• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall medtronic see related information
Date Initiated by Firm May 10, 2023
Date Posted June 16, 2023
Recall Status1 Open3, Classified
Recall Number Z-1737-2023
Recall Event ID 92328
PMA Number P010031S442 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
Product CRT-D DTBA2D4 VIVA XT IS1/DF4 INTL, Model Number DTBA2D4; Implantable Cardioverter Defibrillators

Code Information GTIN 00643169217966, Lot Serial Numbers: BLI613233S, BLI613324S, BLI613699S, BLI614038S, BLI614165S, BLI613293S, BLI613712S, BLI613772S, BLI614301S, BLI613871S, BLI614140S, BLI614826S, BLI612567S, BLI614031S, BLI613259S, BLI613417S, BLI614737S, BLI614194S, BLI614210S, BLI614604S, BLI613371S, BLI612853S, BLI612859S, BLI613234S, BLI613304S, BLI613305S, BLI613994S, BLI614312S, BLI614469S, BLI612711S, BLI612774S, BLI612954S, BLI613449S, BLI613588S, BLI613591S, BLI614002S, BLI614013S, BLI614290S, BLI614537S, BLI614600S, BLI612805S, BLI613296S, BLI614202S, BLI614460S, BLI613765S, BLI613867S, BLI614307S, BLI614462S, BLI614482S, BLI614593S, BLI612754S, BLI612627S, BLI612707S, BLI612922S, BLI613008S, BLI613255S, BLI613312S, BLI613592S, BLI613665S, BLI613874S, BLI614300S, BLI614304S, BLI614478S, BLI614550S, BLI612769S, BLI613229S, BLI613373S, BLI614302S, BLI614836S, BLI613865S, BLI612946S, BLI613750S, BLI614184S, BLI614577S, BLI613754S, BLI613758S, BLI612866S, BLI612961S, BLI614599S, BLI613275S, BLI613418S, BLI613649S, BLI613879S, BLI614325S, BLI612962S, BLI613272S, BLI614595S, BLI613243S, BLI613284S, BLI614472S, BLI612753S, BLI613334S, BLI613664S, BLI612828S, BLI614463S, BLI614487S, BLI614772S, BLI614781S, BLI612721S, BLI612752S, BLI613693S, BLI613790S, BLI612950S, BLI613294S, BLI613331S, BLI613332S, BLI614173S, BLI614535S, BLI614831S, BLI612781S, BLI612956S, BLI613231S, BLI612717S, BLI612902S, BLI614784S, BLI613992S, BLI612937S, BLI614227S, BLI612867S, BLI614167S, BLI613452S, BLI614000S, BLI614043S, BLI614316S, BLI614586S, BLI613336S, BLI613415S, BLI613689S, BLI614182S, BLI612917S, BLI612941S, BLI612952S, BLI614147S, BLI613239S, BLI614587S, BLI614467S, BLI614474S, BLI613222S, BLI614218S, BLI613300S, BLI612908S, BLI614764S, BLI612778S, BLI613728S, BLI614769S, BLI612584S, BLI612764S, BLI614003S, BLI614575S, BLI613307S, BLI613224S, BLI613502S, BLI614759S, BLI612576S, BLI612837S, BLI614180S, BLI612558S, BLI612773S, BLI614027S, BLI614477S, BLI614573S, BLI614024S, BLI614581S, BLI612948S, BLI614321S, BLI614583S, BLI613695S, BLI613590S, BLI612755S, BLI613225S, BLI614569S, BLI614499S, BLI614200S, BLI614486S, BLI612963S, BLI613270S, BLI612772S, BLI612861S, BLI613764S, BLI612868S, BLI613214S, BLI613446S, BLI613368S, BLI614733S, BLI612847S, BLI612600S, BLI612841S, BLI613873S, BLI614295S, BLI614314S, BLI614580S, BLI614775S, BLI612453S, BLI612771S, BLI612783S, BLI612850S, BLI612957S, BLI613228S, BLI613238S, BLI613264S, BLI613279S, BLI613328S, BLI613451S, BLI613490S, BLI613603S, BLI613669S, BLI613677S, BLI613681S, BLI613775S, BLI613997S, BLI614234S, BLI614311S, BLI612942S, BLI613500S, BLI614747S, BLI613999S, BLI614209S, BLI614541S, BLI614724S, BLI612588S, BLI613262S, BLI613422S, BLI613586S, BLI614235S, BLI614306S, BLI612698S, BLI612862S, BLI612966S, BLI613325S, BLI614291S, BLI613420S, BLI612676S, BLI612678S, BLI612679S, BLI612680S, BLI612681S, BLI612686S, BLI612687S, BLI612689S, BLI612693S, BLI612829S, BLI612869S, BLI612870S, BLI612871S, BLI612872S, BLI612873S, BLI612874S, BLI612877S, BLI612883S, BLI612896S, BLI612898S, BLI612901S, BLI613298S, BLI613308S, BLI613330S, BLI613701S, BLI613704S, BLI613707S, BLI613716S, BLI613795S, BLI612840S, BLI612699S, BLI612757S, BLI613216S, BLI613235S, BLI613295S, BLI613717S, BLI614768S, BLI612802S, BLI612907S, BLI613654S, BLI614041S, BLI614162S, BLI614238S, BLI613303S, BLI614187S, BLI613794S, BLI613219S, BLI614774S, BLI614479S, BLI614739S, BLI612719S, BLI614544S, BLI613407S, BLI613673S, BLI613241S, BLI613798S, BLI612596S, BLI613256S, BLI613309S, BLI612909S, BLI613998S, BLI614014S, BLI614163S, BLI612597S, BLI613335S, BLI613990S, BLI614578S, BLI612903S, BLI613991S, BLI614217S, BLI614473S, BLI614297S, BLI613326S, BLI613770S, BLI613870S, BLI614225S, BLI614598S, BLI614815S, BLI613875S, BLI614465S, BLI614178S, BLI613313S, BLI614294S, BLI612581S, BLI613237S, BLI614317S, BLI613287S, BLI614236S, BLI613868S, BLI614582S, BLI614020S, BLI613283S, BLI614007S, BLI614222S, BLI614567S, BLI612951S, BLI613277S, BLI613421S, BLI613217S, BLI612939S, BLI613656S, BLI614017S, BLI614323S, BLI612949S, BLI613779S, BLI614459S, BLI613660S, BLI612803S, BLI612911S, BLI612921S, BLI613218S, BLI613221S, BLI613232S, BLI613242S, BLI613274S, BLI613292S, BLI613611S, BLI613661S, BLI613796S, BLI613861S, BLI612720S, BLI614480S, BLI612923S, BLI613678S, BLI614318S, BLI612459S, BLI612762S, BLI613223S, BLI613314S, BLI614223S, BLI612965S, BLI614470S, BLI614328S, BLI612930S, BLI614745S, BLI613327S, BLI613644S, BLI614143S, BLI614572S, BLI613706S, BLI613721S, BLI612702S, BLI612763S, BLI612846S, BLI614229S, BLI614461S, BLI613447S, BLI614006S, BLI613297S, BLI613443S, BLI613697S, BLI614181S, BLI613732S, BLI612759S, BLI613453S, BLI613714S, BLI614305S, BLI614481S, BLI614601S, BLI613877S, BLI613405S, BLI614475S, BLI614145S, BLI612817S, BLI613782S, BLI614032S, BLI614466S, BLI612730S, BLI614177S, BLI614763S, BLI613445S, BLI613774S, BLI613315S, BLI614313S, BLI613783S, BLI614320S, BLI612845S, BLI613369S, BLI612723S, BLI612835S, BLI613866S, BLI613220S, BLI613226S, BLI613254S, BLI614001S, BLI613323S, BLI613281S, BLI614220S, BLI613878S, BLI612464S, BLI614216S, BLI612715S, BLI612785S, BLI614322S, BLI612842S, BLI614146S, BLI614735S, BLI613236S, BLI612664S, BLI612665S, BLI613273S, BLI613282S, BLI613450S, BLI612751S, BLI613657S, BLI613996S, BLI613719S, BLI614734S, BLI614046S, BLI614752S, BLI614792S, BLI612936S, BLI613253S, BLI614315S, BLI614813S, BLI612573S, BLI612865S, BLI612799S, BLI614155S, BLI614198S, BLI614226S, BLI612766S, BLI612695S, BLI612815S, BLI614299S, BLI614471S, BLI614987S, BLI615026S, BLI614327S, BLI613872S, BLI612792S, BLI614497S, BLI613301S, BLI614208S, BLI614571S, BLI613596S, BLI612701S, BLI612830S, BLI612931S, BLI613291S, BLI613487S, BLI613672S, BLI613762S, BLI614726S, BLI612667S, BLI612669S, BLI612670S, BLI612708S, BLI612710S, BLI612806S, BLI612807S, BLI613227S, BLI614158S, BLI614717S, BLI614783S, BLI612912S, BLI614833S, BLI612658S, BLI612834S, BLI613319S, BLI613289S, BLI613676S, BLI614793S, BLI613333S, BLI614329S, BLI614476S, BLI613306S, BLI614835S, BLI614310S, BLI613600S, BLI614148S, BLI614483S, BLI612716S, BLI613302S, BLI613484S, BLI613720S, BLI613742S, BLI614004S, BLI613414S, BLI614092S, BLI613466S, BLI613278S, BLI614484S, BLI614727S, BLI614035S, BLI614596S, BLI614293S, BLI614319S, BLI614324S, BLI612852S, BLI613862S, BLI613311S, BLI613318S, BLI613320S, BLI613338S, BLI613339S, BLI613340S, BLI614597S, BLI613562S, BLI614008S, BLI614740S, BLI614785S, BLI612724S, BLI613285S, BLI612904S, BLI612910S, BLI614832S, BLI612944S, BLI614464S, BLI613367S, BLI612933S, BLI614015S, BLI613257S, BLI613288S, BLI612827S, BLI614777S, BLI613286S, BLI613464S; GTIN 00643169720282, Lot Serial Numbers: BLI613497S, BLI612633S, BLI612639S, BLI612814S, BLI612816S, BLI612818S, BLI612474S, BLI612565S, BLI612624S, BLI613054S, BLI613064S, BLI612544S, BLI613408S, BLI613062S, BLI613365S, BLI613542S, BLI613547S, BLI612618S, BLI613363S, BLI612562S, BLI613069S, BLI613083S, BLI613530S, BLI613725S, BLI613410S, BLI613548S, BLI613559S, BLI612513S, BLI612663S, BLI612541S, BLI612550S, BLI613082S, BLI613072S, BLI613074S, BLI613075S, BLI612712S, BLI612713S, BLI612450S, BLI612469S, BLI612509S, BLI612171S, BLI612496S, BLI612534S, BLI612482S, BLI612484S, BLI612490S, BLI612580S, BLI613353S, BLI613376S, BLI613057S, BLI612630S, BLI612641S, BLI612645S, BLI612696S, BLI612697S, BLI612455S, BLI612756S, BLI612634S, BLI612637S, BLI613090S, BLI613535S, BLI612542S, BLI613539S, BLI613372S, BLI612540S, BLI613055S, BLI612449S, BLI612685S, BLI613079S, BLI613081S, BLI612552S, BLI612564S, BLI612546S, BLI612561S, BLI613080S, BLI613108S, BLI613529S, BLI614756S, BLI612654S, BLI613375S, BLI613105S, BLI613126S, BLI613127S, BLI613133S, BLI613136S, BLI613140S, BLI612533S, BLI612535S, BLI613825S, BLI614725S, BLI612860S, BLI612563S, BLI612673S, BLI613109S, BLI613110S, BLI613111S, BLI613112S, BLI612553S, BLI613374S, BLI613039S, BLI613685S, BLI613709S, BLI612591S, BLI613183S, BLI613188S, BLI613448S, BLI613659S, BLI613710S, BLI613760S, BLI613797S, BLI614080S, BLI614104S, BLI612593S, BLI612629S, BLI612761S, BLI613016S, BLI613139S, BLI613145S, BLI613456S, BLI613463S, BLI613698S, BLI612589S, BLI612647S, BLI612760S, BLI612784S, BLI613121S, BLI613134S, BLI613156S, BLI613185S, BLI613196S, BLI612789S, BLI613195S, BLI613651S, BLI613690S, BLI613715S, BLI613769S, BLI613791S, BLI614206S, BLI612622S, BLI613162S, BLI614090S, BLI613168S, BLI613170S, BLI613171S, BLI613172S, BLI613175S, BLI612608S, BLI612613S, BLI612614S, BLI612642S, BLI612666S, BLI613071S, BLI613129S, BLI613130S, BLI613135S, BLI613142S, BLI613147S, BLI613191S, BLI613213S, BLI613477S, BLI613646S, BLI613679S, BLI613683S, BLI614086S, BLI614204S, BLI614211S, BLI612612S, BLI613144S, BLI614072S, BLI614085S, BLI614107S, BLI612660S, BLI612770S, BLI613692S, BLI612812S, BLI613076S, BLI613163S, BLI613470S, BLI614207S, BLI612775S, BLI613190S, BLI612765S, BLI612810S, BLI613041S, BLI613053S, BLI613192S, BLI613444S, BLI613471S, BLI613648S, BLI614071S, BLI612598S, BLI612662S, BLI612750S, BLI612811S, BLI613017S, BLI613157S, BLI613197S, BLI613199S, BLI613201S, BLI613206S, BLI612518S, BLI612604S, BLI612672S, BLI612788S, BLI612793S, BLI613066S, BLI613131S, BLI613141S, BLI613184S, BLI613207S, BLI613647S, BLI613688S, BLI614094S, BLI614105S, BLI614213S, BLI614214S, BLI612605S, BLI612638S, BLI613148S, BLI613149S, BLI613150S, BLI613187S, BLI613193S, BLI613482S, BLI613668S, BLI613670S, BLI613777S, BLI613778S, BLI613780S, BLI614089S, BLI613128S, BLI613146S, BLI613210S, BLI613478S, BLI612577S, BLI612626S, BLI612671S, BLI613158S, BLI613161S, BLI613180S, BLI613469S, BLI613479S, BLI613480S, BLI613708S, BLI612640S, BLI613092S, BLI612651S, BLI612809S, BLI613409S, BLI613411S, BLI613540S, BLI612458S, BLI612460S, BLI612683S, BLI613078S, BLI612690S, BLI612536S, BLI612538S, BLI612539S, BLI612547S, BLI612668S, BLI613392S, BLI613645S, BLI613988S, BLI613391S, BLI613390S, BLI613246S, BLI612729S, BLI612740S, BLI612742S, BLI612747S, BLI612863S, BLI612864S, BLI612875S, BLI612915S, BLI612924S, BLI612927S, BLI612929S, BLI612932S, BLI612986S, BLI612987S, BLI612992S, BLI612993S, BLI613000S, BLI613001S, BLI613004S, BLI613005S, BLI613006S, BLI613014S, BLI613018S, BLI613020S, BLI613022S, BLI613023S, BLI613024S, BLI613025S, BLI613027S, BLI613028S, BLI613031S, BLI613035S, BLI613557S, BLI613582S, BLI613583S, BLI613584S, BLI613585S, BLI613587S, BLI613755S, BLI613756S, BLI613820S, BLI613821S, BLI613822S, BLI613823S, BLI613824S, BLI613826S, BLI613827S, BLI613828S, BLI613835S, BLI614033S, BLI614034S, BLI614037S, BLI614079S, BLI614150S, BLI614151S, BLI614152S, BLI614172S, BLI614176S, BLI614183S, BLI614185S, BLI614186S, BLI614212S, BLI614231S, BLI614232S, BLI614273S, BLI614275S, BLI614276S, BLI614278S, BLI614280S, BLI614281S, BLI614282S, BLI614283S, BLI613563S, BLI612503S, BLI613056S, BLI612493S, BLI612519S, BLI612728S, BLI612746S, BLI612895S, BLI613599S, BLI613752S, BLI613837S, BLI613935S, BLI613936S, BLI613947S, BLI613971S, BLI613977S, BLI613983S, BLI613984S, BLI613985S, BLI613986S, BLI614023S, BLI614026S, BLI614096S, BLI614098S, BLI614099S, BLI613061S, BLI613526S, BLI612578S, BLI613059S, BLI613070S, BLI613329S, BLI613337S, BLI613527S, BLI612571S, BLI613342S, BLI612555S, BLI612454S, BLI614720S, BLI612559S, BLI612557S, BLI613341S, BLI613322S, BLI612798S, BLI612988S, BLI612609S, BLI613382S, BLI612486S, BLI612506S, BLI612507S, BLI612508S, BLI612510S, BLI612583S, BLI612592S, BLI612602S, BLI612607S, BLI612625S, BLI612779S, BLI612782S, BLI612786S, BLI612819S, BLI612820S, BLI612822S, BLI612844S, BLI612848S, BLI612849S, BLI612851S, BLI612891S, BLI612928S, BLI613040S, BLI613042S, BLI613043S, BLI613044S, BLI613118S, BLI613119S, BLI613120S, BLI613122S, BLI613123S, BLI613124S, BLI613125S, BLI613132S, BLI613137S, BLI613173S, BLI613174S, BLI613176S, BLI613177S, BLI613194S, BLI613310S, BLI613317S, BLI613321S, BLI613389S, BLI613458S, BLI613459S, BLI613460S, BLI613475S, BLI613495S, BLI613618S, BLI613619S, BLI613620S, BLI613621S, BLI613622S, BLI612777S, BLI612524S, BLI612525S, BLI613036S, BLI613474S, BLI612794S, BLI612796S, BLI612972S, BLI612973S, BLI613604S, BLI612511S, BLI612512S, BLI612514S, BLI612515S, BLI612516S, BLI613735S, BLI613736S, BLI613737S, BLI612991S, BLI612465S, BLI612468S, BLI612470S, BLI612471S, BLI612472S, BLI612473S, BLI612475S, BLI612476S, BLI612477S, BLI612478S, BLI612479S, BLI612480S, BLI612481S, BLI612568S, BLI612569S, BLI612574S, BLI612575S, BLI612628S, BLI612631S, BLI612659S, BLI612675S, BLI612677S, BLI612684S, BLI612688S, BLI612743S, BLI612744S, BLI612943S, BLI612947S, BLI612960S, BLI612970S, BLI613010S, BLI613045S, BLI613050S, BLI613051S, BLI613063S, BLI613094S, BLI613097S, BLI613099S, BLI613100S, BLI613115S, BLI613116S, BLI613117S, BLI613211S, BLI613496S, BLI613747S, BLI613748S, BLI612526S, BLI612527S, BLI612579S, BLI612582S, BLI612646S, BLI612787S, BLI612790S, BLI612791S, BLI612857S, BLI612889S, BLI612890S, BLI612974S, BLI612523S, BLI612727S, BLI613011S, BLI613015S, BLI613019S, BLI613021S, BLI613026S, BLI613029S, BLI613102S, BLI613103S, BLI613104S, BLI613106S, BLI613113S, BLI613202S, BLI613203S, BLI613204S, BLI613205S, BLI613208S, BLI613209S, BLI613413S, BLI612632S, BLI612657S, BLI612741S, BLI612768S, BLI612776S, BLI612975S, BLI612980S, BLI612983S, BLI613046S, BLI613048S, BLI613114S, BLI613151S, BLI613152S, BLI613153S, BLI613154S, BLI613165S, BLI613166S, BLI613167S, BLI613182S, BLI613198S, BLI613467S, BLI613468S, BLI613472S, BLI613473S, BLI613729S, BLI613730S, BLI613743S, BLI613744S, BLI613749S, BLI612466S, BLI612467S, BLI612610S, BLI612616S, BLI612617S, BLI612621S, BLI612705S, BLI612709S, BLI613212S, BLI613244S, BLI613247S, BLI613248S, BLI613249S, BLI613250S, BLI613251S, BLI613258S, BLI613260S, BLI613087S, BLI613089S, BLI613359S, BLI613360S, BLI613361S, BLI613362S, BLI613416S, BLI613419S, BLI613524S, BLI612691S, BLI612692S, BLI612694S, BLI612700S, BLI612703S, BLI612892S, BLI612894S, BLI612899S, BLI612900S, BLI612905S, BLI612913S, BLI612914S, BLI612916S, BLI612918S, BLI612919S, BLI612920S, BLI612925S, BLI613155S, BLI613159S, BLI613160S, BLI613609S, BLI613612S, BLI613613S, BLI613615S, BLI613741S, BLI613498S, BLI612452S, BLI612457S, BLI612461S, BLI612543S, BLI612549S, BLI612653S, BLI612858S, BLI612876S, BLI613058S, BLI613493S, BLI613494S, BLI613864S, BLI613869S, BLI614285S, BLI614286S, BLI614287S, BLI613354S, BLI613356S, BLI613357S, BLI613358S, BLI613378S, BLI613384S, BLI613385S, BLI613388S, BLI613403S, BLI613406S, BLI612821S, BLI612843S, BLI612967S, BLI613012S, BLI613013S, BLI613033S, BLI613555S, BLI613581S, BLI613593S, BLI613711S, BLI613713S, BLI613718S, BLI613753S, BLI613838S, BLI613972S, BLI613987S, BLI614025S, BLI614074S, BLI614075S, BLI614083S, BLI612483S, BLI612485S, BLI612488S, BLI612489S, BLI612491S, BLI612492S, BLI612494S, BLI612495S, BLI612497S, BLI612498S, BLI612499S, BLI612501S, BLI612502S, BLI612505S, BLI612528S, BLI612529S, BLI612530S, BLI612531S, BLI612566S, BLI612623S, BLI612823S, BLI613746S, BLI612548S, BLI612881S, BLI612520S, BLI612521S, BLI612674S, BLI612714S, BLI612718S, BLI612722S, BLI612824S, BLI612825S, BLI612826S, BLI612831S, BLI612832S, BLI613267S, BLI613065S, BLI612885S, BLI612661S, BLI612706S, BLI612725S, BLI612748S, BLI612749S, BLI612795S, BLI612838S, BLI612878S, BLI612879S, BLI612880S, BLI612884S, BLI612888S, BLI612897S, BLI612955S, BLI612958S, BLI612964S, BLI612995S, BLI612996S, BLI612997S, BLI612998S, BLI612999S, BLI613002S, BLI613003S, BLI613007S, BLI613009S, BLI612572S, BLI612745S, BLI612587S, BLI614271S, BLI614274S, BLI613606S, BLI612595S, BLI612615S, BLI613481S, BLI613680S, BLI612893S, BLI613488S, BLI613595S, BLI613763S, BLI613766S, BLI613945S, BLI613946S, BLI613978S, BLI613979S, BLI613982S, BLI614021S, BLI614109S, BLI614279S, BLI614284S, BLI613386S, BLI613186S, BLI613608S, BLI613034S, BLI613533S, BLI613541S, BLI613549S, BLI613574S, BLI613589S, BLI613602S, BLI613652S, BLI613757S, BLI613759S, BLI613768S, BLI613771S, BLI613776S, BLI613781S, BLI613785S, BLI613787S, BLI613789S, BLI613792S, BLI613793S, BLI613830S, BLI613831S, BLI613832S, BLI613833S, BLI613834S, BLI613839S, BLI613930S, BLI613931S, BLI613933S, BLI613934S, BLI613937S, BLI613938S, BLI613939S, BLI613940S, BLI613941S, BLI613942S, BLI613948S, BLI613949S, BLI613970S, BLI613976S, BLI613981S, BLI614016S, BLI614040S, BLI614070S, BLI614073S, BLI614076S, BLI614077S, BLI614078S, BLI614081S, BLI614082S, BLI614087S, BLI614088S, BLI614091S, BLI614100S, BLI614102S, BLI614157S, BLI614164S, BLI614166S, BLI614168S, BLI614169S, BLI614170S, BLI614171S, BLI614174S, BLI614175S, BLI614188S, BLI614189S, BLI614191S, BLI614192S, BLI614195S, BLI614196S, BLI614199S, BLI614219S, BLI614221S, BLI614230S, BLI614233S, BLI614237S, BLI614270S, BLI614272S, BLI614277S, BLI614288S, BLI614289S, BLI613457S, BLI613483S, BLI613485S, BLI613486S, BLI613489S, BLI613491S, BLI613492S, BLI613601S, BLI613860S, BLI613863S, BLI612650S, BLI613086S, BLI613545S, BLI612726S, BLI612926S, BLI612989S, BLI613531S, BLI613932S, BLI613096S, BLI614103S, BLI613614S, BLI612522S, BLI612551S, BLI612554S, BLI612590S, BLI612601S, BLI612619S, BLI612620S, BLI612682S, BLI612731S, BLI612935S, BLI612938S, BLI612940S, BLI612945S, BLI612953S, BLI612968S, BLI612969S, BLI612971S, BLI612976S, BLI612977S, BLI612978S, BLI612981S, BLI612982S, BLI612984S, BLI612985S, BLI612994S, BLI613038S, BLI613049S, BLI613575S, BLI613576S, BLI613577S, BLI613578S, BLI613580S, BLI613662S, BLI613702S, BLI613705S, BLI613943S, BLI613944S, BLI613973S, BLI613974S, BLI613975S, BLI613980S, BLI614011S, BLI613366S, BLI612833S, BLI613364S, BLI612836S, BLI613525S, BLI613030S, BLI613032S, BLI613189S, BLI613252S, BLI613263S, BLI613461S, BLI613462S, BLI613143S, BLI613200S, BLI613597S, BLI613598S, BLI613623S, BLI612886S, BLI612887S, BLI613181S, BLI613594S, BLI613617S, BLI614106S, BLI614201S, BLI614153S, BLI614154S, BLI614156S, BLI614159S, BLI614160S, BLI614161S, BLI614197S, BLI614239S; GTIN 00643169980099, Lot Serial Numbers: BLI614334S, BLI615844S, BLI615933S, BLI615914S, BLI615913S, BLI615677S, BLI613817S, BLI613572S, BLI613521S, BLI613429S, BLI613433S, BLI613629S, BLI613811S, BLI614493S, BLI614435S, BLI615158S, BLI615489S, BLI615413S, BLI614623S, BLI614989S, BLI613509S, BLI615062S, BLI615067S, BLI614682S, BLI615328S, BLI615523S, BLI613625S, BLI614495S, BLI614559S, BLI614614S, BLI614627S, BLI614928S, BLI615202S, BLI615412S, BLI613950S, BLI613887S, BLI614339S, BLI614490S, BLI615129S, BLI613427S, BLI613431S, BLI613807S, BLI613808S, BLI613568S, BLI613569S, BLI613508S, BLI613565S, BLI613566S, BLI613890S, BLI613891S, BLI613893S, BLI613894S, BLI613895S, BLI614418S, BLI614878S, BLI614879S, BLI614882S, BLI614883S, BLI614886S, BLI614887S, BLI614913S, BLI615335S, BLI615336S, BLI615337S, BLI614528S, BLI613435S, BLI613423S, BLI613505S, BLI613814S, BLI614440S, BLI614442S, BLI614450S, BLI614526S, BLI615292S, BLI614420S, BLI614443S, BLI613398S, BLI613396S, BLI614330S, BLI614331S, BLI614332S, BLI614333S, BLI614336S, BLI613510S, BLI614489S, BLI615578S, BLI613882S, BLI614609S, BLI615296S, BLI613960S, BLI614676S, BLI613801S, BLI613518S, BLI613800S, BLI613426S, BLI613428S, BLI613432S, BLI613504S, BLI613511S, BLI613519S, BLI613523S, BLI613573S, BLI613628S, BLI614620S, BLI614621S, BLI614852S, BLI615190S, BLI615400S, BLI613968S, BLI614885S, BLI615069S, BLI613424S, BLI613434S, BLI613503S, BLI613506S, BLI613507S, BLI613840S, BLI613841S, BLI613842S, BLI613843S, BLI613844S, BLI613919S, BLI614376S, BLI614377S, BLI614378S, BLI614515S, BLI614516S, BLI614517S, BLI614518S, BLI614519S, BLI614520S, BLI614521S, BLI614522S, BLI614523S, BLI614524S, BLI614917S, BLI614919S, BLI615063S, BLI613430S, BLI614496S, BLI615216S, BLI615264S, BLI615444S, BLI614607S, BLI614941S, BLI615405S, BLI612854S, BLI612856S, BLI613344S, BLI613345S, BLI613346S, BLI613347S, BLI613348S, BLI613349S, BLI613350S, BLI613351S, BLI613352S, BLI613393S, BLI613401S, BLI613439S, BLI613440S, BLI613441S, BLI613442S, BLI613900S, BLI613901S, BLI613902S, BLI613903S, BLI613913S, BLI613914S, BLI613916S, BLI613924S, BLI613926S, BLI613927S, BLI613928S, BLI613929S, BLI613951S, BLI613957S, BLI613962S, BLI613965S.BLI614122S, BLI614130S, BLI614132S, BLI614134S, BLI614135S, BLI614138S, BLI614139S, BLI613516S, BLI613922S, BLI613436S, BLI613438S, BLI613804S, BLI613845S, BLI614261S, BLI614262S, BLI614264S, BLI614688S, BLI614701S, BLI614863S, BLI614867S, BLI614874S, BLI614875S, BLI615047S, BLI614876S, BLI614246S, BLI615031S, BLI615195S, BLI614250S, BLI614445S, BLI614662S, BLI614689S, BLI614808S, BLI614854S, BLI614856S, BLI615046S, BLI615095S, BLI615174S, BLI615192S, BLI615197S, BLI615199S, BLI615200S, BLI615204S, BLI615211S, BLI615227S, BLI615261S, BLI615266S, BLI614245S, BLI614259S, BLI614706S, BLI614868S, BLI615043S, BLI615055S, BLI615072S, BLI615186S, BLI615234S, BLI615839S, BLI615841S, BLI614267S, BLI614848S, BLI614241S, BLI614242S, BLI614247S, BLI614253S, BLI614849S, BLI614859S, BLI615065S, BLI615105S, BLI615116S, BLI615170S, BLI615206S, BLI615214S, BLI615219S, BLI615270S, BLI615929S, BLI615938S, BLI614265S, BLI614650S, BLI614240S, BLI614256S, BLI614268S, BLI614851S, BLI614853S, BLI614857S, BLI615210S, BLI615215S, BLI614846S, BLI614243S, BLI614248S, BLI614249S, BLI614255S, BLI614257S, BLI614258S, BLI614260S, BLI614423S, BLI614429S, BLI614613S, BLI614828S, BLI614837S, BLI615035S, BLI615041S, BLI615908S, BLI614263S, BLI613564S, BLI613624S, BLI613810S, BLI613897S, BLI613915S, BLI613921S, BLI614492S, BLI614920S, BLI614960S, BLI615098S, BLI615188S, BLI615213S, BLI615258S, BLI615272S, BLI615415S, BLI615449S, BLI615452S, BLI615488S, BLI615566S, BLI615629S, BLI615686S, BLI613923S, BLI613952S, BLI613953S, BLI614491S, BLI614494S, BLI614921S, BLI614929S, BLI614937S, BLI614938S, BLI614942S, BLI614949S, BLI614991S, BLI615075S, BLI615487S, BLI615500S, BLI615580S, BLI615637S, BLI612855S, BLI613343S, BLI615028S, BLI615233S, BLI615262S, BLI615665S, BLI615668S, BLI615936S, BLI614543S, BLI614554S, BLI614615S, BLI614633S, BLI613967S, BLI614137S, BLI614548S, BLI614549S, BLI614632S, BLI614918S, BLI615074S, BLI615077S, BLI615675S, BLI614911S, BLI614939S, BLI613959S, BLI614931S, BLI613955S, BLI613956S, BLI614118S, BLI614449S, BLI614514S, BLI614525S, BLI614646S, BLI614888S, BLI614891S, BLI614892S, BLI614922S, BLI614933S, BLI614943S, BLI614958S, BLI615308S, BLI615309S, BLI615334S, BLI615664S, BLI613969S, BLI615479S, BLI615058S, BLI614335S, BLI614675S, BLI615198S, BLI614995S, BLI614546S, BLI615307S, BLI615068S, BLI615189S, BLI615268S, BLI613394S, BLI614439S, BLI614051S, BLI614948S, BLI615518S, BLI614896S, BLI614897S, BLI614898S, BLI614899S, BLI614900S, BLI614901S, BLI614902S, BLI614903S, BLI614904S, BLI614905S, BLI614946S, BLI614971S, BLI614972S, BLI614973S, BLI614974S, BLI614434S, BLI614117S, BLI614636S, BLI615082S, BLI614608S, BLI614639S, BLI615327S, BLI615853S, BLI613813S, BLI614063S, BLI614064S, BLI614065S, BLI614066S, BLI614067S, BLI614068S, BLI614069S, BLI614355S, BLI614357S, BLI614358S, BLI614391S, BLI614398S, BLI614421S, BLI614424S, BLI614425S, BLI614426S, BLI614427S, BLI614428S, BLI614430S, BLI614431S, BLI614432S, BLI614619S, BLI614622S, BLI614652S, BLI614653S, BLI614655S, BLI614656S, BLI614657S, BLI614673S, BLI614674S, BLI614703S, BLI614704S, BLI614711S, BLI614712S, BLI614713S, BLI614715S, BLI614716S, BLI614819S, BLI614824S, BLI614895S, BLI614965S, BLI614975S, BLI614976S, BLI615071S, BLI615079S, BLI615081S, BLI615277S, BLI615278S, BLI615280S, BLI615290S, BLI615291S, BLI615318S, BLI615319S, BLI615498S, BLI614052S, BLI614053S, BLI614054S, BLI614055S, BLI615293S, BLI615303S, BLI615304S, BLI615306S, BLI615411S, BLI615416S, BLI615445S, BLI615465S, BLI615467S, BLI613961S, BLI613963S, BLI614359S, BLI614394S, BLI614395S, BLI614396S, BLI614397S, BLI614405S, BLI614407S, BLI614408S, BLI614409S, BLI614410S, BLI614411S, BLI614412S, BLI614433S, BLI614441S, BLI614629S, BLI614635S, BLI614637S, BLI614683S, BLI614684S, BLI614685S, BLI614838S, BLI614839S, BLI614893S, BLI614908S, BLI614909S, BLI614910S, BLI614936S, BLI614940S, BLI615171S, BLI615172S, BLI615175S, BLI615187S, BLI615237S, BLI615243S, BLI615244S, BLI615248S, BLI615249S, BLI615251S, BLI615252S, BLI615253S, BLI615257S, BLI615407S, BLI615408S, BLI615409S, BLI615429S, BLI615431S, BLI615447S, BLI615473S, BLI615474S, BLI615499S, BLI615501S, BLI614638S, BLI613908S, BLI613958S, BLI614686S, BLI614884S, BLI614889S, BLI614890S, BLI614658S, BLI614661S, BLI614707S, BLI614804S, BLI614817S, BLI615086S, BLI615089S, BLI615102S, BLI615273S, BLI615276S, BLI615279S, BLI615282S, BLI614056S, BLI614111S, BLI614113S, BLI614114S, BLI614115S, BLI614119S, BLI614120S, BLI614136S, BLI614375S, BLI614379S, BLI614380S, BLI614392S, BLI614393S, BLI614507S, BLI614509S, BLI614616S, BLI614647S, BLI614648S, BLI614649S, BLI614651S, BLI614659S, BLI614660S, BLI614663S, BLI614665S, BLI614666S, BLI614668S, BLI614670S, BLI614671S, BLI614678S, BLI614680S, BLI614698S, BLI614699S, BLI614700S, BLI614702S, BLI614708S, BLI614710S, BLI614841S, BLI614877S, BLI614894S, BLI614944S, BLI614945S, BLI614950S, BLI614951S, BLI614952S, BLI614953S, BLI614954S, BLI614955S, BLI614956S, BLI614978S, BLI614979S, BLI614982S, BLI614984S, BLI614985S, BLI615297S, BLI615298S, BLI615340S, BLI615342S, BLI615343S, BLI615345S, BLI615346S, BLI615347S, BLI615348S, BLI615349S, BLI615160S, BLI615161S, BLI615162S, BLI615163S, BLI615164S, BLI615169S, BLI615323S, BLI615512S, BLI615514S, BLI615530S, BLI615536S, BLI615538S, BLI615539S, BLI615575S, BLI615615S, BLI615517S, BLI614532S, BLI614542S, BLI614558S, BLI614640S, BLI613880S, BLI613881S, BLI613883S, BLI613884S, BLI613885S, BLI613886S, BLI613888S, BLI613889S, BLI613892S, BLI613896S, BLI613917S, BLI613918S, BLI614529S, BLI614530S, BLI614531S, BLI614536S, BLI614557S, BLI614603S, BLI614611S, BLI614642S, BLI614643S, BLI614669S, BLI614672S, BLI614677S, BLI614693S, BLI614807S, BLI615029S, BLI615030S, BLI615032S, BLI615039S, BLI615040S, BLI615049S, BLI615054S, BLI615099S, BLI615107S, BLI615109S, BLI615111S, BLI615112S, BLI615113S, BLI615114S, BLI615115S, BLI615117S, BLI615118S, BLI615124S, BLI615135S, BLI615137S, BLI615148S, BLI615185S, BLI615194S, BLI615205S, BLI615207S, BLI615208S, BLI615209S, BLI615212S, BLI615217S, BLI615223S, BLI615225S, BLI615231S, BLI615235S, BLI615239S, BLI615250S, BLI615254S, BLI615256S, BLI615267S, BLI615269S, BLI615379S, BLI615382S, BLI615388S, BLI615404S, BLI615406S, BLI615414S, BLI615426S, BLI615428S, BLI615430S, BLI615432S, BLI615434S, BLI615435S, BLI615436S, BLI615437S, BLI615439S, BLI615440S, BLI615442S, BLI615446S, BLI615451S, BLI615453S, BLI615466S, BLI615468S, BLI615469S, BLI615470S, BLI615471S, BLI615472S, BLI615475S, BLI615476S, BLI615477S, BLI615480S, BLI615481S, BLI615482S, BLI615483S, BLI615507S, BLI615508S, BLI615516S, BLI615520S, BLI615525S, BLI615526S, BLI615527S, BLI615528S, BLI615529S, BLI615533S, BLI615534S, BLI615535S, BLI615537S, BLI615540S, BLI615553S, BLI615557S, BLI615562S, BLI615569S, BLI615570S, BLI615572S, BLI615573S, BLI615576S, BLI615577S, BLI615579S, BLI615581S, BLI615587S, BLI615588S, BLI615590S, BLI615592S, BLI615604S, BLI615605S, BLI615606S, BLI615607S, BLI615609S, BLI615614S, BLI615616S, BLI615621S, BLI615622S, BLI615623S, BLI615624S, BLI615625S, BLI615626S, BLI615627S, BLI615638S, BLI615639S, BLI615641S, BLI615643S, BLI615644S, BLI615645S, BLI615646S, BLI615650S, BLI615652S, BLI615653S, BLI615666S, BLI615667S, BLI615669S, BLI615672S, BLI615673S, BLI615680S, BLI615681S, BLI615683S, BLI615692S, BLI615765S, BLI615771S, BLI615778S, BLI615781S, BLI615799S, BLI615800S, BLI615802S, BLI615803S, BLI615804S, BLI615805S, BLI615806S, BLI615810S, BLI615812S, BLI615813S, BLI615814S, BLI615815S, BLI615817S, BLI615818S, BLI615819S, BLI615820S, BLI615821S, BLI615822S, BLI615823S, BLI615824S, BLI615825S, BLI615826S, BLI615829S, BLI615833S, BLI615834S, BLI615835S, BLI615836S, BLI615845S, BLI615850S, BLI615851S, BLI615852S, BLI615854S, BLI615859S, BLI615860S, BLI615862S, BLI615864S, BLI615865S, BLI615867S, BLI615871S, BLI615873S, BLI615882S, BLI615884S, BLI615885S, BLI615888S, BLI615889S, BLI615890S, BLI615891S, BLI615894S, BLI615895S, BLI615896S, BLI615897S, BLI615898S, BLI615925S, BLI615926S, BLI615930S, BLI615954S, BLI615958S, BLI615966S, BLI615971S, BLI615977S, BLI615980S, BLI615981S, BLI615989S, BLI615998S, BLI615999S, BLI616000S, BLI616001S, BLI616002S, BLI616003S, BLI616004S, BLI616005S, BLI616008S, BLI616009S, BLI616010S, BLI616011S, BLI616014S, BLI616015S, BLI616016S, BLI616017S, BLI616018S, BLI616019S, BLI616020S, BLI616021S, BLI616022S, BLI616023S, BLI616024S, BLI616025S, BLI616026S, BLI616027S, BLI616028S, BLI616029S, BLI616030S, BLI616031S, BLI616032S, BLI616033S, BLI615317S, BLI613954S, BLI613964S, BLI613966S, BLI614061S, BLI614062S, BLI614110S, BLI614121S, BLI614123S, BLI614124S, BLI614125S, BLI614126S, BLI614127S, BLI614128S, BLI614129S, BLI614131S, BLI614133S, BLI614381S, BLI614382S, BLI614383S, BLI614384S, BLI614385S, BLI614386S, BLI614387S, BLI614388S, BLI614389S, BLI614390S, BLI614402S, BLI614403S, BLI614404S, BLI614444S, BLI614500S, BLI614501S, BLI614502S, BLI614503S, BLI614504S, BLI614505S, BLI614506S, BLI614510S, BLI614511S, BLI614964S, BLI614966S, BLI614967S, BLI614968S, BLI614969S, BLI614970S, BLI614990S, BLI615103S, BLI615275S, BLI615314S, BLI615320S, BLI615321S, BLI615363S, BLI615364S, BLI615398S, BLI615399S, BLI615401S, BLI615402S, BLI615403S, BLI615502S, BLI615617S, BLI615655S, BLI615659S, BLI615661S, BLI615671S, BLI615674S, BLI615693S, BLI615764S, BLI615847S, BLI615849S, BLI615855S, BLI615856S, BLI615858S, BLI615869S, BLI615870S, BLI614116S, BLI615450S, BLI614798S, BLI614252S, BLI614057S, BLI614399S, BLI615224S, BLI615325S, BLI615376S, BLI615380S, BLI614552S, BLI614641S, BLI615104S, BLI615173S, BLI615285S, BLI615377S, BLI615511S, BLI615542S, BLI615684S, BLI616061S, BLI614513S, BLI614957S, BLI615061S, BLI614458S, BLI614617S, BLI613906S, BLI614624S, BLI614873S, BLI614512S, BLI613425S, BLI613812S, BLI614058S, BLI614059S, BLI614060S, BLI614400S, BLI614401S, BLI614508S, BLI614618S, BLI614690S, BLI614705S, BLI614842S, BLI614843S, BLI614844S, BLI614845S, BLI614847S, BLI614850S, BLI614855S, BLI614860S, BLI614861S, BLI614862S, BLI614864S, BLI614865S, BLI614869S, BLI614871S, BLI614872S, BLI614880S, BLI614881S, BLI614906S, BLI614907S, BLI614977S, BLI614992S, BLI615085S, BLI615139S, BLI615140S, BLI615142S, BLI615143S, BLI615146S, BLI615150S, BLI615151S, BLI615152S, BLI615153S, BLI615246S, BLI615299S, BLI615300S, BLI615310S, BLI615311S, BLI615312S, BLI615338S, BLI615350S, BLI615367S, BLI615368S, BLI615369S, BLI615372S, BLI615503S, BLI615504S, BLI615506S, BLI613907S, BLI614527S, BLI613397S, BLI613904S, BLI615767S, BLI614551S, BLI614809S, BLI614810S, BLI614812S, BLI615090S, BLI615155S, BLI615545S, BLI615550S, BLI615586S, BLI615599S, BLI615782S, BLI615816S, BLI615837S, BLI615987S, BLI615988S, BLI614654S, BLI615106S, BLI615125S, BLI615145S, BLI615147S, BLI615179S, BLI615230S, BLI615423S, BLI615438S, BLI615448S, BLI615510S, BLI615515S, BLI615531S, BLI615541S, BLI615548S, BLI615551S, BLI615556S, BLI615558S, BLI615565S, BLI615567S, BLI615571S, BLI615584S, BLI615613S, BLI615628S, BLI615635S, BLI615636S, BLI615663S, BLI615670S, BLI615678S, BLI615679S, BLI615685S, BLI615691S, BLI615830S, BLI615857S, BLI615880S, BLI615881S, BLI615934S, BLI615942S, BLI615991S, BLI615993S, BLI616060S, BLI616062S, BLI615242S, BLI615356S, BLI615394S, BLI615509S, BLI615766S, BLI614547S, BLI614644S, BLI614645S, BLI614664S, BLI614667S, BLI614679S, BLI614800S, BLI614803S, BLI614820S, BLI614829S, BLI615027S, BLI615033S, BLI615034S, BLI615036S, BLI615037S, BLI615038S, BLI615042S, BLI615044S, BLI615048S, BLI615051S, BLI615053S, BLI615093S, BLI615097S, BLI615101S, BLI615120S, BLI615122S, BLI615141S, BLI615144S, BLI615149S, BLI615218S, BLI615220S, BLI615221S, BLI615222S, BLI615226S, BLI615229S, BLI615245S, BLI615263S, BLI615331S, BLI615333S, BLI615339S, BLI615344S, BLI615361S, BLI615371S, BLI615378S, BLI615383S, BLI615384S, BLI615385S, BLI615393S, BLI615459S, BLI615460S, BLI615478S, BLI615505S, BLI615543S, BLI615544S, BLI615546S, BLI615547S, BLI615555S, BLI615559S, BLI615560S, BLI615563S, BLI615564S, BLI615574S, BLI615582S, BLI615583S, BLI615585S, BLI615594S, BLI615595S.BLI615596S, BLI615597S, BLI615598S, BLI615600S, BLI615608S, BLI615610S, BLI615611S, BLI615612S, BLI615618S, BLI615619S, BLI615620S, BLI615631S, BLI615633S, BLI615640S, BLI615647S, BLI615648S, BLI615649S, BLI615656S, BLI615657S, BLI615660S, BLI615662S, BLI615807S, BLI615827S, BLI615828S, BLI615831S, BLI615832S, BLI615838S, BLI615886S, BLI615931S, BLI615941S, BLI615978S, BLI615990S, BLI615992S, BLI615994S, BLI615995S, BLI616006S, BLI616007S, BLI616012S, BLI616013S, BLI616057S, BLI616058S, BLI616059S, BLI614341S, BLI614342S, BLI614343S, BLI614344S, BLI614345S, BLI614346S, BLI614347S, BLI614348S, BLI614349S, BLI614350S, BLI614351S, BLI614352S, BLI614353S, BLI614354S, BLI614360S, BLI614361S, BLI614362S, BLI614363S, BLI614364S, BLI614365S, BLI614366S, BLI614367S, BLI614368S, BLI614369S, BLI614370S, BLI614371S, BLI614372S, BLI614373S, BLI614374S, BLI614406S, BLI614413S, BLI614414S, BLI614415S, BLI614416S, BLI614417S, BLI613905S, BLI614687S, BLI614691S, BLI614692S, BLI614694S, BLI614695S, BLI614696S, BLI614697S, BLI614811S, BLI615088S, BLI615091S, BLI615092S, BLI615108S, BLI615119S, BLI615165S, BLI615329S, BLI615330S, BLI615332S, BLI615365S, BLI615366S, BLI615370S, BLI615373S, BLI615374S, BLI615375S, BLI615381S, BLI615133S, BLI615134S, BLI615136S, BLI615138S, BLI615156S, BLI615176S, BLI615177S, BLI615178S, BLI615180S, BLI615182S, BLI615183S, BLI615184S, BLI615313S, BLI615324S, BLI615417S, BLI615418S, BLI615420S, BLI615421S, BLI615422S, BLI615424S, BLI615425S, BLI615433S, BLI615441S, BLI615443S, BLI615454S, BLI615457S, BLI615458S, BLI615461S, BLI615462S, BLI614912S, BLI614914S, BLI614915S, BLI614916S, BLI614923S, BLI614924S, BLI614925S, BLI614926S, BLI614927S, BLI614934S, BLI614935S, BLI614959S, BLI614961S, BLI614962S, BLI614963S, BLI614986S, BLI614988S, BLI614993S, BLI614994S, BLI615121S, BLI615123S, BLI615126S, BLI615127S, BLI615128S, BLI615130S, BLI615131S, BLI615132S, BLI615157S, BLI615159S, BLI615166S, BLI615167S, BLI615168S, BLI615228S, BLI615232S, BLI615236S, BLI615284S, BLI615286S, BLI615288S, BLI615289S, BLI615294S, BLI615295S, BLI615301S, BLI615302S, BLI615305S, BLI615351S, BLI615353S, BLI615354S, BLI615355S, BLI615357S, BLI615358S, BLI615359S, BLI615360S, BLI615362S, BLI615386S, BLI615387S, BLI615389S, BLI615390S, BLI615391S, BLI615392S, BLI615395S, BLI615396S, BLI615397S, BLI615485S, BLI615486S, BLI615490S, BLI615491S, BLI615492S, BLI615493S, BLI615494S, BLI615495S, BLI615496S, BLI615497S, BLI615532S, BLI615552S, BLI615554S, BLI615561S, BLI615568S, BLI615603S, BLI615630S, BLI615654S, BLI615808S, BLI615809S, BLI615811S, BLI615939S, BLI615984S, BLI615985S, BLI615986S, BLI615996S, BLI615997S, BLI616063S, BLI615154S, BLI615651S, BLI615774S, BLI615775S, BLI615776S, BLI615842S, BLI615872S, BLI615874S, BLI615877S, BLI615905S, BLI615906S, BLI615907S, BLI615909S, BLI615910S, BLI615911S, BLI615912S, BLI615915S, BLI615916S, BLI615917S, BLI615918S, BLI615928S, BLI615932S, BLI615935S, BLI615943S, BLI615949S, BLI615952S, BLI615953S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information Contact Technical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
Action Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce 2904 units
Distribution US Nationwide - Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA Database PMAs with Product Code = NIK and Original Applicant = MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT
PMAs with Product Code = NIK and Original Applicant = Medtronic, Inc.
-
-