| Date Initiated by Firm | July 12, 2023 |
| Date Posted | September 12, 2023 |
| Recall Status1 |
Completed |
| Recall Number | Z-2550-2023 |
| Recall Event ID |
92647 |
| 510(K)Number | K192196 |
| Product Classification |
Accelerator, linear, medical - Product Code IYE
|
| Product | Calibrate (Cs-137) screened nanoDot D2DNS, Model Number 03065-OTO; radiation monitoring dosimeter used with the microSTAR readers |
| Code Information |
UDI/DI 0860003399903, all batch numbers |
| FEI Number |
3008426232
|
Recalling Firm/ Manufacturer |
Landauer
2 Science Rd
Glenwood
IL
60425-1531
|
| For Additional Information Contact | Brian Malone 815-690-1231 |
Manufacturer Reason for Recall | LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Landauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following:
Discontinue use of the affected products, destroy used products, and return unused products. |
| Quantity in Commerce | 0 |
| Distribution | Worldwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
| 510(K) Database | 510(K)s with Product Code = IYE
|