 Class 2 Device Recall Aptio Automationsee related information
Date Initiated by FirmMay 16, 2023
Date PostedJuly 05, 2023
Recall Status1 Open3, Classified
Recall NumberZ-2091-2023
Recall Event ID 92461
510(K)NumberK043546 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductAptio Automation Storage and Retrieval Module (SRM)
Code Information Model No. 10713762 and 10715227; UDI-DI: 00630414596754; All units with software versions prior to 38.
FEI Number 2517506
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
500 Gbc Dr
Newark DE 19702-2466
For Additional Information ContactJeff Semone
800-441-9250
Manufacturer Reason
for Recall
A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm sent a recall letter dated 05/16/2026 to consignees notifying them of the firmware issue presenting with this recall event. Consignees are asked to conduct visual checks of the device's gates at the buffer lane entry daily to ensure there are no obstructions that prevent the correct activation and movement of the divert, the divert looks intact, and the position of the divert is aligned to the profiles when it is diverting a tube without any missing or partial blocking of the carrier. If the visual check does not pass, contact your local technical support for assistance before using the impacted system. Until a service visit to upgrade the firmware of affected devices can be conducted, consignees are asked to maintain awareness of the provided recall notice and conduct daily recommended actions. Consignees are asked to complete and return the provided response form within 7 days of recall notification. Completed forms are to be emailed to escctsfcaecfax.team@siemens-healthineers.com or faxed to (312)274-7795. Customers with any questions are asked to contact their local Siemens Healthineers Customer Care Center or technical support representative.
Quantity in Commerce83 units
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IL, KY, LA, MA, MD, MI, MS, MT, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, & WV. The countries of Australia, Brazil, Canada, China, France, Germany, Greece, Italy, Portugal, Saudi Arabia, Spain, Sweden, & United Kingdom.
Total Product Life CycleTPLC Device Report