• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1766-2023
Recall Event ID 92328
PMA NumberP010031S550 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
ProductCRTD DTMB2D4 AMPLIA MRI OUS DF4, Model Number DTMB2D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00643169720725, Lot Serial Numbers: RPK603310S, RPK606030S, RPK607323S, RPK606692S, RPK605552S, RPK604274S, RPK607156S, RPK607510S, RPK606721S, RPK606379S, RPK606753S, RPK603895S, RPK607997S, RPK608001S, RPK608115S, RPK602608S, RPK602842S, RPK602915S, RPK603155S, RPK603438S, RPK603475S, RPK604209S, RPK604230S, RPK604885S, RPK605001S, RPK605994S, RPK606046S, RPK606664S, RPK607134S, RPK608665S, RPK604002S, RPK604373S, RPK605499S, RPK605680S, RPK603963S, RPK607543S, RPK607544S, RPK607943S, RPK608088S, RPK608203S, RPK607149S, RPK607416S, RPK607418S, RPK607615S, RPK608212S, RPK608297S, RPK608298S, RPK609062S, RPK604646S, RPK605555S, RPK605685S, RPK606797S, RPK604171S, RPK605262S, RPK605291S, RPK605292S, RPK606475S, RPK607086S, RPK603372S, RPK603375S, RPK603380S, RPK603962S, RPK604174S, RPK604351S, RPK604536S, RPK604651S, RPK604654S, RPK604869S, RPK604972S, RPK604973S, RPK605109S, RPK605214S, RPK605233S, RPK605272S, RPK606024S, RPK606053S, RPK606141S, RPK606473S, RPK606792S, RPK606817S, RPK606853S, RPK607088S, RPK607422S, RPK607441S, RPK607442S, RPK607443S, RPK607620S, RPK607647S, RPK607857S, RPK607865S, RPK607972S, RPK608102S, RPK608939S, RPK604642S, RPK604644S, RPK604989S, RPK607684S, RPK608101S, RPK608890S, RPK603423S, RPK603908S, RPK604177S, RPK604228S, RPK604660S, RPK604981S, RPK605848S, RPK606143S, RPK606525S, RPK606535S, RPK607423S, RPK607911S, RPK607917S, RPK607918S, RPK604191S, RPK606048S, RPK606472S, RPK606793S, RPK606813S, RPK606816S, RPK607922S, RPK604227S, RPK604353S, RPK604630S, RPK605297S, RPK606819S, RPK606880S, RPK607479S, RPK604631S, RPK604649S, RPK606045S, RPK606047S, RPK604183S, RPK607541S, RPK605294S, RPK605491S, RPK607420S, RPK607424S, RPK607916S, RPK604169S, RPK607617S, RPK604354S, RPK605972S, RPK607864S, RPK604337S, RPK606799S, RPK607862S, RPK607970S, RPK605255S, RPK606479S, RPK606795S, RPK603498S, RPK607111S, RPK605251S, RPK602530S, RPK602763S, RPK604901S, RPK605486S, RPK606302S, RPK606871S, RPK606924S, RPK607005S, RPK607012S, RPK602647S, RPK603874S, RPK604083S, RPK604250S, RPK604637S, RPK604843S, RPK605472S, RPK605501S, RPK605944S, RPK608697S, RPK608716S, RPK602684S, RPK607155S, RPK608751S, RPK603603S, RPK603793S, RPK604657S, RPK604914S, RPK606445S, RPK608369S, RPK603127S, RPK603736S, RPK603744S, RPK604254S, RPK604761S, RPK604816S, RPK605328S, RPK606467S, RPK607777S, RPK608737S, RPK606595S, RPK607324S, RPK602648S, RPK603632S, RPK608510S, RPK602495S, RPK602759S, RPK603124S, RPK607846S, RPK607699S, RPK602811S, RPK602613S, RPK608087S, RPK603103S, RPK605235S, RPK606577S, RPK602662S, RPK603185S, RPK605397S, RPK605466S, RPK602556S, RPK602590S, RPK603507S, RPK604255S, RPK604426S, RPK605441S, RPK605698S, RPK607293S, RPK608446S, RPK602502S, RPK605307S, RPK605334S, RPK605467S, RPK605469S, RPK606398S, RPK607835S, RPK608429S, RPK605036S, RPK605290S, RPK605487S, RPK605723S, RPK606252S, RPK606490S, RPK606668S, RPK606791S, RPK607004S, RPK607331S, RPK607820S, RPK607976S, RPK608217S, RPK608382S, RPK608513S, RPK608953S, RPK608976S, RPK603433S, RPK606865S, RPK602561S, RPK602883S, RPK606212S, RPK606603S, RPK607653S, RPK608883S, RPK608275S, RPK603636S, RPK603657S, RPK605077S, RPK606947S, RPK608365S, RPK608725S, RPK605327S, RPK606596S, RPK607188S, RPK604573S, RPK606513S, RPK606580S, RPK606604S, RPK607603S, RPK607608S, RPK607686S, RPK602588S, RPK608270S, RPK602851S, RPK605459S, RPK607073S, RPK608450S, RPK608545S, RPK606412S, RPK606978S, RPK607745S, RPK608735S, RPK602595S, RPK602641S, RPK602816S, RPK603264S, RPK603656S, RPK603735S, RPK603883S, RPK604088S, RPK604801S, RPK605570S, RPK607279S, RPK607520S, RPK608016S, RPK608517S, RPK608794S, RPK608835S, RPK604800S, RPK604822S, RPK604950S, RPK605192S, RPK605573S, RPK605761S, RPK606681S, RPK607405S, RPK607725S, RPK607770S, RPK604622S, RPK602591S, RPK603086S, RPK603972S, RPK604363S, RPK605579S, RPK606589S, RPK607161S, RPK606858S, RPK608520S, RPK608804S, RPK608834S, RPK603263S, RPK605210S, RPK608696S, RPK602600S, RPK602712S, RPK603258S, RPK605971S, RPK606930S, RPK608740S, RPK608800S, RPK605935S, RPK606142S, RPK606362S, RPK606399S, RPK608469S, RPK608541S, RPK608574S, RPK608850S, RPK602989S, RPK603041S, RPK603134S, RPK603252S, RPK603544S, RPK603547S, RPK604006S, RPK604136S, RPK604225S, RPK604460S, RPK604578S, RPK604617S, RPK604688S, RPK604734S, RPK604958S, RPK605028S, RPK605076S, RPK605115S, RPK605304S, RPK605370S, RPK605710S, RPK605993S, RPK606189S, RPK606254S, RPK606860S, RPK606984S, RPK607015S, RPK607594S, RPK607736S, RPK607839S, RPK607991S, RPK608072S, RPK608532S, RPK608824S, RPK608853S, RPK608880S, RPK608930S, RPK608958S, RPK603489S, RPK603617S, RPK604434S, RPK605075S, RPK605497S, RPK605904S, RPK606384S, RPK608170S, RPK606568S, RPK608516S, RPK604858S, RPK608882S, RPK608849S, RPK606388S, RPK608172S, RPK608868S, RPK606495S, RPK606723S, RPK607302S, RPK607816S, RPK608425S, RPK602670S, RPK605539S, RPK603545S, RPK605470S, RPK605908S, RPK606794S, RPK607082S, RPK607162S, RPK608391S, RPK605703S, RPK606222S, RPK607283S, RPK607476S, RPK604748S, RPK605117S, RPK605384S, RPK605412S, RPK606830S, RPK607546S, RPK607769S, RPK607169S, RPK602673S, RPK604147S, RPK604604S, RPK604847S, RPK606727S, RPK608493S, RPK603900S, RPK605000S, RPK608171S, RPK608093S, RPK602603S, RPK603609S, RPK606187S, RPK607074S, RPK607337S, RPK606843S, RPK608734S, RPK608959S, RPK604294S, RPK606566S, RPK608747S, RPK606310S, RPK606606S, RPK608840S, RPK608865S, RPK604301S, RPK604556S, RPK606111S, RPK604405S, RPK603655S, RPK608286S, RPK602714S, RPK605369S, RPK606381S, RPK606534S, RPK607832S, RPK608459S, RPK608743S, RPK602667S, RPK603530S, RPK606377S, RPK606938S, RPK607700S, RPK607845S, RPK608809S, RPK608567S, RPK606214S, RPK608485S, RPK608581S, RPK603235S, RPK605007S, RPK606803S, RPK605367S, RPK606488S, RPK606682S, RPK607572S, RPK608851S, RPK608885S, RPK605399S, RPK606068S, RPK605348S, RPK605658S, RPK605968S, RPK606348S, RPK607833S, RPK608629S, RPK604424S, RPK602581S, RPK605219S, RPK605415S, RPK606371S, RPK606487S, RPK606550S, RPK606715S, RPK607206S, RPK607351S, RPK607989S, RPK602808S, RPK602810S, RPK603546S, RPK604415S, RPK604459S, RPK604702S, RPK605332S, RPK605516S, RPK606273S, RPK606414S, RPK606579S, RPK606726S, RPK606855S, RPK606884S, RPK607641S, RPK607828S, RPK608419S, RPK608452S, RPK608837S, RPK603771S, RPK603801S, RPK604087S, RPK604359S, RPK604413S, RPK604707S, RPK605411S, RPK606349S, RPK602826S, RPK602871S, RPK602564S, RPK603490S, RPK603746S, RPK603823S, RPK608538S, RPK603454S, RPK603602S, RPK603670S, RPK606539S, RPK608463S, RPK608854S, RPK603192S, RPK603430S, RPK605122S, RPK606185S, RPK608091S, RPK608161S, RPK605025S, RPK606969S, RPK608857S, RPK603761S, RPK605022S, RPK603486S, RPK604500S, RPK606836S, RPK607326S, RPK603591S, RPK603519S, RPK603762S, RPK603812S, RPK604334S, RPK604412S, RPK604441S, RPK605031S, RPK605230S, RPK606069S, RPK603402S, RPK603871S, RPK605066S, RPK606663S, RPK603637S, RPK604687S, RPK608920S, RPK603510S, RPK603524S, RPK603840S, RPK604504S, RPK605970S, RPK606999S, RPK607192S, RPK608494S, RPK603653S, RPK607166S, RPK608076S, RPK602623S, RPK605211S, RPK608801S, RPK603126S, RPK603836S, RPK604273S, RPK605679S, RPK605913S, RPK608874S, RPK603515S, RPK603882S, RPK604870S, RPK606318S, RPK607407S, RPK608546S, RPK608864S, RPK608972S, RPK603683S, RPK608757S, RPK608351S, RPK602594S, RPK606404S, RPK606699S, RPK609018S, RPK603063S, RPK603726S, RPK603901S, RPK603905S, RPK604711S, RPK606608S, RPK606647S, RPK606771S, RPK606886S, RPK606946S, RPK607342S, RPK607909S, RPK608375S, RPK608519S, RPK608644S, RPK608765S, RPK608841S, RPK603210S, RPK603822S, RPK605034S, RPK608568S, RPK609012S, RPK602881S, RPK603107S, RPK605337S, RPK605341S, RPK606179S, RPK606502S, RPK606686S, RPK602537S, RPK608449S, RPK606950S, RPK607715S, RPK607811S, RPK607827S, RPK607863S, RPK608271S, RPK608277S, RPK605154S, RPK605763S, RPK606402S, RPK607059S, RPK607765S, RPK608027S, RPK602977S, RPK603814S, RPK603819S, RPK604887S, RPK605378S, RPK606018S, RPK606862S, RPK608357S, RPK608986S, RPK603592S, RPK604309S, RPK604668S, RPK605194S, RPK605359S, RPK607554S, RPK607860S, RPK608294S, RPK603562S, RPK605207S, RPK606071S, RPK606168S, RPK606876S, RPK608049S, RPK608952S, RPK609054S, RPK607734S, RPK609040S, RPK604745S, RPK605323S, RPK605704S, RPK606043S, RPK606223S, RPK606347S, RPK606411S, RPK606588S, RPK606637S, RPK606759S, RPK607076S, RPK607352S, RPK607869S, RPK608089S, RPK608092S, RPK608712S, RPK608981S, RPK605335S, RPK605498S, RPK606920S, RPK607089S, RPK607393S, RPK607622S, RPK608187S, RPK606952S, RPK607264S, RPK602217S, RPK602567S, RPK602752S, RPK602831S, RPK603845S, RPK604462S, RPK604615S, RPK608424S, RPK602569S, RPK604814S, RPK606941S, RPK604605S, RPK604606S, RPK604609S, RPK604614S, RPK606311S, RPK606314S, RPK603528S, RPK603788S, RPK603839S, RPK605339S, RPK606198S, RPK606707S, RPK607953S, RPK608674S, RPK608761S, RPK608785S, RPK602566S, RPK606877S, RPK607019S, RPK602894S, RPK603236S, RPK603815S, RPK604245S, RPK604672S, RPK604867S, RPK606116S, RPK606255S, RPK606636S, RPK606942S, RPK607274S, RPK607806S, RPK608105S, RPK608113S, RPK608596S, RPK603278S, RPK604042S, RPK603393S, RPK604861S, RPK606600S, RPK607148S, RPK606498S, RPK604235S, RPK602559S, RPK603410S, RPK608326S, RPK603060S, RPK603179S, RPK603326S, RPK607189S, RPK608970S, RPK602580S, RPK603045S, RPK602797S, RPK603049S, RPK603062S, RPK603080S, RPK603081S, RPK603580S, RPK603581S, RPK603816S, RPK603854S, RPK603856S, RPK604150S, RPK604151S, RPK602998S, RPK602480S, RPK602626S, RPK602633S, RPK604263S, RPK604364S, RPK604725S, RPK605288S, RPK605574S, RPK605715S, RPK602560S, RPK602563S, RPK602570S, RPK602574S, RPK602622S, RPK602624S, RPK602628S, RPK603135S, RPK603142S, RPK603319S, RPK603329S, RPK603331S, RPK603449S, RPK604451S, RPK604547S, RPK604552S, RPK605032S, RPK605118S, RPK605120S, RPK605840S, RPK606005S, RPK606376S, RPK606530S, RPK607078S, RPK607350S, RPK607354S, RPK608870S, RPK604319S, RPK607376S, RPK602523S, RPK602524S, RPK602525S, RPK602527S, RPK602528S, RPK602650S, RPK602711S, RPK602727S, RPK603022S, RPK603024S, RPK603054S, RPK603280S, RPK603282S, RPK603289S, RPK603292S, RPK603293S, RPK603615S, RPK603630S, RPK604361S, RPK604532S, RPK604533S, RPK604534S, RPK606307S, RPK606308S, RPK607627S, RPK607628S, RPK602551S, RPK602653S, RPK603138S, RPK603140S, RPK603158S, RPK608571S, RPK608805S, RPK609079S, RPK603120S, RPK608261S, RPK602511S, RPK603309S, RPK604716S, RPK602836S, RPK602882S, RPK603245S, RPK604142S, RPK604806S, RPK604809S, RPK604810S, RPK605042S, RPK605237S, RPK605947S, RPK606230S, RPK606341S, RPK606554S, RPK606555S, RPK606823S, RPK608547S, RPK603016S, RPK604432S, RPK605788S, RPK602216S, RPK602625S, RPK604093S, RPK604095S, RPK604135S, RPK604720S, RPK604724S, RPK604781S, RPK608464S, RPK602793S, RPK603035S, RPK603421S, RPK603847S, RPK604206S, RPK602499S, RPK603014S, RPK603125S, RPK603561S, RPK603792S, RPK603967S, RPK605360S, RPK603422S, RPK602536S, RPK607294S, RPK606537S, RPK606831S, RPK607041S, RPK607750S, RPK608079S, RPK608080S, RPK608328S, RPK609077S, RPK606309S, RPK606044S, RPK604751S, RPK605143S, RPK605319S, RPK608884S, RPK605673S, RPK602986S, RPK604519S, RPK604796S, RPK606138S, RPK606509S, RPK607812S, RPK608607S, RPK603455S, RPK603564S, RPK603891S, RPK604086S, RPK604238S, RPK608378S, RPK608989S, RPK605946S, RPK602546S, RPK604159S, RPK602935S, RPK608573S, RPK605116S, RPK605563S, RPK602640S, RPK603699S, RPK604802S, RPK603048S, RPK606641S, RPK602645S, RPK602850S, RPK602852S, RPK603023S, RPK603055S, RPK603057S, RPK603159S, RPK603169S, RPK603170S, RPK603596S, RPK603673S, RPK603674S, RPK604275S, RPK604419S, RPK604420S, RPK604666S, RPK604667S, RPK604689S, RPK604790S, RPK604795S, RPK604980S, RPK604982S, RPK604983S, RPK605114S, RPK605127S, RPK605128S, RPK605130S, RPK605136S, RPK605137S, RPK605244S, RPK605270S, RPK605277S, RPK605281S, RPK605285S, RPK605286S, RPK605355S, RPK605356S, RPK605363S, RPK605364S, RPK605436S, RPK605438S, RPK605471S, RPK605592S, RPK605595S, RPK605734S, RPK605750S, RPK605751S, RPK605756S, RPK605910S, RPK605916S, RPK605917S, RPK605951S, RPK605954S, RPK605955S, RPK605965S, RPK605966S, RPK605984S, RPK605985S, RPK605986S, RPK606140S, RPK606231S, RPK606233S, RPK606234S, RPK606235S, RPK606237S, RPK606239S, RPK606305S, RPK606451S, RPK606461S, RPK606469S, RPK606505S, RPK606514S, RPK606515S, RPK606516S, RPK606518S, RPK606520S, RPK606569S, RPK606574S, RPK606594S, RPK606672S, RPK606675S, RPK606678S, RPK606680S, RPK606685S, RPK606716S, RPK606962S, RPK606964S, RPK606966S, RPK606993S, RPK606995S, RPK607075S, RPK607170S, RPK607171S, RPK607275S, RPK607276S, RPK607278S, RPK607281S, RPK607286S, RPK607291S, RPK607301S, RPK607303S, RPK607411S, RPK607415S, RPK607465S, RPK607468S, RPK607470S, RPK607534S, RPK607540S, RPK607619S, RPK607626S, RPK607629S, RPK607683S, RPK607685S, RPK607709S, RPK607714S, RPK607757S, RPK607758S, RPK607759S, RPK607761S, RPK607762S, RPK607793S, RPK607794S, RPK607800S, RPK607802S, RPK607823S, RPK607824S, RPK607825S, RPK607834S, RPK607838S, RPK607842S, RPK607843S, RPK607847S, RPK607848S, RPK607854S, RPK607856S, RPK607858S, RPK607923S, RPK607924S, RPK607925S, RPK607926S, RPK607927S, RPK607929S, RPK607930S, RPK608069S, RPK608073S, RPK608074S, RPK608081S, RPK608083S, RPK608084S, RPK608134S, RPK608143S, RPK608412S, RPK608415S, RPK608420S, RPK608465S, RPK608466S, RPK608472S, RPK608475S, RPK608477S, RPK608479S, RPK608480S, RPK608481S, RPK608484S, RPK608492S, RPK608580S, RPK608582S, RPK608618S, RPK608620S, RPK608621S, RPK608623S, RPK608624S, RPK608628S, RPK608639S, RPK608690S, RPK608692S, RPK608695S, RPK608709S, RPK608713S, RPK608714S, RPK608719S, RPK608720S, RPK608784S, RPK608795S, RPK608842S, RPK608892S, RPK608894S, RPK608898S, RPK608957S, RPK608975S, RPK608977S, RPK608978S, RPK608979S, RPK608980S, RPK608982S, RPK608983S, RPK608984S, RPK605468S, RPK603419S, RPK603420S, RPK603501S, RPK603502S, RPK603936S, RPK603980S, RPK603981S, RPK603982S, RPK604081S, RPK604164S, RPK604165S, RPK604170S, RPK604172S, RPK604213S, RPK604310S, RPK604313S, RPK604317S, RPK604771S, RPK604775S, RPK605276S, RPK605289S, RPK605299S, RPK605978S, RPK605981S, RPK606968S, RPK607033S, RPK608381S, RPK608575S, RPK608646S, RPK608650S, RPK608797S, RPK608872S, RPK608878S, RPK609074S, RPK609075S, RPK607887S, RPK608570S, RPK606983S, RPK607285S, RPK608766S, RPK603259S, RPK603559S, RPK608372S, RPK608843S, RPK608873S, RPK608877S, RPK603133S, RPK603511S, RPK603708S, RPK604851S, RPK604854S, RPK605478S, RPK605494S, RPK606146S, RPK606199S, RPK606822S, RPK606825S, RPK607081S, RPK602514S, RPK602541S, RPK602611S, RPK602688S, RPK602758S, RPK602828S, RPK602847S, RPK602887S, RPK602888S, RPK602897S, RPK602903S, RPK603032S, RPK603123S, RPK603208S, RPK603254S, RPK603522S, RPK603531S, RPK603534S, RPK603568S, RPK603582S, RPK603709S, RPK603825S, RPK603947S, RPK604114S, RPK604126S, RPK604248S, RPK604264S, RPK604298S, RPK604362S, RPK604369S, RPK604389S, RPK604449S, RPK604497S, RPK604565S, RPK604570S, RPK604572S, RPK604576S, RPK604577S, RPK604984S, RPK604988S, RPK605003S, RPK605172S, RPK605190S, RPK605296S, RPK605310S, RPK605311S, RPK605361S, RPK605440S, RPK605473S, RPK605630S, RPK605669S, RPK605672S, RPK605696S, RPK605718S, RPK605730S, RPK605737S, RPK605748S, RPK605876S, RPK606218S, RPK606375S, RPK606866S, RPK607092S, RPK607306S, RPK607371S, RPK608507S, RPK608705S, RPK602577S, RPK606221S, RPK607115S, RPK607855S, RPK608368S, RPK608764S, RPK608889S, RPK603031S, RPK604889S, RPK605144S, RPK605651S, RPK602875S, RPK605860S, RPK606159S, RPK606925S, RPK607191S, RPK608499S, RPK608763S, RPK609076S, RPK603589S, RPK604085S, RPK605222S, RPK605226S, RPK602912S, RPK603100S, RPK603272S, RPK603607S, RPK603731S, RPK603790S, RPK603878S, RPK604138S, RPK604145S, RPK604247S, RPK604408S, RPK604417S, RPK604633S, RPK604685S, RPK604686S, RPK604690S, RPK604727S, RPK604859S, RPK605103S, RPK605215S, RPK605314S, RPK605642S, RPK605707S, RPK605732S, RPK606145S, RPK606147S, RPK606391S, RPK606504S, RPK606676S, RPK606735S, RPK606744S, RPK606903S, RPK607099S, RPK607154S, RPK607356S, RPK607358S, RPK607406S, RPK607523S, RPK607524S, RPK607601S, RPK607767S, RPK607772S, RPK607912S, RPK608005S, RPK608207S, RPK608370S, RPK608395S, RPK608451S, RPK608526S, RPK608583S, RPK608617S, RPK608710S, RPK608739S, RPK608768S, RPK609043S, RPK605085S, RPK605221S, RPK602939S, RPK605268S, RPK605927S, RPK606369S, RPK606464S, RPK607080S, RPK607531S, RPK608919S, RPK603200S, RPK603418S, RPK603732S, RPK603734S, RPK604836S, RPK605012S, RPK605932S, RPK605976S, RPK606007S, RPK606139S, RPK606430S, RPK606743S, RPK606881S, RPK607068S, RPK607778S, RPK607782S, RPK608231S, RPK608544S, RPK608630S, RPK608651S, RPK608660S, RPK608707S, RPK602735S, RPK602911S, RPK605148S, RPK606409S, RPK608125S, RPK603668S, RPK603868S, RPK605385S, RPK607054S, RPK603610S, RPK603678S, RPK605091S, RPK605999S, RPK606982S, RPK602913S, RPK602948S, RPK602966S, RPK603612S, RPK603966S, RPK604082S, RPK604455S, RPK605343S, RPK602694S, RPK603649S, RPK602730S, RPK602907S, RPK602928S, RPK603669S, RPK603677S, RPK603681S, RPK603713S, RPK603860S, RPK604033S, RPK605383S, RPK605458S, RPK605531S, RPK605551S, RPK606211S, RPK606287S, RPK606557S, RPK606710S, RPK608470S, RPK604917S, RPK606737S, RPK607023S, RPK608736S, RPK602732S, RPK603598S, RPK603642S, RPK603848S, RPK603849S, RPK603869S, RPK604015S, RPK604022S, RPK604103S, RPK604905S, RPK605086S, RPK605229S, RPK605518S, RPK606125S, RPK606576S, RPK607001S, RPK607007S, RPK607009S, RPK607207S, RPK603866S, RPK602517S, RPK603861S, RPK603867S, RPK604046S, RPK605050S, RPK605062S, RPK605064S, RPK605242S, RPK605538S, RPK605550S, RPK605625S, RPK606390S, RPK607201S, RPK607220S, RPK603714S, RPK603759S, RPK608987S, RPK603529S, RPK604421S, RPK606544S, RPK606997S, RPK607677S, RPK603611S, RPK605318S, RPK605610S, RPK604084S, RPK605342S, RPK606801S, RPK602960S, RPK603597S, RPK604903S, RPK605454S, RPK606131S, RPK606698S, RPK604655S, RPK604659S, RPK603338S, RPK603567S, RPK602859S, RPK602884S, RPK602995S, RPK603294S, RPK603313S, RPK603949S, RPK604356S, RPK604507S, RPK606483S, RPK603160S, RPK603316S, RPK604960S, RPK605687S, RPK607130S, RPK607675S, RPK604012S, RPK604803S, RPK606976S, RPK607268S, RPK602584S, RPK602602S, RPK602813S, RPK604588S, RPK605873S, RPK603050S, RPK604458S, RPK607378S, RPK603595S, RPK608190S, RPK608738S, RPK608769S, RPK607235S, RPK607260S, RPK604457S, RPK605019S, RPK609011S, RPK609016S, RPK606931S, RPK607528S, RPK603104S, RPK608099S, RPK602612S, RPK602629S, RPK602716S, RPK602785S, RPK602846S, RPK603197S, RPK604330S, RPK604414S, RPK605376S, RPK605692S, RPK606135S, RPK607228S, RPK607763S, RPK604760S, RPK606951S, RPK607771S, RPK608816S, RPK602749S, RPK602775S, RPK602777S, RPK603634S, RPK609024S, RPK605347S, RPK607061S, RPK608266S, RPK602498S, RPK602934S, RPK603392S, RPK604957S, RPK603088S, RPK607374S, RPK602516S, RPK602545S, RPK602553S, RPK602821S, RPK602822S, RPK602992S, RPK603011S, RPK603113S, RPK603202S, RPK603411S, RPK603495S, RPK603540S, RPK603560S, RPK603572S, RPK603585S, RPK603740S, RPK603741S, RPK603820S, RPK603833S, RPK603991S, RPK604007S, RPK604220S, RPK604296S, RPK604342S, RPK604736S, RPK604987S, RPK605070S, RPK605181S, RPK605377S, RPK605646S, RPK605693S, RPK605705S, RPK605709S, RPK605736S, RPK605987S, RPK606188S, RPK606192S, RPK606281S, RPK606496S, RPK606528S, RPK606601S, RPK606878S, RPK606974S, RPK607062S, RPK607063S, RPK607069S, RPK607091S, RPK607116S, RPK607262S, RPK607267S, RPK607305S, RPK607372S, RPK607499S, RPK607526S, RPK607599S, RPK607737S, RPK607836S, RPK607837S, RPK607905S, RPK608239S, RPK608258S, RPK608327S, RPK608901S, RPK608905S, RPK609087S, RPK603027S, RPK604838S, RPK605368S, RPK605569S, RPK603089S, RPK603594S, RPK604232S, RPK603255S, RPK603389S, RPK603440S, RPK607181S, RPK605054S, RPK603323S, RPK604952S, RPK606694S, RPK602889S, RPK603277S, RPK604249S, RPK604620S, RPK603093S, RPK606471S, RPK602898S, RPK602909S, RPK603230S, RPK603527S, RPK604276S, RPK607058S, RPK605991S, RPK602606S, RPK602767S, RPK602780S, RPK602796S, RPK602853S, RPK603067S, RPK603163S, RPK603457S, RPK604124S, RPK604140S, RPK604152S, RPK604872S, RPK605010S, RPK605018S, RPK606034S, RPK607439S, RPK607831S, RPK608156S, RPK608410S, RPK608427S, RPK608602S, RPK603491S, RPK603739S, RPK606918S, RPK607252S, RPK608077S, RPK608591S, RPK603239S, RPK603322S, RPK605041S, RPK605142S, RPK605845S, RPK606286S, RPK606293S, RPK606435S, RPK606439S, RPK602809S, RPK602844S, RPK603474S, RPK603620S, RPK603625S, RPK603831S, RPK608453S, RPK602557S, RPK603233S, RPK606313S, RPK608454S, RPK608686S, RPK608774S, RPK608782S, RPK603623S, RPK604461S, RPK605239S, RPK607788S, RPK602572S, RPK606132S, RPK603194S, RPK603432S, RPK607774S, RPK602638S, RPK604596S, RPK602837S, RPK602891S, RPK603096S, RPK605561S, RPK606559S, RPK608893S, RPK607630S, RPK608428S, RPK608967S, RPK606396S, RPK604090S, RPK604730S, RPK605373S, RPK605464S, RPK605500S, RPK606020S, RPK607753S, RPK608058S, RPK608509S, RPK608527S, RPK608791S, RPK603271S, RPK603451S, RPK603487S, RPK603729S, RPK603872S, RPK603879S, RPK604123S, RPK604465S, RPK604582S, RPK604693S, RPK605017S, RPK605029S, RPK605100S, RPK605123S, RPK605196S, RPK605329S, RPK605381S, RPK606161S, RPK606260S, RPK606834S, RPK607402S, RPK607614S, RPK607730S, RPK607741S, RPK608373S, RPK608531S, RPK608592S, RPK608670S, RPK608771S, RPK608954S, RPK609082S, RPK603964S, RPK605403S, RPK605425S, RPK605444S, RPK602854S, RPK602855S, RPK602856S, RPK602885S, RPK602886S, RPK602976S, RPK602979S, RPK602980S, RPK603763S, RPK603764S, RPK603765S, RPK603766S, RPK603767S, RPK603875S, RPK603876S, RPK603877S, RPK604092S, RPK606717S, RPK605379S, RPK605391S, RPK603065S, RPK603698S, RPK605189S, RPK602932S, RPK607752S, RPK607829S, RPK608263S, RPK608497S, RPK608606S, RPK603177S, RPK603508S, RPK603635S, RPK604678S, RPK607649S, RPK603548S, RPK603641S, RPK606009S, RPK603199S, RPK603448S, RPK603553S, RPK604517S, RPK602579S, RPK602585S, RPK603429S, RPK603555S, RPK604144S, RPK606383S, RPK602597S, RPK602768S, RPK603565S, RPK603619S, RPK604697S, RPK604974S, RPK605661S, RPK605975S, RPK606176S, RPK606929S, RPK606937S, RPK605849S, RPK606533S, RPK607738S, RPK602698S, RPK606094S, RPK607084S, RPK608408S, RPK602982S, RPK603624S, RPK605068S, RPK607296S, RPK604744S, RPK606536S, RPK602654S, RPK602515S, RPK602586S, RPK603539S, RPK604014S, RPK604180S, RPK604195S, RPK604211S, RPK605186S, RPK605257S, RPK605258S, RPK605724S, RPK605753S, RPK605863S, RPK606374S, RPK607048S, RPK607241S, RPK607345S, RPK607644S, RPK607662S, RPK607813S, RPK608129S, RPK608174S, RPK608398S, RPK608505S, RPK609021S, RPK603760S, RPK606279S, RPK608586S, RPK607176S, RPK607185S, RPK602676S, RPK603488S, RPK605023S, RPK605039S, RPK605284S, RPK605287S, RPK607157S, RPK607674S, RPK603710S, RPK607891S, RPK603859S, RPK602803S, RPK602679S, RPK602689S, RPK602786S, RPK605352S, RPK606178S, RPK607814S, RPK608020S, RPK608659S, RPK609013S, RPK609070S, RPK603435S, RPK603526S, RPK603532S, RPK604877S, RPK605711S, RPK605757S, RPK608033S, RPK608039S, RPK608535S, RPK604423S, RPK604442S, RPK604821S, RPK603873S, RPK605677S, RPK605906S, RPK606609S, RPK603400S, RPK605081S, RPK605540S, RPK606296S, RPK606953S, RPK607317S, RPK607458S, RPK607598S, RPK607885S, RPK608748S, RPK608991S, RPK605465S, RPK609081S, RPK606671S, RPK607369S, RPK606955S, RPK608911S, RPK608788S, RPK605071S, RPK605231S, RPK608951S, RPK604131S, RPK606216S, RPK602985S, RPK605365S, RPK606585S, RPK608717S, RPK608811S, RPK603556S, RPK604411S, RPK606109S, RPK607065S, RPK608881S, RPK603477S, RPK603110S, RPK607178S, RPK608047S, RPK603586S, RPK605762S, RPK608723S, RPK602733S, RPK603094S, RPK603606S, RPK608906S, RPK602215S, RPK603145S, RPK605380S, RPK607309S, RPK603478S, RPK605308S, RPK604550S, RPK602605S, RPK603109S, RPK607265S, RPK603082S, RPK603105S, RPK606690S, RPK608085S, RPK604674S, RPK605393S, RPK605485S, RPK605514S, RPK603099S, RPK603999S, RPK604721S, RPK605340S, RPK606151S, RPK608417S, RPK607658S, RPK606988S, RPK607584S, RPK607623S, RPK607624S, RPK607625S, RPK607659S, RPK607664S, RPK607853S, RPK607915S, RPK608095S, RPK608097S, RPK608184S, RPK608214S, RPK608288S, RPK608377S, RPK608379S, RPK608421S, RPK608423S, RPK608500S, RPK608524S, RPK608528S, RPK608529S, RPK608565S, RPK608615S, RPK608616S, RPK608631S, RPK608632S, RPK608643S, RPK608678S, RPK608724S, RPK608825S, RPK608844S, RPK608846S, RPK608961S, RPK608969S, RPK608973S, RPK608974S, RPK608998S, RPK609006S, RPK609008S, RPK609035S, RPK609053S, RPK606074S, RPK602839S, RPK604194S, RPK608727S, RPK604871S, RPK605541S, RPK605741S, RPK606593S, RPK606662S, RPK606665S, RPK606697S, RPK606850S, RPK606856S, RPK607042S, RPK607053S, RPK607066S, RPK607079S, RPK607341S, RPK607346S, RPK607348S, RPK607359S, RPK607362S, RPK607396S, RPK607438S, RPK607532S, RPK607810S, RPK608262S, RPK608276S, RPK608374S, RPK608402S, RPK608542S, RPK608611S, RPK608654S, RPK608847S, RPK608956S, RPK608997S, RPK609007S, RPK609052S, RPK609057S, RPK609069S, RPK606465S, RPK603002S, RPK604117S, RPK608094S, RPK602739S, RPK602919S, RPK603214S, RPK603304S, RPK604223S, RPK604261S, RPK604308S, RPK604694S, RPK604807S, RPK605133S, RPK605185S, RPK605279S, RPK605686S, RPK607085S, RPK607652S, RPK607661S, RPK608096S, RPK608272S, RPK608273S, RPK608274S, RPK608380S, RPK608422S, RPK608496S, RPK608625S, RPK608688S, RPK609004S, RPK606491S, RPK604307S, RPK604422S, RPK604876S, RPK607104S, RPK607322S, RPK608411S, RPK609048S, RPK608530S, RPK608685S, RPK607183S, RPK607187S, RPK603130S, RPK603076S, RPK603243S, RPK603340S, RPK602664S, RPK602696S, RPK602744S, RPK602747S, RPK602865S, RPK602975S, RPK603166S, RPK603178S, RPK603298S, RPK604593S, RPK604755S, RPK604834S, RPK607040S, RPK602607S, RPK603036S, RPK603091S, RPK603150S, RPK603341S, RPK604226S, RPK604258S, RPK604492S, RPK604508S, RPK604579S, RPK605461S, RPK605565S, RPK605647S, RPK605865S, RPK606494S, RPK606607S, RPK606640S, RPK608813S, RPK609005S, RPK602704S, RPK602805S, RPK607095S, RPK607660S, RPK603633S, RPK603724S, RPK604233S, RPK604864S, RPK604949S, RPK605188S, RPK605283S, RPK606493S, RPK606570S, RPK607409S, RPK603152S, RPK605479S, RPK606103S, RPK606280S, RPK606419S, RPK606463S, RPK606598S, RPK606827S, RPK607094S, RPK607103S, RPK607105S, RPK607109S, RPK607145S, RPK607172S, RPK607173S, RPK607287S, RPK607394S, RPK607850S, RPK607861S, RPK608400S, RPK608508S, RPK608666S, RPK608698S, RPK602519S, RPK602535S, RPK602565S, RPK602778S, RPK602799S, RPK602978S, RPK603085S, RPK603090S, RPK603151S, RPK603167S, RPK603168S, RPK603173S, RPK603273S, RPK603275S, RPK603320S, RPK604215S, RPK604350S, RPK604366S, RPK604571S, RPK604587S, RPK604993S, RPK605113S, RPK605301S, RPK605313S, RPK605980S, RPK606148S, RPK606667S, RPK606701S, RPK607056S, RPK607108S, RPK607343S, RPK608042S, RPK608201S, RPK608264S, RPK608301S, RPK608501S, RPK608522S, RPK608866S, RPK608999S, RPK609010S, RPK602927S, RPK602880S, RPK606970S, RPK602582S, RPK607284S, RPK605949S, RPK608585S, RPK608993S, RPK602538S, RPK602910S, RPK607643S, RPK606172S, RPK605481S, RPK608053S, RPK605082S, RPK606215S, RPK606971S, RPK608971S, RPK602702S, RPK604132S, RPK607263S, RPK607447S, RPK608966S, RPK602552S, RPK603074S, RPK603121S, RPK603149S, RPK602697S, RPK604433S, RPK604879S, RPK605002S, RPK605179S, RPK605702S, RPK606861S, RPK608601S, RPK608641S, RPK602661S, RPK603051S, RPK606183S, RPK606864S, RPK607721S, RPK607727S, RPK602717S, RPK602750S, RPK602818S, RPK603241S, RPK603415S, RPK604758S, RPK607112S, RPK605324S, RPK603083S, RPK606428S, RPK603692S, RPK604186S, RPK604574S, RPK604729S, RPK608955S, RPK602554S, RPK606511S, RPK608790S, RPK608403S, RPK608855S, RPK603701S, RPK606670S, RPK605794S, RPK603703S, RPK605933S, RPK603551S, RPK604091S, RPK606552S, RPK606788S, RPK607375S, RPK605326S, RPK603007S, RPK605950S, RPK602643S, RPK602680S, RPK603533S, RPK603557S, RPK606911S, RPK603116S, RPK603268S, RPK606781S, RPK608599S, RPK603240S, RPK608442S, RPK609034S, RPK603880S, RPK603652S, RPK602832S, RPK605728S, RPK603073S, RPK603244S, RPK603554S, RPK607098S, RPK602533S, RPK602583S, RPK607180S, RPK607150S, RPK608780S, RPK603846S, RPK605026S, RPK605316S, RPK606406S, RPK606410S, RPK606693S, RPK607755S, RPK608371S, RPK608810S, RPK608856S, RPK606868S, RPK607344S, RPK607822S, RPK604776S, RPK607383S, RPK607826S, RPK605213S, RPK605309S, RPK605727S, RPK605749S, RPK606137S, RPK606586S, RPK606838S, RPK606854S, RPK605336S, RPK608577S, RPK605787S, RPK606611S, RPK604985S, RPK602869S, RPK605771S, RPK606031S, RPK606274S, RPK607655S, RPK608008S, RPK603742S, RPK607390S, RPK607984S, RPK602548S, RPK604684S, RPK604954S, RPK606842S, RPK607174S, RPK607521S, RPK603573S, RPK605488S, RPK606219S, RPK607017S, RPK608404S, RPK603672S, RPK608908S, RPK606510S, RPK608290S, RPK605474S, RPK605934S, RPK607748S, RPK608498S, RPK608610S, RPK604677S, RPK607597S, RPK608781S, RPK602857S, RPK603516S, RPK603525S, RPK603638S, RPK603639S, RPK603640S, RPK602713S, RPK603009S, RPK604616S, RPK604618S, RPK605004S, RPK605312S, RPK605358S, RPK605725S, RPK605726S, RPK606258S, RPK607047S, RPK607488S, RPK608789S, RPK608960S, RPK603671S, RPK604272S, RPK605475S, RPK605649S, RPK605714S, RPK606173S, RPK608534S, RPK608803S, RPK604489S, RPK606197S, RPK603881S, RPK607391S, RPK605303S, RPK602540S, RPK602618S, RPK602938S, RPK604279S, RPK604280S, RPK604282S, RPK606163S, RPK606195S, RPK606196S, RPK606364S, RPK606368S, RPK606967S, RPK607365S, RPK607551S, RPK607552S, RPK607702S, RPK609097S, RPK609101S, RPK606477S, RPK603705S, RPK602703S, RPK603818S, RPK604003S, RPK604113S, RPK604201S, RPK604743S, RPK604893S, RPK605160S, RPK605177S, RPK605315S, RPK605712S, RPK605716S, RPK605742S, RPK605905S, RPK606226S, RPK606425S, RPK606426S, RPK606543S, RPK606591S, RPK606837S, RPK606883S, RPK606980S, RPK606994S, RPK607025S, RPK607096S, RPK607101S, RPK607514S, RPK607645S, RPK606777S, RPK602609S, RPK602710S, RPK603266S, RPK603817S, RPK604418S, RPK607437S, RPK605306S, RPK608963S, RPK602576S, RPK602592S, RPK602593S, RPK603988S, RPK604283S, RPK604357S, RPK604970S, RPK604971S, RPK605187S, RPK605263S, RPK604728S, RPK604210S, RPK604324S, RPK605158S, RPK602687S, RPK603071S, RPK603301S, RPK603302S, RPK605280S, RPK606610S, RPK602521S, RPK603858S, RPK605040S, RPK608566S, RPK608704S, RPK608871S, RPK604873S, RPK604874S, RPK604875S, RPK604880S, RPK605562S, RPK605868S, RPK607349S, RPK607495S, RPK607633S, RPK607665S, RPK607673S, RPK607977S, RPK607978S, RPK604202S, RPK604203S, RPK605435S, RPK603789S, RPK604271S, RPK604293S, RPK605240S, RPK605243S, RPK606288S, RPK606538S, RPK606583S, RPK607408S, RPK607417S, RPK607960S, RPK608070S, RPK608225S, RPK608950S, RPK609015S, RPK609017S, RPK607297S, RPK602573S, RPK602782S, RPK602845S, RPK602991S, RPK602637S, RPK606758S, RPK607064S, RPK602481S, RPK603312S, RPK604367S, RPK604444S, RPK605112S, RPK605138S, RPK605553S, RPK606560S, RPK606592S, RPK606944S, RPK607436S, RPK607553S, RPK607635S, RPK607654S, RPK608822S, RPK608992S, RPK604902S, RPK605492S, RPK606713S, RPK602479S, RPK602482S, RPK602531S, RPK605739S, RPK607876S, RPK603303S, RPK603728S, RPK604188S, RPK604302S, RPK605265S, RPK607841S, RPK608588S, RPK608055S, RPK608075S, RPK603276S, RPK603012S, RPK607364S, RPK608826S, RPK602652S, RPK602800S, RPK602804S, RPK603095S, RPK603775S, RPK605940S, RPK607395S, RPK607784S, RPK602494S, RPK602660S, RPK602993S, RPK603112S, RPK603186S, RPK603274S, RPK603403S, RPK603437S, RPK603513S, RPK603621S, RPK604009S, RPK605366S, RPK607325S, RPK607708S, RPK602829S, RPK605689S, RPK607388S, RPK602615S, RPK606006S, RPK602493S, RPK602518S, RPK602555S, RPK602562S, RPK602663S, RPK602682S, RPK602838S, RPK602860S, RPK602866S, RPK602867S, RPK603008S, RPK603043S, RPK603114S, RPK603115S, RPK603147S, RPK603485S, RPK603514S, RPK603552S, RPK603768S, RPK604104S, RPK604731S, RPK604832S, RPK605038S, RPK605433S, RPK605543S, RPK605836S, RPK605839S, RPK607163S, RPK607897S, RPK608875S, RPK603087S, RPK605405S, RPK602491S, RPK603450S, RPK607585S, RPK607707S, RPK608910S, RPK609003S, RPK603205S, RPK603737S, RPK602578S, RPK602644S, RPK607604S, RPK603723S, RPK602529S, RPK604000S, RPK606193S, RPK607034S, RPK607744S, RPK602877S, RPK606521S, RPK608988S, RPK604768S, RPK608633S, RPK602599S, RPK602754S, RPK602762S, RPK603247S, RPK603248S, RPK603584S, RPK603608S, RPK603804S, RPK603824S, RPK604436S, RPK605020S, RPK605735S, RPK606102S, RPK606175S, RPK606879S, RPK606916S, RPK607035S, RPK607144S, RPK607295S, RPK607688S, RPK607792S, RPK608065S, RPK608078S, RPK608536S, RPK602598S, RPK602715S, RPK602858S, RPK604425S, RPK605045S, RPK605173S, RPK606433S, RPK607273S, RPK602496S, RPK604435S, RPK605423S, RPK606033S, RPK607621S, RPK603628S, RPK603629S, RPK603700S, RPK605844S, RPK608783S, RPK608482S, RPK602651S, RPK602753S, RPK602990S, RPK603704S, RPK604371S, RPK604440S, RPK605349S, RPK606529S, RPK606977S, RPK607859S, RPK608407S, RPK608653S, RPK603727S, RPK603979S, RPK605009S, RPK606841S, RPK608269S, RPK605483S, RPK609078S, RPK607569S, RPK608627S, RPK608848S, RPK606722S, RPK605700S, RPK604430S, RPK608792S, RPK603446S, RPK603456S, RPK607357S, RPK607052S, RPK602823S, RPK602825S, RPK602827S, RPK602830S, RPK604158S, RPK604182S, RPK604306S, RPK604581S, RPK606283S, RPK606284S, RPK608996S, RPK603249S, RPK603265S, RPK603283S, RPK603284S, RPK603285S, RPK604270S, RPK604344S, RPK604345S, RPK605558S, RPK606599S, RPK606605S, RPK602619S, RPK603176S, RPK603805S, RPK604769S, RPK606851S, RPK606519S, RPK606673S, RPK603172S, RPK603436S, RPK605008S, RPK607050S, RPK607320S, RPK606898S, RPK603566S, RPK604898S, RPK605502S, RPK606597S, RPK603004S, RPK604429S, RPK604584S, RPK605650S, RPK606042S, RPK607382S, RPK608703S, RPK605678S, RPK605846S, RPK602571S, RPK602575S, RPK603122S, RPK603128S, RPK603164S, RPK603165S, RPK603855S, RPK603857S, RPK604763S, RPK604773S, RPK604789S, RPK605049S, RPK605052S, RPK605053S, RPK605056S, RPK605057S, RPK605058S, RPK605059S, RPK605063S, RPK608701S, RPK603180S, RPK603321S, RPK604512S, RPK604916S, RPK605184S, RPK606712S, RPK607631S, RPK608278S, RPK608867S, RPK603885S, RPK604328S, RPK605155S, RPK605869S, RPK603118S, RPK602539S, RPK603738S, RPK603887S, RPK604196S, RPK604208S, RPK605462S, RPK605740S, RPK605758S, RPK605760S, RPK605776S, RPK605852S, RPK606011S, RPK606012S, RPK606014S, RPK603066S, RPK606808S, RPK605271S, RPK602899S, RPK603441S, RPK603802S, RPK604926S, RPK605212S, RPK605508S, RPK605659S, RPK605883S, RPK608626S, RPK606545S, RPK602642S, RPK602646S, RPK603442S, RPK608994S, RPK603800S, RPK603884S, RPK604052S, RPK604055S, RPK604510S, RPK604890S, RPK605152S, RPK605493S, RPK605566S, RPK605568S, RPK605596S, RPK605722S, RPK605729S, RPK606027S, RPK606298S, RPK606649S, RPK606651S, RPK606859S, RPK607117S, RPK607151S, RPK607955S, RPK602870S, RPK603684S, RPK603691S, RPK603834S, RPK604530S, RPK604589S, RPK604892S, RPK605145S, RPK605547S, RPK605690S, RPK605743S, RPK605745S, RPK606395S, RPK606416S, RPK606507S, RPK607539S, RPK608758S, RPK608759S, RPK603026S, RPK602508S, RPK602693S, RPK602695S, RPK603132S, RPK603262S, RPK603318S, RPK603098S, RPK603157S, RPK604278S, RPK604791S, RPK605856S, RPK606303S, RPK606450S, RPK606869S, RPK607087S, RPK607818S, RPK607830S, RPK608284S, RPK608786S, RPK602841S, RPK604842S, RPK605205S, RPK605666S, RPK605855S, RPK606029S, RPK606035S, RPK607821S, RPK608861S, RPK605512S, RPK604634S, RPK604635S, RPK607632S, RPK608061S, RPK608995S, RPK602771S, RPK603156S, RPK603826S, RPK604740S, RPK605269S, RPK605374S, RPK606000S, RPK606229S, RPK606462S, RPK606669S, RPK607107S, RPK608287S, RPK608779S, RPK608787S, RPK608533S, RPK608578S, RPK602895S, RPK603141S, RPK603631S, RPK605278S, RPK605477S, RPK606004S, RPK606186S, RPK606446S, RPK606708S, RPK606940S, RPK606945S, RPK608590S, RPK608652S, RPK608700S, RPK608726S, RPK605035S, RPK602510S, RPK602996S, RPK602997S, RPK603184S, RPK604431S, RPK604969S, RPK603077S, RPK603843S, RPK604484S, RPK607421S, RPK603308S, RPK605490S, RPK604198S, RPK604899S, RPK604948S, RPK605033S, RPK605274S, RPK605275S, RPK605463S, RPK605567S, RPK605708S, RPK605837S, RPK605967S, RPK605974S, RPK606017S, RPK606256S, RPK606484S, RPK606719S, RPK606790S, RPK607790S, RPK606485S, RPK607072S, RPK602568S, RPK602878S, RPK607102S, RPK608706S, RPK608838S, RPK606323S, RPK606981S, RPK602520S, RPK603694S, RPK604125S, RPK604253S, RPK605135S, RPK605259S, RPK605482S, RPK605847S, RPK607648S, RPK606191S, RPK608418S, RPK607272S, RPK608657S, RPK602621S, RPK606679S, RPK603175S, RPK603803S, RPK604343S, RPK602649S, RPK605560S, RPK606294S, RPK606548S, RPK606687S, RPK603046S, RPK603300S, RPK603997S, RPK602700S, RPK603044S, RPK604189S, RPK605330S, RPK606742S, RPK606812S, RPK607292S, RPK607321S, RPK608673S, RPK608985S, RPK606889S, RPK605250S, RPK605713S, RPK606761S, RPK604675S, RPK602509S, RPK602522S, RPK603070S, RPK605506S, RPK605945S, RPK606201S, RPK602874S, RPK606558S, RPK608525S, RPK608474S, RPK608495S, RPK608679S, RPK602999S, RPK602798S, RPK609009S, RPK602587S, RPK602596S, RPK602630S, RPK602631S, RPK602632S, RPK602665S, RPK602666S, RPK602677S, RPK602678S, RPK602765S, RPK603029S, RPK603030S, RPK603034S, RPK603064S, RPK603412S, RPK603414S, RPK603426S, RPK603427S, RPK603428S, RPK603434S, RPK603492S, RPK603493S, RPK603494S, RPK603622S, RPK603651S, RPK603695S, RPK603696S, RPK603697S, RPK603706S, RPK603707S, RPK603769S, RPK603770S, RPK603776S, RPK603777S, RPK603778S, RPK603779S, RPK604127S, RPK604129S, RPK604139S, RPK604179S, RPK604224S, RPK604229S, RPK604231S, RPK604287S, RPK604288S, RPK604290S, RPK605161S, RPK605164S, RPK605168S, RPK605180S, RPK605193S, RPK605195S, RPK605202S, RPK606154S, RPK606155S, RPK606156S, RPK606373S, RPK606378S, RPK606403S, RPK606440S, RPK606441S, RPK606895S, RPK606921S, RPK606998S, RPK607000S, RPK607003S, RPK607006S, RPK607055S, RPK607271S, RPK607577S, RPK607579S, RPK607581S, RPK607817S, RPK608232S, RPK608447S, RPK608448S, RPK608699S, RPK608793S, RPK608876S, RPK608940S, RPK608949S, RPK608947S, RPK608964S, RPK608965S, RPK602757S, RPK602940S, RPK603092S, RPK603952S, RPK603985S, RPK604212S, RPK604237S, RPK604239S, RPK604240S, RPK604299S, RPK604404S, RPK604407S, RPK604409S, RPK605546S, RPK605969S, RPK606022S, RPK606312S, RPK606458S, RPK606728S, RPK606730S, RPK608289S, RPK609051S, RPK604696S, RPK602604S, RPK603059S, RPK603181S, RPK604427S, RPK604645S, RPK605639S, RPK606015S, RPK606363S, RPK606407S, RPK606875S, RPK607705S, RPK608548S, RPK608752S, RPK609072S, RPK602547S, RPK606677S, RPK606948S, RPK603047S, RPK603297S, RPK603314S, RPK603325S, RPK603327S, RPK604437S, RPK605030S, RPK606405S, RPK603153S, RPK604700S, RPK605648S, RPK605867S, RPK607198S, RPK605404S, RPK605510S, RPK605576S, RPK605577S, RPK605600S, RPK605823S, RPK605825S, RPK605834S, RPK605880S, RPK605881S, RPK605882S, RPK605884S, RPK605890S, RPK606086S, RPK606087S, RPK606100S, RPK606101S, RPK606346S, RPK606355S, RPK607205S, RPK607208S, RPK607209S, RPK607210S, RPK607211S, RPK607212S, RPK607419S, RPK607445S, RPK607446S, RPK607448S, RPK607474S, RPK607475S, RPK607477S, RPK607478S, RPK607898S, RPK607899S, RPK608014S, RPK608015S, RPK608130S, RPK608131S, RPK608132S, RPK603373S, RPK604682S, RPK604992S, RPK605557S, RPK606480S, RPK607969S, RPK608935S, RPK604965S, RPK607044S, RPK603148S, RPK603162S, RPK603188S, RPK604121S, RPK604312S, RPK604402S, RPK604754S, RPK604765S, RPK604772S, RPK604779S, RPK605176S, RPK605451S, RPK605824S, RPK605892S, RPK605896S, RPK605901S, RPK606321S, RPK606330S, RPK606755S, RPK606874S, RPK608200S, RPK608223S, RPK608750S, RPK602487S, RPK602489S, RPK602925S, RPK602949S, RPK602956S, RPK602957S, RPK602958S, RPK602961S, RPK602962S, RPK602963S, RPK602964S, RPK602965S, RPK602970S, RPK602971S, RPK602972S, RPK602983S, RPK602988S, RPK602994S, RPK603003S, RPK603006S, RPK603020S, RPK603342S, RPK603343S, RPK603344S, RPK603345S, RPK603346S, RPK603347S, RPK603348S, RPK603349S, RPK603350S, RPK603351S, RPK603352S, RPK603353S, RPK603354S, RPK603355S, RPK603356S, RPK603425S, RPK603465S, RPK603471S, RPK603472S, RPK603484S, RPK603521S, RPK604265S, RPK604266S, RPK604267S, RPK604268S, RPK604269S, RPK604291S, RPK604323S, RPK604326S, RPK604327S, RPK604329S, RPK604333S, RPK604335S, RPK604338S, RPK604339S, RPK604341S, RPK604346S, RPK604348S, RPK604349S, RPK604723S, RPK604934S, RPK604937S, RPK604941S, RPK604998S, RPK605078S, RPK605079S, RPK605087S, RPK605088S, RPK605089S, RPK605092S, RPK605096S, RPK605097S, RPK605098S, RPK605099S, RPK606358S, RPK606360S, RPK606385S, RPK606386S, RPK606393S, RPK606397S, RPK606782S, RPK606785S, RPK606787S, RPK606963S, RPK606965S, RPK607204S, RPK607249S, RPK607403S, RPK607404S, RPK607776S, RPK607779S, RPK607780S, RPK607781S, RPK608393S, RPK608394S, RPK608396S, RPK608397S, RPK608478S, RPK608539S, RPK608647S, RPK608648S, RPK603408S, RPK603424S, RPK603444S, RPK603453S, RPK603888S, RPK604048S, RPK604054S, RPK604058S, RPK604063S, RPK604242S, RPK607732S, RPK607733S, RPK608808S, RPK608817S, RPK608820S, RPK608823S, RPK604094S, RPK604107S, RPK604110S, RPK604490S, RPK604491S, RPK605074S, RPK605169S, RPK605421S, RPK605429S, RPK605796S, RPK605903S, RPK606097S, RPK606319S, RPK606365S, RPK606789S, RPK606835S, RPK606896S, RPK606901S, RPK606917S, RPK606975S, RPK607353S, RPK607773S, RPK607796S, RPK608562S, RPK608564S, RPK608609S, RPK608799S, RPK609093S, RPK609095S, RPK603335S, RPK603336S, RPK603401S, RPK603431S, RPK603452S, RPK603458S, RPK603480S, RPK603750S, RPK603886S, RPK604025S, RPK604120S, RPK604134S, RPK604141S, RPK604382S, RPK604384S, RPK604386S, RPK604393S, RPK604479S, RPK604486S, RPK604487S, RPK604488S, RPK604766S, RPK604767S, RPK604829S, RPK604835S, RPK604837S, RPK604852S, RPK604878S, RPK604891S, RPK604897S, RPK604913S, RPK605044S, RPK605125S, RPK605126S, RPK605141S, RPK605147S, RPK605175S, RPK605362S, RPK605389S, RPK605418S, RPK605437S, RPK605456S, RPK605746S, RPK605754S, RPK605826S, RPK605857S, RPK605895S, RPK605926S, RPK606092S, RPK606099S, RPK606107S, RPK606158S, RPK606167S, RPK606337S, RPK606340S, RPK606366S, RPK606367S, RPK606382S, RPK606754S, RPK606780S, RPK606839S, RPK606849S, RPK606872S, RPK606885S, RPK606890S, RPK606891S, RPK606899S, RPK607316S, RPK607340S, RPK607347S, RPK607414S, RPK607795S, RPK608059S, RPK608543S, RPK608671S, RPK608680S, RPK609085S, RPK603333S, RPK604477S, RPK604137S, RPK605410S, RPK605769S, RPK606757S, RPK606908S, RPK607749S, RPK608392S, RPK608836S, RPK603010S, RPK603013S, RPK604398S, RPK606165S, RPK608770S, RPK602506S, RPK602507S, RPK602812S, RPK603001S, RPK603017S, RPK603334S, RPK603395S, RPK604260S, RPK604485S, RPK604844S, RPK605398S, RPK605419S, RPK605720S, RPK605930S, RPK606144S, RPK606372S, RPK606720S, RPK606852S, RPK606958S, RPK608399S, RPK608675S, RPK603965S, RPK604049S, RPK604128S, RPK604315S, RPK604321S, RPK604506S, RPK604515S, RPK604516S, RPK604520S, RPK604521S, RPK605914S, RPK605915S, RPK606751S, RPK606752S, RPK606763S, RPK606764S, RPK606765S, RPK606766S, RPK606769S, RPK606772S, RPK606773S, RPK606778S, RPK602492S, RPK602497S, RPK604380S, RPK604397S, RPK604856S, RPK605043S, RPK605822S, RPK606894S, RPK607290S, RPK607367S, RPK607368S, RPK607370S, RPK607373S, RPK607380S, RPK607385S, RPK607798S, RPK607804S, RPK608383S, RPK608476S, RPK608760S, RPK608821S, RPK609088S, RPK608555S, RPK608048S, RPK608681S, RPK608755S, RPK606049S, RPK606055S, RPK608753S, RPK605858S, RPK605893S, RPK605894S, RPK606228S, RPK607269S, RPK607399S, RPK608067S, RPK608260S, RPK608773S, RPK605046S, RPK605065S, RPK605069S, RPK605132S, RPK606067S, RPK606078S, RPK606084S, RPK606380S, RPK607400S, RPK607401S, RPK605402S, RPK605407S, RPK605409S, RPK605455S, RPK605759S, RPK606232S, RPK606249S, RPK606339S, RPK606401S, RPK607760S, RPK607797S, RPK604089S, RPK604857S, RPK604863S, RPK605401S, RPK605706S, RPK605900S, RPK606072S, RPK606126S, RPK606244S, RPK606350S, RPK606353S, RPK606357S, RPK606750S, RPK607304S, RPK607783S, RPK608677S, RPK602658S, RPK604073S, RPK604075S, RPK604115S, RPK604381S, RPK604503S, RPK604540S, RPK604883S, RPK605139S, RPK605721S, RPK605925S, RPK606299S, RPK606306S, RPK606867S, RPK608683S, RPK603413S, RPK603417S, RPK603443S, RPK604067S, RPK604401S, RPK604501S, RPK604925S, RPK604928S, RPK605163S, RPK605392S, RPK605897S, RPK605899S, RPK606210S, RPK606295S, RPK606756S, RPK606760S, RPK606762S, RPK606847S, RPK607282S, RPK607298S, RPK607355S, RPK607786S, RPK607789S, RPK607807S, RPK608409S, RPK608563S, RPK608676S, RPK608684S, RPK608746S, RPK608796S, RPK608050S, RPK608051S, RPK608052S, RPK608056S, RPK607791S, RPK607799S, RPK608390S, RPK608413S, RPK608416S, RPK602486S, RPK603018S, RPK603019S, RPK603366S, RPK608762S, RPK602922S, RPK602924S, RPK603357S, RPK604300S, RPK604304S, RPK604522S, RPK604741S, RPK604747S, RPK604753S, RPK602636S, RPK602802S, RPK602806S, RPK602807S, RPK603189S, RPK603394S, RPK603460S, RPK606213S, RPK606238S, RPK606245S, RPK606292S, RPK606361S, RPK606848S, RPK606882S, RPK606954S, RPK606961S, RPK607280S, RPK607689S, RPK608057S, RPK608387S, RPK608556S, RPK608560S, RPK608649S, RPK608658S, RPK608749S, RPK606164S, RPK606344S, RPK608060S, RPK608254S, RPK603182S, RPK604051S, RPK604076S, RPK604494S, RPK605140S, RPK606157S, RPK606320S, RPK608775S, RPK604762S, RPK602478S, RPK602500S, RPK603364S, RPK603404S, RPK603523S, RPK603892S, RPK603906S, RPK604392S, RPK604400S, RPK604499S, RPK604784S, RPK604786S, RPK604787S, RPK604788S, RPK605443S, RPK605752S, RPK605902S, RPK606166S, RPK606236S, RPK606322S, RPK606333S, RPK606973S, RPK607729S, RPK607801S, RPK608066S, RPK608572S, RPK604244S, RPK604257S, RPK604259S, RPK604782S, RPK605354S, RPK606160S, RPK606833S, RPK607389S, RPK608238S, RPK608249S, RPK608655S, RPK605738S, RPK608888S, RPK607160S, RPK603497S, RPK603690S, RPK604739S, RPK605375S, RPK605862S, RPK606532S, RPK608354S, RPK608593S, RPK603828S, RPK608869S, RPK605083S, RPK605093S, RPK606746S, RPK605111S, RPK605676S, RPK603033S, RPK603390S, RPK603693S, RPK604718S, RPK603504S, RPK604252S, RPK606177S, RPK606486S, RPK607706S, RPK608090S, RPK608863S, RPK605996S, RPK603378S, RPK603409S, RPK603476S, RPK603558S, RPK603970S, RPK604001S, RPK604175S, RPK604636S, RPK605293S, RPK605298S, RPK606417S, RPK606800S, RPK606806S, RPK606933S, RPK606949S, RPK607090S, RPK607640S, RPK607919S, RPK607971S, RPK608064S, RPK608887S, RPK603161S, RPK607299S, RPK604105S, RPK604111S, RPK604112S, RPK604241S, RPK604243S, RPK604505S, RPK605209S, RPK605937S, RPK607119S, RPK608914S, RPK604953S, RPK605146S, RPK605786S, RPK608120S, RPK607051S, RPK608044S, RPK603084S, RPK603250S, RPK605476S, RPK605572S, RPK605717S, RPK606174S, RPK606257S, RPK606342S, RPK606683S, RPK607113S, RPK607840S, RPK608862S, RPK604146S, RPK606844S, RPK606845S, RPK609036S, RPK605480S, RPK607491S, RPK605416S, RPK607152S, RPK607153S, RPK607642S, RPK608146S, RPK608152S, RPK604866S, RPK605936S, RPK606619S, RPK608515S, RPK602904S, RPK603106S, RPK603139S, RPK603616S, RPK602861S, RPK602873S, RPK603496S, RPK603626S, RPK603810S, RPK603902S, RPK605006S, RPK605317S, RPK605331S, RPK605484S, RPK606194S, RPK606602S, RPK602589S, RPK603791S, RPK603459S, RPK603889S, RPK603015S, RPK603253S, RPK603339S, RPK603971S, RPK604732S, RPK605149S, RPK605151S, RPK605256S, RPK605489S, RPK605559S, RPK605859S, RPK605861S, RPK605992S, RPK605939S, RPK606285S, RPK603844S, RPK607060S, RPK606010S, RPK602513S, RPK605153S, RPK605564S, RPK605851S, RPK606447S, RPK606922S, RPK607336S, RPK607819S, RPK608098S, RPK607429S, RPK607515S, RPK608455S, RPK603722S, RPK603795S, RPK605199S, RPK605957S, RPK606079S, RPK604161S, RPK607949S, RPK606923S, RPK608389S, RPK608432S, RPK608438S, RPK603367S, RPK603730S, RPK605645S, RPK606093S, RPK606095S, RPK606356S, RPK606242S, RPK606571S, RPK602863S, RPK603719S, RPK604849S, RPK606171S, RPK606476S, RPK606805S, RPK607013S, RPK608605S, RPK608636S, RPK602974S, RPK603599S, RPK603600S, RPK603686S, RPK603687S, RPK605106S, RPK606928S, RPK602981S, RPK603445S, RPK603748S, RPK604927S, RPK604978S, RPK606121S, RPK606540S, RPK603647S, RPK604936S, RPK605527S, RPK606814S, RPK606959S, RPK609022S, RPK609025S, RPK602901S, RPK603758S, RPK604785S, RPK605094S, RPK605960S, RPK606205S, RPK603754S, RPK605426S, RPK609030S, RPK609032S, RPK603517S, RPK606541S, RPK607077S, RPK603646S, RPK606587S, RPK606927S, RPK605431S, RPK607125S, RPK608175S, RPK608342S, RPK608330S, RPK602683S, RPK602719S, RPK602721S, RPK603718S, RPK603745S, RPK603757S, RPK603806S, RPK604289S, RPK604933S, RPK605080S, RPK605522S, RPK605923S, RPK605962S, RPK606546S, RPK606741S, RPK606804S, RPK607158S, RPK608622S, RPK602790S, RPK603658S, RPK602503S, RPK608587S, RPK609086S, RPK609094S, RPK605447S, RPK605266S, RPK606623S, RPK606220S, RPK608444S, RPK603590S, RPK604101S, RPK608443S, RPK609102S, RPK602756S, RPK602772S, RPK603662S, RPK602723S, RPK604848S, RPK605798S, RPK609023S, RPK603288S, RPK606729S, RPK604841S, RPK606661S, RPK609049S, RPK604190S, RPK606786S, RPK607071S, RPK605870S, RPK607451S, RPK607557S, RPK603227S, RPK606725S, RPK602788S, RPK603667S, RPK603781S, RPK602504S, RPK602709S, RPK603190S, RPK603226S, RPK603229S, RPK603256S, RPK603257S, RPK603399S, RPK603482S, RPK603483S, RPK603574S, RPK603577S, RPK603841S, RPK603842S, RPK603957S, RPK605014S, RPK606352S, RPK607237S, RPK607742S, RPK608173S, RPK602969S, RPK603809S, RPK605446S, RPK605452S, RPK609055S, RPK602773S, RPK603659S, RPK602783S, RPK603575S, RPK603829S, RPK604173S, RPK605013S, RPK605267S, RPK606627S, RPK607253S, RPK604157S, RPK605871S, RPK606524S, RPK607070S, RPK607894S, RPK605920S, RPK605922S, RPK606206S, RPK606996S, RPK609067S, RPK609100S, RPK604080S, RPK602686S, RPK603111S, RPK603216S, RPK603794S, RPK604156S, RPK605201S, RPK606272S, RPK606503S, RPK602879S, RPK605791S, RPK608672S, RPK608702S, RPK606038S, RPK606354S, RPK606453S, RPK605428S, RPK602685S, RPK604454S, RPK606040S, RPK606253S, RPK608347S, RPK608589S, RPK608832S, RPK609001S, RPK602706S, RPK603773S, RPK604184S, RPK604778S, RPK605198S, RPK605504S, RPK606001S, RPK607593S, RPK608435S, RPK608518S, RPK608831S, RPK609002S, RPK603570S, RPK606892S, RPK602707S, RPK604447S, RPK605015S, RPK605733S, RPK605874S, RPK605898S, RPK606227S, RPK607455S, RPK607559S, RPK608433S, RPK608576S, RPK609000S, RPK609073S, RPK606656S, RPK607875S, RPK608177S, RPK606327S, RPK609064S, RPK602751S, RPK602770S, RPK602789S, RPK602794S, RPK603576S, RPK603664S, RPK603665S, RPK603666S, RPK602795S, RPK603663S, RPK602987S, RPK603368S, RPK603537S, RPK603538S, RPK603541S, RPK605524S, RPK606522S, RPK609027S, RPK603578S, RPK603721S, RPK605197S, RPK608634S, RPK608827S, RPK603782S, RPK602769S, RPK604840S, RPK603358S, RPK604102S, RPK605016S, RPK606979S, RPK603290S, RPK604178S, RPK607996S, RPK603821S, RPK608445S, RPK608608S, RPK603830S, RPK607893S, RPK608440S, RPK603588S, RPK602705S, RPK603287S, RPK604097S, RPK607168S, RPK607564S, RPK608584S, RPK603772S, RPK603784S, RPK603785S, RPK603286S, RPK603601S, RPK605101S, RPK605528S, RPK606246S, RPK608718S, RPK608721S, RPK603291S, RPK603688S, RPK603542S, RPK603685S, RPK605105S, RPK605338S, RPK605450S, RPK605530S, RPK606203S, RPK606243S, RPK606251S, RPK606459S, RPK606553S, RPK606956S, RPK606987S, RPK609037S, RPK609065S, RPK609098S, RPK609099S, RPK609033S, RPK602738S, RPK603702S, RPK604496S, RPK606986S, RPK603644S, RPK603469S, RPK603473S, RPK605372S, RPK605961S, RPK606248S, RPK606250S, RPK607028S, RPK603359S, RPK603397S, RPK603470S, RPK603509S, RPK603751S, RPK603755S, RPK604931S, RPK604946S, RPK605664S, RPK608663S, RPK609039S, RPK609066S, RPK602745S, RPK605427S, RPK602720S, RPK603749S, RPK604281S, RPK604469S, RPK605554S, RPK605802S, RPK606149S, RPK606204S, RPK606809S, RPK603712S, RPK605107S, RPK605959S, RPK606442S, RPK602743S, RPK602755S, RPK602776S, RPK606169S, RPK607083S, RPK603753S, RPK606674S, RPK606811S, RPK607011S, RPK607100S, RPK608619S, RPK602787S, RPK602862S, RPK603281S, RPK605388S, RPK605529S, RPK605747S, RPK605814S, RPK608604S, RPK608839S, RPK602746S, RPK602968S, RPK603467S, RPK605104S, RPK605095S, RPK602779S, RPK603225S, RPK604166S, RPK605183S, RPK605386S, RPK604277S, RPK603853S, RPK604483S, RPK607164S, RPK603990S, RPK602848S, RPK603832S, RPK605420S, RPK605989S, RPK607657S, RPK608669S, RPK604019S, RPK604020S, RPK608311S, RPK608386S, RPK603468S, RPK605200S, RPK602656S, RPK606919S, RPK607002S, RPK607318S, RPK607397S, RPK607549S, RPK607867S, RPK607881S, RPK608046S, RPK608300S, RPK609041S, RPK609058S, RPK606783S, RPK607610S, RPK607182S, RPK607513S, RPK607535S, RPK607545S, RPK607945S, RPK607946S, RPK608178S, RPK608299S, RPK608385S, RPK608612S, RPK607680S, RPK607964S, RPK608054S, RPK608215S, RPK609047S, RPK607956S, RPK608103S, RPK609061S, RPK606870S, RPK607596S, RPK607600S, RPK608302S, RPK608664S, RPK609059S, RPK609060S, RPK609038S, RPK608401S, RPK607866S, RPK607868S, RPK607427S, RPK604027S, RPK603898S, RPK603903S, RPK603904S, RPK603907S, RPK603910S, RPK603929S, RPK603977S, RPK603992S, RPK603993S, RPK603995S, RPK603996S, RPK604473S, RPK604474S, RPK604559S, RPK604794S, RPK604797S, RPK604798S, RPK604799S, RPK604804S, RPK604805S, RPK604808S, RPK604811S, RPK604812S, RPK604813S, RPK604815S, RPK604817S, RPK604818S, RPK604820S, RPK604827S, RPK604828S, RPK604830S, RPK604846S, RPK604850S, RPK604912S, RPK605918S, RPK605919S, RPK606114S, RPK606115S, RPK606301S, RPK606315S, RPK607110S, RPK607118S, RPK607127S, RPK607256S, RPK607258S, RPK607501S, RPK607502S, RPK607558S, RPK607560S, RPK607590S, RPK607473S, RPK607509S, RPK603583S, RPK604045S, RPK605655S, RPK605656S, RPK607904S, RPK607931S, RPK607932S, RPK607933S, RPK608312S, RPK606269S, RPK606270S, RPK606277S, RPK606289S, RPK606290S, RPK607310S, RPK607314S, RPK607456S, RPK607457S, RPK607459S, RPK607460S, RPK607462S, RPK607463S, RPK607983S, RPK607985S, RPK607986S, RPK607987S, RPK608007S, RPK608009S, RPK608011S, RPK608012S, RPK608028S, RPK608029S, RPK608032S, RPK608189S, RPK608191S, RPK608197S, RPK608198S, RPK608199S, RPK608313S, RPK608314S, RPK608315S, RPK608316S, RPK608318S, RPK608319S, RPK603950S, RPK603953S, RPK604216S, RPK604217S, RPK604218S, RPK604221S, RPK604594S, RPK604595S, RPK604597S, RPK604598S, RPK604599S, RPK605665S, RPK605667S, RPK605671S, RPK605675S, RPK605835S, RPK606037S, RPK606073S, RPK606075S, RPK607194S, RPK607694S, RPK607944S, RPK607948S, RPK608142S, RPK608145S, RPK603733S, RPK604524S, RPK605973S, RPK606036S, RPK606039S, RPK607469S, RPK607471S, RPK607472S, RPK608259S, RPK607498S, RPK607500S, RPK604922S, RPK604390S, RPK604391S, RPK604448S, RPK604450S, RPK604452S, RPK604456S, RPK604463S, RPK604464S, RPK604470S, RPK604471S, RPK604472S, RPK604560S, RPK604561S, RPK604563S, RPK604566S, RPK604567S, RPK605601S, RPK607431S, RPK607444S, RPK607703S, RPK607710S, RPK607711S, RPK607712S, RPK607713S, RPK607716S, RPK607717S, RPK608139S, RPK608140S, RPK608185S, RPK608188S, RPK608202S, RPK608305S, RPK608306S, RPK608307S, RPK604924S, RPK605581S, RPK605594S, RPK605744S, RPK605804S, RPK605807S, RPK605810S, RPK605811S, RPK605812S, RPK605813S, RPK605815S, RPK605817S, RPK607312S, RPK607313S, RPK608043S, RPK608137S, RPK608151S, RPK608153S, RPK608154S, RPK608155S, RPK608157S, RPK608160S, RPK608230S, RPK608240S, RPK608241S, RPK608242S, RPK608243S, RPK608244S, RPK603896S, RPK603911S, RPK603943S, RPK603944S, RPK603945S, RPK603946S, RPK603948S, RPK604028S, RPK604029S, RPK604030S, RPK604031S, RPK604037S, RPK604318S, RPK604468S, RPK604525S, RPK604538S, RPK604554S, RPK604703S, RPK604704S, RPK604705S, RPK604855S, RPK604860S, RPK605586S, RPK605587S, RPK605589S, RPK605593S, RPK605626S, RPK605628S, RPK605629S, RPK605631S, RPK605633S, RPK605634S, RPK606054S, RPK606119S, RPK606122S, RPK606124S, RPK606127S, RPK606129S, RPK606130S, RPK607582S, RPK607583S, RPK607587S, RPK607588S, RPK607589S, RPK607669S, RPK607670S, RPK607697S, RPK607698S, RPK607701S, RPK607718S, RPK607746S, RPK607747S, RPK607934S, RPK607935S, RPK607937S, RPK607938S, RPK607988S, RPK607990S, RPK607992S, RPK607993S, RPK608123S, RPK608126S, RPK608127S, RPK608128S, RPK603955S, RPK603956S, RPK604544S, RPK604764S, RPK606083S, RPK606085S, RPK606691S, RPK606695S, RPK606700S, RPK606702S, RPK606703S, RPK606704S, RPK607497S, RPK607507S, RPK607508S, RPK607872S, RPK607873S, RPK607874S, RPK607877S, RPK607878S, RPK607950S, RPK607951S, RPK607958S, RPK608309S, RPK604580S, RPK607672S, RPK604013S, RPK606444S, RPK604026S, RPK604043S, RPK604558S, RPK604709S, RPK604935S, RPK605507S, RPK605575S, RPK605887S, RPK605888S, RPK605889S, RPK606066S, RPK606117S, RPK606170S, RPK606688S, RPK607126S, RPK607139S, RPK607196S, RPK607199S, RPK607224S, RPK607449S, RPK607452S, RPK607466S, RPK607511S, RPK607512S, RPK607529S, RPK607530S, RPK607533S, RPK607555S, RPK607556S, RPK607634S, RPK607667S, RPK607668S, RPK607681S, RPK607941S, RPK607942S, RPK608068S, RPK608107S, RPK608108S, RPK608109S, RPK608110S, RPK608111S, RPK608112S, RPK608114S, RPK608119S, RPK608121S, RPK608122S, RPK608147S, RPK608367S, RPK606815S, RPK606443S, RPK604783S, RPK603850S, RPK606581S, RPK608434S, RPK606648S, RPK608363S, RPK608376S, RPK605770S, RPK603506S, RPK603894S, RPK603954S, RPK604475S, RPK604476S, RPK604541S, RPK604545S, RPK605249S, RPK605282S, RPK605457S, RPK605604S, RPK605605S, RPK605682S, RPK605683S, RPK605684S, RPK605695S, RPK605829S, RPK605878S, RPK605879S, RPK607195S, RPK607979S, RPK607980S, RPK608006S, RPK608035S, RPK608036S, RPK608037S, RPK608038S, RPK608180S, RPK604640S, RPK606482S, RPK606857S, RPK602774S, RPK603660S, RPK604109S, RPK605047S, RPK605150S, RPK606394S, RPK606421S, RPK606846S, RPK606932S, RPK606939S, RPK606972S, RPK608597S, RPK608598S, RPK608656S, RPK609089S, RPK604192S, RPK604481S, RPK604647S, RPK604648S, RPK604650S, RPK604653S, RPK605556S, RPK605571S, RPK607167S, RPK607307S, RPK607392S, RPK607412S, RPK607871S, RPK607609S, RPK604792S, RPK602906S, RPK603743S, RPK602675S, RPK604251S, RPK605764S, RPK605765S, RPK605766S, RPK606650S, RPK606652S, RPK608691S, RPK609092S, RPK605225S, RPK607550S, RPK606325S, RPK608213S, RPK608104S, RPK607548S, RPK603924S, RPK604548S, RPK604553S, RPK604706S, RPK604708S, RPK604862S, RPK604865S, RPK604923S, RPK606153S, RPK606400S, RPK606420S, RPK607030S, RPK607038S, RPK607039S, RPK607123S, RPK607202S, RPK607505S, RPK607506S, RPK607518S, RPK607226S, RPK609050S, RPK602526S, RPK602868S, RPK603228S, RPK604719S, RPK604938S, RPK604939S, RPK604940S, RPK604942S, RPK604943S, RPK605509S, RPK605511S, RPK605513S, RPK605515S, RPK605519S, RPK605520S, RPK605523S, RPK605526S, RPK605598S, RPK605616S, RPK605617S, RPK605618S, RPK605875S, RPK605877S, RPK605891S, RPK605907S, RPK606058S, RPK606059S, RPK606060S, RPK606061S, RPK606062S, RPK606706S, RPK606709S, RPK606711S, RPK606732S, RPK607014S, RPK607020S, RPK607021S, RPK607024S, RPK607026S, RPK607029S, RPK607213S, RPK607214S, RPK607215S, RPK607216S, RPK607217S, RPK607218S, RPK607219S, RPK607221S, RPK607222S, RPK607311S, RPK607484S, RPK607485S, RPK607486S, RPK607487S, RPK607489S, RPK607490S, RPK607492S, RPK607493S, RPK607494S, RPK607496S, RPK607561S, RPK607562S, RPK607563S, RPK607565S, RPK607566S, RPK607567S, RPK607568S, RPK607570S, RPK607571S, RPK607573S, RPK607870S, RPK607882S, RPK607883S, RPK607888S, RPK607889S, RPK607892S, RPK607895S, RPK607900S, RPK607906S, RPK607907S, RPK607908S, RPK607959S, RPK607961S, RPK607963S, RPK607965S, RPK607966S, RPK607967S, RPK607968S, RPK608013S, RPK608017S, RPK608018S, RPK608019S, RPK608021S, RPK608022S, RPK608023S, RPK608024S, RPK608025S, RPK608026S, RPK608334S, RPK608336S, RPK608337S, RPK608338S, RPK608340S, RPK608341S, RPK608343S, RPK608344S, RPK608345S, RPK608346S, RPK608348S, RPK608349S, RPK608352S, RPK608353S, RPK608355S, RPK608356S, RPK608358S, RPK608359S, RPK608360S, RPK608031S, RPK608167S, RPK607398S, RPK608296S, RPK608668S, RPK605351S, RPK606328S, RPK605432S, RPK604377S, RPK604881S, RPK605350S, RPK602824S, RPK603078S, RPK603295S, RPK604320S, RPK604445S, RPK604962S, RPK605942S, RPK606008S, RPK603069S, RPK603143S, RPK603306S, RPK603579S, RPK603061S, RPK607338S, RPK607426S, RPK602610S, RPK603237S, RPK604176S, RPK607719S, RPK606077S, RPK608436S, RPK603893S, RPK604011S, RPK605602S, RPK605603S, RPK607450S, RPK604040S, RPK607547S, RPK603058S, RPK608124S, RPK603215S, RPK603605S, RPK603799S, RPK603851S, RPK603852S, RPK603925S, RPK603927S, RPK604219S, RPK604379S, RPK604478S, RPK604523S, RPK604528S, RPK607575S, RPK606617S, RPK604187S, RPK605938S, RPK606070S, RPK604079S, RPK604625S, RPK604628S, RPK605621S, RPK605622S, RPK605623S, RPK605624S, RPK605637S, RPK605638S, RPK605640S, RPK606259S, RPK606262S, RPK606268S, RPK606271S, RPK606326S, RPK606500S, RPK606512S, RPK607128S, RPK607131S, RPK607133S, RPK607135S, RPK607146S, RPK607177S, RPK607254S, RPK607480S, RPK607481S, RPK607482S, RPK607525S, RPK607682S, RPK607754S, RPK608034S, RPK608063S, RPK608148S, RPK608150S, RPK608162S, RPK608163S, RPK608165S, RPK608204S, RPK608224S, RPK608226S, RPK608227S, RPK608229S, RPK607315S, RPK604777S, RPK604900S, RPK607764S, RPK602833S, RPK603075S, RPK603234S, RPK603238S, RPK603675S, RPK603813S, RPK603827S, RPK603969S, RPK603984S, RPK603998S, RPK604428S, RPK604632S, RPK604638S, RPK604722S, RPK604770S, RPK604819S, RPK604845S, RPK604915S, RPK605027S, RPK605067S, RPK605119S, RPK605121S, RPK605305S, RPK605345S, RPK605346S, RPK605357S, RPK605371S, RPK605460S, RPK605777S, RPK605931S, RPK605943S, RPK605995S, RPK606028S, RPK606063S, RPK606064S, RPK606200S, RPK606217S, RPK606523S, RPK606531S, RPK606774S, RPK606888S, RPK606904S, RPK606905S, RPK606906S, RPK606990S, RPK607043S, RPK607147S, RPK607384S, RPK607440S, RPK607592S, RPK607731S, RPK607766S, RPK607994S, RPK608186S, RPK608267S, RPK608335S, RPK608384S, RPK608388S, RPK608457S, RPK608514S, RPK608523S, RPK608661S, RPK608662S, RPK608731S, RPK608852S, RPK608886S, RPK608943S, RPK608968S, RPK608990S, RPK609019S, RPK609028S, RPK609031S, RPK609063S, RPK609068S, RPK609091S, RPK604163S, RPK603119S, RPK603136S, RPK603593S, RPK603618S, RPK603899S, RPK604133S, RPK605333S, RPK606873S, RPK607602S, RPK603072S, RPK605990S, RPK603204S, RPK603206S, RPK606096S, RPK607536S, RPK602740S, RPK602908S, RPK603798S, RPK604108S, RPK604932S, RPK604979S, RPK606120S, RPK606247S, RPK606434S, RPK607010S, RPK608600S, RPK608603S, RPK604214S, RPK602916S, RPK603650S, RPK603679S, RPK603680S, RPK603715S, RPK603863S, RPK605055S, RPK605061S, RPK609044S, RPK606474S, RPK602485S, RPK602542S, RPK602543S, RPK602699S, RPK602725S, RPK602817S, RPK602820S, RPK602876S, RPK603005S, RPK603042S, RPK603079S, RPK603108S, RPK603129S, RPK603154S, RPK603260S, RPK603388S, RPK603543S, RPK603569S, RPK604262S, RPK604316S, RPK605697S, RPK605841S, RPK607379S, RPK603550S, RPK607704S, RPK604068S, RPK604715S, RPK604853S, RPK606343S, RPK606907S, RPK608756S, RPK607646S, RPK604143S, RPK602764S, RPK602892S, RPK603837S, RPK603865S, RPK606240S, RPK606739S, RPK603974S, RPK603975S, RPK604372S, RPK606567S, RPK603987S, RPK607413S, RPK605453S, RPK607726S, RPK607914S, RPK603787S, RPK605831S, RPK606666S, RPK608293S, RPK605614S, RPK606389S, RPK604044S, RPK606013S, RPK606134S, RPK607743S, RPK608426S, RPK608504S, RPK603269S, RPK603307S, RPK603986S, RPK604586S, RPK604701S, RPK605395S, RPK606023S, RPK606468S, RPK606508S, RPK606613S, RPK605322S, RPK605653S, RPK605963S, RPK606578S, RPK608488S, RPK602914S, RPK605182S, RPK605208S, RPK606561S, RPK607435S, RPK608280S, RPK602505S, RPK603960S, RPK604053S, RPK606887S, RPK607288S, RPK607360S, RPK608553S, RPK608806S, RPK605909S, RPK605866S, RPK606840S, RPK607676S, RPK608891S, RPK609096S, RPK607962S, RPK608667S, RPK602936S, RPK606076S, RPK606573S, RPK607122S, RPK606563S, RPK606575S, RPK606080S, RPK605191S, RPK606091S, RPK606225S, RPK607022S, RPK608962S, RPK609056S, RPK606241S, RPK606041S, RPK607724S, RPK607735S, RPK607740S, RPK604154S, RPK604155S, RPK604376S, RPK605102S, RPK605816S, RPK608642S, RPK608645S, RPK608467S, RPK603720S, RPK608431S, RPK602920S, RPK603209S, RPK603211S, RPK603212S, RPK603223S, RPK603224S, RPK603328S, RPK603332S, RPK603360S, RPK603361S, RPK603362S, RPK603363S, RPK603365S, RPK603862S, RPK602681S, RPK606624S, RPK606625S, RPK606626S, RPK606628S, RPK606638S, RPK606639S, RPK606642S, RPK606655S, RPK606658S, RPK608506S, RPK608521S, RPK602549S, RPK602616S, RPK602617S, RPK602672S, RPK605157S, RPK605171S, RPK605772S, RPK605773S, RPK605774S, RPK604149S, RPK604205S, RPK604207S, RPK602781S, RPK602942S, RPK602943S, RPK608579S, RPK608285S, RPK602900S, RPK602902S, RPK602905S, RPK602941S, RPK605159S, RPK605162S, RPK605178S, RPK605223S, RPK605588S, RPK605803S, RPK605820S, RPK608693S, RPK608711S, RPK602946S, RPK603213S, RPK603219S, RPK603221S, RPK603222S, RPK604394S, RPK604396S, RPK604995S, RPK606643S, RPK606645S, RPK606646S, RPK602944S, RPK602945S, RPK604793S, RPK605156S, RPK606644S, RPK606654S, RPK602950S, RPK602951S, RPK602952S, RPK602955S, RPK603097S, RPK603102S, RPK603187S, RPK603203S, RPK603217S, RPK603218S, RPK604056S, RPK604100S, RPK604116S, RPK604130S, RPK604168S, RPK604199S, RPK604200S, RPK604204S, RPK604944S, RPK604945S, RPK604996S, RPK604997S, RPK605780S, RPK605781S, RPK605782S, RPK605784S, RPK605789S, RPK605790S, RPK605792S, RPK605793S, RPK605795S, RPK605799S, RPK605800S, RPK602601S, RPK602614S, RPK602748S, RPK602840S, RPK602947S, RPK604041S, RPK604064S, RPK604096S, RPK604098S, RPK604099S, RPK605694S, RPK605779S, RPK606618S, RPK606620S, RPK606621S, RPK606629S, RPK606634S, RPK606635S, RPK606653S, RPK607815S, RPK608169S, RPK608361S, RPK608550S, RPK608551S, RPK608635S, RPK608638S, RPK608640S, RPK608071S, RPK608106S, RPK604679S, RPK608321S, RPK608256S, RPK606633S, RPK606622S, RPK602729S, RPK603193S, RPK603195S, RPK603198S, RPK603201S, RPK603220S, RPK603958S, RPK603959S, RPK602741S, RPK605167S, RPK605170S, RPK605224S, RPK605234S, RPK605238S, RPK605245S, RPK605247S, RPK605767S, RPK605768S, RPK605775S, RPK606630S, RPK606631S, RPK606632S, RPK606659S, RPK607852S, RPK606180S, RPK606181S, RPK606334S, RPK606335S, RPK607129S, RPK607132S, RPK603398S, RPK604005S, RPK602671S, RPK607936S, RPK607606S, RPK606564S, RPK607516S, RPK608176S, RPK606915S, RPK607428S, RPK607612S, RPK607613S, RPK608283S, RPK608282S, RPK607270S, RPK608149S, RPK606584S, RPK606088S, RPK606090S, RPK607223S, RPK605537S, RPK605827S, RPK606082S, RPK607693S, RPK608030S, RPK604034S, RPK604035S, RPK604038S, RPK604069S, RPK604070S, RPK604071S, RPK604072S, RPK604074S, RPK604710S, RPK605396S, RPK605400S, RPK605408S, RPK605430S, RPK607461S, RPK607519S, RPK608166S, RPK608220S, RPK608221S, RPK608222S, RPK603942S, RPK605583S, RPK605584S, RPK603931S, RPK603932S, RPK603933S, RPK603934S, RPK606112S, RPK606113S, RPK607179S, RPK607193S, RPK607240S, RPK607242S, RPK607243S, RPK607464S, RPK607527S, RPK607723S, RPK608117S, RPK608118S, RPK608193S, RPK608194S, RPK608233S, RPK608234S, RPK608235S, RPK608236S, RPK608325S, RPK608329S, RPK608331S, RPK607245S, RPK607246S, RPK607671S, RPK607691S, RPK607722S, RPK608192S, RPK608195S, RPK608196S, RPK608362S, RPK608364S, RPK604681S, RPK606133S, RPK607920S, RPK607616S, RPK603369S, RPK604641S, RPK604652S, RPK605254S, RPK605295S, RPK606418S, RPK606614S, RPK606821S, RPK606824S, RPK606826S, RPK607639S, RPK605110S, RPK605218S, RPK605252S, RPK604549S, RPK604340S, RPK604670S, RPK605300S, RPK607678S, RPK608323S, RPK603627S, RPK604181S, RPK604303S, RPK604365S, RPK605264S, RPK606478S, RPK606660S, RPK606807S, RPK607637S, RPK607910S, RPK603870S, RPK605217S, RPK606989S, RPK604833S, RPK606481S, RPK604990S, RPK606829S, RPK607921S, RPK604325S, RPK604331S, RPK604336S, RPK605129S, RPK608916S, RPK608928S, RPK604658S, RPK604963S, RPK604964S, RPK604966S, RPK604967S, RPK605260S, RPK605643S, RPK605929S, RPK606278S, RPK606415S, RPK608895S, RPK608929S, RPK605644S, RPK608933S, RPK605124S, RPK605261S, RPK608902S, RPK604332S, RPK604077S, RPK604078S, RPK605578S, RPK605580S, RPK605818S, RPK605819S, RPK604673S, RPK606818S, RPK605778S, RPK608798S, RPK604643S, RPK608558S, RPK608569S, RPK608594S, RPK608614S, RPK608689S, RPK608715S, RPK607503S, RPK608489S, RPK603725S, RPK604167S, RPK605321S, RPK604823S, RPK604825S, RPK604831S, RPK605236S, RPK605941S, RPK606136S, RPK606345S, RPK606893S, RPK607186S, RPK608595S, RPK608430S, RPK606897S, RPK606900S, RPK606909S, RPK606910S, RPK606912S, RPK606913S, RPK606914S, RPK606934S, RPK606935S, RPK606936S, RPK607319S, RPK607327S, RPK607328S, RPK607330S, RPK607332S, RPK607333S, RPK607334S, RPK607335S, RPK606413S, RPK607339S, RPK607720S, RPK606828S, RPK608859S, RPK607363S, RPK607366S, RPK607386S, RPK607803S, RPK607805S, RPK608557S, RPK608694S, RPK608830S, RPK605911S, RPK606370S, RPK607538S, RPK608491S, RPK606542S, RPK605024S, RPK605850S, RPK605979S, RPK606021S, RPK606304S, RPK606657S, RPK606408S, RPK607121S, RPK606019S, RPK606957S, RPK607057S, RPK607576S, RPK607692S, RPK607890S, RPK608040S, RPK608141S, RPK608303S, RPK608304S, RPK608317S, RPK608320S, RPK608322S, RPK607879S, RPK604375S, RPK604378S, RPK604383S, RPK604385S, RPK604388S, RPK604757S, RPK604882S, RPK604888S, RPK604894S, RPK606002S, RPK608247S, RPK608248S, RPK606359S, RPK608437S, RPK606016S, RPK607430S, RPK606562S, RPK608292S, RPK607425S, RPK607638S, RPK607522S, RPK608687S, RPK608899S, RPK608295S, RPK604929S, RPK604906S, RPK604742S, RPK604746S, RPK604749S, RPK604752S, RPK604896S, RPK604904S, RPK604907S, RPK604908S, RPK604909S, RPK604910S, RPK604911S, RPK604919S, RPK604920S, RPK604868S, RPK605590S, RPK608133S, RPK605912S, RPK604531S, RPK604537S, RPK604539S, RPK604575S, RPK604583S, RPK604759S, RPK606517S, RPK606802S, RPK608205S, RPK608366S, RPK608253S, RPK608252S, RPK608251S, RPK608245S, RPK608183S, RPK608116S, RPK608003S, RPK608002S, RPK608000S, RPK607999S, RPK607998S, RPK607995S, RPK607687S, RPK607591S, RPK607578S, RPK607574S, RPK607137S, RPK607136S, RPK607037S, RPK607036S, RPK606748S, RPK606747S, RPK606460S, RPK606427S, RPK606331S, RPK605641S, RPK605582S, RPK605535S, RPK605533S; GTIN 00763000206291, Lot Serial numbers: RPK610175S, RPK610302S, RPK612352S, RPK611429S, RPK611837S, RPK611833S, RPK612180S, RPK610478S, RPK610111S, RPK609460S, RPK611355S, RPK612757S, RPK612241S, RPK612760S, RPK612214S, RPK612756S, RPK611554S, RPK612143S, RPK612221S, RPK610236S, RPK612274S, RPK612251S, RPK612066S, RPK612151S, RPK612153S, RPK612229S, RPK611758S, RPK610694S, RPK609844S, RPK610952S, RPK609643S, RPK612602S, RPK611908S, RPK612573S, RPK612585S, RPK611791S, RPK612562S, RPK612024S, RPK609994S, RPK610275S, RPK610574S, RPK610931S, RPK611857S, RPK611923S, RPK611929S, RPK611812S, RPK611436S, RPK611802S, RPK612627S, RPK611166S, RPK612211S, RPK612475S, RPK612477S, RPK612479S, RPK612506S, RPK612431S, RPK612454S, RPK611466S, RPK612640S, RPK611242S, RPK611331S, RPK612529S, RPK612223S, RPK612108S, RPK612645S, RPK612218S, RPK611714S, RPK610739S, RPK610874S, RPK611641S, RPK612281S, RPK612418S, RPK612464S, RPK610870S, RPK610475S, RPK610544S, RPK610595S, RPK610759S, RPK610769S, RPK610877S, RPK612276S, RPK612508S, RPK612542S, RPK612682S, RPK609790S, RPK609797S, RPK611173S, RPK609150S, RPK609694S, RPK609774S, RPK610554S, RPK610758S, RPK611580S, RPK609138S, RPK609260S, RPK609852S, RPK609860S, RPK610146S, RPK610549S, RPK610555S, RPK611358S, RPK611360S, RPK611892S, RPK611893S, RPK612231S, RPK612243S, RPK610901S, RPK609116S, RPK609119S, RPK609709S, RPK609815S, RPK609867S, RPK610882S, RPK610897S, RPK610899S, RPK611356S, RPK609133S, RPK609594S, RPK610267S, RPK611551S, RPK609117S, RPK610223S, RPK610880S, RPK611649S, RPK610900S, RPK609822S, RPK609578S, RPK609114S, RPK609415S, RPK611354S, RPK609112S, RPK609115S, RPK610898S, RPK609868S, RPK609111S, RPK609120S, RPK611737S, RPK611738S, RPK611739S, RPK609320S, RPK609106S, RPK610775S, RPK611060S, RPK610423S, RPK612069S, RPK610426S, RPK610721S, RPK611514S, RPK611690S, RPK612282S, RPK610191S, RPK610108S, RPK611296S, RPK609542S, RPK610827S, RPK609602S, RPK610208S, RPK609601S, RPK609938S, RPK610509S, RPK610320S, RPK610822S, RPK609311S, RPK610472S, RPK610779S, RPK611324S, RPK610912S, RPK610099S, RPK609796S, RPK609625S, RPK610810S, RPK611136S, RPK609474S, RPK610104S, RPK610682S, RPK611520S, RPK611546S, RPK611547S, RPK609545S, RPK611002S, RPK610080S, RPK610109S, RPK610273S, RPK611327S, RPK609171S, RPK610073S, RPK610691S, RPK611243S, RPK612503S, RPK611990S, RPK610132S, RPK610184S, RPK610637S, RPK610808S, RPK610809S, RPK611725S, RPK611750S, RPK609327S, RPK610892S, RPK612261S, RPK609568S, RPK611233S, RPK610095S, RPK609435S, RPK609595S, RPK610068S, RPK611134S, RPK611323S, RPK611995S, RPK611299S, RPK609794S, RPK611513S, RPK610114S, RPK610230S, RPK610891S, RPK611067S, RPK611498S, RPK609357S, RPK611762S, RPK610301S, RPK611512S, RPK612155S, RPK609354S, RPK610079S, RPK611184S, RPK609223S, RPK609355S, RPK609583S, RPK610773S, RPK610024S, RPK610091S, RPK609429S, RPK610806S, RPK609770S, RPK610418S, RPK611001S, RPK612488S, RPK609600S, RPK609394S, RPK609577S, RPK609690S, RPK609702S, RPK609980S, RPK610048S, RPK610134S, RPK610722S, RPK611132S, RPK611518S, RPK611497S, RPK610162S, RPK610815S, RPK611147S, RPK611460S, RPK612216S, RPK612496S, RPK610122S, RPK611770S, RPK612062S, RPK609217S, RPK610167S, RPK609978S, RPK610780S, RPK609722S, RPK609246S, RPK609126S, RPK609766S, RPK610016S, RPK610455S, RPK610888S, RPK611055S, RPK610061S, RPK609533S, RPK609241S, RPK609728S, RPK610631S, RPK610890S, RPK611244S, RPK611515S, RPK610408S, RPK611098S, RPK612210S, RPK609273S, RPK610428S, RPK610627S, RPK611461S, RPK609930S, RPK610232S, RPK610411S, RPK610668S, RPK611190S, RPK611246S, RPK611715S, RPK612246S, RPK610014S, RPK612518S, RPK610129S, RPK610310S, RPK610629S, RPK610817S, RPK611289S, RPK610225S, RPK611226S, RPK609316S, RPK610113S, RPK610510S, RPK610511S, RPK610807S, RPK611004S, RPK611749S, RPK611777S, RPK612672S, RPK609438S, RPK609536S, RPK610828S, RPK611295S, RPK609537S, RPK609772S, RPK609975S, RPK610359S, RPK610390S, RPK610515S, RPK610910S, RPK611304S, RPK611694S, RPK612284S, RPK609626S, RPK610163S, RPK610178S, RPK611068S, RPK609440S, RPK609953S, RPK609984S, RPK610415S, RPK611748S, RPK612018S, RPK610789S, RPK612545S, RPK610783S, RPK610805S, RPK609398S, RPK609603S, RPK610116S, RPK610138S, RPK610369S, RPK611522S, RPK609779S, RPK610410S, RPK611057S, RPK610774S, RPK611701S, RPK609990S, RPK610377S, RPK611187S, RPK611238S, RPK611239S, RPK611273S, RPK609505S, RPK609560S, RPK609562S, RPK609765S, RPK610686S, RPK610826S, RPK611064S, RPK611540S, RPK609306S, RPK609442S, RPK609550S, RPK610206S, RPK609634S, RPK609903S, RPK610656S, RPK610670S, RPK611024S, RPK611026S, RPK611027S, RPK611045S, RPK611047S, RPK611048S, RPK611049S, RPK611078S, RPK611158S, RPK611159S, RPK611161S, RPK611163S, RPK611164S, RPK611165S, RPK611167S, RPK611684S, RPK611686S, RPK609598S, RPK610365S, RPK609109S, RPK609696S, RPK610337S, RPK610751S, RPK609368S, RPK609353S, RPK610269S, RPK609813S, RPK611342S, RPK609893S, RPK611116S, RPK609303S, RPK611711S, RPK611300S, RPK609691S, RPK610454S, RPK609358S, RPK609719S, RPK610150S, RPK611282S, RPK611320S, RPK611536S, RPK610494S, RPK609251S, RPK609145S, RPK609541S, RPK609876S, RPK612072S, RPK609641S, RPK610085S, RPK610149S, RPK610447S, RPK610045S, RPK610481S, RPK610655S, RPK610788S, RPK610715S, RPK610059S, RPK609789S, RPK610654S, RPK609341S, RPK609585S, RPK609745S, RPK610417S, RPK611661S, RPK611689S, RPK609252S, RPK609254S, RPK609255S, RPK609257S, RPK609343S, RPK609350S, RPK609351S, RPK609369S, RPK609392S, RPK609395S, RPK609396S, RPK609443S, RPK609446S, RPK609448S, RPK609449S, RPK609450S, RPK609575S, RPK609607S, RPK609609S, RPK609610S, RPK609627S, RPK609633S, RPK609700S, RPK609726S, RPK609749S, RPK609750S, RPK609751S, RPK609759S, RPK609816S, RPK609827S, RPK609832S, RPK609833S, RPK609834S, RPK609907S, RPK609909S, RPK609911S, RPK609926S, RPK609927S, RPK609928S, RPK610067S, RPK610070S, RPK610071S, RPK610075S, RPK610089S, RPK610090S, RPK610092S, RPK610094S, RPK610101S, RPK610235S, RPK610238S, RPK610241S, RPK610245S, RPK610254S, RPK610331S, RPK610333S, RPK610334S, RPK610419S, RPK610420S, RPK610434S, RPK610435S, RPK610436S, RPK610438S, RPK610445S, RPK610446S, RPK610507S, RPK610526S, RPK610530S, RPK610531S, RPK610532S, RPK610577S, RPK610641S, RPK610643S, RPK610644S, RPK610645S, RPK610646S, RPK610647S, RPK610648S, RPK610649S, RPK610650S, RPK610652S, RPK610653S, RPK610658S, RPK610689S, RPK610690S, RPK610734S, RPK610735S, RPK610736S, RPK610737S, RPK610738S, RPK610742S, RPK610743S, RPK610835S, RPK610836S, RPK610837S, RPK610838S, RPK610839S, RPK610840S, RPK610841S, RPK610843S, RPK610850S, RPK610851S, RPK610852S, RPK610855S, RPK610856S, RPK610858S, RPK610859S, RPK610860S, RPK610861S, RPK610862S, RPK610863S, RPK610864S, RPK610865S, RPK611044S, RPK611080S, RPK611105S, RPK611182S, RPK611194S, RPK611196S, RPK611200S, RPK611279S, RPK611309S, RPK611310S, RPK611311S, RPK611312S, RPK611477S, RPK611479S, RPK611480S, RPK611481S, RPK611483S, RPK611501S, RPK611621S, RPK611622S, RPK611623S, RPK611624S, RPK611625S, RPK611662S, RPK611663S, RPK611664S, RPK611665S, RPK611666S, RPK611669S, RPK611670S, RPK611671S, RPK611672S, RPK611673S, RPK612419S, RPK612421S, RPK612427S, RPK612438S, RPK612439S, RPK612441S, RPK612442S, RPK612446S, RPK609496S, RPK609539S, RPK609540S, RPK609945S, RPK610271S, RPK610313S, RPK610315S, RPK610316S, RPK610501S, RPK610503S, RPK610546S, RPK610667S, RPK610744S, RPK610844S, RPK610845S, RPK610846S, RPK610848S, RPK611629S, RPK611630S, RPK611676S, RPK611759S, RPK611991S, RPK612000S, RPK612001S, RPK610791S, RPK610064S, RPK611506S, RPK610293S, RPK609549S, RPK609218S, RPK609310S, RPK610635S, RPK609559S, RPK609865S, RPK609977S, RPK611492S, RPK609786S, RPK610155S, RPK611704S, RPK610677S, RPK611109S, RPK611326S, RPK609894S, RPK609899S, RPK610015S, RPK610055S, RPK610069S, RPK611667S, RPK612453S, RPK609122S, RPK609235S, RPK609758S, RPK609933S, RPK610651S, RPK610847S, RPK610853S, RPK611660S, RPK611988S, RPK609328S, RPK609584S, RPK609889S, RPK610051S, RPK610327S, RPK610372S, RPK610748S, RPK610818S, RPK610819S, RPK611013S, RPK611321S, RPK611500S, RPK611775S, RPK612003S, RPK612347S, RPK612449S, RPK609661S, RPK609280S, RPK609710S, RPK609755S, RPK609905S, RPK610019S, RPK610406S, RPK610642S, RPK610701S, RPK610740S, RPK610849S, RPK611247S, RPK611626S, RPK611632S, RPK612059S, RPK612095S, RPK612436S, RPK610857S, RPK611351S, RPK611619S, RPK611620S, RPK611627S, RPK611722S, RPK612060S, RPK612256S, RPK610695S, RPK609650S, RPK609563S, RPK609670S, RPK609413S, RPK609530S, RPK609920S, RPK609964S, RPK610304S, RPK610305S, RPK610697S, RPK610699S, RPK612364S, RPK610041S, RPK610698S, RPK611865S, RPK609962S, RPK609639S, RPK609756S, RPK609538S, RPK610706S, RPK611900S, RPK610832S, RPK611712S, RPK611713S, RPK611761S, RPK611772S, RPK610023S, RPK610830S, RPK609948S, RPK611131S, RPK611494S, RPK610363S, RPK609213S, RPK609528S, RPK609754S, RPK609937S, RPK610136S, RPK610413S, RPK610450S, RPK610484S, RPK610684S, RPK610685S, RPK611264S, RPK611319S, RPK611329S, RPK612146S, RPK612252S, RPK612675S, RPK609534S, RPK610803S, RPK611097S, RPK611139S, RPK610180S, RPK609318S, RPK609244S, RPK609630S, RPK609873S, RPK609936S, RPK610106S, RPK610123S, RPK610135S, RPK610186S, RPK610441S, RPK610535S, RPK610733S, RPK611018S, RPK611063S, RPK611100S, RPK612040S, RPK612081S, RPK609118S, RPK609129S, RPK609437S, RPK609462S, RPK609548S, RPK610831S, RPK611066S, RPK611340S, RPK611986S, RPK611997S, RPK612012S, RPK609461S, RPK610713S, RPK611290S, RPK610782S, RPK610821S, RPK609321S, RPK610804S, RPK609543S, RPK611720S, RPK609127S, RPK610177S, RPK611095S, RPK611235S, RPK609250S, RPK609304S, RPK609430S, RPK609546S, RPK609776S, RPK609942S, RPK609979S, RPK610025S, RPK610063S, RPK610103S, RPK610168S, RPK610264S, RPK610353S, RPK610486S, RPK610597S, RPK610664S, RPK610669S, RPK610812S, RPK611010S, RPK611280S, RPK611334S, RPK611335S, RPK611489S, RPK611535S, RPK611462S, RPK611488S, RPK610729S, RPK611523S, RPK610814S, RPK611236S, RPK611700S, RPK610078S, RPK610295S, RPK611099S, RPK610156S, RPK610457S, RPK611332S, RPK609963S, RPK610107S, RPK610375S, RPK610449S, RPK610728S, RPK610876S, RPK610906S, RPK611337S, RPK611341S, RPK611367S, RPK611727S, RPK611781S, RPK611992S, RPK611993S, RPK612417S, RPK609265S, RPK610529S, RPK611017S, RPK611996S, RPK610470S, RPK610630S, RPK609879S, RPK609231S, RPK609363S, RPK609804S, RPK610181S, RPK610416S, RPK611499S, RPK611745S, RPK610703S, RPK611062S, RPK610873S, RPK610384S, RPK610203S, RPK611297S, RPK609781S, RPK609880S, RPK611301S, RPK611456S, RPK610573S, RPK611539S, RPK609554S, RPK611726S, RPK609431S, RPK609221S, RPK610750S, RPK612242S, RPK612262S, RPK612667S, RPK610047S, RPK610050S, RPK610137S, RPK610224S, RPK610905S, RPK611275S, RPK611325S, RPK611357S, RPK612661S, RPK608211S, RPK609372S, RPK609444S, RPK609581S, RPK609590S, RPK609591S, RPK609606S, RPK609622S, RPK609847S, RPK609913S, RPK610056S, RPK610100S, RPK610179S, RPK610325S, RPK610380S, RPK610409S, RPK610443S, RPK610516S, RPK610633S, RPK610679S, RPK610908S, RPK610988S, RPK611006S, RPK611177S, RPK611306S, RPK611529S, RPK611708S, RPK611709S, RPK611731S, RPK611734S, RPK611754S, RPK612010S, RPK612428S, RPK611655S, RPK609592S, RPK609949S, RPK609969S, RPK610200S, RPK610661S, RPK610662S, RPK610726S, RPK610747S, RPK610768S, RPK610770S, RPK610911S, RPK611069S, RPK611241S, RPK611303S, RPK611408S, RPK611420S, RPK611730S, RPK611747S, RPK611760S, RPK612343S, RPK612345S, RPK612420S, RPK612423S, RPK612525S, RPK612536S, RPK610895S, RPK610461S, RPK610712S, RPK610717S, RPK611322S, RPK609420S, RPK609971S, RPK610076S, RPK610125S, RPK610326S, RPK610407S, RPK611769S, RPK612056S, RPK612444S, RPK611469S, RPK610522S, RPK610110S, RPK610065S, RPK612625S, RPK609764S, RPK610340S, RPK610437S, RPK610760S, RPK611012S, RPK611767S, RPK612105S, RPK609970S, RPK609991S, RPK610680S, RPK610725S, RPK612656S, RPK611654S, RPK611656S, RPK611657S, RPK610057S, RPK610066S, RPK610151S, RPK610378S, RPK610903S, RPK611191S, RPK611240S, RPK611336S, RPK610480S, RPK610521S, RPK609767S, RPK610266S, RPK610894S, RPK611771S, RPK612497S, RPK610681S, RPK610088S, RPK610212S, RPK612217S, RPK610157S, RPK611774S, RPK609576S, RPK610097S, RPK610346S, RPK610811S, RPK610373S, RPK610586S, RPK610414S, RPK611277S, RPK611476S, RPK610771S, RPK611106S, RPK610429S, RPK609689S, RPK610585S, RPK610188S, RPK609795S, RPK610425S, RPK610794S, RPK610386S, RPK609678S, RPK609775S, RPK610474S, RPK610636S, RPK612057S, RPK609687S, RPK609778S, RPK610096S, RPK611658S, RPK611705S, RPK611742S, RPK611743S, RPK611732S, RPK611733S, RPK611496S, RPK609787S, RPK609792S, RPK610283S, RPK610284S, RPK610663S, RPK610915S, RPK612445S, RPK612659S, RPK612660S, RPK610493S, RPK609557S, RPK610465S, RPK610518S, RPK612132S, RPK611302S, RPK611463S, RPK611065S, RPK611317S, RPK609631S, RPK610060S, RPK610127S, RPK611573S, RPK612361S, RPK612362S, RPK610553S, RPK609983S, RPK610731S, RPK609242S, RPK610459S, RPK612053S, RPK610174S, RPK610328S, RPK609985S, RPK609278S, RPK609377S, RPK609404S, RPK609286S, RPK609180S, RPK609624S, RPK609858S, RPK611059S, RPK612433S, RPK609346S, RPK610463S, RPK611333S, RPK611680S, RPK611728S, RPK610987S, RPK609780S, RPK611011S, RPK610483S, RPK611318S, RPK610133S, RPK609433S, RPK610829S, RPK611008S, RPK611508S, RPK609348S, RPK609835S, RPK610657S, RPK610405S, RPK610802S, RPK609211S, RPK610202S, RPK610793S, RPK610889S, RPK612669S, RPK610795S, RPK611003S, RPK609326S, RPK609898S, RPK611267S, RPK611330S, RPK610281S, RPK611058S, RPK611283S, RPK611679S, RPK612481S, RPK610182S, RPK610322S, RPK610469S, RPK611056S, RPK611178S, RPK609561S, RPK610781S, RPK612016S, RPK611285S, RPK609314S, RPK609632S, RPK610341S, RPK610402S, RPK610798S, RPK611544S, RPK610683S, RPK611232S, RPK609731S, RPK610528S, RPK610723S, RPK610776S, RPK610468S, RPK609370S, RPK610777S, RPK610842S, RPK609402S, RPK612009S, RPK612490S, RPK609406S, RPK609615S, RPK610749S, RPK611307S, RPK611315S, RPK610338S, RPK611115S, RPK609818S, RPK609955S, RPK609173S, RPK610527S, RPK611101S, RPK612052S, RPK610520S, RPK610676S, RPK609258S, RPK609956S, RPK610309S, RPK610368S, RPK610711S, RPK611234S, RPK610084S, RPK610492S, RPK610164S, RPK610166S, RPK610424S, RPK610834S, RPK611345S, RPK609256S, RPK609768S, RPK609817S, RPK609886S, RPK610102S, RPK610332S, RPK610412S, RPK611051S, RPK611053S, RPK611262S, RPK611263S, RPK611288S, RPK611344S, RPK611471S, RPK612130S, RPK612487S, RPK609447S, RPK610058S, RPK610183S, RPK609172S, RPK609946S, RPK610229S, RPK611148S, RPK611291S, RPK609365S, RPK609897S, RPK610545S, RPK609869S, RPK610005S, RPK611783S, RPK611110S, RPK610342S, RPK611305S, RPK609432S, RPK612236S, RPK609207S, RPK610525S, RPK609896S, RPK609405S, RPK609895S, RPK610081S, RPK610185S, RPK609451S, RPK609452S, RPK610077S, RPK610519S, RPK610634S, RPK610666S, RPK609174S, RPK609360S, RPK610130S, RPK610001S, RPK611061S, RPK610448S, RPK609225S, RPK610508S, RPK609347S, RPK609196S, RPK609240S, RPK609514S, RPK611718S, RPK612228S, RPK610692S, RPK609791S, RPK610541S, RPK609412S, RPK610086S, RPK610173S, RPK611293S, RPK609247S, RPK609262S, RPK609519S, RPK609777S, RPK609798S, RPK609848S, RPK609856S, RPK609973S, RPK610082S, RPK610339S, RPK610371S, RPK610495S, RPK610500S, RPK610523S, RPK610579S, RPK610660S, RPK610665S, RPK610868S, RPK610869S, RPK610871S, RPK612224S, RPK610442S, RPK611237S, RPK611511S, RPK611682S, RPK611473S, RPK611820S, RPK609165S, RPK609167S, RPK609197S, RPK609199S, RPK609209S, RPK609212S, RPK609214S, RPK609295S, RPK609298S, RPK609337S, RPK609338S, RPK609374S, RPK609384S, RPK609393S, RPK609399S, RPK609484S, RPK609511S, RPK609961S, RPK610453S, RPK610456S, RPK610543S, RPK610559S, RPK610560S, RPK610932S, RPK610977S, RPK610978S, RPK610979S, RPK611030S, RPK611041S, RPK611042S, RPK611043S, RPK611218S, RPK611447S, RPK611448S, RPK611449S, RPK611450S, RPK611922S, RPK611924S, RPK611933S, RPK609134S, RPK609381S, RPK609382S, RPK609821S, RPK609824S, RPK610187S, RPK610765S, RPK611526S, RPK609375S, RPK609464S, RPK609809S, RPK610195S, RPK610231S, RPK610601S, RPK610613S, RPK611016S, RPK611089S, RPK612071S, RPK609705S, RPK609914S, RPK610035S, RPK610036S, RPK610323S, RPK610360S, RPK610374S, RPK610376S, RPK610388S, RPK610389S, RPK610391S, RPK610392S, RPK610393S, RPK610394S, RPK610696S, RPK610925S, RPK610935S, RPK611143S, RPK611170S, RPK611175S, RPK611176S, RPK611179S, RPK611186S, RPK611195S, RPK611197S, RPK611198S, RPK611205S, RPK611211S, RPK611464S, RPK611467S, RPK611528S, RPK612144S, RPK612225S, RPK612227S, RPK612260S, RPK609508S, RPK609663S, RPK609692S, RPK609807S, RPK609808S, RPK610324S, RPK610397S, RPK610399S, RPK610403S, RPK610476S, RPK610491S, RPK610608S, RPK610615S, RPK610619S, RPK611083S, RPK611085S, RPK611135S, RPK611509S, RPK611517S, RPK611521S, RPK612147S, RPK612249S, RPK609233S, RPK609701S, RPK609769S, RPK609829S, RPK610400S, RPK610506S, RPK610593S, RPK610611S, RPK611157S, RPK611160S, RPK611505S, RPK611567S, RPK609210S, RPK609216S, RPK609407S, RPK609524S, RPK609551S, RPK609564S, RPK609599S, RPK609738S, RPK609788S, RPK609823S, RPK609960S, RPK610190S, RPK610385S, RPK610589S, RPK610599S, RPK610600S, RPK610614S, RPK610623S, RPK610625S, RPK611033S, RPK611050S, RPK611084S, RPK611129S, RPK611145S, RPK611151S, RPK611172S, RPK611605S, RPK611607S, RPK610398S, RPK610622S, RPK611077S, RPK611557S, RPK612272S, RPK610387S, RPK609201S, RPK610922S, RPK611111S, RPK611114S, RPK611128S, RPK611155S, RPK611482S, RPK611487S, RPK609390S, RPK609397S, RPK609611S, RPK609866S, RPK610031S, RPK610032S, RPK610218S, RPK610244S, RPK610602S, RPK610612S, RPK609352S, RPK609651S, RPK610502S, RPK610590S, RPK611121S, RPK611138S, RPK611162S, RPK611210S, RPK611470S, RPK611606S, RPK612178S, RPK611090S, RPK611169S, RPK610317S, RPK610618S, RPK611094S, RPK611472S, RPK611571S, RPK609810S, RPK610464S, RPK610422S, RPK610542S, RPK611203S, RPK611475S, RPK611485S, RPK612119S, RPK612137S, RPK611538S, RPK610033S, RPK610034S, RPK610039S, RPK610046S, RPK611452S, RPK609125S, RPK610471S, RPK609361S, RPK609362S, RPK609383S, RPK609741S, RPK611199S, RPK612078S, RPK609332S, RPK609386S, RPK610396S, RPK610458S, RPK609132S, RPK609635S, RPK609659S, RPK609870S, RPK609871S, RPK609941S, RPK610395S, RPK610404S, RPK610591S, RPK610605S, RPK610617S, RPK610620S, RPK610991S, RPK611007S, RPK611204S, RPK611455S, RPK609760S, RPK610569S, RPK612286S, RPK609245S, RPK610152S, RPK611052S, RPK611081S, RPK611217S, RPK611516S, RPK612065S, RPK612740S, RPK609147S, RPK609345S, RPK611038S, RPK611510S, RPK609268S, RPK609376S, RPK609454S, RPK609802S, RPK609924S, RPK610265S, RPK611130S, RPK610148S, RPK610504S, RPK611468S, RPK611614S, RPK609243S, RPK609297S, RPK610321S, RPK609270S, RPK610624S, RPK609825S, RPK609828S, RPK610607S, RPK610825S, RPK611092S, RPK611141S, RPK611156S, RPK611647S, RPK609553S, RPK609445S, RPK611181S, RPK611193S, RPK610209S, RPK611228S, RPK609110S, RPK609410S, RPK609826S, RPK610126S, RPK610176S, RPK610878S, RPK609312S, RPK609349S, RPK610854S, RPK611675S, RPK611688S, RPK609356S, RPK612017S, RPK609570S, RPK610708S, RPK610778S, RPK610727S, RPK609104S, RPK609274S, RPK609535S, RPK612311S, RPK612312S, RPK609103S, RPK609850S, RPK611955S, RPK609156S, RPK609629S, RPK609686S, RPK609419S, RPK609423S, RPK609523S, RPK609527S, RPK609637S, RPK609640S, RPK612561S, RPK612604S, RPK612605S, RPK612606S, RPK612607S, RPK612608S, RPK612609S, RPK609531S, RPK609642S, RPK611032S, RPK609184S, RPK609456S, RPK609458S, RPK611823S, RPK609470S, RPK609501S, RPK609572S, RPK611034S, RPK611270S, RPK611271S, RPK610927S, RPK611074S, RPK611817S, RPK609232S, RPK609334S, RPK609441S, RPK609479S, RPK609547S, RPK609552S, RPK609565S, RPK609567S, RPK610160S, RPK610161S, RPK610792S, RPK609623S, RPK611435S, RPK609771S, RPK611214S, RPK609762S, RPK609513S, RPK610158S, RPK610170S, RPK609735S, RPK611875S, RPK611603S, RPK611604S, RPK609881S, RPK610887S, RPK611102S, RPK610221S, RPK610884S, RPK609586S, RPK609840S, RPK610886S, RPK609743S, RPK609688S, RPK609401S, RPK609277S, RPK609784S, RPK609883S, RPK610214S, RPK609742S, RPK611593S, RPK609469S, RPK609486S, RPK610165S, RPK609699S, RPK609491S, RPK609494S, RPK611209S, RPK611801S, RPK611822S, RPK609340S, RPK609367S, RPK609506S, RPK609589S, RPK610787S, RPK609904S, RPK609275S, RPK609411S, RPK610291S, RPK610989S, RPK609259S, RPK609675S, RPK609676S, RPK610990S, RPK611646S, RPK609261S, RPK609679S, RPK609922S, RPK609923S, RPK610288S, RPK610074S, RPK610292S, RPK610299S, RPK609426S, RPK609532S, RPK609763S, RPK610286S, RPK609836S, RPK612019S, RPK609366S, RPK609230S, RPK611596S, RPK609335S, RPK609336S, RPK609463S, RPK609471S, RPK609515S, RPK609517S, RPK609571S, RPK610907S, RPK611201S, RPK611202S, RPK611206S, RPK611427S, RPK611577S, RPK611610S, RPK611798S, RPK609467S, RPK609472S, RPK609475S, RPK609476S, RPK609498S, RPK609509S, RPK609510S, RPK609516S, RPK609587S, RPK610159S, RPK610799S, RPK610985S, RPK610986S, RPK611365S, RPK611609S, RPK611787S, RPK609227S, RPK611212S, RPK609588S, RPK609617S, RPK610904S, RPK611036S, RPK609204S, RPK609228S, RPK609473S, RPK609521S, RPK609566S, RPK610154S, RPK609478S, RPK609518S, RPK609733S, RPK611245S, RPK610105S, RPK609695S, RPK611368S, RPK609154S, RPK609746S, RPK609855S, RPK609859S, RPK610757S, RPK611419S, RPK609142S, RPK609149S, RPK609877S, RPK610145S, RPK611903S, RPK609139S, RPK609148S, RPK610147S, RPK610350S, RPK609131S, RPK609373S, RPK609753S, RPK609853S, RPK609861S, RPK610018S, RPK610329S, RPK610345S, RPK610347S, RPK610505S, RPK610552S, RPK610752S, RPK610753S, RPK610754S, RPK611894S, RPK611899S, RPK610755S, RPK609851S, RPK610022S, RPK609151S, RPK609434S, RPK610219S, RPK610761S, RPK611590S, RPK609757S, RPK609857S, RPK611361S, RPK612007S, RPK609315S, RPK609891S, RPK609900S, RPK609901S, RPK609121S, RPK609124S, RPK609499S, RPK609503S, RPK609507S, RPK610007S, RPK610008S, RPK610009S, RPK610010S, RPK610011S, RPK610012S, RPK610013S, RPK610017S, RPK610297S, RPK610298S, RPK610303S, RPK610306S, RPK610307S, RPK610936S, RPK610937S, RPK610938S, RPK610939S, RPK610940S, RPK610961S, RPK610962S, RPK610964S, RPK611369S, RPK611370S, RPK611371S, RPK611412S, RPK611413S, RPK611930S, RPK611932S, RPK609717S, RPK609723S, RPK609864S, RPK610361S, RPK611921S, RPK610357S, RPK610358S, RPK610673S, RPK610675S, RPK612324S, RPK612325S, RPK612326S, RPK612327S, RPK612328S, RPK612329S, RPK612336S, RPK612337S, RPK611813S, RPK611814S, RPK609299S, RPK609301S, RPK609305S, RPK609457S, RPK609466S, RPK609468S, RPK609682S, RPK609683S, RPK609684S, RPK609712S, RPK609713S, RPK609714S, RPK609715S, RPK609716S, RPK610918S, RPK610919S, RPK610920S, RPK610921S, RPK610923S, RPK611415S, RPK611416S, RPK611417S, RPK611425S, RPK611426S, RPK611451S, RPK612331S, RPK612332S, RPK612333S, RPK612334S, RPK609573S, RPK609582S, RPK609593S, RPK609596S, RPK609597S, RPK609605S, RPK609616S, RPK609618S, RPK609619S, RPK609739S, RPK609740S, RPK609931S, RPK609932S, RPK609935S, RPK609939S, RPK609940S, RPK610052S, RPK610053S, RPK610054S, RPK609192S, RPK609198S, RPK609202S, RPK609339S, RPK609342S, RPK609385S, RPK609387S, RPK609388S, RPK609389S, RPK609391S, RPK609673S, RPK610207S, RPK610211S, RPK610255S, RPK610256S, RPK610257S, RPK610290S, RPK610294S, RPK610296S, RPK610366S, RPK610367S, RPK610561S, RPK610563S, RPK610564S, RPK610565S, RPK610570S, RPK611414S, RPK611421S, RPK611422S, RPK611423S, RPK611844S, RPK611845S, RPK611846S, RPK611847S, RPK611849S, RPK611850S, RPK611851S, RPK611904S, RPK611935S, RPK611936S, RPK611937S, RPK611938S, RPK611940S, RPK611965S, RPK611966S, RPK611087S, RPK611088S, RPK611091S, RPK611093S, RPK611530S, RPK611531S, RPK609380S, RPK611587S, RPK611790S, RPK611887S, RPK611888S, RPK611911S, RPK611950S, RPK609302S, RPK609427S, RPK609697S, RPK611185S, RPK611328S, RPK611729S, RPK612038S, RPK612649S, RPK610632S, RPK610217S, RPK609130S, RPK609140S, RPK609141S, RPK610348S, RPK610344S, RPK611793S, RPK611617S, RPK611792S, RPK611902S, RPK609146S, RPK610343S, RPK609153S, RPK609159S, RPK609160S, RPK609161S, RPK609162S, RPK609163S, RPK609164S, RPK609168S, RPK609170S, RPK609175S, RPK609176S, RPK609177S, RPK609178S, RPK609179S, RPK609181S, RPK609182S, RPK609183S, RPK609185S, RPK609186S, RPK609187S, RPK609307S, RPK609477S, RPK609480S, RPK609481S, RPK609482S, RPK609483S, RPK609488S, RPK609620S, RPK609621S, RPK609644S, RPK609645S, RPK609646S, RPK609647S, RPK609648S, RPK609653S, RPK609654S, RPK609655S, RPK609656S, RPK609660S, RPK609665S, RPK609666S, RPK609667S, RPK609668S, RPK609669S, RPK609671S, RPK609672S, RPK609925S, RPK609929S, RPK610261S, RPK610262S, RPK610263S, RPK610268S, RPK610335S, RPK610354S, RPK610431S, RPK610433S, RPK610534S, RPK610536S, RPK610537S, RPK610540S, RPK610980S, RPK610981S, RPK610982S, RPK610984S, RPK610274S, RPK609371S, RPK610551S, RPK609319S, RPK609736S, RPK609783S, RPK611122S, RPK609422S, RPK610451S, RPK611866S, RPK612634S, RPK612635S, RPK609293S, RPK610220S, RPK610499S, RPK610756S, RPK610763S, RPK611585S, RPK611915S, RPK611916S, RPK610026S, RPK609799S, RPK609800S, RPK609872S, RPK609887S, RPK610140S, RPK610141S, RPK610571S, RPK611439S, RPK611440S, RPK611441S, RPK611442S, RPK611443S, RPK611841S, RPK611858S, RPK611943S, RPK611944S, RPK610349S, RPK610557S, RPK611611S, RPK611612S, RPK611956S, RPK611957S, RPK611977S, RPK610172S, RPK610638S, RPK610272S, RPK610381S, RPK610383S, RPK610401S, RPK610707S, RPK611248S, RPK611249S, RPK611250S, RPK611252S, RPK611253S, RPK611254S, RPK611255S, RPK611256S, RPK611257S, RPK611258S, RPK611259S, RPK611260S, RPK611261S, RPK611287S, RPK611979S, RPK609113S, RPK609220S, RPK609224S, RPK609234S, RPK609555S, RPK609558S, RPK610062S, RPK610083S, RPK610087S, RPK610093S, RPK610210S, RPK610640S, RPK610659S, RPK610678S, RPK610688S, RPK610730S, RPK610824S, RPK611000S, RPK611005S, RPK611073S, RPK611192S, RPK611224S, RPK611230S, RPK611231S, RPK611740S, RPK611741S, RPK612070S, RPK609272S, RPK610866S, RPK611313S, RPK611659S, RPK611763S, RPK611782S, RPK612015S, RPK612348S, RPK609236S, RPK609752S, RPK610741S, RPK609137S, RPK609862S, RPK610027S, RPK610311S, RPK610314S, RPK610558S, RPK611215S, RPK610222S, RPK609424S, RPK610049S, RPK610351S, RPK610547S, RPK609152S, RPK609206S, RPK609226S, RPK611829S, RPK609944S, RPK609947S, RPK609950S, RPK609951S, RPK609952S, RPK609954S, RPK609957S, RPK609959S, RPK610705S, RPK611983S, RPK611556S, RPK610382S, RPK610432S, RPK609814S, RPK609143S, RPK609249S, RPK609711S, RPK609782S, RPK611284S, RPK609267S, RPK610427S, RPK609157S, RPK610785S, RPK611877S, RPK611878S, RPK612630S, RPK609526S, RPK609636S, RPK611952S, RPK611869S, RPK610216S, RPK609556S, RPK610598S, RPK609331S, RPK611618S, RPK611876S, RPK611220S, RPK611953S, RPK611868S, RPK612346S, RPK609313S, RPK609489S, RPK611037S, RPK611216S, RPK610941S, RPK610945S, RPK610583S, RPK610584S, RPK610956S, RPK610957S, RPK610954S, RPK610948S, RPK610958S, RPK611372S, RPK611373S, RPK611380S, RPK611381S, RPK610556S, RPK611588S, RPK611589S, RPK611912S, RPK611978S, RPK611980S, RPK611075S, RPK611180S, RPK609544S, RPK609693S, RPK610548S, RPK612610S, RPK609638S, RPK609727S, RPK610762S, RPK611615S, RPK609885S, RPK611359S, RPK611362S, RPK611418S, RPK611575S, RPK611795S, RPK611796S, RPK611917S, RPK611918S, RPK611920S, RPK609875S, RPK610566S, RPK611934S, RPK610466S, RPK610581S, RPK610967S, RPK610968S, RPK610969S, RPK610970S, RPK610976S, RPK610460S, RPK610462S, RPK610479S, RPK610482S, RPK610485S, RPK610488S, RPK610489S, RPK610490S, RPK610576S, RPK610578S, RPK610580S, RPK610582S, RPK610965S, RPK610966S, RPK610971S, RPK610972S, RPK610973S, RPK610974S, RPK610975S, RPK610924S, RPK610926S, RPK610928S, RPK611861S, RPK610929S, RPK610930S, RPK611927S, RPK611928S, RPK612341S, RPK609403S, RPK611071S, RPK609215S, RPK609219S, RPK609455S, RPK609806S, RPK609902S, RPK609906S, RPK609997S, RPK610250S, RPK611808S, RPK611809S, RPK611926S, RPK609263S, RPK609264S, RPK609266S, RPK609269S, RPK609271S, RPK609281S, RPK609282S, RPK609283S, RPK609284S, RPK609285S, RPK609490S, RPK609512S, RPK609987S, RPK609988S, RPK610213S, RPK610226S, RPK610330S, RPK610336S, RPK610567S, RPK611363S, RPK611364S, RPK611446S, RPK611859S, RPK611942S, RPK611967S, RPK611968S, RPK611969S, RPK611970S, RPK611971S, RPK611972S, RPK611973S, RPK611974S, RPK611975S, RPK611976S, RPK612319S, RPK612320S, RPK612321S, RPK612322S, RPK612323S, RPK612338S, RPK610913S, RPK610242S, RPK610287S, RPK610289S, RPK610362S, RPK610364S, RPK610671S, RPK611424S, RPK611816S, RPK611842S, RPK612317S, RPK612340S, RPK609144S, RPK609612S, RPK609580S, RPK609819S, RPK610189S, RPK610883S, RPK609408S, RPK609608S, RPK609846S, RPK609604S, RPK609830S, RPK610902S, RPK609136S, RPK609107S, RPK609965S, RPK609967S, RPK610879S, RPK609108S, RPK609135S, RPK610169S, RPK611366S, RPK611562S, RPK611563S, RPK611564S, RPK611565S, RPK611566S, RPK612219S, RPK612222S, RPK612226S, RPK612247S, RPK612597S, RPK612598S, RPK611572S, RPK609409S, RPK609725S, RPK609344S, RPK610672S, RPK611642S, RPK609878S, RPK609882S, RPK610318S, RPK611031S, RPK609974S, RPK610764S, RPK609208S, RPK609237S, RPK609239S, RPK609329S, RPK609359S, RPK609520S, RPK611189S, RPK612553S, RPK612554S, RPK609628S, RPK609291S, RPK609322S, RPK610473S, RPK610603S, RPK610604S, RPK610616S, RPK610875S, RPK610983S, RPK611103S, RPK611107S, RPK611183S, RPK611569S, RPK609155S, RPK609972S, RPK609123S, RPK609919S, RPK609921S, RPK610117S, RPK610118S, RPK610193S, RPK610194S, RPK610538S, RPK610568S, RPK610572S, RPK611444S, RPK611445S, RPK611905S, RPK611925S, RPK611941S, RPK612316S, RPK612330S, RPK612335S, RPK611570S, RPK611815S, RPK611853S, RPK611854S, RPK611855S, RPK611856S, RPK611880S, RPK609890S, RPK609730S, RPK609912S, RPK610153S, RPK612363S, RPK611981S, RPK609169S, RPK609309S, RPK609323S, RPK609324S, RPK609333S, RPK609493S, RPK609522S, RPK609649S, RPK609680S, RPK609681S, RPK609801S, RPK609803S, RPK609805S, RPK609916S, RPK609917S, RPK609918S, RPK609998S, RPK609999S, RPK610002S, RPK610004S, RPK610006S, RPK610142S, RPK610143S, RPK610144S, RPK610201S, RPK610204S, RPK610205S, RPK610252S, RPK610260S, RPK610282S, RPK610285S, RPK611803S, RPK611804S, RPK611805S, RPK611806S, RPK611807S, RPK611862S, RPK611863S, RPK611864S, RPK611870S, RPK611886S, RPK611949S, RPK611958S, RPK611960S, RPK611961S, RPK611962S, RPK611963S, RPK611848S, RPK609414S, RPK609958S, RPK610072S, RPK611353S, RPK611910S, RPK611794S, RPK612043S, RPK610171S, RPK612578S, RPK612571S, RPK612564S, RPK609485S, RPK609718S, RPK610963S, RPK611931S, RPK610933S, RPK610934S, RPK611411S, RPK609652S, RPK609657S, RPK610277S, RPK610279S, RPK610280S, RPK609737S, RPK610119S, RPK610192S, RPK610197S, RPK610198S, RPK610227S, RPK610233S, RPK610234S, RPK610243S, RPK610247S, RPK610248S, RPK610249S, RPK610251S, RPK610253S, RPK610259S, RPK610270S, RPK610278S, RPK610319S, RPK610370S, RPK610440S, RPK610444S, RPK610487S, RPK610498S, RPK610512S, RPK610517S, RPK610524S, RPK610550S, RPK610587S, RPK610594S, RPK610596S, RPK610609S, RPK610621S, RPK610687S, RPK610702S, RPK610704S, RPK610720S, RPK610745S, RPK610766S, RPK610767S, RPK610796S, RPK610801S, RPK610816S, RPK610833S, RPK610872S, RPK610881S, RPK610916S, RPK610992S, RPK611025S, RPK611028S, RPK611029S, RPK611035S, RPK611046S, RPK611072S, RPK611079S, RPK611104S, RPK611117S, RPK611123S, RPK611125S, RPK611137S, RPK611140S, RPK611149S, RPK611150S, RPK611153S, RPK611154S, RPK611168S, RPK611171S, RPK611207S, RPK611208S, RPK611213S, RPK611222S, RPK611223S, RPK611274S, RPK611281S, RPK611298S, RPK611339S, RPK611346S, RPK611347S, RPK611350S, RPK611430S, RPK611431S, RPK611433S, RPK611437S, RPK611438S, RPK611454S, RPK611458S, RPK611465S, RPK611478S, RPK611490S, RPK611519S, RPK611524S, RPK611527S, RPK611532S, RPK611533S, RPK611543S, RPK611545S, RPK611550S, RPK611555S, RPK611558S, RPK611559S, RPK611560S, RPK611561S, RPK611568S, RPK611579S, RPK611586S, RPK611598S, RPK611599S, RPK611602S, RPK611608S, RPK611613S, RPK611616S, RPK611628S, RPK611633S, RPK611634S, RPK611635S, RPK611636S, RPK611668S, RPK611678S, RPK611693S, RPK611699S, RPK611703S, RPK611719S, RPK611744S, RPK611751S, RPK611752S, RPK611764S, RPK611766S, RPK611773S, RPK611778S, RPK611779S, RPK611789S, RPK611799S, RPK611824S, RPK611825S, RPK611826S, RPK611828S, RPK611830S, RPK611831S, RPK611834S, RPK611835S, RPK611836S, RPK611838S, RPK611839S, RPK611843S, RPK611860S, RPK611882S, RPK611883S, RPK611884S, RPK611885S, RPK611889S, RPK611890S, RPK611895S, RPK611901S, RPK611907S, RPK611914S, RPK611948S, RPK611951S, RPK611964S, RPK611984S, RPK611985S, RPK611987S, RPK611998S, RPK612002S, RPK612008S, RPK612011S, RPK612058S, RPK612067S, RPK612094S, RPK612096S, RPK612097S, RPK612099S, RPK612125S, RPK612139S, RPK612158S, RPK612159S, RPK612166S, RPK612168S, RPK612169S, RPK612181S, RPK612182S, RPK612183S, RPK612184S, RPK612212S, RPK612234S, RPK612237S, RPK612239S, RPK612240S, RPK612248S, RPK612250S, RPK612255S, RPK612257S, RPK612258S, RPK612266S, RPK612268S, RPK612289S, RPK612318S, RPK612339S, RPK612351S, RPK612354S, RPK612355S, RPK612357S, RPK612358S, RPK612360S, RPK612366S, RPK612367S, RPK612368S, RPK612369S, RPK612370S, RPK612426S, RPK612430S, RPK612432S, RPK612435S, RPK612437S, RPK612443S, RPK612447S, RPK612448S, RPK612450S, RPK612451S, RPK612456S, RPK612457S, RPK612458S, RPK612459S, RPK612460S, RPK612461S, RPK612462S, RPK612463S, RPK612465S, RPK612466S, RPK612467S, RPK612468S, RPK612469S, RPK612470S, RPK612471S, RPK612472S, RPK612473S, RPK612474S, RPK612478S, RPK612489S, RPK612499S, RPK612500S, RPK612501S, RPK612510S, RPK612520S, RPK612521S, RPK612522S, RPK612524S, RPK612528S, RPK612530S, RPK612534S, RPK612535S, RPK612537S, RPK612538S, RPK612540S, RPK612541S, RPK612547S, RPK612548S, RPK612550S, RPK612552S, RPK612555S, RPK612556S, RPK612559S, RPK612563S, RPK612565S, RPK612566S, RPK612567S, RPK612568S, RPK612569S, RPK612570S, RPK612572S, RPK612574S, RPK612582S, RPK612583S, RPK612584S, RPK612586S, RPK612590S, RPK612591S, RPK612592S, RPK612603S, RPK612615S, RPK612620S, RPK612628S, RPK612639S, RPK612643S, RPK612653S, RPK612654S, RPK612662S, RPK612670S, RPK612676S, RPK612693S, RPK612705S, RPK612711S, RPK612717S, RPK612721S, RPK612727S, RPK612729S, RPK612733S, RPK612735S, RPK612736S, RPK612737S, RPK612738S, RPK612739S, RPK612742S, RPK612743S, RPK612746S, RPK612758S, RPK612759S, RPK612761S, RPK612764S, RPK612765S, RPK611696S, RPK611706S, RPK612315S, RPK612549S, RPK612713S, RPK612715S, RPK612744S, RPK612745S, RPK612747S, RPK612748S, RPK612749S, RPK612750S, RPK610258S, RPK610626S, RPK610867S, RPK610993S, RPK611040S, RPK611113S, RPK611174S, RPK611188S, RPK611219S, RPK611286S, RPK611338S, RPK611410S, RPK611484S, RPK611640S, RPK611653S, RPK611702S, RPK611784S, RPK611797S, RPK611879S, RPK611913S, RPK611919S, RPK612023S, RPK612042S, RPK612356S, RPK612422S, RPK612476S, RPK612516S, RPK612551S, RPK612613S, RPK612629S, RPK612636S, RPK612677S, RPK612722S, RPK611906S, RPK610356S, RPK610355S, RPK610246S, RPK610240S, RPK610237S, RPK610196S, RPK610139S, RPK610121S, RPK610120S, RPK609993S, RPK609992S, RPK609989S, RPK609910S, RPK609497S, RPK609495S, RPK609200S, RPK609193S, RPK609158S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce8575 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = NIK
PMAs with Product Code = NIK
-
-