSO600189S, RSO600429S, RSO600436S, RSO600439S, RSO600440S, RSO600444S, RSO600451S, RSO600569S, RSO601241S, RSO601320S, RSO601699S, RSO601734S, RSO601735S, RSO601736S, RSO601768S, RSO601769S, RSO601770S, RSO601771S, RSO601772S, RSO601773S, RSO602813S, RSO602837S, RSO602901S, RSO602982S, RSO603040S, RSO603219S, RSO603250S, RSO603602S, RSO603822S, RSO603902S, RSO607903S, RSO600832S, RSO600904S, RSO601354S, RSO605502S, RSO606348S, RSO607629S, RSO610381S, RSO610390S, RSO610391S, RSO610545S, RSO610902S, RSO611651S, RSO611725S, RSO604586S, RSO610857S, RSO612214S, RSO603827S, RSO605348S, RSO605673S, RSO606673S, RSO607018S, RSO607781S, RSO607886S, RSO602830S, RSO602831S, RSO602836S, RSO602840S, RSO609403S, RSO609406S, RSO609410S, RSO609412S, RSO611689S, RSO611690S, RSO611924S, RSO611925S, RSO611926S, RSO611927S, RSO611928S, RSO610416S, RSO602835S, RSO605093S, RSO601491S, RSO603363S, RSO612220S, RSO612221S, RSO600920S, RSO611586S, RSO601821S, RSO603609S, RSO604653S, RSO605414S, RSO600381S, RSO600382S, RSO600688S, RSO600689S, RSO601707S, RSO601978S, RSO602728S, RSO602853S, RSO602854S, RSO602856S, RSO602859S, RSO603467S, RSO603772S, RSO604150S, RSO605034S, RSO605052S, RSO605560S, RSO606620S, RSO607433S, RSO607942S, RSO607943S, RSO609710S, RSO604157S, RSO609025S, RSO604504S, RSO610680S, RSO610681S, RSO610682S, RSO610683S, RSO611536S, RSO611537S, RSO611887S, RSO607238S, RSO607269S, RSO607444S, RSO607447S, RSO607448S, RSO611535S, RSO611692S, RSO611795S, RSO611888S, RSO612538S, RSO612539S, RSO611691S, RSO601509S, RSO607070S, RSO607565S, RSO607574S, RSO608081S, RSO608332S, RSO608334S, RSO608335S, RSO608345S, RSO608346S, RSO608895S, RSO608896S, RSO609238S, RSO609239S, RSO609240S, RSO609241S, RSO609243S, RSO611238S, RSO611263S, RSO611292S, RSO611303S, RSO611319S, RSO602545S, RSO603234S, RSO605865S, RSO606195S, RSO607965S, RSO607966S, RSO610172S, RSO610173S, RSO610176S, RSO610177S, RSO610178S, RSO610179S, RSO610180S, RSO610181S, RSO610182S, RSO610183S, RSO610188S, RSO610189S, RSO610190S, RSO610191S, RSO610192S, RSO610193S, RSO610194S, RSO610195S, RSO610196S, RSO610198S, RSO610199S, RSO610200S, RSO610691S, RSO610991S, RSO610992S, RSO610993S, RSO611148S, RSO611149S, RSO611150S, RSO611151S, RSO611152S, RSO611154S, RSO611155S, RSO611156S, RSO611157S, RSO611158S, RSO611159S, RSO611160S, RSO611161S, RSO611162S, RSO611163S, RSO611165S, RSO611166S, RSO611167S, RSO611168S, RSO611169S, RSO611170S, RSO611173S, RSO611176S, RSO611177S, RSO611181S, RSO611183S, RSO611186S, RSO611187S, RSO611189S, RSO611190S, RSO611191S, RSO611193S, RSO611194S, RSO611195S, RSO611196S, RSO611197S, RSO611902S, RSO611903S, RSO611905S, RSO611906S, RSO611929S, RSO611932S, RSO611937S, RSO611941S, RSO611943S, RSO603281S, RSO603677S, RSO603678S, RSO603682S, RSO603718S, RSO603719S, RSO603720S, RSO603721S, RSO603731S, RSO603733S, RSO603734S, RSO603740S, RSO603745S, RSO603746S, RSO603758S, RSO603770S, RSO604589S, RSO604591S, RSO604592S, RSO605771S, RSO605773S, RSO605780S, RSO605785S, RSO605787S, RSO605788S, RSO605789S, RSO605790S, RSO605791S, RSO605792S, RSO605794S, RSO605797S, RSO605799S, RSO605800S, RSO605801S, RSO605802S, RSO605806S, RSO605810S, RSO605813S, RSO605814S, RSO605815S, RSO605816S, RSO607154S, RSO607156S, RSO607157S, RSO607354S, RSO607381S, RSO607402S, RSO607403S, RSO607406S, RSO607409S, RSO607411S, RSO607501S, RSO607502S, RSO607503S, RSO607504S, RSO607505S, RSO607506S, RSO607507S, RSO607508S, RSO607509S, RSO607510S, RSO607511S, RSO607512S, RSO607513S, RSO607514S, RSO607516S, RSO607517S, RSO607518S, RSO607519S, RSO607522S, RSO607836S, RSO607837S, RSO607838S, RSO607843S, RSO607844S, RSO607845S, RSO607846S, RSO607847S, RSO607921S, RSO608087S, RSO608088S, RSO608089S, RSO608090S, RSO608091S, RSO608094S, RSO608095S, RSO608096S, RSO608098S, RSO608099S, RSO608127S, RSO608128S, RSO608129S, RSO608130S, RSO608131S, RSO608132S, RSO608133S, RSO608134S, RSO608135S, RSO608136S, RSO608137S, RSO608138S, RSO608139S, RSO608140S, RSO608142S, RSO608143S, RSO608144S, RSO608146S, RSO608147S, RSO608148S, RSO608149S, RSO608151S, RSO608152S, RSO608153S, RSO608155S, RSO608156S, RSO608157S, RSO608158S, RSO608159S, RSO608160S, RSO608161S, RSO608162S, RSO608163S, RSO608410S, RSO608897S, RSO608898S, RSO608899S, RSO608901S, RSO608902S, RSO608903S, RSO609042S, RSO609043S, RSO609044S, RSO609045S, RSO609046S, RSO609047S, RSO609048S, RSO609049S, RSO609050S, RSO609052S, RSO609054S, RSO609058S, RSO609069S, RSO609071S, RSO609081S, RSO609087S, RSO609090S, RSO609095S, RSO609096S, RSO609097S, RSO609098S, RSO609099S, RSO609100S, RSO609102S, RSO610175S, RSO610197S, RSO610201S, RSO610202S, RSO611164S, RSO611171S, RSO611904S, RSO611931S, RSO611933S, RSO611934S, RSO611935S, RSO611936S, RSO611938S, RSO611939S, RSO611940S, RSO611942S, RSO601865S, RSO603948S, RSO605514S, RSO609712S, RSO609713S, RSO610036S, RSO610342S, RSO610988S, RSO601482S, RSO601483S, RSO603810S, RSO603812S, RSO604587S, RSO604718S, RSO604726S, RSO604729S, RSO604730S, RSO604977S, RSO604982S, RSO604984S, RSO605596S, RSO605599S, RSO605600S, RSO605601S, RSO606158S, RSO606159S, RSO606161S, RSO607078S, RSO607079S, RSO607080S, RSO607116S, RSO607117S, RSO607118S, RSO607834S, RSO607835S, RSO608038S, RSO608039S, RSO608041S, RSO608042S, RSO608043S, RSO608294S, RSO607004S, RSO600918S, RSO600125S, RSO600745S, RSO612070S, RSO605842S, RSO607678S, RSO610000S, RSO610483S, RSO610484S, RSO610485S, RSO610597S, RSO610599S, RSO610600S, RSO610601S, RSO610602S, RSO610938S, RSO611004S, RSO611111S, RSO611113S, RSO611115S, RSO611118S, RSO611119S, RSO611120S, RSO611121S, RSO611124S, RSO611126S, RSO600586S, RSO602634S, RSO602866S, RSO602989S, RSO602992S, RSO602993S, RSO602994S, RSO603270S, RSO603271S, RSO603555S, RSO603559S, RSO603560S, RSO603724S, RSO603725S, RSO604037S, RSO604038S, RSO604514S, RSO604516S, RSO604598S, RSO604599S, RSO604600S, RSO604610S, RSO604611S, RSO604612S, RSO604617S, RSO604618S, RSO604619S, RSO604620S, RSO604622S, RSO605211S, RSO605214S, RSO605295S, RSO605296S, RSO605297S, RSO607457S, RSO607459S, RSO607463S, RSO607465S, RSO607466S, RSO607468S, RSO607470S, RSO607473S, RSO607477S, RSO607478S, RSO607479S, RSO607480S, RSO607481S, RSO607486S, RSO608116S, RSO608117S, RSO608372S, RSO608374S, RSO608380S, RSO608383S, RSO608384S, RSO608391S, RSO608394S, RSO608395S, RSO611003S, RSO611112S, RSO611114S, RSO611116S, RSO611122S, RSO611123S, RSO611125S, RSO609958S, RSO609959S, RSO609960S, RSO610534S, RSO610537S, RSO610765S, RSO610776S, RSO610777S, RSO610784S, RSO610786S, RSO610791S, RSO610799S, RSO610813S, RSO601806S, RSO603032S, RSO603273S, RSO605209S, RSO606194S, RSO607044S, RSO607464S, RSO610668S, RSO607253S, RSO613169S, RSO613171S, RSO613173S, RSO613174S, RSO613175S, RSO613178S, RSO609769S, RSO609813S, RSO609814S, RSO609899S, RSO609962S, RSO610279S, RSO610432S, RSO610448S, RSO611836S, RSO611837S, RSO603097S, RSO606166S, RSO606253S, RSO606256S, RSO606483S, RSO607047S, RSO607048S, RSO607106S, RSO607359S, RSO607744S, RSO607852S, RSO607853S, RSO608358S, RSO608508S, RSO609055S, RSO609393S, RSO609396S, RSO600227S, RSO600932S, RSO600939S, RSO601296S, RSO601903S, RSO603399S, RSO603610S, RSO603625S, RSO603628S, RSO604475S, RSO605131S, RSO605328S, RSO608768S, RSO608913S, RSO609261S, RSO601545S, RSO602867S, RSO602966S, RSO603086S, RSO603451S, RSO605432S, RSO610468S, RSO610656S, RSO605486S, RSO605866S, RSO608669S, RSO603839S, RSO604964S, RSO610506S, RSO611712S, RSO612342S, RSO612666S, RSO600815S, RSO600177S, RSO600680S, RSO610935S, RSO601457S, RSO610550S, RSO610713S, RSO611703S, RSO611994S, RSO612106S, RSO612107S, RSO612169S, RSO612297S, RSO612567S, RSO603688S, RSO606287S, RSO607570S, RSO607579S, RSO607677S, RSO608489S, RSO608594S, RSO600512S, RSO600536S, RSO605819S, RSO601753S, RSO609405S, RSO610868S, RSO601907S, RSO603178S, RSO605715S, RSO600600S, RSO601150S, RSO604148S, RSO604580S, RSO605287S, RSO606524S, RSO607270S, RSO607360S, RSO611737S, RSO612475S, RSO612587S, RSO603299S, RSO604451S, RSO604452S, RSO604453S, RSO605036S, RSO605345S, RSO606309S, RSO606401S, RSO607035S, RSO607036S, RSO607362S, RSO607363S, RSO608799S, RSO609308S, RSO609624S, RSO609657S, RSO609754S, RSO609848S, RSO609903S, RSO609927S, RSO610645S, RSO610823S, RSO602492S, RSO604537S, RSO606534S, RSO606747S, RSO606776S, RSO607007S, RSO607008S, RSO607029S, RSO607030S, RSO607031S, RSO607146S, RSO607150S, RSO607495S, RSO607904S, RSO607957S, RSO608220S, RSO608221S, RSO608295S, RSO608638S, RSO608734S, RSO609265S, RSO609266S, RSO601157S, RSO607561S, RSO601057S, RSO611797S, RSO605809S, RSO603025S, RSO603077S, RSO603078S, RSO603080S, RSO603081S, RSO604079S, RSO604164S, RSO612200S, RSO600677S, RSO600759S, RSO601210S, RSO601411S, RSO601459S, RSO601668S, RSO601782S, RSO601794S, RSO603337S, RSO605082S, RSO605270S, RSO605377S, RSO605437S, RSO605626S, RSO606006S, RSO606072S, RSO606074S, RSO606084S, RSO606264S, RSO606325S, RSO606575S, RSO608529S, RSO603450S, RSO611788S, RSO600206S, RSO607568S, RSO611606S, RSO612104S, RSO603780S, RSO604059S, RSO605912S, RSO609696S, RSO606801S, RSO600265S, RSO602894S, RSO603407S, RSO601494S, RSO605195S, RSO605666S, RSO605798S, RSO606439S, RSO609566S, RSO610661S, RSO610921S, RSO611657S, RSO611726S, RSO612237S, RSO600217S, RSO600221S, RSO601149S, RSO601802S, RSO604127S, RSO607229S, RSO608989S, RSO610924S, RSO600837S, RSO601218S, RSO601393S, RSO601427S, RSO601512S, RSO601558S, RSO601868S, RSO602783S, RSO603135S, RSO603218S, RSO605081S, RSO605115S, RSO606237S, RSO606355S, RSO606389S, RSO608528S, RSO609534S, RSO610544S, RSO610916S, RSO600593S, RSO600610S, RSO601550S, RSO603583S, RSO605851S, RSO606427S, RSO607281S, RSO608037S, RSO608526S, RSO609342S, RSO608328S, RSO600615S, RSO600670S, RSO600843S, RSO600867S, RSO600888S, RSO601222S, RSO601290S, RSO601291S, RSO601315S, RSO601351S, RSO601388S, RSO601432S, RSO601517S, RSO601554S, RSO601594S, RSO601596S, RSO601599S, RSO601600S, RSO601601S, RSO601602S, RSO601702S, RSO601764S, RSO601808S, RSO602962S, RSO603142S, RSO603144S, RSO603280S, RSO603383S, RSO603473S, RSO603793S, RSO603964S, RSO604760S, RSO605033S, RSO605051S, RSO605397S, RSO605448S, RSO605556S, RSO605603S, RSO605621S, RSO605906S, RSO606009S, RSO606437S, RSO607639S, RSO607640S, RSO607641S, RSO607683S, RSO607706S, RSO608219S, RSO608343S, RSO608344S, RSO608641S, RSO608642S, RSO604981S, RSO609516S, RSO605258S, RSO605869S, RSO606770S, RSO607361S, RSO608766S, RSO600121S, RSO600198S, RSO601182S, RSO601614S, RSO602595S, RSO603627S, RSO603901S, RSO604283S, RSO606291S, RSO606751S, RSO612253S, RSO606278S, RSO607132S, RSO607275S, RSO610352S, RSO612026S, RSO612174S, RSO611735S, RSO612201S, RSO612632S, RSO608569S, RSO610557S, RSO611608S, RSO612308S, RSO612641S, RSO603613S, RSO605229S, RSO605615S, RSO606830S, RSO601637S, RSO610591S, RSO609617S, RSO609809S, RSO610700S, RSO611505S, RSO611709S, RSO601155S, RSO601160S, RSO601224S, RSO603105S, RSO603321S, RSO604155S, RSO605260S, RSO602799S, RSO601159S, RSO605198S, RSO606985S, RSO609859S, RSO604118S, RSO605118S, RSO605897S, RSO606299S, RSO607153S, RSO603878S, RSO605945S, RSO606007S, RSO606654S, RSO607855S, RSO608704S, RSO601002S, RSO611774S, RSO610504S, RSO603112S, RSO603338S, RSO606245S, RSO609506S, RSO610361S, RSO610362S, RSO610424S, RSO610425S, RSO610571S, RSO610626S, RSO610869S, RSO610870S, RSO610874S, RSO611407S, RSO611565S, RSO611628S, RSO611629S, RSO611630S, RSO611631S, RSO611792S, RSO612031S, RSO612032S, RSO612033S, RSO612074S, RSO612079S, RSO612180S, RSO612216S, RSO612217S, RSO612338S, RSO612339S, RSO612356S, RSO612361S, RSO612362S, RSO612595S, RSO612598S, RSO612600S, RSO600123S, RSO600127S, RSO600129S, RSO600130S, RSO600199S, RSO600200S, RSO600222S, RSO600223S, RSO600861S, RSO600899S, RSO600981S, RSO601021S, RSO601033S, RSO601037S, RSO601162S, RSO601591S, RSO601610S, RSO602569S, RSO602589S, RSO602868S, RSO602915S, RSO602941S, RSO602943S, RSO602948S, RSO602952S, RSO602960S, RSO602999S, RSO603033S, RSO603117S, RSO603162S, RSO603220S, RSO603248S, RSO603297S, RSO603425S, RSO603500S, RSO603522S, RSO603820S, RSO603836S, RSO603870S, RSO603877S, RSO603914S, RSO603966S, RSO604009S, RSO604028S, RSO604153S, RSO604174S, RSO604296S, RSO604316S, RSO604480S, RSO604502S, RSO604503S, RSO604513S, RSO604786S, RSO604813S, RSO604895S, RSO605002S, RSO605171S, RSO605192S, RSO605482S, RSO605500S, RSO605524S, RSO605530S, RSO605589S, RSO605634S, RSO605635S, RSO605669S, RSO605777S, RSO605793S, RSO605972S, RSO606180S, RSO606296S, RSO606601S, RSO606734S, RSO606866S, RSO607010S, RSO607034S, RSO607094S, RSO607144S, RSO607205S, RSO607656S, RSO607665S, RSO608075S, RSO608076S, RSO608077S, RSO608078S, RSO608186S, RSO608312S, RSO608571S, RSO608572S, RSO608709S, RSO608710S, RSO608711S, RSO608712S, RSO608782S, RSO608891S, RSO608951S, RSO609269S, RSO609291S, RSO609292S, RSO609306S, RSO609418S, RSO609419S, RSO612594S, RSO613143S, RSO613256S, RSO610764S, RSO606010S, RSO608064S, RSO610423S, RSO600182S, RSO601109S, RSO601595S, RSO602583S, RSO605424S, RSO605628S, RSO605994S, RSO606005S, RSO606603S, RSO606924S, RSO607244S, RSO608780S, RSO601556S, RSO602361S, RSO603922S, RSO604013S, RSO605341S, RSO605941S, RSO605982S, RSO606008S, RSO606067S, RSO606372S, RSO606523S, RSO606895S, RSO608057S, RSO608063S, RSO608480S, RSO603251S, RSO606081S, RSO609901S, RSO600749S, RSO600767S, RSO601840S, RSO609347S, RSO610358S, RSO600126S, RSO600209S, RSO605894S, RSO605895S, RSO605896S, RSO605899S, RSO600839S, RSO600840S, RSO602554S, RSO607239S, RSO607241S, RSO608884S, RSO602742S, RSO606391S, RSO608645S, RSO608699S, RSO608481S, RSO600550S, RSO601298S, RSO605573S, RSO607936S, RSO601289S, RSO603767S, RSO610607S, RSO600584S, RSO600585S, RSO600666S, RSO600667S, RSO600764S, RSO600980S, RSO601269S, RSO601306S, RSO601419S, RSO601791S, RSO601879S, RSO602807S, RSO608923S, RSO609695S, RSO609890S, RSO610582S, RSO610925S, RSO610926S, RSO611728S, RSO612132S, RSO600851S, RSO600883S, RSO603414S, RSO603440S, RSO604976S, RSO605089S, RSO605200S, RSO605363S, RSO605366S, RSO605843S, RSO606351S, RSO606352S, RSO606353S, RSO606953S, RSO607323S, RSO607815S, RSO608676S, RSO608925S, RSO609271S, RSO603689S, RSO603690S, RSO603691S, RSO605050S, RSO605274S, RSO606563S, RSO602764S, RSO604783S, RSO605847S, RSO606012S, RSO606185S, RSO606640S, RSO608940S, RSO601424S, RSO601657S, RSO604025S, RSO604146S, RSO602953S, RSO603561S, RSO603684S, RSO600632S, RSO602804S, RSO603755S, RSO609897S, RSO610662S, RSO600649S, RSO600896S, RSO601680S, RSO603222S, RSO603697S, RSO604304S, RSO602655S, RSO603799S, RSO609819S, RSO609603S, RSO609896S, RSO601186S, RSO607649S, RSO612272S, RSO609500S, RSO609502S, RSO609651S, RSO609658S, RSO609679S, RSO609718S, RSO609719S, RSO609861S, RSO610334S, RSO610750S, RSO610752S, RSO611680S, RSO612040S, RSO612041S, RSO612273S, RSO612314S, RSO612508S, RSO612606S, RSO601630S, RSO601635S, RSO603257S, RSO603774S, RSO603963S, RSO604188S, RSO604233S, RSO604234S, RSO604235S, RSO604236S, RSO604237S, RSO604238S, RSO604239S, RSO604413S, RSO604867S, RSO605048S, RSO605049S, RSO605222S, RSO605766S, RSO605919S, RSO606958S, RSO607419S, RSO607772S, RSO607773S, RSO607776S, RSO607779S, RSO608010S, RSO608011S, RSO608183S, RSO608184S, RSO608190S, RSO608556S, RSO608612S, RSO608613S, RSO608683S, RSO608758S, RSO608759S, RSO608883S, RSO608963S, RSO612604S, RSO613092S, RSO608204S, RSO610834S, RSO603269S, RSO603465S, RSO603489S, RSO607978S, RSO600340S, RSO600359S, RSO600393S, RSO601486S, RSO601905S, RSO603194S, RSO603829S, RSO605314S, RSO606013S, RSO607298S, RSO607302S, RSO607752S, RSO607927S, RSO608681S, RSO610714S, RSO601445S, RSO602743S, RSO602928S, RSO603302S, RSO605456S, RSO607422S, RSO608539S, RSO608665S, RSO608885S, RSO613195S, RSO600658S, RSO600106S, RSO610914S, RSO606365S, RSO606430S, RSO607074S, RSO608213S, RSO600633S, RSO603633S, RSO604010S, RSO604111S, RSO604249S, RSO604253S, RSO604627S, RSO605329S, RSO605427S, RSO605428S, RSO605645S, RSO605654S, RSO605655S, RSO606258S, RSO606276S, RSO606369S, RSO607009S, RSO607657S, RSO607660S, RSO607932S, RSO607933S, RSO610426S, RSO600651S, RSO601522S, RSO601932S, RSO602570S, RSO602851S, RSO603147S, RSO604315S, RSO605247S, RSO605394S, RSO606531S, RSO608582S, RSO605995S, RSO609518S, RSO609652S, RSO609673S, RSO609674S, RSO609816S, RSO609818S, RSO610314S, RSO611733S, RSO612013S, RSO612157S, RSO612223S, RSO612224S, RSO612542S, RSO612617S, RSO600487S, RSO600503S, RSO600598S, RSO601055S, RSO601435S, RSO602756S, RSO602823S, RSO603123S, RSO603125S, RSO603210S, RSO603320S, RSO603604S, RSO604615S, RSO604784S, RSO604881S, RSO605138S, RSO605194S, RSO605489S, RSO605495S, RSO605551S, RSO605579S, RSO605686S, RSO605805S, RSO605807S, RSO605837S, RSO606177S, RSO606223S, RSO606292S, RSO606502S, RSO606608S, RSO606609S, RSO606635S, RSO606797S, RSO607071S, RSO607075S, RSO607076S, RSO607077S, RSO607090S, RSO607633S, RSO607670S, RSO607671S, RSO607672S, RSO607884S, RSO607913S, RSO608187S, RSO608188S, RSO608798S, RSO612925S, RSO601481S, RSO605691S, RSO609797S, RSO610644S, RSO600724S, RSO601484S, RSO601748S, RSO601838S, RSO606078S, RSO610920S, RSO600726S, RSO600773S, RSO601003S, RSO602788S, RSO602819S, RSO603132S, RSO603391S, RSO603448S, RSO606688S, RSO600457S, RSO600772S, RSO601722S, RSO603841S, RSO604163S, RSO606332S, RSO600249S, RSO608340S, RSO610206S, RSO610207S, RSO601425S, RSO605751S, RSO605856S, RSO608397S, RSO609639S, RSO612631S, RSO601053S, RSO601065S, RSO601266S, RSO601796S, RSO609490S, RSO610503S, RSO612588S, RSO603083S, RSO604552S, RSO606410S, RSO607137S, RSO607215S, RSO607216S, RSO608603S, RSO603614S, RSO602613S, RSO603352S, RSO609508S, RSO609843S, RSO610846S, RSO611738S, RSO612146S, RSO612569S, RSO600142S, RSO600931S, RSO600949S, RSO600954S, RSO600957S, RSO600958S, RSO601071S, RSO601076S, RSO601230S, RSO601759S, RSO602735S, RSO602797S, RSO602798S, RSO603226S, RSO603433S, RSO603586S, RSO603707S, RSO603796S, RSO604044S, RSO604152S, RSO604953S, RSO605023S, RSO605025S, RSO605057S, RSO605059S, RSO605070S, RSO605285S, RSO605380S, RSO605662S, RSO605903S, RSO606106S, RSO606526S, RSO606682S, RSO606732S, RSO607013S, RSO607045S, RSO607145S, RSO607430S, RSO607538S, RSO607906S, RSO607907S, RSO607952S, RSO608008S, RSO608169S, RSO608515S, RSO608567S, RSO608601S, RSO608773S, RSO608775S, RSO609471S, RSO609616S, RSO609921S, RSO610614S, RSO610712S, RSO611995S, RSO612112S, RSO612170S, RSO612176S, RSO601935S, RSO603846S, RSO605987S, RSO607110S, RSO607287S, RSO608695S, RSO609255S, RSO606252S, RSO610862S, RSO612574S, RSO612575S, RSO607061S, RSO610396S, RSO610496S, RSO611747S, RSO612027S, RSO600641S, RSO600692S, RSO602815S, RSO603574S, RSO605191S, RSO605633S, RSO606897S, RSO607175S, RSO608173S, RSO612021S, RSO603753S, RSO612386S, RSO605231S, RSO605540S, RSO607033S, RSO609578S, RSO610646S, RSO611076S, RSO611077S, RSO611139S, RSO611140S, RSO611753S, RSO612204S, RSO612205S, RSO612235S, RSO600150S, RSO600154S, RSO600642S, RSO600711S, RSO600720S, RSO600729S, RSO601019S, RSO601038S, RSO601080S, RSO601142S, RSO601225S, RSO601696S, RSO601786S, RSO603348S, RSO603493S, RSO603699S, RSO603952S, RSO603962S, RSO604073S, RSO604656S, RSO604959S, RSO605030S, RSO605504S, RSO605604S, RSO605623S, RSO605735S, RSO606438S, RSO606617S, RSO606625S, RSO606659S, RSO606698S, RSO608217S, RSO607953S, RSO604084S, RSO608840S, RSO610824S, RSO601497S, RSO605202S, RSO610449S, RSO610450S, RSO610452S, RSO611827S, RSO611828S, RSO611829S, RSO609544S, RSO609561S, RSO609627S, RSO609690S, RSO609770S, RSO609827S, RSO610660S, RSO610911S, RSO611830S, RSO611831S, RSO611832S, RSO612123S, RSO612249S, RSO601101S, RSO601125S, RSO601703S, RSO605728S, RSO605846S, RSO606451S, RSO607349S, RSO607682S, RSO607824S, RSO608207S, RSO608324S, RSO608527S, RSO608986S, RSO609373S, RSO610005S, RSO610007S, RSO610008S, RSO610010S, RSO610141S, RSO610144S, RSO610148S, RSO610150S, RSO610157S, RSO610158S, RSO610159S, RSO610161S, RSO610163S, RSO611049S, RSO611052S, RSO611056S, RSO611057S, RSO611061S, RSO611062S, RSO611064S, RSO611069S, RSO611128S, RSO611129S, RSO611130S, RSO611131S, RSO611132S, RSO611133S, RSO611134S, RSO611135S, RSO611136S, RSO611137S, RSO602482S, RSO602483S, RSO602484S, RSO602488S, RSO602489S, RSO602491S, RSO604535S, RSO604538S, RSO604542S, RSO604543S, RSO604603S, RSO604605S, RSO604614S, RSO605983S, RSO607062S, RSO607064S, RSO607068S, RSO607488S, RSO607489S, RSO607490S, RSO607492S, RSO607494S, RSO607497S, RSO607862S, RSO608278S, RSO608279S, RSO608280S, RSO608461S, RSO608750S, RSO608751S, RSO609014S, RSO609015S, RSO609016S, RSO609202S, RSO609203S, RSO609204S, RSO609207S, RSO609209S, RSO609210S, RSO609211S, RSO609212S, RSO611050S, RSO611051S, RSO611054S, RSO611055S, RSO611058S, RSO611059S, RSO611060S, RSO611065S, RSO611066S, RSO611067S, RSO611068S, RSO611070S, RSO611071S, RSO611072S, RSO611907S, RSO611908S, RSO610377S, RSO612073S, RSO603655S, RSO605677S, RSO607152S, RSO607158S, RSO607160S, RSO608657S, RSO612709S, RSO609853S, RSO610415S, RSO601506S, RSO602362S, RSO602364S, RSO602365S, RSO602367S, RSO602369S, RSO602374S, RSO602375S, RSO602376S, RSO603101S, RSO606666S, RSO607438S, RSO607471S, RSO607521S, RSO607736S, RSO607757S, RSO607774S, RSO607818S, RSO607819S, RSO607822S, RSO608166S, RSO608167S, RSO608281S, RSO608284S, RSO608396S, RSO608402S, RSO608405S, RSO608406S, RSO609930S, RSO610978S, RSO611021S, RSO611022S, RSO611023S, RSO611024S, RSO611025S, RSO611026S, RSO611529S, RSO611530S, RSO611531S, RSO611532S, RSO611533S, RSO611534S, RSO611538S, RSO611539S, RSO611540S, RSO611541S, RSO611542S, RSO611543S, RSO605658S, RSO606406S, RSO610511S, RSO600391S, RSO603100S, RSO603113S, RSO603199S, RSO604713S, RSO604995S, RSO605147S, RSO607181S, RSO608767S, RSO600229S, RSO608562S, RSO608650S, RSO610908S, RSO601045S, RSO601293S, RSO601378S, RSO602522S, RSO604692S, RSO607089S, RSO600715S, RSO600790S, RSO600727S, RSO611585S, RSO601012S, RSO610002S, RSO610003S, RSO610004S, RSO610027S, RSO610030S, RSO610032S, RSO610033S, RSO610034S, RSO610663S, RSO610664S, RSO610689S, RSO610982S, RSO611696S, RSO611698S, RSO602979S, RSO604285S, RSO604287S, RSO605044S, RSO605056S, RSO605058S, RSO605064S, RSO605071S, RSO606578S, RSO606579S, RSO606581S, RSO607170S, RSO607337S, RSO607339S, RSO607442S, RSO607446S, RSO607450S, RSO607540S, RSO607543S, RSO607549S, RSO607556S, RSO607558S, RSO607560S, RSO608439S, RSO608453S, RSO608753S, RSO608754S, RSO609375S, RSO609377S, RSO609379S, RSO609381S, RSO610690S, RSO611005S, RSO611695S, RSO611697S, RSO611962S, RSO611963S, RSO611964S, RSO611965S, RSO611966S, RSO611967S, RSO611968S, RSO611969S, RSO611970S, RSO611971S, RSO600149S, RSO601688S, RSO601816S, RSO602326S, RSO602775S, RSO602929S, RSO602967S, RSO602973S, RSO603858S, RSO603859S, RSO603861S, RSO607709S, RSO608047S, RSO608996S, RSO609033S, RSO609036S, RSO609041S, RSO610733S, RSO600786S, RSO601444S, RSO601662S, RSO601900S, RSO601909S, RSO610120S, RSO610122S, RSO610125S, RSO610127S, RSO610128S, RSO610133S, RSO610135S, RSO611951S, RSO611952S, RSO611955S, RSO602494S, RSO602495S, RSO602498S, RSO602500S, RSO602506S, RSO602532S, RSO602533S, RSO602535S, RSO602536S, RSO602538S, RSO602539S, RSO602542S, RSO602546S, RSO602598S, RSO602601S, RSO602602S, RSO602606S, RSO602607S, RSO602608S, RSO602609S, RSO602610S, RSO602612S, RSO602627S, RSO602638S, RSO602640S, RSO602643S, RSO602646S, RSO602647S, RSO602650S, RSO602653S, RSO604680S, RSO604710S, RSO604717S, RSO604728S, RSO604734S, RSO604739S, RSO604755S, RSO604757S, RSO604774S, RSO605955S, RSO605959S, RSO605961S, RSO607183S, RSO607184S, RSO607185S, RSO607187S, RSO607188S, RSO607191S, RSO607192S, RSO607197S, RSO607209S, RSO607462S, RSO607474S, RSO607475S, RSO607476S, RSO607482S, RSO607487S, RSO607496S, RSO607499S, RSO607515S, RSO608228S, RSO608229S, RSO608230S, RSO608411S, RSO608416S, RSO608417S, RSO608418S, RSO608419S, RSO608420S, RSO608421S, RSO608423S, RSO608425S, RSO608429S, RSO608430S, RSO608431S, RSO608434S, RSO608435S, RSO608920S, RSO608921S, RSO608937S, RSO608938S, RSO609008S, RSO609009S, RSO609022S, RSO609023S, RSO609024S, RSO609121S, RSO609122S, RSO609123S, RSO609124S, RSO609126S, RSO609127S, RSO609128S, RSO609129S, RSO609131S, RSO609132S, RSO609133S, RSO609134S, RSO609135S, RSO609136S, RSO609137S, RSO609138S, RSO609139S, RSO609140S, RSO609141S, RSO609143S, RSO609144S, RSO609148S, RSO609149S, RSO609150S, RSO609152S, RSO609153S, RSO609154S, RSO609160S, RSO609245S, RSO609252S, RSO609253S, RSO609383S, RSO609384S, RSO609385S, RSO609386S, RSO609387S, RSO609389S, RSO609391S, RSO609392S, RSO610121S, RSO610123S, RSO610124S, RSO610126S, RSO610129S, RSO610130S, RSO610132S, RSO610134S, RSO610136S, RSO610137S, RSO610138S, RSO610981S, RSO611557S, RSO611558S, RSO611559S, RSO611560S, RSO611950S, RSO611953S, RSO611954S, RSO611956S, RSO611957S, RSO611958S, RSO611959S, RSO611961S, RSO600397S, RSO602493S, RSO602615S, RSO603061S, RSO603255S, RSO603744S, RSO603941S, RSO604098S, RSO605381S, RSO605503S, RSO606509S, RSO607491S, RSO609963S, RSO609972S, RSO609980S, RSO609986S, RSO609993S, RSO609994S, RSO610080S, RSO610081S, RSO610082S, RSO610083S, RSO610084S, RSO610085S, RSO610086S, RSO610087S, RSO610089S, RSO610090S, RSO610091S, RSO610093S, RSO610094S, RSO610096S, RSO610097S, RSO610098S, RSO610099S, RSO610100S, RSO610101S, RSO610102S, RSO610103S, RSO610104S, RSO610105S, RSO610107S, RSO611036S, RSO611037S, RSO611038S, RSO611039S, RSO611042S, RSO611043S, RSO601519S, RSO601520S, RSO603438S, RSO603439S, RSO603775S, RSO603776S, RSO604335S, RSO604738S, RSO604740S, RSO604744S, RSO604747S, RSO604749S, RSO604751S, RSO604754S, RSO604769S, RSO604773S, RSO604777S, RSO604785S, RSO604788S, RSO604791S, RSO604799S, RSO604800S, RSO604803S, RSO604810S, RSO604996S, RSO605000S, RSO605470S, RSO605471S, RSO605473S, RSO605476S, RSO605477S, RSO605722S, RSO605786S, RSO605848S, RSO606221S, RSO606222S, RSO606224S, RSO606225S, RSO606230S, RSO606231S, RSO607449S, RSO607451S, RSO607454S, RSO607455S, RSO607458S, RSO607461S, RSO608446S, RSO608449S, RSO608452S, RSO608468S, RSO608483S, RSO608533S, RSO608547S, RSO609218S, RSO609219S, RSO609220S, RSO609221S, RSO609222S, RSO609223S, RSO609224S, RSO609226S, RSO609227S, RSO609228S, RSO609229S, RSO609230S, RSO609231S, RSO609233S, RSO609235S, RSO610092S, RSO610095S, RSO610106S, RSO611000S, RSO611041S, RSO611044S, RSO611045S, RSO611046S, RSO611047S, RSO613235S, RSO603263S, RSO604334S, RSO604978S, RSO605233S, RSO610013S, RSO610014S, RSO610016S, RSO610017S, RSO610018S, RSO610110S, RSO610111S, RSO610112S, RSO610113S, RSO610114S, RSO610115S, RSO610117S, RSO610119S, RSO610976S, RSO610977S, RSO610979S, RSO610980S, RSO610983S, RSO610984S, RSO611079S, RSO611080S, RSO611085S, RSO611086S, RSO611089S, RSO611090S, RSO611091S, RSO611095S, RSO611096S, RSO611097S, RSO611099S, RSO611102S, RSO611105S, RSO611108S, RSO601318S, RSO601507S, RSO602617S, RSO603002S, RSO604243S, RSO604244S, RSO604355S, RSO604356S, RSO604566S, RSO604569S, RSO604949S, RSO604952S, RSO604954S, RSO605734S, RSO605738S, RSO605749S, RSO606488S, RSO606489S, RSO606490S, RSO606491S, RSO607083S, RSO607084S, RSO607085S, RSO607126S, RSO607127S, RSO607493S, RSO607500S, RSO607520S, RSO607531S, RSO607534S, RSO607599S, RSO607619S, RSO608080S, RSO608114S, RSO608271S, RSO608272S, RSO608274S, RSO608350S, RSO608355S, RSO610116S, RSO610118S, RSO611093S, RSO611109S, RSO609654S, RSO603943S, RSO603944S, RSO604661S, RSO604663S, RSO604664S, RSO604665S, RSO604668S, RSO604671S, RSO605001S, RSO605011S, RSO608035S, RSO608349S, RSO611523S, RSO600735S, RSO600803S, RSO610592S, RSO610593S, RSO610594S, RSO611016S, RSO602631S, RSO602632S, RSO603365S, RSO603366S, RSO603368S, RSO607547S, RSO607548S, RSO608226S, RSO609012S, RSO609013S, RSO609213S, RSO609217S, RSO600703S, RSO600903S, RSO601882S, RSO601899S, RSO602768S, RSO603221S, RSO608860S, RSO608864S, RSO604951S, RSO609800S, RSO603016S, RSO609487S, RSO609501S, RSO609540S, RSO611019S, RSO608120S, RSO608121S, RSO608122S, RSO608124S, RSO608125S, RSO609254S, RSO609450S, RSO611001S, RSO611002S, RSO611034S, RSO611035S, RSO612349S, RSO612896S, RSO612897S, RSO612898S, RSO612901S, RSO612902S, RSO611604S, RSO607941S, RSO608741S, RSO606552S, RSO608501S, RSO610695S, RSO611006S, RSO611007S, RSO611008S, RSO611009S, RSO611010S, RSO611198S, RSO611199S, RSO611200S, RSO611201S, RSO611204S, RSO611205S, RSO611206S, RSO611207S, RSO611209S, RSO611210S, RSO611211S, RSO611213S, RSO611214S, RSO611215S, RSO604337S, RSO604338S, RSO604339S, RSO604340S, RSO604341S, RSO604342S, RSO604343S, RSO604344S, RSO604345S, RSO604405S, RSO604406S, RSO604407S, RSO604408S, RSO604409S, RSO604410S, RSO604411S, RSO604412S, RSO604629S, RSO604633S, RSO604647S, RSO604649S, RSO607452S, RSO607680S, RSO608100S, RSO608109S, RSO608154S, RSO608373S, RSO608427S, RSO608436S, RSO608437S, RSO608438S, RSO608441S, RSO608442S, RSO608445S, RSO608460S, RSO608506S, RSO608507S, RSO609112S, RSO609113S, RSO609114S, RSO609115S, RSO609116S, RSO609117S, RSO609118S, RSO609119S, RSO610697S, RSO611203S, RSO611208S, RSO611212S, RSO611216S, RSO611217S, RSO611889S, RSO611890S, RSO611891S, RSO611892S, RSO611893S, RSO611909S, RSO611910S, RSO611911S, RSO611912S, RSO611913S, RSO611914S, RSO611915S, RSO611916S, RSO611917S, RSO611918S, RSO611919S, RSO611920S, RSO611921S, RSO611922S, RSO611923S, RSO612286S, RSO612287S, RSO612288S, RSO612289S, RSO612290S, RSO612291S, RSO612802S, RSO612804S, RSO612805S, RSO612806S, RSO612807S, RSO612808S, RSO612809S, RSO612810S, RSO612811S, RSO612813S, RSO612814S, RSO612815S, RSO612816S, RSO612817S, RSO612818S, RSO612819S, RSO612820S, RSO612821S, RSO612822S, RSO612823S, RSO612824S, RSO612826S, RSO612827S, RSO612828S, RSO612829S, RSO612831S, RSO612832S, RSO612833S, RSO612834S, RSO612835S, RSO612836S, RSO612837S, RSO612838S, RSO612839S, RSO612840S, RSO612841S, RSO612842S, RSO612843S, RSO612844S, RSO612845S, RSO612848S, RSO612851S, RSO612856S, RSO612857S, RSO612866S, RSO612867S, RSO612870S, RSO612873S, RSO612877S, RSO609111S, RSO610666S, RSO610671S, RSO610672S, RSO610674S, RSO610675S, RSO610676S, RSO607218S, RSO607221S, RSO607222S, RSO607223S, RSO607224S, RSO607227S, RSO607231S, RSO607233S, RSO607271S, RSO607272S, RSO607273S, RSO607274S, RSO607276S, RSO607277S, RSO607280S, RSO607285S, RSO607286S, RSO607288S, RSO607292S, RSO607293S, RSO607296S, RSO607299S, RSO607300S, RSO607304S, RSO607316S, RSO609247S, RSO609248S, RSO609250S, RSO609251S, RSO610667S, RSO610673S, RSO609717S, RSO604376S, RSO604377S, RSO611663S, RSO611975S, RSO611976S, RSO611977S, RSO609586S, RSO609764S, RSO609765S, RSO609766S, RSO609767S, RSO609850S, RSO609851S, RSO609866S, RSO610019S, RSO610021S, RSO610184S, RSO610185S, RSO610186S, RSO610187S, RSO610693S, RSO610694S, RSO611030S, RSO611031S, RSO611032S, RSO611033S, RSO611320S, RSO611525S, RSO611526S, RSO611527S, RSO611546S, RSO611652S, RSO611653S, RSO611654S, RSO611810S, RSO611811S, RSO612294S, RSO612295S, RSO612381S, RSO603031S, RSO603041S, RSO603051S, RSO603328S, RSO603394S, RSO603396S, RSO603920S, RSO603921S, RSO603936S, RSO604279S, RSO607372S, RSO607373S, RSO607592S, RSO607593S, RSO607595S, RSO607753S, RSO607755S, RSO607841S, RSO607842S, RSO608164S, RSO608165S, RSO608265S, RSO608269S, RSO608270S, RSO608584S, RSO608930S, RSO608931S, RSO608974S, RSO608976S, RSO608977S, RSO608978S, RSO609092S, RSO609094S, RSO609103S, RSO609105S, RSO609206S, RSO609397S, RSO609398S, RSO611803S, RSO612318S, RSO612382S, RSO612753S, RSO612754S, RSO612765S, RSO612766S, RSO612767S, RSO612768S, RSO612769S, RSO612770S, RSO611944S, RSO611945S, RSO611946S, RSO611947S, RSO611949S, RSO611948S, RSO601579S, RSO612138S, RSO612139S, RSO607642S, RSO602790S, RSO602791S, RSO602793S, RSO602795S, RSO602796S, RSO603084S, RSO603325S, RSO603326S, RSO603329S, RSO603330S, RSO603332S, RSO603333S, RSO603336S, RSO603355S, RSO603356S, RSO603358S, RSO603361S, RSO603362S, RSO603415S, RSO603417S, RSO603419S, RSO603420S, RSO604267S, RSO604273S, RSO604906S, RSO604907S, RSO604908S, RSO604931S, RSO604946S, RSO604958S, RSO609619S, RSO609709S, RSO609937S, RSO610669S, RSO610670S, RSO610732S, RSO611662S, RSO612101S, RSO600157S, RSO600162S, RSO603485S, RSO603513S, RSO603537S, RSO603823S, RSO605284S, RSO605383S, RSO607026S, RSO607040S, RSO607596S, RSO607763S, RSO607764S, RSO607914S, RSO608225S, RSO608910S, RSO608911S, RSO600750S, RSO600870S, RSO604596S, RSO605385S, RSO612236S, RSO603241S, RSO610548S, RSO610934S, RSO600868S, RSO604570S, RSO605139S, RSO606407S, RSO608216S, RSO609264S, RSO610964S, RSO611595S, RSO611769S, RSO603679S, RSO604181S, RSO604196S, RSO604224S, RSO605240S, RSO605370S, RSO605379S, RSO606360S, RSO606379S, RSO610022S, RSO610023S, RSO610025S, RSO610108S, RSO610109S, RSO604987S, RSO605764S, RSO605964S, RSO606103S, RSO606486S, RSO606767S, RSO607392S, RSO607394S, RSO607545S, RSO607546S, RSO607728S, RSO607734S, RSO608286S, RSO609063S, RSO609414S, RSO611822S, RSO611838S, RSO600723S, RSO601034S, RSO601036S, RSO601257S, RSO601708S, RSO601904S, RSO603024S, RSO603406S, RSO605436S, RSO610595S, RSO610596S, RSO611017S, RSO611018S, RSO600120S, RSO610867S, RSO603119S, RSO609756S, RSO607622S, RSO610041S, RSO610042S, RSO610045S, RSO610046S, RSO610047S, RSO610049S, RSO610053S, RSO610054S, RSO610055S, RSO610056S, RSO610057S, RSO610059S, RSO610060S, RSO610061S, RSO610062S, RSO610063S, RSO610064S, RSO610065S, RSO610066S, RSO610686S, RSO610687S, RSO610688S, RSO611012S, RSO611013S, RSO611014S, RSO611544S, RSO611545S, RSO611555S, RSO611556S, RSO611562S, RSO611563S, RSO611682S, RSO611683S, RSO611684S, RSO611685S, RSO611686S, RSO611687S, RSO611693S, RSO611694S, RSO611845S, RSO611849S, RSO611852S, RSO611853S, RSO611854S, RSO611856S, RSO611858S, RSO611863S, RSO611865S, RSO611878S, RSO611879S, RSO611880S, RSO611883S, RSO601442S, RSO601700S, RSO605523S, RSO605525S, RSO605529S, RSO605531S, RSO605534S, RSO605538S, RSO606189S, RSO606440S, RSO606442S, RSO606599S, RSO606600S, RSO606742S, RSO606746S, RSO606928S, RSO606930S, RSO606931S, RSO606979S, RSO607308S, RSO607310S, RSO607311S, RSO607312S, RSO607314S, RSO607425S, RSO607426S, RSO607434S, RSO607435S, RSO607437S, RSO607973S, RSO607974S, RSO607975S, RSO608239S, RSO608240S, RSO608241S, RSO608243S, RSO608244S, RSO608251S, RSO608252S, RSO608260S, RSO608263S, RSO608266S, RSO608267S, RSO608268S, RSO608273S, RSO608277S, RSO608282S, RSO608283S, RSO608285S, RSO608289S, RSO608290S, RSO608293S, RSO608298S, RSO608300S, RSO608301S, RSO608306S, RSO608307S, RSO608308S, RSO608311S, RSO609020S, RSO609021S, RSO609175S, RSO609176S, RSO609178S, RSO609179S, RSO609180S, RSO609181S, RSO609182S, RSO609184S, RSO609185S, RSO609186S, RSO609187S, RSO609188S, RSO609189S, RSO609190S, RSO609191S, RSO609192S, RSO609193S, RSO609196S, RSO609197S, RSO609198S, RSO611817S, RSO611818S, RSO611841S, RSO611842S, RSO611843S, RSO611844S, RSO611846S, RSO611847S, RSO611848S, RSO611850S, RSO611851S, RSO611855S, RSO611857S, RSO611859S, RSO611860S, RSO611861S, RSO611862S, RSO611864S, RSO611866S, RSO611867S, RSO611868S, RSO611869S, RSO611870S, RSO611871S, RSO611872S, RSO611873S, RSO611874S, RSO611875S, RSO611876S, RSO611877S, RSO611881S, RSO611882S, RSO611884S, RSO605142S, RSO606423S, RSO606424S, RSO610821S, RSO611063S; GTIN 00763000178406, Lot Serial Numbers: RSO609039S, RSO600386S, RSO600452S, RSO604660S, RSO604662S, RSO612464S, RSO609687S, RSO609512S, RSO610411S, RSO613299S, RSO606800S, RSO613055S, RSO613319S, RSO612872S, RSO613563S, RSO611396S, RSO611324S, RSO613027S, RSO612879S, RSO611412S, RSO612450S, RSO613097S, RSO611269S, RSO610267S, RSO613322S, RSO610009S, RSO613004S, RSO612581S, RSO612996S, RSO613377S, RSO600286S, RSO600529S, RSO600537S, RSO600539S, RSO600541S, RSO604690S, RSO604691S, RSO606765S, RSO606842S, RSO606844S, RSO606845S, RSO606846S, RSO606881S, RSO606883S, RSO606884S, RSO606935S, RSO608960S, RSO609467S, RSO609483S, RSO610015S, RSO610261S, RSO611236S, RSO611367S, RSO611437S, RSO611598S, RSO612376S, RSO613037S, RSO613450S, RSO609401S, RSO613335S, RSO604796S, RSO610822S, RSO606091S, RSO608858S, RSO611117S, RSO606156S, RSO602670S, RSO604655S, RSO604657S, RSO604658S, RSO600322S, RSO600325S, RSO600507S, RSO602594S, RSO606066S, RSO606780S, RSO610726S, RSO611237S, RSO606037S, RSO608869S, RSO610051S, RSO611450S, RSO613017S, RSO611372S, RSO606672S, RSO608874S, RSO608857S, RSO609681S, RSO600289S, RSO600290S, RSO600411S, RSO602409S, RSO602730S, RSO604823S, RSO606093S, RSO606399S, RSO606715S, RSO606729S, RSO602429S, RSO611406S, RSO602682S, RSO608861S, RSO602368S, RSO606196S, RSO607821S, RSO608847S, RSO610224S, RSO611454S, RSO605546S, RSO610212S, RSO602694S, RSO609433S, RSO609524S, RSO610217S, RSO611172S, RSO611277S, RSO611278S, RSO611403S, RSO612152S, RSO612395S, RSO612457S, RSO613096S, RSO606003S, RSO610218S, RSO610748S, RSO611272S, RSO611460S, RSO612434S, RSO612528S, RSO612884S, RSO613354S, RSO600280S, RSO602712S, RSO602713S, RSO604672S, RSO606129S, RSO606315S, RSO606678S, RSO606699S, RSO606787S, RSO606909S, RSO609066S, RSO609082S, RSO609407S, RSO610246S, RSO610257S, RSO610258S, RSO606403S, RSO606807S, RSO608985S, RSO611267S, RSO613366S, RSO600456S, RSO600464S, RSO600467S, RSO600502S, RSO600505S, RSO600506S, RSO602467S, RSO602620S, RSO602676S, RSO602677S, RSO604826S, RSO604828S, RSO604830S, RSO605778S, RSO605920S, RSO606686S, RSO606687S, RSO606702S, RSO608862S, RSO609030S, RSO609338S, RSO610028S, RSO610029S, RSO610031S, RSO612436S, RSO612437S, RSO612438S, RSO613046S, RSO613047S, RSO613048S, RSO604640S, RSO604641S, RSO604642S, RSO600293S, RSO600350S, RSO600583S, RSO602306S, RSO602353S, RSO602428S, RSO602475S, RSO602633S, RSO604646S, RSO606082S, RSO606198S, RSO606821S, RSO609395S, RSO609468S, RSO609485S, RSO610253S, RSO611249S, RSO611253S, RSO611301S, RSO612400S, RSO612439S, RSO613000S, RSO613024S, RSO600042S, RSO600519S, RSO611202S, RSO600025S, RSO600055S, RSO602421S, RSO611321S, RSO611322S, RSO611323S, RSO611397S, RSO611400S, RSO611401S, RSO611411S, RSO600053S, RSO609470S, RSO600029S, RSO600036S, RSO600039S, RSO600304S, RSO600307S, RSO602312S, RSO604092S, RSO604636S, RSO604871S, RSO606038S, RSO606705S, RSO606731S, RSO606818S, RSO606829S, RSO606912S, RSO606936S, RSO607812S, RSO609431S, RSO609949S, RSO610271S, RSO610763S, RSO611310S, RSO611383S, RSO611444S, RSO611449S, RSO613003S, RSO613039S, RSO613043S, RSO600023S, RSO600046S, RSO600288S, RSO600314S, RSO600317S, RSO602426S, RSO602437S, RSO602654S, RSO602681S, RSO602715S, RSO602752S, RSO604693S, RSO604722S, RSO604748S, RSO604822S, RSO604861S, RSO604882S, RSO606073S, RSO606108S, RSO606472S, RSO606692S, RSO606908S, RSO606929S, RSO608819S, RSO609319S, RSO609380S, RSO609382S, RSO609447S, RSO610755S, RSO610770S, RSO611239S, RSO611281S, RSO611325S, RSO611328S, RSO611346S, RSO611371S, RSO611404S, RSO611445S, RSO611478S, RSO612208S, RSO612429S, RSO612458S, RSO609075S, RSO610238S, RSO600043S, RSO600045S, RSO600175S, RSO600344S, RSO600396S, RSO602422S, RSO604632S, RSO604684S, RSO606142S, RSO606695S, RSO606719S, RSO609104S, RSO609429S, RSO610817S, RSO611423S, RSO611442S, RSO602698S, RSO602699S, RSO604762S, RSO606030S, RSO606032S, RSO611357S, RSO611359S, RSO609344S, RSO600575S, RSO600576S, RSO602477S, RSO604753S, RSO604759S, RSO604837S, RSO604840S, RSO606017S, RSO606028S, RSO611284S, RSO611358S, RSO611360S, RSO611362S, RSO602372S, RSO602379S, RSO602512S, RSO602648S, RSO602664S, RSO602680S, RSO602692S, RSO602761S, RSO604686S, RSO604696S, RSO604704S, RSO604725S, RSO604792S, RSO604852S, RSO606202S, RSO606940S, RSO608477S, RSO608849S, RSO609157S, RSO609166S, RSO609367S, RSO609388S, RSO609481S, RSO609947S, RSO610239S, RSO610273S, RSO611228S, RSO611345S, RSO611350S, RSO611385S, RSO611420S, RSO611425S, RSO611481S, RSO613321S, RSO613349S, RSO600370S, RSO600373S, RSO600389S, RSO600412S, RSO600443S, RSO600446S, RSO600570S, RSO602320S, RSO602400S, RSO602443S, RSO602671S, RSO604683S, RSO604795S, RSO606205S, RSO606890S, RSO604643S, RSO600022S, RSO600362S, RSO600380S, RSO600383S, RSO600432S, RSO600474S, RSO600475S, RSO602335S, RSO602352S, RSO602384S, RSO602387S, RSO602402S, RSO602403S, RSO602427S, RSO602465S, RSO602652S, RSO602657S, RSO602658S, RSO602679S, RSO604801S, RSO604819S, RSO604853S, RSO604854S, RSO606079S, RSO606172S, RSO606175S, RSO606453S, RSO606454S, RSO606664S, RSO606740S, RSO606774S, RSO606892S, RSO609072S, RSO609156S, RSO609442S, RSO609477S, RSO610215S, RSO610227S, RSO610775S, RSO610809S, RSO611222S, RSO611260S, RSO611381S, RSO611408S, RSO611417S, RSO611743S, RSO612403S, RSO612449S, RSO612859S, RSO612881S, RSO612998S, RSO612999S, RSO613012S, RSO613041S, RSO613056S, RSO613063S, RSO613370S, RSO609089S, RSO604809S, RSO609364S, RSO606139S, RSO600345S, RSO600485S, RSO600491S, RSO600493S, RSO600582S, RSO602385S, RSO602391S, RSO602414S, RSO602416S, RSO602661S, RSO602668S, RSO602714S, RSO606358S, RSO606675S, RSO606927S, RSO600528S, RSO602386S, RSO606109S, RSO606901S, RSO609324S, RSO602523S, RSO602540S, RSO608820S, RSO608848S, RSO609952S, RSO600006S, RSO604731S, RSO608853S, RSO609521S, RSO609525S, RSO610769S, RSO611327S, RSO600027S, RSO600067S, RSO600402S, RSO600521S, RSO602691S, RSO604746S, RSO606119S, RSO606876S, RSO609028S, RSO609475S, RSO610229S, RSO611242S, RSO612951S, RSO600073S, RSO609437S, RSO608475S, RSO609311S, RSO609507S, RSO611246S, RSO612409S, RSO609511S, RSO609321S, RSO609322S, RSO600319S, RSO600484S, RSO600072S, RSO606903S, RSO607754S, RSO607814S, RSO609469S, RSO610233S, RSO610270S, RSO602360S, RSO602363S, RSO602366S, RSO604644S, RSO604645S, RSO604707S, RSO604879S, RSO606098S, RSO606676S, RSO606910S, RSO608518S, RSO608852S, RSO608877S, RSO609035S, RSO609088S, RSO609194S, RSO609310S, RSO609339S, RSO609361S, RSO609459S, RSO610242S, RSO604712S, RSO604720S, RSO606149S, RSO606350S, RSO606387S, RSO609357S, RSO609362S, RSO611273S, RSO611274S, RSO600316S, RSO602322S, RSO602327S, RSO602445S, RSO602674S, RSO602675S, RSO604631S, RSO604675S, RSO604771S, RSO606694S, RSO606906S, RSO606937S, RSO609448S, RSO610222S, RSO610254S, RSO610262S, RSO611247S, RSO611279S, RSO611473S, RSO600523S, RSO600525S, RSO602381S, RSO602462S, RSO602656S, RSO602689S, RSO606087S, RSO602338S, RSO606636S, RSO606713S, RSO606726S, RSO609390S, RSO610774S, RSO611485S, RSO600532S, RSO600534S, RSO604745S, RSO610263S, RSO611248S, RSO602461S, RSO602473S, RSO604790S, RSO609365S, RSO609400S, RSO611229S, RSO611370S, RSO611443S, RSO611492S, RSO611813S, RSO612455S, RSO604848S, RSO602434S, RSO604772S, RSO610241S, RSO611413S, RSO611447S, RSO602430S, RSO602431S, RSO602433S, RSO602672S, RSO602762S, RSO604607S, RSO604701S, RSO604732S, RSO604864S, RSO606479S, RSO606643S, RSO606707S, RSO606812S, RSO610231S, RSO610904S, RSO611264S, RSO611331S, RSO611343S, RSO606458S, RSO606763S, RSO609443S, RSO611304S, RSO600387S, RSO600405S, RSO600421S, RSO600425S, RSO600433S, RSO600479S, RSO600577S, RSO602456S, RSO602472S, RSO604651S, RSO604782S, RSO606138S, RSO606671S, RSO606703S, RSO610274S, RSO611286S, RSO611441S, RSO612423S, RSO612459S, RSO611225S, RSO611226S, RSO602436S, RSO600324S, RSO600483S, RSO602411S, RSO604877S, RSO609086S, RSO600509S, RSO600513S, RSO602399S, RSO602499S, RSO602515S, RSO602687S, RSO604635S, RSO606690S, RSO606730S, RSO611218S, RSO611430S, RSO613058S, RSO609436S, RSO611451S, RSO611452S, RSO612443S, RSO612444S, RSO612445S, RSO612446S, RSO612447S, RSO612852S, RSO612853S, RSO613184S, RSO613298S, RSO613297S, RSO613307S, RSO600413S, RSO600414S, RSO602383S, RSO602441S, RSO602662S, RSO602665S, RSO604750S, RSO604850S, RSO604862S, RSO606137S, RSO609349S, RSO609435S, RSO610174S, RSO610223S, RSO611250S, RSO611251S, RSO611379S, RSO612431S, RSO613013S, RSO613014S, RSO602468S, RSO602469S, RSO606063S, RSO606083S, RSO608867S, RSO609077S, RSO611469S, RSO611470S, RSO611471S, RSO611472S, RSO612418S, RSO612419S, RSO612420S, RSO612421S, RSO606665S, RSO610245S, RSO606804S, RSO609074S, RSO609486S, RSO610268S, RSO611257S, RSO611335S, RSO611410S, RSO611446S, RSO611474S, RSO612452S, RSO613065S, RSO604609S, RSO606794S, RSO609460S, RSO610225S, RSO611223S, RSO611252S, RSO611463S, RSO612451S, RSO612847S, RSO613002S, RSO602424S, RSO602439S, RSO604089S, RSO604855S, RSO606638S, RSO606710S, RSO609446S, RSO610219S, RSO600343S, RSO602382S, RSO602404S, RSO602405S, RSO602406S, RSO602423S, RSO602447S, RSO602690S, RSO604727S, RSO604817S, RSO604818S, RSO606068S, RSO606176S, RSO609067S, RSO609083S, RSO609162S, RSO609337S, RSO610272S, RSO610751S, RSO611275S, RSO611276S, RSO611344S, RSO611477S, RSO612121S, RSO612456S, RSO613312S, RSO613374S, RSO606102S, RSO606169S, RSO609034S, RSO609038S, RSO609060S, RSO609065S, RSO609214S, RSO609329S, RSO609330S, RSO609465S, RSO610234S, RSO610235S, RSO611262S, RSO611268S, RSO611270S, RSO611424S, RSO600040S, RSO600048S, RSO602513S, RSO604843S, RSO606586S, RSO600051S, RSO609445S, RSO611254S, RSO611483S, RSO600297S, RSO602337S, RSO602419S, RSO602460S, RSO602727S, RSO604688S, RSO604689S, RSO606736S, RSO600335S, RSO600342S, RSO602380S, RSO602442S, RSO602622S, RSO608878S, RSO609363S, RSO609520S, RSO611342S, RSO600564S, RSO600568S, RSO602459S, RSO604087S, RSO604833S, RSO604884S, RSO606077S, RSO606112S, RSO606164S, RSO606679S, RSO606811S, RSO606859S, RSO607816S, RSO608863S, RSO608970S, RSO609078S, RSO609359S, RSO609531S, RSO610213S, RSO610214S, RSO610779S, RSO610788S, RSO611256S, RSO611480S, RSO613040S, RSO613057S, RSO613059S, RSO606065S, RSO607785S, RSO607790S, RSO607802S, RSO609327S, RSO609328S, RSO610208S, RSO610721S, RSO611219S, RSO611224S, RSO611230S, RSO611332S, RSO600049S, RSO600064S, RSO606053S, RSO611336S, RSO600346S, RSO600074S, RSO600333S, RSO604706S, RSO604733S, RSO604832S, RSO606117S, RSO606728S, RSO608823S, RSO608933S, RSO610256S, RSO610789S, RSO611476S, RSO600062S, RSO602307S, RSO602466S, RSO602476S, RSO604741S, RSO604885S, RSO606775S, RSO609476S, RSO610244S, RSO611266S, RSO611326S, RSO611465S, RSO612980S, RSO611434S, RSO602471S, RSO600330S, RSO609316S, RSO600284S, RSO600331S, RSO600374S, RSO600375S, RSO600517S, RSO602359S, RSO602415S, RSO602669S, RSO602749S, RSO604090S, RSO604715S, RSO604845S, RSO606086S, RSO606446S, RSO606878S, RSO609079S, RSO609151S, RSO609331S, RSO609345S, RSO610236S, RSO610237S, RSO611240S, RSO611271S, RSO611309S, RSO611382S, RSO611402S, RSO612402S, RSO613330S, RSO602345S, RSO609366S, RSO601555S, RSO602678S, RSO604780S, RSO606911S, RSO609458S, RSO609472S, RSO610220S, RSO600276S, RSO600283S, RSO600296S, RSO602401S, RSO602410S, RSO602412S, RSO602463S, RSO602716S, RSO604667S, RSO604695S, RSO606714S, RSO606725S, RSO606737S, RSO606764S, RSO606788S, RSO606896S, RSO607759S, RSO609147S, RSO609200S, RSO609496S, RSO611241S, RSO611282S, RSO611316S, RSO611416S, RSO600355S, RSO602486S, RSO602751S, RSO604770S, RSO604806S, RSO606263S, RSO606744S, RSO609145S, RSO610986S, RSO611227S, RSO611349S, RSO612428S, RSO602487S, RSO602497S, RSO602725S, RSO604650S, RSO611285S, RSO612868S, RSO613060S, RSO604873S, RSO604875S, RSO606717S, RSO606718S, RSO606785S, RSO609351S, RSO609352S, RSO609353S, RSO609798S, RSO609941S, RSO610228S, RSO610727S, RSO610737S, RSO610738S, RSO610739S, RSO610743S, RSO611233S, RSO611234S, RSO611337S, RSO611338S, RSO611339S, RSO611340S, RSO611427S, RSO611428S, RSO611660S, RSO611763S, RSO611791S, RSO612405S, RSO612406S, RSO612994S, RSO612995S, RSO613332S, RSO613333S, RSO613364S, RSO613365S, RSO609309S, RSO606831S, RSO609317S, RSO609318S, RSO611626S, RSO611633S, RSO611658S, RSO613049S, RSO613050S, RSO602697S, RSO604742S, RSO604804S, RSO606757S, RSO609323S, RSO602392S, RSO606720S, RSO606748S, RSO600012S, RSO600024S, RSO600026S, RSO600028S, RSO602502S, RSO602504S, RSO602508S, RSO602773S, RSO602780S, RSO602792S, RSO606039S, RSO609093S, RSO609125S, RSO609161S, RSO610210S, RSO610211S, RSO610275S, RSO610276S, RSO611438S, RSO611439S, RSO611440S, RSO613021S, RSO613568S, RSO613569S, RSO609998S, RSO602695S, RSO604719S, RSO606745S, RSO604868S, RSO606668S, RSO600498S, RSO602453S, RSO606162S, RSO606412S, RSO606727S, RSO600020S, RSO600031S, RSO600032S, RSO600033S, RSO600034S, RSO600035S, RSO600075S, RSO600321S, RSO602390S, RSO602420S, RSO602455S, RSO602470S, RSO602667S, RSO602696S, RSO602706S, RSO604721S, RSO604836S, RSO606200S, RSO606723S, RSO606753S, RSO606755S, RSO606760S, RSO608830S, RSO608832S, RSO609032S, RSO609040S, RSO602343S, RSO602349S, RSO602351S, RSO602458S, RSO602623S, RSO602693S, RSO602704S, RSO602708S, RSO602709S, RSO602726S, RSO602769S, RSO602772S, RSO604724S, RSO606042S, RSO606115S, RSO606701S, RSO606721S, RSO606733S, RSO606791S, RSO608834S, RSO610398S, RSO610818S, RSO610820S, RSO611296S, RSO611317S, RSO611329S, RSO612271S, RSO612398S, RSO611421S, RSO613329S, RSO602393S, RSO602396S, RSO604866S, RSO608859S, RSO602474S, RSO606415S, RSO606449S, RSO606684S, RSO606722S, RSO600406S, RSO604768S, RSO609053S, RSO610723S, RSO610829S, RSO600017S, RSO600018S, RSO600558S, RSO606199S, RSO600327S, RSO604673S, RSO604778S, RSO606712S, RSO602659S, RSO604711S, RSO610221S, RSO606181S, RSO602356S, RSO602446S, RSO606070S, RSO600050S, RSO600326S, RSO600328S, RSO600339S, RSO600431S, RSO604702S, RSO604865S, RSO604883S, RSO606052S, RSO608851S, RSO609019S, RSO602686S, RSO602717S, RSO606204S, RSO600007S, RSO600076S, RSO600357S, RSO600430S, RSO604677S, RSO604765S, RSO604811S, RSO606792S, RSO609942S, RSO611231S, RSO602673S, RSO602688S, RSO604676S, RSO604789S, RSO609426S, RSO611184S, RSO600434S, RSO602511S, RSO609343S, RSO611308S, RSO609315S, RSO611429S, RSO600277S, RSO600278S, RSO600281S, RSO600302S, RSO600384S, RSO606823S, RSO609056S, RSO609513S, RSO609515S, RSO609955S, RSO610860S, RSO611244S, RSO611245S, RSO612407S, RSO612408S, RSO604639S, RSO604652S, RSO604758S, RSO606019S, RSO606040S, RSO607810S, RSO609945S, RSO611369S, RSO610438S, RSO610441S, RSO610447S, RSO610514S, RSO610517S, RSO609978S, RSO610035S, RSO610369S, RSO606769S, RSO606771S, RSO606773S, RSO609943S, RSO609944S, RSO609964S, RSO610518S, RSO610001S, RSO608829S, RSO609314S, RSO609456S, RSO609457S, RSO609461S, RSO609965S, RSO609967S, RSO609968S, RSO609969S, RSO609973S, RSO609974S, RSO609975S, RSO609976S, RSO609979S, RSO609981S, RSO609982S, RSO609983S, RSO609984S, RSO609985S, RSO609987S, RSO609988S, RSO609989S, RSO609990S, RSO609991S, RSO609995S, RSO609997S, RSO604094S, RSO604095S, RSO604096S, RSO604097S, RSO604099S, RSO604100S, RSO604101S, RSO604102S, RSO604103S, RSO604104S, RSO604106S, RSO604108S, RSO604110S, RSO604113S, RSO604114S, RSO604119S, RSO604120S, RSO604554S, RSO604572S, RSO604573S, RSO604574S, RSO604576S, RSO604584S, RSO604597S, RSO604604S, RSO600285S, RSO612424S, RSO600378S, RSO602388S, RSO602448S, RSO604709S, RSO604824S, RSO609409S, RSO609473S, RSO612981S, RSO606872S, RSO608822S, RSO608831S, RSO600500S, RSO600580S, RSO600581S, RSO606833S, RSO606837S, RSO606853S, RSO606861S, RSO608821S, RSO609536S, RSO609568S, RSO609585S, RSO609662S, RSO609683S, RSO600287S, RSO600352S, RSO600353S, RSO600354S, RSO600497S, RSO600499S, RSO600510S, RSO600559S, RSO600560S, RSO600571S, RSO600579S, RSO606840S, RSO606854S, RSO606862S, RSO606869S, RSO608762S, RSO608966S, RSO609535S, RSO609539S, RSO609541S, RSO609569S, RSO609579S, RSO609582S, RSO609590S, RSO609694S, RSO606207S, RSO606843S, RSO608837S, RSO609320S, RSO604859S, RSO604899S, RSO606824S, RSO608833S, RSO609425S, RSO606206S, RSO606208S, RSO606817S, RSO606858S, RSO608836S, RSO609336S, RSO609404S, RSO610838S, RSO606210S, RSO600351S, RSO600461S, RSO600462S, RSO600511S, RSO606850S, RSO606855S, RSO606863S, RSO606865S, RSO608827S, RSO608846S, RSO609543S, RSO600552S, RSO600574S, RSO606860S, RSO609555S, RSO609584S, RSO609656S, RSO608825S, RSO609537S, RSO609550S, RSO602457S, RSO600347S, RSO600348S, RSO600470S, RSO609787S, RSO609788S, RSO609789S, RSO609790S, RSO609791S, RSO609792S, RSO610287S, RSO610288S, RSO610289S, RSO610290S, RSO610291S, RSO610292S, RSO610293S, RSO610294S, RSO610295S, RSO610296S, RSO610297S, RSO610531S, RSO610814S, RSO610909S, RSO610933S, RSO613434S, RSO613435S, RSO613439S, RSO613440S, RSO613441S, RSO604634S, RSO604723S, RSO604763S, RSO604812S, RSO604816S, RSO604834S, RSO604835S, RSO604844S, RSO606716S, RSO606738S, RSO606761S, RSO606934S, RSO608835S, RSO608838S, RSO608982S, RSO609312S, RSO609313S, RSO609422S, RSO609424S, RSO609441S, RSO609462S, RSO610209S, RSO610725S, RSO611300S, RSO600054S, RSO600068S, RSO600069S, RSO600070S, RSO600308S, RSO600310S, RSO600312S, RSO602425S, RSO602685S, RSO604091S, RSO604679S, RSO604781S, RSO604808S, RSO606806S, RSO606926S, RSO608470S, RSO608971S, RSO609101S, RSO609432S, RSO609474S, RSO610226S, RSO610998S, RSO611258S, RSO611259S, RSO611305S, RSO611418S, RSO611457S, RSO611475S, RSO612351S, RSO613026S, RSO602683S, RSO605921S, RSO606696S, RSO606743S, RSO606798S, RSO606891S, RSO606938S, RSO609159S, RSO609174S, RSO609340S, RSO609341S, RSO609354S, RSO609355S, RSO609451S, RSO609463S, RSO609464S, RSO609482S, RSO610044S, RSO610048S, RSO610050S, RSO610232S, RSO611290S, RSO611294S, RSO611307S, RSO611312S, RSO611313S, RSO611314S, RSO611315S, RSO611348S, RSO611459S, RSO611464S, RSO611484S, RSO611592S, RSO612864S, RSO612865S, RSO613015S, RSO600041S, RSO600279S, RSO606825S, RSO606827S, RSO606857S, RSO609957S, RSO608875S, RSO600420S, RSO604637S, RSO606822S, RSO609946S, RSO610266S, RSO610792S, RSO612849S, RSO600546S, RSO600551S, RSO604878S, RSO606141S, RSO606752S, RSO600014S, RSO600015S, RSO600021S, RSO600399S, RSO600409S, RSO602464S, RSO604896S, RSO604898S, RSO611185S, RSO611456S, RSO611494S, RSO611495S, RSO611496S, RSO606784S, RSO610989S, RSO613407S, RSO613408S, RSO613409S, RSO613410S, RSO609411S, RSO610440S, RSO610451S, RSO610458S, RSO610467S, RSO610500S, RSO610793S, RSO610795S, RSO610803S, RSO610805S, RSO610806S, RSO610810S, RSO610812S, RSO610359S, RSO610367S, RSO606810S, RSO606813S, RSO606815S, RSO611048S, RSO611153S, RSO611174S, RSO611515S, RSO611516S, RSO604892S, RSO610954S, RSO613405S, RSO606783S, RSO604888S, RSO604889S, RSO604891S, RSO609439S, RSO610368S, RSO610382S, RSO610394S, RSO610434S, RSO610437S, RSO607762S, RSO611175S, RSO609774S, RSO609775S, RSO609776S, RSO609777S, RSO609778S, RSO609779S, RSO609785S, RSO609786S, RSO610277S, RSO610278S, RSO610280S, RSO610281S, RSO610282S, RSO610283S, RSO610284S, RSO610285S, RSO610286S, RSO610298S, RSO610299S, RSO610300S, RSO610301S, RSO610302S, RSO610305S, RSO610306S, RSO610307S, RSO610308S, RSO610309S, RSO610310S, RSO610311S, RSO610864S, RSO610898S, RSO610901S, RSO611178S, RSO611179S, RSO611180S, RSO611182S, RSO611501S, RSO613384S, RSO613385S, RSO613389S, RSO613392S, RSO613393S, RSO613394S, RSO613395S, RSO613397S, RSO613398S, RSO613404S, RSO604886S, RSO606808S, RSO606809S, RSO609780S, RSO609782S, RSO609783S, RSO609784S, RSO613067S, RSO613069S, RSO613070S, RSO613071S, RSO613072S, RSO613073S, RSO613074S, RSO613075S, RSO613076S, RSO613077S, RSO613078S, RSO613079S, RSO613086S, RSO613088S, RSO613090S, RSO600010S, RSO600011S, RSO611455S, RSO613061S, RSO613062S, RSO613528S, RSO604857S, RSO605962S, RSO606002S, RSO608828S, RSO610758S, RSO610762S, RSO611265S, RSO612860S, RSO612861S, RSO612862S, RSO602438S, RSO606947S, RSO610250S, RSO604831S, RSO606272S, RSO610990S, RSO606944S, RSO600047S, RSO606779S, RSO602444S, RSO602626S, RSO606397S, RSO609051S, RSO609334S, RSO609335S, RSO609420S, RSO610040S, RSO610076S, RSO609155S, RSO609215S, RSO609325S, RSO609370S, RSO609376S, RSO609484S, RSO609563S, RSO609971S, RSO609977S, RSO609999S, RSO610216S, RSO610230S, RSO610240S, RSO610251S, RSO610269S, RSO610373S, RSO610519S, RSO610521S, RSO611188S, RSO611192S, RSO611235S, RSO611255S, RSO611261S, RSO611295S, RSO611302S, RSO611311S, RSO611351S, RSO611395S, RSO611414S, RSO611415S, RSO611422S, RSO611433S, RSO611448S, RSO611453S, RSO611462S, RSO611466S, RSO611467S, RSO611468S, RSO611491S, RSO611497S, RSO611498S, RSO611590S, RSO611593S, RSO611650S, RSO612064S, RSO612088S, RSO612187S, RSO612401S, RSO612414S, RSO612415S, RSO612416S, RSO612417S, RSO612427S, RSO612432S, RSO612433S, RSO612441S, RSO612442S, RSO612448S, RSO612453S, RSO612454S, RSO612460S, RSO612461S, RSO612462S, RSO612463S, RSO612479S, RSO612481S, RSO612812S, RSO612830S, RSO612850S, RSO612854S, RSO612855S, RSO612863S, RSO612869S, RSO612874S, RSO612882S, RSO612883S, RSO612887S, RSO612892S, RSO612899S, RSO612900S, RSO612978S, RSO612983S, RSO612984S, RSO612997S, RSO613001S, RSO613007S, RSO613008S, RSO613009S, RSO613010S, RSO613011S, RSO613018S, RSO613020S, RSO613025S, RSO613028S, RSO613029S, RSO613033S, RSO613038S, RSO613042S, RSO613045S, RSO613053S, RSO613066S, RSO613093S, RSO613100S, RSO613101S, RSO613313S, RSO613315S, RSO613318S, RSO613323S, RSO613324S, RSO613327S, RSO613336S, RSO613337S, RSO613338S, RSO613339S, RSO613341S, RSO613342S, RSO613343S, RSO613345S, RSO613346S, RSO613347S, RSO613348S, RSO613350S, RSO613351S, RSO613352S, RSO613356S, RSO613357S, RSO613360S, RSO613361S, RSO613367S, RSO613371S, RSO613375S, RSO613424S, RSO613425S, RSO613426S, RSO613427S, RSO613428S, RSO613429S, RSO613431S, RSO613433S, RSO613549S, RSO613550S, RSO613551S, RSO613552S, RSO613553S, RSO613554S, RSO613555S, RSO613557S, RSO613560S, RSO613562S, RSO613564S, RSO613566S, RSO613567S, RSO613570S, RSO613571S, RSO613572S, RSO613573S, RSO613574S, RSO613577S, RSO613579S, RSO613581S, RSO613589S, RSO613590S, RSO613595S, RSO613599S, RSO613601S, RSO613606S, RSO613607S, RSO613608S, RSO613610S, RSO613612S, RSO613613S, RSO613614S, RSO613616S, RSO613596S, RSO613598S, RSO613378S, RSO613379S, RSO613380S, RSO613382S, RSO613383S, RSO613412S, RSO613413S, RSO613418S, RSO613419S, RSO613421S, RSO613422S, RSO613423S, RSO613603S, RSO613604S, RSO613615S, RSO613617S, RSO613619S, RSO613620S, RSO613621S, RSO607441S, RSO607738S, RSO608403S, RSO608519S, RSO612825S, RSO613022S, RSO613051S, RSO613340S, RSO613358S, RSO613359S, RSO613373S, RSO613556S, RSO613580S, RSO613586S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce11,645 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LWS
PMAs with Product Code = LWS
-
-