| | Class 2 Device Recall Sofsilk Braided Silk sutures |  |
| Date Initiated by Firm | December 21, 2023 |
| Date Posted | January 29, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0845-2024 |
| Recall Event ID |
93768 |
| 510(K)Number | K981128 |
| Product Classification |
Suture, nonabsorbable, silk - Product Code GAP
|
| Product | Sofsilk Braided Silk sutures:
CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36
GS-831 SOFSILK* 4-0 BLK 75CM V20 X36
S-176 SOFSILK* 0 BLK 12X45CM PCT X24
S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36
S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36
S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36
S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36
S-246 SOFSILK* 0 BLK 6X60CM PCT X36
S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA
S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36
S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36
S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36
S-316 SOFSILK* 0 BLK 12X75CM PCT X24
S-317 SOFSILK* 1 BLK 12X75CM PCT X24
S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36
S-606 SOFSILK* 0 BLK 2X150CM PCT X36
S606-12 SOFSILK* 0 BLK 2X150CM PCT X12
S-610 SOFSILK* 5 BLK 2X150CM PCT X12
SOFSILK 2-0 30 BK CV-24 GS453-2
SOFSLK 4-0 24 BK CV-15 DA VS581-2
***Added 4/1/24***
CS-792 SOFSILK* 0 BLK 100CM GCC90 X36
S-1172 SOFSILK* 6-0 BLK 45CM C1 X36
S-1735 SOFSILK* 4-0 BLK 30CM C1 DA X12
S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12
S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12
S-1789K SOFSILK* 4-0 BLK 45CM HE3 X12
S-187 SOFSILK* 1 BLK 6X45CM PCT X36
S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36
S-2780K SOFSILK* 6-0 BLK 45CM SS-24DA
S-282 SOFSILK* 5 BLK 3X60CM PCT X36
S-318 SOFSILK* 2 BLK 12X75CM PCT X24
S-608 SOFSILK* 2 BLK 2X150CM PCT X36
VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36
VS872-2 SOFSILK 3-0 30 BK CV-23 VS872-2
SS-1694G SOFSILK 4/0 BLK 45CM P24 X12
SS-5641 SOFSILK* 4-0 BLK 45CM P13 X36
SS-5649G SOFSILK* 6-0 BLK 45CM P13 X12
SS-645 SOFSILK* 4-0 BLK 75CM C12 X36
SS-675 SOFSILK* 3-0 BLK 75CM C16 X36
SS-681 SOFSILK* 6-0 BLK 45CM C12 X36
VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36
XX-5280 SOFSILK* 0 BLK 150CM PCT X36
--
Ti-Cron Coated Braided Polyester Sutures:
3087-31 TICRON* 4-0 BLU 75CM CV-331DA
3226-31 TICRON* 4-0 BLU 60CM CV330 DA
88863026-71 TICRON* 1 BLU 75CM HOS12X36
88863070-51 TICRON* 2-0 BLU 75CM C15X36
88863087-31 TICRON* 4-0 BLU 75CM CV33DA
88863092-71 TICRON* 1 BLU 75CM KV37X36
88863147-83 TICRON 2 BLU 5X75CM KV37X12
88863185-41 TICRON* 3-0 BLU 75CM CV305
88863186-31 TICRON* 4-0 BLU 75CM CV331
88863226-31 TICRON* 4-0 BLU 60CM CV330DA
88863226-41 TICRON* 3-0 BLU 60CM CV330DA
88863229-31 TICRON* 4-0 BLU 75CM CV307DA
88863271-31 TICRON* 4-0 BLU 90CM CV331DA
88863309-71 TICRON* 1 BLU 75CM GS11X36
--
Monosof Monofilament Nylon Sutures:
N-63 MONOSOF* 3-0 BLK 12X45CM PCT X36
SN-3695 MONOSOF* 5-0 BLK 45CM P10 X12
--
Steel Monofilament Stainless Steel Sutures
88862224-49 STEEL 7 4X45CM SCC X12 |
| Code Information |
Model CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36
UDI-DI 20884521083629
Lot
D1L1576RY
Model GS-831 SOFSILK* 4-0 BLK 75CM V20 X36
UDI-DI 20884521085203
Lot D2B1216RY
Model S-176 SOFSILK* 0 BLK 12X45CM PCT X24
UDI-DI 20884521085746
Lot D2D0612RY
Model S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36
UDI-DI 20884521085944
Lot D2B0728RY
Model S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36
UDI-DI 20884521085968
Lots
D2D0972RY
D2E2396RY
Model S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36
UDI-DI 20884521085975
Lots
D2D0957RY
D2D0968RY
D2D0971RY
D2D0984RY
D2D0985RY
D2E0900RY
D2E0902RY
Model S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36
UDI-DI 10884521086197
Lot D2E0674RY
UDI-DI 20884521086194
Lots
D2D1871RY
D2E0674RY
D2E0901RY
Model S-246 SOFSILK* 0 BLK 6X60CM PCT X36
UDI-DI 20884521086200
Lot D2D0980RY
Model S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA
UDI-DI 20884521086422
Lot D2F2101RY
Model S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36
UDI-DI 20884521086460
Lots
D2D1872RY
D2D2120RY
Model S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36
UDI-DI 20884521086477
Lot D2D1991RY
Model S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36
UDI-DI 20884521086484
Lots
D2E0907RY
D2E2685RY
Model S-316 SOFSILK* 0 BLK 12X75CM PCT X24
UDI-DI 20884521086491
Lot D2D2436RY
Model S-317 SOFSILK* 1 BLK 12X75CM PCT X24
UDI-DI 20884521086514
Lot D2E0587RY
Model S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36
UDI-DI 20884521086620
Lot D2E0676RY
Model S-606 SOFSILK* 0 BLK 2X150CM PCT X36
UDI-DI 10884521086630
Lot D2B2638RY
UDI-DI 20884521086637
Lots
D2B2638RY
D2D0961RY
Model S606-12 SOFSILK* 0 BLK 2X150CM PCT X12
UDI-DI 10884521637887
Lot D2D0961RY
Model S-610 SOFSILK* 5 BLK 2X150CM PCT X12
UDI-DI 20884521086675
Lot D2B1535RY
Model SOFSILK 2-0 30 BK CV-24 GS453-2
UDI-DI 20884521758121
Lots
D2D0254RY
D2D1576RY
Model SOFSLK 4-0 24 BK CV-15 DA VS581-2
UDI-DI 20884521757971
Lot D2E1431RY
***Added 4/1/24***
Model CS-792 SOFSILK* 0 BLK 100CM GCC90 X36
UDI-DI 20884521084756
Lot D2C0576RY
Model S-1172 SOFSILK* 6-0 BLK 45CM C1 X36
UDI-DI 20884521085470
Lot D2J0691RY
Model S-1735 SOFSILK* 4-0 BLK 30CM C1 DA X12
UDI-DI 20884521085630
Lot D1L3318FRY
Model S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12
UDI-DI 20884521085739
Lot D2E2348RY
Model S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12
UDI-DI 20884521085784
Lot D2B2217RY
Model S-1789K SOFSILK* 4-0 BLK 45CM HE3 X12
UDI-DI 20884521085876
Lot D2D0262RY
Model S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36
UDI-DI 20884521085975
Lots
D2B2878RY
D2E2680RY
D2E2769RY
Model S-187 SOFSILK* 1 BLK 6X45CM PCT X36
UDI-DI 20884521085982
Lot D2A1585RY
Model S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36
UDI-DI 20884521086170
Lots
D1M1020RY
D2D2348RY
D2E1492RY
Model S-246 SOFSILK* 0 BLK 6X60CM PCT X36
UDI-DI 10884521086203
Lot D2E1375RY
UDI-DI 20884521086200
Lots D2E1375RY
D2E1488RY
Model S-2780K SOFSILK* 6-0 BLK 45CM SS-24DA
UDI-DI 20884521086408
Lot D2C1916RY
Model S-282 SOFSILK* 5 BLK 3X60CM PCT X36
UDI-DI 20884521086453
Lot D2E1470RY
Model S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36
UDI-DI 20884521086460
Lots D2E1495RY
D2E1496RY
Model S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36
UDI-DI 20884521086477
Lot D2E1472RY
Model S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36
UDI-DI 20884521086484
Lots D2C1479RY
D2E2772RY
Model S-318 SOFSILK* 2 BLK 12X75CM PCT X24
UDI-DI 20884521086538
Lots D1M0153RY
D2A1677RY
Model SS-1694G SOFSILK 4/0 BLK 45CM P24 X12
UDI-DI 20884521176406
Lot D1M0141FRY
Model S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36
UDI-DI 20884521086620
Lot D2E1491RY
Model S-608 SOFSILK* 2 BLK 2X150CM PCT X36
UDI-DI 20884521086668
Lots D1M0173RY
D1M0635RY
D2E1210RY
Model VS872 SOFSILK* 3-0 BLK 75CM CV23 X36
UDI-DI 20884521090184
Lot D1M1837RY
Model VS872-2 SOFSILK 3-0 30 BK CV-23 VS872-2
UDI-DI 20884521758060
Lot D2B1783RY
Model SS-5641 SOFSILK* 4-0 BLK 45CM P13 X36
UDI-DI 20884521086774
Lot D2B2181FRY
Model SS-5649G SOFSILK* 6-0 BLK 45CM P13 X12
UDI-DI 20884521086804
Lot D2C1987FRY
Model SS-645 SOFSILK* 4-0 BLK 75CM C12 X36
UDI-DI 20884521087580
Lot D1M2531FRY
Model SS-675 SOFSILK* 3-0 BLK 75CM C16 X36
UDI-DI 20884521087726
Lot D1K1916FRY
Model SS-681 SOFSILK* 6-0 BLK 45CM C12 X36
UDI-DI 20884521087924
Lot D2G0795FRY
Model VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36
UDI-DI 20884521090184
Lot D1M1837RY
Model XX-5280 SOFSILK* 0 BLK 150CM PCT X36
UDI-DI 20884521106045
Lot D2E1482RY
--
Ti-Cron Coated Braided Polyester Sutures:
3087-31 TICRON* 4-0 BLU 75CM CV-331DA
UDI-DI 10884521632271
Lot D1M2840RY
3226-31 TICRON* 4-0 BLU 60CM CV330 DA
UDI-DI 10884521632547
Lot D2G1007RY
88863026-71 TICRON* 1 BLU 75CM HOS12X36
UDI-DI 20884521059129
Lot D2C1802RY
88863070-51 TICRON* 2-0 BLU 75CM C15X36
UDI-DI 20884521059570
Lot D2C3250FRY
88863087-31 TICRON* 4-0 BLU 75CM CV33DA
UDI-DI 10884521059832
Lot D1M2840RY
88863092-71 TICRON* 1 BLU 75CM KV37X36
UDI-DI 20884521059990
Lot D2C2450RY
88863147-83 TICRON 2 BLU 5X75CM KV37X12
UDI-DI 10884521060340
Lot D1K0110RY
88863185-41 TICRON* 3-0 BLU 75CM CV305
UDI-DI 20884521061191
Lot D2C0774RY
88863186-31 TICRON* 4-0 BLU 75CM CV331
UDI-DI 10884521061279
Lot D2G1004RY
UDI-DI 20884521061276
Lot D2G1004RY
88863226-31 TICRON* 4-0 BLU 60CM CV330DA
UDI-DI 20884521061634
Lot D2G1007RY
88863226-41 TICRON* 3-0 BLU 60CM CV330DA
UDI-DI 20884521061641
Lot D2A0451RY
88863229-31 TICRON* 4-0 BLU 75CM CV307DA
UDI-DI 20884521061689
Lot D2G1307RY
88863271-31 TICRON* 4-0 BLU 90CM CV331DA
UDI-DI 10884521062177
Lot D2F2152RY
UDI-DI 20884521062174
Lot D2F2152RY
88863309-71 TICRON* 1 BLU 75CM GS11X36
UDI-DI 20884521062488
Lot D2D2188RY
--
Monosof Monofilament Nylon Sutures:
N-63 MONOSOF* 3-0 BLK 12X45CM PCT X36
UDI-DI 20884521076607
Lot D2A0862RY
SN-3695 MONOSOF* 5-0 BLK 45CM P10 X12
UDI-DI 20884521077994
Lot D2C1333FRY
D2H1541FRY
--
Steel Monofilament Stainless Steel Sutures
88862224-49 STEEL 7 4X45CM SCC X12
UDI-DI 20884521102351
Lot D3B1366RY |
| FEI Number |
1282497
|
Recalling Firm/ Manufacturer |
Covidien LP
15 Hampshire St
Mansfield
MA
02048-1113
|
| For Additional Information Contact | Technical Service 800-962-9888 Ext. 2 |
Manufacturer Reason for Recall | Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level. |
FDA Determined Cause 2 | Process control |
| Action | On December 21, the firm notified affected consignees via URGENT MEDICAL DEVICE RECALL letters.
Customers were instructed to identify, quarantine, and return all unused, non-expired affected sutures to Medtronic. Please contact rs.covidienfeedbackcustomerservice@medtronic.com for the Return Good Authorization (RGA). Customers will receive credit for returned affected product. If purchased from a distributor, please contact your distributor directly to arrange for the return of product back to your distributor.
Pass on the recall notice to all those who need to be aware within your organization, or to any organization where the affected sutures have been transferred or distributed.
If you have questions regarding this recall, please call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1
***Updated 4/1/2024***
Firm identified additional models and lots affected by the gamma radiation issue and sent additional notification to customers on January 25, 2024. |
| Quantity in Commerce | 46,812 US; 68,976 OUS |
| Distribution | Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = GAP
|
|
|
|