|
Class 2 Device Recall Alinity |
|
Date Initiated by Firm |
October 19, 2023 |
Date Posted |
January 09, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-0680-2024 |
Recall Event ID |
93355 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product |
Alinity i Processing Module REF 03R65-01
The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples. |
Code Information |
List Number: 03R65-01
Serial Numbers / UDI-DI codes (for U.S. systems only):
AI01413 (01)00380740137366(21)AI01413
AI01414 (01)00380740137366(21)AI01414
AI01434 (01)00380740137366(21)AI01434
AI01447 (01)00380740137366(21)AI01447
AI01467 (01)00380740137366(21)AI01467
AI01468 (01)00380740137366(21)AI01468
AI01599 (01)00380740137366(21)AI01599
AI01607 (01)00380740137366(21)AI01607
AI01796 (01)00380740137366(21)AI01796
AI01823 (01)00380740137366(21)AI01823
AI01824 (01)00380740137366(21)AI01824
AI01827 (01)00380740137366(21)AI01827
AI01830 (01)00380740137366(21)AI01830
AI01837 (01)00380740137366(21)AI01837
AI01839 (01)00380740137366(21)AI01839
AI01873 (01)00380740137366(21)AI01873
AI01874 (01)00380740137366(21)AI01874
AI01991 (01)00380740137366(21)AI01991
AI02055 (01)00380740137366(21)AI02055
AI02056 (01)00380740137366(21)AI02056
AI02057 (01)00380740137366(21)AI02057
AI02058 (01)00380740137366(21)AI02058
AI02109 (01)00380740137366(21)AI02109
AI02110 (01)00380740137366(21)AI02110
AI02111 (01)00380740137366(21)AI02111
AI02117 (01)00380740137366(21)AI02117
AI02199 (01)00380740137366(21)AI02199
AI02201 (01)00380740137366(21)AI02201
AI02208 (01)00380740137366(21)AI02208
AI02232 (01)00380740137366(21)AI02232
AI02234 (01)00380740137366(21)AI02234
AI02288 (01)00380740137366(21)AI02288
AI02292 (01)00380740137366(21)AI02292
AI02293 (01)00380740137366(21)AI02293
AI02345 (01)00380740137366(21)AI02345
AI02368 (01)00380740137366(21)AI02368
AI02391 (01)00380740137366(21)AI02391
AI02392 (01)00380740137366(21)AI02392
AI02423 (01)00380740137366(21)AI02423
AI02433 (01)00380740137366(21)AI02433
AI02444 (01)00380740137366(21)AI02444
AI02448 (01)00380740137366(21)AI02448
AI02493 (01)00380740137366(21)AI02493
AI02514 (01)00380740137366(21)AI02514
AI02536 (01)00380740137366(21)AI02536
AI02555 (01)00380740137366(21)AI02555
AI02569 (01)00380740137366(21)AI02569
AI02592 (01)00380740137366(21)AI02592
AI02594 (01)00380740137366(21)AI02594
AI02595 (01)00380740137366(21)AI02595
AI02750 (01)00380740137366(21)AI02750
AI02784 (01)00380740137366(21)AI02784
AI02802 (01)00380740137366(21)AI02802
AI02808 (01)00380740137366(21)AI02808
AI02812 (01)00380740137366(21)AI02812
AI02821 (01)00380740137366(21)AI02821
AI02827 (01)00380740137366(21)AI02827
AI02849 (01)00380740137366(21)AI02849
AI02850 (01)00380740137366(21)AI02850
AI02859 (01)00380740137366(21)AI02859
AI02877 (01)00380740137366(21)AI02877
AI02881 (01)00380740137366(21)AI02881
AI02883 (01)00380740137366(21)AI02883
AI02894 (01)00380740137366(21)AI02894
AI02921 (01)00380740137366(21)AI02921
AI02922 (01)00380740137366(21)AI02922
AI02923 (01)00380740137366(21)AI02923
AI02939 (01)00380740137366(21)AI02939
AI02981 (01)00380740137366(21)AI02981
AI02997 (01)00380740137366(21)AI02997
AI02999 (01)00380740137366(21)AI02999
AI03007 (01)00380740137366(21)AI03007
AI03035 (01)00380740137366(21)AI03035
AI03036 (01)00380740137366(21)AI03036
AI03047 (01)00380740137366(21)AI03047
AI03060 (01)00380740137366(21)AI03060
AI03070 (01)00380740137366(21)AI03070
AI03073 (01)00380740137366(21)AI03073
AI03075 (01)00380740137366(21)AI03075
AI03077 (01)00380740137366(21)AI03077
AI03081 (01)00380740137366(21)AI03081
AI03082 (01)00380740137366(21)AI03082
AI03084 (01)00380740137366(21)AI03084
AI03085 (01)00380740137366(21)AI03085
AI03086 (01)00380740137366(21)AI03086
AI03088 (01)00380740137366(21)AI03088
AI03089 (01)00380740137366(21)AI03089
AI03093 (01)00380740137366(21)AI03093
AI03094 (01)00380740137366(21)AI03094
AI03141 (01)00380740137366(21)AI03141
AI03143 (01)00380740137366(21)AI03143
AI03145 (01)00380740137366(21)AI03145
AI03146 (01)00380740137366(21)AI03146
AI03153 (01)00380740137366(21)AI03153
AI03157 (01)00380740137366(21)AI03157
AI03160 (01)00380740137366(21)AI03160
AI03188 (01)00380740137366(21)AI03188
AI03390 (01)00380740137366(21)AI03390
AI03392 (01)00380740137366(21)AI03392
AI03393 (01)00380740137366(21)AI03393
AI03397 (01)00380740137366(21)AI03397
AI03398 (01)00380740137366(21)AI03398
AI03487 (01)00380740137366(21)AI03487
AI03498 (01)00380740137366(21)AI03498
AI03513 (01)00380740137366(21)AI03513
AI03518 (01)00380740137366(21)AI03518
AI03523 (01)00380740137366(21)AI03523
AI03526 (01)00380740137366(21)AI03526
AI03541 (01)00380740137366(21)AI03541
AI03545 (01)00380740137366(21)AI03545
AI03548 (01)00380740137366(21)AI03548
AI03607 (01)00380740137366(21)AI03607
AI03639 (01)00380740137366(21)AI03639
AI03642 (01)00380740137366(21)AI03642
AI03697 (01)00380740137366(21)AI03697
AI03707 (01)00380740137366(21)AI03707
AI03712 (01)00380740137366(21)AI03712
AI03713 (01)00380740137366(21)AI03713
AI03727 (01)00380740137366(21)AI03727
AI03729 (01)00380740137366(21)AI03729
AI03736 (01)00380740137366(21)AI03736
AI03739 (01)00380740137366(21)AI03739
AI03744 (01)00380740137366(21)AI03744
AI03747 (01)00380740137366(21)AI03747
AI03748 (01)00380740137366(21)AI03748
AI03749 (01)00380740137366(21)AI03749
AI03751 (01)00380740137366(21)AI03751
AI03753 (01)00380740137366(21)AI03753
AI03755 (01)00380740137366(21)AI03755
AI03756 (01)00380740137366(21)AI03756
AI03758 (01)00380740137366(21)AI03758
AI03762 (01)00380740137366(21)AI03762
AI03763 (01)00380740137366(21)AI03763
AI03765 (01)00380740137366(21)AI03765
AI03767 (01)00380740137366(21)AI03767
AI03768 (01)00380740137366(21)AI03768
AI03801 (01)00380740137366(21)AI03801
AI03802 (01)00380740137366(21)AI03802
AI03809 (01)00380740137366(21)AI03809
AI03810 (01)00380740137366(21)AI03810
AI03811 (01)00380740137366(21)AI03811
AI03812 (01)00380740137366(21)AI03812
AI03814 (01)00380740137366(21)AI03814
AI03815 (01)00380740137366(21)AI03815
AI03820 (01)00380740137366(21)AI03820
AI03848 (01)00380740137366(21)AI03848
AI03852 (01)00380740137366(21)AI03852
AI03859 (01)00380740137366(21)AI03859
AI03860 (01)00380740137366(21)AI03860
AI03861 (01)00380740137366(21)AI03861
AI03862 (01)00380740137366(21)AI03862
AI03866 (01)00380740137366(21)AI03866
AI03867 (01)00380740137366(21)AI03867
AI03882 (01)00380740137366(21)AI03882
AI03883 (01)00380740137366(21)AI03883
AI03884 (01)00380740137366(21)AI03884
AI03885 (01)00380740137366(21)AI03885
AI03886 (01)00380740137366(21)AI03886
AI03887 (01)00380740137366(21)AI03887
AI03888 (01)00380740137366(21)AI03888
AI03890 (01)00380740137366(21)AI03890
AI03891 (01)00380740137366(21)AI03891
AI03894 (01)00380740137366(21)AI03894
AI03895 (01)00380740137366(21)AI03895
AI03898 (01)00380740137366(21)AI03898
AI03899 (01)00380740137366(21)AI03899
AI03900 (01)00380740137366(21)AI03900
AI04016 (01)00380740137366(21)AI04016
AI04030 (01)00380740137366(21)AI04030
AI04031 (01)00380740137366(21)AI04031
AI04032 (01)00380740137366(21)AI04032
AI04033 (01)00380740137366(21)AI04033
AI04034 (01)00380740137366(21)AI04034
AI04035 (01)00380740137366(21)AI04035
AI04037 (01)00380740137366(21)AI04037
AI04038 (01)00380740137366(21)AI04038
AI04040 (01)00380740137366(21)AI04040
AI04058 (01)00380740137366(21)AI04058
AI04062 (01)00380740137366(21)AI04062
AI04068 (01)00380740137366(21)AI04068
AI04088 (01)00380740137366(21)AI04088
AI04199 (01)00380740137366(21)AI04199
AI04216 (01)00380740137366(21)AI04216
AI04218 (01)00380740137366(21)AI04218
AI04219 (01)00380740137366(21)AI04219
AI04220 (01)00380740137366(21)AI04220
AI04221 (01)00380740137366(21)AI04221
AI04222 (01)00380740137366(21)AI04222
AI04224 (01)00380740137366(21)AI04224
AI04226 (01)00380740137366(21)AI04226
AI04235 (01)00380740137366(21)AI04235
AI04237 (01)00380740137366(21)AI04237
AI04283 (01)00380740137366(21)AI04283
AI04284 (01)00380740137366(21)AI04284
AI04289 (01)00380740137366(21)AI04289
AI04290 (01)00380740137366(21)AI04290
AI04291 (01)00380740137366(21)AI04291
AI04293 (01)00380740137366(21)AI04293
AI04295 (01)00380740137366(21)AI04295
AI04297 (01)00380740137366(21)AI04297
AI04301 (01)00380740137366(21)AI04301
AI04302 (01)00380740137366(21)AI04302
AI04328 (01)00380740137366(21)AI04328
AI04340 (01)00380740137366(21)AI04340
AI04342 (01)00380740137366(21)AI04342
AI04349 (01)00380740137366(21)AI04349
AI04398 (01)00380740137366(21)AI04398
AI04458 (01)00380740137366(21)AI04458
AI04459 (01)00380740137366(21)AI04459
AI04460 (01)00380740137366(21)AI04460
AI04461 (01)00380740137366(21)AI04461
AI04462 (01)00380740137366(21)AI04462
AI04463 (01)00380740137366(21)AI04463
AI04464 (01)00380740137366(21)AI04464
AI04465 (01)00380740137366(21)AI04465
AI04467 (01)00380740137366(21)AI04467
AI04476 (01)00380740137366(21)AI04476
AI04477 (01)00380740137366(21)AI04477
AI04480 (01)00380740137366(21)AI04480
AI04481 (01)00380740137366(21)AI04481
AI04482 (01)00380740137366(21)AI04482
AI04484 (01)00380740137366(21)AI04484
AI04485 (01)00380740137366(21)AI04485
AI04486 (01)00380740137366(21)AI04486
AI04487 (01)00380740137366(21)AI04487
AI04660 (01)00380740137366(21)AI04660
AI04661 (01)00380740137366(21)AI04661
AI04662 (01)00380740137366(21)AI04662
AI04663 (01)00380740137366(21)AI04663
AI04666 (01)00380740137366(21)AI04666
AI04667 (01)00380740137366(21)AI04667
AI04701 (01)00380740137366(21)AI04701
AI04702 (01)00380740137366(21)AI04702
AI04703 (01)00380740137366(21)AI04703
AI04708 (01)00380740137366(21)AI04708
AI04709 (01)00380740137366(21)AI04709
AI04710 (01)00380740137366(21)AI04710
AI04711 (01)00380740137366(21)AI04711
AI04712 (01)00380740137366(21)AI04712
AI04713 (01)00380740137366(21)AI04713
AI04715 (01)00380740137366(21)AI04715
AI04716 (01)00380740137366(21)AI04716
AI04721 (01)00380740137366(21)AI04721
AI04724 (01)00380740137366(21)AI04724
AI04726 (01)00380740137366(21)AI04726
AI04732 (01)00380740137366(21)AI04732
AI04733 (01)00380740137366(21)AI04733
AI04734 (01)00380740137366(21)AI04734
AI04735 (01)00380740137366(21)AI04735
AI04736 (01)00380740137366(21)AI04736
AI04737 (01)00380740137366(21)AI04737
AI04738 (01)00380740137366(21)AI04738
AI04739 (01)00380740137366(21)AI04739
AI04740 (01)00380740137366(21)AI04740
AI04742 (01)00380740137366(21)AI04742
AI04743 (01)00380740137366(21)AI04743
AI04744 (01)00380740137366(21)AI04744
AI04745 (01)00380740137366(21)AI04745
AI04746 (01)00380740137366(21)AI04746
AI04747 (01)00380740137366(21)AI04747
AI04748 (01)00380740137366(21)AI04748
AI04749 (01)00380740137366(21)AI04749
AI04759 (01)00380740137366(21)AI04759
AI04762 (01)00380740137366(21)AI04762
AI04763 (01)00380740137366(21)AI04763
AI04764 (01)00380740137366(21)AI04764
AI04769 (01)00380740137366(21)AI04769
AI04770 (01)00380740137366(21)AI04770
AI04771 (01)00380740137366(21)AI04771
AI04774 (01)00380740137366(21)AI04774
AI04775 (01)00380740137366(21)AI04775
AI04776 (01)00380740137366(21)AI04776
AI04777 (01)00380740137366(21)AI04777
AI04778 (01)00380740137366(21)AI04778
AI04782 (01)00380740137366(21)AI04782
AI04783 (01)00380740137366(21)AI04783
AI04784 (01)00380740137366(21)AI04784
AI04785 (01)00380740137366(21)AI04785
AI04786 (01)00380740137366(21)AI04786
AI04787 (01)00380740137366(21)AI04787
AI04788 (01)00380740137366(21)AI04788
AI04789 (01)00380740137366(21)AI04789
AI04792 (01)00380740137366(21)AI04792
AI04793 (01)00380740137366(21)AI04793
AI04794 (01)00380740137366(21)AI04794
AI04805 (01)00380740137366(21)AI04805
AI04806 (01)00380740137366(21)AI04806
AI04807 (01)00380740137366(21)AI04807
AI04809 (01)00380740137366(21)AI04809
AI04817 (01)00380740137366(21)AI04817
AI04818 (01)00380740137366(21)AI04818
AI04819 (01)00380740137366(21)AI04819
AI04820 (01)00380740137366(21)AI04820
AI04821 (01)00380740137366(21)AI04821
AI04825 (01)00380740137366(21)AI04825
AI04829 (01)00380740137366(21)AI04829
AI04830 (01)00380740137366(21)AI04830
AI04831 (01)00380740137366(21)AI04831
AI04832 (01)00380740137366(21)AI04832
AI04833 (01)00380740137366(21)AI04833
AI04834 (01)00380740137366(21)AI04834
AI04835 (01)00380740137366(21)AI04835
AI04836 (01)00380740137366(21)AI04836
AI04837 (01)00380740137366(21)AI04837
AI04849 (01)00380740137366(21)AI04849
AI04850 (01)00380740137366(21)AI04850
AI04851 (01)00380740137366(21)AI04851
AI04852 (01)00380740137366(21)AI04852
AI04853 (01)00380740137366(21)AI04853
AI04854 (01)00380740137366(21)AI04854
AI04856 (01)00380740137366(21)AI04856
AI04858 (01)00380740137366(21)AI04858
AI04869 (01)00380740137366(21)AI04869
AI04870 (01)00380740137366(21)AI04870
AI04871 (01)00380740137366(21)AI04871
AI04872 (01)00380740137366(21)AI04872
AI04882 (01)00380740137366(21)AI04882
AI04883 (01)00380740137366(21)AI04883
AI04884 (01)00380740137366(21)AI04884
AI04885 (01)00380740137366(21)AI04885
AI04886 (01)00380740137366(21)AI04886
AI04887 (01)00380740137366(21)AI04887
AI20046 (01)00380740137366(21)AI20046
AI20051 (01)00380740137366(21)AI20051
AI20053 (01)00380740137366(21)AI20053
AI20055 (01)00380740137366(21)AI20055
AI20056 (01)00380740137366(21)AI20056
AI20069 (01)00380740137366(21)AI20069
AI20073 (01)00380740137366(21)AI20073
AI20114 (01)00380740137366(21)AI20114
AI20115 (01)00380740137366(21)AI20115
AI20118 (01)00380740137366(21)AI20118
AI20357 (01)00380740137366(21)AI20357
AI20358 (01)00380740137366(21)AI20358
AI20360 (01)00380740137366(21)AI20360
AI20366 (01)00380740137366(21)AI20366
AI20368 (01)00380740137366(21)AI20368
AI20369 (01)00380740137366(21)AI20369
AI20370 (01)00380740137366(21)AI20370
AI20371 (01)00380740137366(21)AI20371
AI20374 (01)00380740137366(21)AI20374
AI20376 (01)00380740137366(21)AI20376
AI20416 (01)00380740137366(21)AI20416
AI20501 (01)00380740137366(21)AI20501
AI20503 (01)00380740137366(21)AI20503
AI20511 (01)00380740137366(21)AI20511
AI20519 (01)00380740137366(21)AI20519
AI20521 (01)00380740137366(21)AI20521
AI20525 (01)00380740137366(21)AI20525
AI20527 (01)00380740137366(21)AI20527
AI20528 (01)00380740137366(21)AI20528
AI20530 (01)00380740137366(21)AI20530
AI20531 (01)00380740137366(21)AI20531
AI20532 (01)00380740137366(21)AI20532
AI20535 (01)00380740137366(21)AI20535
AI20536 (01)00380740137366(21)AI20536
AI20539 (01)00380740137366(21)AI20539
AI20541 (01)00380740137366(21)AI20541
AI20543 (01)00380740137366(21)AI20543
AI20578 (01)00380740137366(21)AI20578
AI20582 (01)00380740137366(21)AI20582
AI20606 (01)00380740137366(21)AI20606
AI20607 (01)00380740137366(21)AI20607
AI20609 (01)00380740137366(21)AI20609
AI20674 (01)00380740137366(21)AI20674
AI20676 (01)00380740137366(21)AI20676
AI20680 (01)00380740137366(21)AI20680
AI20700 (01)00380740137366(21)AI20700
AI20797 (01)00380740137366(21)AI20797
AI20798 (01)00380740137366(21)AI20798
AI20799 (01)00380740137366(21)AI20799
AI20800 (01)00380740137366(21)AI20800
AI20861 (01)00380740137366(21)AI20861
AI20868 (01)00380740137366(21)AI20868
AI20871 (01)00380740137366(21)AI20871
AI20885 (01)00380740137366(21)AI20885
AI20887 (01)00380740137366(21)AI20887
AI20900 (01)00380740137366(21)AI20900
AI20904 (01)00380740137366(21)AI20904
AI20994 (01)00380740137366(21)AI20994
AI20996 (01)00380740137366(21)AI20996
AI20999 (01)00380740137366(21)AI20999
AI21106 (01)00380740137366(21)AI21106
AI21107 (01)00380740137366(21)AI21107
AI21108 (01)00380740137366(21)AI21108
AI21126 (01)00380740137366(21)AI21126
AI21151 (01)00380740137366(21)AI21151
AI21152 (01)00380740137366(21)AI21152
AI21154 (01)00380740137366(21)AI21154
AI21156 (01)00380740137366(21)AI21156
AI21157 (01)00380740137366(21)AI21157
AI21171 (01)00380740137366(21)AI21171
AI21173 (01)00380740137366(21)AI21173
AI21174 (01)00380740137366(21)AI21174
AI21177 (01)00380740137366(21)AI21177
AI21179 (01)00380740137366(21)AI21179
AI21182 (01)00380740137366(21)AI21182
AI21183 (01)00380740137366(21)AI21183
AI21193 (01)00380740137366(21)AI21193
AI21199 (01)00380740137366(21)AI21199
AI21200 (01)00380740137366(21)AI21200
AI21201 (01)00380740137366(21)AI21201
AI21217 (01)00380740137366(21)AI21217
AI21293 (01)00380740137366(21)AI21293
AI21301 (01)00380740137366(21)AI21301
AI21302 (01)00380740137366(21)AI21302
AI21311 (01)00380740137366(21)AI21311
AI21325 (01)00380740137366(21)AI21325
AI21326 (01)00380740137366(21)AI21326
AI21327 (01)00380740137366(21)AI21327
AI21360 (01)00380740137366(21)AI21360
AI21385 (01)00380740137366(21)AI21385
AI21395 (01)00380740137366(21)AI21395
AI21396 (01)00380740137366(21)AI21396
AI21398 (01)00380740137366(21)AI21398
AI21402 (01)00380740137366(21)AI21402
AI21405 (01)00380740137366(21)AI21405
AI21407 (01)00380740137366(21)AI21407
AI21408 (01)00380740137366(21)AI21408
AI21410 (01)00380740137366(21)AI21410
AI21411 (01)00380740137366(21)AI21411
AI21412 (01)00380740137366(21)AI21412
AI21413 (01)00380740137366(21)AI21413
AI21414 (01)00380740137366(21)AI21414
AI21430 (01)00380740137366(21)AI21430
AI21431 (01)00380740137366(21)AI21431
AI21432 (01)00380740137366(21)AI21432
AI21433 (01)00380740137366(21)AI21433
AI21445 (01)00380740137366(21)AI21445
AI21446 (01)00380740137366(21)AI21446
AI21451 (01)00380740137366(21)AI21451
AI21454 (01)00380740137366(21)AI21454
AI21463 (01)00380740137366(21)AI21463
AI21494 (01)00380740137366(21)AI21494
AI21495 (01)00380740137366(21)AI21495
AI21496 (01)00380740137366(21)AI21496
AI21497 (01)00380740137366(21)AI21497
AI21498 (01)00380740137366(21)AI21498
AI21499 (01)00380740137366(21)AI21499
AI21507 (01)00380740137366(21)AI21507
AI21508 (01)00380740137366(21)AI21508
AI21511 (01)00380740137366(21)AI21511
AI21512 (01)00380740137366(21)AI21512
AI21515 (01)00380740137366(21)AI21515
AI21521 (01)00380740137366(21)AI21521
AI21522 (01)00380740137366(21)AI21522
AI21523 (01)00380740137366(21)AI21523
AI21524 (01)00380740137366(21)AI21524
AI21525 (01)00380740137366(21)AI21525
AI21526 (01)00380740137366(21)AI21526
AI21527 (01)00380740137366(21)AI21527
AI21528 (01)00380740137366(21)AI21528
AI21531 (01)00380740137366(21)AI21531
AI21532 (01)00380740137366(21)AI21532
AI21535 (01)00380740137366(21)AI21535
AI21536 (01)00380740137366(21)AI21536
AI21537 (01)00380740137366(21)AI21537
AI21539 (01)00380740137366(21)AI21539
AI21541 (01)00380740137366(21)AI21541
AI21542 (01)00380740137366(21)AI21542
AI21543 (01)00380740137366(21)AI21543
AI21581 (01)00380740137366(21)AI21581
AI21592 (01)00380740137366(21)AI21592
AI21594 (01)00380740137366(21)AI21594
AI21595 (01)00380740137366(21)AI21595
AI21607 (01)00380740137366(21)AI21607
AI21623 (01)00380740137366(21)AI21623
AI21627 (01)00380740137366(21)AI21627
AI21628 (01)00380740137366(21)AI21628
AI21629 (01)00380740137366(21)AI21629
AI21630 (01)00380740137366(21)AI21630
AI21659 (01)00380740137366(21)AI21659
AI21665 (01)00380740137366(21)AI21665
AI21666 (01)00380740137366(21)AI21666
AI21668 (01)00380740137366(21)AI21668
AI21669 (01)00380740137366(21)AI21669
AI21684 (01)00380740137366(21)AI21684
AI21687 (01)00380740137366(21)AI21687
AI21690 (01)00380740137366(21)AI21690
AI21691 (01)00380740137366(21)AI21691
AI21697 (01)00380740137366(21)AI21697
AI21699 (01)00380740137366(21)AI21699
AI21753 (01)00380740137366(21)AI21753
AI21754 (01)00380740137366(21)AI21754
AI21810 (01)00380740137366(21)AI21810
AI21811 (01)00380740137366(21)AI21811
AI21814 (01)00380740137366(21)AI21814
AI21815 (01)00380740137366(21)AI21815
AI21817 (01)00380740137366(21)AI21817
AI21818 (01)00380740137366(21)AI21818
AI21847 (01)00380740137366(21)AI21847
AI21848 (01)00380740137366(21)AI21848
AI21852 (01)00380740137366(21)AI21852
AI21855 (01)00380740137366(21)AI21855
AI21856 (01)00380740137366(21)AI21856
AI21857 (01)00380740137366(21)AI21857
AI21864 (01)00380740137366(21)AI21864
AI21866 (01)00380740137366(21)AI21866
AI21867 (01)00380740137366(21)AI21867
AI21876 (01)00380740137366(21)AI21876
AI21884 (01)00380740137366(21)AI21884
AI21885 (01)00380740137366(21)AI21885
AI21892 (01)00380740137366(21)AI21892
AI21896 (01)00380740137366(21)AI21896
AI21897 (01)00380740137366(21)AI21897
AI21900 (01)00380740137366(21)AI21900
AI21904 (01)00380740137366(21)AI21904
AI21907 (01)00380740137366(21)AI21907
AI21985 (01)00380740137366(21)AI21985
AI21986 (01)00380740137366(21)AI21986
AI21987 (01)00380740137366(21)AI21987
AI21988 (01)00380740137366(21)AI21988
AI21989 (01)00380740137366(21)AI21989
AI21990 (01)00380740137366(21)AI21990
AI21991 (01)00380740137366(21)AI21991
AI21992 (01)00380740137366(21)AI21992
AI21993 (01)00380740137366(21)AI21993
AI21996 (01)00380740137366(21)AI21996
AI22000 (01)00380740137366(21)AI22000
AI22060 (01)00380740137366(21)AI22060
AI22075 (01)00380740137366(21)AI22075
AI22076 (01)00380740137366(21)AI22076
AI22077 (01)00380740137366(21)AI22077
AI22078 (01)00380740137366(21)AI22078
AI22104 (01)00380740137366(21)AI22104
AI22106 (01)00380740137366(21)AI22106
AI22107 (01)00380740137366(21)AI22107
AI22109 (01)00380740137366(21)AI22109
AI22110 (01)00380740137366(21)AI22110
AI22111 (01)00380740137366(21)AI22111
AI22126 (01)00380740137366(21)AI22126
AI22131 (01)00380740137366(21)AI22131
AI22132 (01)00380740137366(21)AI22132
AI22159 (01)00380740137366(21)AI22159
AI22160 (01)00380740137366(21)AI22160
AI22217 (01)00380740137366(21)AI22217
AI22222 (01)00380740137366(21)AI22222
AI22226 (01)00380740137366(21)AI22226
AI22227 (01)00380740137366(21)AI22227
AI22228 (01)00380740137366(21)AI22228
AI22231 (01)00380740137366(21)AI22231
AI22249 (01)00380740137366(21)AI22249
AI22250 (01)00380740137366(21)AI22250
AI22284 (01)00380740137366(21)AI22284
AI22289 (01)00380740137366(21)AI22289
AI22342 (01)00380740137366(21)AI22342
AI22343 (01)00380740137366(21)AI22343
AI22345 (01)00380740137366(21)AI22345
AI22347 (01)00380740137366(21)AI22347
AI22348 (01)00380740137366(21)AI22348
AI22349 (01)00380740137366(21)AI22349
AI22371 (01)00380740137366(21)AI22371
AI22408 (01)00380740137366(21)AI22408
AI22441 (01)00380740137366(21)AI22441
AI22469 (01)00380740137366(21)AI22469
AI22472 (01)00380740137366(21)AI22472
AI22473 (01)00380740137366(21)AI22473
AI22477 (01)00380740137366(21)AI22477
AI22480 (01)00380740137366(21)AI22480
AI22482 (01)00380740137366(21)AI22482
AI22489 (01)00380740137366(21)AI22489
AI22491 (01)00380740137366(21)AI22491
AI22499 (01)00380740137366(21)AI22499
AI22501 (01)00380740137366(21)AI22501
AI22511 (01)00380740137366(21)AI22511
AI22516 (01)00380740137366(21)AI22516
AI22549 (01)00380740137366(21)AI22549
AI22640 (01)00380740137366(21)AI22640
AI22654 (01)00380740137366(21)AI22654
AI22659 (01)00380740137366(21)AI22659
AI22661 (01)00380740137366(21)AI22661
AI22669 (01)00380740137366(21)AI22669
AI22688 (01)00380740137366(21)AI22688
AI22699 (01)00380740137366(21)AI22699
AI22705 (01)00380740137366(21)AI22705
AI22713 (01)00380740137366(21)AI22713
AI22714 (01)00380740137366(21)AI22714
AI22756 (01)00380740137366(21)AI22756
AI22759 (01)00380740137366(21)AI22759
AI22760 (01)00380740137366(21)AI22760
AI22762 (01)00380740137366(21)AI22762
AI22774 (01)00380740137366(21)AI22774
AI22775 (01)00380740137366(21)AI22775
AI22776 (01)00380740137366(21)AI22776
AI22789 (01)00380740137366(21)AI22789
AI22797 (01)00380740137366(21)AI22797
AI22799 (01)00380740137366(21)AI22799
AI22800 (01)00380740137366(21)AI22800
AI22802 (01)00380740137366(21)AI22802
AI22929 (01)00380740137366(21)AI22929
AI22954 (01)00380740137366(21)AI22954
AI22966 (01)00380740137366(21)AI22966
AI22968 (01)00380740137366(21)AI22968
AI22974 (01)00380740137366(21)AI22974
AI22976 (01)00380740137366(21)AI22976
AI22977 (01)00380740137366(21)AI22977
AI22978 (01)00380740137366(21)AI22978
AI22979 (01)00380740137366(21)AI22979
AI22990 (01)00380740137366(21)AI22990
AI22992 (01)00380740137366(21)AI22992
AI22994 (01)00380740137366(21)AI22994
AI22997 (01)00380740137366(21)AI22997
AI23075 (01)00380740137366(21)AI23075
AI23076 (01)00380740137366(21)AI23076
AI23081 (01)00380740137366(21)AI23081
AI23095 (01)00380740137366(21)AI23095
AI23100 (01)00380740137366(21)AI23100
AI23105 (01)00380740137366(21)AI23105
AI23114 (01)00380740137366(21)AI23114
AI23128 (01)00380740137366(21)AI23128
AI23131 (01)00380740137366(21)AI23131
AI23157 (01)00380740137366(21)AI23157
AI23241 (01)00380740137366(21)AI23241
AI23274 (01)00380740137366(21)AI23274
AI23303 (01)00380740137366(21)AI23303
AI23307 (01)00380740137366(21)AI23307
AI23321 (01)00380740137366(21)AI23321
AI23329 (01)00380740137366(21)AI23329
AI23330 (01)00380740137366(21)AI23330
AI23331 (01)00380740137366(21)AI23331
AI23399 (01)00380740137366(21)AI23399
AI23563 (01)00380740137366(21)AI23563
AI23577 (01)00380740137366(21)AI23577
AI23582 (01)00380740137366(21)AI23582
AI23587 (01)00380740137366(21)AI23587
AI23588 (01)00380740137366(21)AI23588
AI23591 (01)00380740137366(21)AI23591
AI23594 (01)00380740137366(21)AI23594
AI23595 (01)00380740137366(21)AI23595
AI23601 (01)00380740137366(21)AI23601
AI23603 (01)00380740137366(21)AI23603
AI23606 (01)00380740137366(21)AI23606
AI23609 (01)00380740137366(21)AI23609
AI23610 (01)00380740137366(21)AI23610
AI23617 (01)00380740137366(21)AI23617
AI23620 (01)00380740137366(21)AI23620
AI23622 (01)00380740137366(21)AI23622
AI23630 (01)00380740137366(21)AI23630
AI23674 (01)00380740137366(21)AI23674
AI23675 (01)00380740137366(21)AI23675
AI23678 (01)00380740137366(21)AI23678
AI23686 (01)00380740137366(21)AI23686
AI23701 (01)00380740137366(21)AI23701
AI23704 (01)00380740137366(21)AI23704
AI23708 (01)00380740137366(21)AI23708
AI23711 (01)00380740137366(21)AI23711
AI23713 (01)00380740137366(21)AI23713
AI23717 (01)00380740137366(21)AI23717
AI23719 (01)00380740137366(21)AI23719
AI23724 (01)00380740137366(21)AI23724
AI23728 (01)00380740137366(21)AI23728
AI23731 (01)00380740137366(21)AI23731
AI23738 (01)00380740137366(21)AI23738
AI23739 (01)00380740137366(21)AI23739
AI23743 (01)00380740137366(21)AI23743
AI23748 (01)00380740137366(21)AI23748
AI23752 (01)00380740137366(21)AI23752
AI23756 (01)00380740137366(21)AI23756
AI23757 (01)00380740137366(21)AI23757
AI23758 (01)00380740137366(21)AI23758
AI23759 (01)00380740137366(21)AI23759
AI23762 (01)00380740137366(21)AI23762
AI23763 (01)00380740137366(21)AI23763
AI23773 (01)00380740137366(21)AI23773
AI23774 (01)00380740137366(21)AI23774
AI23775 (01)00380740137366(21)AI23775
AI23781 (01)00380740137366(21)AI23781
AI23783 (01)00380740137366(21)AI23783
AI23785 (01)00380740137366(21)AI23785
AI23787 (01)00380740137366(21)AI23787
AI23797 (01)00380740137366(21)AI23797
AI23806 (01)00380740137366(21)AI23806
AI23808 (01)00380740137366(21)AI23808
AI23873 (01)00380740137366(21)AI23873
AI23902 (01)00380740137366(21)AI23902
AI23913 (01)00380740137366(21)AI23913
AI23921 (01)00380740137366(21)AI23921
AI23932 (01)00380740137366(21)AI23932
AI23935 (01)00380740137366(21)AI23935
AI23938 (01)00380740137366(21)AI23938
AI24021 (01)00380740137366(21)AI24021
AI24034 (01)00380740137366(21)AI24034
AI24060 (01)00380740137366(21)AI24060
AI24061 (01)00380740137366(21)AI24061
AI24064 (01)00380740137366(21)AI24064
AI24071 (01)00380740137366(21)AI24071
AI24072 (01)00380740137366(21)AI24072
AI24074 (01)00380740137366(21)AI24074
AI24091 (01)00380740137366(21)AI24091
AI24093 (01)00380740137366(21)AI24093
AI24094 (01)00380740137366(21)AI24094
AI24097 (01)00380740137366(21)AI24097
AI24099 (01)00380740137366(21)AI24099
AI24100 (01)00380740137366(21)AI24100
AI24122 (01)00380740137366(21)AI24122
AI24124 (01)00380740137366(21)AI24124
AI24125 (01)00380740137366(21)AI24125
AI24126 (01)00380740137366(21)AI24126
AI24130 (01)00380740137366(21)AI24130
AI24131 (01)00380740137366(21)AI24131
AI24132 (01)00380740137366(21)AI24132
AI24133 (01)00380740137366(21)AI24133
AI24135 (01)00380740137366(21)AI24135
AI24137 (01)00380740137366(21)AI24137
AI24138 (01)00380740137366(21)AI24138
AI24159 (01)00380740137366(21)AI24159
AI24169 (01)00380740137366(21)AI24169
AI24172 (01)00380740137366(21)AI24172
AI24184 (01)00380740137366(21)AI24184
AI24189 (01)00380740137366(21)AI24189
AI24199 (01)00380740137366(21)AI24199
AI24236 (01)00380740137366(21)AI24236
AI24264 (01)00380740137366(21)AI24264
AI24281 (01)00380740137366(21)AI24281
AI24282 (01)00380740137366(21)AI24282
AI24287 (01)00380740137366(21)AI24287
AI24293 (01)00380740137366(21)AI24293
AI24294 (01)00380740137366(21)AI24294
AI24295 (01)00380740137366(21)AI24295
AI24296 (01)00380740137366(21)AI24296
AI24297 (01)00380740137366(21)AI24297
AI24298 (01)00380740137366(21)AI24298
AI24299 (01)00380740137366(21)AI24299
AI24303 (01)00380740137366(21)AI24303
AI24306 (01)00380740137366(21)AI24306
AI24311 (01)00380740137366(21)AI24311
AI24317 (01)00380740137366(21)AI24317
AI24318 (01)00380740137366(21)AI24318
AI24319 (01)00380740137366(21)AI24319
AI24324 (01)00380740137366(21)AI24324
AI24330 (01)00380740137366(21)AI24330
AI24334 (01)00380740137366(21)AI24334
AI24336 (01)00380740137366(21)AI24336
AI24339 (01)00380740137366(21)AI24339
AI24345 (01)00380740137366(21)AI24345
AI24346 (01)00380740137366(21)AI24346
AI24347 (01)00380740137366(21)AI24347
AI24348 (01)00380740137366(21)AI24348
AI24349 (01)00380740137366(21)AI24349
AI24350 (01)00380740137366(21)AI24350
AI24351 (01)00380740137366(21)AI24351
AI24354 (01)00380740137366(21)AI24354
AI24356 (01)00380740137366(21)AI24356
AI24362 (01)00380740137366(21)AI24362
AI24363 (01)00380740137366(21)AI24363
AI24364 (01)00380740137366(21)AI24364
AI24366 (01)00380740137366(21)AI24366
AI24369 (01)00380740137366(21)AI24369
AI24383 (01)00380740137366(21)AI24383
AI24390 (01)00380740137366(21)AI24390
AI24479 (01)00380740137366(21)AI24479
AI24480 (01)00380740137366(21)AI24480
AI24481 (01)00380740137366(21)AI24481
AI24482 (01)00380740137366(21)AI24482
AI24485 (01)00380740137366(21)AI24485
AI24486 (01)00380740137366(21)AI24486
AI24487 (01)00380740137366(21)AI24487
AI24488 (01)00380740137366(21)AI24488
AI24494 (01)00380740137366(21)AI24494
AI24495 (01)00380740137366(21)AI24495
AI24497 (01)00380740137366(21)AI24497
AI24504 (01)00380740137366(21)AI24504
AI24505 (01)00380740137366(21)AI24505
AI24506 (01)00380740137366(21)AI24506
AI24509 (01)00380740137366(21)AI24509
AI24510 (01)00380740137366(21)AI24510
AI24514 (01)00380740137366(21)AI24514
AI24518 (01)00380740137366(21)AI24518
AI24520 (01)00380740137366(21)AI24520
AI24521 (01)00380740137366(21)AI24521
AI24540 (01)00380740137366(21)AI24540
AI24544 (01)00380740137366(21)AI24544
AI24545 (01)00380740137366(21)AI24545
AI24548 (01)00380740137366(21)AI24548
AI24550 (01)00380740137366(21)AI24550
AI24551 (01)00380740137366(21)AI24551
AI24552 (01)00380740137366(21)AI24552
AI24559 (01)00380740137366(21)AI24559
AI24560 (01)00380740137366(21)AI24560
AI24574 (01)00380740137366(21)AI24574
AI24575 (01)00380740137366(21)AI24575
AI24581 (01)00380740137366(21)AI24581
AI24584 (01)00380740137366(21)AI24584
AI24591 (01)00380740137366(21)AI24591
AI24593 (01)00380740137366(21)AI24593
AI24603 (01)00380740137366(21)AI24603
AI24605 (01)00380740137366(21)AI24605
AI24615 (01)00380740137366(21)AI24615
AI24889 (01)00380740137366(21)AI24889
AI24892 (01)00380740137366(21)AI24892
AI24936 (01)00380740137366(21)AI24936
AI24943 (01)00380740137366(21)AI24943
AI24950 (01)00380740137366(21)AI24950
AI24980 (01)00380740137366(21)AI24980
AI24991 (01)00380740137366(21)AI24991
AI24995 (01)00380740137366(21)AI24995
AI24998 (01)00380740137366(21)AI24998
AI25008 (01)00380740137366(21)AI25008
AI25009 (01)00380740137366(21)AI25009
AI25011 (01)00380740137366(21)AI25011
AI25012 (01)00380740137366(21)AI25012
AI25018 (01)00380740137366(21)AI25018
AI25029 (01)00380740137366(21)AI25029
AI25036 (01)00380740137366(21)AI25036
AI25042 (01)00380740137366(21)AI25042
AI25054 (01)00380740137366(21)AI25054
AI25058 (01)00380740137366(21)AI25058
AI25059 (01)00380740137366(21)AI25059
AI25065 (01)00380740137366(21)AI25065
AI25073 (01)00380740137366(21)AI25073
AI25132 (01)00380740137366(21)AI25132
AI25138 (01)00380740137366(21)AI25138
AI25141 (01)00380740137366(21)AI25141
AI25203 (01)00380740137366(21)AI25203
AI25204 (01)00380740137366(21)AI25204
AI25206 (01)00380740137366(21)AI25206
AI25207 (01)00380740137366(21)AI25207
AI25208 (01)00380740137366(21)AI25208
AI25214 (01)00380740137366(21)AI25214
AI25216 (01)00380740137366(21)AI25216
AI25219 (01)00380740137366(21)AI25219
AI25254 (01)00380740137366(21)AI25254
AI25257 (01)00380740137366(21)AI25257
AI25261 (01)00380740137366(21)AI25261
AI25262 (01)00380740137366(21)AI25262
AI25276 (01)00380740137366(21)AI25276
AI25336 (01)00380740137366(21)AI25336
AI25339 (01)00380740137366(21)AI25339
AI25342 (01)00380740137366(21)AI25342
AI25358 (01)00380740137366(21)AI25358
AI25360 (01)00380740137366(21)AI25360
AI25366 (01)00380740137366(21)AI25366
AI25368 (01)00380740137366(21)AI25368
AI25369 (01)00380740137366(21)AI25369
AI25371 (01)00380740137366(21)AI25371
AI25373 (01)00380740137366(21)AI25373
AI25384 (01)00380740137366(21)AI25384
AI25385 (01)00380740137366(21)AI25385
AI25439 (01)00380740137366(21)AI25439
AI25451 (01)00380740137366(21)AI25451
AI25460 (01)00380740137366(21)AI25460
AI25461 (01)00380740137366(21)AI25461
AI25462 (01)00380740137366(21)AI25462
AI25463 (01)00380740137366(21)AI25463
AI25478 (01)00380740137366(21)AI25478
AI25484 (01)00380740137366(21)AI25484
AI25524 (01)00380740137366(21)AI25524
AI25526 (01)00380740137366(21)AI25526
AI25528 (01)00380740137366(21)AI25528
AI25536 (01)00380740137366(21)AI25536
AI25537 (01)00380740137366(21)AI25537
AI25539 (01)00380740137366(21)AI25539
AI25540 (01)00380740137366(21)AI25540
AI25541 (01)00380740137366(21)AI25541
AI25542 (01)00380740137366(21)AI25542
AI25543 (01)00380740137366(21)AI25543
AI25544 (01)00380740137366(21)AI25544
AI25545 (01)00380740137366(21)AI25545
AI25547 (01)00380740137366(21)AI25547
AI25549 (01)00380740137366(21)AI25549
AI25555 (01)00380740137366(21)AI25555
AI25566 (01)00380740137366(21)AI25566
AI25612 (01)00380740137366(21)AI25612
AI25658 (01)00380740137366(21)AI25658
AI25660 (01)00380740137366(21)AI25660
AI25661 (01)00380740137366(21)AI25661
AI25662 (01)00380740137366(21)AI25662
AI25663 (01)00380740137366(21)AI25663
AI25664 (01)00380740137366(21)AI25664
AI25666 (01)00380740137366(21)AI25666
AI25667 (01)00380740137366(21)AI25667
AI25668 (01)00380740137366(21)AI25668
AI25669 (01)00380740137366(21)AI25669
AI25670 (01)00380740137366(21)AI25670
AI25671 (01)00380740137366(21)AI25671
AI25672 (01)00380740137366(21)AI25672
AI25676 (01)00380740137366(21)AI25676
AI25677 (01)00380740137366(21)AI25677
AI25738 (01)00380740137366(21)AI25738
AI25741 (01)00380740137366(21)AI25741
AI25748 (01)00380740137366(21)AI25748
AI25751 (01)00380740137366(21)AI25751
AI25753 (01)00380740137366(21)AI25753
AI25754 (01)00380740137366(21)AI25754
AI25756 (01)00380740137366(21)AI25756
AI25759 (01)00380740137366(21)AI25759
AI25761 (01)00380740137366(21)AI25761
AI25762 (01)00380740137366(21)AI25762
AI25764 (01)00380740137366(21)AI25764
AI25765 (01)00380740137366(21)AI25765
AI25768 (01)00380740137366(21)AI25768
AI25769 (01)00380740137366(21)AI25769
AI25777 (01)00380740137366(21)AI25777
AI25779 (01)00380740137366(21)AI25779
AI25780 (01)00380740137366(21)AI25780
AI25788 (01)00380740137366(21)AI25788
AI25798 (01)00380740137366(21)AI25798
AI25844 (01)00380740137366(21)AI25844
AI25851 (01)00380740137366(21)AI25851
AI25854 (01)00380740137366(21)AI25854
AI25863 (01)00380740137366(21)AI25863
AI25864 (01)00380740137366(21)AI25864
AI25867 (01)00380740137366(21)AI25867
AI25868 (01)00380740137366(21)AI25868
AI25871 (01)00380740137366(21)AI25871
AI25873 (01)00380740137366(21)AI25873
AI25874 (01)00380740137366(21)AI25874
AI25875 (01)00380740137366(21)AI25875
AI25876 (01)00380740137366(21)AI25876
AI25883 (01)00380740137366(21)AI25883
AI25884 (01)00380740137366(21)AI25884
AI25885 (01)00380740137366(21)AI25885
AI25886 (01)00380740137366(21)AI25886
AI25887 (01)00380740137366(21)AI25887
AI25888 (01)00380740137366(21)AI25888
AI25889 (01)00380740137366(21)AI25889
AI25890 (01)00380740137366(21)AI25890
AI25892 (01)00380740137366(21)AI25892
AI25901 (01)00380740137366(21)AI25901
AI25903 (01)00380740137366(21)AI25903
AI25919 (01)00380740137366(21)AI25919
AI25924 (01)00380740137366(21)AI25924
AI25932 (01)00380740137366(21)AI25932
AI25936 (01)00380740137366(21)AI25936
AI25938 (01)00380740137366(21)AI25938
AI25941 (01)00380740137366(21)AI25941
AI25942 (01)00380740137366(21)AI25942
AI25946 (01)00380740137366(21)AI25946
AI25972 (01)00380740137366(21)AI25972
AI25994 (01)00380740137366(21)AI25994
AI26013 (01)00380740137366(21)AI26013
AI26014 (01)00380740137366(21)AI26014
AI26015 (01)00380740137366(21)AI26015
AI26021 (01)00380740137366(21)AI26021
AI26025 (01)00380740137366(21)AI26025
AI26028 (01)00380740137366(21)AI26028
AI26034 (01)00380740137366(21)AI26034
AI26035 (01)00380740137366(21)AI26035
AI26038 (01)00380740137366(21)AI26038
AI26039 (01)00380740137366(21)AI26039
AI26042 (01)00380740137366(21)AI26042
AI26043 (01)00380740137366(21)AI26043
AI26045 (01)00380740137366(21)AI26045
AI26046 (01)00380740137366(21)AI26046
AI26049 (01)00380740137366(21)AI26049
AI26051 (01)00380740137366(21)AI26051
AI26058 (01)00380740137366(21)AI26058
AI26062 (01)00380740137366(21)AI26062
AI26067 (01)00380740137366(21)AI26067
AI26068 (01)00380740137366(21)AI26068
AI26074 (01)00380740137366(21)AI26074
AI26075 (01)00380740137366(21)AI26075
AI26088 (01)00380740137366(21)AI26088
AI26122 (01)00380740137366(21)AI26122
AI26123 (01)00380740137366(21)AI26123
AI26124 (01)00380740137366(21)AI26124
AI26126 (01)00380740137366(21)AI26126
AI26127 (01)00380740137366(21)AI26127
AI26128 (01)00380740137366(21)AI26128
AI26131 (01)00380740137366(21)AI26131
AI26133 (01)00380740137366(21)AI26133
AI26139 (01)00380740137366(21)AI26139
AI26140 (01)00380740137366(21)AI26140
AI26142 (01)00380740137366(21)AI26142
AI26145 (01)00380740137366(21)AI26145
AI26156 (01)00380740137366(21)AI26156
AI26162 (01)00380740137366(21)AI26162
AI26165 (01)00380740137366(21)AI26165
AI26172 (01)00380740137366(21)AI26172
AI26173 (01)00380740137366(21)AI26173
AI26175 (01)00380740137366(21)AI26175
AI26179 (01)00380740137366(21)AI26179
AI26186 (01)00380740137366(21)AI26186
AI26191 (01)00380740137366(21)AI26191
AI26194 (01)00380740137366(21)AI26194
AI26196 (01)00380740137366(21)AI26196
AI26197 (01)00380740137366(21)AI26197
AI26199 (01)00380740137366(21)AI26199
AI26229 (01)00380740137366(21)AI26229
AI26232 (01)00380740137366(21)AI26232
AI26237 (01)00380740137366(21)AI26237
AI26240 (01)00380740137366(21)AI26240
AI26241 (01)00380740137366(21)AI26241
AI26242 (01)00380740137366(21)AI26242
AI26247 (01)00380740137366(21)AI26247
AI26249 (01)00380740137366(21)AI26249
AI26250 (01)00380740137366(21)AI26250
AI26251 (01)00380740137366(21)AI26251
AI26252 (01)00380740137366(21)AI26252
AI26254 (01)00380740137366(21)AI26254
AI26259 (01)00380740137366(21)AI26259
AI26261 (01)00380740137366(21)AI26261
|
Recalling Firm/ Manufacturer |
Abbott Laboratories 1915 Hurd Dr Irving TX 75038-4313
|
For Additional Information Contact |
Suchin Song 224-668-0343
|
Manufacturer Reason for Recall |
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
On 10/19/2023, the firm sent an "Urgent Medical Device Correction" Letter via FedEx priority overnight express to customers informing them that Abbott has determined that, A labeling update is required for the products listed above to meet US regulatory requirement 21 CFR 801.437 (User labeling for devices that contain natural rubber) and Brazilian requirement RDC 37/2015 (medical devices and IVDs to include in the label a Standardized Latex Warning in Portuguese CONTAIN NATURAL LATEX. MAY CAUSE ALLERGIES ). These Abbott products include internal hardware parts and/or subassemblies that contain dry natural rubber (latex) which may cause allergic reactions, but do not include the required precautionary label. Additionally, the Alinity s System Solid Waste Container Rubber Band (LN 04U97-01), used to secure the biohazard bag to the solid waste container, contains dry natural rubber (latex).
Customer are instructed to:
1) Follow good laboratory practice by wearing impervious gloves and other personal protective equipment when working on the above listed Abbott product(s).
2) If you have forwarded the product(s) listed above to other laboratories, please inform them of this Product Correction and provide a copy of this letter.
3) Please retain this letter for your laboratory records.
A precautionary label will be applied to the affected products. An Abbott representative will schedule a service visit when the instrument label is available for application.
For questions or concerns, contact Customer Service at 1-877-4ABBOTT (24 hr/7 days a week). |
Quantity in Commerce |
1,325 systems |
Distribution |
U.S.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV
O.U.S.: Brazil |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|