Class 2 Device Recall Alinity |
|
Date Initiated by Firm |
October 19, 2023 |
Date Posted |
January 09, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-0682-2024 |
Recall Event ID |
93355 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product |
ARCHITECT c4000 Processing Module REF: 02P24-01 02P24-02 02P24-40 01R24-56 01R25-56
The Abbott ARCHITECT c4000 Processing Module is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. |
Code Information |
List Number:
02P24-01
Serial Numbers / UDI Codes:
C400911 (01)00380740003746(21)C400911
C402913 (01)00380740003746(21)C402913
C460004 (01)00380740003746(21)C460004
C460033 (01)00380740003746(21)C460033
C460055 (01)00380740003746(21)C460055
C460066 (01)00380740003746(21)C460066
C460082 (01)00380740003746(21)C460082
C460158 (01)00380740003746(21)C460158
C460200 (01)00380740003746(21)C460200
C460375 (01)00380740003746(21)C460375
C460385 (01)00380740003746(21)C460385
C460396 (01)00380740003746(21)C460396
C460397 (01)00380740003746(21)C460397
C460408 (01)00380740003746(21)C460408
C460410 (01)00380740003746(21)C460410
C460422 (01)00380740003746(21)C460422
C460435 (01)00380740003746(21)C460435
C460439 (01)00380740003746(21)C460439
C460523 (01)00380740003746(21)C460523
C460532 (01)00380740003746(21)C460532
C460588 (01)00380740003746(21)C460588
C460728 (01)00380740003746(21)C460728
C460767 (01)00380740003746(21)C460767
C460797 (01)00380740003746(21)C460797
C460798 (01)00380740003746(21)C460798
C460800 (01)00380740003746(21)C460800
C460826 (01)00380740003746(21)C460826
C460827 (01)00380740003746(21)C460827
C460830 (01)00380740003746(21)C460830
C460832 (01)00380740003746(21)C460832
C460833 (01)00380740003746(21)C460833
C460834 (01)00380740003746(21)C460834
C460846 (01)00380740003746(21)C460846
C460849 (01)00380740003746(21)C460849
C460857 (01)00380740003746(21)C460857
C460868 (01)00380740003746(21)C460868
C460875 (01)00380740003746(21)C460875
C460886 (01)00380740003746(21)C460886
C460927 (01)00380740003746(21)C460927
C460928 (01)00380740003746(21)C460928
C460939 (01)00380740003746(21)C460939
C460958 (01)00380740003746(21)C460958
C460993 (01)00380740003746(21)C460993
C461042 (01)00380740003746(21)C461042
C461092 (01)00380740003746(21)C461092
C461093 (01)00380740003746(21)C461093
C461100 (01)00380740003746(21)C461100
C461105 (01)00380740003746(21)C461105
C461112 (01)00380740003746(21)C461112
C461122 (01)00380740003746(21)C461122
C461124 (01)00380740003746(21)C461124
C461129 (01)00380740003746(21)C461129
C461130 (01)00380740003746(21)C461130
C461132 (01)00380740003746(21)C461132
C461133 (01)00380740003746(21)C461133
C461138 (01)00380740003746(21)C461138
C461148 (01)00380740003746(21)C461148
C461150 (01)00380740003746(21)C461150
C461151 (01)00380740003746(21)C461151
C461155 (01)00380740003746(21)C461155
C461172 (01)00380740003746(21)C461172
C461173 (01)00380740003746(21)C461173
C461175 (01)00380740003746(21)C461175
C461176 (01)00380740003746(21)C461176
C461192 (01)00380740003746(21)C461192
C461195 (01)00380740003746(21)C461195
C461208 (01)00380740003746(21)C461208
C461212 (01)00380740003746(21)C461212
C461220 (01)00380740003746(21)C461220
C461222 (01)00380740003746(21)C461222
C461247 (01)00380740003746(21)C461247
C461250 (01)00380740003746(21)C461250
C461271 (01)00380740003746(21)C461271
C461272 (01)00380740003746(21)C461272
C461408 (01)00380740003746(21)C461408
C461412 (01)00380740003746(21)C461412
C461413 (01)00380740003746(21)C461413
C461414 (01)00380740003746(21)C461414
C461420 (01)00380740003746(21)C461420
C461425 (01)00380740003746(21)C461425
C461446 (01)00380740003746(21)C461446
C461451 (01)00380740003746(21)C461451
C461453 (01)00380740003746(21)C461453
C461455 (01)00380740003746(21)C461455
C461457 (01)00380740003746(21)C461457
C461459 (01)00380740003746(21)C461459
C461462 (01)00380740003746(21)C461462
C461473 (01)00380740003746(21)C461473
C461474 (01)00380740003746(21)C461474
C461475 (01)00380740003746(21)C461475
C461476 (01)00380740003746(21)C461476
C461477 (01)00380740003746(21)C461477
C461491 (01)00380740003746(21)C461491
C461492 (01)00380740003746(21)C461492
C461495 (01)00380740003746(21)C461495
C461496 (01)00380740003746(21)C461496
C461497 (01)00380740003746(21)C461497
C461498 (01)00380740003746(21)C461498
C461528 (01)00380740003746(21)C461528
C461553 (01)00380740003746(21)C461553
C461554 (01)00380740003746(21)C461554
C461555 (01)00380740003746(21)C461555
C461562 (01)00380740003746(21)C461562
C461563 (01)00380740003746(21)C461563
C461587 (01)00380740003746(21)C461587
C461597 (01)00380740003746(21)C461597
C461598 (01)00380740003746(21)C461598
C461621 (01)00380740003746(21)C461621
C461656 (01)00380740003746(21)C461656
C461658 (01)00380740003746(21)C461658
C461659 (01)00380740003746(21)C461659
C461660 (01)00380740003746(21)C461660
C461662 (01)00380740003746(21)C461662
C461663 (01)00380740003746(21)C461663
C461664 (01)00380740003746(21)C461664
C461667 (01)00380740003746(21)C461667
C461668 (01)00380740003746(21)C461668
C461674 (01)00380740003746(21)C461674
C461675 (01)00380740003746(21)C461675
C461676 (01)00380740003746(21)C461676
C461678 (01)00380740003746(21)C461678
C461680 (01)00380740003746(21)C461680
C461683 (01)00380740003746(21)C461683
C461684 (01)00380740003746(21)C461684
C461712 (01)00380740003746(21)C461712
C461713 (01)00380740003746(21)C461713
C461714 (01)00380740003746(21)C461714
C461716 (01)00380740003746(21)C461716
C461717 (01)00380740003746(21)C461717
C461720 (01)00380740003746(21)C461720
C461721 (01)00380740003746(21)C461721
C461723 (01)00380740003746(21)C461723
C461724 (01)00380740003746(21)C461724
C461751 (01)00380740003746(21)C461751
C461752 (01)00380740003746(21)C461752
C461770 (01)00380740003746(21)C461770
C461771 (01)00380740003746(21)C461771
C461772 (01)00380740003746(21)C461772
C461781 (01)00380740003746(21)C461781
C461782 (01)00380740003746(21)C461782
C461783 (01)00380740003746(21)C461783
C461800 (01)00380740003746(21)C461800
C461801 (01)00380740003746(21)C461801
C461802 (01)00380740003746(21)C461802
C461804 (01)00380740003746(21)C461804
C461820 (01)00380740003746(21)C461820
C461822 (01)00380740003746(21)C461822
C461823 (01)00380740003746(21)C461823
C461835 (01)00380740003746(21)C461835
C461837 (01)00380740003746(21)C461837
C461839 (01)00380740003746(21)C461839
C461840 (01)00380740003746(21)C461840
C461843 (01)00380740003746(21)C461843
C461846 (01)00380740003746(21)C461846
C461847 (01)00380740003746(21)C461847
C461848 (01)00380740003746(21)C461848
C461855 (01)00380740003746(21)C461855
C461856 (01)00380740003746(21)C461856
C461857 (01)00380740003746(21)C461857
C461873 (01)00380740003746(21)C461873
C461875 (01)00380740003746(21)C461875
C461881 (01)00380740003746(21)C461881
C461894 (01)00380740003746(21)C461894
C461895 (01)00380740003746(21)C461895
C461906 (01)00380740003746(21)C461906
C461919 (01)00380740003746(21)C461919
C461924 (01)00380740003746(21)C461924
C461931 (01)00380740003746(21)C461931
C461932 (01)00380740003746(21)C461932
C461934 (01)00380740003746(21)C461934
C461958 (01)00380740003746(21)C461958
C462013 (01)00380740003746(21)C462013
C462014 (01)00380740003746(21)C462014
C462015 (01)00380740003746(21)C462015
C462070 (01)00380740003746(21)C462070
C462081 (01)00380740003746(21)C462081
C462100 (01)00380740003746(21)C462100
C462108 (01)00380740003746(21)C462108
C462109 (01)00380740003746(21)C462109
C462116 (01)00380740003746(21)C462116
C462117 (01)00380740003746(21)C462117
C462123 (01)00380740003746(21)C462123
C462134 (01)00380740003746(21)C462134
C462135 (01)00380740003746(21)C462135
C462139 (01)00380740003746(21)C462139
C462144 (01)00380740003746(21)C462144
C462154 (01)00380740003746(21)C462154
C462156 (01)00380740003746(21)C462156
C462215 (01)00380740003746(21)C462215
C462216 (01)00380740003746(21)C462216
C462217 (01)00380740003746(21)C462217
C462218 (01)00380740003746(21)C462218
C462219 (01)00380740003746(21)C462219
C462220 (01)00380740003746(21)C462220
C462222 (01)00380740003746(21)C462222
C462234 (01)00380740003746(21)C462234
C462236 (01)00380740003746(21)C462236
C462239 (01)00380740003746(21)C462239
C462240 (01)00380740003746(21)C462240
C462241 (01)00380740003746(21)C462241
C462245 (01)00380740003746(21)C462245
C462246 (01)00380740003746(21)C462246
C462247 (01)00380740003746(21)C462247
C462249 (01)00380740003746(21)C462249
C462250 (01)00380740003746(21)C462250
C462254 (01)00380740003746(21)C462254
C462255 (01)00380740003746(21)C462255
C462256 (01)00380740003746(21)C462256
C462257 (01)00380740003746(21)C462257
C462309 (01)00380740003746(21)C462309
C462312 (01)00380740003746(21)C462312
C462313 (01)00380740003746(21)C462313
C462347 (01)00380740003746(21)C462347
C462384 (01)00380740003746(21)C462384
C462386 (01)00380740003746(21)C462386
C462402 (01)00380740003746(21)C462402
C462403 (01)00380740003746(21)C462403
C462404 (01)00380740003746(21)C462404
C462434 (01)00380740003746(21)C462434
C462436 (01)00380740003746(21)C462436
C462437 (01)00380740003746(21)C462437
C462439 (01)00380740003746(21)C462439
C462441 (01)00380740003746(21)C462441
C462446 (01)00380740003746(21)C462446
C462450 (01)00380740003746(21)C462450
C462506 (01)00380740003746(21)C462506
C462536 (01)00380740003746(21)C462536
02P24-02
Serial Numbers / UDI codes:
C460391 (01)00380740081225(21)C460391
C460394 (01)00380740081225(21)C460394
C460395 (01)00380740081225(21)C460395
C460430 (01)00380740081225(21)C460430
C460434 (01)00380740081225(21)C460434
C461123 (01)00380740081225(21)C461123
02P24-40
Serial Numbers / UDI codes:
C400062 (01)00380740003753(21)C400062
C400251 (01)00380740003753(21)C400251
C400268 (01)00380740003753(21)C400268
C400288 (01)00380740003753(21)C400288
C400350 (01)00380740003753(21)C400350
C400392 (01)00380740003753(21)C400392
C400399 (01)00380740003753(21)C400399
C400404 (01)00380740003753(21)C400404
C400412 (01)00380740003753(21)C400412
C400416 (01)00380740003753(21)C400416
C400430 (01)00380740003753(21)C400430
C400431 (01)00380740003753(21)C400431
C400445 (01)00380740003753(21)C400445
C400479 (01)00380740003753(21)C400479
C400491 (01)00380740003753(21)C400491
C400500 (01)00380740003753(21)C400500
C400515 (01)00380740003753(21)C400515
C400552 (01)00380740003753(21)C400552
C400561 (01)00380740003753(21)C400561
C400578 (01)00380740003753(21)C400578
C400624 (01)00380740003753(21)C400624
C400660 (01)00380740003753(21)C400660
C400668 (01)00380740003753(21)C400668
C400673 (01)00380740003753(21)C400673
C400680 (01)00380740003753(21)C400680
C400686 (01)00380740003753(21)C400686
C400689 (01)00380740003753(21)C400689
C400709 (01)00380740003753(21)C400709
C400710 (01)00380740003753(21)C400710
C400716 (01)00380740003753(21)C400716
C400725 (01)00380740003753(21)C400725
C400727 (01)00380740003753(21)C400727
C400741 (01)00380740003753(21)C400741
C400749 (01)00380740003753(21)C400749
C400750 (01)00380740003753(21)C400750
C400756 (01)00380740003753(21)C400756
C400757 (01)00380740003753(21)C400757
C400760 (01)00380740003753(21)C400760
C400761 (01)00380740003753(21)C400761
C400799 (01)00380740003753(21)C400799
C400804 (01)00380740003753(21)C400804
C400809 (01)00380740003753(21)C400809
C400811 (01)00380740003753(21)C400811
C400812 (01)00380740003753(21)C400812
C400816 (01)00380740003753(21)C400816
C400818 (01)00380740003753(21)C400818
C400836 (01)00380740003753(21)C400836
C400839 (01)00380740003753(21)C400839
C400842 (01)00380740003753(21)C400842
C400844 (01)00380740003753(21)C400844
C400847 (01)00380740003753(21)C400847
C400848 (01)00380740003753(21)C400848
C400857 (01)00380740003753(21)C400857
C400861 (01)00380740003753(21)C400861
C400870 (01)00380740003753(21)C400870
C400881 (01)00380740003753(21)C400881
C400884 (01)00380740003753(21)C400884
C400890 (01)00380740003753(21)C400890
C400927 (01)00380740003753(21)C400927
C400948 (01)00380740003753(21)C400948
C400993 (01)00380740003753(21)C400993
C400994 (01)00380740003753(21)C400994
C401003 (01)00380740003753(21)C401003
C401008 (01)00380740003753(21)C401008
C401010 (01)00380740003753(21)C401010
C401025 (01)00380740003753(21)C401025
C401027 (01)00380740003753(21)C401027
C401052 (01)00380740003753(21)C401052
C401080 (01)00380740003753(21)C401080
C401082 (01)00380740003753(21)C401082
C401083 (01)00380740003753(21)C401083
C401113 (01)00380740003753(21)C401113
C401119 (01)00380740003753(21)C401119
C401131 (01)00380740003753(21)C401131
C401149 (01)00380740003753(21)C401149
C401161 (01)00380740003753(21)C401161
C401180 (01)00380740003753(21)C401180
C401181 (01)00380740003753(21)C401181
C401182 (01)00380740003753(21)C401182
C401188 (01)00380740003753(21)C401188
C401189 (01)00380740003753(21)C401189
C401199 (01)00380740003753(21)C401199
C401204 (01)00380740003753(21)C401204
C401212 (01)00380740003753(21)C401212
C401213 (01)00380740003753(21)C401213
C401214 (01)00380740003753(21)C401214
C401217 (01)00380740003753(21)C401217
C401219 (01)00380740003753(21)C401219
C401231 (01)00380740003753(21)C401231
C401233 (01)00380740003753(21)C401233
C401234 (01)00380740003753(21)C401234
C401237 (01)00380740003753(21)C401237
C401251 (01)00380740003753(21)C401251
C401252 (01)00380740003753(21)C401252
C401253 (01)00380740003753(21)C401253
C401257 (01)00380740003753(21)C401257
C401263 (01)00380740003753(21)C401263
C401264 (01)00380740003753(21)C401264
C401265 (01)00380740003753(21)C401265
C401275 (01)00380740003753(21)C401275
C401282 (01)00380740003753(21)C401282
C401283 (01)00380740003753(21)C401283
C401284 (01)00380740003753(21)C401284
C401289 (01)00380740003753(21)C401289
C401294 (01)00380740003753(21)C401294
C401295 (01)00380740003753(21)C401295
C401301 (01)00380740003753(21)C401301
C401321 (01)00380740003753(21)C401321
C401341 (01)00380740003753(21)C401341
C401364 (01)00380740003753(21)C401364
C401367 (01)00380740003753(21)C401367
C401368 (01)00380740003753(21)C401368
C401369 (01)00380740003753(21)C401369
C401372 (01)00380740003753(21)C401372
C401373 (01)00380740003753(21)C401373
C401387 (01)00380740003753(21)C401387
C401388 (01)00380740003753(21)C401388
C401397 (01)00380740003753(21)C401397
C401399 (01)00380740003753(21)C401399
C401405 (01)00380740003753(21)C401405
C401406 (01)00380740003753(21)C401406
C401408 (01)00380740003753(21)C401408
C401409 (01)00380740003753(21)C401409
C401416 (01)00380740003753(21)C401416
C401417 (01)00380740003753(21)C401417
C401424 (01)00380740003753(21)C401424
C401425 (01)00380740003753(21)C401425
C401435 (01)00380740003753(21)C401435
C401452 (01)00380740003753(21)C401452
C401453 (01)00380740003753(21)C401453
C401456 (01)00380740003753(21)C401456
C401463 (01)00380740003753(21)C401463
C401468 (01)00380740003753(21)C401468
C401470 (01)00380740003753(21)C401470
C401484 (01)00380740003753(21)C401484
C401485 (01)00380740003753(21)C401485
C401486 (01)00380740003753(21)C401486
C401487 (01)00380740003753(21)C401487
C401488 (01)00380740003753(21)C401488
C401490 (01)00380740003753(21)C401490
C401492 (01)00380740003753(21)C401492
C401513 (01)00380740003753(21)C401513
C401517 (01)00380740003753(21)C401517
C401520 (01)00380740003753(21)C401520
C401522 (01)00380740003753(21)C401522
C401524 (01)00380740003753(21)C401524
C401532 (01)00380740003753(21)C401532
C401537 (01)00380740003753(21)C401537
C401538 (01)00380740003753(21)C401538
C401540 (01)00380740003753(21)C401540
C401543 (01)00380740003753(21)C401543
C401546 (01)00380740003753(21)C401546
C401548 (01)00380740003753(21)C401548
C401551 (01)00380740003753(21)C401551
C401552 (01)00380740003753(21)C401552
C401553 (01)00380740003753(21)C401553
C401555 (01)00380740003753(21)C401555
C401557 (01)00380740003753(21)C401557
C401558 (01)00380740003753(21)C401558
C401565 (01)00380740003753(21)C401565
C401568 (01)00380740003753(21)C401568
C401574 (01)00380740003753(21)C401574
C401575 (01)00380740003753(21)C401575
C401578 (01)00380740003753(21)C401578
C401585 (01)00380740003753(21)C401585
C401590 (01)00380740003753(21)C401590
C401599 (01)00380740003753(21)C401599
C401604 (01)00380740003753(21)C401604
C401617 (01)00380740003753(21)C401617
C401620 (01)00380740003753(21)C401620
C401624 (01)00380740003753(21)C401624
C401626 (01)00380740003753(21)C401626
C401635 (01)00380740003753(21)C401635
C401636 (01)00380740003753(21)C401636
C401639 (01)00380740003753(21)C401639
C401641 (01)00380740003753(21)C401641
C401642 (01)00380740003753(21)C401642
C401645 (01)00380740003753(21)C401645
C401648 (01)00380740003753(21)C401648
C401649 (01)00380740003753(21)C401649
C401653 (01)00380740003753(21)C401653
C401664 (01)00380740003753(21)C401664
C401666 (01)00380740003753(21)C401666
C401672 (01)00380740003753(21)C401672
C401674 (01)00380740003753(21)C401674
C401682 (01)00380740003753(21)C401682
C401683 (01)00380740003753(21)C401683
C401684 (01)00380740003753(21)C401684
C401685 (01)00380740003753(21)C401685
C401687 (01)00380740003753(21)C401687
C401689 (01)00380740003753(21)C401689
C401691 (01)00380740003753(21)C401691
C401693 (01)00380740003753(21)C401693
C401697 (01)00380740003753(21)C401697
C401698 (01)00380740003753(21)C401698
C401700 (01)00380740003753(21)C401700
C401703 (01)00380740003753(21)C401703
C401704 (01)00380740003753(21)C401704
C401705 (01)00380740003753(21)C401705
C401706 (01)00380740003753(21)C401706
C401708 (01)00380740003753(21)C401708
C401710 (01)00380740003753(21)C401710
C401711 (01)00380740003753(21)C401711
C401715 (01)00380740003753(21)C401715
C401716 (01)00380740003753(21)C401716
C401732 (01)00380740003753(21)C401732
C401733 (01)00380740003753(21)C401733
C401772 (01)00380740003753(21)C401772
C401773 (01)00380740003753(21)C401773
C401774 (01)00380740003753(21)C401774
C401783 (01)00380740003753(21)C401783
C401784 (01)00380740003753(21)C401784
C401793 (01)00380740003753(21)C401793
C401825 (01)00380740003753(21)C401825
C401826 (01)00380740003753(21)C401826
C401828 (01)00380740003753(21)C401828
C401838 (01)00380740003753(21)C401838
C401839 (01)00380740003753(21)C401839
C401849 (01)00380740003753(21)C401849
C401850 (01)00380740003753(21)C401850
C401855 (01)00380740003753(21)C401855
C401856 (01)00380740003753(21)C401856
C401857 (01)00380740003753(21)C401857
C401880 (01)00380740003753(21)C401880
C401881 (01)00380740003753(21)C401881
C401882 (01)00380740003753(21)C401882
C401885 (01)00380740003753(21)C401885
C401886 (01)00380740003753(21)C401886
C401887 (01)00380740003753(21)C401887
C401888 (01)00380740003753(21)C401888
C401889 (01)00380740003753(21)C401889
C401895 (01)00380740003753(21)C401895
C401897 (01)00380740003753(21)C401897
C401906 (01)00380740003753(21)C401906
C401946 (01)00380740003753(21)C401946
C401947 (01)00380740003753(21)C401947
C401949 (01)00380740003753(21)C401949
C401950 (01)00380740003753(21)C401950
C401951 (01)00380740003753(21)C401951
C401954 (01)00380740003753(21)C401954
C401955 (01)00380740003753(21)C401955
C401956 (01)00380740003753(21)C401956
C401957 (01)00380740003753(21)C401957
C401958 (01)00380740003753(21)C401958
C401959 (01)00380740003753(21)C401959
C401975 (01)00380740003753(21)C401975
C401976 (01)00380740003753(21)C401976
C401977 (01)00380740003753(21)C401977
C401978 (01)00380740003753(21)C401978
C401979 (01)00380740003753(21)C401979
C401986 (01)00380740003753(21)C401986
C401988 (01)00380740003753(21)C401988
C401989 (01)00380740003753(21)C401989
C401990 (01)00380740003753(21)C401990
C401993 (01)00380740003753(21)C401993
C401996 (01)00380740003753(21)C401996
C401997 (01)00380740003753(21)C401997
C401998 (01)00380740003753(21)C401998
C401999 (01)00380740003753(21)C401999
C402007 (01)00380740003753(21)C402007
C402008 (01)00380740003753(21)C402008
C402009 (01)00380740003753(21)C402009
C402011 (01)00380740003753(21)C402011
C402012 (01)00380740003753(21)C402012
C402013 (01)00380740003753(21)C402013
C402024 (01)00380740003753(21)C402024
C402025 (01)00380740003753(21)C402025
C402026 (01)00380740003753(21)C402026
C402027 (01)00380740003753(21)C402027
C402029 (01)00380740003753(21)C402029
C402036 (01)00380740003753(21)C402036
C402037 (01)00380740003753(21)C402037
C402038 (01)00380740003753(21)C402038
C402039 (01)00380740003753(21)C402039
C402040 (01)00380740003753(21)C402040
C402041 (01)00380740003753(21)C402041
C402042 (01)00380740003753(21)C402042
C402043 (01)00380740003753(21)C402043
C402044 (01)00380740003753(21)C402044
C402045 (01)00380740003753(21)C402045
C402046 (01)00380740003753(21)C402046
C402047 (01)00380740003753(21)C402047
C402048 (01)00380740003753(21)C402048
C402049 (01)00380740003753(21)C402049
C402050 (01)00380740003753(21)C402050
C402051 (01)00380740003753(21)C402051
C402052 (01)00380740003753(21)C402052
C402053 (01)00380740003753(21)C402053
C402054 (01)00380740003753(21)C402054
C402055 (01)00380740003753(21)C402055
C402056 (01)00380740003753(21)C402056
C402057 (01)00380740003753(21)C402057
C402058 (01)00380740003753(21)C402058
C402059 (01)00380740003753(21)C402059
C402060 (01)00380740003753(21)C402060
C402067 (01)00380740003753(21)C402067
C402068 (01)00380740003753(21)C402068
C402069 (01)00380740003753(21)C402069
C402070 (01)00380740003753(21)C402070
C402071 (01)00380740003753(21)C402071
C402072 (01)00380740003753(21)C402072
C402073 (01)00380740003753(21)C402073
C402074 (01)00380740003753(21)C402074
C402075 (01)00380740003753(21)C402075
C402076 (01)00380740003753(21)C402076
C402079 (01)00380740003753(21)C402079
C402080 (01)00380740003753(21)C402080
C402081 (01)00380740003753(21)C402081
C402083 (01)00380740003753(21)C402083
C402086 (01)00380740003753(21)C402086
C402087 (01)00380740003753(21)C402087
C402088 (01)00380740003753(21)C402088
C402106 (01)00380740003753(21)C402106
C402107 (01)00380740003753(21)C402107
C402110 (01)00380740003753(21)C402110
C402119 (01)00380740003753(21)C402119
C402124 (01)00380740003753(21)C402124
C402126 (01)00380740003753(21)C402126
C402148 (01)00380740003753(21)C402148
C402149 (01)00380740003753(21)C402149
C402150 (01)00380740003753(21)C402150
C402165 (01)00380740003753(21)C402165
C402166 (01)00380740003753(21)C402166
C402167 (01)00380740003753(21)C402167
C402168 (01)00380740003753(21)C402168
C402171 (01)00380740003753(21)C402171
C402172 (01)00380740003753(21)C402172
C402173 (01)00380740003753(21)C402173
C402182 (01)00380740003753(21)C402182
C402183 (01)00380740003753(21)C402183
C402184 (01)00380740003753(21)C402184
C402202 (01)00380740003753(21)C402202
C402203 (01)00380740003753(21)C402203
C402204 (01)00380740003753(21)C402204
C402206 (01)00380740003753(21)C402206
C402210 (01)00380740003753(21)C402210
C402233 (01)00380740003753(21)C402233
C402234 (01)00380740003753(21)C402234
C402235 (01)00380740003753(21)C402235
C402236 (01)00380740003753(21)C402236
C402237 (01)00380740003753(21)C402237
C402238 (01)00380740003753(21)C402238
C402247 (01)00380740003753(21)C402247
C402248 (01)00380740003753(21)C402248
C402249 (01)00380740003753(21)C402249
C402251 (01)00380740003753(21)C402251
C402252 (01)00380740003753(21)C402252
C402253 (01)00380740003753(21)C402253
C402254 (01)00380740003753(21)C402254
C402256 (01)00380740003753(21)C402256
C402267 (01)00380740003753(21)C402267
C402268 (01)00380740003753(21)C402268
C402280 (01)00380740003753(21)C402280
C402281 (01)00380740003753(21)C402281
C402282 (01)00380740003753(21)C402282
C402284 (01)00380740003753(21)C402284
C402285 (01)00380740003753(21)C402285
C402286 (01)00380740003753(21)C402286
C402287 (01)00380740003753(21)C402287
C402288 (01)00380740003753(21)C402288
C402289 (01)00380740003753(21)C402289
C402290 (01)00380740003753(21)C402290
C402291 (01)00380740003753(21)C402291
C402292 (01)00380740003753(21)C402292
C402296 (01)00380740003753(21)C402296
C402297 (01)00380740003753(21)C402297
C402298 (01)00380740003753(21)C402298
C402299 (01)00380740003753(21)C402299
C402301 (01)00380740003753(21)C402301
C402302 (01)00380740003753(21)C402302
C402304 (01)00380740003753(21)C402304
C402305 (01)00380740003753(21)C402305
C402306 (01)00380740003753(21)C402306
C402307 (01)00380740003753(21)C402307
C402308 (01)00380740003753(21)C402308
C402311 (01)00380740003753(21)C402311
C402326 (01)00380740003753(21)C402326
C402327 (01)00380740003753(21)C402327
C402331 (01)00380740003753(21)C402331
C402332 (01)00380740003753(21)C402332
C402333 (01)00380740003753(21)C402333
C402335 (01)00380740003753(21)C402335
C402336 (01)00380740003753(21)C402336
C402337 (01)00380740003753(21)C402337
C402338 (01)00380740003753(21)C402338
C402339 (01)00380740003753(21)C402339
C402346 (01)00380740003753(21)C402346
C402348 (01)00380740003753(21)C402348
C402349 (01)00380740003753(21)C402349
C402350 (01)00380740003753(21)C402350
C402354 (01)00380740003753(21)C402354
C402355 (01)00380740003753(21)C402355
C402356 (01)00380740003753(21)C402356
C402357 (01)00380740003753(21)C402357
C402358 (01)00380740003753(21)C402358
C402359 (01)00380740003753(21)C402359
C402361 (01)00380740003753(21)C402361
C402390 (01)00380740003753(21)C402390
C402391 (01)00380740003753(21)C402391
C402392 (01)00380740003753(21)C402392
C402393 (01)00380740003753(21)C402393
C402394 (01)00380740003753(21)C402394
C402395 (01)00380740003753(21)C402395
C402396 (01)00380740003753(21)C402396
C402397 (01)00380740003753(21)C402397
C402402 (01)00380740003753(21)C402402
C402403 (01)00380740003753(21)C402403
C402404 (01)00380740003753(21)C402404
C402405 (01)00380740003753(21)C402405
C402406 (01)00380740003753(21)C402406
C402407 (01)00380740003753(21)C402407
C402410 (01)00380740003753(21)C402410
C402411 (01)00380740003753(21)C402411
C402412 (01)00380740003753(21)C402412
C402413 (01)00380740003753(21)C402413
C402416 (01)00380740003753(21)C402416
C402417 (01)00380740003753(21)C402417
C402419 (01)00380740003753(21)C402419
C402422 (01)00380740003753(21)C402422
C402423 (01)00380740003753(21)C402423
C402424 (01)00380740003753(21)C402424
C402430 (01)00380740003753(21)C402430
C402431 (01)00380740003753(21)C402431
C402432 (01)00380740003753(21)C402432
C402434 (01)00380740003753(21)C402434
C402435 (01)00380740003753(21)C402435
C402436 (01)00380740003753(21)C402436
C402437 (01)00380740003753(21)C402437
C402438 (01)00380740003753(21)C402438
C402439 (01)00380740003753(21)C402439
C402440 (01)00380740003753(21)C402440
C402441 (01)00380740003753(21)C402441
C402455 (01)00380740003753(21)C402455
C402456 (01)00380740003753(21)C402456
C402457 (01)00380740003753(21)C402457
C402458 (01)00380740003753(21)C402458
C402459 (01)00380740003753(21)C402459
C402479 (01)00380740003753(21)C402479
C402480 (01)00380740003753(21)C402480
C402481 (01)00380740003753(21)C402481
C402483 (01)00380740003753(21)C402483
C402484 (01)00380740003753(21)C402484
C402485 (01)00380740003753(21)C402485
C402486 (01)00380740003753(21)C402486
C402487 (01)00380740003753(21)C402487
C402488 (01)00380740003753(21)C402488
C402490 (01)00380740003753(21)C402490
C402491 (01)00380740003753(21)C402491
C402492 (01)00380740003753(21)C402492
C402507 (01)00380740003753(21)C402507
C402508 (01)00380740003753(21)C402508
C402509 (01)00380740003753(21)C402509
C402510 (01)00380740003753(21)C402510
C402511 (01)00380740003753(21)C402511
C402512 (01)00380740003753(21)C402512
C402525 (01)00380740003753(21)C402525
C402527 (01)00380740003753(21)C402527
C402528 (01)00380740003753(21)C402528
C402531 (01)00380740003753(21)C402531
C402532 (01)00380740003753(21)C402532
C402533 (01)00380740003753(21)C402533
C402534 (01)00380740003753(21)C402534
C402536 (01)00380740003753(21)C402536
C402537 (01)00380740003753(21)C402537
C402538 (01)00380740003753(21)C402538
C402540 (01)00380740003753(21)C402540
C402559 (01)00380740003753(21)C402559
C402560 (01)00380740003753(21)C402560
C402562 (01)00380740003753(21)C402562
C402563 (01)00380740003753(21)C402563
C402564 (01)00380740003753(21)C402564
C402565 (01)00380740003753(21)C402565
C402567 (01)00380740003753(21)C402567
C402569 (01)00380740003753(21)C402569
C402575 (01)00380740003753(21)C402575
C402576 (01)00380740003753(21)C402576
C402577 (01)00380740003753(21)C402577
C402581 (01)00380740003753(21)C402581
C402589 (01)00380740003753(21)C402589
C402590 (01)00380740003753(21)C402590
C402592 (01)00380740003753(21)C402592
C402593 (01)00380740003753(21)C402593
C402594 (01)00380740003753(21)C402594
C402595 (01)00380740003753(21)C402595
C402596 (01)00380740003753(21)C402596
C402605 (01)00380740003753(21)C402605
C402607 (01)00380740003753(21)C402607
C402608 (01)00380740003753(21)C402608
C402611 (01)00380740003753(21)C402611
C402612 (01)00380740003753(21)C402612
C402613 (01)00380740003753(21)C402613
C402615 (01)00380740003753(21)C402615
C402616 (01)00380740003753(21)C402616
C402618 (01)00380740003753(21)C402618
C402622 (01)00380740003753(21)C402622
C402623 (01)00380740003753(21)C402623
C402624 (01)00380740003753(21)C402624
C402635 (01)00380740003753(21)C402635
C402638 (01)00380740003753(21)C402638
C402639 (01)00380740003753(21)C402639
C402642 (01)00380740003753(21)C402642
C402651 (01)00380740003753(21)C402651
C402658 (01)00380740003753(21)C402658
C402659 (01)00380740003753(21)C402659
C402660 (01)00380740003753(21)C402660
C402661 (01)00380740003753(21)C402661
C402662 (01)00380740003753(21)C402662
C402663 (01)00380740003753(21)C402663
C402665 (01)00380740003753(21)C402665
C402668 (01)00380740003753(21)C402668
C402669 (01)00380740003753(21)C402669
C402671 (01)00380740003753(21)C402671
C402672 (01)00380740003753(21)C402672
C402673 (01)00380740003753(21)C402673
C402674 (01)00380740003753(21)C402674
C402675 (01)00380740003753(21)C402675
C402692 (01)00380740003753(21)C402692
C402705 (01)00380740003753(21)C402705
C402707 (01)00380740003753(21)C402707
C402708 (01)00380740003753(21)C402708
C402709 (01)00380740003753(21)C402709
C402720 (01)00380740003753(21)C402720
C402721 (01)00380740003753(21)C402721
C402722 (01)00380740003753(21)C402722
C402739 (01)00380740003753(21)C402739
C402741 (01)00380740003753(21)C402741
C402754 (01)00380740003753(21)C402754
C402755 (01)00380740003753(21)C402755
C402756 (01)00380740003753(21)C402756
C402769 (01)00380740003753(21)C402769
C402770 (01)00380740003753(21)C402770
C402773 (01)00380740003753(21)C402773
C402774 (01)00380740003753(21)C402774
C402775 (01)00380740003753(21)C402775
C402776 (01)00380740003753(21)C402776
C402780 (01)00380740003753(21)C402780
C402781 (01)00380740003753(21)C402781
C402782 (01)00380740003753(21)C402782
C402784 (01)00380740003753(21)C402784
C402785 (01)00380740003753(21)C402785
C402786 (01)00380740003753(21)C402786
C402787 (01)00380740003753(21)C402787
C402788 (01)00380740003753(21)C402788
C402790 (01)00380740003753(21)C402790
C402791 (01)00380740003753(21)C402791
C402792 (01)00380740003753(21)C402792
C402793 (01)00380740003753(21)C402793
C402795 (01)00380740003753(21)C402795
C402796 (01)00380740003753(21)C402796
C402797 (01)00380740003753(21)C402797
C402806 (01)00380740003753(21)C402806
C402807 (01)00380740003753(21)C402807
C402808 (01)00380740003753(21)C402808
C402828 (01)00380740003753(21)C402828
C402829 (01)00380740003753(21)C402829
C402831 (01)00380740003753(21)C402831
C402834 (01)00380740003753(21)C402834
C402835 (01)00380740003753(21)C402835
C402836 (01)00380740003753(21)C402836
C402837 (01)00380740003753(21)C402837
C402838 (01)00380740003753(21)C402838
C402840 (01)00380740003753(21)C402840
C402845 (01)00380740003753(21)C402845
C402846 (01)00380740003753(21)C402846
C402848 (01)00380740003753(21)C402848
C402849 (01)00380740003753(21)C402849
C402850 (01)00380740003753(21)C402850
C402851 (01)00380740003753(21)C402851
C402852 (01)00380740003753(21)C402852
C402853 (01)00380740003753(21)C402853
C402854 (01)00380740003753(21)C402854
C402855 (01)00380740003753(21)C402855
C402856 (01)00380740003753(21)C402856
C402864 (01)00380740003753(21)C402864
C402865 (01)00380740003753(21)C402865
C402866 (01)00380740003753(21)C402866
C402867 (01)00380740003753(21)C402867
C402868 (01)00380740003753(21)C402868
C402895 (01)00380740003753(21)C402895
C402896 (01)00380740003753(21)C402896
C402898 (01)00380740003753(21)C402898
C402899 (01)00380740003753(21)C402899
C402900 (01)00380740003753(21)C402900
C402901 (01)00380740003753(21)C402901
C402902 (01)00380740003753(21)C402902
C402903 (01)00380740003753(21)C402903
C402904 (01)00380740003753(21)C402904
C402911 (01)00380740003753(21)C402911
C402912 (01)00380740003753(21)C402912
C402914 (01)00380740003753(21)C402914
C402938 (01)00380740003753(21)C402938
C402939 (01)00380740003753(21)C402939
C402940 (01)00380740003753(21)C402940
C402941 (01)00380740003753(21)C402941
C402953 (01)00380740003753(21)C402953
C402954 (01)00380740003753(21)C402954
C402955 (01)00380740003753(21)C402955
C402956 (01)00380740003753(21)C402956
C402958 (01)00380740003753(21)C402958
C402959 (01)00380740003753(21)C402959
C402960 (01)00380740003753(21)C402960
C402961 (01)00380740003753(21)C402961
C402962 (01)00380740003753(21)C402962
C402963 (01)00380740003753(21)C402963
C402982 (01)00380740003753(21)C402982
C402983 (01)00380740003753(21)C402983
C402985 (01)00380740003753(21)C402985
C402996 (01)00380740003753(21)C402996
C402997 (01)00380740003753(21)C402997
C402998 (01)00380740003753(21)C402998
C403006 (01)00380740003753(21)C403006
C403007 (01)00380740003753(21)C403007
C403008 (01)00380740003753(21)C403008
C403009 (01)00380740003753(21)C403009
C403018 (01)00380740003753(21)C403018
C403019 (01)00380740003753(21)C403019
C403025 (01)00380740003753(21)C403025
C403026 (01)00380740003753(21)C403026
C403027 (01)00380740003753(21)C403027
C403036 (01)00380740003753(21)C403036
C403060 (01)00380740003753(21)C403060
C403061 (01)00380740003753(21)C403061
C403062 (01)00380740003753(21)C403062
C403070 (01)00380740003753(21)C403070
C403071 (01)00380740003753(21)C403071
C403072 (01)00380740003753(21)C403072
C403073 (01)00380740003753(21)C403073
C403074 (01)00380740003753(21)C403074
C403075 (01)00380740003753(21)C403075
C403096 (01)00380740003753(21)C403096
C403128 (01)00380740003753(21)C403128
C403129 (01)00380740003753(21)C403129
C403130 (01)00380740003753(21)C403130
C403131 (01)00380740003753(21)C403131
C403132 (01)00380740003753(21)C403132
C403133 (01)00380740003753(21)C403133
C403154 (01)00380740003753(21)C403154
C403155 (01)00380740003753(21)C403155
C403156 (01)00380740003753(21)C403156
C403174 (01)00380740003753(21)C403174
C403178 (01)00380740003753(21)C403178
C403179 (01)00380740003753(21)C403179
C403180 (01)00380740003753(21)C403180
C403186 (01)00380740003753(21)C403186
C403187 (01)00380740003753(21)C403187
C403188 (01)00380740003753(21)C403188
C403191 (01)00380740003753(21)C403191
C403192 (01)00380740003753(21)C403192
C403193 (01)00380740003753(21)C403193
C403202 (01)00380740003753(21)C403202
C403203 (01)00380740003753(21)C403203
C403204 (01)00380740003753(21)C403204
C403214 (01)00380740003753(21)C403214
C403215 (01)00380740003753(21)C403215
C403220 (01)00380740003753(21)C403220
C403221 (01)00380740003753(21)C403221
C403224 (01)00380740003753(21)C403224
C403226 (01)00380740003753(21)C403226
C403227 (01)00380740003753(21)C403227
C403228 (01)00380740003753(21)C403228
C403262 (01)00380740003753(21)C403262
C403310 (01)00380740003753(21)C403310
C460089 (01)00380740003753(21)C460089
C460433 (01)00380740003753(21)C460433
01R24-56
01R25-56
Serial Numbers and UDI codes were not provided. These two products were only distributed O.U.S. (Brazil) |
Recalling Firm/ Manufacturer |
Abbott Laboratories 1915 Hurd Dr Irving TX 75038-4313
|
For Additional Information Contact |
Suchin Song 224-668-0343
|
Manufacturer Reason for Recall |
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
On 10/19/2023, the firm sent an "Urgent Medical Device Correction" Letter via FedEx priority overnight express to customers informing them that Abbott has determined that, A labeling update is required for the products listed above to meet US regulatory requirement 21 CFR 801.437 (User labeling for devices that contain natural rubber) and Brazilian requirement RDC 37/2015 (medical devices and IVDs to include in the label a Standardized Latex Warning in Portuguese CONTAIN NATURAL LATEX. MAY CAUSE ALLERGIES ). These Abbott products include internal hardware parts and/or subassemblies that contain dry natural rubber (latex) which may cause allergic reactions, but do not include the required precautionary label. Additionally, the Alinity s System Solid Waste Container Rubber Band (LN 04U97-01), used to secure the biohazard bag to the solid waste container, contains dry natural rubber (latex).
Customer are instructed to:
1) Follow good laboratory practice by wearing impervious gloves and other personal protective equipment when working on the above listed Abbott product(s).
2) If you have forwarded the product(s) listed above to other laboratories, please inform them of this Product Correction and provide a copy of this letter.
3) Please retain this letter for your laboratory records.
A precautionary label will be applied to the affected products. An Abbott representative will schedule a service visit when the instrument label is available for application.
For questions or concerns, contact Customer Service at 1-877-4ABBOTT (24 hr/7 days a week). |
Quantity in Commerce |
995 systems |
Distribution |
U.S.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV
O.U.S.: Brazil |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|