| Class 2 Device Recall Gallant/Neutrino/Entrant CRTD |  |
Date Initiated by Firm | August 16, 2023 |
Date Posted | September 19, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-2615-2023 |
Recall Event ID |
92923 |
PMA Number | P030054S374 |
Product Classification |
Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
|
Product | Gallant HF Cardiac Resynchronization Therapy Defibrillator:
HF CRT-D REF CDHFA500Q
Neutrino HF Cardiac Resynchronization Therapy Defibrillator:
NxT HF CRT-D REF CDHFA600Q
Entrant HF Cardiac Resynchronization Therapy Defibrillator :
HF CRT-D REF CDHFA300Q
The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV.
Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique. |
Code Information |
Gallant:
Model Number: HF CRT-D REF CDHFA500Q
UDI-DI Code: 05415067032010
Serial Numbers:
111023714
111033397
111026077
111008528
810007304
810023031
111009077
810007164
111036262
111026540
111013180
810020124
111040419
111042048
810022176
111029687
111006406
111017718
810008795
111012956
111016597
810022867
810018823
810006029
810027443
810000072
810016983
111011034
111015410
810006013
111009628
111018199
111030383
810026178
111037835
111038995
111039205
111039169
111039213
111039103
111039227
111039172
111039154
111039298
111038919
111040351
111037720
111037178
111037142
111005031
810019888
810007234
111028528
111001728
111001715
111000681
111000233
111000202
111000235
111001707
111001777
111000294
111015428
111000224
111000494
111016493
111015792
111001752
111001687
111021218
111021101
111009661
111010339
111010612
111010599
111010615
111036821
111043197
111037844
111014472
111005935
111014496
111006045
111006001
111016094
810027416
810000369
810003198
111028551
111028566
111028556
810019969
810011841
810011914
111042858
111008829
810013774
810018766
810017010
810015486
810009746
810005988
810006054
810023114
111012399
810009476
810011649
111011036
810027614
810012107
810024932
111021623
111012101
111003756
810020343
810003891
111009278
810023576
810013854
111015681
111003733
810009559
810014126
111010585
810024290
810022925
111008862
111012548
810016189
810017522
810009554
111015794
810019965
810011865
111040526
810009812
810012821
111015675
111011312
111011387
111011336
111011393
810026670
810026649
810026529
810010664
810011672
810011882
810005592
810006102
111015355
111019534
111010215
111011884
810005106
810020548
810027286
810009557
810011904
810020259
111022242
810009288
810008353
111022498
810010454
810010579
810006036
111009330
111010065
111015392
111011278
111009334
111010584
111009290
111021177
810003958
810012930
810003393
111002912
810011728
810013669
111015718
111003914
111003895
111022218
111022176
111021632
810004142
810003754
810003969
810020452
810004923
810023056
111019916
111016496
111003747
111009342
810003352
810022597
810017501
810005626
111005857
111011043
810012796
810026382
810004289
111017256
810009843
810004185
111012096
810026615
111011508
810024005
810024143
111010413
810009501
810003379
810026380
810011567
111004463
111004347
111020270
111010053
810023220
111015771
111012508
111012533
111011321
111011664
111011671
111011660
111011277
111012141
111010395
111010208
810022788
810022914
810022951
810023117
810020559
111001720
101000009
810018735
810018810
810015704
810016881
111016614
111003711
810024010
810025112
810005461
810005513
810003954
810003896
810003890
810003783
810003322
810024961
810024302
810026390
810026353
810026405
111040521
810026631
810008266
810026620
810027604
810010652
810014146
810009840
111003720
111008882
111005848
111005862
111003911
111008897
111005795
111010086
111003585
111017283
111017282
111003203
111011860
810004489
111012566
810003631
810024154
810008374
810010504
810013660
810013900
810020172
810016892
111008830
111009273
810022926
810015357
111042850
810027289
810006366
810005983
810006005
111011058
111009689
111012563
111011850
111015664
810011902
810009831
810016974
810022616
111002245
111009306
810006040
810026359
810024115
111041429
111004378
810004095
810013886
111009651
111014610
810003397
810005670
810009631
810003965
111010998
810026366
810015426
111009681
111015007
111015773
810009833
810010528
111008924
111008908
111008922
810023083
810025119
810024292
810009810
810015484
810012757
810012772
810013666
810016936
810022887
810020536
810020382
810009765
810009946
810009253
810009264
810005589
810003389
810003473
810004275
810004101
810004517
810004580
111009626
111009672
111010357
111003814
810004378
810006381
111022487
111009331
111009319
111011051
810004481
810004446
810008382
810020534
111011879
810009243
111011677
111011656
111011098
111009632
810015724
111015372
810013610
111012559
810006009
810006167
810025111
810020507
111022258
111020916
810016004
810024269
810003395
810006038
810006071
810010542
810022922
810009286
111021727
111040535
111022254
810016993
810015492
810020445
810020542
810013764
111021735
111041442
810025097
810026387
810023997
111041420
810020148
810009509
810009276
810010443
111001769
810014091
810014172
810013798
810004671
810017003
810016731
111017271
810016597
810015592
810020752
810018834
810018837
111022248
810012770
810003686
111019514
810004181
810005530
111011498
810013853
810004677
810006014
111010252
111020934
810006032
810013995
111005806
810009234
111002771
810006411
111004023
111004032
111012350
810005141
111042833
111004014
810013668
810011898
810015504
111011303
111011888
810005602
111004022
810008369
810003398
810010433
810010545
810015360
810010592
810010484
810015437
810009285
810004521
111012556
810010511
810010497
810009792
810015487
810015274
111020259
111010343
810010512
810013815
810012815
111002318
111020262
111022734
111020168
810009652
810023573
810022774
810023599
810013997
810022602
111004020
111017258
810004905
810027634
111002886
810004268
810009235
111012108
111010344
810013987
810005937
111010349
810015584
810023597
810022896
810026419
810015997
111017275
810016519
111041444
810015230
810012784
810018571
810016587
810005507
810020168
111009670
810005544
810004290
111004352
810016870
111001914
111001801
111042837
810011754
810016950
111010578
810023077
111004346
111022205
810016507
111000873
810009912
111012553
111010398
810027300
810026355
111004339
810004432
111000857
111012388
111010245
810012783
111017274
111004159
111010415
810016515
810013885
111003876
111009326
111008904
111001781
111004019
810010661
810011678
111021420
111011093
810005651
810013580
111021508
111012562
810012633
810005916
111012519
810009271
810023230
810026376
810009654
810014148
810020512
810024003
810009270
810009475
810016876
111012347
810026400
111011083
111012542
810018764
810024264
810004769
111021629
810003476
111011680
111015012
810013904
810006153
810018575
111015374
111012136
810024140
111010219
810009572
111012424
810004367
810003630
810010647
111012543
810013998
810027603
810014007
810010665
111001939
111002319
111012111
111021080
111020179
111021214
810020376
111001937
111021387
111008852
810013742
111020918
111004024
810004980
111008846
111015656
111010050
810023067
810010470
810027309
111021607
810028868
810020496
810009610
111041433
111010055
810015713
111011469
111009280
111003717
111012356
810027283
810015609
810020237
810020346
111004124
111002228
111002114
810013924
810016190
810024939
111004373
810009236
810005612
810014149
111004010
810015581
111021523
111003989
111010598
111017287
111011067
111003983
810005625
111011053
810013969
810009275
810006417
111022193
111011885
810023079
810016501
111012401
111003749
111003754
111003607
111015008
111042709
111024158
810006363
111004138
111012421
111041464
111022177
810016298
810016297
810016984
810016929
810016891
810016910
810016960
810016717
810016716
810016735
111004338
111015763
111011661
111004357
111012554
111011287
111009294
111004345
111002252
111004366
111004358
111015719
111004106
111008569
111041425
810004479
111021408
111011672
810012776
810018569
810009257
810015365
810022557
810022918
810013937
810025118
810024946
810028884
810028879
111003810
111002230
111021184
810010531
810010568
810010422
810010449
111024152
810003685
810005640
810005919
111009695
810015427
810015226
810013903
111004354
111008892
111003801
111003907
111003997
111012129
111003305
111002567
111011077
111012135
111012547
111008831
111004123
810004585
111001607
111001908
111015390
111016480
810010541
111020338
111021201
111004107
810016905
111012152
111009301
810011748
111003978
810009545
810003892
111011673
111001428
111019547
810026404
111041430
111019484
810026360
111003701
810009850
810009259
810008721
810020162
111004001
810020751
810015519
111012123
810003683
810026619
810009565
111010075
111017264
810027260
810026618
111008832
111012430
111011475
111021529
810010475
810006138
810008349
810010462
111003987
111022157
810009617
810012803
810015324
810008727
810004366
111011890
810008351
810003414
111015871
810016988
810009829
810008728
810008720
810012842
810011628
111020271
810006051
810026417
810003323
111021081
810016010
810011832
810015236
111010645
810018825
810009290
810003528
810013618
810006196
810019968
810009240
810005648
111022161
810018740
111010212
810015238
111020375
810003405
810024931
810019976
111011392
111012391
810004893
111014616
810023055
810013952
810010437
810018798
111041451
111011313
810020418
810005150
810006008
810004516
810011653
111021514
810004572
810004704
111019496
810004974
810024160
810011694
111011591
111011866
810013920
111012511
810008262
810011778
111012383
810023120
810016908
111009325
810015474
810023567
810013794
810015613
810013917
810013840
810026604
810024000
810020537
111004114
810014014
810010546
810010602
810010595
810014094
810004983
810024153
810024163
111010241
111010239
810004667
810015711
810004793
111010356
810022965
810010524
810013986
810027269
810010657
810006392
810012800
810013651
111011646
111009669
810009256
810022576
810011763
111012544
111021421
810015348
810006030
810023234
111000589
810010459
810010645
111010632
111002477
810018802
111004343
810006171
111022229
810010577
810012691
111003994
111042704
810019993
810026426
810020394
810018750
810013600
810009608
810006380
810009460
111021211
111015375
810010650
810015439
810026648
810012824
810011879
111008874
810020554
810009926
810009763
810009795
810022622
810016730
810024953
810022599
810024963
810027173
810011854
810020406
810022894
810017543
810003526
810014134
810014032
810015639
810016295
810014138
111019480
810020503
111022220
810009932
810009914
810022775
810012799
111042847
810011582
111011090
810015233
810004518
810015566
810015567
810009842
111020267
111011652
810004557
111021734
111042707
810020266
810013671
111000631
810023587
810020268
111003743
111008898
111000343
810009558
810018839
810003692
810003562
111012147
111015391
810004924
111022160
810003563
810013934
111005814
111015662
111003593
111015365
111009309
810009446
810009468
810015630
810004363
810013814
111017272
810015602
810016184
810024108
111010070
810024273
111015660
111002891
810020747
810003761
810010565
810020541
810022799
111010625
810023590
111002867
111020335
810011711
810009811
810015229
111035639
111036824
111035636
111036808
111035659
810003971
111011276
810015342
810022613
810022598
810022603
111009298
111040541
810015332
111010372
111020279
810011798
111008855
111000186
111008833
810003524
810024280
111001182
111015367
810004507
810003400
111042848
111001742
810009265
810003930
111004458
810022903
810024329
111011492
810012814
810011585
810011618
810018751
111011071
810012759
111010380
810020272
810022606
810020545
111020182
810027315
810010619
810006134
111015043
111015677
810004439
810009549
810026625
810026617
810004584
810019963
810020743
810012695
810004552
810011846
111022153
810005618
810012831
810024956
111009662
111020329
111019557
810011920
111008879
810018713
111009315
111021506
111011384
810028893
111010377
810027641
810003319
810004447
810013633
810006166
810013604
810011750
111015042
111021084
810009620
810006050
810018592
810018603
810005986
810027172
810006060
810019997
810022947
810016956
810010482
810013588
810009920
810020250
810020230
111020265
810010506
111011052
810011859
810003320
810004513
810010451
111003220
111020334
810003306
810005614
810020160
111010634
810011588
111016492
111010090
810004183
810011581
810018595
810016729
810005667
810018805
810018762
810026601
810013770
111012407
810026603
810020178
810004476
810011612
810016289
810018743
111019536
111019526
111019504
111011283
810015234
111020189
810005470
810003894
810003887
111019550
810009636
810024157
810012817
111015359
111009674
111002880
111002758
111011343
111009305
810025105
111015784
810013916
810005616
810009291
810027648
810023245
111011466
810013673
810015575
810018737
810018795
810013662
810006353
810009786
111020318
810005154
810003407
810004926
111012429
111021086
810005590
810023115
111009693
111012131
111021515
111000392
111000393
111016486
111009623
111012166
111009654
810023063
111008576
810004474
810018815
810014101
111009694
111015776
810013962
810004858
810013561
810009598
111012393
111000332
810017498
111008857
111002866
111022237
810012908
810026427
810012829
810015367
810006078
810016715
810026413
111008573
111012520
111012420
111000614
111012387
111009676
810005955
111000478
810024966
810010495
810009619
111020252
111000918
810004523
810014121
810003318
810012836
111020249
810005003
810003312
810005109
111004462
111012153
111021400
111004442
111010261
111010227
111003755
810011845
810003682
810014174
111010213
810018581
810006129
810018579
810006170
810014005
111010202
810011722
111011681
810009292
810005885
111003713
810009267
810016992
810026409
810018767
810003392
111021194
810004035
810017537
111021419
810015327
810020557
810008248
810016702
810027329
810020436
810016982
810015628
810027288
111009688
111011073
810006058
810009582
111010069
810009759
810010485
810026367
810009770
111021418
810024282
111011669
111015788
111003980
111003608
111003716
810014104
810012962
810011719
810015362
810004526
810003899
810004931
810009632
810009575
810006165
111003204
111003903
111011032
111012510
111011041
810020419
111016498
810027609
111010048
810012826
810020725
810011734
111016503
810011909
810022794
810009470
810013784
111001358
111003731
111003700
111005797
810013851
810011848
810013842
111003730
810003691
111021104
810004702
111015405
810026392
810015509
810020506
810016736
810012632
810013796
810003477
810013725
111012549
111012518
810013936
810003567
111004009
810003522
111012534
810008232
810008250
111019524
810005647
810012760
111015758
111021725
810003316
111021729
810004090
111005813
810015505
810012692
810004676
111010218
111010061
810005887
111002775
111010408
111040288
810005532
111004115
810004436
111012382
111011683
111001938
111004117
810023585
810014108
111042703
111015874
810004547
810004982
810016510
111011849
810022617
810005511
111012104
111042702
810020338
111010622
810009441
810010445
810012792
810026661
810027647
111008896
810009524
810020352
810010458
810015718
810015583
810018564
810014033
810004207
810004672
810004190
810003968
111020261
111008899
111022173
810005999
810015330
810005101
111011324
111010573
810015359
810006039
810004396
111011145
810006096
810008722
810022792
810004586
111011873
111004122
810013622
111012558
810013783
810020241
810003750
111004363
111042708
111012397
111011016
111020357
111009337
810022574
810022605
810022578
810024150
810018709
810015572
810012795
810012762
810026659
111008850
111009633
111009340
111015394
111015692
111022251
111021117
810004435
111012410
810006004
111009646
111015398
810004271
810010498
810009751
810009607
111021503
111011081
810010436
810003325
111011000
810004511
111011275
810004144
111022155
810006351
810006017
810020455
111016506
810020922
810016964
111003993
810013865
810015300
111001789
810019975
810022891
810009568
111001763
111000488
111002936
810024102
810003469
111000915
810005918
810010629
810009456
810015349
810003785
810013654
810016294
810005600
111010604
810003751
810011894
810015351
810004219
810018555
810011630
111005861
810011651
810020159
810004143
810004081
111009321
810027555
810003353
810017517
810009451
810010439
810015515
810009902
810020255
111002109
111005860
810027626
111002903
810026348
111009679
111009666
810009521
810004930
810009523
810020444
810020493
111016513
111004367
111009289
111011394
111004084
111003982
111004086
111011403
111016514
111010246
810022769
111015872
810003487
810005047
810015594
810004373
810003386
111011887
810024001
810015356
810025122
810015571
810010529
810018557
810009805
810018562
810011871
810012765
810012786
810005004
810003377
810023129
810016686
810004448
111017251
810003305
810014090
111011676
111020932
111022261
111021188
111021026
810015286
810013775
810013862
810020920
810009625
810011698
810011869
810011867
810006000
810020336
810009600
111012100
810015338
810019996
810003570
810020386
810005998
810020563
810010481
810025095
810024148
810003900
810003953
111004457
810013933
810014023
111000875
111010428
810012835
810012841
111001194
111000858
111001171
810014156
810023085
810016999
810016582
810009529
810009480
810015510
111041459
810020736
111000874
111000481
111000266
111001626
111001452
111004446
810003488
111000928
111011875
101000005
810027330
111002344
111000894
111000856
810005644
810004429
810013761
111021208
111001370
810015472
810026668
810006154
810003394
810004441
111005867
111016489
810025093
810009645
810009928
111017279
111016495
111000185
111000225
111000595
111009313
111020928
810014175
810009938
810004975
111000391
810006149
810015495
111004349
111009343
111020315
810003747
810004036
111012392
111011310
111002930
101000004
111000223
111001366
111000931
111000929
111001186
810022722
111001614
810004182
810012840
810016948
111001179
810009246
111016505
111021186
810011657
111001929
111000893
810020183
111040378
810003385
810012768
111005859
810016188
810013659
111020331
111010411
111005868
111016485
810012775
810024152
810020154
810006018
810020430
810011802
810013821
111009678
810004192
810024117
810004018
810004280
111017281
810005001
810011895
810005002
111020341
810015346
810017521
810012919
810013616
810013690
111021381
111010596
111009318
111009329
111017284
111011487
111010226
111009627
111015028
810017538
111023915
810006393
111041446
810009530
810008735
810014011
810010584
810011716
810004509
810004365
111004102
810004850
810023110
810026354
810022621
111011101
810009800
810015586
810014118
111010401
810024960
810022733
810022615
810015585
810020191
810020385
810020524
810015641
810016287
810018829
810011921
810010478
111009637
111000841
111012426
810004921
111012161
111011692
810010514
810009221
810005605
810026389
111012425
810004089
810013810
810015488
810013974
810008254
111012375
810010550
810011622
810008263
111017267
810012823
810013723
111010997
111011856
111022227
810006001
810024158
111010063
111010365
111005799
111012349
111011861
111012557
111015027
810015288
810004713
810006080
810009934
810024283
111019501
810010556
111019529
810020552
111010348
810004933
810009537
810022624
810004545
810004551
810023123
810011614
810018787
810022614
810003782
810012850
810009903
810027265
810018576
810023126
810018726
810013688
810015440
111011388
111011678
111011402
111002754
111017255
111012363
111011288
810014009
810011890
810009500
111011685
810006068
810006052
810022753
111012545
810016725
111008759
810004910
111009297
111009675
111012564
810008269
111010253
111010626
111011687
810020232
810006356
810018746
810011825
810009927
111003209
111009332
111009275
111015783
810006042
810003472
111010351
111012414
810004673
810010612
111002892
810004902
810006378
111001925
810015996
810022782
810010620
810009930
810020758
810011924
810010648
810009798
111012120
111015793
810015428
810026611
810012777
810020495
810024295
810004852
810011568
111012539
810026358
810020389
810013743
810013738
810016598
810010607
810006152
810015434
111016481
111009317
810015354
111004145
111004112
810004092
810003396
810003307
810011639
111015356
111012422
810012780
111021604
111011862
810003304
111011050
810019971
111004443
810010537
111003738
111009659
111011120
111022206
810026664
810005636
810010628
810016894
810016980
810005606
111010404
810009266
810020267
810009799
810006272
810011584
810013739
111022187
111022488
810024101
810009779
810014102
810004498
810023595
111012119
111009320
810013819
810013677
111010581
810018606
111008875
810013860
111010381
810018816
810016506
111010419
810004502
810015320
111015418
810010489
810010510
810009516
810018729
810010611
810004987
810024138
810003529
810015231
111020170
810011834
810024327
111011893
111002355
111042827
810009273
111022208
810011861
810011931
810023104
111012536
810020459
810013911
810028852
111002251
111012529
810015361
810009750
111008838
810010525
810018574
810018589
810003515
810017500
810010515
810009777
810009543
111015040
111015425
810010516
111019505
810023103
111010414
810011818
111008841
111012165
111015383
810027629
810016866
810011620
111021171
810005649
810005917
111011648
810024276
810014154
810022587
810018590
810009280
111008825
810015280
810004494
111011501
810005658
810011617
111011068
111011128
111004344
111011688
111011516
810009947
810006161
111021122
111016615
111012525
111000231
111000917
810026408
111000341
111001460
111001625
111011395
111000896
111015701
111004444
111021407
111001127
111008844
101000012
810009566
111019556
810008756
810022571
810022725
810022562
810022572
810003402
111011653
111022214
111001159
111001389
111001699
111001373
111000204
111002115
111000230
111001164
111000308
111001172
101000059
111004368
111000491
111001189
111001726
111001421
111001809
101000006
111001383
111001429
111001795
111000653
101000050
111000289
111000297
111001601
111001149
111000305
111010346
111015011
111021423
810009940
111004111
111003193
111012151
810022742
111000860
111004135
111015034
810004928
111009296
810020448
111015017
111001759
111000891
111020197
111000328
810006367
810009533
810006094
810006088
111000390
810020505
810009289
810011621
111020163
810014144
111010217
810020540
810014019
810004084
810009252
111004385
810006143
810020422
810004368
810004431
810011650
111015658
810009935
111015666
810003956
111020933
810019948
111021621
810022566
111001902
810011776
810020426
810025115
111011329
111001810
101000052
111011309
111004108
111001174
111020929
111015761
111015786
111000611
810020538
810020543
810020546
111011290
810009839
111000895
810017509
111022221
111001832
111012390
111000129
111000338
111001130
111001129
111001408
111011091
810018818
810016008
111000632
111000933
111000292
101000049
810016001
810009294
810018803
111001905
111000136
111022151
111000889
111000313
810024103
111022253
111000138
810004424
810022634
111002905
810015340
111022256
111000645
101000056
111022165
810009458
111001145
810020741
810018838
111000339
111000122
111020937
810026609
810015511
810026675
810020497
810020753
111001139
101000013
111001744
810018719
111022213
111010611
111011510
810005915
810006015
810005957
810005980
810004210
810004173
810004215
810015318
810013870
810013847
810013843
810004478
810004472
111001430
111001917
111002241
111042856
111042855
111042853
111021112
810017494
810017514
810015603
810008381
111004375
810006409
111021393
810004579
111002581
111005849
111015683
111001952
810004364
111012501
810006354
810015588
111012522
810004857
111015430
111021180
810017001
810009837
810005146
111008877
810024110
810006349
810008330
810008264
810024118
810024156
810027605
810027602
810027611
810025098
810026662
810026349
810026361
810026638
810020373
810022932
810022936
810022751
810023074
810023560
810013746
810020234
810019995
810003387
810010474
111009314
111001831
111011293
111002118
810004496
810006360
111015436
111008821
810004484
810010483
810016605
810011664
810022630
810020417
810016971
111001711
810004208
810015326
810020428
810016938
810016904
111000644
810003690
810006425
111000386
111000181
810004034
810022741
111009682
111040291
810006415
810006364
810009268
810003756
810004372
810005655
810004992
810022740
111000206
111010397
810010446
111000899
810011901
810008732
810010438
111015714
111000309
810016917
810015335
810015345
810014165
810014171
810013735
111042845
111042860
810024125
810024128
810027628
810027625
810026621
810024935
111040520
810024011
810024104
810022930
810022888
810022911
810022900
810022743
810024012
111016509
111016613
111002130
111001445
810018827
810018799
810018748
810018742
810020181
810020395
810020565
111010609
111012418
111010620
111003901
111002769
111015395
111008915
111012357
111012355
111011119
111004116
111005801
111003979
111017270
810015576
111011300
810011752
111022244
111003757
111021524
111011094
810003888
810003889
810004093
810004082
810004286
111022232
111022234
111022217
111021616
111021516
810011675
111002123
111004104
111020196
111020191
111020199
111002862
810009489
111020365
111021107
111011308
111011070
111011460
111012139
111021520
111011643
111011518
111000647
111000281
810010427
111000274
111002953
111000291
111000264
111000226
111004125
111004002
111000295
810011833
810013578
111000301
111000293
810012947
111000327
111000191
111000590
111000648
810011781
810011772
111000625
810011673
810010464
810011671
810011637
111000397
111000867
111002337
111000878
111001187
111002255
111001940
111000870
111000854
111000395
111000593
111002125
111001910
810009580
111000197
111000265
111001591
111001784
111001740
111000286
111000651
111001782
111000843
111000712
111000298
111000121
111000284
111002332
111000190
101000007
111001391
111001157
111001616
111003916
810010471
810010522
111000119
101000057
111000135
810010567
810010623
111000883
111000288
111000852
111000596
111000267
111001919
111000616
111002333
111000862
111000650
111000183
111000192
111001392
111000326
111000591
111000134
111000132
111000310
111000654
111001690
111000342
111000283
111004350
111001621
810009445
810009447
111001166
810009506
810009573
810009591
810009577
810009562
810009464
111021605
111002902
111000629
111000485
810009471
111015673
111015689
111015778
810013664
810004211
810004707
810009891
111021426
810013606
111010056
111010046
111012408
111010206
111009323
810015563
810018568
810017499
810012781
810011929
810010508
810005593
810008261
810009596
810009931
810010526
111012170
810026658
810026646
810024287
111041456
810023571
810023572
111015005
111009677
111012132
101000010
810003962
810003752
111016403
810022583
111015400
111011401
111010606
111019481
810014021
111015705
810009237
810012809
810005159
810005152
810005153
810005158
111001931
810026347
810011912
111009643
810027264
111040296
810024925
810011652
810020490
810014117
111009630
810018786
810013973
810015431
810023051
810023563
111010600
810015232
111011023
810020270
810022579
810020174
810015514
810023100
810024962
810009222
810009229
111002757
111000182
111002945
810019978
810019954
111029125
810027617
810016182
810015722
810008747
810004998
810010617
810026412
810016889
810016874
810026402
810012689
111020195
810011774
111010088
111012540
111010064
111021409
111000932
111002959
111011493
111020183
810016878
810018584
810006418
810016002
810016914
810016990
810009239
111022192
810020756
111009660
810005938
810004445
810020739
810003300
810017518
810013752
810010502
111010072
810020397
111010627
111015654
810009623
111022252
810005595
810015424
810011892
111016511
111021219
810011744
810011742
810011765
810009922
810010467
810017493
810017533
810022627
810022628
810013653
111005870
810013599
810004768
111008903
111010237
111011657
810015615
810013729
810027297
111011082
810011880
810004381
810010505
810015631
810014010
111002103
111008871
810009923
111000936
810004770
111001903
810004772
810011851
111011096
111011871
810020242
810027277
810020254
810024142
810015621
810026600
111022236
111022731
810013975
810014131
111022733
810011736
810003483
111000628
111000126
111000180
111000124
111008910
111008562
810009916
810020163
111002773
111003915
111021176
810011589
111008858
810006022
111020171
111004139
111015424
111015385
810004434
111011339
810010532
810003527
810012837
111008864
810026414
810013907
111004096
810006359
810009919
111004016
810016687
111004018
111015422
111012515
810009531
810005622
810015608
111011010
111011667
111003874
111011007
810009248
810020511
810003897
111010093
810006084
810006163
111008902
810024284
810024289
810012812
810010599
810004273
810016187
810013950
810013999
810003535
810027613
810013951
810003786
810005545
810003753
810005505
810022575
810011761
810009560
810006144
810015478
810016981
810018745
810004705
810011918
810013808
810010610
810013596
810020750
111022265
810009272
810005504
810023558
810013656
810020516
810010596
810010642
810013750
810004587
810026651
810003516
810015561
810026672
810022784
810009561
810027276
810003486
810015316
810010572
111020188
810022796
111011386
810009590
810013566
810015358
810006061
810024147
810009552
111011642
810004442
111010360
111015002
810026416
810020526
810013881
810026406
810009226
111011891
810015298
810027560
810004506
111010258
111022182
111011017
810004099
810025110
810027272
810005985
810022899
810022728
810005671
111015018
111010345
111001775
810024015
810016898
810004548
810012943
111005793
810026657
810011757
810011643
810025116
810020381
810014092
810026641
810013931
810020251
810016952
810009547
810018752
810009787
111042710
810016868
810003689
810011768
810026614
111021497
810009473
810003310
111012164
810022573
810016723
810003627
810004896
111010235
111005864
810026608
810018841
810003468
810009486
810009499
810019981
810015702
810014034
111021635
810022632
810005914
111008920
810003536
111003973
810022724
111015781
810024967
810016867
111022179
810025113
810020919
810003687
810013678
810019962
111008885
810014017
810011788
810005466
810016935
810026669
810011632
111004109
111002926
111011459
111011499
111021726
810022623
111022200
810006156
810010578
111003617
810026645
810015284
810022787
810023580
810013844
810014168
111000232
810013954
111011662
810012771
111012169
810010594
810028877
111021212
111001741
111042851
111000623
111000617
111000303
111015423
111001378
111001799
111002574
810015493
111012361
111010232
111000615
111011389
111000934
101000003
810009791
810016604
810006104
111015297
111015416
111015020
111015294
111000884
111001441
111001148
111001168
111001142
111000927
111001162
111001153
111001465
111000930
111001473
111012385
111012160
111029129
111008883
111015388
111015399
111016501
111009303
111009335
111009308
111016512
111009333
111001451
111001593
111001608
111001425
111001414
111001443
111001748
111001819
111001800
111001794
111035646
111035635
111000711
111001160
111001620
111001610
111001152
111001385
111000187
111000275
111035638
111023918
111023933
111009663
111009658
111029118
111031821
111036822
111036818
111029131
111029115
111000788
111000497
111000791
111000396
111000785
111000787
111000790
111029128
111029134
111029116
111029133
111035650
111015782
111029130
111029124
111029123
111029126
111029117
111029119
111031816
111031813
111029121
111011297
111011084
111000892
111001606
111001438
111024275
111030113
111030111
111031819
111036817
111036804
111036823
111029132
111029120
111031812
111029127
111031824
111015015
111015009
111015003
111015021
111001898
111001943
111001945
111001946
111001928
111008560
810015564
810005991
810024131
810015513
810015436
111004134
111002951
111009286
111008905
111015709
111012373
111011507
111010431
810013573
810013563
810018717
810013558
810018736
111000914
111000912
111001433
810024009
111000277
111000394
111000350
810027599
810027601
810027618
810027616
810013608
810015368
810015369
810015370
810015366
810003478
810003479
810024159
810020360
810013601
810013811
810003623
810027284
111042849
810020535
810004014
810027258
810009250
810026379
810009801
810009908
111022175
810013897
111012506
810020177
111021630
111004101
111010263
810013849
810004097
810019982
810012787
810005062
810022898
810012925
111015033
810004100
111015408
111004105
810018828
111017286
810020731
810020449
810019955
810015623
810027298
810027335
810022564
810009228
111015031
810015285
810009225
810013894
810004274
810005095
111015429
810018731
810009633
101000011
810004546
810009814
810004204
810004096
810011783
810003741
810003955
810026398
810004576
810011939
111004026
810012838
810015605
810023592
810022596
810019992
111010242
111011497
810011911
810005672
810013857
111015873
810004085
111010433
810004767
810011887
111004381
111010366
111011311
111012148
111011468
810026676
810017540
810003748
111041440
111015680
111015795
810003404
810010656
810009503
810024303
810003298
810004856
111022490
810010603
111015363
810022591
810026383
111011655
111010262
810016877
111010379
111016488
810009525
111016507
810004473
111012528
810022763
810009845
111021526
810011737
810024331
111015373
810015587
810015596
810023237
111022250
810022955
810005654
810005598
810013611
810009595
810009647
810004015
810003412
810026654
810026667
810024013
810026650
111004440
810016909
810020530
810018596
810020229
810020528
810024134
111015775
810024129
810022738
810008233
810013855
810013949
111010264
111010214
810005615
111021517
810005653
111011481
810011819
810011889
810018763
111015868
810009638
810017002
111022239
810015580
810012097
810020372
810009802
810015565
810008733
810013790
111015863
810010429
810013646
111009279
111009282
810004859
810027608
111008847
111004152
111015369
111015360
111016499
111009288
810005994
810006020
810004178
111020159
810020443
810026610
810026606
810011665
111019488
111015690
111015779
111004348
111019495
111010211
111012514
111004143
111008884
111011139
810003309
810020447
111022168
111015401
810023089
810015425
810014139
810014170
810006350
810004222
810013641
810009492
111008574
810011580
810004423
810016991
810018754
810003970
810015422
810004026
810009806
111015695
111015869
111008837
111004005
810005056
810023069
810023116
810022904
810005633
810011792
810004674
810012884
111017250
810009771
810011723
810008376
111021383
810006142
111015694
111003906
810022931
810017510
810020501
810015650
810009491
810013635
810011709
810011922
810023066
111010374
111010386
111010394
111009292
111009274
111008881
111009324
111009277
810026613
810005601
810005524
111008551
111008926
111008900
111008893
111008565
111011398
111011674
111011482
810027280
810027296
810027292
810027293
810027266
810027306
810027256
810027290
111011060
111011456
111011405
810003749
810009821
111020184
810018821
111022149
111012419
810024301
810004986
810026630
810023099
810004206
810024270
810009822
810013737
810004086
810004493
810016000
810004480
810012805
810011674
810020491
810009603
810004524
810025117
111001133
111000296
111001870
111001623
111000194
111000189
111001812
810012769
810009254
111002363
810016921
810020755
810018812
810020437
810022892
810026394
810003895
111004160
111002368
111005852
810010499
111016490
111000229
111000280
111000201
810004395
810022889
111011509
810006370
111012517
810009796
111015393
111021103
111040290
111041463
810024106
111020367
111012163
111021606
810005987
810004486
111015774
810009542
111011049
111011116
111011272
111011515
111010254
111010240
111010334
111010341
111010233
111010385
810009522
810009944
810008260
810004912
810006365
810006098
810009941
810009830
111012550
111012527
111012500
111003872
111010047
810004209
111005812
111001163
111001158
111023929
810016710
810023589
810018757
111011020
810015475
810009906
810006173
111040299
111002888
111002865
810008345
111011140
810015333
810015317
111040294
111024130
810023076
111000849
111000921
810009907
111004149
111008818
111000844
101000008
810015325
111000907
111001603
810015319
111000847
111001165
111012130
810022560
810008338
810017013
111000299
111000311
111001700
111011107
810006006
810004370
111011046
111000300
111000337
810006083
810006063
111001594
810004027
810004088
111042834
111020369
810016945
810016899
111008824
111000861
810026636
111000348
810023057
810022770
810022893
111016482
111001822
111001853
111001779
810015423
810014004
810008737
111000903
111000331
810009583
810013816
810013841
111021205
810005979
810024951
111002117
810018833
111010338
111010412
810003960
111004383
810015485
111001955
810005629
810005659
111015669
111004120
111008819
111010369
810014177
111010074
810006028
111042863
810006416
810026671
810009541
810006074
810026602
810011654
810022765
810006157
810022761
810006151
810027257
810009284
810020353
810011862
810010533
111015768
111011859
810004371
810009578
810008739
111021495
111020364
111015717
810024296
111001904
111004027
111004369
111001909
111002514
111011298
111002250
810018715
111002312
810009283
810015480
810013872
111021614
810009574
810009528
111002949
111002933
810022916
810017529
810008755
111020912
810004392
810020431
810018561
810015343
810009261
810026605
810026362
111008559
810020550
810026396
810020561
111020349
111004456
111019487
810003688
810020435
810009497
810010637
810013968
810013947
810009769
810020502
810013644
111012107
810003512
111015864
111019503
111015798
111015799
810016872
810005582
111019549
111011106
111010054
111021179
810011697
810020734
810016902
810008331
810004291
810004903
111003606
111004371
111004382
111021111
810015479
810010479
810011839
111009631
111009668
810009295
810011740
111018842
810019987
111021207
810016684
810022736
111009638
810004438
810014012
111009291
810004393
111015376
810011863
810013689
810009615
810005584
810016186
810013850
111012110
111020930
111021533
111019494
111008914
810013893
111015703
111008886
810011706
810009772
111008568
111010049
810003566
810003484
810011936
810012103
810012091
810027319
810027314
810027324
810023562
810023570
810023060
111021096
810011900
810012681
111011319
111011650
810025090
111011485
111010641
111010406
810023598
810004475
810004525
810004495
810024959
810024965
810024933
810013768
810013763
111023939
111023928
111023938
111023937
111015711
111015295
810004377
111002253
111002334
111002110
111016504
810009548
810009535
810023124
810010581
810010491
810010460
810010662
810009828
810011729
810012911
810012959
810012883
810012889
810012958
810012928
810009937
810010519
810010539
810010649
810011809
810011803
810012882
810012903
810023070
810023113
111004015
111004012
111003971
810009601
810009592
810024936
111020313
111021181
810003388
810003390
810024107
810006168
810004194
810004375
810004032
111009625
111018835
810017012
810024122
810004284
810015597
111002362
810003530
810009300
810011913
810011611
111020277
111021631
111018836
111008873
111000887
111008888
810018809
810018560
111010336
111004155
810027631
810022635
111012157
810006085
810016003
111011328
810008372
111015022
111015032
810009244
111001764
111015765
810004510
810009768
111011150
810012684
810027334
111018841
810004681
111015866
810008333
810003378
111015702
810005107
111003890
810009892
111004008
810009544
111008901
810004553
810011708
810012095
810008253
810009260
111015693
111003912
810017531
810003929
810004174
111000880
810004444
810009255
810015715
111016621
810011881
810009621
810009634
810009749
111002356
810016951
810009649
810018794
810011586
810005642
810016965
810022901
810026407
111011345
810009263
810004581
810011683
810008367
810020145
810020169
810014133
810015709
810020420
810004508
111012127
111024156
111009283
111001935
810004202
810013642
111002324
810015606
111010082
810020161
111022197
111015370
810004278
810004505
810009238
111017273
111000199
111001135
111003194
111000334
111000594
111000619
111000489
111010373
111010367
111020255
810008344
810014115
111004360
810009788
111022231
810015598
810022789
810024929
810027262
810012954
810009945
111010376
810024277
111001624
810011680
810011801
810027304
810022580
810015601
111001154
810022608
810020738
810020383
810020729
111001783
810027270
810027337
810011806
111001140
810025094
810025100
810027317
111003616
810011767
111000877
810011758
810020337
810008383
111020925
111023913
111021403
111021732
810003525
810012822
111012103
810010469
810013785
810005617
810022933
810010634
111010424
810016176
810009278
810020265
810009599
810011727
810003514
111011857
810009637
810008724
810003513
810022963
810020438
111017254
111015790
810011826
810015260
810015296
810023119
810010527
810011687
810013638
810013802
111015704
111016625
810004543
111002365
111011400
810004679
111003736
111011011
111008870
111018837
810010600
810009587
810017527
810004908
111016508
810003470
810013598
810012968
111020169
810016005
810017005
810017523
810004934
810003375
810008342
810022620
810022568
810024111
810024112
810018806
810009609
810015627
810017519
810012907
111004119
810013876
810010543
111012125
111005804
111005807
111012124
111011004
111011644
111003570
810006044
111012503
111002585
810024120
810026635
810027607
810027333
111021397
111002374
810003961
111011019
810012096
810004422
810013602
111010060
111022195
111022204
810009570
111011127
810012844
810018604
810013882
111042857
810024297
810024943
810003401
810005363
810025101
810024121
810005608
810005624
111024133
810009579
810008259
810006092
111040289
810026666
810026377
810026632
810026527
111015006
111008869
111020248
111008866
111008907
111004356
111008826
111022219
111003992
810013864
810004376
810003355
810003693
810020525
810027643
810022781
810016737
111011026
810008758
810023127
111020275
810008752
810017015
111011327
111009322
111022732
810023108
111041449
111011658
810008726
111021027
111010403
111022738
810024291
810011686
810005637
111012386
111012428
111003612
111010375
111010399
111020185
111012371
111010058
111011855
810003313
810005529
111011684
810004083
810027623
810010562
810023109
810005888
810004504
810004855
810006133
810004922
810022745
810004894
111042852
810024940
810010551
810010605
810009648
810015624
810010598
810015648
810004427
810020239
810013883
810009807
810010616
810014024
111010393
810013927
111008912
111004017
111022169
810006137
111012099
810011629
810022964
810020520
810016985
810008760
810020518
111005794
810009589
810023216
810006372
111021626
810011780
810006368
810004669
810004214
111004148
810026628
810023092
111003889
111012405
810013892
111015379
810016986
810009293
111012523
810020375
111008895
810009949
111018840
810004197
810004433
810009296
810018587
111021082
810023579
810023581
810011787
810013634
111022198
111015382
111002881
810011799
810013869
810024146
810005934
810009584
111002685
111002626
111012398
810008730
111002582
111011039
111002122
810005603
111002870
111002675
810016593
111009655
810017520
111018838
810003901
810008366
111022167
810009773
810009448
810009454
810005099
111011865
810015364
810020514
810023596
810006076
810026370
810009628
810020223
810019990
810020224
810005151
810005537
111021599
810008356
810024323
111015004
111015409
810004180
810005619
810009808
810008757
810013888
810011766
810009490
111022189
810015258
111015357
810016997
810016926
810020415
111015668
111009300
810004854
111015421
810006273
111000349
111000496
111021622
810004193
111012505
111003815
810020144
810019973
111012509
810005990
810013824
111020317
810013617
810020260
810004024
810016594
810018727
810013663
810013667
111015663
810005302
810005299
111015417
810008740
111008567
810009776
111008842
111011131
111011483
810004853
810013745
810020342
810020357
111010426
810026655
810004906
111008909
810004714
810008341
810016596
111005866
810003626
810006106
111010198
111041436
810020515
810015714
810015599
810015723
810027323
810020354
111000621
111001150
111017261
111001151
810012967
111015384
810004269
810012937
810005662
111022207
810003411
810006270
810013782
810023594
111019548
111022210
810003297
810004556
111019482
111019497
810008347
810009917
111021399
810008377
111022491
810020246
810020439
111042843
810009588
810010553
810003959
111011490
111015014
810023568
111015862
810015490
810018570
111015030
810016584
810010670
810022772
111015016
810011684
111021095
111015035
111020910
810018573
111020926
810009504
810020742
111012156
810022734
810009457
810009622
810009748
810022601
810006034
810010425
810027307
111010576
111041434
810003966
810027636
111010608
810004019
810027650
111009341
810012763
111042832
111042706
111023927
810004020
810020744
810013574
810004033
810016178
810027639
810013879
810017495
810015618
810011691
111041461
111012376
810005143
111015674
810022755
810022791
810025845
810026059
810026027
810026269
111041080
810004544
810013730
810013817
810015646
810023578
810016895
810004098
810027278
810013771
810005510
810005464
111024129
810012845
810012790
810012811
810020166
810020416
810020457
810027646
111011295
810024941
810020451
810020519
810003565
810016714
810013861
810004440
810012948
810016708
810016996
810012846
810026616
810027295
810015622
810013591
810013789
810004993
810004025
810022559
810020236
111015696
111003596
111003883
810013833
810003474
810020180
810006025
810006023
111021172
810010606
111014614
111021384
810006016
111021404
810018797
810018722
810010609
810020380
111020246
810016732
111019509
111016623
810026642
111010591
111010597
111010199
810010559
810024265
810020184
810013822
810012761
810012827
810008373
810009780
810009564
810009569
810009519
810009299
810015430
810020522
810018552
810016915
111003314
810006053
810005635
111018846
111024157
810004703
111010244
111002107
810024964
810024950
111017280
111014613
111020167
111010248
111002873
111022247
111003812
810006371
810004079
111001138
810006172
810009836
111003737
810005996
111019500
810024927
810010639
111003999
111021531
810020560
111003806
810026607
810013691
111040522
810010663
810020188
810019979
111011306
810004997
810011764
111019914
111015434
111011880
810010513
810025120
810020384
810027554
111021525
111021020
810012764
810012807
810011747
810011816
810020248
111041466
111022263
810005462
810006024
111011463
810010549
111023926
810015283
111017252
810011713
111022496
810027612
111011494
810009455
810008340
810012101
810003311
111002767
111002766
810009540
810020350
111010078
810027561
810027557
810013929
810013988
810015520
111016617
810003475
111011121
810027638
111012159
111011008
810003927
111012140
111010432
810003757
810015491
810016513
810003957
810011714
111021528
810008360
810013687
810009484
810014001
810014109
810009911
111014609
111011867
810012832
810012111
810006107
111020911
111021600
111010230
810015438
111011148
810010426
810009790
111001948
111022163
810015421
810006070
111001900
111009664
111003886
810009809
111014615
111021502
810004175
810004773
111003905
810014122
810022739
810026385
810024942
111021615
810011634
810011717
111015682
810018711
810011915
111011074
111009293
111003212
111001173
111020176
810006027
111000282
111022156
111000198
810004999
810009803
810009817
810009755
810009924
810015503
810012970
810012092
810011893
810011666
810011817
111010067
111009648
111010352
111011066
810009551
810009567
810013749
810015483
810016599
810016697
810016504
810017504
810017532
111011645
810022908
111010221
111010637
810027281
810026425
111040287
810025121
810023084
111021491
111022188
810004221
810003622
111022183
810006271
810006361
810005886
810005587
111011679
111002480
810004796
810003354
810004935
111003586
810016975
111002247
810026640
810020917
810011636
810015611
810010455
810022957
810003628
810024155
111021116
111021190
810009520
111021382
111016616
111016622
810006097
111021411
810005632
810023591
810018723
810005639
111040293
111021731
111009272
810004907
810009789
810012913
810010476
810004187
810011769
810006062
810015227
810013731
810018747
810018605
810020348
111011654
111000886
810003299
111010384
810022906
111009338
810012924
111020366
111012168
111015433
810015508
810018602
111008918
810017511
810005935
111009302
111015685
111008889
810003467
810017512
111008880
111011461
810006048
111012409
111015364
810011648
111017263
810026356
810003532
810011857
810008235
810012953
810015289
810017513
810005000
810017507
810008359
111010405
810020748
810004929
111017285
810009242
810019994
111021200
111004004
111003618
111008845
810026634
810004282
810012965
111009636
111017265
810023600
111042844
111015426
810004198
810024324
810003755
111011694
111010362
810016704
810003738
810022581
810022920
810004195
810003301
810004031
810018608
810022785
111021634
810015477
810027311
111015657
111011889
810004213
111015686
111010618
810008267
111016502
111003718
810016897
810016007
810026652
810016893
810019950
810004218
810013873
810003758
810015257
810026384
810005992
810015716
111021209
810015725
810006041
810004077
810013575
810027316
810003784
810023577
810004145
810024109
111014783
111012348
810024294
810009838
111008836
810013895
810003964
810016603
810009517
111011149
111009644
810022584
111012531
111040295
810008723
810025091
810013829
810013619
810022773
810013674
810006049
810012104
810025108
810005623
810020446
810020434
111011273
810020531
810015728
111012353
111008571
111008561
810003302
810020754
111004384
810022600
810005936
111015019
111010577
810024272
810024322
810020521
810020261
810020404
111005796
111005815
810004554
111000484
111000486
111000304
111000335
111015698
111009339
111012121
111011496
111000840
810010544
810010547
810005663
111022497
111000897
111011458
810027261
810025092
810011820
810012828
810012834
810009747
810005981
810003324
810006158
810003308
810009462
810005596
810022629
810016292
810024969
111040375
810005105
810022790
810011578
810004212
810004270
810004574
810003571
810011705
810011638
810011633
810016288
810020517
810024275
810024286
810011870
810009816
810009604
810009834
810009827
810018826
810008346
810026526
810015619
810020551
111021102
111015687
111021182
111002783
810004080
810024006
111024810
111000287
111005858
810015473
111016497
111020178
111004150
810016698
810016699
111011279
810015637
810009640
111015770
810006194
810006132
810008327
810010573
810010575
810024974
810024008
111002232
111000713
111002911
111005816
111005855
810011793
810020727
810010636
810024426
111021173
810005535
810006148
111019538
810008329
810004909
810008268
810003408
810004775
111020377
810023588
810011668
810015595
810013846
810004176
810009532
810004555
810024293
111010238
810008354
810018785
810027598
810008375
810004200
111009642
810004542
111010358
810011715
810016998
810003952
810018831
810018824
810003410
810009651
111008868
810005628
810009804
810012931
810024123
810024132
810024114
810015264
111008853
111021627
810020238
810009784
810005989
810025102
810022924
810005599
810009624
111004137
810012106
111011462
111011863
810005046
810025106
810014176
810004426
810009762
111010417
111010201
111019507
810013568
810005078
111015697
810004030
810016942
111011504
111011881
111020931
111011686
810004668
111015777
810009495
810027259
810022561
111015024
810009644
111020193
810010651
111015029
810020423
810003482
810004369
810015507
111012530
810022749
111012158
111009276
111011280
111011088
810009921
810010646
810008265
111011474
111015671
111015013
810004380
810016689
810010430
810010444
810016911
111022152
111012105
111022737
810023068
810011702
810009794
810018599
810010517
810014173
810005527
111011512
111021202
111012521
810006419
111021098
810024326
810020363
111011086
810004177
810015568
810005995
111003214
810023247
810008237
810011720
810027325
111018839
111018844
810012830
810027563
810023052
810012788
810013555
810012120
111015038
111018845
111020253
810022618
111008827
810004979
111008856
111010429
810006147
810005610
810006155
810020408
111008823
810015614
810005920
810015708
810024119
810004911
810027287
810015506
111009665
111022494
810022699
810017539
111015396
810004996
810011690
810011699
810016888
111004128
810004483
111017277
810005528
111004130
810005884
810011631
111011338
111016491
810010563
111015867
111015039
810025107
111008839
111040540
111002884
810022944
810004575
111001890
111003592
810020432
810004285
111021598
810023996
810024937
810022768
810010560
810004029
810016995
810024938
810022631
111015766
810005591
810016906
111012541
111020362
810020356
111011318
111020316
810020335
810004522
810013971
810020510
111020278
111020355
810016726
810022746
810023602
810024133
810020454
810020393
810024165
810020244
810020379
111011115
111004144
810010432
810004709
810015707
111022216
810011782
810009824
810003490
111020332
810008371
111015362
810016721
810009478
810004716
111020370
111011670
810015336
111011108
111010205
111004370
810004201
810018610
111003882
810004541
810020252
810009258
111002942
810016722
810018593
810022905
111022262
810009761
810004927
810026656
810016700
810009444
810003356
810016601
810016586
810015633
810003317
810020737
810020377
810026363
810004519
810009513
810003534
810006033
111021206
810020424
810008748
111021530
810024934
810013866
810023121
810020175
810008251
810012785
810013955
111015368
111000620
111005802
101000058
810026660
810006159
111000195
111001184
111000853
101000060
111011069
111001134
810011724
111001871
111000236
101000054
101000002
111015665
111001390
810013839
810005526
810025103
810019959
810012683
810012939
111001815
810020539
111042074
111042098
111042067
111042029
111042242
111042071
111042055
111042092
111042097
111042030
111042080
111042060
111042637
111042235
111042086
111042145
111042666
111035611
810028876
810010473
810010487
810018811
810018807
810028858
111015381
810004771
810015516
810020225
810016502
810004514
810020555
810015498
810005660
111019512
810004680
810011579
810004711
810015562
111015377
111017259
111020260
810027649
810009813
111019518
810024930
111012143
111022180
111015001
810004937
810004492
111002931
810004970
810011814
810011827
810005508
810013626
111002354
810019977
810020271
810011741
111001949
111019521
111003589
111002468
111003293
111002882
810017502
111022246
111019489
111019551
810020401
111011114
111003742
111022159
810005149
111003802
111002361
111010350
111002897
111002326
810013875
810011730
111003303
810022565
810013836
111002879
810010638
111002899
111003803
810023224
111012128
111002925
810013915
810019986
810005452
810020745
111019478
111012532
810013837
810010472
810020392
810023561
111011882
810018721
111011292
810011813
111004342
810012896
810011733
111002868
111024307
810019974
810012881
111011035
810013845
111010077
111011322
810013982
111011056
810018753
810020564
810016728
111002950
810024975
810023082
810022952
810022950
810018793
810023236
810009752
111019543
111012423
111003294
810022943
111021193
810012956
111002961
111003199
111011335
111008849
111002894
111011294
810016518
810022912
111011895
810022567
111002934
810016923
810012915
810018585
111003741
111020198
111002116
810011626
810010570
111011018
810028873
810023097
111001951
111002872
111002320
810004916
810013648
111019554
111001915
111010586
810009896
810019958
111002315
810011749
111002682
810011866
111011055
111019530
810023219
111002915
111002889
810011566
810011908
111008865
810015277
810022595
111003292
111003885
810014116
111002762
810027621
111002960
810022910
111002939
810016727
111005791
810015730
810020396
111002577
111021730
111021083
810009538
111020254
111022190
810011935
111002927
810010615
111021203
111019552
810009895
111002929
810004981
810010586
111015652
111020337
111011138
810014152
111012555
810018577
111003207
111020923
111024310
111003419
111003210
111023919
810012922
111012427
810016293
111011143
810005108
111010220
111002871
810016706
810018563
111003573
810022929
810013827
111011100
810016961
111010646
111020175
111002922
111004006
810010501
111012109
810010608
111022264
111002627
810011701
810005531
810018572
810022593
810020263
810009449
111020165
810012940
111003728
111010418
111003904
111002753
810012929
111001950
111011305
111003745
111015432
111010595
810018733
810011829
810017496
111003310
111002474
111002329
111019486
111005853
810012951
810016739
810024971
111003746
810013852
810010587
111002676
810016585
111002956
111021115
111002917
111024144
111015797
111020268
810011656
810016516
111003206
810022897
111011872
111012368
810013834
111002759
810013649
111002779
111003704
810010643
111003748
111022154
810013868
111024311
111021395
111015796
810013935
111003899
810014132
111023921
810023226
111011390
810010523
111001934
810023222
810013922
111009673
111003215
111020273
111003816
111002955
111010223
810013928
111005865
111021498
111001916
111001944
810023583
111020333
810013640
111012149
111003213
111003196
111019493
111011103
810011726
111021024
810013970
111015407
810012920
111020360
111002932
111023935
111001924
111010200
111011099
810022594
810013628
111002579
111002943
111004153
810009539
111002229
810019989
810004990
810028849
111024153
111020323
810020391
810010589
111002784
111003304
810016977
810005147
111010071
111002918
810012902
810011938
810023075
111024132
810011934
111021092
111002885
810016588
810009849
111021178
810016994
111002751
810011855
111003902
111012162
111011383
111003900
111021412
111021185
810009758
810022780
810009282
111003908
810016884
111011693
810015999
111002940
810004904
111024125
810013859
810017497
111003218
810014020
111003703
111020256
810005103
810010654
111011125
810011916
810013863
810017009
810022958
111002919
111003800
111017268
111003724
810013994
111009645
810009909
810013983
810010668
111002921
111021398
111003311
810012914
810014002
810010644
111003216
810009893
111012367
111023923
810005145
810013953
810013830
810009835
111002680
810013838
810013793
111010052
111011647
111024146
810018598
111011038
111021424
810013856
111011301
111003887
810005471
810016976
810010557
810004973
810011877
111019545
810009915
111003729
810011824
111009639
810023062
111002098
810005060
810011928
111011457
810016918
810005045
111004088
111003752
111005803
810017535
810011843
810018712
111021114
810013926
810018578
810018609
111004157
810005064
810013977
810014164
810009262
810011789
111012366
111002948
111011471
810009556
111009328
810011771
810011853
810016738
810023223
810020358
111012411
810022778
111002787
111020192
810009898
810011755
810010421
111002924
111021089
111010368
810012888
111020927
111019479
810012961
810016682
810005077
810005076
810016683
111021391
111024143
111021427
111020372
810022949
810011762
111002941
810014015
810013754
810012897
111002914
111008556
111003217
810005631
111009690
810011635
111002311
810005061
111011271
810012894
810016901
111002325
111002774
810011807
111020322
810013948
111002761
111021527
810014028
111011037
810005465
111003219
111012116
810009764
810016900
810020170
810011647
810016931
810013990
810009526
810024958
111021388
111003599
810015647
810028874
111001913
810011575
111012138
111011135
810020351
111020914
111002913
111022194
810022758
810011682
810005469
111001954
111011315
111009634
111011044
810005050
810005102
810005457
810005148
810013991
111011852
810004932
810010630
111002129
111008554
111021204
810018814
111024139
810009643
810022890
810011874
810011847
111012516
111011883
111019532
810018768
810010457
111024151
111003735
810018588
810013957
810013978
111012507
810009950
810012926
111011057
810010461
111024126
810011800
810023073
810013779
111003197
810004984
810012899
111003574
810010625
810011619
810019957
111011344
810020440
810005054
810005453
810012944
111008822
810013925
111011105
111023936
810012901
111012118
111003977
810013825
810018554
111011129
810016919
111021174
810016740
111005871
810005075
810011731
810004925
810011570
111011877
810011897
810009496
111002101
810016907
111002471
810004972
111002341
111011087
810017525
810015429
111012412
810017541
111002673
111003708
111003751
111011015
111015397
810013801
111023934
111011682
810023565
810022915
810005063
111012512
111010234
810023071
111023922
810024955
111019492
111008913
810014137
111002962
810009782
111020342
111004455
111012360
810009281
810023101
810016928
810015629
111002343
810012946
810016690
810012917
810018601
111020336
111011289
111002328
111002473
111011076
111003712
810020240
810014129
111019483
111019515
111002907
810013905
810013959
810011844
810013639
810022793
111011136
810023098
111002890
810005303
810011759
810005473
810005467
810005666
810010540
810005142
111002764
111002366
810016972
111003613
111003614
111011864
111011307
111012369
810012904
810011927
810023118
810023228
111003597
111022170
810011745
810023231
810004897
111020345
111020250
810009820
810013831
111003715
810013828
810016973
810012895
111002901
111002878
810016589
111003297
111010642
111002900
111012416
111002906
810018566
111002571
111010207
810022934
810012934
810013958
810024968
810014147
810012933
810013795
810013979
810009469
111005856
111010068
111002583
111015762
111021385
111001797
810009646
111023916
111012551
111022150
810010627
810010604
111002331
111000848
111024136
810010624
111010209
111001923
111004094
111003577
111002120
810010503
810010669
111021406
111024148
111002469
810005613
111011062
810012900
111002254
111011006
810023106
111021532
810020192
810025096
111019537
111021085
111010407
810012886
111002121
111011846
810018718
111012537
810005155
810019952
810012892
111003744
810005096
111022241
111003202
810015705
810009461
111022162
111002513
111010594
111012155
810011732
810012918
810013921
111011488
111001918
111001936
810004919
810028850
111011113
810009483
810012912
111019553
111010076
111003702
111019499
810015721
810011838
810016512
810018725
111011065
111002896
111004453
111011075
810018732
111011316
810011661
111024138
810005049
810016009
111023920
111029122
810022771
810009948
810012898
111011331
111024150
810005463
111002108
810013914
111011323
810009815
810013993
810023593
111002238
810004978
111002476
810017004
810016508
810018756
810016962
810013760
111011110
111003811
810011933
810011864
111002104
810005048
810017545
111011001
111003891
111011341
111023917
111020920
111002624
810016734
111011064
111010089
810016924
111021088
111010342
111020319
111024135
810013792
111024141
810003326
810017536
111020272
111002946
111011092
810012893
111010623
810022558
810010576
111020348
111002944
810020355
111021413
111003609
111009667
111010095
810016705
810012957
810009823
111011012
111015772
810023564
111024147
111010371
810022948
810011836
111004353
111024149
111002515
810018714
810012905
810010538
111002322
810012118
111021493
111021215
111011029
111020922
810011906
111019533
810022585
810009774
111008555
810023058
810012909
810016954
810005079
810013939
111002340
111002112
111012358
111003698
810005059
810015600
111001920
111020917
810009904
111011391
111003909
111012377
810009224
810009899
810016505
111024137
111011342
111002864
810022942
810023128
810016879
810014018
810017007
111012535
810011777
810022730
111002335
111004448
810013981
810020339
111019485
810012941
111021402
810011676
810005058
111002102
810013964
111012403
111002239
810011739
810019988
810018822
111002860
810018551
810010655
111002367
111002875
111021191
111020938
810018708
810016963
810005468
810011794
810016886
111021090
111002231
111015720
111016484
810010566
810005359
111003797
111003299
111002947
111012504
810010534
810009274
111019539
111019540
111003601
111019513
810011804
111024155
810011646
111002099
810005104
111004003
111011021
810005055
810009832
111010400
810023253
111011130
810011662
111020180
111005850
810005583
111010383
810018558
111004451
810022732
810014016
111003805
111002478
111002877
810020185
111003722
810023227
810023225
111008570
810005051
111003898
111004131
810015240
810023122
810018730
810022626
111003740
111002908
111019490
810009905
810005157
810022586
810004901
810020365
810020371
111020919
810005536
810022588
111005810
111011304
810013781
111024312
111010347
810004988
111004090
111019555
111003308
111011296
810005456
810010585
111022186
111010370
111019546
810010590
810022592
810017524
810016913
810016925
810013932
810018553
810017508
810018843
810020164
810023241
810023096
810004994
810020407
810024972
810016932
111020363
111002516
111002858
111002565
810005300
111002242
810013884
810023998
810013809
111020913
810009629
810013614
810018741
810009233
810011923
111020245
111001774
111008828
810013636
810023086
810012949
810005156
810016916
111023914
111003578
111010197
810012964
111022257
810016707
810016590
111002370
810004891
111011285
111002785
810012906
810010622
111012561
810022913
810020344
810023088
111003892
111011104
111008552
111011134
810015518
810011860
810012688
111001957
810028855
111019517
810023255
810023239
111001393
111001459
111003205
810028871
111001719
810026381
111001596
111001448
810009775
111035657
111040382
111040286
111041426
810009611
111015870
810011645
810020549
810012687
810011858
810004971
810027642
810011613
111021110
810015341
810016709
810016946
810011627
810027627
810020442
810024162
810026633
810024161
810028859
810009279
111010624
810027600
810012758
810012690
810022777
810004220
111011396
810005638
810010580
111017257
810017492
810017542
810004425
810027620
810004573
810020441
810004443
810020427
810011625
810009550
810008384
810008239
810009593
810022607
810015271
810009251
810010486
810014029
810011795
810012686
810016873
111000482
111000612
111001364
111000904
111000344
111000851
111001776
111000490
111000871
111000306
111001474
111000680
810022909
810003409
111000487
111020939
111012546
111008835
111000480
810022747
111021500
111024308
810004172
111002244
111020344
111019491
111018843
111008566
111008867
111000340
111014608
111011874
111004147
810016890
111000626
111000682
810015591
111001824
111001701
111001597
111001396
111001730
111001444
111001899
111000598
111000346
111000302
111000387
111016021
111001760
810006077
810013908
810026399
111001689
810012952
111003727
810014151
810011849
810013583
810011925
810013877
810025109
111001721
810015265
810012098
810015634
810022786
810022756
810011888
810024136
810020399
810024116
810011831
810008240
810009230
111011061
111011665
810013800
810013820
810004670
810010561
810010521
810013788
810009514
810009297
810008729
810020235
810020147
810004497
111011111
111011033
810008362
810008380
810017490
810017516
810009841
810009901
111005811
111016619
810022779
810022556
111011118
810004477
111002893
111002883
111010592
111020166
111041423
111011102
111011337
111015667
111002678
111002674
810024928
810004708
111001933
111002770
810011884
111008923
810004795
111015655
111020358
111020174
111021603
111021597
111010644
111010601
111010628
111008919
111009656
111009653
111009624
111009685
111010423
111004091
111004362
111004364
111003981
111005863
111031914
810026235
810026044
810026128
111031589
810025838
111002578
111009647
111011847
810010456
111000920
111024309
810018738
810011658
810011688
810012806
810012843
810012813
810012825
111001360
111001132
111001394
111001411
111001412
111001380
111000916
111000924
111001188
111021422
111001463
111001377
111001595
810006108
810006081
810006086
810006128
810014022
111003198
810006422
810006382
810006373
111012524
111010359
111010575
111010638
111003208
111003306
111003211
111010243
111010617
111002243
111001416
111001382
111001359
111001466
111011078
111011089
111011085
111011892
111011651
111012106
111011326
810006385
810006387
111010387
111004447
111003734
111003910
111010631
810012778
810012773
810010548
810010667
810010569
111003723
111003699
111003611
810006379
810006377
810006384
111004361
111021610
111021513
111021494
111004127
111004461
111004452
111004151
111004099
111003873
111003974
111004359
111010225
111042846
111004021
111021519
111004013
111042839
111011385
111021192
111042705
111021021
111000923
111012145
111042700
111010639
111041438
810008718
810026612
111011495
810018804
111011479
810018788
810026374
111021518
810006047
111008572
810006427
810008270
111020173
111010081
111011080
810006391
810020398
810009481
810009472
810013878
810015649
810024099
810011574
810006424
810020245
810020186
810020257
810020256
810012848
810012794
111000872
111000869
810006414
810006412
111001371
111001399
111001365
111002937
111002778
810008247
810014123
810014100
810013605
810022902
810022935
810006035
810006090
810006124
810006099
810014031
810014135
810014136
810014120
810013620
111003221
111003300
111003316
111002309
810022590
111010353
111004133
810010571
810010535
810006386
810006388
810006428
810006389
810023054
111011514
111011870
810010477
810010480
111001791
810008734
810015235
810013755
810024137
810023091
810018801
111041443
810011910
810027255
111011137
810011840
111012394
111019502
111010222
111011666
810009844
810015574
810008370
111003579
810015643
810004205
810013803
111040285
810020553
810023059
810024267
810026623
111041465
810026663
810015500
111020320
111002510
111010073
810005621
111021105
111011286
810013665
810009502
810016180
111012417
111009287
810016300
111017253
810011615
810012767
810016927
810004582
810022921
111015041
111008834
810005534
111011477
111005032
111041372
111031650
810026109
810026096
810025996
810026152
810020412
810004191
810017000
111023932
810008743
810004583
810012793
810012779
810018565
810018556
810003376
810003314
810024285
810024325
810010588
810004917
111017276
111017278
810005585
810005611
810005634
810004899
111022259
111022495
111022222
810009597
810008379
810013832
810013871
810014124
810014161
810013976
810014027
810013655
810013748
810022729
810022754
810022766
810016291
810020345
810020562
810020362
810020388
810020253
810020171
810020374
810026424
810026418
810026350
810024007
810024004
810024105
810023569
810011681
810004976
111011675
111011281
810004675
810003893
810013734
810004287
810010621
810020231
810009913
111040539
111012396
810026643
111009311
810027268
810027303
810011773
810012833
810008365
111041445
111011112
810004715
810006150
810012114
810020458
810020456
810020413
810020349
810020453
810020378
810019984
810019960
810027615
810027606
810027597
810019991
810019967
810019985
810019949
111022185
810019966
111022238
111022202
111022181
111011333
111011476
111011291
111009652
111009650
111009691
111008820
810009494
810009467
810009477
810009518
810009507
810009443
810009510
810009487
810009512
810009505
111008872
111021079
111022191
111001830
111022249
111022245
111010066
111010231
111010255
111010224
111010196
810027610
810027619
111001190
111008564
111008563
810018836
810018789
810018790
111036820
810011907
111001746
111002573
810009498
111003298
111001768
810016920
111041418
810023061
111004136
810009508
111004460
810016953
810015612
111001897
810005984
111001912
810004189
810027640
111001826
111011503
111008876
111011054
111011868
111015037
111015676
810011790
810005630
810009793
810010613
810023251
810005645
810026665
810016944
111012102
810003624
111012513
111020321
111042854
111010333
111014611
810020269
810010440
111015371
111012122
810016958
810009754
810004226
810010496
810009594
111012374
111012372
810015476
810020500
111016494
111016618
111020274
111002472
810020558
810005656
810006141
810018720
810016934
810020498
810022625
111009310
810011583
111004031
810008246
810009778
810015329
810018759
810018761
810018728
810018594
810018832
810018817
111019511
810024948
810024949
111015678
810020740
810004487
810004501
810004578
810004706
810004488
111018834
111015713
810013632
810013593
810013559
810012923
810013623
111018833
810008749
810008750
810008742
810008741
111042840
111042862
111019510
810011753
810011770
810011785
111000285
810018716
810018758
810018607
810022956
810022946
810022928
810004379
111022174
111000483
810004915
111015688
111000207
111002863
111021196
111020921
111021019
810020556
111009629
111009671
111021121
111021025
111012499
111015787
111003584
810022610
111011695
111023925
111011404
111011467
810004267
810012116
810012113
810012112
111012379
111016483
111019535
111019542
111019527
111015699
111000336
111003581
810004272
810010555
810010530
810010632
810009452
810009459
810009298
810009466
810009450
810009511
111009295
111005869
111003881
111008851
111000203
111004028
810004428
111000624
810014141
810014003
810020923
111015767
111012406
111011478
111015415
810008355
810013744
111011500
111003571
810013756
810013758
111002684
810011856
810011685
111001722
810008717
111011473
810009245
810024144
810024151
810013874
111011851
111000618
111015420
810013772
111010390
810004549
810013776
810022797
810022795
111000193
810014162
810024014
111015791
810019970
810022959
111003988
810013902
810023087
810016179
810015701
810016181
810016299
810016290
810013675
810023093
810023078
810011695
810011712
810011821
810012094
810011700
810011692
810023072
810023065
111041427
111041453
810015640
810015636
810015617
810015626
810015610
810008234
810008335
810008332
810008336
810008249
810008746
810008745
810008725
810008231
810008378
810008238
810008244
810012910
810017491
810005669
810005665
810005588
111011022
810009527
810011805
810012891
111002904
111041462
111031822
111010607
111035651
111002339
111040529
111022493
111001792
111002321
111003975
111041428
111031814
111000205
111000128
111009327
111036825
111003315
111021507
111021612
111003706
111040524
111015427
111004158
111001820
111001682
111003913
111010583
111000333
111000130
111001177
111001413
111008887
111008861
111035643
111020327
111041457
111040527
111031823
111035641
111035655
111004025
111001697
111003986
111001458
111031810
111041435
111011123
111041447
111008553
111035648
111021108
111040523
111012113
111002100
111001470
111041437
111040532
111035656
111020915
111035649
111008575
111040542
111040537
111040533
111016510
111015651
111004029
111002097
111009299
111015358
111019525
111000131
111041424
111040519
111031815
111001847
111001836
111035645
111042831
111042835
111001754
111001790
111001170
111001156
111040534
111040543
810004203
111020935
111022230
810015573
810022927
111002928
111010203
810013661
810011937
111036806
111036802
111008854
111001823
111000911
111041450
111003991
111036809
111036812
111001876
111001735
111020924
111001446
111002935
111035654
111040528
111001161
111035640
111001811
111015036
111040383
111010382
111008911
111008848
111001756
111035653
111002475
111010080
111035637
111021197
111015386
111002106
810009781
810010466
111021189
810005140
810016947
810020400
810015579
810011660
810011786
810014008
810026674
810024113
111021100
111021119
810004028
111021386
111021522
810003625
810020733
810004491
810016733
810020369
111022215
111016500
111016620
111024154
810012885
111002227
111011505
810011659
810013658
810017506
810005057
111040292
810012921
810015293
111001850
111004118
111031818
111008916
111001683
810018813
810020735
810016509
111022211
111012354
810004277
810024954
111011397
810018819
111042838
111011465
111011511
810006057
810011899
111012142
111042836
810020421
810011784
111022209
810009846
810006079
810024973
810024952
810009925
810024139
111003799
810003403
810004184
111021093
810004995
111010636
810004939
111003893
111024145
810003399
111020266
111021521
810003931
111015411
111009686
111001771
111001727
111001734
111004121
111004132
111011480
111002342
111010396
111002096
111002575
111011314
111015865
111004376
810009247
810009241
111012144
111001442
111003575
111004146
111008894
111020251
810004279
810004188
111015419
111015659
111015380
111015661
111015389
810012945
810016701
810005472
810005450
810005458
810005454
810005459
810012936
111003878
810015481
111010388
810004895
111020351
810011796
111011854
111010610
810003963
111011122
810026644
810026401
111011690
810003485
810015353
810004022
111021198
111020347
810012797
111010250
111011517
111012146
111011302
810026629
810012820
810027624
810006007
810018744
111011876
810011930
111004141
111012560
111022212
111021023
111004445
810024279
810013956
810020509
111021175
111011858
810020182
810011791
810013587
111041419
810027630
810009555
810012789
810013727
810009641
111005851
810006369
810005657
810005144
810005923
810027632
111002105
111008843
111020373
111020352
111020330
111011132
810013906
810005650
810004490
810013747
810004016
810024947
810019956
810008326
810020409
111014618
810013586
111011284
810014103
111010630
810013966
111011334
111012362
111015707
810006162
810015433
810015435
810015703
810013556
111012415
810022563
111041452
810026653
111021405
810010494
111003753
810011735
810010626
111010409
111003726
810010463
810026410
810004890
810004936
810024124
810020433
810010435
810015496
810015499
810016720
810006043
111012167
111002782
810020527
111017266
810015350
111036803
111011502
111001618
111001840
810004774
111001953
111000279
111000330
111001376
111001842
111000278
111000314
111015716
111001395
111001461
111001803
111001829
111003760
111010210
111011317
111011399
111003604
111010605
810003391
810013670
810015604
810008744
810013676
111001691
111020186
810004217
111001464
810015321
810011576
111001848
810017515
810012804
810003740
810016514
810024970
111011470
111003580
111005792
111011014
111015670
810011569
111022240
111022235
810019972
810018835
111015431
111015361
111002512
810014105
810011815
810004196
810009287
810004374
111021094
111003817
810013657
111002517
111003732
111004372
810027308
810008738
810004216
111020368
810015717
111012098
111011005
810016875
810011822
111002572
810022589
111003808
111003758
111003897
111015025
111011878
810008751
810020146
810018600
111010051
810018591
810024288
810022953
810020915
810013612
111001169
111021210
810026388
111015721
810003760
810003737
810004223
111024142
810015616
810014160
810018792
810020732
810013889
810026622
810020153
810026530
810006055
810024274
810011725
810020730
810026599
810027273
810027327
111004340
810003413
810008328
111024131
111010079
111008860
111017269
111012538
810013603
810011837
810013582
810026351
111004459
810027637
810027644
810009232
810009269
810009826
810009585
810009613
810009571
810009639
810009576
810023105
810010666
810009223
810009818
810010423
810010492
810010552
810009936
810009942
810009757
810009847
810009848
810024298
810023586
810009929
810027318
810022907
111004454
111004450
111011117
111004154
111002580
111004092
111009285
111005809
111020376
111008890
111036816
111000312
111001475
111001686
111001778
111015404
111002345
111003705
111004351
111009344
111002233
111002509
111001796
111001362
111002479
810013835
810008256
810013848
111021608
111011299
810020179
810027285
111001709
810027562
810009819
810004023
111000731
111000276
111000732
111000730
810013724
810011919
810011885
810012849
810011876
810009227
810010633
810013787
810004091
810011756
810006031
810013890
810012756
810018734
810010614
810009534
111015769
111005854
111021120
111015387
111011506
810006093
810020158
810027326
810023574
111003996
810003569
111023924
810010631
111012370
810011775
810019953
810010465
111001873
111024127
810013733
810020359
810024100
810023999
810026647
810026637
810026627
810026624
810027622
810010536
810009515
810009485
810013741
810010488
810006021
810005362
810022940
810015355
810013753
810013766
111004113
111010355
111001175
810012847
111011869
810020508
810020916
810011738
111004126
111004093
810015582
111024277
810018765
810020233
810015337
810020414
810027558
810016696
810016591
810016718
810016685
810016703
810016602
810016719
810016688
810016869
810015726
810015720
111008863
810010431
810020249
810020247
810020390
111021113
111041441
810020364
111008558
111002958
810014179
111003818
810022569
810022776
810012801
810016520
810008358
111012097
810016503
111000888
810013645
111004365
810024145
810010452
810015719
810024135
111035658
810024167
111041133
810005301
810004794
810022919
111031817
810006358
111009687
111010094
111012112
810011850
111003888
810011940
810013960
810014159
810013732
810013887
810005360
111001780
810014128
111003976
111009680
810004281
810004276
111031825
111009284
111001828
111002373
111001369
111004341
111036814
111021106
111021097
111009307
810026009
111011144
111010999
111003707
111002369
111015764
111020350
111021195
810004094
111002375
810023221
810023111
810005053
810006357
810009894
810009493
810009563
810009630
810009581
810008348
810020165
111011691
810016583
810016712
810017014
810015625
810011642
810010434
810013992
810013880
810013629
111002372
111010416
810011677
810012890
810004991
111036815
810027305
810027328
810027332
810027336
810013592
810008736
111012384
111015354
111020340
810005958
810004186
810010635
111021620
111010087
810006075
111022489
810011905
810012938
111021099
810022923
810022917
111011486
111002954
111009304
111010603
111021213
111005798
111011048
810011872
111010643
111002237
810009614
810012887
810016711
111011484
111010085
111010084
111004380
111011274
111003605
810016883
111024128
111001835
810005525
810005533
810005538
810005512
810005609
810005652
810005620
810005661
810024957
810005627
810010640
810026528
810015638
111009683
810018808
810010659
810011718
810015706
810011696
810009536
810018586
111022243
111015708
810004985
810017505
810015432
810006198
810012108
810012694
810004394
111009336
810020410
810011932
111009657
111010229
810004087
810004199
111021504
810005982
111015366
810011883
810023125
810004989
810011886
810005586
111004129
810027331
810024266
111003598
111002364
111002952
810023235
810022577
810012927
810011917
810023053
810023584
111003969
810011669
810014114
810009616
111010621
111003807
810013804
111003576
111003970
111003984
810012819
810008252
111003750
810006410
111003588
111005805
111004374
111011464
111011886
111010059
810005956
111002920
111014612
111003295
810018755
111005800
111010402
810015512
111015785
810011721
810008352
111001926
810010583
810012766
810016600
810011587
810018784
810017528
810003967
810004900
810004898
810003759
810003928
111041460
810022723
810020152
810020340
810016286
810016978
810015517
810020533
810015645
810011667
810012802
810011878
810013647
810013581
810011903
810006011
810013584
810010520
810010509
810010593
810009277
810008364
111021501
111024134
810004892
111021401
111011489
810004776
111020378
111020343
810026639
810008719
810008243
810010428
810005997
810005607
810006413
111021118
810013736
810013637
810012099
111011097
810020523
810018791
111012567
111012400
111004007
111012552
111003877
111011079
111011042
810015569
810015279
810010447
111011320
111011848
111011894
810003480
810016595
810005455
111019506
111022172
810011760
810011743
810015292
810013913
810027279
810027271
810013765
111009641
111009640
810006066
810006037
810006067
810015494
810004500
111022735
810027322
810004482
810004430
810006426
810006421
810006420
810006376
810006383
810027299
810027267
810027171
810020544
810020494
810015299
810015729
810014143
111002248
810020532
810013589
810013609
810013595
810027321
810015502
810015589
810020411
810016933
810020728
810016922
111021609
111037899
111008557
810006362
111021611
810003629
810027274
810020366
111015700
111002131
111004110
810004512
810004515
810014113
810008257
111008906
810003564
810015489
810006087
810026673
111003972
810009463
810027263
810018739
111015712
810008753
810027302
810009249
111011491
111002756
111002568
810003489
810025099
810027291
810013858
810015352
810024268
810024002
111042859
810023601
810020368
810024300
810006169
810020566
810013560
810004503
810024328
810028857
810018707
111021109
810006136
810011828
810020243
111010083
810020258
810009642
810009650
810009618
810009655
810009553
810020370
810020189
810009602
810020367
810020264
810020405
810011577
810011891
111010057
810020402
810012119
810011830
111021511
810009488
111004011
810023090
111011141
810011873
810011751
810009627
810012950
111021028
810013672
810013751
810008759
810008255
111019508
810005921
810005954
810017006
111024279
810026415
810010641
810010493
810013985
810008236
810004017
810020227
810020176
810020167
111010574
111010602
111012381
111012364
810019980
810019983
810026397
111002313
111001468
111002240
111001839
111015715
810008245
810005597
810006026
111022201
810018760
810018840
810011875
810023095
810024127
111040525
810027556
111011045
111020314
111015406
810004437
111015780
111020356
111011031
810013564
810018842
810004914
810022570
810016930
810016989
810024130
810024141
810022604
810017526
810016955
810017011
810027294
810016581
810016511
810016592
810023575
810016939
810026373
111012114
810015607
111002336
111004142
810016912
810003471
111019544
810024330
111001872
810003406
111020269
810026378
810026421
810005509
810004920
810013621
810020492
810022962
810009442
810017008
810018559
111019541
810009231
810022938
111012389
111011040
111009635
111011126
111020936
810004918
810003481
111011027
810013627
111011472
111012115
810018820
111035647
111012526
111035642
111004449
111002126
111000329
111000120
111010364
111002376
810016871
111021499
810004288
111017260
111015023
810023250
810023107
810011590
111003582
111004098
111003725
111003603
111003572
111003600
111024278
111024276
111024280
810013594
810013579
810013590
810012839
810006046
810012798
810012791
810012816
111003595
111003602
810012810
810012808
810012818
111021613
111001773
111003813
810015642
810015620
810016185
111001733
111003880
111003879
111003809
111010363
111010228
810008357
810008242
810008350
810008339
810005993
810006375
111010256
810010518
111010425
111011513
111003569
111003583
111010249
810010660
810010597
111021390
810006091
810006126
810006100
111000937
111012150
810016713
810017530
810014163
810014142
810014153
810014178
111010635
810018830
810018710
810018724
111004156
810016183
111002861
111003591
111021091
810018749
111021394
810016979
810016885
810016896
810016943
810013607
810013624
810014155
810014145
810014107
810013773
810014130
111003721
111003610
111000855
111001374
111001732
810006140
810006095
810009918
810013767
810006131
810006127
810006089
810019951
810019961
810020173
810020187
810020228
111021636
111002470
111004441
111000881
810005052
111004087
111002625
810009766
111020172
810020759
810011693
810009586
111010378
810028862
111002781
111022199
810015228
111042712
810014093
810013813
810018706
810014026
810004550
111002887
111011649
111010259
810022937
111010251
111010247
810018800
111015026
111020312
111002569
810011655
810013740
810015635
111003714
810024126
111011024
111012134
810004283
111014617
810004146
810003315
111022184
111015403
810011704
810022961
810027275
111003594
810006195
111002506
111003739
111004103
111011325
810004577
810027301
810011746
810020918
810018796
810022895
810013896
810009653
810022582
810005506
111012137
810003898
810011797
810009939
810003491
111009684
810011623
111003709
111020276
111011133
810004485
111011028
111010091
111012154
111015706
810013759
111003918
810020361
810020749
810005643
810005594
111041467
810024149
111010062
810013805
111001704
810011812
810004851
111001688
111011142
810022886
111004095
111012380
111012351
111011109
810010500
810006374
111040297
111040377
810015334
810012685
810015344
810006197
810009479
810010601
111011072
810013570
810012963
810015501
810015497
111022255
111022203
111022228
111022233
810008754
810008731
810003531
810003533
810009910
810009933
810006101
810006139
810015322
810015471
111042701
111042711
810020757
810020924
810016177
810015998
810020425
810020547
810022735
810016887
810016959
810015323
810024926
810023102
810016724
111010410
111011147
111011025
810011779
111015402
810013778
810015363
810004913
810003568
111024140
111024124
810015281
810015331
810022960
810018580
111012378
810005451
111011519
111002923
810005449
810008334
111001136
111015653
111040376
111022158
111006458
810023246
810023130
810023080
810023081
810013807
810013806
810013797
810013818
810013769
810013777
810023217
810015290
810015256
810017534
810005646
810005460
111035652
111036807
111004000
111011124
810023582
810024164
111003998
111031820
810020262
810009605
810024976
810024278
810020504
810009626
810016681
810015328
111000389
111000234
810013780
810024945
810020513
810014025
810013762
810013826
111040530
111040538
111036810
111036819
810022731
810022744
810022750
810022609
810022612
810020921
810020529
111003615
810016006
810013938
810013799
810013823
810013728
810016937
810016987
810016941
810016949
810016957
810006390
810006103
810005922
111010421
111011002
111041421
810008343
810014035
111041454
810015590
810014030
810010591
810011835
810005361
810013567
111003990
810012916
810016940
810015339
111041458
810015482
810016517
111041422
810010564
810023064
111040380
810009546
111041455
810005664
111040536
810022748
810013989
810015282
111040379
810027320
810020450
810020726
111016546
111011689
111001403
111011282
111011520
111011030
810006423
111002124
111010579
111001921
111011146
111009316
810011808
111004355
111002338
810008363
810015347
810012932
111003590
111003587
111042842
111042828
810011896
810013919
111001609
810020347
810020746
810011868
810020429
111011340
810020341
810020403
810020190
111001375
111001367
111002111
810014119
810014006
810004362
810004520
810004499
111001705
111041448
810010448
810010441
810010453
111036811
111036805
111036801
111036813
810008241
810008385
810008337
810008361
111040284
111040298
111040531
810001601
111039239
810008955
810019662
111026044
810023163
111035612
111027356
111041398
111014761
111040851
810008978
111038800
111025213
111016600
111016468
810022186
111037185
810002283
111038982
111005770
111031607
111029577
111039122
111008424
111013223
111007763
810025356
111006875
111039415
111033266
111034790
111034950
111034697
111016166
111005959
111013179
111033019
111024774
111041369
111024256
111006883
810000120
111018352
810026278
111013312
111036447
111032894
111035946
111012919
810009185
111018345
111025206
111006027
111019070
111032915
111034730
111025899
111027640
111018157
111014336
111026532
111006476
111014452
111027485
111022472
111016968
810007085
111037153
111037873
111032945
810028281
111037356
111025984
111025271
111037730
111023617
111037137
111029243
111023757
111037642
111029201
810007137
111023632
111026766
111034005
810007109
111038949
111027314
111039071
111027638
111022448
111027151
111042804
810001871
111030096
111041059
111034151
111006413
810008693
111027303
111023756
111034762
111024493
111033973
111038750
111035799
111034669
111037042
111034155
111036879
111029374
111034749
111035866
111034704
810024756
810025831
111043141
810027356
111024104
111024228
111039020
810000340
111037704
111024462
111029245
111040823
111030310
111036127
111012766
111013218
111037813
810027362
810000102
810020084
111043142
111033107
111029308
111039425
111025286
111037039
111016986
810008975
111034376
111029357
111007753
111007745
111033067
111034216
810008776
810026299
111030418
111027647
111039677
111027477
111030422
111032988
111033142
810024215
810026143
111042673
111039584
111017799
111019069
111031590
111031681
810024189
111035962
111029415
111014322
111022876
810008792
810022983
810022811
810026188
810026061
111034088
111005614
111016325
111008532
810008852
111029600
111034543
810009000
111019227
111034236
111034013
111025160
111013372
111031587
111031561
111037200
111034270
810007632
810023028
111031633
111028520
111014546
111025903
810026431
810027404
111019123
111039389
111016375
111031583
111039076
810007005
810007070
810023014
810022714
810026155
111040573
810020295
810028609
111007256
810008817
111019054
111037739
810026080
810007612
111037610
810026108
810008180
810028286
111016458
111037697
111039499
810000209
810001769
810023310
810028309
111030030
111040493
111037692
111030078
810007306
111041200
111037371
111025185
810001369
111014535
810026324
810000978
111027322
111034166
111034129
111028632
111035606
111036188
111038582
111034107
111029561
111029309
111029516
810028317
810026288
111037562
111016045
111036003
111036054
111025362
810026282
111039162
111028657
810026082
111028589
111034718
111027503
111042670
111033272
810025333
111032964
111040480
810006112
111040887
810000353
810006348
111041395
111027354
111025804
111033211
111025926
111014419
810002585
111042719
111032062
810001057
810005971
111025842
111039297
111039477
810025222
111034390
111040731
810002649
810002652
111025925
111014475
111025830
810025178
111025836
111033317
111014455
810025207
111014460
810002716
810025181
810007187
111027515
810002579
111032889
111025784
111025939
111026030
111025833
810002597
810022854
111025922
111005364
111014316
111024967
111033332
111037637
111039482
810022884
111025823
111027158
111027174
111035607
111027199
810020290
810000342
810000472
111012922
810000348
111036027
111038743
111023208
111027497
810007329
111023584
111029347
111014744
111032930
111040513
111034525
810002249
111034078
810001711
111039241
111037176
810000206
111014815
810027372
810025769
111031578
111036857
111041056
111016146
111006013
810000198
810024212
111040344
810001818
810024191
111014489
111023147
111035949
111034655
111024242
111013196
810000454
111029753
810019713
111029705
111029755
111019207
111016373
810025821
111037624
111029650
111031719
111023296
810006342
111037192
111027448
111016331
810008271
111020704
111029568
111024264
111020681
111016129
111008544
111023593
111040501
111019058
111037724
111014531
111006436
111027339
111025979
810022865
111037101
810022182
111037593
111029350
111008539
111020404
111030692
111024117
111034732
111006475
810007301
111024799
111019629
111006272
111034646
111005260
810000370
111038683
810026157
111007774
111008405
111043190
810020298
810025314
810019814
810002135
111034575
111042131
111042765
111039077
810022669
810026295
810020280
111042128
111024904
111040332
810026242
111037157
111034488
111042664
111026654
111037896
111042113
111041351
111034164
111034185
111039277
111030230
111030390
111034417
111019135
810007339
111034709
810026460
810007264
111039264
111039244
111025999
111039289
111037805
111016089
111023294
111017719
810020323
111027202
111037789
111016040
111015991
111014522
111013397
111017729
111035905
810020194
111034448
810000919
111034357
111036844
810007226
810023164
111036889
111014166
111020387
810020328
111014094
810002651
111014986
111017797
810006110
111035904
810006114
111017708
810022693
111026776
810022648
111037890
111023231
111005265
810020926
111025992
111005322
111034571
111034528
810000869
111036212
111020478
810022707
810020028
111038589
810028303
111039621
111036020
810003100
111006888
111036150
111027153
111032885
111036181
111031797
111017825
810007150
111019162
111019163
111014726
111043138
810016811
111042783
111037881
111035814
111042749
810028606
111034562
111037876
111034747
810009020
111029188
111017780
810018861
111023174
111019128
111040750
111029599
111029244
111014342
810025194
810007613
111037636
111034197
111025356
111014474
111034196
111025162
810016801
111016167
111027282
810019807
111034578
111039506
111036022
111023186
111022899
111040555
111020380
111022891
810025859
111038622
111025289
810002590
810018858
810019653
111034368
111039483
810026323
111025018
810023020
810025830
111019208
111027427
810020047
111024100
111035790
111036018
111043205
111025229
111022797
810007041
810001561
111025809
111030088
810022864
111027323
111027149
111032908
810019132
111035893
111020445
810001443
111030275
111012778
111039453
111005937
111034140
111006891
111035868
810026472
111034154
111019063
810008313
111020696
111024103
810002889
111043260
111025280
810003260
111029208
810025486
111036839
111041374
111038572
111022474
111026056
810025802
111036016
111016341
810003266
111035991
111030250
111005970
111039642
111034500
810002130
810026177
111034303
111034327
111038994
111038950
111030314
111038857
111029229
111035855
111026019
111035732
810000364
111031924
111023306
111024246
111037590
810000159
810006437
111008398
111039088
111023188
111034721
810006434
111024243
111027243
810001820
111023695
111008483
111040453
111033384
111033446
111026013
111033990
111016606
810027444
111040340
111040331
111035629
810007098
111026011
111007761
810007335
111037558
111039151
111015351
111017137
111016091
810019090
810023161
111014554
810026345
111005944
111034804
111022630
111006052
111019048
111038925
111033302
111005932
111038595
810007321
111037384
111036006
111034533
111036004
111041128
111034470
111027188
111034531
111031615
111042781
111031995
111039079
111026087
810007334
810007310
810022167
111039131
111025998
111039198
111030146
810024027
111037862
111037882
111037760
111038903
111037770
111037877
111039118
111037846
111042222
111042078
111042050
111042112
111031934
111030103
111042133
810024086
111031918
111039144
111030085
111039098
810024058
111042246
111031910
111025887
111042144
111013371
111039276
111025723
111042085
810008308
111042223
111039236
111039263
111039102
111022898
111039219
111039221
111038740
111023197
111022619
810022975
111015296
810009145
111030658
111032890
111009100
111012896
111016564
111033390
111029332
111008410
111036261
111041115
111039288
111029471
111013316
111029382
111015349
111023241
111015312
111023598
111025920
111035709
111037381
111037029
810002393
111043167
111036074
111027361
111038905
111036029
810025388
111028656
111019200
111028598
810000895
810002338
111016473
111029359
810026065
111027504
111041327
810023147
111023580
111024097
111016033
810002343
111024282
111031859
111031941
111032900
111031860
111032080
111019186
111014098
810003185
810008866
810008771
810003192
810003271
810008708
111022470
111005774
111038824
111033967
111035844
810002892
111030591
810000088
111028599
111031588
810008284
111031847
111025867
111028549
111031649
810007153
111028628
111029340
810008208
111030106
810007229
111025797
111031645
810002699
111006672
111035837
111034356
111025783
111014801
111014745
111040329
111015352
111038830
111034480
111012740
111029435
111025014
111004974
810001805
111027459
111016073
111026016
111029436
111023362
111036050
810009006
111036446
111026626
111030286
810027367
111040270
810022849
111028550
111031884
810025235
810025322
111029370
111028433
111031647
111028629
810025343
810025819
111029303
111036170
810025823
111039193
810025762
810025796
810025844
111039250
810006470
111012946
111014772
111013351
810000119
111014961
111016540
111017130
810008962
111029752
810000572
810002493
111015982
111019910
810023206
111037169
111006674
111030086
111035999
111036065
111020414
810006117
111024928
111039612
810001874
111012953
810023327
111027179
111037368
111016360
111036715
111020482
111027313
810018902
111036862
111029343
810009018
111035920
111028420
810007120
111009115
111027506
111024937
111024457
111031767
111034485
810020116
111037727
810017110
111033303
111038879
111042070
111037080
111038656
810026485
111032063
111039136
810003120
111029666
810028598
111035805
111023175
111006067
111041198
810026094
111035810
111040813
111039456
810002586
111037104
111036165
810026073
111034662
111034211
111034311
111040497
810022968
111034573
111034441
111024122
810000704
111038804
111034638
810000393
111042716
111040932
111014501
810000718
111025900
810002712
111024250
111016996
810000566
111014971
111013350
111008437
111026069
111006877
111039589
111026007
111025222
111038969
111026032
111017733
111039570
111006846
810008794
111027502
111006432
111027373
111014508
111017114
111030110
810018899
810024515
111034171
111020476
810000217
111014479
810000338
111015344
810001828
111025889
111037758
111033018
111016095
111015946
111008510
111014524
111038575
111015318
810020196
111039610
111038902
810020308
111034458
111027431
111030039
111025881
111035998
111025182
111037079
111031943
111042059
810002723
111035759
111029256
111025194
810025400
810007148
111042737
810008206
111029223
111006806
111032956
111014109
810001675
111033439
111024941
111033296
111040364
111023689
111034754
111038826
111030419
111030317
111034016
111034521
111027462
111027395
810019904
111026059
810007331
111039672
111028647
111020472
111017706
111037575
810007078
111035995
111037370
810007297
111023165
111033305
111030659
111040707
111028441
111031756
810018870
111027159
810027401
810003091
810019087
111036082
111032899
111042233
810027446
810022649
810020203
111016459
111040498
111038855
111041330
111031620
111031581
111033059
810003200
810009195
810025244
111037662
810006343
111016153
111030285
810026241
111009080
111013222
111024795
111040700
111023298
810024423
111030405
111022808
111005618
111036038
111027261
111031753
810002717
111019224
111030684
111033352
810008849
810024064
810025818
810023144
810027339
111040505
111025171
111027334
111039337
111029468
810026524
810028273
111027435
111028558
111016595
111038684
111039419
111014341
111014525
111016948
111027471
810001604
810024183
111016542
111005616
111032892
111032004
111032007
111027285
111031603
111037726
111038749
810000714
111005551
111034806
111034681
111006253
111037055
810023153
810024210
810026284
111038671
111042743
111037656
810001055
810025173
810000207
111019073
111038953
111025297
111031799
810007597
810018676
810018852
810026055
810026051
810026238
810026149
810018770
810018772
810025249
810008189
111032078
111015956
111005790
111039245
111005771
111029202
810018685
111008484
810000202
111040358
111022456
111040426
111033010
111034277
111030151
111040504
810027347
111027342
111033279
111006287
111019196
111017138
111040456
810022828
111030351
111035956
810026308
111037530
111031897
111015341
111040435
111032094
111030035
810009186
111031839
111031989
810008984
810009003
111038644
810025309
810000373
111016950
810003118
111031861
111037065
111029220
111006273
111034225
111034141
111026680
810002115
111025146
810002700
111037714
111034396
111035600
111030269
111034330
111037197
111040698
111034233
111036901
111039318
810000469
111039598
810001136
111031699
810025203
810001355
111030679
111034803
810009019
111031903
810008979
810027398
111029681
111041093
111039214
810026203
111042102
810027338
111043248
111029261
111039623
810000810
111019217
111022806
111029420
111035615
111033290
810006945
810001563
810002125
111027268
810008317
111033053
810024056
111019900
810002461
111029403
111030276
111036109
810022665
111033025
111040437
111022624
111042063
111033108
111033380
810000401
111036896
111040741
111019152
111006438
111022897
810022817
810000550
111024109
810002397
111006871
111016550
810019878
111026487
111023669
111031723
111029479
111013189
111031611
111015959
111033045
810022819
111016020
111024933
111020435
111020397
111033090
111008469
111032099
111023603
111005001
111039540
111033212
111033346
111032120
111028615
111033347
810025865
111016607
111014360
111006815
111043199
111041103
111023592
111033084
111022458
810001143
111006039
111007766
111038605
111039583
810025345
111042633
111012917
810008324
810018886
111031609
810006334
111039340
810026074
111036032
111007258
111037621
111030157
111029363
111022622
111022610
111026657
111024801
111030579
111039594
111016009
111034442
111024772
111014992
111012954
810020195
111037452
111033003
111028469
111039335
111038585
111035883
111025977
111043235
810023328
810002281
111038561
810017162
111027439
111039470
810005871
111040754
111040311
810019891
810001892
111035873
111023213
111034705
111016979
111022477
810001657
111035952
810001606
111030245
810020099
111016450
111039533
111042134
810023544
111034653
111040489
810000087
810019705
111033197
111027217
810002119
111041197
111017136
810007002
111025004
111030120
111030595
111015975
111027437
111041326
810001889
111030155
111006425
111026529
111034554
111031674
111023269
111023151
810002190
111015306
810002123
111020692
810025335
810027340
111037186
810007163
810026466
111034194
111029323
810008178
810002112
111035698
810024070
111006878
111012789
111030584
111032091
111026513
111027226
111039656
111017142
810007082
111037566
111025810
111023696
111023685
111030107
810006450
111039647
111020426
111039044
810001831
111017713
810007136
111016469
111006416
810001365
810001110
111020382
111012963
111030128
111028644
111029369
111036999
111037547
111027312
111030716
810001851
111030313
810001749
111043191
111039622
111037087
111042117
810019831
111030057
111037719
111037205
810024071
111041108
810019870
111015960
111032931
810002672
810026506
810026505
111013168
810027345
111038973
111027309
810020311
111040805
111042785
111037612
810000529
111034371
111016541
810020325
111028674
111027363
810002909
810025807
111006470
111032901
111042083
111040474
810019836
111016001
111025134
810023043
810026179
111039510
810020108
810019818
810022979
810022216
111037585
111037372
810022821
810001489
810001171
111039016
111031706
810008702
111034351
111033048
111040727
111036149
111031865
111025216
111017116
111042111
111014766
810022981
810007087
111042068
111040928
111009119
810023543
810006989
810001138
111005016
111030319
810000388
111007750
810019704
810026484
111038843
111025872
810001436
810005973
111034150
810028496
111016008
111040465
111024894
810008183
111040721
810002750
111023159
111020383
810028288
810028316
111024794
111007752
111039204
111004982
810025858
111031597
111005352
111014422
111027644
111038996
111030091
111005010
111023152
111016537
111031981
111040347
111024983
111039159
111024751
111015957
810022683
810002593
111006823
810023526
111014762
810001374
111023636
111024951
111027374
810007291
810001401
111019905
111025291
111025711
810026151
810017153
111032905
111026489
810026500
111018359
111006251
111025981
111025970
111024733
111014110
810001079
111027376
111017133
810000337
111023281
810024202
810023259
111034560
810024529
810024527
111034326
810008213
111025929
810007627
111039542
111039555
111007755
111033282
810026174
810026195
810001872
111024790
810017106
111024767
810023025
111016984
111025987
111024773
111035986
111024240
111016975
810022196
111030122
810028293
810008220
111028455
111037534
111024813
111015343
111039104
111035708
111025813
810024214
111019147
810001613
111029383
111029565
111029645
810001367
111036042
810018867
111026674
810018884
810000092
810001144
111028442
111039366
810008664
111038992
111014733
111024976
111029788
810000443
810006337
111028531
111040810
810017179
810003196
111013354
111014324
111015986
111016006
810000410
810000090
111016163
111016369
810003191
111025716
111023691
111023784
810000442
810000426
810001487
810001078
810001497
810001486
111013388
111025718
111013381
111013356
810003208
111014367
810001380
810000085
810000099
810003250
810020942
111042724
111027154
111027150
111040734
111023774
810009148
111037797
810008941
111033285
111033367
810006340
111028660
111034338
111027240
810020941
111034297
111031713
810008272
111042069
111027350
810025361
111040801
111040316
111035766
111035783
111029179
810017173
810025399
111034379
111042041
111040464
111034382
111028584
111031789
111023716
810001373
111015993
111015976
111036988
111035797
810007019
111040676
111029797
810024351
111037188
810002726
111039397
111032009
111035945
111025822
111032041
111031917
810019820
111028438
810025498
111013379
111038742
810023143
810022835
111020442
111034738
111006870
111030042
111039551
111006821
111033975
111040546
810025487
111027346
111031905
111024108
810002728
810019808
111026027
111042244
111028652
111017813
810026085
111034627
111025938
111026507
111038704
111027315
111014439
111014492
111019074
111019044
111028591
111035936
111033276
111039033
111033181
810026292
810019706
111038708
810020019
111028637
810023133
111014551
111035608
111043163
111024251
810019126
111034650
810003110
111019201
111035984
111038570
111037215
111029456
111027274
111041060
111019199
111027383
111005924
111012966
111031956
111039019
111034319
111025009
111025261
810022217
810002492
111025846
111034339
810006461
111031662
111031731
810002488
810025180
810022219
111037701
111039051
111029355
810002449
111035978
810018906
810018880
111029476
111037804
111035819
111024460
111034636
111037041
111029432
111033417
810025232
111034810
111030367
111037120
111033956
111037366
111030358
810006482
111030144
810002465
810002466
810002456
111037810
111035924
111031866
111030409
111025200
810009200
111025033
810018878
111038730
111039465
111025150
111039404
111034432
810008224
810006478
111024992
810008819
810022218
111029463
111039142
111024752
810008870
111039347
111043201
810028478
111034515
810026042
111038825
111031639
111029225
111027456
810008851
111012795
810019884
111035922
111016588
111009104
111037165
111008502
111034189
810007170
810006295
810023292
810008960
111039110
111037629
810026286
111027163
111020456
111040756
111037352
111030231
810001712
111031968
810008818
111016361
111023212
111025897
111038698
111037132
111039217
111029199
111029518
111033031
111040793
111030134
810007212
111030590
111039167
810020015
810000334
810024254
810022718
810006285
111026079
111039493
111037362
810025382
111039330
810025357
111035786
111036877
810000387
111036977
810000568
111036120
111023706
111035940
111006069
111030020
111034580
111032985
111038755
810001137
111029453
111020401
810001863
111018356
111030674
111030672
111027319
810003109
111034193
111023265
810001383
111031745
810022991
111042104
111027528
111027481
810002919
111016349
810027403
111014740
111005961
111018353
111036855
810024257
111006840
111014469
111031804
810002820
111039211
810025213
111033402
111028634
810008705
111035831
111031748
111031677
111037650
111030381
111037088
810020006
111036272
111035875
111034494
810001842
810003151
111017796
111017749
111038877
111040859
111042729
111020687
810007162
111029394
111034394
810022877
111013332
810006952
810002931
810007317
111043239
111019093
810009014
111029423
810002880
111026764
810026326
111036026
111039421
810020949
111039473
111039010
111038780
810025482
810019647
111015945
111036024
111031737
111039522
810025765
111035951
111028594
111013394
111028610
810008666
111008536
810020083
810025808
810007315
810025761
111025888
111018148
111037541
111025026
111034549
111038689
111041052
111006664
111036056
111029662
810002691
810025327
111028595
111031798
111039027
111013352
810025827
810008290
111031577
111031969
111032957
810007261
810009154
810023032
111023798
810026285
111039649
111039181
810026062
810026314
111040429
111029381
810007634
810026522
111039691
111029305
111029451
111034077
111029535
111031837
111029294
111032020
111029459
111014093
810019701
810000490
111034300
111038939
111034220
111037799
810001394
111020703
111024476
111030149
111033961
111023787
111020418
111015353
111033049
111006021
111024252
111035733
111024742
111038762
810024336
810006457
111014099
810026176
111041096
111028619
111025941
111027359
111031679
111017736
810024200
111025906
111017821
111020432
111041119
111037183
111012747
111029287
111037053
111029282
111029615
111029631
810026076
111041131
111028445
810024190
810006992
111031772
111041110
111031621
810025833
810025344
810025366
810025191
810008202
111025730
111039315
111028587
111029222
810025766
810008218
810008297
810025179
810025764
810008312
111029252
111029231
111030724
810025499
111033366
810006462
810000512
810007608
111016317
810028292
111024237
111004980
810001759
111019239
810001788
810024345
111036154
111019166
810026018
111038548
111034503
810027405
111037017
111025008
810008321
111037623
810025177
111029649
111037594
810027448
810022880
111032019
111036081
810025476
111026531
111029643
111015310
111043174
111024236
111020682
111024260
111036206
111036093
810024755
111006820
111025880
810020112
111041121
111020470
111020430
111039538
111037109
810008687
111037620
111022633
111006025
810022976
111029688
111035813
810016793
810019712
111019143
810028588
810001857
810001064
111014989
810026319
111025188
111040762
111039509
111031950
111014409
111037630
810020274
111012748
111023181
810007630
111036984
810007596
111012772
111008455
111007245
111012774
111030340
111016026
810023333
111039049
111032000
111032134
111040452
111025912
111026633
111034567
111026632
111008476
810009015
111030641
810024258
111043175
810025797
111013204
111006839
111025728
810007095
111033131
111006246
111008506
810020114
111012751
111015317
111013385
111008498
111030671
111025792
810002193
111037216
810024078
810001397
111029550
111024885
111029723
810022702
111033207
111039174
111038976
111008480
111027645
111039435
111029433
111037398
111023748
111027139
111036125
111034176
111034737
810020149
810025832
111031864
111037402
810002182
111029380
111023219
111014784
111006890
111014779
111015988
111006451
810000525
810026201
111030718
810022212
810007324
810007305
111039194
111033431
111036893
111036194
111033412
810008274
810026336
810026079
810022820
111040763
111035967
111031655
111030717
111035689
111034286
111030423
111037203
111036089
111037034
111037008
111037085
810019889
810018874
111042755
111032949
111035913
810007115
111038563
810020285
111037012
111033354
111030729
111029780
111035944
810019813
111034657
111034210
111036036
111022881
111032911
111042651
810009162
111031946
111009084
111034739
111014975
111031629
810007309
810000524
111040850
111034116
111034359
111032076
111024974
111032028
111020398
111037649
111008451
810008182
111016412
111029191
810005965
111031730
810001756
111031617
111028426
111036169
111024816
810020115
111037021
810019656
111040881
111030130
810025176
111029622
111038978
111016947
111022605
111019906
111033062
111007249
810022859
111032939
810025315
111033370
111039037
810019679
810017156
810019699
111039018
111037579
111034126
111036002
111036827
810008698
111036197
111039270
111025189
111029242
810002279
111039032
810002887
111035901
111025161
810002896
810007322
111039671
111041048
111034348
111022467
111026008
111032046
111042776
810023557
111039091
111017822
111017743
810000812
810001492
111039199
111014346
810020334
111027474
111042105
111028524
810001431
810009016
810002255
810001850
111023261
111012950
810000527
810026228
111025157
810022873
111023368
111032918
111033111
111013182
111025350
111028640
111029196
111030119
810006345
810028470
111030046
111013333
810017154
111035870
111040835
810024061
810019642
810002490
810001667
810027391
111034143
111034163
111041049
810000868
111042118
111025969
111037582
111040890
810019819
810007173
111017727
810019815
111016071
111015000
810001441
111019644
111027227
111019234
111042742
111016075
111027178
111029367
111029312
810002731
810000376
810002655
111030399
111019085
111035833
111027509
810007635
111014325
810008230
111027457
111025705
111014331
111027147
111035752
810025330
111039361
111033252
111016966
111034342
810022221
111024938
111023143
111029699
111025364
111006895
111012964
111016440
111018143
111037647
111023356
810000815
111025005
810019129
810019837
810009149
111013155
810026447
111039123
810008196
111039409
111030665
810002903
810008177
111041353
111031933
810026341
111039317
111039383
111025923
111027223
810007617
111034170
111026750
111030023
111040491
111040740
111041335
810026307
111041392
111039224
810007114
111041367
111041321
111040820
810026309
810007228
111039132
111007756
810026171
111030353
111006834
111007743
111009081
810007020
810019109
111014557
111019911
111016604
111038735
111014996
111025921
111004997
111016065
111042768
111016145
810025994
111014424
111029726
111034687
810002637
111023271
810002642
810000564
810002643
111023276
111032040
810007248
810002605
111016162
111036066
111029551
810007119
111014566
810000399
810000589
810000569
111029486
111034514
810000604
111014363
111014494
810008210
111034436
810008690
111038717
111034723
111016584
111031830
111034534
810002648
111036161
810026090
111014719
111014737
111032044
111038783
810007294
810007257
111038751
111038727
111032027
111036014
111014451
111031718
111023291
810026056
111030267
111012776
810001676
111037340
111006434
111024487
810009188
810019846
111034495
810019658
810002403
111022461
810026510
111040439
111040779
111040566
111035711
810025395
111040786
111033118
111035928
810022848
111040516
111036073
111034256
111030271
810026293
111029253
810026273
810026072
810026006
111039427
810026140
111031742
810009193
111017124
111042756
111028659
810007222
111012900
111030097
810020314
111035968
111040814
111034413
111024914
111030397
111037712
810007231
111027529
111038957
111027237
111032952
810024089
111040839
810000679
111019160
111033080
810007235
810001562
111035776
111030639
111019127
111029627
111038579
111024917
111005753
111026739
111016143
111028593
111014332
111028623
111029798
111029545
111030043
810022662
810006291
810006344
111013141
111038628
810019645
111040548
111019084
111006397
111037360
111033072
810027428
111018338
111025893
111029647
111019075
111030256
810023039
111030660
111039179
810028593
111005035
111042052
111029598
810002941
810008996
810017175
111030710
111031734
810018873
111040699
810024050
810000793
111027281
111029570
111039306
111014343
111038715
111016036
111022478
111039497
111041308
111031769
111029557
111028451
111014329
111015339
111029619
111034624
810009158
111023243
111023621
111033318
111029648
111026671
111034748
111014477
810000379
111037149
111015969
111038667
810024201
810025317
111040712
810001873
111027403
111036882
111016012
111019645
111008505
111029400
111034249
111036123
111022874
810016791
111017119
111040889
111034574
111018347
111018350
111005558
111023600
111034672
810001750
111005563
810000391
111023148
810002939
111023601
810006947
111030061
111039208
810002603
111037025
111025712
810028477
111008513
810028483
111029449
111027192
111020451
111008491
111026625
810022870
111034564
111022608
810002572
111034541
111006270
111005931
111006259
810000675
810000683
810000703
810000366
810000210
810000791
810000680
111026018
810002278
111025915
810022986
111022479
111022618
810002575
810023009
810002587
111031567
111033307
111025184
111033360
111031729
111014311
810017163
810024046
111023779
111026498
111036155
810001672
810000405
111023158
111024270
810000199
111016051
111020481
111034605
111030308
111031637
111041138
111038990
810023281
111039536
111022446
111037103
111025821
111036172
111036095
111027305
111039592
810024043
111036863
111026677
111025236
111034373
111023205
111023749
810027354
810026430
810026525
810008187
810026432
111039360
810000817
111024803
810026476
111018168
810008987
111024989
111039376
111039356
111040713
111042225
810000335
111036115
111033950
111033393
111008540
111025193
111027180
810026220
810025204
111042652
111039039
111042627
111040930
111039210
111009120
111034445
111038577
111034516
111025256
810003098
111033116
111033375
111006409
810007178
810023549
810018898
810024055
111033097
111032122
111029430
810006118
111029717
111020708
111028616
111034231
810007277
111030282
111034384
810026159
111014767
810008854
111025968
810007063
111030133
111033248
111015942
111040923
111020392
810002852
111030150
810024262
111005626
111034735
810027364
111008432
810001664
111020474
810008294
810000380
111014337
111043192
111040846
111014366
111023199
111030726
111038815
111040891
111015307
111014550
111039135
111039336
111032010
111039385
111036086
111037633
810023185
111041137
111037023
111033291
111008481
810006458
810025852
111040766
810001771
111031926
111029414
111029354
111014306
810028592
111037451
111006866
111012791
111018209
810002679
810003268
810019108
111018153
810000816
810000706
111024919
111028580
111014465
111004991
111027331
111037351
810020950
111033175
111032045
810002487
111006812
111017761
111037576
111039070
111017778
810000747
111034008
111038586
111029314
810026499
111014727
810022713
111006460
111006471
111013360
810000355
111013147
111013308
111016448
810000218
111016444
810000213
111027275
111035908
810003119
111025348
111035910
111020464
810001843
810000161
111025870
111025972
111025934
111017811
111017770
111014982
111039590
111029503
111031973
810006432
810020129
810028596
111037127
111037141
111029720
111030152
111037133
111033295
111040687
111037849
111033288
111025151
111041101
111032969
111012769
111040826
111040770
810025211
810018783
111015930
810002118
111019050
111013348
111013315
810002906
810001879
111016080
111037338
810020310
111004959
111016092
111025932
111015348
810000156
810020318
111013370
810025491
111022799
111023230
111023386
111013156
111015305
111033400
111013206
111013152
810006483
810026209
111043187
111031670
111034156
111035917
111038613
810026489
111040751
111037099
111037096
111038549
111038593
810022667
810007107
111042640
111014444
111006835
111040854
810000746
111040834
111015337
810007040
111037645
111025982
810000180
111016998
111035618
111014556
111025790
111025856
111020417
810025339
111014443
111022809
111006060
111013199
111037439
111029602
111040782
810025477
111038706
111029250
111005997
111025038
810000098
111014720
111006010
111040899
111012762
111037093
810022189
111041136
111035753
111033277
111032103
111025704
810001429
111033008
111030594
111033300
111033089
111020400
810000403
111006447
111042802
111034349
111006461
111037208
111025803
111034471
111018160
111025807
111019197
111041140
810003262
111039311
111040373
810001808
111019046
111018139
111040557
810007011
111029390
810017176
111023705
810022195
111025799
111023194
111020454
111026091
111026080
111013154
810002715
810019670
111029795
111014353
111029771
111029796
111024267
810002447
111037086
111031703
111032966
810000096
111030696
111040421
111034519
111029178
111030344
111030033
111014519
810026338
810026263
810026218
810026035
111024984
810000573
111005356
810003108
111020428
111027391
810019707
810027361
111019141
111014514
111037740
111034314
111006415
111033150
111022475
111027377
810022672
111039627
111016056
111005625
111019229
111025795
111016100
810025828
111037204
111033145
810019091
111037387
810007270
810007111
111023377
111029402
111028665
810000945
810001605
810028320
810000970
810001738
810002595
111034103
111025913
111025719
111036849
810001717
111036148
111027155
810027395
111005009
810015267
111014807
810022973
111039176
111005939
810027359
111039184
810018690
111028642
810008192
111039286
810020102
111038980
810019833
111006002
810028604
111016596
111016608
111014738
111043154
810027349
111016589
111006441
111014333
810007206
111014417
111015304
111030577
810007205
810002127
111005566
810001764
111034491
111029571
810002394
111030302
111029461
111036848
111015965
111008524
111029653
111040274
111037066
111027451
810028294
111019902
810009175
111030045
111027533
111015948
111025351
111037190
111036013
111043254
810022208
810001741
111037898
111038868
111017716
111040856
810019654
111036057
111037535
111014518
111014431
111040883
111029480
810025820
111042642
111036309
810020945
111016353
111025727
111041355
111022635
111024102
810006454
111040909
111042635
111036164
810001733
111020467
111029547
111036136
111024283
111032097
810001736
810022696
111034374
111029728
111037084
111039235
111033213
111034136
111035748
111042243
810023149
810001713
111040783
810018856
810003258
810023277
111019078
111033423
111031778
111029239
111025713
111028554
111029654
111038827
111030688
111029725
810003121
810002885
810007290
810008788
111034401
810002886
810005872
810007232
111027399
810007110
111013386
810027445
810026509
810002111
111037563
111036049
111024777
111040486
111026037
111036997
111018185
111030568
810024092
111028555
810023553
111017788
810019703
111030368
111033401
111013391
111017751
111028546
111028552
111032919
111017793
111022625
111022597
111022600
810019714
810019835
111017728
111022617
111017777
111033361
111042035
111022621
111024500
810019673
111033437
111030643
111042142
810023554
111028563
111030645
810023537
111030357
810024097
111033389
111030653
111032032
111033427
111022607
810019669
111004968
111005784
810000773
111040276
111023667
810008709
111024934
111019133
111025197
111023375
111022798
111030079
111008472
111041350
810020954
810027456
111034386
111034527
111020707
111016130
111008385
111026749
111039663
111029693
810025383
810000135
111008473
111008499
111030396
111024463
111024723
111016414
111025025
111006399
111035881
111027434
111037140
111034095
810001662
111034523
810001955
111026755
111034568
111014483
111042038
810028304
810001839
810000208
111024502
111014421
111016337
111014966
111016452
111037725
111024949
810002253
111005014
810008276
111005030
111029670
111025786
810001766
810028274
810002181
111029319
810000116
810008219
111024769
111032891
111016058
111014985
111034222
111034075
810002185
111019083
810023162
111038959
111030141
111036134
111034237
111036142
111034459
111029566
111023166
810000455
810019840
810020077
111037794
111023781
111029668
111023257
111016025
111012897
111017773
111016555
810001737
111029335
810024031
111019173
111037609
111014998
111032907
810006122
111026629
111034440
111036163
111038889
111029576
810028314
111009078
111029628
111006280
111023746
111033372
111037709
810022708
810000777
111036047
111030121
111029776
111024807
810000350
111024232
810020024
111013188
111042773
810008275
111016340
810002391
111039617
810018903
810026207
111037713
810023322
111035602
111006467
111008427
111016323
111030142
111029387
111036033
111037343
111037094
111017768
111031619
111016342
111036846
111030332
111016567
111034729
111038726
111030018
810020113
111029406
111029474
111029713
810000179
111034772
810008868
111037852
111037900
810019845
810017141
111027536
111027172
111014310
111023773
810022838
810020206
111033176
111036087
810017140
810017165
810017172
111029500
111029279
111034350
111030327
111029702
111029259
111006455
111033200
810024168
111033414
810007308
111039652
111037122
111024740
111016013
111034361
111038720
111038719
810018776
810022834
111035729
111035714
810023016
111016569
810025350
810022830
111040483
111035736
111016964
810018686
111036105
111034798
111035788
111006462
111039576
111017783
810025394
111019120
111036210
111029558
810028591
810028590
810019862
111025153
111037879
111038837
111037854
111025843
111017807
810027435
111023300
111030082
111017732
810000774
111015325
111013185
111025209
810027440
111039080
111015952
111042143
111035821
810027369
810023280
111024961
111008390
810007054
810007099
111036179
111039521
111038958
111024814
111030722
111024966
810003190
810028295
111038764
111036865
111027208
810022866
111023384
111039478
111039011
111039407
111032116
810017152
111022807
810024054
810024059
111023692
111042631
111026538
111005966
111013364
111013401
111013166
111013178
111035758
111030565
111027393
111030394
111022880
111029719
111022890
111039314
810001398
111027455
111029739
111015347
111023369
810002388
810023531
810002460
111013144
810001075
810026439
810026462
111035784
111008529
111042634
111036138
111040552
810023307
111020713
111033063
810028287
111038899
111038901
111038968
111026522
111033179
111035792
810007246
111032971
111033283
111030309
111037171
111042065
111031938
810001719
111012773
111038694
111005404
111020388
111038659
111038860
111020410
111026737
810002730
111036884
810001666
111039063
810026332
111030334
111013361
810020304
111039580
111014751
810025849
810006456
111024978
111023583
111016004
111033180
810006473
111029262
111038608
111038555
111030339
111031852
111031915
810026087
111031954
810026003
111039108
111025190
810001754
810001767
111031929
810003199
111029593
111030415
810027358
111037014
111039222
810001684
810019885
111006456
111034665
810022642
810024068
111034298
111020402
810023010
810028473
111024254
111019908
111006437
111029731
111023373
810007267
111034291
111038811
111024986
111036121
111037180
111034198
111031591
810007075
111040553
111040765
810022173
111037181
111035903
111033171
810020319
810020287
810019130
111031565
810020013
810003184
111036214
111031667
111014416
111037098
810023548
810027343
111037700
111033022
111030575
111033451
810026244
111034332
810027344
111029730
111033368
111033371
111032114
111040434
810007327
111034773
111039591
810024425
111016611
111028569
111039207
111037738
111032011
111028583
810020937
810020293
810022878
111038580
810025254
810026487
810007275
111025812
111033178
111034701
111032057
111033298
111025141
810019127
810009182
810019810
111033100
111033124
810020938
111034364
810022199
111040427
810023042
810024377
810019858
111035760
111030158
111031774
111041365
111026559
111034628
810007060
111006419
111023195
810009017
111016049
111029389
111033208
111039599
111033035
810006957
111040345
111042798
111029750
111042136
111029740
111042039
111030017
111037706
111040469
810002694
111037716
111042730
111022476
111019051
111027479
111033962
111027421
810024042
111040268
111038591
111013150
111037054
810001169
111034795
810002392
810022875
810023181
111033428
111035717
810028282
111030325
111030145
111030360
111019202
111018365
111027301
111037765
111014806
111019124
111025817
111015962
111020460
111025840
111027484
111033429
810007188
810000713
111016044
111016347
111015964
810025387
111018307
810000219
111036178
810000607
111033993
111034469
111033994
810020093
111043259
111014466
111005783
111037026
111037782
111027327
810006480
111024229
111008489
810000421
111039164
111037154
111033325
111024760
810005882
111023388
810005875
111023595
111022444
111024955
810008998
111035843
810022827
111031595
810020060
810028301
111005400
111029440
111034010
810009151
810020064
111040558
111033345
111042096
111037603
111038602
810008820
111038556
810000943
111037769
810000947
111035896
111031627
111034391
111037893
111020434
111023185
111005367
810000374
111036990
810026139
111040845
111040885
810022184
810019124
810007190
810007325
810025175
111037126
111035954
111039249
111015932
111034217
111037043
111034728
111015983
111034743
111032088
111041310
111016018
111014757
111034733
111034782
810002924
111020391
111036868
111012765
111006440
111006026
111034499
810020118
111033447
810022690
111030569
111023709
111026684
810007068
111016358
111037710
111016355
111041089
111006043
111031982
111041341
111039267
111033331
111037796
111033011
111030402
810019095
111019042
111040697
810001485
111033113
111029572
111030335
111035688
111017003
810022192
111040338
111039411
111031920
111028530
111030646
810007158
111029356
111032996
111032095
111020715
111027212
810003188
111034331
111032893
111023687
810025801
111033030
111025226
810007314
111035762
111030328
111039085
111031735
810008195
111031948
810003149
810025365
111042744
810018864
810026175
111034343
111025729
111026548
810009005
111009108
111018180
810003256
111038597
111025800
111037572
810008304
111037589
810000060
111016174
111032979
111037655
111039485
111034091
111004961
111006403
810001852
111014309
810026333
111024908
111030650
111029790
111034265
111037749
111034329
111033327
111034542
111019041
111016408
111016329
111039530
810000147
111020440
810007181
111015313
111037838
111018341
111020379
810025862
111015308
810019640
810000974
111029706
810000155
111037155
111018207
111031664
810025806
111031827
111016978
111034801
810023182
111037885
111043181
111024120
111024943
810007282
111024094
810001063
810007284
111036100
810020012
111039447
810019848
111039134
810020079
810022705
810020953
111043211
810023166
810022850
111035996
810000416
810007194
111026726
111026545
111029302
111033183
810000214
111015990
111025225
111038590
111034460
111040306
111038686
111034147
810022651
111037135
810019907
810002580
810007059
111024494
111008442
111014548
111029560
810008665
111033037
111033155
111036835
810002921
111040784
111031916
111040475
111005775
111006412
111039008
810007127
111034603
810022213
810000975
111040809
810027375
111027423
111035926
111034486
111034703
111035898
810000358
111031984
810022679
810024194
810022682
111037771
111037538
111028431
111039106
111029342
111040681
810002311
111014505
111039585
111037886
810019120
111037671
111038853
810024204
810026507
111030398
810026503
111035994
111025834
111040711
111042777
111035988
810020321
111034779
810000412
111023177
810001410
810001422
810007233
111042806
111042747
810023170
111042110
810019693
111029181
810023156
810002706
810022205
111032951
810020080
111039395
111019219
111022601
111016031
111037791
111016474
810002866
111019130
810001722
111008443
111030572
111029716
810008704
810000976
810001428
111037788
111028641
111035730
111025928
111035740
111028621
111025240
111034700
111023244
111033458
111006032
111016424
111035620
111016433
111034615
111033081
111023383
111018167
111029538
111039399
111013184
810009159
810001165
810026226
111006277
111033424
111038972
111029608
810001888
111036976
111034096
810009166
111017767
111023361
111023370
111036864
111028620
810008203
111024968
111006869
111006868
111020711
111015346
111016370
111043228
810026038
810026024
810008948
810006293
111027229
810001106
111033395
111042044
111039553
111006852
111008433
111030102
111033349
810023267
111026009
111035963
810007193
111038851
810000748
111016436
111025022
111032031
810027414
111038835
111037728
111022626
810017169
111013368
111040255
111012901
810024380
111025796
111024802
111037851
810002938
111016372
111014958
111014454
111024793
111017730
111029324
111028465
111028541
111015336
111013310
111032938
111014408
810008684
810008789
810002601
111022796
111014749
810002463
111014565
810000593
111013339
810001366
810001747
810001787
111025354
111014502
111006391
111027534
111006386
111006387
810000565
810000901
810001806
111019165
111025859
111020485
111020475
111016088
111014457
111014335
111037129
111027386
111013378
111034583
810002386
111020408
111016594
111013366
111013403
111015342
111020407
111025275
111024744
111032982
810008699
111016078
111030235
810000491
111005353
111013404
810007302
111024477
111023681
111026681
111025911
111023751
111027419
111030060
810008876
810000592
111015979
111026751
111029344
111014478
810002458
810007079
111016087
111033007
111026043
111023753
111025801
111018164
111026061
111027189
111024787
111024736
111037755
111041061
111025831
111015951
111034228
810005962
111006385
111032928
810001811
810005970
810002681
111040789
111023750
810007307
810017151
111025265
111036853
111024259
111005928
111019637
111027200
111033382
111033444
111040894
111026742
810007088
810007069
111040511
111027480
810023215
111033088
111031988
111036007
111030410
111042054
111036184
810020291
111039014
111041390
810003150
111033073
111041337
111041385
810007610
111041346
111040775
111039173
111039251
111024981
111027472
111039327
111025019
111041352
111031971
111034539
111039255
111025218
111019913
111031669
810026429
111039481
111027177
810026069
810026471
111039084
111030301
111019122
810001858
111039274
111041387
111039296
111029296
111034157
810008184
111035849
111025930
111040840
111042793
810026193
810007003
111030668
810009172
111039492
111033015
111034365
111009102
111026017
111026002
810005874
111007246
111038626
111005998
111039500
111014504
111014549
111014773
111007251
111031608
111034452
111028611
111031708
111031727
810022881
810007089
111014770
111033399
111030214
111017765
111029272
810026205
111028460
111008525
111019203
810028311
111031689
810023145
111035888
111042769
111034682
111038693
111034208
810008976
111039390
810024045
111023665
810025866
111025041
111023765
111025001
810025332
111027631
111006017
810007006
111025710
810008773
111026490
111034421
810026521
111037832
810007210
810001661
111026034
111037784
111032995
111043216
810000375
111043166
111037168
111030587
111028462
111033077
810025488
111006439
111017809
111019167
111033012
111037112
111043165
810027419
111025032
810007086
111018169
810009205
111023619
111038920
111029483
111034370
111024235
111020393
111034796
111037163
111014537
111005554
111005767
810009008
810009179
810002405
111031634
810024525
111038816
810005964
810005961
111033101
111006444
810024521
111022473
111030346
111023797
111023747
810025848
810001720
111014106
111041126
810022858
111042770
111024996
810007081
810001596
111023179
111037024
111040911
810000423
111033034
111027161
111027236
111032991
810000196
810001731
111025143
111042752
111029548
111041377
111030159
111029708
810028305
111016969
810016798
111036061
111022794
111024915
111005768
111034205
810007611
810007620
810006336
111040689
111023771
111025283
111023674
810001663
111029210
111029514
810022823
810007134
111036892
810000383
810002905
111024730
810001864
810000804
111031605
111036979
111033457
111037114
111039686
810026210
111026035
111024091
111009092
810007123
111012793
111026090
111040704
810009004
810008695
111036971
111034707
111030135
111024238
111023283
111029655
111032124
111039613
111029213
810026311
111037824
111008389
111025000
111025908
111033086
111033009
111019169
810026458
810026454
810008879
111033324
111029427
111035706
111034306
810020211
810002865
111024241
111014420
111005318
111005261
111004970
810002684
111024763
810001564
810002247
111026777
810002331
810002727
111042141
111037194
111034080
810001435
111037170
111014348
111026062
111043198
111033996
111013209
111034226
111033153
111006658
111034142
810006988
111040875
111013344
111030705
111035682
111027464
111017141
810000191
111037110
810022190
111014445
111016177
111033407
111039662
810025224
111043169
111030657
810028587
111035782
111029635
810024176
111029766
111040371
111034708
111030270
810009198
111029786
111043164
111029760
111030262
111025296
810009176
111026666
810028489
111034497
111027487
111019090
111026076
111042654
111025365
111034454
810000571
111024470
111019630
810018905
111038621
111034477
810002920
810016796
111029291
111030372
810023194
111040349
810020030
111038866
810025842
810026259
810019883
111039180
810026204
111017131
810020276
810027388
810000440
111025139
111040925
111013392
111005028
111037626
810001734
111004979
111039567
111007748
111039543
111007773
111033308
111007751
111016544
810001677
810023303
810000341
111024973
810024251
111039454
111039412
111032024
111034493
111030055
111028622
810023172
111027389
810001951
111037872
111034119
111013227
111038797
810023134
111013191
810002225
111034127
111030089
810024197
111036143
810023283
111016429
111032075
111027512
810002194
810000806
111023702
111029341
810007196
111035756
111039349
111035880
111037038
111034160
810023285
810018904
111035895
111034420
111033068
111039362
111031959
810025835
810025799
111037212
111043151
810025331
810025328
111006824
111043208
111031630
111014450
111016426
111039658
111023769
810022211
111024233
111036861
810007191
810007049
111008475
810020202
111033020
111005619
810007182
111014759
810020302
810000117
111019638
111026075
111026041
111006030
111014977
111014781
111023582
111033356
111004992
111035795
111036268
111014995
810000354
111014959
111029592
810007258
111025249
111025793
111033114
111030279
111034783
810024250
111037798
111034273
111031976
111029789
111033398
111038567
111029774
111031856
810024424
111034173
810020027
111033032
111028436
810000805
810024514
810027425
111033246
111033169
111018159
111030420
111033130
810002116
111026785
111026738
111006838
810006469
810026445
810009141
111039496
810022803
111038553
810009196
111026040
111042072
111041383
111029495
111029295
111031610
111024110
810008965
111023572
810003263
810003103
810002904
111031631
810008959
111033945
111024918
111026651
111031755
111036030
111034018
111040341
111039201
111013359
810001890
111027335
111027639
111032980
111031923
810024081
111008436
810001067
111006668
810024348
111043170
111032068
111039396
111026731
111016147
810026331
111036091
111034345
111037651
111031752
111033373
111034715
111027205
111030338
111039537
111038568
111029559
111031794
111034532
111033377
111024923
111016022
111030222
111016579
111039575
111038598
111040837
111027628
111013311
111033984
111013175
810008673
111019171
111013377
810008714
111036124
111033974
810026283
810023191
111006007
810007161
111034777
111026668
111019148
111006006
111006066
810002650
111020694
111016156
810008977
111037182
111034289
111034599
810020073
111039678
111019204
111041125
111026747
111035798
111026006
111013309
111013214
111023767
111019087
111016574
111032922
810019107
111024781
111012739
111030284
111005623
111037363
111019206
111028639
111018349
111024118
111042241
111019218
111019222
111016365
111040865
810002936
111041129
111038895
111014533
810009010
111040422
111035737
111038871
111038629
810000368
111023605
111020385
810000333
111029625
111030356
111034663
810023308
810020009
111035990
111043225
111008426
111017792
810000866
111023293
111026026
810022710
111026029
111013338
810000075
111008407
810000059
111036031
111037723
111028450
810020931
111023292
810022638
111015978
111016377
111025155
111040691
111006856
111032976
111031658
111038842
111023379
810025233
111027414
111037354
111035716
810000681
810023198
111028466
111023587
810000903
111013160
810007076
111008496
111038807
111023591
111014780
810008785
111024997
111030116
810007145
111036878
111022636
810020081
111037600
111027162
111031942
111022883
810026488
111039069
111039355
111041394
111039391
111026015
111033289
111039380
111039386
111006063
111017714
810001381
111028447
810001376
111025214
111023196
810002132
111015309
810019692
111043189
810007017
111034526
111024464
111040418
111020458
111040275
111037837
810001894
810005880
111028590
111035961
111033391
111017829
111029601
111031642
111039433
810026040
810026232
111031625
111034556
111027392
111012781
111023367
111035738
111008404
111020412
111020427
111036985
111029457
111033249
111029770
111035684
111012779
111043180
111029525
810019099
111016438
111014797
111040863
111006400
111039357
810000710
810027421
111025789
810022178
810000390
111035874
111027157
810001425
810003295
111030246
111023786
111024115
111016016
810026491
111029458
111014357
111025905
111037708
111031949
111024248
111035751
111027436
111016117
111034158
111027465
810001891
810008949
810023140
111034106
111024231
810001848
810025348
810002598
111037131
111037628
111026732
111036078
111043215
111029751
810009212
810020946
810023036
810020279
111034199
111023162
111025993
111025277
810024256
111025863
111024970
810023301
111032051
810016800
111037081
810007618
810007273
810001602
111038691
810007144
111031613
810028279
111042713
810027365
111023571
111041393
111006421
810007195
111038687
810026266
810025401
111016423
111006401
111040922
111042130
111023279
111026780
111023189
111026045
111018165
111029685
810020210
111032035
111019055
111041099
810007130
111042051
810023538
111041055
111041144
111036112
810001418
810025221
111042148
111042028
111029411
111029368
111040816
810028610
111019132
810024018
810025197
810027357
810027392
111037102
810003141
111024753
111024896
111027210
810008671
111030261
111034123
111020701
111037033
111006064
111009072
111035930
111035830
810008982
810028474
111028576
111033972
111012912
111029585
111014786
111030014
111034507
111037056
111040485
810026133
111034302
810027415
111037209
810020294
111043160
111028429
111036202
810002241
111008435
810018889
111034775
810027438
111037006
111014964
111029633
111026667
111029534
111023671
111032002
111008501
111026042
810022868
111026074
810001503
810001364
810001488
810001498
111005945
111026023
810024172
111038813
810001362
111038971
810025186
111026010
111039687
810008947
111034766
810023551
111029218
111035597
810008956
111032121
111030296
810015266
810020208
810022663
810019641
111026036
810023555
111027324
111014441
111027507
111027496
111019639
810019823
111023776
111016566
111032022
111030307
111033315
111023686
111007754
111027417
111006263
810022168
810022845
111026630
111023575
810008786
810009171
111026073
810008968
111041324
810006976
810007288
111031566
810007274
111041322
111034260
111034268
111027341
810025183
111041064
111039275
111027255
111025850
111016132
810026208
111009071
111031919
111031806
810026289
111034304
810024037
810019903
810023318
111034240
810025347
810022639
810026192
111009097
111040494
111025016
810007074
111036137
111040706
810028486
810008712
111036826
111013170
111040915
111024498
810002725
111008448
810000441
111020411
111020468
111034576
810006455
810023157
111027629
111013323
111038891
111038782
111040924
111008460
810000371
111031564
111024905
111031668
810023546
111025904
111016099
111014307
810002589
810001495
111012925
111038752
810024211
810019103
111025268
111015992
111008495
810002695
111018344
111008470
810003195
111014414
111005940
810020040
111006833
111023611
111008464
111033104
111023273
111025140
810002285
111037447
111016003
111031844
111015944
111038937
111023628
810018853
810019893
111023666
810025497
111029667
111025861
111040724
111023266
111023233
111025879
111032959
111014495
111014312
111004978
111035993
111029524
111017827
111037685
111029470
111033986
111036189
111017806
810009168
111025172
810009201
111036873
111025196
810002581
111008403
111009088
111027380
111031977
111038921
111037089
111027643
111036132
111038645
810025377
810009189
111006275
810008696
111030223
111027390
111029734
111035617
111027494
111016338
111017754
111028650
111033997
111038922
111033422
810023541
111042662
111027250
111029270
111040726
810023168
111016007
111038673
111033061
111019188
111017748
111017791
111036058
810020056
810020058
111029536
111029663
810026145
111031857
111032073
810003204
111033202
810019128
111029532
810026160
111028603
111029587
111027410
810024253
111036995
810005972
111037816
111028444
111036141
111039641
810002858
810026120
810001669
810027346
111034411
111019235
810020053
810020042
810020041
810007072
111022445
111030698
111025358
111023248
810001748
111030040
810001745
111029481
810023264
111013044
111041095
111023790
111019907
111008462
111030416
111031877
111025276
111029298
810022215
111025219
111030721
111016381
111031685
810001494
111038927
111037108
810020288
111039133
111026004
810003209
810008692
810027393
111005624
810002818
111028648
810003142
111005401
810025826
810019998
111038965
810023001
111006036
111038809
810022842
111034654
111034764
111029581
111039081
111038712
111029677
111034634
111025985
111029618
111029583
111025983
111034800
111029620
810022977
111038998
111029596
810022809
111034780
810019880
111038818
111034581
111030701
111030699
111034696
111030709
111027446
810022885
810022982
111038986
810023027
111034443
111039238
111034640
810001853
111028633
810003202
111035731
111008518
810002935
111022485
111034122
111016111
111028578
810002113
810002927
111014499
810001810
810002705
810022720
111028577
810020150
111037201
810002345
810002900
111037659
111038839
810001114
810025230
111019139
810008221
111008416
111034512
111005000
111043186
810023552
810023279
111016967
111039226
111029329
111029195
111035599
111033362
111029333
810019665
111032112
111035815
111042721
111039601
111025174
111034377
810006449
111030228
810023268
111039170
111036312
111029515
111039527
111018171
111023372
111039443
111034802
111040776
111040279
111035914
111024895
111029543
111026503
111024731
111032061
111014800
111006894
111031961
111016999
111025886
111022886
111014776
810000339
111014167
810022703
111039648
111024503
810009153
111025361
111025170
111023694
111023206
111025201
111027256
111039643
111040821
111037531
810028306
810020198
810016804
111031678
810025174
810025253
111022805
111017762
111022612
810007124
111022800
111029745
810008225
111007255
111019631
111039284
111009118
111039519
111038856
810024021
111032925
111032888
111027245
111022629
111037151
810024207
810008191
810022882
111016161
810008296
111016364
111012916
810007295
810026435
111040755
111016467
111015328
111014983
810025371
111039293
810001407
810008681
111042748
810023000
111023634
111007768
810006481
810024072
810023141
810023136
111040720
810003114
810020101
111029778
810006453
111029338
810025205
111035894
111041058
111017726
111040842
111025896
111006407
111016125
111025924
111016435
111019045
111023626
111039175
810002387
111024262
111033147
810000485
111005617
111031935
111036072
810003269
111020705
111006048
810026340
810008974
111033135
810006999
810008986
111024499
111023633
111043242
810002638
111029193
810007237
111016070
111008444
810026130
810026461
111040852
810026216
111016413
111014434
111022486
111016142
810026147
111030662
111039216
810018681
810026103
111042139
111027194
111037214
111029497
111024957
111030382
810002719
111029549
111027416
810000221
111015999
111013167
111028425
111029588
111036067
111035723
111026662
111043162
111042126
111027406
111040370
111034024
111008441
111014795
111027169
810020128
111016972
810028613
111016318
111029675
111016141
111027252
111027235
810006467
111016035
111027176
810006468
111016097
111016138
111019240
111027165
111017738
111017750
810001743
111016165
810000407
111024273
111019211
111017739
111034271
111034076
111034734
111033119
810020278
810007001
111029388
111024946
111034647
111026753
111026523
111030703
111033162
111041134
111031851
111040487
111042794
111042734
111031945
111039625
111038912
810002932
810017142
111034546
111034354
810017103
111033187
111038705
111039050
810025479
810000411
810024076
111023363
111030333
111027259
111018212
810008999
111023255
111038739
810002457
810008969
810001893
111039529
810016797
111035773
111029644
810024244
111038787
111033050
111039531
111023703
111037891
111033979
111033971
111035864
810001493
810001841
810001170
810001610
810001838
810007048
111033448
111037721
810007112
810024238
111026055
111033043
111042648
111040438
111037812
111041120
111006009
111023701
111017782
810001360
111005957
111027280
111023614
111041331
111029729
111031998
810019084
810020205
111018340
111037617
111027258
810000408
111018364
810025319
810002149
111035943
111039468
111031986
810026187
111023240
111040565
111015950
810006983
111036117
111033174
810002188
111028612
111039129
111035965
111023223
111006281
810018872
111026565
111018362
111025013
111037646
111019215
111034741
810020096
111039393
111014567
111006826
111009109
111019088
810027397
111012928
810027377
111005627
111032100
111032077
111022790
111033087
111026549
111023698
111012908
810022966
111040717
111016055
111035840
111032955
111038778
111016120
810000772
111034612
111024766
810003254
810001840
111035705
111029541
810018769
111017812
111038633
810019801
111035982
111026736
810005876
810026457
111039248
810006443
111042674
111016612
111013148
111030059
111026763
810026197
111034214
111023145
111029542
111018193
111033071
111037395
111029539
111023378
810025362
810020193
810020309
111026064
810006982
111009070
111040343
810006338
810018881
810024166
111030032
111034951
111033995
111022896
111012763
111025135
111029233
810007128
111043217
111039163
111030315
810007172
111039562
810020137
111037605
111025255
111033152
111023256
111029454
111032133
111029365
111020710
111016024
111006886
810000807
810008679
111027164
111031624
810001819
810025188
111013162
810026134
810000470
810006335
111025148
810001870
810001603
111040333
111025031
810001111
111038560
111034584
810023275
111023586
111036310
111008530
810000133
111014536
111008467
810000134
111036900
111037378
111033330
111037119
111039692
810020197
111036085
111029318
111036183
111029429
111007778
810003292
111008452
111016093
111008534
111016061
111008453
810000677
111007758
810003296
111042646
111034520
111033112
111020461
111038603
810025360
810024758
810020948
111039484
111029484
111037359
111020709
810008676
810002854
111031657
111026495
810000073
111019193
111034622
810008961
111026760
111030707
111037336
810023201
810025182
111037386
810024063
111038732
111037095
111030695
810023200
111038946
810026007
810025999
810024523
111039446
810024049
111028586
111024484
111024796
810020020
810024082
111012771
111026071
111028627
111028579
111032008
111025936
111037809
111033275
111013396
111040781
810000898
111016159
111043143
111043176
111025858
111025024
111027530
111016384
810024181
111026053
810026310
111027298
111034558
810008277
111036982
111005954
111030266
810020123
810026202
810001370
111024099
111030571
111032131
810017149
810026275
810001496
111040334
111038574
810008985
111034447
810008993
111033312
111031868
111032974
111036867
810008703
810002183
111029186
111029311
810002134
111034135
111034511
810007067
111016114
111038754
810009211
111025202
111005266
111040450
111007776
810020134
111040773
111030123
111043265
111018198
111018181
810025312
810000077
111014487
111040265
111008394
111018342
111019225
810000484
111025154
111037545
810020022
111040448
111034712
810026433
810024530
810002718
111022885
810002697
111037860
111029623
111023275
111026084
111027461
810020004
111005781
111005776
111030417
111019220
111014304
810001400
111019209
111005325
810008316
111033416
810007623
111027517
810000944
111024501
111034631
810023261
111030215
111036833
111035601
111007764
111015316
111012744
111012768
111007260
111004994
810001772
111014449
111024274
111027244
111030124
111014748
111012797
111004977
111039384
810000138
111040425
111006802
111016077
111005964
111037616
111034725
111028459
111039017
111037364
111029784
111025212
111029563
111016543
111023762
111016113
111026088
111024112
111027146
111013153
810006465
111029278
111043212
111009087
111006810
810000487
111015995
111039023
111038729
111039628
810020045
111040462
111006805
111030371
111016554
111014415
810007241
810019659
111032047
111014314
111017817
810000673
111040272
810025248
111024496
111024497
111037549
810001721
810028594
111023365
810009147
111030676
810026222
111034025
810023046
111031777
111034279
111027311
111033115
111033989
111036025
111024891
111014791
111037035
111024268
111036258
810025313
810025192
810023528
810025234
810002341
810024079
810020107
111035735
111035856
111030386
111029638
111035872
111027486
810007217
111012952
810027408
111037091
111034598
111039363
111032105
111025806
111033948
810002895
810023205
111029349
111014538
810000385
111028428
111032904
810000332
810001405
111019230
111037332
111041132
111018190
111031842
810020936
810019102
111039556
810027411
111006452
111014165
810007239
111012764
810026474
111014753
111030137
111029637
111017758
111040836
111039220
111027167
111023249
111014411
111036829
111042109
111008408
111008522
111038703
111030288
111029609
111029504
810025867
111039141
810016802
111043182
111029661
111027478
810002576
111027443
111016581
111039620
111028523
111016997
111038601
111034677
810002539
810009156
111014112
111030234
810003293
111037764
810002541
111013387
810002189
810003294
111020702
111012752
111040471
111040719
810017178
111019627
111016122
111014817
111016032
111020700
810001174
111016028
111023778
111029487
111031881
111034296
111031623
111030239
111037394
111042753
810026301
810025218
111040312
810019896
111031632
111014970
111035626
810019803
111030248
810001678
810019657
111036871
810009197
111032924
111035824
111037350
111016558
810007225
111029742
111030652
111034169
810025229
111031746
810000471
111024729
810022647
111014969
111023594
111034253
111035980
111016330
810002452
111033964
810007026
111028535
111013216
111024244
111027375
111027491
111029513
111027368
111031580
810027442
810019661
810027386
111033359
111025266
111032070
111026530
111031932
810023317
111031784
810022225
111041311
111034274
111014993
111023761
111009121
810000331
111014354
810026291
111025975
111018191
111029501
810008287
111028649
111025726
111026066
111026086
111037015
111040454
111030311
111027402
111029265
810025824
111033036
111032946
111029209
111039319
111027493
111030029
111024902
810025496
111040910
111028592
111025873
111014344
111041342
111035884
810019856
810024025
111005760
111042036
111032950
111036035
810025404
111013405
810003189
810026148
111022895
111027270
111013324
810023540
111029462
111023606
111018366
810025478
111006274
111036198
111029442
810007255
111016133
810020126
810009007
111016447
810006346
111033387
111013349
810001816
810005873
810007066
111009096
111037597
810006949
810002280
111027239
111031936
111029263
111029512
810020127
111038692
111027489
111024969
810019900
810007171
111026648
111037785
111039282
810007180
111039212
810026167
111039471
111041313
810006954
111032014
111031893
111039155
111032128
111031891
111031901
111028564
111031887
111033310
111032110
111028606
111025852
810007094
111040920
111019065
111036051
810001438
810001377
111031922
111029614
111038838
111014777
111037553
111036034
810007027
810026270
111030586
111038867
111038833
111013343
111038774
111041309
111036182
111038836
111038795
111016949
111006033
810001356
111031870
810000336
810026233
111038794
111029595
111038771
111036128
810000918
111006247
111040479
810002686
810000082
810026279
111032106
111038777
111016118
111016971
111013326
111013151
810025841
111036193
111037673
810024195
111016602
810001827
810001706
810001847
111033268
111019076
111037691
111039541
111037711
111017803
111024931
111018196
111033460
111028432
111028446
810001682
810026498
810023180
111025988
810020929
111031704
810026231
111039294
111031747
111031771
111031733
111039287
111039055
111034587
111031705
111037587
810019865
810026031
111034768
111031571
810019113
810020057
111039256
810027389
111038859
111038938
111033965
810008972
111040913
111035907
111039232
111030058
810008694
111040339
111040477
111034406
810022836
810019802
111036144
111037732
810000526
810023023
111038806
111040259
111031927
111006671
111030581
810001387
111016995
111005950
111014785
111024909
111034195
810020322
810017157
810025352
810018882
111039566
810022716
810019698
111025883
111006271
810025834
111025145
111041079
111006239
810022990
111030243
111034400
111030638
810000162
111027358
810022210
111012757
111012770
111039424
111006843
111032990
111019909
111027278
810001415
111038631
111029724
111039231
810018782
111006012
111034023
111041092
111005323
810002682
810002150
111040372
810000792
111029190
810025250
810024038
111006808
111025868
810024349
111038906
111039229
111038801
111039247
111012949
111039113
111006865
810002698
111009128
111019047
111026683
111017723
111020480
111037532
111034765
111014544
111016359
810023024
111013164
111033196
111016023
111006053
111019138
111037076
111035739
111039119
111040874
111024478
111024887
111030299
111025349
111037106
111025220
111036046
111024788
810019874
111013400
111039505
111041091
111020680
111012742
111015322
111037707
111023618
810002930
111030044
111004990
111033425
111019232
111024732
810022689
111016563
810002645
111026678
111008445
111032965
111019241
111007760
111014725
111034242
810003249
111027449
111024258
810001834
111023149
111015332
111023228
111041067
111042146
810027455
111015940
111032113
111014491
810019122
111035777
111032940
111015954
111040357
111016158
111029267
111030404
111023287
111040313
810007333
111030093
111034139
111026533
111038954
111014320
111016985
111027438
111033464
111027430
111034346
111038924
111038760
111033256
810008878
111038766
111038892
111038840
111032025
810023138
111034179
111035610
111031666
111030369
810019853
810022879
111013201
810022967
810001740
810024094
810023545
810018680
111033374
810007151
111040812
810025490
111023254
111023777
111039462
111015996
810028298
111034596
111039554
810017145
111027330
111029764
111030663
111031616
111016346
810001708
111034787
111043263
111037757
111035847
111020471
111034416
111033353
111033205
111017814
111034293
111037400
111034083
111034093
111005013
111042759
111005552
111029502
810008966
111019168
111016571
810022970
111042778
111034720
810020000
111025986
111037694
111031760
810024048
810020121
111037903
111025844
810007186
810023049
111023164
810008707
111039125
111032072
111039200
111032021
111039353
810007184
810008678
111041051
111025919
111028534
111029704
810008309
810002707
111027221
111033260
111025862
111028538
111024743
810019886
810006297
111034418
810000178
111037050
111016430
111016327
111029671
111037568
111024950
810001424
111034545
111034481
111025177
111034267
111016041
111012755
111023297
810002675
111022889
111042659
810027390
810019901
111034369
111027344
810017147
111032968
111038974
111031972
111038796
111027183
111020455
111040466
810000682
810002918
111041124
810000678
810006996
810022974
111043184
810023177
111034104
111035632
810002464
111017816
810000775
810020039
111029727
810026303
111040366
810023040
111037867
810025397
111041384
111014803
111041375
810025403
111040916
111040878
810019861
111036084
111042826
111029336
111040903
810025336
111031707
111027190
810026067
111018205
111018162
111014517
111036888
111041117
111027326
111029290
111006268
111013320
111031711
111031710
111005947
111039030
810007318
111031661
111017794
111012923
111041094
111037136
111029464
111031775
111040678
111005789
810019696
810023289
111006418
111037756
111023690
111016538
111037001
111035621
111029527
810002277
111023389
810000771
111027191
810009144
111029398
111030156
111040866
111032994
111025355
810000790
810002897
111027214
810020331
111030318
111025848
111039682
111034752
810000948
111028430
111031962
111039137
111042135
810008289
111031889
111039325
111026663
111027310
111037666
111037856
111039458
111042125
111031964
111025811
111030114
111039147
111030237
111031907
810008301
111030274
810008318
810008310
111042237
810008207
111042084
111039127
810008215
111031909
111030076
111015985
810028291
111040684
111034422
810001433
810000131
810008824
810020016
111036045
111041380
111013212
111043253
810019664
111039307
111036098
111029737
810008227
111038569
111030218
111034101
111038988
810020036
111031739
111029746
111029613
810006951
111026544
111031879
810020283
111012903
111006255
111014509
111014321
111018354
111023211
810002891
111020688
810007064
111030324
111038607
111036040
810018850
810007046
810007211
111031743
111036190
111032087
111040709
111040931
810024085
111025857
111038993
111033102
111040736
810024032
111038829
111037766
111022613
810008176
111035603
111039253
111035852
810008188
111040753
111034508
111030597
810019674
111009090
111037369
111006862
111036088
111033151
810022675
810008185
111036094
810020289
111032039
810019652
111039532
810000946
111036028
810024334
111020394
111039130
111040788
111035848
111004960
111005621
810001419
111037551
111005925
111031911
111016039
810026115
111014775
111025284
810002389
111034504
810007058
111014562
810001683
111039082
810022654
111030385
111024093
810027454
111025042
111033408
111042647
111040778
111038679
111024960
111026068
111035613
810026473
111038696
111014997
111034570
810001674
111024105
810015269
111039372
111025138
111032923
111033033
111032981
111033024
111028548
810003205
111035683
810008667
111033959
810023539
810008685
111025709
810001109
111006423
111036079
810023300
111030083
111026049
111040861
810026116
111023173
810007609
810026127
810008222
111034130
111005040
810002128
111029695
111032943
810006472
111033442
111016562
111029624
111025147
111009076
111006057
111019144
111034645
810028277
111022599
111042240
111036841
111036980
810008706
111027249
111040352
111033198
111037743
111029226
810008983
111037687
111024468
810028485
111030099
111023799
810008850
111040512
111030563
111043261
810023556
111031732
111025820
111039661
111038915
111038700
111016170
810002751
111026748
810026083
810008661
810007285
111008454
111034490
810023213
111035628
111035725
111042066
810002860
810007004
810020005
810007220
111042771
111031593
111039674
810009157
810025375
111026631
810007245
810000427
111029521
111031653
111014339
111039595
111034261
111039316
111020406
111024765
111026685
111042116
810007117
111034132
111039402
111023176
810020109
111042632
111025017
810019890
111025994
111023791
810007330
111030686
111030712
111022628
111006479
111015323
810023527
111030693
111029567
810007121
111024893
111038926
810023186
111025230
111039568
111024988
111034559
111027138
111031908
810023319
111041142
111031931
111027632
111030592
111030268
810008869
111009125
111009075
111035925
810018890
111037125
810027452
111037138
111037397
810024182
810022183
111037546
111038775
111036052
111029192
810000056
810000900
810019672
111036160
111024936
810001880
111039345
111031640
111014990
810008790
810024261
111018186
810006967
810006986
810002192
810002282
810006942
810002243
810006985
111008396
810006946
810006991
111009083
810001826
810006993
810007013
810006966
111016462
810006968
111016378
810002344
810006974
810002250
810002398
810002254
810006984
810006940
111014426
111016066
111016593
810002248
111014512
111016570
810007010
111034238
810019695
810025337
810025196
810025215
810025390
810024522
111004985
111009089
111014458
810006997
810007332
810002186
111016354
111014545
111016363
810002124
810025324
810025493
810025184
810024757
810025310
810023050
810025311
810025195
810025385
111030292
810019792
810025323
810025216
810025236
810025220
810024047
810019702
810025500
810025201
810025504
111036151
810001065
111031952
111033021
111033322
111033299
111033314
111031782
810020067
111035912
111041336
111040484
810024338
810024034
111029346
111033056
111033052
111029232
111029177
111031724
810023271
111023253
111027279
111036875
111041381
111027284
111016545
111037032
111027500
810008791
111014812
111016148
111040506
810002599
111033344
111037004
111008411
111016082
111005559
810008305
111029419
810016817
111040898
810001142
111029692
810001141
810001056
111034566
810001673
810018900
810020098
810026199
810020947
111043139
111041370
810019855
111014723
111013208
810002496
111037845
810025247
111039225
111008521
111036264
810001430
111016586
111020483
810002184
810019106
111040368
111043247
111029639
111034089
111037158
111033184
111043156
810019086
111034691
111034175
810019910
810019826
111039694
111039491
810026512
111031688
111039666
111025199
111039664
111039138
111038820
111039183
810026342
111022469
810007105
111033264
111035850
111039228
810026320
111035755
111038776
810026211
111029276
111031944
111039059
810026268
810002574
111039646
810026053
111039285
810006121
111032066
111031899
111039060
111032125
111039644
810026523
111034604
111029477
111039618
810026514
111039582
111031980
111008430
810026255
111025195
111024987
810026492
810008193
111039046
810008179
810008319
111029478
111029175
111039637
111031800
111031660
111035604
111026536
111016019
810008278
111025011
111031786
810026508
111031741
810002489
111039116
111039683
111039685
111039688
111039654
111029640
111029626
111034337
111032933
810020142
111038991
111039463
111043145
111024123
810023337
111027408
810019863
111033293
111025854
111037022
810009192
111034668
111040729
111034518
111042643
111038885
111036192
111035693
111026733
111033262
810019716
111023305
810022198
810022188
111034403
111040860
111037128
111037831
810024246
810019832
111039115
111038874
111034084
111030015
810026256
111039303
810020201
810026025
810026005
111034404
810026084
111035992
111038718
810008229
111034651
111025878
810003193
111023754
111033210
111037207
111038781
810007185
810022984
111017795
810025800
111024761
111042667
111043155
810019877
111042763
810020008
111040838
810017146
111030147
810024198
810007165
111031768
810024178
111034648
111031584
111031749
111037776
111038987
111039432
111035859
111006207
810001069
111028673
111036060
111029594
111028561
810008994
111029641
810000716
111032954
111034510
111019238
111037698
810007303
810007179
111037591
111034474
111037631
111036102
810007125
810019892
111038985
111039629
810019867
111036156
111014811
810022831
111029283
810001058
111043146
111037059
111033095
810020297
111008413
111039388
111018195
111029237
111035816
810007266
111026057
810008285
111030160
111034223
810007140
111035627
810007045
111037627
111014163
111016350
111008401
111024953
111029376
111042106
111025352
111031635
111033247
111036195
111006427
111024797
111038635
111033156
111018161
111024492
810008697
111023788
111030153
111034726
810020025
111034464
111034188
111040416
111024107
111038792
810019663
810002853
810025760
111034671
111039426
810008973
111033329
111006860
810027355
111037570
111032903
111040929
111037668
111034224
810024029
111037696
111040892
810024077
111027348
111039090
111040905
810026243
111019198
111024770
810025209
111006240
810026459
111036876
111033144
111019153
111034203
111008418
111016993
111035743
810009146
111036904
111037150
111025847
111013187
111020405
111037005
810019125
111032026
111031570
111034292
111030421
111037615
111033047
810000669
810026294
111037392
111032058
111034788
810006994
111036859
810000511
111035918
111030221
810002878
111020698
111023715
111027447
111025228
111013224
111039444
111039416
111025198
810007253
111027340
111006819
111014808
111023272
111034462
111015947
111026669
111042751
810023534
810003140
111025837
810027380
111040414
810028467
810019821
111012902
111013331
111022481
111013142
111040685
111016037
111034230
111034340
810026480
810019112
111016465
111014486
111040348
111024249
111039262
111036063
111034760
810020076
111007747
810023287
111031773
111035876
810020282
111041063
810008780
810028308
111034774
111012787
111027531
111027187
111037821
111005946
810002342
810002284
810019648
810018901
111015350
111014362
810000874
810006438
810024180
111027248
111038966
111038916
810025226
111038917
111028596
810006981
111028653
111032074
111031853
111033313
111031898
111031762
810026135
810026068
111039278
111039099
111041314
810026132
111031894
111031841
111031970
111039190
810026028
111041070
810025995
810026043
111031569
111039242
111031992
111031939
111040844
111032096
111032978
111016580
111016140
111033426
111040855
111014467
111014432
111032071
810026483
810022840
111009079
111009105
111036096
111043214
111033415
810020329
111043150
810026125
111017734
810026077
810020106
111026060
111030081
111031709
111015931
111035812
111031614
810019879
810000972
111025890
111025366
111033199
111034713
810022846
111006260
810001427
111027144
111040415
810002462
810026217
111020473
111031691
111029714
111039596
111041097
111006279
111033023
810001139
111020443
810022170
111039191
810028615
810020065
111014448
111034232
111006858
111029300
111006863
111006454
810008964
111029460
111029537
111023264
111024952
111033953
810003107
111041147
810009001
111033126
111027220
111025785
810009178
810002640
111025030
111024488
111042062
111031599
111034284
810026251
111043177
810026206
111033214
111016160
111030414
111027433
810026504
111025725
111033284
111029793
111040745
111039669
810023323
111041323
111040829
111039408
111039281
111038960
111031563
111039434
111039041
111006019
111026672
111027299
111025231
111012962
810007199
111017742
111013398
111005029
111039524
111034252
111036996
111043159
111025282
111030131
111034473
111028562
111040777
111042095
111026682
810001382
111037013
111036008
111006289
111017769
111026660
111005786
111027219
111008463
111018208
810026214
111029787
111040354
111005941
111005926
111008537
810023132
111005951
111005357
111016428
111006832
111030095
111038670
810002592
111005938
810008963
111034352
111035746
111005324
111040273
111041364
111040716
111034184
111029249
810001555
111038999
810002901
111019172
111006242
810002913
810002926
111029247
111033403
111015938
111028567
810002907
111018170
810009160
111040500
810007012
810007224
111030640
111038983
111013172
111035598
111036119
810020130
111033316
111019195
111016416
111014105
111027196
111019066
111024469
111005630
111024234
111034757
111005764
810025190
111039480
810002914
810027400
810002714
810001861
111020422
111015970
111009114
810007007
111038930
111039218
111029691
111037438
810000076
111020420
810019849
111031693
111038942
111027211
810023029
111039572
111036969
111040693
111034134
810018771
111004964
111036974
111023380
111031586
810009143
111042807
111029315
111020685
111025895
111020679
111016319
111023677
111012780
810000372
111012767
810002703
111007262
111029274
111020453
810026189
111031622
111029584
111031601
111012907
810023187
111029493
810007292
111014802
810026088
111016451
111014319
111038748
810002656
810007240
810007337
810007293
111040728
111040880
810007336
810008200
111034597
111014527
111023263
810007249
111015936
810007326
111023280
111006670
111039332
111034003
111026550
810003187
111040467
111023708
810028284
111042032
810027406
810019999
810022179
810022181
111042049
810018866
111034689
111025805
111029452
810020138
111025290
111016568
810026276
111026014
111040790
810026443
111037682
111025707
111016005
111026025
111027307
810000078
810002540
810001595
111019126
111020429
111037643
111020424
810000473
111039197
810001172
111029555
111027181
111025989
111034660
111025165
111034692
111036131
111026778
810002925
810002916
111016442
111014540
111041082
111005007
111025205
111031892
111031883
111022480
111022440
111042656
111042064
111022455
111008421
810023183
810023212
111008477
111040799
810023175
111040743
111022447
111022465
810008825
111037052
111034522
111034182
111033190
111022451
111022460
810023195
111040787
111022483
111040742
111006414
111036135
111027201
111029533
111035633
111013319
111042727
111036070
111012775
111037548
111029757
111006261
111034395
810002911
810008990
810008779
810009009
111026046
810026249
111029580
111042731
111039331
111039111
810005879
111040871
810024091
111037661
111025239
111034241
111031978
111033419
111034012
810022833
111037870
111037678
111041317
111030362
111029767
810007197
111037652
111036983
810006980
111039608
111037167
111034313
810006298
111042766
111038624
111040346
111034234
810003265
810009183
111028661
111042047
111031692
810008855
810002817
810028491
111034015
111034239
111026782
111032079
111022875
111032052
810019843
111037175
810023047
111016102
111024466
111016965
111024483
111030691
810023207
810022209
810023316
111039495
111016382
810020204
810022659
111027352
810001135
111014547
111035701
111037773
810019094
111036850
111036313
111034229
810028488
810018879
810018895
111035871
810028495
111032958
111032029
111012753
111039578
810022220
111034676
111031612
111034616
111034551
810018876
810002537
111025724
111020479
111037811
111039312
111035829
810007204
111033358
810006977
111034609
111031582
810002678
111032967
111037722
810007009
810025223
111034781
111005628
111035749
810002722
111008512
111008515
810000220
810008995
810008988
810008991
111034299
111038933
810009199
810027420
810009170
111014750
111023608
810020200
111034112
111008402
111034577
111013321
111040718
111012920
111037556
810007202
810001597
111024471
111035721
111041116
111020436
111034619
810007269
111039504
111025714
111035937
810025483
810024531
111014480
111025244
111034219
111030080
111025808
111034435
111034180
810019655
111014438
111028597
111008474
111006879
111023578
810000702
810000154
111014413
111018154
111030053
111018197
111040853
111035897
111014447
111017759
111025869
111030063
111024964
111041389
111005958
111024792
111008447
111025819
111030363
810024052
111034152
111020459
111023629
111030293
111029761
810000344
111016046
111023758
111014345
810017136
111026024
111023360
111006443
111017808
111016560
810008992
111006243
111034776
810000351
111028421
111034457
111036443
111006062
111034334
810026227
111031792
111029676
111032886
111029660
111038884
810024340
111024263
111016068
111024812
810000674
111025039
111025227
111034419
111016351
111016445
111033957
111040696
111025935
111006813
111028658
810020216
810001830
111034438
111006442
111016559
111006849
810027402
111006847
111014430
111012914
810003144
111012743
810026441
810001670
810006284
111023599
111023390
111019061
111033157
111041143
111031726
111008508
810006339
111026642
111032034
111016955
111005758
810008691
111019150
111027467
810022841
111037440
810020043
111030253
810024090
810007168
111016461
810028275
111006881
810007201
111031867
111033445
111041113
111025270
111039254
111035745
810020934
810006987
111023203
111004988
111006046
111005022
111027152
111037533
111025838
111029185
111023221
111014476
810025805
111034426
111017774
111019119
111023183
111039160
111039157
111020431
111034427
111042750
111032932
111017823
111031863
111033388
111039105
111035605
111026514
111025156
810026112
111025942
111037148
810003280
111025871
111037143
111017121
111039258
111037116
111038831
810028607
111023222
111017128
111017785
111030100
111035933
111013181
111032963
111040752
810000148
810000409
111013389
810000103
810023533
810024381
111008527
111007264
111032961
111033946
111012741
111038617
111042645
111033123
111029591
111004958
111040749
810001417
810001440
111029194
111005965
111023627
111031572
810026337
111029651
111031626
111023623
810019644
810025484
111039632
111027521
111025927
111041361
111016431
111025191
810019681
111026670
111012909
111012911
111014732
810009180
111039074
111034004
810019694
111034022
810017155
810001656
111034191
111027300
111017707
111034602
111039460
111038710
111038716
111031770
111037437
111012786
111031641
111034372
111041338
111019040
810001501
111033355
810007149
111027329
111033257
111030251
111005017
111029665
111023796
111038596
810002928
111037894
111037000
111040258
810008711
111030031
111019154
111040886
111008386
111034007
111034111
810024075
111033154
810017170
111035823
810026456
111034105
111035624
810007047
111029652
111023782
111024811
111028527
111029680
111040807
111020689
111030401
111037665
810022851
810007014
111029391
111035761
111019170
810002713
810001753
111033265
111027328
111040278
111035857
111025964
810022719
810025206
111006059
111042738
810023209
111027232
111014356
111031975
111006880
111031987
111013165
111040771
111036069
810023048
111032909
111034433
111031573
810025393
111033186
111033159
810020277
111027394
111014774
810025305
111006663
111037160
111009093
111039420
111025918
111006656
111037592
111026761
111022803
810000428
810000093
810027374
111027388
810001854
111023167
111028456
111028437
111028435
111039513
810026455
111009129
111029310
111006054
111023699
810026515
111014765
111006396
111028601
111005773
810020217
810001375
111029510
111038935
111041333
111039100
111031832
111033281
111029505
111023204
810000486
810000513
810020087
111040544
111036039
111026072
111016551
810016815
111042075
111034168
111037388
111042127
111026506
111036099
111017753
111040441
111026783
810017158
810017168
111033055
111035947
111033297
111024748
810007015
111031651
111037654
111023366
111029531
111006848
111018203
111024903
111005969
111039358
111030412
111040304
111030634
111016144
111039445
810024022
111035794
111016410
111036118
111039631
111033292
111031672
111029327
111036166
111016169
111034092
111033185
810007022
111041054
111017715
111024780
111017126
111035800
810000149
111025295
111038681
111041098
111035948
810018885
111036256
111039350
111027525
111023239
111019228
810002606
810002693
111020444
810007230
810026142
111014315
111029612
111035681
111030677
810026253
111025826
111042764
111029590
810007154
111038646
810022816
810006436
810022874
810023008
810023015
810000899
111037344
111016366
111029697
810008226
111019901
810000770
810022985
111038634
810023038
111029673
111028422
810023044
810000813
810022837
111033411
111042792
111039512
810022717
111036083
111039338
810022701
111034001
111038914
810026219
111038967
111038648
111026078
111039058
111039659
111029240
810023262
111033947
111030361
111034381
111005023
111012957
810007328
111033005
111012951
111005568
111023672
810000145
810000157
111027283
111019125
111006285
111026510
111032127
810024422
111034450
111007765
810000201
111028581
111034761
810025853
810002246
111030644
111025158
111024495
111027216
810003197
111035997
111035889
111015935
111039526
111039365
111005779
111005267
810009194
111013177
111016598
810023189
111037868
111034555
111029689
810020082
810002881
111030233
111017710
111027345
111038668
111039461
111023153
111025248
111005788
111037037
111037684
111005613
111006665
111030036
111042045
111023237
111032082
111030403
111033983
810000101
111029187
111006005
810001168
111016149
810027426
111026767
111023785
111016116
111020409
111035741
111016135
111016150
111030074
111029206
810001812
810006973
111032126
111037604
111030306
810019816
111031957
111025237
111018150
111022632
810027370
111026052
111029307
111014103
111018140
810025340
111026555
111006435
111035595
111005953
111037010
111014788
810006962
111037675
111041328
111034407
111034121
111035724
810003210
111025791
810000091
111026775
111034439
111036970
810001658
111036010
111025262
111026771
111014981
810026037
111034716
111039370
111040723
111005359
111024886
111037123
111006257
111018189
111026673
111024111
810003248
111036843
111038547
111038610
111005948
810007216
810007156
810007175
111014771
111025002
111024804
111034746
111030391
111032937
111018173
111034751
111006464
111042099
810008967
111039581
810007279
111023670
111040872
111013384
810007000
810006969
111031793
810009155
810026463
810026444
810003089
810026344
810003253
810008186
810008848
111029636
111006887
810026495
810026000
111024504
810026501
111035686
111016994
111022888
111016601
111023207
111016573
111027458
111034667
111005005
111027501
111041081
111038886
111023673
111027238
810002748
810001389
111024116
111024985
111026058
111013345
810001393
111005037
111016014
810024096
810024350
111039252
111039593
111040677
111040849
111042132
111013341
810024192
111030217
111034685
810023336
111039005
111032936
810003255
111029445
111016963
111014503
111040478
111026547
810002863
111041106
111038962
111034753
810022676
111005943
111030656
810007251
111037695
810008306
111034496
111027444
111033376
111033129
111019185
111034750
810028484
810022637
810026184
111023590
111039565
111024786
111036869
111029736
810020100
111034756
111037346
111031862
111032108
111031716
111039394
111039237
111034021
111025181
111033149
810009187
111030108
111026516
111030105
111030139
810008781
810007033
111025731
810000417
810020072
810020078
111018355
111027271
111040927
111024479
810018848
810002459
111022627
111024467
111025828
111025835
111026740
810027348
810026438
111039520
810024024
810024041
111037044
111024977
810009202
810020031
111037152
111031698
111040808
810000357
111019156
111038788
111033459
111029255
111038822
810009173
111025973
810018875
111043158
111018363
111029254
111020691
111033319
111022466
810000345
810001439
111016960
810006290
111006844
810025355
111024922
111038854
111020462
111018351
111042810
111042641
810025795
111038642
810027450
111036140
111006018
810001060
111008397
810027433
111030090
111038571
111038772
810019864
111030359
111005769
111020390
111019136
810001077
111036832
810025338
111036017
111020419
111034355
810009184
111020686
111029266
111006468
111034389
111023775
111034537
111023285
111027440
810020296
111040722
111030733
111033060
810001729
111017746
111024947
810003186
810024532
111030670
111027355
810020136
111035703
111035734
111016981
111008392
810027427
810002591
810006477
111038897
810019834
111034582
111037077
111033158
810024255
111035625
111027353
111016339
111019640
111027454
810002915
810002126
810003251
111042234
111027476
111033076
111031646
111041086
810022802
111037046
810000776
111039029
111028568
111037817
111039498
810023178
810020299
810020284
111014412
810007062
810027453
111020457
111015327
111005756
111009123
111030148
111030580
111042774
810023282
810025243
111032064
111034282
111037339
810015270
111030126
810002244
111029337
111026652
810016813
111037189
810007169
111029417
111037818
111020695
810023324
111006404
111043178
111006474
111031600
810018854
810008293
810008299
810007108
810007141
810026265
810026317
111039054
111039012
111040260
810026047
810026093
810026026
810026343
111038910
111031828
111039609
111031871
111038701
810026450
111031831
111035713
111039268
111038948
810026075
111029377
111029375
111039028
111038763
111039140
111034618
810006442
111039121
111040264
111026063
810006431
111038970
111039607
111029466
111039588
111039634
111033976
810026452
810026246
111034673
111038896
111039161
111029508
810008204
111039487
111039633
111038947
111039001
111039640
111039696
111035878
111031826
111008546
111041347
810005869
810019830
111008450
111023711
111037888
810006476
111006818
111040307
111030395
810026182
810007338
111029422
111027370
111006058
111036168
111032012
111014534
810007177
111015338
111006822
111016406
111041078
111016592
111041057
810023142
111030304
111040488
111040309
810025480
810017143
111042775
810026138
111030675
111004972
111027182
810008950
111024757
111036062
111031766
111036158
111039352
111041316
810019088
111037648
111034283
111039545
111035727
810023211
111040900
111034588
810026032
111039302
111037595
111031834
111035919
111041382
111033134
810001363
111026765
810002899
810026482
111040359
111039097
810009174
111029385
810002404
111027347
111040733
111023574
810020092
111023374
111026515
111019626
111033434
810026150
111030702
111037559
111005015
111035989
111025877
111036055
810023139
810016792
111040507
111024924
111014746
111007746
111004976
111036208
111014559
111024485
111012915
810002396
111029322
111030037
111009069
111034544
111027228
111026679
111024995
810008205
111015980
111016043
111005782
111006867
810026451
810002390
111014100
810025228
810025231
111038652
111038661
111022792
810007192
111014497
111031751
111006282
111019086
111006278
111016069
111016050
810000567
810007132
810007091
111034463
810002334
111035818
111039304
810016805
111039441
111014482
111039300
111037199
111038611
810028603
810028310
810028472
810028476
111041391
111038662
111039507
810002495
111024247
810017148
810008688
810000177
111030661
810008847
111016038
111032920
111034702
111024779
111036266
111008493
810022222
111031673
111006853
111026020
111022801
111008457
810006464
111030352
111026781
111007259
111019233
810007056
111034102
111027143
111027453
111018155
810023274
111025898
810009161
111016992
810020062
810019875
810008952
111018357
111025254
810027422
111025159
111039280
810001877
810007260
111034430
111037577
111035892
810000979
111032067
111027262
810025334
111042031
810000717
111037774
810026137
111006264
111028423
111027175
111034109
111036260
111014352
111033146
111023218
111023214
810023257
111043196
111014490
111033339
810024170
111033980
111034451
111038592
111035726
111024784
111025818
111030378
111025875
111040269
111037078
111034487
111034209
810023263
111016575
111041075
111040502
111006065
111034305
111008419
810000778
111027206
810020944
810006961
111035973
810000549
111035806
111009111
111023245
111009112
810024528
810024069
111036177
111034347
111043213
111027510
111037028
810006941
111030732
111006008
111008465
111017830
111028533
111023168
111017115
111040692
111024734
111024930
111020437
111032895
810026213
111033966
111034613
111035747
111025703
111006250
111006122
111027526
111035886
111016324
111027277
111023680
111029610
810000158
111036044
111041139
111020381
111014743
810024205
111025211
810025837
810025804
111043183
810022697
111027276
111029762
810008823
111033455
810026118
810026304
111033364
810025856
111016060
810026225
111025885
111014111
111037777
810023203
810022660
810022650
810002673
111017779
111017709
111014792
810022700
111027156
810002680
810022715
111016108
111029227
111029241
111023210
111037731
111037753
111023229
111036840
111036894
810006111
111014107
111036903
111014095
111040893
111037878
810023167
111022884
111034278
111028448
111006206
111020477
111026028
810007096
111022604
111020438
111037206
111014735
111035959
810002538
111027384
111014305
111017118
111018142
810000094
111005761
111025278
810026306
111027304
810002864
111023220
111026754
111038676
111039093
810001765
111039177
111039145
111034579
111034505
111008545
111012756
810022677
111018163
810026110
111013161
111012905
111031579
810000384
111029582
111016126
810022692
111023678
111039203
111015340
810000216
111042800
111023190
111031990
810024098
111023675
111012910
111013219
111034227
111040906
810008282
111032065
111038952
111038844
810022187
111013402
810001390
810001392
810001395
810001414
810001403
111038584
111041345
111005317
111034468
111038677
111038637
111038638
810028468
111038566
111034610
111038643
111031592
111039547
111031702
111034611
810000419
111013362
111013373
111004986
810001385
810001409
111013336
111016982
111016951
111041343
810001399
111013337
810001413
810026224
111027362
111038623
810024095
111032987
111033105
111032941
810019667
810025822
111033433
111009130
810026298
111039573
111039000
111039564
111039112
111038963
111025203
810008223
111029467
111029499
111032111
111014813
111039149
111039645
111039048
111040374
111039061
111039684
810026318
810026316
111039655
111039139
810007131
111038964
111039626
111039650
111039442
111031913
111031880
111039056
111029511
111031846
810026493
810026194
111035802
111031780
111029437
111029492
111029489
111032005
111024269
111038984
111039600
111038941
111028440
810008201
111039528
111031874
111038943
111031875
111039260
111025990
111039333
111039653
810026022
111035845
111032006
111029441
810000509
111029286
111039630
111039563
111039072
111039680
111037795
111035825
111038961
111039009
111024756
810008288
810002729
111039004
111025021
111031994
111023573
111034553
111039171
111039675
111039668
810026122
810026034
810026168
810026196
111039114
111038900
111038724
111039587
111039107
111039269
111039066
810002890
111040355
810026449
111039089
111039635
111039031
111033340
111039057
810026164
111027266
810007218
111031993
111027400
111035869
111039494
111031876
810007209
111039124
111038881
111039243
810026297
111031695
111039043
111026659
111039660
111040353
810008181
111040431
810001108
111016404
111040481
810019697
111033992
810025214
111029491
810019806
111029690
111039534
810007028
111016371
810018855
111015981
810017137
111031652
111036107
111017722
111030689
111042720
810000150
111036975
111029732
111014963
810023311
111019642
810003147
111039502
111022877
111032069
111028443
111029328
111025183
111030387
111029326
111006395
111014804
111037842
111040884
111030220
111037075
111030019
111037454
111035975
111023161
111034316
111025853
111037746
111035827
810007106
111035882
810008782
111034684
111035754
111034607
111034724
111029205
111016405
111037068
810007038
111019236
111034410
111034711
111037904
111037365
111024764
111038612
111029674
111036101
111034408
111034335
111034572
111038675
810000971
111025974
111034759
111013399
111023289
111022452
111023215
111038977
111039324
111026734
111034698
810007030
810001714
810023196
810023011
111041068
111040798
111043173
111042628
111038588
111037445
111013376
111037752
111041201
810002133
111029221
111027272
810025843
111034149
111035750
111037761
810007626
810026237
810026111
810007621
810007615
810005966
810026312
810026172
111040864
111005615
111030583
810008816
810016810
111039486
111026730
810002644
810000590
111024271
111027426
111014456
810000745
810001444
111027379
111032104
111014359
111024806
111023156
810019838
111040283
111030241
111028543
810026428
111029523
111034642
111029258
111037564
111037901
111028669
810018860
111039006
111024929
111037031
810026436
111038690
810024062
111042108
111038734
111034679
111032972
111033165
111037187
111029200
111033350
111035811
111043188
810026468
111040888
111027141
111039273
810024213
810025379
111037361
111037639
111037850
810026490
810009190
810019708
111030260
111033172
810022810
111038979
111033311
111036048
810020151
111030354
810007133
111042056
810000215
111016453
810025329
810001825
111013202
810026098
111040832
111041329
111016958
810000143
810025320
111025976
111015320
111019146
111016379
111015998
810007289
111039096
810000163
810001881
111018183
810000176
111013171
810000532
111016548
810000223
810019115
111025841
111035838
810006296
111025720
111026486
810000205
111014511
111023144
810001768
810001884
111039259
810006286
111006254
111038890
111037905
111034658
111031776
111039414
810001824
810001862
111031644
111031805
111031722
111041141
111028461
810000552
810002749
111019064
810000356
810008772
111017140
111019057
111006283
111019912
111032048
111039202
111016425
111037689
111006265
111016476
810019117
810023034
810001739
810007259
810022695
111030711
111037436
111033404
111033351
111014747
111040918
810006971
111040547
111031602
810025217
111035804
810007097
111029251
111026661
810026190
111029496
810018781
111030094
810018865
111043140
810024019
810002937
111037596
810006294
810019824
810005883
111030337
810019828
810017101
111035863
111028547
810003111
111039265
810020951
810026252
810022876
810020317
111019190
111028439
111036883
111034215
111040518
111033206
810019905
810000492
810007198
111024907
111025003
111042663
810026513
111024884
810019798
810007147
111039474
111025260
111025894
111030370
111013176
111006014
111024092
111026083
810019110
111008507
111040447
111039295
111039182
810024344
111036851
111037390
810003267
111030598
111014790
111013146
111031684
111031568
111020463
111038663
111029621
111030232
810006970
810006978
111033066
111034484
111035796
810003281
111006841
111030138
111028575
111026000
111006804
111025257
111023225
111025359
111006882
111030226
111028671
111006874
111038852
111006315
111027429
810000510
810020044
111016027
111014418
111014164
810007183
111028427
111013217
111022482
111038923
111039065
111008449
810020029
111022457
111022464
111037554
111043200
111027520
111014782
111035939
111015333
111029234
111033177
810024510
810024243
111005763
810002879
111030291
111030047
111033365
111039379
111039459
810001599
111039301
111032043
111030129
810000867
111037211
111028557
111029293
810024346
810024179
111037768
111034125
111033338
810001655
810000870
111038682
111031638
111027343
111027333
111025175
111028607
111028559
111015943
111016443
111016961
810020037
111038619
111039450
810023286
810026328
810019689
111042725
111041344
111029658
810024088
111029264
111024954
810019817
111034547
810022684
111034235
111036209
111029603
111041396
111028614
111034629
810023176
810024177
111029362
111030244
111032001
111041127
111034793
111030227
111031781
111041130
111027523
810023340
111030321
111036167
111037373
111016464
810018847
111005765
810018777
810026095
810018689
111036053
810017105
111032049
810026008
810025245
810026114
810026048
810007080
111027366
810003277
810007319
111032086
810007628
111004981
111040350
111040468
111025006
111033017
111035977
111040817
111033191
111037737
111031736
111034643
810022641
111034635
810024236
111042745
111038945
111040470
111030706
810019121
111033462
111027273
810000605
111020714
111016973
111041069
111029211
111034307
111019635
111029216
111016175
111026759
111033369
810026486
810027457
111033194
111029325
111026509
810006963
111029686
111034172
111039261
111038695
111030272
111027222
111039308
111008526
111013383
810020943
111039206
111013198
111019049
810023012
111029364
111005361
111027231
111035909
810024379
810019105
111016478
111018145
111008414
810001598
111037045
111020699
111033254
810020215
810017108
810025501
810023135
810023525
111037715
810017104
810026245
111033435
111033294
111039024
810001499
111023150
111031788
111025235
111023295
111016547
111031750
111030265
111035860
111023752
810022193
810026162
111037614
111014991
810023332
111014101
810020315
111027253
111023358
111014102
111005316
810026191
810006113
810007116
111036019
810000424
810020320
111023284
111037808
810000915
810006123
810022681
111032942
111015987
111023278
111033161
111016134
111017820
810020273
111023277
810022664
810007283
111034769
111034755
810020930
111014965
111033259
111017731
111017737
111017824
111013358
111018210
810008212
111005315
810019660
111029748
810002912
111037353
810025346
111033274
111040847
111036986
111005039
810018891
111022795
810019133
111006408
810025847
810026250
111029330
111028668
111028662
111033028
111033051
111031997
111028617
111029371
810026029
111031906
810008951
111028666
111030700
111025178
810008944
810007250
111025234
810003252
111032989
810008775
111006477
111029703
111033057
111029772
810001608
111037198
111031796
111029372
111009101
111016590
111014728
111008492
111014754
111008533
111022793
111037117
810026039
111036129
111043219
111031604
810026064
111036973
111023274
111014104
111024911
111017000
111031690
111038823
111038680
111033064
810023179
810024248
111041325
111005787
111040328
111036175
111040499
111030041
810006289
111016434
810006475
111018339
111017744
111032130
810006292
111023795
111038888
111038736
111027224
111038893
810006460
111029656
111030364
111017815
111016411
111019223
111005405
111016422
810009191
810008880
111018137
810024051
810019911
111027246
810020075
111005363
111029498
810022998
810024044
111006431
810025485
810006939
111034082
111035972
810025378
810022843
111023625
111030027
111023251
111018138
111034517
810023321
810023041
810026010
111038741
810000917
111029284
111031765
111031787
111040815
111025721
810007189
810003138
111012792
810007077
111019056
111030342
111029509
111030601
111014506
810020300
810025489
810019887
810020935
111042661
111042119
111042658
111042718
810023160
810022643
111042228
111033420
111032083
810024249
111029212
111031656
111033304
111037003
111034269
810025813
111041349
111027369
111016599
111038630
111034353
111040841
810022680
810025495
810019899
111038861
111029682
810001770
111042649
810023003
111037574
111014440
111033250
111036890
111036041
111041320
111036994
111040266
111029554
111037767
810022200
111034428
111036981
111040517
810027342
111026054
810001560
111024239
111040315
111038951
111034794
111037653
111033385
111042140
111031967
111030104
810018688
810023550
810023018
810000418
111030589
111030680
810023197
111008541
111025980
810001761
111041358
111029530
810026247
111030115
810019811
810028480
111029424
111033138
111015330
111005033
111027320
111027349
111039400
810023302
111036011
111034177
111027492
111019643
810001411
111022631
111028565
111025243
111025353
810019804
111024738
810007215
111030649
111037858
111023234
111023382
111033136
111007740
111030075
111005021
111036126
111025168
111027263
810025997
111018188
111006047
111019079
111017123
111017712
810008209
111034131
111016374
111037606
111029578
111025933
111033109
810019097
111025142
810002114
111035687
111023385
111039605
111030087
111025217
810028614
111041348
810020940
810006948
111030025
810007625
111035887
111035631
810001755
810008713
111037441
111041102
111005972
111027215
810023288
111042801
111031848
111031740
111038955
111034159
111029679
810024030
111033337
111039165
111039569
111031872
111037565
111039233
111038858
111005963
111034245
810007203
810020023
111007252
111022468
111037382
810002400
111029488
111031869
810017174
111025036
111014303
111034415
111023200
111039448
111037435
111039230
111030667
810007286
810000422
810000458
111008517
810007051
810000439
810017180
111040737
810000105
810022687
111036972
111022892
111035775
810027351
810022813
810027373
111034791
111023631
111005555
111028464
111038878
111006023
111005262
111035911
111026762
111034301
111040879
111037688
810002120
810026330
810026223
111030143
810002701
111013158
111005785
111026081
111005780
111040550
111042089
111016552
810020059
810028297
111039146
810020026
111042226
111012918
111022893
111018202
111034641
111040792
111018361
111013140
111005927
111005923
810001724
111026744
810003270
810003261
111013143
111017786
111004973
111004962
810003139
111025891
111006390
810007207
111032992
111016466
111040902
111024775
111026769
111024739
810018871
111040914
111030573
111033069
111041145
810018893
810025372
111004984
111024746
810023273
111026779
111033137
111006466
111017784
111033320
111033432
810024020
111017752
111016345
810007073
810002604
111023260
111029709
111014351
111038759
810008214
810026057
111038744
111040715
111034678
810025809
810000533
111040725
810027441
111032902
111028625
111033002
111027234
810002721
810019688
111030263
111023745
111030283
111034466
111037184
111016048
111024892
810002689
111016107
111033968
111020450
810002583
111016151
111038805
111039346
111037857
810003090
111037855
111034694
111037613
111031858
810001105
111040551
111033960
810008784
111024480
810026121
111025967
810026212
810027382
810017102
111038808
810007213
810025855
111016383
111005560
111023602
111023609
111039283
111012745
111014752
111020416
111014521
810024335
111029792
111017766
111042107
111016446
111012746
810008295
111023616
810007113
810003278
111022471
810022826
810000349
111033991
111016155
111031930
111017117
810019854
111017755
111023596
810020007
111041072
111016137
111012749
111034097
111024106
810009167
111020425
111037907
111023290
111034162
111030596
810025199
111030118
111037874
810001744
810001822
111033006
111036204
111037780
111023612
111030682
810024188
111035702
111037453
810008663
810008683
111007742
111033167
810007622
810008197
810026036
810026086
111037658
810018773
810019101
810026163
810007619
810018674
810018774
810018671
810018678
810026229
810026033
810018849
810026052
111037736
810018683
810018684
111022484
810008669
810026101
810019119
111037664
810008198
810000769
111025246
810024252
810024341
810007008
810023004
111042232
111027186
111042754
111016980
810001402
810028300
111036012
111033044
810000404
810001391
111028449
810001746
810001773
111037848
111034006
111037146
111035807
111033328
810020003
111025914
810028299
111039087
810000949
111029756
810008877
111034699
111036211
111036854
810001406
111033286
111030350
111025917
810002884
111042790
111016119
810002747
810000386
111029207
111009091
111017132
810001752
111018200
111016326
111016990
111024474
810007320
111033410
111042138
111042093
111039501
810025219
111030051
111033977
111042100
111030026
810007103
111024257
111012784
111039558
111037082
111038664
111031574
111015329
111016074
111017804
810018907
111029634
111036891
810020928
111037375
111039271
111031785
111023579
111033443
111023697
111030326
810007143
111041062
111029589
810002455
111041359
111026627
810008971
111032123
810023184
810020119
111034144
111036257
810022686
111023576
111037693
111034295
111029630
111029455
111028646
810007247
810026049
111039368
111030389
111035902
810024332
111040803
111034465
810008946
111029277
111037391
810024187
111037100
111038655
111030092
111040560
111036162
111029744
111042122
111025865
810024093
111034744
810000940
111025874
111014530
111035764
111016356
111038790
111039075
810024060
810001887
111017740
810002646
111025722
111026033
111038898
810027447
111028635
810025200
111008431
810000204
111009113
111014789
111008393
111009126
810020939
810001728
810026517
111030252
111035938
810023030
111040746
810020330
111018175
111034086
810000789
111017735
810019812
111019157
810007262
810022191
111006872
111006885
111017772
111005034
111005018
111005036
810024193
111014810
111006855
810028589
810028605
111006845
111004998
810023339
810019800
111005025
111006831
810022175
111037580
111004989
810028616
810019116
111029439
111017703
111032119
810028315
111037729
111014563
111029574
111034161
111037640
111005611
111005402
810023131
111006889
111037561
111013317
111023187
111006449
810016803
111035757
111020706
111030254
810018868
111026664
111035834
111006398
111037063
111024910
111030054
111006655
111036860
111015989
111023371
810022706
111005553
111028664
111034569
111027287
111006433
111014428
111006422
111006393
111016096
810019089
111014326
111036270
810022847
810024199
111023364
111006472
810001378
810022987
810022807
810007152
810019685
111029768
111018178
111034221
111033357
111016176
111036847
810016809
111006392
111040473
111025207
111006267
810009163
111031783
111012958
111006249
111005019
810026183
111016090
111038770
111027407
111020449
111006038
111005003
810024259
111006851
111038768
810019797
111034565
810022852
111036215
111031715
111037669
810007604
810018851
111029238
111030028
111017828
111012924
111037355
111035808
810007061
111036831
111028525
111027173
111016578
111032999
810003106
111023258
810020033
111042728
111016553
810002653
111016098
111035942
111030294
111024098
111034019
810020069
111014468
111024901
111024266
111034145
111037833
111038669
111042732
111043179
810017166
111029215
111029412
111024281
111026537
111016063
111006269
111028618
111034153
810002710
111037586
111016572
111027316
810023171
111037703
810019687
810020051
111016083
810000456
111018204
111027306
810006965
111024890
111016470
111029718
111012806
111024990
111039689
111037775
111038609
111028624
111025963
111027415
111043264
111030038
111039508
111028582
111026741
111015298
810000457
111037869
111037599
810023150
111033098
111033085
111023772
111014816
111024737
111038573
111027404
111015971
111027535
111038738
111027524
111037607
111006426
810007256
810027360
810007208
111036176
810001730
111027325
111025971
810017185
111016983
111016344
810025864
111037644
810019683
111019214
810001357
111013157
810006960
111040360
111006830
111039188
111039240
111040458
111034695
111037450
111025876
111040563
111034688
111028424
111028452
111026012
111037156
810008672
111040335
111034789
111030642
111032977
810007280
111020697
810018857
111040336
111022887
111038651
111043203
111042077
111042076
810022805
111033955
810024196
111042623
111033958
111018147
111018201
111006410
111034626
810026146
111030348
111027519
111005403
810006990
810003143
111024783
111037092
111033336
111026031
810002252
111034375
111025802
111042043
111029507
111024791
810002336
111018206
111040825
810007138
111019216
111037820
111024230
810019711
111030329
111036147
111033078
111033335
111025829
810009150
111027412
810001372
111013374
111027264
111040509
111042762
111019899
810026141
111039651
111012947
111023622
111033173
111039603
111024472
111036445
810001758
810026154
810007316
111033987
111027269
111033456
111042789
111037174
810022997
111033270
111035927
111029552
810022988
111037567
111025839
111028522
111025015
111040310
810009203
111039616
111037342
810009169
810027352
111034489
111040308
111035851
810027439
810007071
111024982
810026262
111042671
111039422
810025840
810020085
111025035
111014473
111034148
111014809
111040271
111030704
111037829
810007092
111034087
111022894
111005949
810000192
810025392
111037699
111008399
810002877
810024339
111005778
111005020
111027409
111034595
810023190
111041378
111035770
810025210
111037866
111006837
810023214
111037602
111016427
111009098
111016328
111014308
111039321
111005327
111014313
810000896
810002688
111006667
810000212
810001066
111005620
810000194
111023676
111006266
810019860
111031873
810020286
810019882
111036870
111037750
111014724
111026001
111006041
111013190
111007770
111006015
810001681
111026655
111037683
111035867
111030685
810023297
111033453
810028312
111005952
111038576
810022869
111037861
810017171
111031999
111039064
111035619
111004995
111014960
111025169
810000058
111034263
810023536
111038862
111026757
111029700
111034785
111031904
810020326
111032050
111040482
111027646
810019639
810026161
111030347
111037800
111019632
111024900
810001742
111034017
111033970
111022459
810002940
111039209
810007311
111035934
111032092
111029659
810028612
111041366
111031791
111025253
111031888
111014334
111008535
111028537
111026768
810007244
111024113
810007122
111008438
810022204
111033463
111015324
111038550
111037865
810023151
810019715
111041076
111023170
810007243
111032953
111005399
810000400
111013342
810022668
111029280
111025995
111025292
111039475
111023247
111026656
111033038
111025241
111039344
111039670
111036866
111008428
111027267
111025281
111027336
810019906
810025189
111028585
111028463
111024883
111019071
111034530
111016456
111034367
810025846
810007236
111040843
810025342
111029520
111029182
111029197
111023713
810026494
111029297
111033122
111042094
111030022
810026277
111037778
111037677
810024245
810022856
810024209
111018146
111029528
810002187
810007083
111029522
111029358
111029183
111032948
111038814
111042715
111036265
111035966
111032017
111040680
111040758
111019159
111035941
111008523
111004983
111006284
111006004
111008509
111008485
111032089
111019053
111015941
111008440
111041111
810024023
111040277
810019682
111039186
111034246
111037735
111034767
111035680
111034429
810001504
111037393
111034212
810022174
111034366
111037619
810000705
111037073
111037569
111037118
111037335
111036187
111025247
111034693
111025144
810006451
810000140
111027242
111027204
111035778
111029273
111034674
810007100
111023610
111031963
111039320
111039153
111039067
810019809
111036130
111038731
111013205
111005024
111032059
111037196
111033132
111018211
111023585
111034710
111041373
810025208
810019850
111034455
111026556
111027378
810027366
111028526
111024975
111031594
111039246
111019068
810022206
111039359
111038697
111039452
810022808
111030112
810001807
111024245
111006035
111039611
810027437
111034259
111034281
810001829
111025892
111038767
810000872
810001833
111027634
111006829
111006031
810026327
810022214
111040796
810024240
111037702
111034630
111034513
111026638
810025227
810019710
111034341
810001609
111022616
111023607
111032897
111005565
111033273
111032970
111037843
810003105
111034387
111023570
111040760
111005557
111034315
810002129
810022201
111037389
111009122
111037049
111035853
810008774
111006827
111016585
111030678
810025374
111030048
111006252
111042224
111030264
111029546
111025208
111035707
111024889
111026650
810002677
111034187
111027532
111025798
111038600
111037058
810001953
111030593
111038828
111040833
111038756
810008777
111016057
810000977
111038779
111038872
810001404
111027142
111026637
111029281
111016104
111016407
111027470
111035696
810003146
810003148
111019072
810023329
111016454
111014096
111024505
810007146
111035828
810000084
111036180
111036122
111027372
810024337
810028290
111037871
810019805
810020207
810019118
111037601
111032929
111032906
111020683
111040769
111029749
111037742
111034108
810020103
111042665
810003102
111034248
111023794
810001442
111034623
111012904
810023320
111012754
111027360
810007044
810008273
810026107
810026097
111036108
810007167
111016054
111034621
810000606
111036075
111038722
111019121
111014529
810000551
111023303
810002602
810007252
111025251
111012790
111024490
111033042
111030303
111031955
111014722
111006448
111019052
111034218
111026665
111019140
810002337
111042809
111033193
111037745
810020049
111030341
810001379
111034308
111034792
111032056
111032098
111037686
111040356
111030683
810023272
111030365
111006068
111034285
810020131
111019043
111029540
111042123
111034294
111034434
810019839
111014459
111037827
810025359
810023045
111037090
810026239
111038640
111033075
111032898
810020014
111041090
810024016
111038558
111005968
111029292
111041356
111038614
111039329
111015974
111016171
810016794
111038636
111038587
111013327
810026281
111014318
810019851
111016368
810008314
111007261
111013149
111032132
111037341
810019085
810016808
111040744
111040907
810026520
111032983
810026156
111037060
111024255
111042741
810009215
111006892
111026676
810001612
111024482
111028453
111034358
111025272
111037048
111022620
111035985
111039681
111016152
111041202
111023216
810002674
111016322
111036110
111036174
810019872
111014340
111028670
111030259
111041105
111027209
810026254
111035718
111025037
111023302
111037539
810002335
111036311
111038876
111038654
111023146
111039369
810001396
810022857
111039549
111040710
111033963
810020209
810028482
810017164
810022825
111026518
810002251
810006435
111035987
111034501
111039428
111025152
111015331
111037442
111040267
810019947
111040367
111027428
810001707
111014778
810002922
111016415
111016367
111027488
111023688
111027637
111005759
111039195
111040460
810024759
111040436
810001611
111014764
111027513
810022883
810020122
810007200
111026003
111017745
111029421
111029407
111037396
810024036
111038761
111009106
111006241
111038620
111024755
111024920
810026181
810026315
810020095
111019628
111037751
111040361
111023268
111012921
810003279
111034118
810019898
810006998
111033343
111017139
111016977
111037349
111030566
111038773
111027432
111034633
111016010
111019192
111016352
111034686
111033954
111006034
111008439
111008538
810001718
111040867
111039436
111020423
111040428
810018844
810022829
810025814
111025223
111037863
111035921
111016587
111034165
810019093
810001859
810001785
111039560
111033014
111036153
111014446
111017760
111039559
810002340
111018348
111040330
111016417
111023171
111043148
810020018
111029317
810023294
810022224
111039667
810023293
111033449
111029392
111012777
111033981
111024942
111024481
810025481
111037864
111037130
111034727
810028296
111035976
111041199
111027442
111033082
111031576
810026180
111023288
810022644
111034538
111034808
111031890
810022666
810022685
111029642
111036000
111036897
111031757
111027630
111032016
111016954
111029785
111037611
111034478
111013200
111005933
810022861
111026746
810023266
810000222
810003115
111029777
111027411
111028529
810008668
111012926
810002121
111027466
810001705
111038757
810006447
810001679
810025815
810001809
111022873
810001846
111038799
810022704
111023301
810023309
810017161
111008409
111027490
111037444
810028319
111037134
810017184
111029759
111027473
111035964
111014799
111041148
111025991
111027317
111037802
810020105
111028588
111038578
111009124
111023766
111043249
111018174
111026624
111013369
111038564
810025475
810028608
111028643
111028645
111030101
810008943
810007312
810008300
810007313
111034098
111030140
111041203
111030376
111042788
111037173
111009127
111008417
111040559
111040701
810008283
111034392
111037815
111036001
111037836
810028602
111032018
810019700
111032107
111034778
111018177
111009074
810000193
111014317
111008434
111006673
810001491
111036830
111023759
111006061
810018862
810019668
810020932
111034540
810008302
111012794
810001732
111030225
111005757
111014427
111004955
111013194
111016011
111004965
111023760
111018152
111037679
810017150
111037717
810022223
111034110
111008503
810000818
810002711
111032944
111034784
111026745
111025910
810007084
810022824
111016064
111009085
111035787
111022611
810026257
111023169
111034661
111014433
810022804
111030016
111029769
111014358
111029396
111033215
111016962
810019902
111019155
810001384
111039548
111008520
111036043
810000709
111006028
111022463
810001607
111026505
111024473
810026198
810026272
111038832
810020068
810017177
111013226
111031925
111039092
111029418
810001856
810007614
810008216
810007605
810008194
810007606
111027197
810026045
810007607
810026105
810026058
810017160
810026264
111025028
810007602
810007603
810018775
810018677
111030323
810026016
810026124
810026126
810018682
810026100
810007633
810026004
810008217
111039418
810001434
810020213
810020327
810022646
111033999
111034652
810023146
111030277
810001821
111026561
111038865
111014755
111026675
810002647
111008391
111029606
111029573
111027230
111008415
111036205
810024073
810001432
111014515
810019842
810019857
111027171
111040257
111027635
111039086
111027321
111039606
810001836
111024956
111024939
111006020
810002888
111036895
810002898
111037670
810019114
111041332
111030574
111017818
111037446
111016970
111016052
111030224
111041050
810002573
111020395
111034552
111041205
810002451
111016062
111037331
111040848
111016989
810026236
111040822
111034797
111039351
111039440
111039375
111031763
111032038
810003211
111032023
810026091
111039431
111030666
111041334
810019793
111009095
111027387
111037681
810019671
111029397
111037674
111037622
111025978
111026005
111042229
111033079
810025187
111014558
810023202
111032115
810025825
111032914
111040904
111023209
810027363
111039636
810001875
111034243
111033326
111030320
111027468
810003099
111042808
111043226
810001849
810018897
810001113
111031575
111026658
111034380
111039015
810020292
111006029
111019091
810027376
810006944
810006979
111015929
810000672
111040561
111034333
111034309
111020452
810000142
111012759
111012961
111039168
111026021
810000811
810001061
111006459
111019129
810001814
111006428
111016101
111026758
810001832
810006943
810007263
111018176
111034632
111016610
111034475
111042091
111040572
810022995
111031717
111040443
810024519
111033195
111037403
111034378
810002685
810026519
111043207
111035781
810000074
111029694
111022603
111005321
810022709
111033452
111037663
810000531
111040831
810024511
111023198
810003207
810002934
810019859
111039676
111039490
810006288
111034186
111019212
111027166
810020046
810007057
111006850
111034592
111031808
111038940
810003257
111037019
111032927
111040926
111027367
111030570
111029399
111025909
810020048
810001559
810025396
111039040
111025259
111038834
810005967
810024520
810008303
111038918
111025029
111024724
111030316
111029353
111038786
111031675
111038802
111031855
111023246
111031895
111038931
111031885
111037357
111038956
111038981
111027398
111037705
111037744
111028654
111037841
111029214
111029235
111025263
111038785
111031665
111039073
111038936
111025299
810026021
810008320
810024524
810026017
810008286
111030280
111025264
111026039
111026085
111025040
111029257
111039042
111039003
810026023
810007155
111019191
111042088
111034535
810026440
810023035
111017129
810008989
810020132
111030278
111027511
810002582
111014760
810017109
111038599
810020301
111038594
810003259
111040818
111024486
111013221
111006024
810000591
810008958
111032055
810025367
810008953
111039101
111029204
111040461
810019643
111037219
111023180
111030366
111016457
111039166
810022207
111023755
810024241
810019096
111024096
111023154
111040870
111025186
111006816
111008504
810000079
111014796
810000100
111034206
111038618
810024053
111025020
111033070
810020139
111007253
810002692
810026469
111024959
111014984
111042626
111014429
111042053
111030602
111038554
111040800
111037159
111038616
111024979
810024169
810020332
111005406
111026752
810025817
111034807
111024095
111040901
810020010
111006314
111034479
111042120
810000330
810025321
111037839
111032960
111037555
810022994
810026296
810022806
111019081
810002702
111022442
810009214
810008980
111024809
111025814
111016362
111016110
111024090
111014437
810008292
810020002
810008957
111041073
111029269
111018166
111023217
111013210
111008494
810008675
111039234
111038746
111006854
111019145
111040490
810026518
810023295
111031596
111035839
111042236
111034625
111029684
111034529
111035780
810000671
111019158
810000528
111039339
111037345
810002857
111017122
111040420
111030247
111014484
111007777
810006439
111006814
111016017
111014560
810027353
111024745
111022882
810023155
111032984
810023165
111005967
111006809
111036015
111014999
111016168
111015314
111032916
810020094
111019904
111016557
111026038
111025007
810001886
111022634
111023282
111030322
111042245
111019142
810001490
810020305
111024921
111004987
111006288
111017776
111006825
111029426
111029506
111005038
111033204
111038546
111029416
810025381
810001500
810002491
111005365
810002454
111034600
111019641
111038674
810022653
111034120
111023262
111027508
810027430
810000670
111023299
111037083
111026047
810001885
810007223
111015984
111014442
810019675
111016956
810025351
810008793
111013225
111014462
111016321
111013203
111024261
810001837
111038870
111038845
111034124
111035690
111042638
111039574
810001423
111023710
111033271
810025225
111028458
111026786
111006258
111026646
111032117
111026082
111035960
810026001
111033099
111030249
111031562
111029288
111040562
810027424
111029765
111042639
810027418
111042121
111025907
810008307
111037847
111037786
111025279
111037814
810007159
111040917
111031951
111042147
111031829
111032042
111042137
111039156
111037680
111037772
111037790
111034113
111020415
810005881
111008497
111038883
111027145
111043172
810018877
111038882
111035817
111038887
111037537
810019909
111032896
810020952
810002131
111026787
111033041
111025827
810002929
810022652
810023137
111029299
111038793
111033092
810000152
111040748
111017001
111029386
111037069
111034591
111037124
810026434
111037162
111006003
810000707
111013159
111038688
111025245
111020413
111036106
810026300
111033409
111037638
111042761
111042787
810002596
111005561
111006463
111017763
810002639
810001709
111034213
810017183
111014731
111017826
111023770
111036993
810001107
111037635
111039586
111036080
111007248
111007247
810023298
810020120
810018894
111034424
111043209
111030281
111039517
810023529
111023682
111024456
810017144
111015949
111009116
111039002
111014365
810008981
111031643
810023192
111034736
111027445
810007176
810003206
111036845
111006000
111028553
111012965
111028545
111025179
111039035
111014962
111039341
111040767
111037334
810002687
111042733
111034476
111042221
111035809
111014787
111037819
111014814
111026756
111013313
111016103
111037906
111005351
111028630
810000195
111025164
810007034
111023387
111026543
111024735
111014730
111039215
810007035
111015963
111036267
111025274
111019187
111029184
111037405
111037806
810003104
111020446
111016336
111023157
111015973
810024378
111016441
810028469
111022462
810027399
810002402
810002704
111004975
810001952
810026465
810002690
111033128
810006995
111018358
111013380
111015968
111033342
111017819
111033413
111034409
810001876
111016084
111029529
111031694
111035953
810001845
810000420
111037793
111030647
111042799
111034585
810001140
111005766
111014978
111007744
111014974
111020486
810003201
111034094
111034805
111006042
111006394
810019680
111018144
111033461
810020104
111027424
111034412
111033348
111024980
111033189
111034722
810006474
111039437
111005754
810000530
111040514
810026302
810000493
111013322
111014435
810023002
810009204
111025166
810009177
111041084
111038907
810019894
111037407
111034178
810008783
810002494
111035695
810007031
111031663
111029799
111033016
111027338
111037618
111037657
810020312
810023291
111038817
111034606
111033110
111038913
111038672
111006051
111033998
111040858
810001556
810022862
111036886
111034272
111034745
810023193
111026065
111038932
810025237
111027265
810002450
111029735
810023037
111027193
111025824
111013213
810007025
810007055
810006445
810022671
111038869
111032037
111039488
111016059
111033251
810006433
810028302
111029646
810026248
810002709
111039472
810007227
810026334
810026070
810008279
111031721
111038864
810026200
810022655
111037401
810001166
810020055
111024265
111015955
111037448
111029775
111040556
111017790
810001659
111041112
810018888
111037830
111041074
111034740
111027642
111006859
111022450
111040876
111031720
111034608
111027420
111026022
810000808
111037027
111029738
111032912
810023006
111031953
111041399
111020384
111031996
111039279
111039382
111030408
111041388
111025204
111041360
111039185
111033436
111041363
111028600
111005002
810009152
810008700
111036203
111043195
111030564
810008945
810022971
111034446
111042780
111043137
111037748
111006286
111033454
111031712
111028626
111025250
810028599
111037783
111039544
111034656
111040747
111024945
111042772
111019221
111023252
111020684
111020439
111024897
111039158
810025325
111034009
111020433
111036885
111026628
111029438
111014968
810008686
810000225
111028532
111029306
111017002
111023172
111029425
810023547
111029575
111039272
111030715
810020275
111016582
111030336
111005006
111032036
111018179
111014461
111017787
810000081
810020324
810019684
111038562
111042115
111034201
111013357
111042040
810026089
111014721
810026050
111039403
111032129
111006313
111040369
111032085
111006666
810028493
810008298
810028490
111039036
111025293
810008674
111036872
111037408
111026770
111034262
810025803
810025767
810026335
111014979
810009165
111024888
111006661
810001723
111035803
111034247
111006857
810025863
810027429
111034719
810008662
810001835
111040896
111020396
111013193
111024722
111029754
111023792
111029722
810008325
111006662
111029334
111025294
111036838
111027198
111029741
111025149
111030024
111037404
111037632
111030238
111038841
111035771
810022636
810001757
111005956
810023535
111039511
810007242
111037409
111039624
111037074
111034002
111035981
111031585
111027254
111040282
111016539
111024272
111007254
111014973
111032975
111031947
111023236
111037747
111030127
810022172
111015933
111038944
111035822
111031835
111034799
111030162
111031659
810020117
111029176
111037552
111030723
810027432
810001068
810008846
111024725
111025937
111025298
111029401
111029321
111025163
111013174
111034683
111039152
111038758
111029707
111038789
810026020
810020097
810022711
810027383
111036992
810022656
111042087
810025246
111036146
111035891
810019795
810019827
111035694
810006441
810026081
111042033
111006473
810000200
111023604
111033394
111033258
810028307
111016418
810001735
810006463
111019625
111037543
810002708
111032033
111039025
810001855
810024087
111038552
111042057
810028481
111016072
111027148
111037781
111037718
111037213
810022999
111025851
810020011
111037380
111039187
111029448
111039449
111035785
111039413
111031849
111037020
111037557
111014498
111040496
111014327
111014350
111006402
810006975
810007036
111033406
810019104
810020021
810024067
111006450
111039022
111040459
111024759
111042636
111035846
810024208
111041357
111027184
111036171
810024035
111005268
111015937
111023270
111024726
111030713
810024260
810026267
111027450
810019678
111039223
111031636
810001671
111015319
111008387
111029611
111031845
111038707
810007214
810022678
111023182
111029428
111016105
111006893
810020333
111034250
111037202
111036092
810009181
111036090
111034482
111031671
111029490
111038721
810005969
810022872
111006445
111006049
111020448
111006056
111017747
111019077
111031728
810023335
111029553
111005027
111023267
111037071
810007268
111014768
111006405
111006420
111019226
111025180
111014347
111014542
111014553
111014471
111008429
810009164
111036009
111037348
111030273
111037177
810028283
111042796
111037875
810025492
810022993
810000151
111039657
111014108
111018184
111013192
111040702
810007174
810000553
111039410
111023581
810005963
111039310
810019881
111041114
810023017
111031779
111016121
111015326
111006424
111035826
111036139
111035969
810020140
111035801
111027483
111040739
111035971
810002395
810025386
810028272
111033988
111029604
111033418
111025815
111014976
111039192
111035719
111005349
111032986
111024789
111031686
111028605
810025341
111024927
111027241
111031850
111024747
810002861
111025732
111005350
111043171
111026070
111037787
810019825
111019903
111017781
810000083
111040342
810001786
111004957
111035685
111005258
111016029
810001726
111018141
810007254
111039257
810007278
810006116
111041077
111034014
111025832
111039038
111042124
810025860
111036021
111039128
810025198
111026649
810026448
810025307
111037139
810006347
810027410
111025210
810022812
111012737
810008228
810027436
810022640
111028434
111036444
111035630
810025816
810006444
111025187
111024101
810019709
111029431
111037009
111031680
111028454
111035879
111029352
111013139
111040757
111033091
111023679
111008500
111007257
111039398
111032998
111039602
810026437
111015928
810006341
810025851
111013340
111013195
810026131
810025768
810026325
810024084
810022969
810008997
111020466
810002191
810026464
111013207
111029669
111033392
111042669
810023169
111037895
111025715
111022454
111034639
111026653
111038685
810005877
810000367
111034456
111037011
111027514
111039126
111014523
111039535
111030380
111005971
111033168
810020091
810000414
810001408
111012955
111015315
111033396
111040476
111035983
111029203
111030125
111039095
111040683
111029395
810024039
810022980
810022822
111006453
111017789
810002339
810024080
111025882
111013355
810007093
111040797
810026158
111005962
810007101
810001416
111031838
111032093
111024458
810007042
111038821
111013363
111025367
111025287
111008388
111014423
111038847
111041368
111041104
810001716
111004956
810026165
111024963
111030407
111036201
810001388
111038791
111029632
111030393
111030229
810022171
111034449
111029316
111028672
111040303
111042714
111040568
810006955
810007102
111034675
111031683
111019134
111019149
111019174
111029444
111034190
111022879
111006836
111038723
111037337
111009086
111024459
111024741
111018343
111031687
111024785
111029404
111024758
810000211
111008395
111037047
111023783
111015934
111006884
111024798
111033441
111038665
111034809
111040457
111040495
810008682
111007767
111005362
111019062
810002724
111034344
111034383
111022791
810018687
810022658
111029447
810020050
111034601
111039120
810022674
111034952
111038846
810026066
111033093
111040735
810025202
810025861
810000203
111025866
111019189
111023155
111040761
111016136
111034011
810017139
810001173
111027247
810008867
810003101
111038565
111030056
111038653
111012782
111031696
111030021
111013186
810008822
111024913
111039479
810020035
810027387
111041123
111025706
111036023
111035614
810007323
111025252
111006660
111037792
111025242
111025285
111024461
111025902
810028601
111029698
111016974
111029351
810025363
111030727
111023589
111042101
111023201
111028608
111019080
810001076
810026117
111039464
111028539
111036263
810019799
111042805
111041319
111037061
111037347
810022202
111034251
810019666
111035835
111035774
111040510
111023232
810007271
111041397
111039094
111037634
810001557
810006115
111039373
111038894
810023299
111031902
810022698
111025849
111038765
111041146
111031790
810027378
111034717
111033106
810007166
810001412
111017798
810001762
111022614
111019137
111016164
111014972
810025326
111029672
111031759
810002917
810008954
111029405
111038908
111032947
111030725
810000488
111033269
111027185
111024971
111038769
810000141
111007263
810022860
111016076
111026492
111016106
810006429
111042797
111030373
111009117
111023588
111023717
111033000
111024926
111008516
111033096
111037406
111027636
111026735
111042129
111039571
111027251
111023359
111005570
111030255
810001774
111040314
111040802
111035769
111030384
111037333
111035704
111036874
111035974
111037583
810020017
810026113
810026136
810026099
810026173
111031843
111013183
111037823
111006256
810024206
111028540
111039417
111030034
810023258
111005934
111028636
111029710
111033120
111040254
111033253
810026260
111016953
111006469
810006446
111040440
111025224
111029519
111024932
810026271
111034472
111029473
111024944
810007021
111032053
111029711
111005755
111028542
810000814
111027441
111039614
810022996
810008677
810019895
111035955
810022832
111027351
111036133
810022818
111025232
111035832
111037115
111039514
810008715
111033267
111037540
111038911
111023577
111029443
111040795
111037193
111025966
111029446
111031738
810019876
111016042
111031725
111026774
810008942
810007142
111016988
111028631
810007053
111027207
810007299
111007741
810001760
111008511
111024916
111039693
810028597
111040569
111035691
111040503
111013318
111004967
111017704
810007624
111038581
810019100
111033341
810020933
111024778
111019633
810007024
810003203
111034192
111027522
111016085
111022449
111029224
111016112
111034360
111025023
111032973
111027518
111030161
111042655
810020925
111018156
111008412
810019691
111033255
111038615
111034200
810027449
810026046
111040882
111034287
111037166
111030258
111033026
111027413
111037030
810022978
111040897
111023307
111034414
111041107
810018845
111030730
111038625
810023208
111039466
810025380
111030655
111016432
810007219
111042758
810027423
111040430
810019690
111036989
111034405
810027434
111040690
111018360
111030050
111040791
111041066
810019796
111040492
111025137
111042722
111009073
111035861
111040679
111040675
810024174
810002720
111033245
111035957
111037803
810000346
111036987
111042760
111035763
111036842
810008311
810001420
111016576
111036191
111035596
111037067
111036097
111038975
111006022
810019869
111036200
810022180
111034336
111036978
111037179
111039469
111041312
111029758
810023013
111034637
810023007
111038745
111040571
111037834
111030687
810028276
810026169
810018779
111039516
111034000
111037195
111031958
111037560
111033951
111032934
111034320
111036898
810024173
111030098
111029701
111020399
111042046
111040674
111035931
111042784
111029485
111039292
810026041
810024263
111037542
810024028
111042786
810022645
810022712
111034706
810016795
111039619
810019646
810025384
111017717
111030392
810022839
810022694
111029494
111041318
111034321
111041315
111039679
111037578
111030673
111031701
111030669
111040281
810024185
111030298
111030242
111039665
111037064
111029747
111034431
111034423
111034620
111040811
111034318
810008211
111031700
111031965
111029517
111029366
111029526
810026280
810024242
810024017
810020070
111041085
111041088
810009002
111031833
111037573
111037581
810023330
810025358
111039525
810026106
111039334
111039290
810022972
111039457
111036308
111039364
111039546
111025238
111037801
111039078
810022863
810016806
111037625
810018887
111033321
111034133
111039266
810025373
111031754
111024475
810007126
111035793
111035715
111024727
111037908
111037779
111028655
111027505
111022849
111037672
111032109
810020032
111034402
111043152
111039552
810020125
111037550
111037367
111034425
810020143
810028278
111038660
111034362
111037145
810018869
111040714
810006964
111033334
111025997
810017159
810006972
810017138
111038849
111038934
111036157
111035789
111032101
111032081
111038880
111034586
111034258
111034453
111034181
111042082
810025836
111033405
810024239
111030690
111033333
111035858
111033083
810024074
111037007
810018892
111037807
111031598
111036836
111039062
111040730
111027286
111033381
111038904
111023683
111028536
810000916
111016419
111006876
111035700
111041379
810026496
111034322
810001167
111039439
111039371
111042757
111033969
111035916
810015268
111035970
810002933
111035772
111016959
111029348
111006803
111020690
810007296
111033192
111037057
111034758
810024333
111042231
111023704
111035609
810000378
810000389
111006411
111031654
111036207
810027412
810023304
810001421
111014481
810024534
111039377
111034506
810019841
810026339
111042653
810025238
111035929
810026479
810026329
111039178
111040433
111041204
111031761
810025798
111034204
111030411
111039045
810018859
111016591
111036064
111034770
111020441
810019650
111034081
111014532
111032887
111039083
111025176
111014736
111033117
111040824
111033164
111007749
111025221
111033040
111032997
111034100
810022853
111035692
810025239
111008479
111036887
111022802
810020927
111024999
111012783
810001112
111015335
111029271
111026508
111008468
111028468
111040705
810022814
111029268
111037897
111025825
111037002
111030413
111035779
111038699
810026322
111037734
810002883
810020307
111025864
810007118
810005878
111029586
111016139
111014756
810002577
111015972
111016460
111014528
810001368
810002859
111019092
111023286
111008384
111014510
111014980
111022804
111027140
810002276
111014769
111027499
111006248
810001361
111014349
111016053
111029285
810028595
111035720
111037828
111039367
810025241
111033430
111033450
810026446
111034324
111034288
111034363
111026784
111019634
810006956
810018846
111037667
111012913
810000415
111009103
111038583
111030400
111039518
810019908
111022453
111028604
810019873
111015321
810001660
810020001
111034557
810020281
810020066
111040908
111030636
111023693
810000343
111005772
111015967
111034498
111029217
111040827
111031628
111040442
810008322
111005762
111040862
111016124
810002242
111005562
810000489
111023764
111033163
810027394
111037111
111040263
111007775
111030349
111037759
111038559
810023248
810027409
810013996
111003302
111001849
111019915
810005960
111002776
111001477
111042238
810011811
111036145
111029696
111033094
111014564
111042795
111030635
810025242
111029791
111017805
810008821
810000942
111033029
810000136
111016421
111034393
111006864
111016549
111005556
810024184
111018182
111017125
111019164
111037113
111014739
111006861
111039476
111039342
111029246
111042717
111030406
111012785
111038666
111037147
111016079
111018172
111013375
111012898
111031840
111013330
111029180
111019213
111033363
111019205
810019123
810026221
810000708
111024994
111041065
111027385
111040261
111027225
111036271
111037660
111033188
111027460
810007065
111039673
810023158
810002578
111014493
111033440
111039322
111034328
111024912
111027452
111031921
111032060
810001371
810000676
810023305
810019649
810002117
111039150
111013346
810026166
111037105
810007598
111033278
111038737
810028321
111027401
111036880
111037144
111030240
810023276
810025391
111042103
111036856
111032054
111009110
111026635
111034666
810008280
111034312
111034483
810008281
111027364
810019844
111008466
111015966
810026258
111029331
111024121
111016127
111012758
111028602
111033261
111014561
111034275
111027498
810002894
810025353
810001882
111014541
111014793
111008514
111017120
111019059
111020421
111024958
111015997
810007160
111016128
111035820
111016081
111014994
111027371
810003112
810006448
111032118
111042825
111037374
111012959
810001715
111024776
810006430
810006466
810018672
111024993
111017764
810024040
111034502
111027302
111030694
111024754
111031985
111030585
111016463
111038650
111034257
111023192
111033978
111016556
810020074
810025364
810001502
111013211
111020465
810028479
111033140
111029361
111030588
111023235
111041087
111030648
111030379
810018675
810026104
810007104
810003087
111037040
111037399
810026123
111020447
111032102
111029657
111001770
111001767
111021199
111000913
810013961
111008878
810010468
111001128
111022736
810006073
810006056
111001758
810006072
810011640
810013631
111021187
810005100
810013554
810022760
810013923
111001692
111001361
810009612
810013613
111001424
111001401
111021414
111001406
111001167
810023296
111035699
111035710
111041118
111037763
810009897
111001844
111001788
111001896
111010427
111001457
111001602
810013980
810012105
111001437
111001834
111000900
810010450
111001143
111001755
111001906
111001894
810026423
111021216
111002576
111001708
111020328
111001440
111001738
111001422
810006064
111019531
810004938
810013569
810004792
111010572
810015273
111021087
111021510
810006059
111021628
810005097
111002623
111002235
810010554
810012102
111021617
111002127
111000901
111002679
810022737
111021509
810026420
810012960
810013967
810022633
810006012
810012693
111000865
810014169
111001725
810023244
810014125
111000919
810013984
111000885
111000879
111001710
111001772
111010619
111001804
810023559
810015275
810015278
810015263
111001456
111001386
810015262
111001604
111021022
111010616
111020181
810026365
111001613
111001397
111001455
810022726
810012100
111000898
810007043
111036213
111027425
810023530
810025354
111036037
111036199
810003088
111033421
810020038
810023252
111001753
111010587
111001144
111002584
111010590
111020257
810026369
111002909
810009606
810023229
111001750
810012969
810013918
810010424
810022752
111013328
111029556
111002119
111027382
111006055
111034763
810026078
111002128
111001472
111020353
111021417
111001942
111002768
111042624
111016605
111013334
810028494
810003183
111037097
111040515
111033379
111030219
111023259
111034254
111008420
111034174
810001725
111004966
111008487
111006807
111013335
111015939
111037892
810002910
111030132
111040868
111038909
111032921
810018780
111032910
111014330
111033160
810002908
810022185
111025787
111030651
111029413
111033386
111038747
111019231
111040686
810025193
810017181
810016799
810016816
111025192
111040759
810022169
111038928
111033306
111040363
111040280
810000715
111023615
111016320
111037383
111005366
111034563
111035722
810024175
111030599
111030216
111029597
111037741
111034659
111037902
810009142
111030052
111033209
111016115
111006016
111019060
111030343
810020063
111042079
111030290
810025185
111027469
111033383
111027418
111029678
810007601
111034593
810028600
111038657
810007616
111030720
111035765
810007039
810023334
810024247
810006459
810024342
111026645
111014987
810019676
111030236
111041354
810019822
111016015
810002332
810001359
111024972
111024768
111024998
810024526
111005612
111005629
111005622
111025167
111033301
111039615
111035768
111035616
111038641
111033103
810024343
111040785
810001358
810024533
111040694
810024512
810024516
111040774
810024513
111040682
111041083
111008547
810000382
810028475
111022623
810026497
810000377
111031676
111023624
111035841
810000402
810019829
111035862
111036991
810026215
111042090
810026060
111029260
810026054
810025368
111034117
111033985
810025349
111009099
810001844
111012796
111031979
810020089
111013367
111034437
111024948
810006471
111016577
111006465
111037051
111027260
810008787
111036071
810019092
111037072
111030312
111032013
810000712
111035744
111037036
111039047
111034649
111030212
111035728
111039597
111041053
111036116
111042037
111040772
111017800
111029721
111030714
111012948
810024347
111014507
111027218
111025360
810020316
810020086
111014425
810023188
111038798
810009012
111026051
111036114
111036837
111030062
111035932
111029248
111025965
111024962
810008970
810022871
111039604
111032935
111038997
111040472
810022844
810007276
810019131
111030345
111026512
111023635
111040446
810023173
111032993
111014453
111030117
111029469
810023022
111005354
111038863
810000701
810008689
810008680
111029408
810003145
111036076
111035842
111037884
111034594
111034670
111039579
111026089
810023019
810028471
111014526
111027213
111016472
111034183
111023184
111008482
810006119
810028285
810006479
111014805
111036881
810025240
810006109
111027257
810002893
810001860
111014741
111027318
810016820
111036858
111038711
111028651
810000153
111016987
111016034
810008190
111008478
111008486
810006958
810006950
810000086
810000392
810020212
111037330
810022657
111014328
111014463
810001614
810020088
810028487
111037358
111039387
111039557
810001665
111029373
111024805
111040463
810026030
810025316
111034731
111033121
810007050
810025369
111036269
111025215
810027384
111031974
111025233
810023542
810026470
111033280
810022855
810003113
111039305
111034786
111037210
111035854
810024083
111037544
111038551
111023191
111029275
111016437
810001668
111034617
111033287
111029605
810024203
111036104
111032030
810003264
111024906
111025258
111024815
111005777
810007157
111037377
111024925
111034310
111006457
111013215
111039489
111032926
111018194
810023154
111043194
810008710
111023238
810022670
111013393
111005355
810003116
810019868
111030637
111039026
111030297
111043258
111030731
111031795
111005004
111030681
111030049
111016477
111016002
810008853
111039007
111027641
111034167
111019210
810001751
111036196
111014364
810007032
111035623
111015994
810018863
810017167
810017107
111030287
810025829
111006478
111040877
111040804
111040362
111032015
111040764
111027203
111026564
111034388
111040873
111029289
111031882
111029384
111033378
111014470
111034202
111026773
111014555
111031928
111014520
111042657
111042660
111042668
111006037
111042058
111042034
111040365
111037754
111030388
810000570
111016343
111037018
810023199
810026129
111031836
810024186
111034771
111037889
810015272
111002677
111001761
111038784
111001874
111003917
810015239
111002780
111010582
810012682
810023249
111003894
111001185
810013565
111020371
810010582
111003884
111002786
810026357
111001384
111019520
111001435
810018567
111019528
810012935
810015570
810006082
111020374
111020190
810012115
111001615
111019522
111001423
810009797
810006045
111002508
810013930
111021416
111003200
810022611
111003201
810006125
810011679
810011670
111001932
810006069
111001147
111001737
810009783
111001798
111001183
111003759
111001907
810013571
810023243
810022767
111021619
810011703
111001717
810012966
111002316
810022759
111019523
111010392
810014157
111021217
111001851
810011852
810011823
111001807
111001420
111001176
111002507
111001843
111012502
111001415
111001786
111001757
111002760
111010589
810022619
111001195
111000925
111021410
111020161
111019519
111001845
111021624
111021496
111001404
111001363
111001891
111010588
111021601
810013891
810012955
111001400
111001469
111010216
810023242
810013625
111001846
111001693
810009785
111001698
810015237
810013912
810026346
810014106
111000906
111001814
111000909
810009635
810009756
111002755
111003291
810011842
111002566
810026372
111021415
111001619
111021380
810023254
111003309
111002895
111019917
810026368
111001695
810010558
111019516
111020160
111001718
810026234
111035979
111021728
810022721
810010490
810013643
111002859
111000910
810013577
111001600
810026393
810012942
111001817
810023233
810020034
111034742
111035877
810002594
810001426
810002641
810023269
111023789
111030567
111032913
111017771
111023630
111033074
111025717
111023620
111016609
111028667
111006040
111006276
810025811
111033013
810024171
810006287
111016086
111029301
111037443
111016348
810028280
111040830
111038753
111014355
111019237
111040780
810024237
111034589
111037379
111003296
810018597
111021123
810023240
111001450
111001785
111021183
111037887
111029304
111026727
111030576
111018187
810022177
810000897
111027332
810002882
810000973
111034115
810026422
810013898
111000902
111020359
810006146
111020177
111002910
810009943
111001592
810011641
111031900
111019194
111029763
810015259
111031764
111035890
111001146
810024026
111033001
111012565
111001696
810012093
111001622
810013910
810026305
810025998
111001381
111021492
810006135
810010442
111001729
810005098
111001922
810013946
810026063
810020110
111038873
111035865
111040256
111033263
111034266
111036899
810010574
810022764
810026386
810022783
111042061
810026511
111034325
111040455
810027407
810013901
810005668
111014361
810023265
810025494
111037121
111029617
810026102
111001706
111042042
810002401
111027495
111023304
810001813
111014485
111023597
111016376
810002902
810026442
810023532
810001062
111038658
111035622
111021389
111021625
810023112
810018582
810012117
810023159
111039013
111039515
810026516
111030708
111037191
111042767
810023306
111033166
810023331
111031896
810024065
111040549
111036968
810019871
111036173
810002855
111033054
111025027
111014758
111023768
111039291
111025855
111014818
111016131
111029393
810022691
111020187
111001813
111001837
111002246
111001808
810012774
111001181
111001821
810006105
810006352
810022798
810009760
810011644
810011689
810011710
810006145
810006269
810004179
810013786
810026403
111020258
111003195
111002249
810013615
810006130
810010653
111001605
111001155
111001372
111001852
810013963
111001736
111002236
810013585
810026375
111020162
810015276
810026395
111001723
111001191
111002683
810011616
810011707
810006065
810014127
810013867
810004977
810009753
810012109
111000939
810011810
111015334
111001956
111001714
111005569
810007139
810016819
810022762
111016455
111024728
810023026
810001059
111027357
111024489
810020135
111001131
111000868
111020311
111010422
111001825
810022757
810023238
810027396
111001703
111001747
111019498
810014166
111001713
810013965
810013562
111001766
111002234
111001196
111040732
810026170
111027463
810025812
111015958
111025288
810002819
111001745
111002310
810026261
111038606
111040432
111025884
111041071
111019161
111037376
810014150
810013899
810013909
810007029
111001462
810026364
810026352
810014000
111001806
111002898
111001387
810022800
111037449
111002765
111001841
111001731
111006050
810022815
111023242
111035885
810023174
111020693
111027381
810002856
810023204
111027396
111018158
111027308
111037608
111029320
111008461
111024991
111040708
111040819
111005960
111040570
111001612
111030728
810000329
810002245
111023712
810028289
111031809
111034397
111027482
111006873
111016603
111026067
111023613
111037584
111037062
810023290
810026371
111029579
111038639
111008456
111016420
111007759
111030331
810026477
810020214
111031912
111029482
111034114
111042779
111031966
111029409
810012782
810023232
810014140
810019794
810022166
111031758
810014013
111030154
111002314
111013325
111042081
111039354
Neutrino:
Model Number: NxT HF CRT-D REF CDHFA600Q
UDI-DI Code: 05415067034359
Serial Numbers:
111022069
111019302
111018899
810012879
111018587
810016858
810017086
810010751
111018970
111018966
111017647
111017444
111021311
810024542
810028044
810028767
810016856
810014201
111017618
810028037
111020776
810010726
111019291
810023346
810017059
810014252
111019676
111017476
111020301
810024573
810024559
810016825
810012675
810012677
111019980
111017656
111017461
810024894
810017056
111018736
810014225
111017617
111017421
111017442
111019101
111018971
111019275
111018537
810024913
810027907
111017443
810021008
810020475
810014198
810017058
111019303
111017480
111019870
810024908
810027897
111019096
810028455
111019180
111017634
810028443
810017064
810027922
810027932
111021858
111019290
111017484
810027911
810027895
810027915
810016824
111017485
111017680
810020484
111018975
111021654
810014224
111017681
111019866
111020206
111020223
810016865
111021293
111020225
111021894
111021768
111017665
111020018
111017606
111017438
111017478
111021781
111018718
111018586
111018958
111022068
111022070
111022065
111018896
111019184
111021862
111021761
111021857
111021890
111017482
111021888
111021765
111022062
111021891
111021878
111022067
111021897
111022061
111021774
810028792
810028444
810028446
810028780
810012873
810012666
810014234
810012680
810010708
810015899
810028401
810024876
810010721
810024922
111018016
111021771
111019861
111020015
111020006
111019669
810010768
111017699
810017047
810005199
111020215
111019675
111019677
810028457
810027928
810024887
111018014
111017674
111018015
810010774
111019857
111019986
810024544
111019183
810013721
111021298
111019989
810017046
111021301
810010748
810010711
111017428
810014205
810014303
111021893
810015218
111018711
810010732
810016841
810017044
810014220
810015219
810016844
810012880
810010702
810017188
810014272
810014290
810014292
810016970
810013018
111017445
810015217
810016850
810017191
810017069
810017189
810017068
810016862
810016842
810014208
810028449
810016829
810017200
810016854
111021873
810024545
810027921
111020299
810028819
810027913
111020019
810017090
810017072
810017194
111021655
810020723
111019983
810013020
810004068
810013708
810015312
810015416
810015922
810015902
810003600
111020013
111019997
810003658
810004257
810004888
810013716
810024924
111021305
810024885
810018938
111018898
810014296
810004328
810003876
810014242
810028461
810014186
810028100
810027924
810028784
810027887
810014275
810024547
111017689
810014211
810004624
111022058
810028422
810021006
810017207
810024874
810024579
810024896
810018919
810020157
810015460
810028098
810020219
810024897
810020218
810027804
810027748
810028423
810023360
810027792
810027790
810014227
810028405
810020461
810028407
810014248
111020770
810020221
810028400
810027746
810027787
111020230
810028430
810028433
111019293
111017637
111019667
810016864
810016832
810016821
810017192
810017187
810028408
810028424
111017422
111018909
111018726
810003719
111017672
810020479
111019864
810004069
810014243
111021772
810012678
810012859
810028398
810014260
810003873
810023313
810014239
810024921
111018727
111018976
810013694
810027800
111021867
111018738
810018932
810018923
810020460
810028080
810016859
111017669
111021644
810018911
111019682
111019809
111018540
810023373
111018588
810003598
810004745
810014218
111020296
111018959
111017624
111018749
111021642
810015459
810028397
810003781
810028089
111017664
111019179
810024895
810028049
111018740
810020465
111020792
810020463
111017440
111017694
111017697
810013701
810012674
810017091
810028440
810027737
810021018
810028017
810028095
810028066
810018703
810016632
810016846
810016863
810028090
810028075
810028043
810028030
810027889
111019686
111018967
111018730
111020778
111021299
810014099
810014238
111018539
111017685
810014181
111021640
810003874
810004128
810004066
111017684
810028086
810028102
810003661
111017610
111017701
810014267
111018744
111018545
111018597
111017488
111019285
810013702
111021308
111021871
111022060
810023350
810018700
810015311
810015923
810016834
810004264
810014314
810015461
810004261
111019865
810013011
810004259
810014276
810003601
810013692
810027719
810013004
810015315
810004133
810016847
810016838
810003720
810024546
111020782
111018962
810004130
810024587
111017666
810003451
810014262
810004129
810015414
810015420
111019690
111019095
111021235
810004258
111020216
810028402
111019854
111020310
810003910
810014288
810014193
810004965
810003906
810016695
810014210
810015397
810014195
810024920
810010740
810004005
111018912
810014206
810015897
810015455
810024576
810004132
810028071
810027711
111018592
111018707
111020796
111019277
810004263
810017042
111021882
111021875
111021877
111042876
111042824
111042821
111042869
111042867
111042819
111042865
111042875
111042866
111042872
111018731
111018715
111018742
111018584
111042820
111042873
111042864
111042870
111042874
111042868
111020797
111020298
111020211
111020290
111020284
111020292
810028771
111019684
111020224
111020775
810004330
810004004
810017088
810020467
810027803
810028064
810016830
810018697
810015909
810028437
810024540
810024538
810028105
810003721
111020023
810024907
810024902
810003597
111019280
810027912
810010762
810024891
810027902
810014274
111019856
810018909
810004622
810014187
810004620
810027721
111021309
810027751
810010772
111019283
810010760
810015216
810016861
810027760
810028442
810024567
810018705
810028438
810024914
810024911
810028818
810027740
111019281
111019289
111019689
810027707
810028428
810015409
810013001
810013016
111021646
810020220
810015402
111022063
111020002
111019808
111017615
810018934
810004621
810024879
810028810
810013712
111017448
810003871
810013714
111019292
810010747
111017660
810018917
810024886
111017486
111020236
111021304
111019282
111019683
111019672
111019296
810010754
810028027
810028040
810003909
810015898
810015924
810004626
111017481
810024883
111019673
810024555
810018698
810024590
810018699
810028409
810004323
810024556
111019693
810023352
810024916
810028783
810027917
810016691
810023355
810027749
810024889
810027774
810028046
111018542
810020473
810015446
810010765
111018021
810028047
111018908
111018741
810003905
111018961
111019094
810017043
810028026
810023368
111017605
810017051
810028057
111017679
810004322
810014241
810004327
810012742
810017204
810017080
810005198
810005196
810015410
810015411
810021030
810021009
111021236
810028038
111019867
111018894
810014188
810013713
111020294
810024549
810028036
810024904
810017045
810028431
810010738
810015441
111019862
810018702
810018691
810015467
810018704
810018937
810018924
810027728
810018929
810018920
810018930
810027745
810027714
111019850
810027712
810015462
810018696
810024537
111017675
810021023
810014244
111020791
111020222
111020212
810017202
810017097
111020306
810010715
111020771
810028835
810004806
810010745
810016968
111020798
810012673
810017048
111017430
810004969
111018536
810024552
111017492
810014285
810010707
810016967
111019103
111022057
111017464
111017686
111019685
810003864
810020488
810014301
810024903
810014298
111022066
111020293
810015394
810014251
111020014
111021896
810012708
810027765
810028845
810024572
810023366
810023364
810027739
810024875
810028068
810020485
111020788
810004265
810024586
111018721
111022064
810003866
810003862
810021007
810020477
810018931
810010753
810014204
810028404
810027796
810024917
810028072
810010728
810027776
111018714
810004070
111020790
810014289
810017075
810017063
810027770
810028069
810027781
810010752
810027733
111017657
111019110
810021029
111017489
111017652
810014111
810024893
810027747
810024923
810024535
810027758
810028768
810024905
111019979
111018544
111018750
810004260
111018713
810003865
111018594
810012863
111017621
810004001
111019998
111020229
111017419
111017459
810013706
810015220
810015907
111019298
810014264
111020227
810005138
810010734
810004885
810004262
111021764
111019810
810010743
111018589
111021294
810027726
810013707
810024551
810028024
810013010
810027753
810016826
810028025
810028403
810027750
810027773
810015392
810024574
810027802
810014202
810027801
810027764
810027766
810027741
810010705
810024919
810023378
810016631
810003779
810028032
810004064
810004625
111020289
111021651
810012754
810013017
810018913
810017055
810005041
810005044
810005043
810018922
810028028
810028022
810015904
810015906
810017096
810015901
111018596
111017659
810014096
111020777
810018915
810014232
810028084
111017614
810027954
111020302
810003998
111018969
111018904
810015456
111018709
810016694
810023371
810028062
810012858
810012867
810003872
810013014
111020304
810012878
810012676
111020009
810016823
810028050
810018933
111020237
810013697
111021885
111019112
111019806
810005197
111018732
111018739
810004967
810015395
111020772
810028048
810024563
810024536
111020203
810003777
810023312
810018693
810017050
810015405
111019175
111019992
810012749
810023369
810028041
810027944
111019994
810015913
810012861
810027799
111020228
810028465
810016828
111019678
111019691
810027788
111018907
810027908
810028806
810016848
810028082
810010723
810027946
111018593
111018965
810016839
810018921
810004003
810004804
810010719
810012998
810004006
810003660
111019852
810010714
810010703
810010701
810016857
810024877
810014185
810028054
810028065
111017646
111019985
810027955
810003599
810020478
810003904
810021010
810024541
810028782
810020476
810024557
810012744
810012856
810010736
810012874
810012868
111019287
810012860
810012753
810012745
810004966
810004968
111018974
111019114
111017454
810012996
810010755
111019099
111017418
810013007
810013005
111019181
111021296
810010713
810010776
111017468
810010777
810015457
810028452
810028448
810010739
111017682
810004000
810028774
810004746
111020003
111019813
111019681
810028436
111017467
810012709
111019294
810013718
810015419
111020011
111017623
810027891
810010761
111017673
111017639
111017450
810024583
810024585
111017630
810004131
810004324
810010763
810013006
810004886
810020482
810003807
810016693
810016634
810016635
810012667
810024561
111017643
111017635
111019822
810014231
810014246
810014308
810014212
810028053
810028085
810024592
810027732
810028103
810015469
810017062
810028458
810028101
810028800
810024591
810023361
810015458
810020466
810024570
810024550
810027920
810024577
810028441
810027894
810012869
810028052
810028034
810027772
810015453
810014311
810021019
810028045
111021863
111021860
810012877
810012876
810012862
810013699
810013698
810013719
111021881
111021865
111021775
111021770
810017094
810017061
810017079
810017041
810017190
810024880
111042822
810015464
111019100
111019106
111019118
111019108
111019860
810024892
111017677
111019679
111017611
810015470
810010750
810012746
810010722
810027757
111017692
111019674
810015400
810023349
810028797
810004331
111017625
810024915
810028104
810012854
810024539
810003810
810014250
111017452
111021295
111019300
810015918
111020303
111017627
810010756
810020486
810003875
111018747
111021650
111019113
810013022
810012743
111018541
810016852
810027717
810010729
810024562
810027937
810004325
111017603
111017653
111018538
810016835
111018712
111018734
111018745
111018728
111018723
111018729
810017198
810017210
810017203
111017497
111017423
111020281
111017451
111017439
810028099
810028078
810028092
810028096
810028079
810028091
810028074
810012853
810012751
810012851
810012855
810012852
810028087
810028088
810028097
111019868
111019991
111020221
111020004
111019988
111017435
111017632
111017651
111017676
111017471
111017455
111017446
111017437
111017434
111017447
111017456
111017495
111017433
810023356
810023341
810028055
810003659
810015398
810014110
810015468
810012747
810016827
111017619
111019116
111019102
111017426
810004266
810015418
810015399
111018746
111017475
111019177
111019817
111019859
111018968
810010733
810012739
810012668
810014281
810014184
810028808
111018598
810028838
810028805
111018751
111020785
111020285
111020286
810004065
111019858
810028840
810028839
810014235
810014209
810014300
810014269
810003997
810014190
810014207
810010737
810010730
810012857
810012866
810023367
810027709
810027798
810015445
810016840
810028790
810028804
810014304
111018590
810028439
810015417
810014305
810014258
810014222
810014191
810027951
810010716
810013015
810003717
810028435
810020487
810012710
810012748
810028056
111019304
810028427
810015448
111021864
810014277
111020242
810024882
810023351
810027777
111019098
111020297
111020232
810012875
111020300
111018719
810028464
810014216
810021017
111019855
810027884
111019987
810003811
810028789
111021892
810004889
111020233
810028786
810017098
810018916
810028822
810018926
810027724
111019276
810014203
810023344
810003863
810010712
810015450
810028781
810028773
810013709
111019299
810027756
810023342
810023357
810024578
810023345
810028063
810027900
810024898
111017472
111017466
111017671
810023343
111017645
111017483
111017609
111017629
111017633
111017642
111017616
111017691
111017490
810024580
810027795
810027923
810028453
810028454
810027906
810023353
810024582
810023314
810027926
810027909
810028031
810010759
810015403
810015401
810027892
111018710
810028020
810003867
810028451
810015443
111019115
111021297
810027763
810027933
810028106
111017479
810014097
810028093
810027734
810028406
810020722
810014299
810020468
810015406
111018591
111018743
810024575
111017667
810021032
810028051
810021012
111022059
111019176
810018692
810021028
810024900
810023359
810028035
111018973
810010717
810018695
810021013
810021021
111017661
111017695
810021027
810023348
810023347
810020474
810027785
810020480
810015463
810014197
810027755
810015415
810005136
111014063
111014078
111014082
111014087
111014081
111014088
111014067
111021900
111022226
111014089
111014075
111014072
111014079
111020280
111022178
111014083
111014084
111012821
111012816
111021903
111022225
111024167
111024163
111024160
111013534
111025962
111019264
111013535
111013536
111013533
111024166
111019260
111012825
111012832
111012822
111024165
111014085
111014091
111014074
111014064
111014070
111014066
111014086
111018505
111018491
111019258
111012831
111012810
111012835
111012819
111012818
111012814
111012836
111019263
111012812
111018508
111018509
111018504
111018506
111018502
111018503
111012837
111012813
111022819
111021901
111016333
111012823
111012820
111014068
111012834
111012808
111012809
111012830
111012824
111012815
111012829
111012811
111014065
111012826
111014077
111012827
111014092
111012833
111012817
111019261
111015826
111022818
111022817
111022810
111022811
111022814
111015827
111024161
111024168
111018496
111018494
111018492
111018493
111018498
111018497
111018499
111018500
111018501
111019259
111021902
111021899
111024164
111022815
111022813
111022812
111022816
111018495
111018510
111022224
111025954
111025957
111025960
111025961
111025953
111025958
111025959
111025955
111025956
111018507
111019262
111035660
111035669
111027295
111035668
111035678
111027289
111027297
111035665
111035675
111035670
111035672
111027292
111027290
111027288
111030375
111028571
111028573
111028574
111028570
111035674
111035663
111035662
111035661
111035666
111030374
111035677
111027293
111027291
111027294
111027296
111028572
111035671
111035673
111035676
111035667
810018914
810017089
810017053
810024571
810024910
810024560
810028770
810028813
810028796
810024881
810012865
810021026
810010770
810028067
810004067
810004071
810027914
111018585
111020779
810020155
810027806
111018717
810020483
810027731
810020470
810010741
810003903
111021876
810014229
810013009
810004802
111021886
810013711
111021648
810020724
111020780
810014223
111021310
111021773
111021643
810013013
810004135
810004002
810015916
810015920
810015917
810014259
810015908
810015915
810027929
810013025
810014255
810015910
810027744
810027918
810027936
810015466
810027888
810027752
810027903
810014199
810027948
810012997
810015900
810015896
810015912
810015905
810027945
810027934
810014263
810014306
810010718
810014278
810012999
810015911
810015903
810015921
810013021
810010749
810014214
810014268
810012706
810012737
810015919
810010704
111019105
810014265
111019286
810005042
810004329
111019821
111019853
111019849
111019816
810018918
810024568
810005135
810004744
810010727
111020786
810018694
810010764
810023354
810014257
810021031
810021022
810010724
111017649
810003778
810015914
111020016
111017425
111020007
810020156
810005137
810027742
111020291
810027793
111020784
810024594
810016845
111020789
810021020
810021024
810027708
111020243
810014286
810004326
810014095
111019305
111020219
111020283
111021307
111020795
810013693
111017668
810014189
810027722
111021883
810004063
810015408
810004742
111020287
810012672
111021641
810010735
111021874
810014228
810015925
810014253
810010766
810027729
810004062
810021025
111020308
810016853
810017071
810028826
111018546
810024589
111018963
810005040
810014226
810015452
810024909
810015444
810015447
810015413
111021879
810013710
810015412
810028787
810028837
810027925
810027797
810027789
810027805
810014313
810018927
111019812
111019107
810014247
810027754
111021237
810013002
810015314
810003902
111020235
810028429
111021656
810015442
810004887
810012736
111021895
810028801
810004805
810017100
810027786
111018595
111019688
111020218
810028094
111021763
810021016
810028019
810013696
810004743
810027779
810014256
810028432
810027943
810027743
810024566
810027784
111019097
810015454
810027759
810028460
810027769
810012669
810014215
810024884
810027767
810014280
810028059
810024564
810003907
810014217
111021647
810017205
810018910
810027938
810014302
111021898
111018583
111018018
810014213
810027713
111021884
111018547
111021868
810004623
810027716
810028803
810003870
810018925
810018935
810014182
810027791
810028023
111018735
111021657
810014237
810027775
810004127
810015393
810004134
810013012
810014291
810014279
810018928
111019104
810028039
111017449
810013705
111021889
810027730
810014282
111017469
810013003
111019869
810013008
111021652
810024878
810003808
810027901
111021866
810027939
111021638
810027898
111020295
810005139
810027905
810013700
810028791
810005195
810010771
111021766
111020021
111018017
111021637
810013000
810024901
111021300
810003809
111021880
111017453
810027768
810027762
810015313
810003869
111018708
810013019
810021015
810013704
111017655
111019182
810012670
810011926
111017687
111042818
810024553
111018012
810027718
810017070
810024912
810028777
810016833
810016633
810021014
810027886
810027952
810027896
810017060
810003868
810014307
810012864
810003908
111017693
111021767
810003657
810015407
810027910
810010778
111019295
810023358
111021870
810027735
810027940
810017092
810017208
810017186
111020213
810013023
111018903
111020000
810003999
810027736
810014196
810020471
810010779
810018936
810012679
810014287
810014230
111020307
810010709
810010757
810014236
810014270
810015391
810012871
810010746
810012870
810010744
111018737
810024918
810003780
810013715
111018900
111018725
810028434
810010773
111018901
810015396
111019109
810016849
111021769
111021859
810024581
810016851
810028411
810003718
810013024
111019668
111021777
810015449
810021011
810027780
810027725
810020472
810020222
810027723
810027715
810027771
810017195
810018912
810027738
111018902
111018906
111019117
810017052
810017193
810017099
810016692
810014266
111017631
111017622
111017663
111019863
111017457
111018964
111018910
810028456
810012738
810017049
111017465
810017209
111019670
810010710
111020305
111018748
810014293
111020794
111020204
111017690
111020020
111019301
111021762
810017073
810027904
810027890
111018543
111018716
111019820
810016831
810014283
810014200
810016966
810017087
810017066
810016836
810016855
111017429
111020288
111017496
810016822
111018722
111019111
111020220
810020469
810010706
810028042
810017197
810027947
810010758
810017196
111019999
810020464
810020462
111017700
810014271
810024890
810027761
810027893
810027899
810028769
810017054
810027931
810016843
810028073
810024584
810024565
111017636
810028462
810015465
810027949
810027885
810024554
810028463
810028466
111019178
810016860
111020022
810010767
810013720
810015451
810016969
810028798
810028825
810015404
111019288
810028821
111020202
111019982
111019851
810010725
810010720
810010742
111019279
111019284
810017057
111020793
111020773
111020783
111020787
810028459
810014284
810027941
810010775
111018724
111017441
810018701
111019278
111020008
810010780
111020231
810024548
810010769
810010731
810027935
111018895
111018972
111018733
111020234
111020214
810024888
810017206
810017201
810017076
810017074
111019680
111018582
111019687
810017078
810017095
810016837
810017065
810017093
810017077
810017199
810013695
810014240
810014261
111020781
111020774
111018897
111018960
810014254
810014273
810014233
810014194
810027919
111020201
111017658
810027956
810028426
810028425
810017067
810024558
810028793
810003911
810028060
810028077
111017648
111021869
810028018
810028083
810028081
810028076
810028021
111021776
111021779
111021306
111019815
111019984
111019814
111020005
111021645
111021303
810013717
810013703
111020241
810014112
810014245
810014192
810014180
810014183
810014249
810014221
810014309
111020027
111021861
111021887
111020239
111019995
111019819
111019996
111019981
111020025
111020244
810024543
810024588
810012707
810012671
810012740
810012750
810023362
111020207
111020209
111020017
111020028
111020026
111017491
111017477
111017604
111017640
111018752
111018023
111018022
111018013
111017650
111018025
111018026
111020208
111020205
111020001
111020217
111019807
111019811
111019818
111019848
111014080
111014090
111014069
111014076
111014073
111019990
111020200
810023315
810023380
810023375
810023376
810023377
810023379
810023363
810023370
810023365
810023372
810027953
810027782
810027916
810027950
810027942
111017473
111017702
111017462
111017494
111017420
111017620
111017470
111017493
111017698
111017628
111017641
111017460
111017608
Entrant:
Model Number: HF CRT-D REF CDHFA300Q
UDI-DI Code: 05415067032850
Serial Numbers:
810024784
111006209
111041472
111041484
111041470
111041479
111041477
111041478
111041488
111041485
111041487
111040400
111042676
111042693
111042689
111042695
111042694
111042688
111042685
111042690
111042687
111040401
111011798
111011778
111022037
111022040
111022053
111022052
111035584
111035572
111035580
111035564
111035567
111011838
111009858
111010165
111008963
111013513
111015899
111010990
111011771
111008361
810004607
111008364
810024872
111008933
111005058
810024619
810024869
111016289
111016242
111008169
111005574
111006587
111016306
810004459
111004315
810016787
111003406
810003838
810003941
810003832
810003939
810004050
810003840
810003827
810003981
810016625
810017549
810017552
810017555
810017546
810016614
810016743
810017033
810017016
810016609
810016754
810016784
810016747
810016758
810016626
111016304
111006633
111008194
111003404
111008377
810015659
810015304
810016781
810005069
111017880
111007104
111008191
111005736
111005305
111019313
111004702
810016618
810016619
111006359
111019314
111019311
111019315
111019319
111016277
810015225
111005411
810005033
111016883
111018047
111012485
111012493
111006604
111012344
111009871
111009848
111009872
111009449
111006349
111004329
111006643
111008358
810004361
111001115
810004300
111009503
111005610
111011793
111009873
111009513
111009467
111016887
111008367
111034828
111010327
111004323
111006335
810024609
111012343
111005747
111012325
111007241
111036785
111021744
810006394
111010325
111034851
111012333
111011820
111011834
810015555
810016763
111017929
111009853
111016225
111004044
111005383
111006648
111017888
810015536
810015662
111010982
111018309
810025078
111011753
111008160
810015535
810024614
810015751
810016624
810024780
810016630
810006404
111011794
111005738
111008162
111008171
111001103
111004697
111005603
111004319
111004672
810004870
810004357
810004740
111011809
111012498
111012483
111013527
810024654
810024599
810024600
111005308
111005302
810015523
111015732
111015727
111009855
810004157
810003977
810003940
810003829
810004153
810004241
111009498
111018310
111018308
111012345
111009865
111018034
111018038
111018045
111018046
111018054
111009861
810003830
111006595
810015551
810015554
810015532
810004611
111010721
111012346
111008376
111042925
111005373
111042895
111042877
111042878
111042880
111042897
111042908
111042904
111042902
111005598
810016623
111006367
810005976
111016235
111011759
111005168
810004461
111011766
111009043
810017020
111009867
111018051
810004115
810004047
111006236
111015918
111006380
111006344
111008375
111017926
111017013
810003935
810005065
111015760
111017958
810003767
111015884
111013517
111013528
111010752
810004355
111017974
111009518
111006985
810016742
111017953
111004045
111012328
111017990
111036795
111011843
111035591
810005032
810015384
111005550
111011818
111009471
111009462
111009497
111009895
111009894
111015882
111005587
111011780
111015734
111015725
111004040
111015736
111017055
111017006
111016287
111015908
111016854
111009514
111020840
111019317
111019307
111019312
111004327
111004066
111004035
111004054
111017987
810016755
111006596
111010729
810004940
810005233
111017971
810024978
111015735
111009850
111004669
810024607
810024977
810015383
111016875
111017032
810004453
111008139
810004450
111008382
111004430
111004039
810004350
111016863
810025076
111042921
810015550
810015654
111006225
111006622
111007232
111007235
111007228
111009464
810015736
810015661
810015733
810015680
810015744
810015747
810015731
810017021
810017017
810017031
810017027
810017034
111006606
111005749
111006355
111005050
111005377
111017923
111017856
111017011
111016876
111018050
111005467
810015675
111009066
111006642
111006382
111006366
111005296
111009516
111010170
111005582
810024632
111004306
111004425
111012477
810004048
111017884
111006342
810015307
111010312
111010268
111016253
111016270
111010179
810017024
111005464
111011779
810016786
810024786
810016773
810025073
111006616
111016281
111016856
810024793
810015547
111004292
810024611
111019316
111015923
111016259
810015537
111009166
111022023
111009507
111006347
111013510
111004435
111004705
111005595
111017875
111036793
111006582
111004417
111016303
111006358
111008209
810017548
111008137
111006620
111018053
111017852
111016237
111016262
810024980
111017967
111017892
111011760
810004299
111013505
111017878
810015243
810024641
111010983
111016857
111016848
111009859
111008180
111008927
111009463
111009474
111005394
810003583
810024783
810003932
111005547
111004426
810024787
810024791
810016761
810015389
111017985
111004695
810015655
111016847
111005446
111004686
111011816
111005469
111006222
111005593
810016785
111004439
111004298
111016898
810015253
810015656
810004608
810004737
810004617
810004613
810004612
810004614
810004561
810004451
810005068
111008150
111008193
111020848
111020852
111011787
111011770
111011776
111011835
111003391
111011836
111017925
111008176
111004438
111021753
111008155
111006621
810006002
111004321
111021760
810024639
810024652
810004042
810004160
810004044
810024658
810015682
810024651
810024634
111010988
111010724
111010986
111010318
111010733
111011797
111016250
111016269
111016245
111016280
111016274
111011817
111036798
111004678
111004334
111008203
111009161
111008381
111005456
111005444
111004301
111005491
111005591
810024606
111008195
111003384
111008945
111004033
111022047
111005372
111010753
111004424
111004333
111004047
111021739
810004107
111005046
111016263
810004616
111009451
111005741
810005008
810004454
810005191
111015888
111019306
111015723
111004318
111016310
111004423
111003400
810015531
810024656
111003401
111009067
111002095
111004700
111001105
810005070
810003704
111022035
111015886
111016873
111013523
111008964
111010726
111010725
111016257
111009052
111006601
810015665
810003834
111022028
810005073
111008172
810024627
111004326
111005398
111004691
111004305
810016776
111010186
111005599
111008947
111042814
810024870
810024790
810015735
810015303
111011805
111009856
111010193
111015741
111005392
111008152
111008366
111011840
111017995
111017989
111005746
810003705
111042881
111042882
111042884
111034836
111015924
111034820
111034853
111041489
111042923
111042891
111042914
111042893
111042890
111042924
111042887
111042883
111042899
111042896
111042915
111042900
111017964
111041492
111041471
111009491
111009461
111009458
111009509
111012482
111011806
111016233
111016261
111035587
111035547
111016265
111016276
111035571
111035575
111035545
111016283
111016294
111034842
111034811
111016264
111041482
111041469
111012481
111012487
111012490
111012488
111035568
111035590
111034831
111035555
111004680
111004324
111004303
111004042
111004314
111010177
111035546
111004664
111015916
111015897
111009456
111015912
111015913
111040387
111040394
111040390
111040300
111040386
111040384
111013496
111013497
111013525
111013519
111007240
111035563
111034813
111035559
111035562
111005375
111040397
111040396
111041473
111034823
111034832
111041481
111034812
111041474
111041468
111004687
111004688
111004675
111004311
111040403
111040393
111040388
111040391
111040302
111040389
111041486
111040399
111008362
111004663
111022049
111022039
111022046
111022033
111022030
111022031
111022027
111022025
111022036
111022050
111022048
111024338
111024341
111024339
111024337
111024340
111024336
111021755
111021742
111021743
111021745
111021741
111021757
111021750
111021737
111021749
111007105
111008178
111008161
111008175
111010996
111010993
111010995
111004704
111004332
111004692
111004296
111004681
111004673
111035574
111036776
111041480
111034816
111035548
111035556
111015925
111015926
111017988
111012486
111012491
111035578
111036783
111036791
111036799
111042699
111042679
111036777
111010985
111010720
111010738
111010723
111040301
111010320
111010313
111010321
111010316
111013522
111013530
111013516
111013514
111008179
111034837
111008170
111011801
111034844
111034843
111042681
111035554
111011808
111012337
111011839
111012326
111041491
111041490
111042677
111042691
111040404
111035588
111035576
111042697
111042680
111042698
111042684
111042682
111042675
111042678
111022038
111001322
111010187
111010185
111003388
111034819
810017019
111022042
810005127
111004068
111011824
111010744
111011757
111015744
810004683
810006406
111005492
111009469
111016850
111016894
111009892
111011823
810015306
111010183
810015249
810003979
810015683
810025074
810024613
810024597
810024604
810024598
810024610
810024782
810005932
810015737
810024647
810024779
111005602
111005578
111004682
810017039
111009044
810017026
810015529
810015560
810015381
111004694
810004609
111016307
111017879
111008949
111008939
111008950
111009493
111008153
111009849
111010323
111011795
111011781
111012342
111011825
111017992
111005408
111006901
111007230
111016295
111007233
111009053
111005452
111005415
111005381
111005412
111006585
111006637
111006617
111004670
111015742
111018035
111003390
111009473
111004679
111005601
111009510
111008141
111011775
111012340
111011799
111011761
111009476
111016268
111016279
111016234
111015738
111015906
810015373
111003395
111018312
111004070
111018048
810004610
810005162
810005120
111004046
810006405
810006399
111009050
111015898
111016254
111016300
111007220
111007231
111006654
111006636
111006591
111006238
111006351
111006340
810015752
810015746
810015653
810015685
810016748
810017037
810017553
810024789
810024608
111008188
111008928
111001318
810015530
111020845
810004242
111008167
111012329
111012336
810004152
111013508
111009484
111010745
111005066
111006374
111006363
810016777
810024649
810024643
810024871
810003793
810003792
111016853
810015668
111018055
111016899
810005126
810004238
810024629
111013509
111005470
111018032
111021751
111008942
111004706
810017454
111004437
111015892
111008138
111035586
810017018
111011827
111003392
810024659
111006605
810004871
810003645
111006653
111017921
810004113
810024867
810025079
111006360
111006646
111010182
111005743
111011837
111011826
111012330
111016315
111016864
111004041
111006590
111004429
111005604
111009453
111008205
111008144
111008135
810017551
810004046
810004351
111017916
810004352
111016223
810015377
810017550
810016779
111008943
111017854
111009881
810024640
810005123
810016620
810015549
810015672
810015670
810017455
810016770
810015251
810024638
111034849
810015669
810017025
810015245
111005545
111009520
111009165
111009870
111008145
111009506
111005449
111008192
810005006
810004240
810003770
810004159
111005395
810004458
111006375
810015667
810015526
810024628
810016766
111036794
810004156
810015255
111015907
111017058
111016288
111016298
111016292
111016243
111015881
810016765
111017902
111017061
810004110
111016888
111005453
111004060
810004347
810006400
111006644
111006357
810004243
111016228
810015559
810004297
810004358
810004348
810004455
810015385
111015919
111005608
111008941
111017908
810024617
111009068
111006602
810015660
810015241
810015540
111016308
111009055
111006368
810005160
111005493
810024645
111008373
111004711
111004698
111011785
111011758
810015248
111016273
111011769
810015375
111003383
111008199
810015252
111009064
810015250
810004798
111036779
810004449
810004049
111017035
111016266
111011830
111036789
111006235
111006230
111021747
810015254
111010184
111003385
111008372
111008185
111004336
810015308
810005164
111003407
111006613
111006337
111005055
810016613
111015877
810016762
111010329
111006652
111021736
111011777
111011774
111005378
111008107
111008196
111015880
111010994
111004300
111010747
810024777
111013529
810004114
111034847
111016290
111008383
111008184
810016746
810016760
111011803
810004942
810024785
810004874
111005407
810004301
111010319
810016752
111011819
111015745
111011841
810004245
111004036
810017456
111012492
810017038
111010315
810004360
111003394
111008149
111010167
111008969
111009167
111005388
111006377
111005306
111016855
111016889
810004615
111010984
111005607
111016865
810004043
810003985
810016745
111009857
111008146
111011783
810024868
111009852
111008186
111005549
111017993
111006237
810003986
810003794
111017947
810015742
111016248
810003984
810016775
810016608
810016772
111021748
111008159
111005060
111005410
111016241
111005577
111005589
810015522
111012324
111010740
111007226
111007244
111008207
111008206
111009847
111004079
111005495
111005494
111012327
111008166
111034840
111036796
111004076
111042686
111042879
111042892
111042922
111042916
111042911
111042907
111042894
111042886
111042920
111005368
111009045
810024776
810024612
810024635
810015673
111015920
111004062
111003393
111003405
111003402
111004328
111006583
111016309
810015374
810004452
111006383
111005371
111005462
111006352
111017994
111015748
111006369
111017881
111008108
111017045
810004239
810015652
111008953
810015738
810015658
111008368
111006348
111008142
111021759
111017040
111004313
111004316
111013501
111017858
111004065
810024644
111005384
111006364
111010324
111010267
810015387
111013511
810015684
810016782
111004051
111004708
111004671
810004294
810003937
810004359
810004457
810005067
810005072
810005071
810004349
810004293
810006408
810025083
810024981
810025089
810024984
810016612
810016621
111007219
111007225
111009042
111009048
810004944
111004074
111004703
111009874
111010174
111008369
111009487
111011845
111010166
111009448
111011832
111005600
810015739
810015748
810015674
111020846
111009882
111019308
111008940
810003584
111008958
111008959
111008965
111011788
810003703
111010728
111005742
111006598
810003444
810004684
111005498
111016296
111016231
111016312
111015728
111009489
111011762
111008380
111008363
111008379
111008968
111009164
111008934
111020844
111006634
111004291
111005457
111016900
111009454
810003768
810004741
810004799
111022056
111022051
810024657
810015750
810015558
810015379
810015244
111016271
111016311
810016606
810004456
111016301
111004419
111005745
810004941
810004108
810004116
810004154
810004051
810004109
810004161
810004158
810004111
810005034
111017954
111010176
111012334
111011791
111011800
111003382
111016278
111016297
111016251
810015524
111004082
111004331
111009468
810017452
111003398
810016750
111005370
111008202
111013515
810015753
111016897
810003836
810006003
810003982
810005929
810016616
810003938
810004045
810024642
111004302
111015917
111016852
111016229
111010330
111011767
111011796
111009494
111003397
810024781
111011765
111008208
810003933
810003980
810004112
111022055
810016629
810016610
810015553
810017032
810016749
810017035
111015759
111017905
111015751
810015676
111013526
111013498
111013506
111017991
111017986
111015915
111004330
810015749
111005451
111009490
111012484
810015557
810015528
111004674
111003386
810015542
111009472
111008158
111015757
111015876
111015755
111015730
810025086
810024601
111005397
111008962
111008961
111009169
111008931
810024595
111005065
111005750
111015890
111006346
111006638
810015310
810024603
111004048
111005390
111003389
111009455
810016769
111009495
111016286
111015911
111004422
810015223
810003837
810003936
111011755
111004295
810015242
111003403
111006336
111006350
810005129
111011784
810015679
810016759
810017036
810015681
810016617
810004800
111034846
810015538
810004739
810015755
810015301
111016884
810005122
810004738
810025071
810015732
810004872
810025084
810015657
810015651
111016291
111009500
111017855
111004667
111009481
810016753
810004292
810003978
111010189
111015726
111010266
111006597
111006600
111007224
111017915
111013504
810016628
810005924
111017863
810006396
810006401
810024637
111004709
111005309
111005573
810017028
111008189
111009486
111005465
111010717
111009488
111009508
111005376
111018037
111008157
111016869
111003399
810024631
111009879
111009862
111009877
111017922
810024661
111015922
810015745
111004668
111005310
111011768
111011773
111004677
111021746
111001315
111005069
810024616
111012332
111009168
111010188
111009460
111008929
810004237
111008952
111004077
111004058
111015750
111010722
111021752
111011815
810015740
111016275
111015879
111004075
111015752
111005307
810006407
810015663
810005931
111016313
111017867
111017874
111008147
111011763
111010190
111010169
111004072
810025077
111008936
111009452
111009478
111009459
111015893
111016258
111016267
111022034
111022026
111021740
111015749
810005230
810005978
111009163
111008966
810005977
810015743
810005930
810005190
111009483
111009517
111005543
111008938
111020850
111008374
111004416
111005583
111013524
810015521
810005234
111009878
111004037
810005928
810005927
111009466
111008937
111008970
111008198
111005585
111008136
111005586
111009162
111008197
111005576
111005497
111004308
111008956
111004038
111020847
111005542
111008967
111008182
111009511
111005393
111004061
111004078
111005575
810005232
810005225
111005594
810005226
810005125
810005926
810006403
810005231
111004064
810005933
111005596
111001116
111009846
111008200
810005229
111011831
810005228
810005974
111005391
810005224
111011756
111004073
111004294
111004059
111004304
111010328
810004244
111010742
111013507
810016789
111005382
111006384
111007216
111017877
810003934
810004685
111004418
111011764
111018039
810024792
810005121
810004353
810004296
111011833
810004246
111017861
810004797
111022022
810003833
111020838
810005074
111004581
111006227
111011810
111012339
111011844
111012338
111020855
111015902
111010164
111010194
810025072
111009499
111015885
810016778
810016767
810016622
810004354
111016872
810016771
111022043
111017009
810015546
810015556
111007222
111034838
111020843
111010732
810016751
810024618
111010730
810024625
111013512
111011804
111011772
810015741
111015883
810006398
810006402
111005751
111011822
111009887
111004696
111006624
111016881
111022045
111005592
810015372
111007242
111008156
810024636
810005194
810024620
111008177
111006213
111005385
111016239
111012335
810004155
111016305
111016282
111008134
111010734
810004298
111008164
111015739
111015733
111004701
111005374
111006581
810003983
111004063
111010175
111008181
111034827
810016768
111006612
111034830
810005193
810005007
111006341
111006353
111005447
111015914
111007010
111004420
810015371
810015527
810003771
111007243
810015548
111004043
111016255
111018033
111012497
810004686
111004685
810025080
111011814
111008954
111005579
111009889
111004432
111005572
111016858
111004293
810004560
111013499
111008960
111004057
111004676
111004421
111009041
810025085
111009885
111017951
810003702
810003795
810003769
810003582
810003839
111004335
111008951
111005450
111010195
111015729
810005192
111019310
111035561
111035558
111036797
111036784
111036788
111036782
111036778
111036800
111036790
111035592
111035589
111035566
111036787
111010265
111005597
111015895
111042683
111009868
111001319
810024982
111000826
111036786
111036792
111001124
111001307
111001308
111018041
111034841
111001317
111000837
111006356
111001314
111001309
111009479
111001313
111042906
111000834
111034850
111035594
111016880
111001111
111034817
111017871
111008932
111001118
111004683
111040395
111042903
111000830
111034822
111035550
111041483
111000833
111035573
111011754
111001114
810025075
111009890
111005496
111005609
111004053
111040392
111001104
111034839
111008187
111008948
111009470
111009465
111034834
111041475
111035583
111001119
111034821
111034818
111009851
111009883
111001321
111001356
111001102
111010741
111001117
111009863
111001123
111008371
111001311
111009501
111034815
111010727
111034835
810025082
111008370
111008946
810015541
810015390
111015746
111017870
111008204
111006599
111016246
111016256
810003796
810024622
111042901
111001120
111009884
111042817
111035569
111011782
111010749
111010718
111035565
111042815
111001121
111035551
111000825
111009860
111006611
111042913
111001122
111008148
111009886
111000829
111000838
111000820
111009475
111015894
111000832
111000823
111001306
111010192
111035549
111003396
111009515
111035582
111000827
111009866
111000822
111036780
111035560
111034854
111034829
111016224
111035552
111009875
111000824
111004310
111008165
111010735
111035579
111005546
111008935
111004312
111001320
111001109
111042692
111000836
111008168
111001110
111018044
111001357
111005068
810024605
810015247
111016882
810024624
810024655
111001125
111004055
111004427
111005580
111004436
111005396
111008359
111009170
111009482
111017889
810015376
810015388
111017914
111008378
810015309
111004428
111004052
111004056
111004320
111015722
111005588
111004050
810004943
810005005
810004460
111004707
111005541
111022029
111010171
111004080
111007237
111008183
111006630
810017023
111010191
111006993
111010322
810015380
111017973
111009504
111004322
111011812
111005606
111005369
111013520
111010754
111010989
111008957
810005066
810024626
111017960
111015887
111015904
111015901
810024602
810024615
111017901
810017029
111022054
810005030
111012341
111020851
111004431
111015743
111019318
810024788
111006381
111009496
810016615
810015221
111006627
111008143
810015545
111016860
111001113
111006645
810016764
111008210
111015737
810015305
111013500
111009065
810024650
810024983
111016236
111017885
810005031
111016316
111006629
111005304
111017903
810005223
810017030
111008174
810005124
810005161
810015754
810015677
810016744
810017022
111005300
111015740
111006649
111015724
111017036
111010311
111021758
111022032
810005128
810004873
111008944
111009502
111010746
111016859
111008360
810006395
810005163
111004665
111010172
111016302
111022024
111017868
111006338
111004666
111005387
111005499
810016780
111004689
111004699
111004710
111021738
810024660
111000835
111011807
111008151
111004297
111009480
111012489
111001323
111001324
111035570
111010748
111011790
111042889
111034845
111016272
111009505
111004317
810015246
111009492
111010991
810024648
111001106
111005047
810015222
111009864
810015544
111006379
810003642
810015525
111006231
810004682
810015671
111009512
111005067
111009046
111005298
111006378
111006614
111015756
111009869
810024778
111006626
810015539
810015533
111019309
810003706
111007007
111016226
810015302
111004067
810016611
111008140
111018042
111010987
111011842
111006610
111004573
111018052
111017996
111042885
111001112
111008163
111042898
111042912
111009893
111000821
111010181
111034848
111040385
111001316
111001310
111034825
111010755
111004049
111034824
111006584
111042813
810024630
810025081
111042888
111012495
111001126
111035577
111009485
111015875
111015754
810015543
111017019
111010739
111004662
111003387
111010750
810006397
111017900
111006588
810005227
111004690
111034852
111000831
111009888
810024621
111022044
810005009
810003841
111005454
810015534
810003831
810003835
111004433
111011829
111008190
111005455
111006594
111013521
111008201
111009854
111017025
111035585
111001108
810024775
810015382
810016607
111009880
111017047
111009876
111010173
111042696
111005605
111010751
111004434
111004325
111015878
111008154
111035581
810015552
810005925
111005070
810004295
111004071
111016890
111035593
111007221
111010168
111016868
111004299
111010737
111000828
810016756
810017547
810024646
111009056
111040398
810015224
810015378
810016741
810015664
111005544
810016757
111004309
111017895
111042917
111042918
111006370
111006339
111017920
111017911
810025087
810025088
810024653
810024623
810017040
810017453
810016774
810016790
810016788
810015386
111010178
111010326
111010992
810017554
111021754
810003585
810003644
810003643
810003646
810003586
810004356
111020854
111040320
111042926
111006640
111006628
111005057
111005468
111006586
111004571
111040408
111040327
111040322
111040409
111042933
111042927
111040407
111005461
111004566
111008119
111008106
111042940
111042929
111015900
111004580
111006211
111008109
111005303
111005052
111005459
111005735
111006608
111006580
111006229
111005733
111005071
111004568
111006641
111006234
111006232
111016249
111008120
111040411
111042928
111005466
111008114
111006373
111006632
111006365
111040326
111006212
111005734
111006210
111006376
111006635
111042935
111042934
111042937
111015905
111016232
111016230
111006609
111005414
111005048
111042930
111007223
111005409
111008132
111006345
111005752
111006208
111006219
111005748
111004574
111006343
111006639
111005063
111005311
111004565
111006593
111005062
111004569
111005043
111005049
111006619
111005739
111006354
111042931
111016247
111040412
111040317
111040318
111042936
111006589
111040323
111005744
111009061
111006217
111040324
111004576
111004575
111005379
111008130
111005061
111042939
111040406
111008127
111004570
111005056
111006371
111006215
111006216
111005458
111009054
111005460
111011828
111005413
|
Recalling Firm/ Manufacturer |
Abbott Medical 15900 Valley View Ct Sylmar CA 91342-3577
|
For Additional Information Contact | Alicia Swanson 408-845-3427 |
Manufacturer Reason for Recall | Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity. |
FDA Determined Cause 2 | Component design/selection |
Action | On 08/16/2023, the firm sent to U.S. customers an "URGENT MEDICAL DEVICE CORRECTION" Letter (via mail, email or hand-delivery) informing physicians and healthcare professionals that there is a potential for a Bluetooth (BLE) circuit component issue on a subset of Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators manufactured prior to April 2022. The issue is associated with a risk of loss of Bluetooth communication (and therefore a loss of remote monitoring), higher-than-normal current consumption, and reduced device battery longevity.
Patient Management Recommendations:
1. Prophylactic device replacement is not recommended as new firmware version pr00.10.87.04 eliminates the potential for loss of therapy between follow-up due to unrecognized decreased device longevity.
2. Determine the firmware version of devices followed at clinics by reviewing the instructions in Appendix A and the device list in Appendix B.
3. Patients with firmware version pr00.10.87.00 or firmware version undetermined, upgrade devices to firmware version pr00.10.87.04:
-Prioritize in-clinic firmware upgrade for the specific devices from the 1,500 device sub-group (listed in Appendix B)
-For remaining patients, schedule the next follow-up in clinic to complete firmware upgrade.
3. Following firmware upgrade, continue to follow patients routinely
4. If device experience a loss of Bluetooth communication, contact Abbott Technical Support at 1-800-722-3774 (U.S.). In addition, a list of Abbott advisories is available at https://www.cardiovascular.abbott/us/en/hcp/product-advisories.html
Abbott has also include a web version of the Customer Notification Letter at
https://www.cardiovascular.abbott/content/dam/cv/cardiovascular/pdf/reports/US-Abbott-Communication-BLE-Communication-Final-FDA-v2.pdf |
Quantity in Commerce | 29,088 devices |
Distribution | U.S.: Nation-wide
O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Nepal, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.
Latin America - Pending |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
PMA Database | PMAs with Product Code = NIK
|
|
|
|