 Class 2 Device Recall VITROS 3600see related information
Date Initiated by FirmOctober 10, 2023
Date PostedDecember 13, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0543-2024
Recall Event ID 93366
510(K)NumberK083173 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductVITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0
Code Information Product Code: 6802783, UDI-DI: 10758750002979; Product Code: 6802914, UDI-DI: 10758750007103; Serial Numbers: J36000109 J36000294 J36001012 J36001489 J36000292 J36000340 J36001024 J36001495 J36000484 J36000348 J36001027 J36001513 J36000489 J36000446 J36001051 J36001515 J36000955 J36000511 J36001088 J36001519 J36001151 J36000517 J36001108 J36001528 J36001167 J36000580 J36001111 J36001535 J36001222 J36000688 J36001112 J36001570 J36001283 J36000753 J36001117 J36001573 J36001330 J36000861 J36001237 J36001575 J36001348 J36000877 J36001245 J36001579 J36000920 J36001246 J36001581 J36000923 J36001401 J36001582 J36000929 J36001403 J36001583 J36000933 J36001427 J36001600 J36000941 J36001451 J36001602 J36000942 J36001488 J36001618 J36001000
FEI Number 1000136573
Recalling Firm/
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Dr
Rochester NY 14626-5101
For Additional Information ContactJoe Falvo
585-453-3452
Manufacturer Reason
for Recall
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn URGENT PRODUCT CORRECTION NOTIFICATION letter dated 10/10/23 was sent to customers. REQUIRED ACTIONS " Review the Customer Actions for the Condition Codes described in this notification. " Acknowledge your understanding of this notification by completing the enclosed Confirmation of Receipt form no later than October 18, 2023. " Save this notification with your User Documentation or post this notification by each VITROS 3600/4600/5600/XT 3400/XT 7600 System until this issue is resolved. " If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization (Previously Ortho Care). Resolution QuidelOrtho is investigating root cause and will communicate again after root cause has been determined. A software update to address this issue is currently in development. Contact Information We apologize for any inconvenience this may cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Quantity in Commerce63 units
DistributionUS Nationwide. Global Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.