U.S. Department of Health and Human Services

Class 2 Device Recall QUADROXi Adult

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall QUADROXi Adultsee related information
Date Initiated by FirmMay 18, 2023
Date PostedMay 26, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1638-2023
Recall Event ID 92003
510(K)NumberK082117 
Product Classification Oxygenator, cardiopulmonary bypass - Product Code DTZ
ProductQUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item no. HMO 70000-USA 701067820 HMO 71000-USA 701067823
Code Information Model Item no. UDI-DI HMO 70000-USA 701067820 4058863019147 HMO 71000-USA 701067823 4058863017341 All lots
FEI Number 3008355164
Recalling Firm/
Manufacturer
Maquet Medical Systems USA
45 Barbour Pond Dr
Wayne NJ 07470-2094
For Additional Information ContactMaryanna Krivak
973-709-7483
Manufacturer Reason
for Recall
1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.
FDA Determined
Cause 2
Pending
ActionOn May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.
Quantity in Commerce0 (US)
DistributionDomestic distribution nationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = DTZ
-
-