| | Class 2 Device Recall Medfusion Model 4000 Syringe Pump |  |
| Date Initiated by Firm | December 19, 2023 |
| Date Posted | April 15, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1541-2024 |
| Recall Event ID |
93771 |
| 510(K)Number | K111386 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers:
a) 4000-0101-249,
b) 4000-0101-50,
c) 4000-0101-51,
d) 4000-0105-249,
e) 4000-0105-50,
f) 4000-0105-51,
g) 4000-0105-78,
h) 4000-0106-00,
i) 4000-0106-01,
j) 4000-0106-231, |
| Code Information |
All serial numbers Medfusion Model 3500 pumps manufactured or serviced with these potentially affected Barrel Clamp Guides between Jul-2016 and Apr-2021.
a) 4000-0101-249, UDI/DI 10610586040078, Serial Numbers: 2035426;
b) 4000-0101-50, UDI/DI 10610586040054, Serial Numbers: 2033894;
c) 4000-0101-51, UDI/DI 10610586040818, Serial Numbers: 2002706, 2004098, 2006133, 2006382, 2007411, 2008160, 2009040, 2009070, 2009661, 2009709, 2009723, 2009734, 2009997, 2010837, 2011049, 2011262, 2012096, 2013299, 2013805, 2020628, 2020630, 2020631, 2020972, 2023768, 2023983, 2023984, 2023985, 2024844, 2024913, 2024914, 2025040, 2026921, 2026922, 2026923, 2026924, 2026925, 2026926, 2026927, 2026928, 2026929, 2026930, 2026931, 2026934, 2026935, 2026936, 2026937, 2026938, 2026939, 2026940, 2026941, 2026942, 2026943, 2026944, 2026945, 2026946, 2026947, 2026948, 2026949, 2026950, 2026951, 2026952, 2026953, 2030198, 2031172, 2031173, 2032385, 2032622, 2034615, 2034616, 2034617, 2034618, 2034619, 2034620, 2034621, 2034622, 2034623, 2034624, 2034625, 2034716, 2034717, 2034718, 2034719, 2034720, 2034721, 2034722, 2034723, 2034724, 2034725, 2034726, 2034727, 2034728, 2034729, 2034730, 2034731, 2034732, 2034733, 2034734, 2034735, 2034736, 2034737, 2034738, 2034739, 2034740, 2034741, 2034742, 2034743, 2034744, 2034745, 2034926, 2034927, 2034928, 2034929, 2034930, 2034931, 2034932;
d) 4000-0105-249, UDI/DI 10610586043581, Serial Numbers: 2026365, 2027405,
2035029;
e) 4000-0105-50, UDI/DI none, Serial Numbers: 2022912, 2022913, 2022914;
f) 4000-0105-51, UDI/DI 10610586043567, Serial Numbers: 2019150, 2019283, 2019287, 2019294, 2019299, 2019304, 2019309, 2019369, 2019471, 2019694, 2019729, 2019761, 2019777, 2019807, 2019808, 2019883, 2020096, 2020132, 2020153, 2020154, 2020211, 2020218, 2020259, 2020328, 2020348, 2020355, 2020365, 2020423, 2020426, 2020496, 2020528, 2020529, 2020530, 2020531, 2020532, 2020533, 2020534, 2020535, 2020536, 2020537, 2020543, 2020583, 2020592, 2020626, 2021417, 2022163, 2022883, 2022885, 2022930, 2022932, 2022936, 2022937, 2022940, 2022941, 2022946, 2022947, 2022948, 2022949, 2022950, 2022952, 2022953, 2022954, 2022964, 2022967, 2022968, 2022969, 2022971, 2022972, 2022973, 2022974, 2022975, 2022977, 2022982, 2022987, 2022988, 2022990, 2023112, 2023211, 2023333, 2023355, 2023358, 2023382, 2023400, 2023454, 2023465, 2023466, 2023467, 2023468, 2023469, 2023470, 2023472, 2023473, 2023474, 2023475, 2023476, 2023477, 2023478, 2023479, 2023480, 2023481, 2023482, 2023483, 2023484, 2023485, 2023486, 2023487, 2023488, 2023489, 2023490, 2023491, 2023492, 2023493, 2023494, 2023725, 2023834, 2024241, 2024308, 2024356, 2024417, 2024544, 2024545, 2024546, 2024547, 2024548, 2024549, 2024551, 2024552, 2024553, 2024554, 2024555, 2024556, 2024557, 2024558, 2024559, 2024560, 2024561, 2024562, 2024563, 2024564, 2024565, 2024567, 2024568, 2024569, 2024570, 2024571, 2024572, 2024574, 2024575, 2024576, 2024577, 2024578, 2024579, 2024580, 2024581, 2024583, 2024664, 2024725, 2024727, 2024731, 2024812, 2024984, 2024985, 2025003, 2025053, 2025054, 2025057, 2025133, 2025144, 2025161, 2025219, 2025220, 2025244, 2025255, 2025270, 2025282, 2025304, 2025550, 2025559, 2025776, 2025799, 2025999, 2026046, 2026047, 2026048, 2026049, 2026050, 2026051, 2026052, 2026053, 2026054, 2026055, 2026056, 2026057, 2026058, 2026059, 2026060, 2026061, 2026062, 2026063, 2026064, 2026065, 2026066, 2026067, 2026068, 2026069, 2026070, 2026071, 2026072, 2026073, 2026074, 2026075, 2026076, 2026077, 2026078, 2026079, 2026080, 2026081, 2026082, 2026083, 2026084, 2026085, 2026086, 2026087, 2026088, 2026089, 2026090, 2026091, 2026092, 2026093, 2026094, 2026095, 2026096, 2026097, 2026098, 2026099, 2026100, 2026101, 2026102, 2026103, 2026104, 2026105, 2026106, 2026107, 2026108, 2026109, 2026110, 2026111, 2026112, 2026113, 2026114, 2026115, 2026116, 2026117, 2026118, 2026119, 2026120, 2026121, 2026122, 2026123, 2026124, 2026125, 2026126, 2026127, 2026128, 2026129, 2026130, 2026131, 2026132, 2026133, 2026134, 2026135, 2026136, 2026137, 2026138, 2026139, 2026140, 2026141, 2026142, 2026143, 2026145, 2026146, 2026147, 2026148, 2026149, 2026150, 2026151, 2026152, 2026153, 2026154, 2026156, 2026157, 2026158, 2026159, 2026160, 2026161, 2026162, 2026163, 2026164, 2026165, 2026166, 2026167, 2026168, 2026169, 2026170, 2026171, 2026173, 2026174, 2026175, 2026176, 2026177, 2026178, 2026179, 2026180, 2026181, 2026182, 2026183, 2026184, 2026185, 2026186, 2026187, 2026188, 2026189, 2026190, 2026191, 2026192, 2026193, 2026194, 2026195, 2026196, 2026197, 2026198, 2026199, 2026200, 2026201, 2026202, 2026203, 2026204, 2026205, 2026206, 2026207, 2026208, 2026209, 2026210, 2026211, 2026212, 2026213, 2026214, 2026215, 2026216, 2026217, 2026218, 2026219, 2026220, 2026221, 2026222, 2026223, 2026224, 2026225, 2026226, 2026227, 2026228, 2026229, 2026230, 2026231, 2026232, 2026233, 2026234, 2026235, 2026236, 2026237, 2026238, 2026239, 2026240, 2026241, 2026243, 2026244, 2026245, 2026246, 2026247, 2026248, 2026249, 2026250, 2026251, 2026252, 2026253, 2026254, 2026255, 2026256, 2026257, 2026258, 2026259, 2026260, 2026261, 2026262, 2026263, 2026265, 2026266, 2026267, 2026268, 2026269, 2026270, 2026271, 2026272, 2026273, 2026275, 2026276, 2026277, 2026278, 2026279, 2026280, 2026281, 2026282, 2026283, 2026284, 2026285, 2026286, 2026287, 2026288, 2026289, 2026290, 2026291, 2026294, 2026295, 2026296, 2026297, 2026298, 2026299, 2026300, 2026301, 2026302, 2026303, 2026304, 2026305, 2026306, 2026307, 2026308, 2026309, 2026310, 2026311, 2026312, 2026313, 2026315, 2026316, 2026317, 2026318, 2026319, 2026320, 2026321, 2026322, 2026323, 2026324, 2026325, 2026326, 2026327, 2026328, 2026329, 2026330, 2026331, 2026332, 2026333, 2026334, 2026335, 2026336, 2026337, 2026338, 2026339, 2026340, 2026341, 2026342, 2026343, 2026344, 2026345, 2026346, 2026347, 2026348, 2026349, 2026350, 2026351, 2026352, 2026353, 2026354, 2026355, 2026356, 2026357, 2026358, 2026359, 2026360, 2026361, 2026362, 2026363, 2026364, 2026366, 2026367, 2026368, 2026369, 2026370, 2026371, 2026372, 2026374, 2026375, 2026376, 2026377, 2026378, 2026379, 2026380, 2026381, 2026382, 2026383, 2026384, 2026386, 2026387, 2026388, 2026389, 2026390, 2026391, 2026392, 2026393, 2026394, 2026395, 2026396, 2026397, 2026398, 2026399, 2026400, 2026401, 2026402, 2026403, 2026405, 2026406, 2026407, 2026408, 2026409, 2026410, 2026411, 2026412, 2026413, 2026414, 2026415, 2026416, 2026417, 2026418, 2026419, 2026420, 2026422, 2026423, 2026424, 2026425, 2026426, 2026427, 2026428, 2026429, 2026430, 2026431, 2026432, 2026434, 2026435, 2026436, 2026437, 2026438, 2026439, 2026440, 2026441, 2026442, 2026443, 2026444, 2026445, 2026446, 2026447, 2026448, 2026449, 2026450, 2026451, 2026452, 2026453, 2026454, 2026455, 2026456, 2026457, 2026458, 2026459, 2026460, 2026461, 2026462, 2026463, 2026464, 2026465, 2026466, 2026467, 2026468, 2026469, 2026470, 2026471, 2026472, 2026473, 2026474, 2026475, 2026476, 2026477, 2026478, 2026479, 2026480, 2026481, 2026482, 2026483, 2026484, 2026485, 2026486, 2026487, 2026488, 2026489, 2026490, 2026491, 2026492, 2026493, 2026494, 2026495, 2026496, 2026497, 2026498, 2026499, 2026500, 2026501, 2026502, 2026503, 2026504, 2026505, 2026506, 2026507, 2026508, 2026509, 2026510, 2026511, 2026512, 2026513, 2026514, 2026515, 2026516, 2026517, 2026518, 2026519, 2026520, 2026521, 2026522, 2026523, 2026524, 2026525, 2026526, 2026527, 2026528, 2026529, 2026530, 2026531, 2026532, 2026533, 2026534, 2026535, 2026536, 2026537, 2026538, 2026539, 2026540, 2026541, 2026542, 2026543, 2026544, 2026545, 2026546, 2026547, 2026548, 2026549, 2026550, 2026551, 2026552, 2026553, 2026554, 2026555, 2026556, 2026557, 2026558, 2026559, 2026560, 2026561, 2026562, 2026563, 2026564, 2026565, 2026566, 2026567, 2026568, 2026569, 2026570, 2026571, 2026572, 2026573, 2026574, 2026575, 2026576, 2026577, 2026578, 2026579, 2026580, 2026581, 2026582, 2026583, 2026584, 2026585, 2026586, 2026587, 2026588, 2026589, 2026590, 2026591, 2026592, 2026593, 2026594, 2026595, 2026596, 2026597, 2026598, 2026599, 2026600, 2026601, 2026602, 2026603, 2026604, 2026605, 2026606, 2026607, 2026608, 2026609, 2026610, 2026611, 2026612, 2026613, 2026614, 2026615, 2026616, 2026617, 2026618, 2026619, 2026620, 2026621, 2026622, 2026623, 2026624, 2026625, 2026626, 2026627, 2026628, 2026629, 2026630, 2026631, 2026632, 2026633, 2026634, 2026635, 2026636, 2026637, 2026638, 2026639, 2026640, 2026641, 2026642, 2026643, 2026644, 2026645, 2026646, 2026647, 2026648, 2026649, 2026650, 2026651, 2026652, 2026653, 2026654, 2026655, 2026656, 2026657, 2026658, 2026659, 2026660, 2026661, 2026662, 2026663, 2026664, 2026665, 2026666, 2026667, 2026668, 2026669, 2026670, 2026671, 2026672, 2026673, 2026674, 2026675, 2026676, 2026677, 2026678, 2026679, 2026680, 2026681, 2026682, 2026683, 2026684, 2026685, 2026686, 2026687, 2026688, 2026689, 2026692, 2026693, 2026694, 2026695, 2026696, 2026697, 2026698, 2026699, 2026700, 2026701, 2026702, 2026703, 2026704, 2026705, 2026706, 2026707, 2026708, 2026709, 2026710, 2026711, 2026712, 2026713, 2026714, 2026715, 2026716, 2026717, 2026718, 2026719, 2026720, 2026721, 2026722, 2026723, 2026724, 2026725, 2026726, 2026727, 2026728, 2026729, 2026730, 2026731, 2026732, 2026733, 2026734, 2026735, 2026736, 2026737, 2026738, 2026739, 2026740, 2026741, 2026742, 2026743, 2026744, 2026745, 2026746, 2026747, 2026748, 2026749, 2026750, 2026751, 2026752, 2026753, 2026754, 2026755, 2026756, 2026757, 2026758, 2026759, 2026760, 2026761, 2026762, 2026763, 2026764, 2026765, 2026766, 2026767, 2026768, 2026769, 2026770, 2026771, 2026772, 2026773, 2026774, 2026775, 2026776, 2026777, 2026778, 2026779, 2026780, 2026781, 2026782, 2026783, 2026784, 2026785, 2026786, 2026787, 2026788, 2026789, 2026790, 2026791, 2026792, 2026793, 2026794, 2026795, 2026796, 2026797, 2026798, 2026799, 2026800, 2026801, 2026802, 2026803, 2026804, 2026805, 2026806, 2026807, 2026808, 2026809, 2026810, 2026811, 2026812, 2026813, 2026814, 2026815, 2026816, 2026817, 2026818, 2026819, 2026820, 2026821, 2026822, 2026823, 2026824, 2026825, 2026826, 2026827, 2026828, 2026829, 2026830, 2026831, 2026832, 2026833, 2026834, 2026835, 2026836, 2026837, 2026838, 2026839, 2026840, 2026841, 2026842, 2026843, 2026844, 2026845, 2026846, 2026847, 2026848, 2026849, 2026850, 2026851, 2026852, 2026853, 2026854, 2026855, 2026856, 2026857, 2026858, 2026859, 2026860, 2026861, 2026862, 2026863, 2026864, 2026865, 2026866, 2026867, 2026868, 2026869, 2026870, 2026871, 2026872, 2026873, 2026874, 2026875, 2026876, 2026877, 2026878, 2026879, 2026880, 2026881, 2026882, 2026883, 2026884, 2026885, 2026886, 2026887, 2026888, 2026889, 2026890, 2026891, 2026892, 2026893, 2026894, 2026895, 2026896, 2026897, 2026898, 2026899, 2026900, 2026901, 2026902, 2026903, 2026904, 2026905, 2026906, 2026907, 2026908, 2026909, 2026910, 2026911, 2026912, 2026913, 2026914, 2026915, 2026916, 2026917, 2026918, 2026919, 2026920, 2026954, 2026955, 2026956, 2026957, 2026958, 2026959, 2026960, 2026961, 2026962, 2026963, 2026964, 2026965, 2026966, 2026967, 2026968, 2026969, 2026970, 2026971, 2026972, 2026973, 2026974, 2026975, 2026976, 2026977, 2026978, 2026979, 2026980, 2026981, 2026982, 2026983, 2026984, 2026985, 2026986, 2026987, 2026988, 2026989, 2026990, 2026991, 2026992, 2026993, 2026994, 2026995, 2026996, 2026997, 2026998, 2026999, 2027000, 2027001, 2027002, 2027003, 2027004, 2027005, 2027006, 2027007, 2027008, 2027009, 2027010, 2027011, 2027012, 2027013, 2027014, 2027015, 2027016, 2027017, 2027018, 2027019, 2027020, 2027021, 2027022, 2027023, 2027024, 2027026, 2027027, 2027028, 2027029, 2027030, 2027031, 2027032, 2027033, 2027034, 2027035, 2027036, 2027037, 2027038, 2027039, 2027040, 2027041, 2027042, 2027043, 2027044, 2027045, 2027046, 2027047, 2027048, 2027049, 2027050, 2027051, 2027052, 2027053, 2027054, 2027055, 2027056, 2027057, 2027058, 2027059, 2027060, 2027061, 2027062, 2027063, 2027064, 2027065, 2027066, 2027067, 2027068, 2027069, 2027070, 2027071, 2027072, 2027073, 2027074, 2027075, 2027076, 2027077, 2027078, 2027079, 2027080, 2027081, 2027082, 2027083, 2027084, 2027085, 2027086, 2027087, 2027088, 2027089, 2027090, 2027091, 2027092, 2027093, 2027094, 2027095, 2027096, 2027097, 2027099, 2027100, 2027101, 2027102, 2027103, 2027104, 2027105, 2027106, 2027107, 2027108, 2027109, 2027110, 2027111, 2027112, 2027113, 2027114, 2027115, 2027116, 2027117, 2027118, 2027119, 2027120, 2027121, 2027122, 2027123, 2027124, 2027125, 2027126, 2027127, 2027128, 2027129, 2027130, 2027131, 2027132, 2027133, 2027134, 2027135, 2027136, 2027137, 2027138, 2027139, 2027140, 2027141, 2027142, 2027143, 2027144, 2027145, 2027146, 2027147, 2027148, 2027149, 2027150, 2027151, 2027152, 2027153, 2027154, 2027155, 2027156, 2027157, 2027158, 2027159, 2027160, 2027161, 2027162, 2027164, 2027165, 2027166, 2027167, 2027168, 2027169, 2027170, 2027171, 2027172, 2027173, 2027174, 2027175, 2027176, 2027177, 2027178, 2027179, 2027180, 2027181, 2027182, 2027183, 2027184, 2027185, 2027186, 2027187, 2027188, 2027189, 2027190, 2027191, 2027192, 2027193, 2027194, 2027195, 2027196, 2027197, 2027198, 2027199, 2027200, 2027201, 2027202, 2027203, 2027204, 2027205, 2027206, 2027207, 2027208, 2027209, 2027210, 2027211, 2027212, 2027213, 2027214, 2027215, 2027216, 2027217, 2027218, 2027219, 2027220, 2027221, 2027222, 2027223, 2027226, 2027227, 2027228, 2027229, 2027230, 2027231, 2027232, 2027233, 2027234, 2027235, 2027236, 2027237, 2027238, 2027239, 2027240, 2027241, 2027242, 2027243, 2027244, 2027245, 2027246, 2027247, 2027248, 2027249, 2027250, 2027251, 2027252, 2027253, 2027254, 2027255, 2027256, 2027257, 2027258, 2027259, 2027260, 2027261, 2027262, 2027263, 2027264, 2027265, 2027266, 2027267, 2027269, 2027270, 2027271, 2027272, 2027273, 2027274, 2027275, 2027276, 2027277, 2027278, 2027279, 2027280, 2027281, 2027282, 2027283, 2027284, 2027285, 2027286, 2027287, 2027288, 2027289, 2027290, 2027291, 2027292, 2027293, 2027294, 2027295, 2027296, 2027297, 2027298, 2027299, 2027300, 2027301, 2027302, 2027303, 2027304, 2027305, 2027306, 2027307, 2027308, 2027309, 2027310, 2027311, 2027312, 2027313, 2027314, 2027315, 2027316, 2027317, 2027318, 2027319, 2027320, 2027321, 2027322, 2027323, 2027324, 2027325, 2027326, 2027327, 2027328, 2027329, 2027330, 2027334, 2027335, 2027336, 2027337, 2027338, 2027339, 2027340, 2027341, 2027342, 2027343, 2027344, 2027345, 2027346, 2027347, 2027348, 2027349, 2027350, 2027351, 2027352, 2027353, 2027354, 2027355, 2027356, 2027357, 2027358, 2027359, 2027360, 2027361, 2027362, 2027363, 2027364, 2027365, 2027366, 2027367, 2027368, 2027369, 2027370, 2027371, 2027372, 2027373, 2027374, 2027375, 2027376, 2027377, 2027378, 2027379, 2027380, 2027381, 2027382, 2027383, 2027384, 2027385, 2027386, 2027387, 2027388, 2027389, 2027390, 2027391, 2027392, 2027393, 2027394, 2027395, 2027397, 2027398, 2027399, 2027400, 2027401, 2027402, 2027403, 2027404, 2027406, 2027407, 2027408, 2027409, 2027410, 2027411, 2027412, 2027413, 2027414, 2027415, 2027416, 2027417, 2027418, 2027419, 2027420, 2027421, 2027422, 2027423, 2027424, 2027425, 2027426, 2027427, 2027428, 2027429, 2027430, 2027431, 2027432, 2027433, 2027434, 2027435, 2027436, 2027437, 2027438, 2027439, 2027440, 2027441, 2027442, 2027443, 2027444, 2027445, 2027446, 2027447, 2027448, 2027449, 2027450, 2027451, 2027452, 2027453, 2027454, 2027455, 2027457, 2027458, 2027460, 2027461, 2027462, 2027463, 2027464, 2027465, 2027466, 2027467, 2027468, 2027469, 2027470, 2027471, 2027472, 2027473, 2027474, 2027475, 2027476, 2027477, 2027478, 2027479, 2027480, 2027481, 2027482, 2027483, 2027484, 2027485, 2027486, 2027487, 2027488, 2027489, 2027490, 2027491, 2027492, 2027493, 2027494, 2027495, 2027496, 2027497, 2027498, 2027499, 2027500, 2027501, 2027502, 2027503, 2027504, 2027505, 2027506, 2027507, 2027508, 2027509, 2027510, 2027511, 2027513, 2027514, 2027515, 2027516, 2027517, 2027518, 2027519, 2027520, 2027521, 2027522, 2027523, 2027524, 2027525, 2027526, 2027527.2027528, 2027529, 2027530, 2027531, 2027532, 2027533, 2027534, 2027535, 2027536, 2027537, 2027538, 2027539, 2027540, 2027541, 2027542, 2027543, 2027544, 2027545, 2027546, 2027547, 2027548, 2027549, 2027550, 2027551, 2027552, 2027553, 2027554, 2027555, 2027556, 2027557, 2027558, 2027559, 2027560, 2027561, 2027562, 2027563, 2027564, 2027565, 2027566, 2027567, 2027568, 2027569, 2027570, 2027571, 2027572, 2027573, 2027574, 2027575, 2027576, 2027577, 2027578, 2027579, 2027580, 2027581, 2027582, 2027583, 2027584, 2027585, 2027586, 2027587, 2027588, 2027589, 2027590, 2027591, 2027592, 2027593, 2027595, 2027597, 2027598, 2027599, 2027600, 2027602, 2027603, 2027604, 2027605, 2027606, 2027607, 2027608, 2027609, 2027610, 2027611, 2027612, 2027613, 2027614, 2027615, 2027616, 2027617, 2027618, 2027619, 2027620, 2027621, 2027622, 2027623, 2027625, 2027626, 2027627, 2027628, 2027629, 2027630, 2027631, 2027632, 2027633, 2027634, 2027635, 2027636, 2027637, 2027638, 2027639, 2027640, 2027641, 2027642, 2027643, 2027644, 2027645, 2027646, 2027647, 2027648, 2027649, 2027650, 2027651, 2027652, 2027653, 2027654, 2027655, 2027656, 2027657, 2027658, 2027659, 2027660, 2027661, 2027662, 2027663, 2027664, 2027665, 2027666, 2027667, 2027668, 2027669, 2027670, 2027671, 2027672, 2027673, 2027675, 2027676, 2027677, 2027678, 2027705, 2027729, 2027777, 2027899, 2027917, 2028047, 2028224, 2028296, 2028317, 2028363, 2028419, 2028473, 2028522, 2028525, 2028690, 2028701, 2028705, 2028766, 2028790, 2028820, 2028871, 2028878, 2028885, 2028918, 2028952, 2028953, 2028976, 2029022, 2029033, 2029073, 2029076, 2029091, 2029092, 2029098, 2029115, 2029129, 2029164, 2029177, 2029183, 2029192, 2029200, 2029258, 2029282, 2029283, 2029324, 2029342, 2029389, 2029435, 2029438, 2029527, 2029891, 2029915, 2029939, 2030067, 2030116, 2030119, 2030151, 2030174, 2030175, 2030176, 2030177, 2030214, 2030276, 2030292, 2030321, 2030329, 2030634, 2030635, 2030636, 2030664, 2030665, 2030669, 2030670, 2030671, 2030672, 2030678, 2030681, 2030682, 2030684, 2030714, 2030715, 2030720, 2030744, 2030783, 2030788, 2030895, 2030896, 2030925, 2030927, 2030955, 2035670, 2035671, 2035672, 2035673, 2035674, 2035675, 2035676, 2035677, 2035678, 2035679, 2035680, 2035681, 2035682, 2035683, 2035684, 2035685, 2035686, 2035687, 2035688, 2035689, 2035690, 2035691, 2036026, 2036252, 2036253;
g) 4000-0105-78, UDI/DI 10610586043574, Serial Numbers: 2026172, 2026433, 2032847, 2032881, 2033072, 2033132, 2035215, 2035443, 2035448, 2035454, 2035457, 2035458, 2035501, 2036239, 2036483, 2036494, 2036626;
h) 4000-0106-00, UDI/DI 15019517070750, Serial Numbers: 2026144, 2026691, 2027396, 2032794, 2032811, 2032870, 2032889, 2032952, 2032957, 2032958, 2032992, 2033005, 2033065, 2033073, 2033123, 2033595, 2033596, 2033597, 2033598, 2035027, 2035028, 2035030, 2035032, 2035034, 2035037, 2035038, 2035424, 2035436, 2035463, 2035464, 2035478, 2035518, 2035540, 2035636, 2035669, 2036330, 2036378, 2036402, 2036478, 2036641, 2036671, 2036940, 2037084, 2037180, 2037234, 2037361, 2037472, 2037477, 2037523, 2037710, 2037828, 2037987, 2038016, 2038038;
i) 4000-0106-01, UDI/DI 15019517069624, Serial Numbers: 2022965, 2026155, 2026242, 2026264, 2026274, 2026292, 2026293, 2026314, 2026373, 2026385, 2026404, 2026421, 2026932, 2026933, 2027025, 2027098, 2027163, 2027224, 2027225, 2027268, 2027331, 2027332, 2027333, 2027456, 2027459, 2027512, 2027594, 2027596, 2027601, 2027624, 2027674, 2027750, 2028841, 2029104, 2029133, 2030409, 2030418, 2030421, 2030435, 2030464, 2030466, 2030467, 2030492, 2030493, 2030500, 2030501, 2030509, 2030514, 2030522, 2030554, 2030560, 2030563, 2030573, 2030579, 2030580, 2030594, 2030599, 2030601, 2030606, 2030617, 2030621, 2030804, 2030805, 2030818, 2030824, 2030831, 2030846, 2030850, 2030858, 2030860, 2030865, 2030868, 2030882, 2030889, 2030991, 2030992, 2030997, 2031008, 2031010, 2031016, 2031021, 2031031, 2031040, 2031045, 2031065, 2031071, 2031095, 2031103, 2031119, 2031131, 2031134, 2031159, 2031189, 2031201, 2031213, 2031217, 2031227, 2031230, 2031235, 2031236, 2031243, 2031251, 2031264, 2031265, 2031268, 2031269, 2031277, 2031287, 2031290, 2031294, 2031302, 2031312, 2031313, 2031324, 2031325, 2031339, 2031344, 2031345, 2031350, 2031351, 2031353, 2031383, 2031384, 2031467, 2031473, 2031510, 2031512, 2031515, 2031630, 2031728, 2031749, 2031860, 2031885, 2031916, 2031956, 2032010, 2032209, 2032309, 2032314, 2032363, 2032486, 2032634, 2032635, 2032636, 2032637, 2032638, 2032639, 2032640, 2032641, 2032642, 2032643, 2032644, 2032645, 2032646, 2032647, 2032648, 2032649, 2032650, 2032651, 2032652, 2032653, 2032654, 2032655, 2032656, 2032657, 2032658, 2032659, 2032660, 2032661, 2032662, 2032663, 2032664, 2032665, 2032666, 2032667, 2032668, 2032669, 2032670, 2032671, 2032672, 2032673, 2032674, 2032675, 2032676, 2032677, 2032678, 2032679, 2032680, 2032681, 2032682, 2032683, 2032684, 2032685, 2032686, 2032687, 2032688, 2032689, 2032690, 2032691, 2032692, 2032693, 2032694, 2032695, 2032696, 2032697, 2032698, 2032699, 2032700, 2032701, 2032702, 2032703, 2032704, 2032705, 2032706, 2032707, 2032708, 2032709, 2032710, 2032711, 2032712, 2032713, 2032714, 2032715, 2032716, 2032717, 2032718, 2032719, 2032720, 2032721, 2032722, 2032723, 2032724, 2032725, 2032726, 2032727, 2032728, 2032729, 2032730, 2032731, 2032732, 2032733, 2032734, 2032735, 2032736, 2032737, 2032738, 2032739, 2032740, 2032741, 2032742, 2032743, 2032744, 2032745, 2032746, 2032747, 2032748, 2032749, 2032750, 2032751, 2032752, 2032753, 2032754, 2032755, 2032756, 2032757, 2032758, 2032759, 2032760, 2032761, 2032762, 2032763, 2032764, 2032765, 2032766, 2032767, 2032768, 2032769, 2032770, 2032771, 2032772, 2032773, 2032774, 2032775, 2032776, 2032777, 2032778, 2032779, 2032780, 2032781, 2032782, 2032783, 2032784, 2032785, 2032786, 2032787, 2032788, 2032789, 2032790, 2032791, 2032792, 2032793, 2032795, 2032796, 2032797, 2032798, 2032799, 2032800, 2032801, 2032802, 2032803, 2032804, 2032805, 2032806, 2032807, 2032808, 2032809, 2032810, 2032812, 2032813, 2032814, 2032815, 2032816, 2032817, 2032818, 2032819, 2032820, 2032821, 2032822, 2032823, 2032824, 2032825, 2032826, 2032827, 2032828, 2032829, 2032830, 2032831, 2032832, 2032833, 2032834, 2032835, 2032836, 2032837, 2032838, 2032839, 2032840, 2032841, 2032842, 2032843, 2032844, 2032845, 2032846, 2032848, 2032849, 2032850, 2032851, 2032852, 2032853, 2032854, 2032855, 2032856, 2032857, 2032858, 2032859, 2032860, 2032861, 2032862, 2032863, 2032864, 2032865, 2032866, 2032867, 2032868, 2032869, 2032871, 2032872, 2032873, 2032874, 2032875, 2032876, 2032877, 2032878, 2032879, 2032880, 2032882, 2032883, 2032884, 2032885, 2032886, 2032887, 2032888, 2032890, 2032891, 2032892, 2032893, 2032894, 2032895, 2032896, 2032897, 2032898, 2032899, 2032900, 2032901, 2032902, 2032903, 2032904, 2032905, 2032906, 2032907, 2032908, 2032909, 2032910, 2032911, 2032912, 2032913, 2032914, 2032915, 2032916, 2032917, 2032918, 2032919, 2032920, 2032921, 2032922, 2032923, 2032924, 2032925, 2032926, 2032927, 2032928, 2032929, 2032930, 2032931, 2032932, 2032933, 2032934, 2032935, 2032936, 2032937, 2032938, 2032939, 2032940, 2032941, 2032942, 2032943, 2032944, 2032945, 2032946, 2032947, 2032948, 2032949, 2032950, 2032951, 2032953, 2032954, 2032955, 2032956, 2032959, 2032960, 2032961, 2032962, 2032963, 2032964, 2032965, 2032966, 2032967, 2032968, 2032969, 2032970, 2032971, 2032972, 2032973, 2032974, 2032975, 2032976, 2032977, 2032978, 2032979, 2032980, 2032981, 2032982, 2032983, 2032984, 2032985, 2032986, 2032987, 2032988, 2032989, 2032990, 2032991, 2032993, 2032994, 2032995, 2032996, 2032997, 2032998, 2032999, 2033000, 2033001, 2033002, 2033003, 2033004, 2033006, 2033007, 2033008, 2033009, 2033010, 2033011, 2033012, 2033013, 2033014, 2033015, 2033016, 2033017, 2033018, 2033019, 2033020, 2033021, 2033022, 2033023, 2033024, 2033025, 2033026, 2033027, 2033028, 2033029, 2033030, 2033031, 2033032, 2033033, 2033034, 2033035, 2033036, 2033037, 2033038, 2033039, 2033040, 2033041, 2033042, 2033043, 2033044, 2033045, 2033046, 2033047, 2033048, 2033049, 2033050, 2033051, 2033052, 2033053, 2033054, 2033055, 2033056, 2033057, 2033058, 2033059, 2033060, 2033061, 2033062, 2033063, 2033064, 2033066, 2033067, 2033068, 2033069, 2033070, 2033071, 2033074, 2033075, 2033076, 2033077, 2033078, 2033079, 2033080, 2033081, 2033082, 2033083, 2033084, 2033085, 2033086, 2033087, 2033088, 2033089, 2033090, 2033091, 2033092, 2033093, 2033094, 2033095, 2033096, 2033097, 2033098, 2033099, 2033100, 2033101, 2033102, 2033103, 2033104, 2033105, 2033106, 2033107, 2033108, 2033109, 2033110, 2033111, 2033112, 2033113, 2033114, 2033115, 2033116, 2033117, 2033118, 2033119, 2033120, 2033121, 2033122, 2033124, 2033125, 2033126, 2033127, 2033128, 2033129, 2033130, 2033131, 2033133, 2033134, 2033135, 2033136, 2033137, 2033138, 2033139, 2033140, 2033141, 2033142, 2033143, 2033144, 2033145, 2033146, 2033147, 2033148, 2033149, 2033150, 2033151, 2033152, 2033153, 2033154, 2033155, 2033156, 2033157, 2033158, 2033159, 2033160, 2033161, 2033162, 2033163, 2033164, 2033165, 2033166, 2033167, 2033168, 2033169, 2033170, 2033171, 2033172, 2033173, 2033174, 2033175, 2033176, 2033177, 2033178, 2033179, 2033180, 2033181, 2033182, 2033183, 2033184, 2033185, 2033186, 2033187, 2033188, 2033189, 2033190, 2033191, 2033192, 2033193, 2033194, 2033195, 2033196, 2033197, 2033198, 2033199, 2033200, 2033201, 2033202, 2033203, 2033204, 2033205, 2033206, 2033207, 2033208, 2033209, 2033210, 2033211, 2033212, 2033213, 2033214, 2033215, 2033216, 2033217, 2033218, 2033219, 2033220, 2033221, 2033222, 2033223, 2033224, 2033225, 2033226, 2033227, 2033228, 2033229, 2033230, 2033231, 2033232, 2033233, 2033234, 2033235, 2033236, 2033237, 2033238, 2033239, 2033240, 2033241, 2033242, 2033243, 2033244, 2033245, 2033246, 2033247, 2033248, 2033249, 2033250, 2033251, 2033252, 2033253, 2033254, 2033255, 2033256, 2033257, 2033258, 2033259, 2033260, 2033261, 2033262, 2033263, 2033264, 2033265, 2033266, 2033267, 2033268, 2033269, 2033270, 2033271, 2033272, 2033273, 2033274, 2033275, 2033276, 2033277, 2033278, 2033279, 2033280, 2033281, 2033282, 2033283, 2033284, 2033285, 2033286, 2033287, 2033288, 2033289, 2033290, 2033291, 2033292, 2033293, 2033294, 2033295, 2033296, 2033297, 2033298, 2033299, 2033300, 2033301, 2033302, 2033303, 2033304, 2033305, 2033306, 2033307, 2033308, 2033309, 2033310, 2033311, 2033312, 2033313, 2033314, 2033315, 2033316, 2033317, 2033318, 2033319, 2033320, 2033321, 2033322, 2033323, 2033324, 2033325, 2033326, 2033327, 2033328, 2033329, 2033330, 2033331, 2033332, 2033333, 2033334, 2033335, 2033336, 2033337, 2033338, 2033339, 2033340, 2033341, 2033342, 2033343, 2033344, 2033345, 2033346, 2033347, 2033348, 2033349, 2033350, 2033351, 2033352, 2033353, 2033354, 2033355, 2033356, 2033357, 2033358, 2033359, 2033360, 2033361, 2033362, 2033363, 2033364, 2033365, 2033366, 2033367, 2033368, 2033369, 2033370, 2033371, 2033372, 2033373, 2033374, 2033375, 2033376, 2033377, 2033378, 2033379, 2033380, 2033381, 2033382, 2033383, 2033384, 2033385, 2033386, 2033387, 2033388, 2033389, 2033390, 2033391, 2033392, 2033393, 2033394, 2033395, 2033396, 2033397, 2033398, 2033399, 2033400, 2033401, 2033402, 2033403, 2033404, 2033405, 2033406, 2033407, 2033408, 2033409, 2033410, 2033411, 2033412, 2033413, 2033414, 2033415, 2033416, 2033417, 2033418, 2033419, 2033420, 2033421, 2033422, 2033423, 2033424, 2033425, 2033426, 2033427, 2033428, 2033429, 2033430, 2033431, 2033432, 2033433, 2033434, 2033435, 2033436, 2033437, 2033438, 2033439, 2033440, 2033441, 2033442, 2033443, 2033444, 2033445, 2033446, 2033447, 2033448, 2033449, 2033450, 2033451, 2033452, 2033453, 2033454, 2033455, 2033456, 2033457, 2033458, 2033459, 2033460, 2033461, 2033462, 2033463, 2033464, 2033465, 2033466, 2033467, 2033468, 2033469, 2033470, 2033471, 2033472, 2033473, 2033474, 2033475, 2033476, 2033477, 2033478, 2033479, 2033480, 2033481, 2033482, 2033483, 2033484, 2033485, 2033486, 2033487, 2033488, 2033489, 2033490, 2033491, 2033492, 2033493, 2033494, 2033495, 2033496, 2033497, 2033498, 2033499, 2033500, 2033501, 2033502, 2033503, 2033504, 2033505, 2033506, 2033507, 2033508, 2033509, 2033510, 2033511, 2033512, 2033513, 2033514, 2033515, 2033516, 2033517, 2033518, 2033519, 2033520, 2033521, 2033522, 2033523, 2033524, 2033525, 2033526, 2033527, 2033528, 2033529, 2033530, 2033531, 2033532, 2033533, 2033534, 2033535, 2033536, 2033537, 2033538, 2033539, 2033540, 2033541, 2033542, 2033543, 2033544, 2033545, 2033546, 2033547, 2033548, 2033549, 2033550, 2033551, 2033552, 2033553, 2033554, 2033555, 2033556, 2033557, 2033558, 2033559, 2033560, 2033561, 2033562, 2033563, 2033564, 2033565, 2033566, 2033567, 2033568, 2033569, 2033570, 2033571, 2033572, 2033573, 2033574, 2033575, 2033576, 2033577, 2033578, 2033579, 2033580, 2033581, 2033582, 2033583, 2033584, 2033585, 2033586, 2033587, 2033588, 2033589, 2033590, 2033591, 2033592, 2033593, 2033594, 2033599, 2033600, 2033601, 2033602, 2033603, 2033604, 2033605, 2033606, 2033607, 2033608, 2033609, 2033610, 2033611, 2033612, 2033613, 2033614, 2033615, 2033616, 2033617, 2033618, 2033619, 2033620, 2033621, 2033622, 2033623, 2033624, 2033625, 2033626, 2033627, 2033628, 2033629, 2033630, 2033631, 2033632, 2033633, 2033634, 2033635, 2033636, 2033637, 2033638, 2033639, 2033640, 2033641, 2033642, 2033643, 2033644, 2033645, 2033646, 2033647, 2033648, 2033649, 2033650, 2033651, 2033652, 2033653, 2033654, 2033655, 2033656, 2033657, 2033658, 2033659, 2033660, 2033661, 2033662, 2033663, 2033664, 2033665, 2033666, 2033667, 2033668, 2033669, 2033670, 2033671, 2033672, 2033673, 2033674, 2033675, 2033676, 2033677, 2033678, 2033679, 2033680, 2033681, 2033682, 2033683, 2033684, 2033685, 2033686, 2033687, 2033688, 2033689, 2033690, 2033691, 2033692, 2033693, 2033694, 2033695, 2033696, 2033697, 2033698, 2033699, 2033700, 2033701, 2033702, 2033703, 2033704, 2033705, 2033706, 2033707, 2033708, 2033709, 2033710, 2033711, 2033712, 2033713, 2033714, 2033715, 2033716, 2033717, 2033718, 2033719, 2033720, 2033721, 2033722, 2033723, 2033724, 2033725, 2033726, 2033727, 2033728, 2033729, 2033730, 2033731, 2033732, 2033733, 2033734, 2033735, 2033736, 2033737, 2033738, 2033739, 2033740, 2033741, 2033742, 2033743, 2033744, 2033745, 2033746, 2033747, 2033748, 2033749, 2033750, 2033751, 2033752, 2033753, 2033754, 2033755, 2033756, 2033757, 2033758, 2033759, 2033760, 2033761, 2033762, 2033763, 2033764, 2033765, 2033766, 2033767, 2033768, 2033769, 2033770, 2033771, 2033772, 2033773, 2033774, 2033775, 2033776, 2033777, 2033778, 2033779, 2033780, 2033781, 2033782, 2033783, 2033784, 2033785, 2033786, 2033787, 2033788, 2033789, 2033790, 2033791, 2033792, 2033793, 2033794, 2033795, 2033796, 2033797, 2033798, 2033799, 2033800, 2033801, 2033802, 2033803, 2033804, 2033805, 2033806, 2033807, 2033808, 2033809, 2033810, 2033811, 2033812, 2033813, 2033814, 2033815, 2033816, 2033817, 2033818, 2033819, 2033820, 2033821, 2033822, 2033823, 2033824, 2033825, 2033826, 2033827, 2033828, 2033829, 2033830, 2033831, 2033832, 2033833, 2033834, 2033835, 2033836, 2033837, 2033838, 2033839, 2033840, 2033841, 2033842, 2033843, 2033844, 2033845, 2033846, 2033847, 2033848, 2033849, 2033850, 2033851, 2033852, 2033853, 2033854, 2033855, 2033856, 2033857, 2033858, 2033859, 2033860, 2033861, 2033862, 2033863, 2033864, 2033865, 2033866, 2033867, 2033868, 2033869, 2033870, 2033871, 2033872, 2033873, 2033874, 2033875, 2033876, 2033877, 2033878, 2033879, 2033880, 2033881, 2033882, 2033883, 2033884, 2033885, 2033886, 2033887, 2033888, 2033889, 2033890, 2033891, 2033892, 2033893, 2033895, 2033896, 2033897, 2033898, 2033899, 2033900, 2033901, 2033902, 2033903, 2033904, 2033905, 2033906, 2033907, 2033908, 2033909, 2033910, 2033911, 2033912, 2033913, 2033914, 2033915, 2033916, 2033917, 2033918, 2033919, 2033920, 2033921, 2033922, 2033923, 2033924, 2033925, 2033926, 2033927, 2033928, 2033929, 2033930, 2033931, 2033932, 2033933, 2033934, 2033935, 2033936, 2033937, 2033938, 2033939, 2033940, 2033941, 2033942, 2033943, 2033944, 2033945, 2033946, 2033947, 2033948, 2033949, 2033950, 2033951, 2033952, 2033953, 2033954, 2033955, 2033956, 2033957, 2033958, 2033959, 2033960, 2033961, 2033962, 2033963, 2033964, 2033965, 2033966, 2033967, 2033968, 2033969, 2033970, 2033971, 2033972, 2033973, 2033974, 2033975, 2033976, 2033977, 2033978, 2033979, 2033980, 2033981, 2033982, 2033983, 2033984, 2033985, 2033986, 2033987, 2033988, 2033989, 2033990, 2033991, 2033992, 2033993, 2033994, 2033995, 2033996, 2033997, 2033998, 2033999, 2034000, 2034001, 2034002, 2034003, 2034004, 2034005, 2034006, 2034007, 2034008, 2034009, 2034010, 2034011, 2034012, 2034013, 2034014, 2034015, 2034016, 2034017, 2034018, 2034019, 2034020, 2034021, 2034022, 2034023, 2034024, 2034025, 2034026, 2034027, 2034028, 2034029, 2034030, 2034031, 2034032, 2034033, 2034034, 2034035, 2034036, 2034037, 2034038, 2034039, 2034040, 2034041, 2034042, 2034043, 2034044, 2034045, 2034046, 2034047, 2034048, 2034049, 2034050, 2034051, 2034052, 2034053, 2034054, 2034055, 2034056, 2034057, 2034058, 2034059, 2034060, 2034061, 2034062, 2034063, 2034064, 2034065, 2034066, 2034067, 2034068, 2034069, 2034070, 2034071, 2034072, 2034073, 2034074, 2034075, 2034076, 2034077, 2034078, 2034079, 2034080, 2034081, 2034082, 2034083, 2034084, 2034085, 2034086, 2034087, 2034088, 2034089, 2034090, 2034091, 2034092, 2034093, 2034094, 2034095, 2034096, 2034097, 2034098, 2034099, 2034100, 2034101, 2034102, 2034103, 2034104, 2034105, 2034106, 2034107, 2034108, 2034109, 2034110, 2034111, 2034112, 2034113, 2034114, 2034115, 2034116, 2034117, 2034118, 2034119, 2034120, 2034121, 2034122, 2034123, 2034124, 2034125, 2034126, 2034127, 2034128, 2034129, 2034130, 2034131, 2034132, 2034133, 2034134, 2034135, 2034136, 2034137, 2034138, 2034139, 2034140, 2034141, 2034142, 2034143, 2034144, 2034145, 2034146, 2034147, 2034148, 2034149, 2034150, 2034151, 2034152, 2034153, 2034154, 2034155, 2034156, 2034157, 2034158, 2034159, 2034160, 2034161, 2034162, 2034163, 2034164, 2034165, 2034166, 2034167, 2034168, 2034169, 2034170, 2034171, 2034172, 2034173, 2034174, 2034175, 2034176, 2034177, 2034178, 2034179, 2034180, 2034181, 2034182, 2034183, 2034184, 2034185, 2034186, 2034187, 2034188, 2034189, 2034190, 2034191, 2034192, 2034193, 2034194, 2034195, 2034196, 2034197, 2034198, 2034199, 2034200, 2034201, 2034202, 2034203, 2034204, 2034205, 2034206, 2034207, 2034208, 2034209, 2034210, 2034211, 2034212, 2034213, 2034214, 2034215, 2034216, 2034217, 2034218, 2034219, 2034220, 2034221, 2034222, 2034223, 2034224, 2034225, 2034226, 2034227, 2034228, 2034229, 2034230, 2034231, 2034232, 2034233, 2034234, 2034235, 2034236, 2034237, 2034238, 2034239, 2034240, 2034241, 2034242, 2034243, 2034244, 2034245, 2034246, 2034247, 2034248, 2034249, 2034250, 2034251, 2034252, 2034253, 2034254, 2034255, 2034256, 2034257, 2034258, 2034259, 2034260, 2034261, 2034262, 2034263, 2034264, 2034265, 2034266, 2034267, 2034268, 2034269, 2034270, 2034271, 2034272, 2034273, 2034274, 2034275, 2034276, 2034277, 2034278, 2034279, 2034280, 2034281, 2034282, 2034283, 2034284, 2034285, 2034286, 2034287, 2034288, 2034289, 2034290, 2034291, 2034292, 2034293, 2034294, 2034295, 2034296, 2034297, 2034298, 2034299, 2034300, 2034301, 2034302, 2034303, 2034304, 2034305, 2034306, 2034307, 2034308, 2034309, 2034310, 2034311, 2034312, 2034313, 2034314, 2034315, 2034316, 2034317, 2034318, 2034319, 2034320, 2034321, 2034322, 2034323, 2034324, 2034325, 2034326, 2034327, 2034328, 2034329, 2034330, 2034331, 2034332, 2034333, 2034334, 2034335, 2034336, 2034337, 2034338, 2034339, 2034340, 2034341, 2034342, 2034343, 2034344, 2034345, 2034346, 2034347, 2034348, 2034349, 2034350, 2034351, 2034352, 2034353, 2034354, 2034355, 2034356, 2034357, 2034358, 2034359, 2034360, 2034361, 2034362, 2034363, 2034364, 2034365, 2034366, 2034367, 2034368, 2034369, 2034370, 2034371, 2034372, 2034373, 2034374, 2034375, 2034376, 2034377, 2034378, 2034379, 2034380, 2034381, 2034382, 2034383, 2034384, 2034385, 2034386, 2034387, 2034388, 2034389, 2034390, 2034391, 2034392, 2034393, 2034394, 2034395, 2034396, 2034397, 2034398, 2034399, 2034400, 2034401, 2034402, 2034403, 2034404, 2034405, 2034406, 2034407, 2034408, 2034409, 2034410, 2034411, 2034412, 2034413, 2034414, 2034415, 2034416, 2034417, 2034418, 2034419, 2034420, 2034421, 2034422, 2034423, 2034424, 2034425, 2034426, 2034427, 2034428, 2034429, 2034430, 2034431, 2034432, 2034433, 2034434, 2034435, 2034436, 2034437, 2034438, 2034439, 2034440, 2034441, 2034442, 2034443, 2034444, 2034445, 2034446, 2034447, 2034448, 2034449, 2034450, 2034451, 2034452, 2034453, 2034454, 2034455, 2034456, 2034457, 2034458, 2034459, 2034460, 2034461, 2034462, 2034463, 2034464, 2034465, 2034466, 2034467, 2034468, 2034469, 2034470, 2034471, 2034472, 2034473, 2034474, 2034475, 2034476, 2034477, 2034478, 2034479, 2034480, 2034481, 2034482, 2034483, 2034484, 2034485, 2034486, 2034487, 2034488, 2034489, 2034490, 2034491, 2034492, 2034493, 2034494, 2034495, 2034496, 2034497, 2034498, 2034499, 2034500, 2034501, 2034502, 2034503, 2034504, 2034505, 2034506, 2034507, 2034508, 2034509, 2034510, 2034511, 2034512, 2034513, 2034514, 2034515, 2034516, 2034517, 2034518, 2034519, 2034520, 2034521, 2034522, 2034523, 2034524, 2034525, 2034526, 2034527, 2034528, 2034529, 2034530, 2034531, 2034532, 2034533, 2034534, 2034535, 2034536, 2034537, 2034538, 2034539, 2034540, 2034541, 2034542, 2034543, 2034544, 2034545, 2034546, 2034547, 2034548, 2034549, 2034550, 2034551, 2034552, 2034553, 2034554, 2034555, 2034556, 2034557, 2034558, 2034559, 2034560, 2034561, 2034562, 2034563, 2034564, 2034565, 2034566, 2034567, 2034568, 2034569, 2034570, 2034571, 2034572, 2034573, 2034574, 2034575, 2034576, 2034577, 2034578, 2034579, 2034580, 2034581, 2034582, 2034583, 2034584, 2034585, 2034586, 2034587, 2034588, 2034589, 2034590, 2034591, 2034592, 2034593, 2034594, 2034595, 2034596, 2034597, 2034598, 2034599, 2034600, 2034601, 2034602, 2034603, 2034604, 2034605, 2034606, 2034607, 2034608, 2034609, 2034610, 2034611, 2034612, 2034613, 2034614, 2034626, 2034627, 2034628, 2034629, 2034630, 2034631, 2034632, 2034633, 2034634, 2034635, 2034636, 2034637, 2034638, 2034639, 2034640, 2034641, 2034642, 2034643, 2034644, 2034645, 2034646, 2034647, 2034648, 2034649, 2034650, 2034651, 2034652, 2034653, 2034654, 2034655, 2034656, 2034657, 2034658, 2034659, 2034660, 2034661, 2034662, 2034663, 2034664, 2034665, 2034666, 2034667, 2034668, 2034669, 2034670, 2034671, 2034672, 2034673, 2034674, 2034675, 2034676, 2034677, 2034678, 2034679, 2034680, 2034681, 2034682, 2034683, 2034684, 2034685, 2034686, 2034687, 2034688, 2034689, 2034690, 2034691, 2034692, 2034693, 2034694, 2034695, 2034696, 2034697, 2034698, 2034699, 2034700, 2034701, 2034702, 2034703, 2034704, 2034705, 2034706, 2034707, 2034708, 2034709, 2034710, 2034711, 2034712, 2034713, 2034714, 2034715, 2034746, 2034747, 2034748, 2034749, 2034750, 2034751, 2034752, 2034753, 2034754, 2034755, 2034756, 2034757, 2034758, 2034759, 2034760, 2034761, 2034762, 2034763, 2034764, 2034765, 2034766, 2034767, 2034768, 2034769, 2034770, 2034771, 2034772, 2034773, 2034774, 2034775, 2034776, 2034777, 2034778, 2034779, 2034780, 2034781, 2034782, 2034783, 2034784, 2034785, 2034786, 2034787, 2034788, 2034789, 2034790, 2034791, 2034792, 2034793, 2034794, 2034795, 2034796, 2034797, 2034798, 2034799, 2034800, 2034801, 2034802, 2034803, 2034804, 2034805, 2034806, 2034807, 2034808, 2034809, 2034810, 2034811, 2034812, 2034813, 2034814, 2034815, 2034816, 2034817, 2034818, 2034819, 2034820, 2034821, 2034822, 2034823, 2034824, 2034825, 2034826, 2034827, 2034828, 2034829, 2034830, 2034831, 2034832, 2034833, 2034834, 2034835, 2034836, 2034837, 2034838, 2034839, 2034840, 2034841, 2034842, 2034843, 2034844, 2034845, 2034846, 2034847, 2034848, 2034849, 2034850, 2034851, 2034852, 2034853, 2034854, 2034855, 2034856, 2034857, 2034858, 2034859, 2034860, 2034861, 2034862, 2034863, 2034864, 2034865, 2034866, 2034867, 2034868, 2034869.2034870, 2034871, 2034872, 2034873, 2034874, 2034875, 2034876, 2034877, 2034878, 2034879, 2034880, 2034881, 2034882, 2034883, 2034884, 2034885, 2034886, 2034887, 2034888, 2034889, 2034890, 2034891, 2034892, 2034893, 2034894, 2034895, 2034896, 2034897, 2034898, 2034899, 2034900, 2034901, 2034902, 2034903, 2034904, 2034905, 2034906, 2034907, 2034908, 2034909, 2034910, 2034911, 2034912, 2034913, 2034914, 2034915, 2034916, 2034917, 2034918, 2034919, 2034920, 2034921, 2034922, 2034923, 2034924, 2034925, 2034933, 2034934, 2034935, 2034936, 2034937, 2034938, 2034939, 2034940, 2034941, 2034942, 2034943, 2034944, 2034945, 2034946, 2034947, 2034948, 2034949, 2034950, 2034951, 2034952, 2034953, 2034954, 2034955, 2034956, 2034957, 2034958, 2034959, 2034960, 2034961, 2034962, 2034963, 2034964, 2034965, 2034966, 2034967, 2034968, 2034969, 2034970, 2034971, 2034972, 2034973, 2034974, 2034975, 2034976, 2034977, 2034978, 2034979, 2034980, 2034981, 2034982, 2034983, 2034984, 2034985, 2034986, 2034987, 2034988, 2034989, 2034990, 2034991, 2034992, 2034993, 2034994, 2034995, 2034996, 2034997, 2034998, 2034999, 2035000, 2035001, 2035002, 2035003, 2035004, 2035005, 2035006, 2035007, 2035008, 2035009, 2035010, 2035011, 2035012, 2035013, 2035014, 2035015, 2035016, 2035017, 2035018, 2035019, 2035020, 2035021, 2035022, 2035039, 2035040, 2035041, 2035042, 2035043, 2035044, 2035045, 2035046, 2035047, 2035048, 2035049, 2035050, 2035051, 2035052, 2035053, 2035054, 2035055, 2035056, 2035057, 2035058, 2035059, 2035060, 2035061, 2035062, 2035063, 2035064, 2035065, 2035066, 2035067, 2035068, 2035069, 2035070, 2035071, 2035072, 2035073, 2035074, 2035075, 2035076, 2035077, 2035078, 2035079, 2035080, 2035081, 2035082, 2035083, 2035084, 2035085, 2035086, 2035087, 2035088, 2035089, 2035090, 2035091, 2035092, 2035093, 2035094, 2035095, 2035096, 2035097, 2035098, 2035099, 2035100, 2035101, 2035102, 2035103, 2035104, 2035105, 2035106, 2035107, 2035108, 2035109, 2035110, 2035111, 2035112, 2035113, 2035114, 2035115, 2035116, 2035117, 2035118, 2035119, 2035120, 2035121, 2035122, 2035123, 2035124, 2035125, 2035126, 2035127, 2035128, 2035129, 2035130, 2035131, 2035132, 2035133, 2035134, 2035135, 2035136, 2035137, 2035138, 2035139, 2035140, 2035141, 2035142, 2035143, 2035144, 2035145, 2035146, 2035147, 2035148, 2035149, 2035150, 2035151, 2035152, 2035153, 2035154, 2035155, 2035156, 2035157, 2035158, 2035159, 2035160, 2035161, 2035162, 2035163, 2035164, 2035165, 2035166, 2035167, 2035168, 2035169, 2035170, 2035171, 2035172, 2035173, 2035174, 2035175, 2035176, 2035177, 2035178, 2035179, 2035180, 2035181, 2035182, 2035183, 2035184, 2035185, 2035186, 2035187, 2035188, 2035189, 2035190, 2035191, 2035192, 2035193, 2035194, 2035195, 2035196, 2035197, 2035198, 2035199, 2035200, 2035201, 2035202, 2035203, 2035204, 2035205, 2035206, 2035207, 2035208, 2035209, 2035210, 2035211, 2035212, 2035213, 2035214, 2035216, 2035217, 2035218, 2035219, 2035220, 2035221, 2035222, 2035223, 2035224, 2035225, 2035226, 2035227, 2035228, 2035229, 2035230, 2035231, 2035232, 2035233, 2035234, 2035235, 2035236, 2035237, 2035238, 2035239, 2035240, 2035241, 2035242, 2035243, 2035244, 2035245, 2035246, 2035247, 2035248, 2035249, 2035250, 2035251, 2035252, 2035253, 2035254, 2035255, 2035256, 2035257, 2035258, 2035259, 2035260, 2035261, 2035262, 2035263, 2035264, 2035265, 2035266, 2035267, 2035268, 2035269, 2035270, 2035271, 2035272, 2035273, 2035274, 2035275, 2035276, 2035277, 2035278, 2035279, 2035280, 2035281, 2035282, 2035283, 2035284, 2035285, 2035286, 2035287, 2035288, 2035289, 2035290, 2035291, 2035292, 2035293, 2035294, 2035295, 2035296, 2035297, 2035298, 2035299, 2035300, 2035301, 2035302, 2035303, 2035304, 2035305, 2035306, 2035307, 2035308, 2035309, 2035310, 2035311, 2035312, 2035313, 2035314, 2035315, 2035316, 2035317, 2035318, 2035319, 2035320, 2035321, 2035322, 2035323, 2035324, 2035325, 2035326, 2035327, 2035328, 2035329, 2035330, 2035331, 2035332, 2035333, 2035334, 2035335, 2035336, 2035337, 2035338, 2035339, 2035340, 2035341, 2035342, 2035343, 2035344, 2035345, 2035346, 2035347, 2035348, 2035349, 2035350, 2035351, 2035352, 2035353, 2035354, 2035355, 2035356, 2035357, 2035358, 2035359, 2035360, 2035361, 2035362, 2035363, 2035364, 2035365, 2035366, 2035367, 2035368, 2035369, 2035370, 2035371, 2035372, 2035373, 2035374, 2035375, 2035376, 2035377, 2035378, 2035379, 2035380, 2035381, 2035382, 2035383, 2035384, 2035385, 2035386, 2035387, 2035388, 2035389, 2035390, 2035391, 2035392, 2035393, 2035394, 2035395, 2035396, 2035397, 2035398, 2035399, 2035400, 2035401, 2035402, 2035403, 2035404, 2035405, 2035406, 2035407, 2035408, 2035409, 2035410, 2035411, 2035412, 2035413, 2035414, 2035415, 2035416, 2035417, 2035418, 2035419, 2035420, 2035421, 2035422, 2035423, 2035425, 2035427, 2035428, 2035429, 2035430, 2035431, 2035432, 2035433, 2035434, 2035435, 2035437, 2035438, 2035439, 2035440, 2035441, 2035442, 2035444, 2035445, 2035446, 2035447, 2035449, 2035450, 2035451, 2035452, 2035453, 2035455, 2035456, 2035459, 2035460, 2035461, 2035462, 2035465, 2035466, 2035467, 2035468, 2035469, 2035470, 2035471, 2035472, 2035473, 2035474, 2035475, 2035476, 2035477, 2035479, 2035480, 2035481, 2035482, 2035483, 2035484, 2035485, 2035486, 2035487, 2035488, 2035489, 2035490, 2035491, 2035492, 2035493, 2035494, 2035495, 2035496, 2035497, 2035498, 2035499, 2035500, 2035502, 2035503, 2035504, 2035505, 2035506, 2035507, 2035508, 2035509, 2035510, 2035511, 2035512, 2035513, 2035514, 2035515, 2035516, 2035517, 2035519, 2035520, 2035521, 2035522, 2035523, 2035524, 2035525, 2035526, 2035527, 2035528, 2035529, 2035530, 2035531, 2035532, 2035533, 2035534, 2035535, 2035536, 2035537, 2035538, 2035539, 2035541, 2035542, 2035543, 2035544, 2035545, 2035546, 2035547, 2035548, 2035549, 2035550, 2035551, 2035552, 2035553, 2035554, 2035555, 2035556, 2035557, 2035558, 2035559, 2035560, 2035561, 2035562, 2035563, 2035564, 2035565, 2035566, 2035567, 2035568, 2035569, 2035570, 2035571, 2035572, 2035573, 2035574, 2035575, 2035576, 2035577, 2035578, 2035579, 2035580, 2035581, 2035582, 2035583, 2035584, 2035585, 2035586, 2035587, 2035588, 2035589, 2035590, 2035591, 2035592, 2035593, 2035594, 2035595, 2035596, 2035597, 2035598, 2035599, 2035600, 2035601, 2035602, 2035603, 2035604, 2035605, 2035606, 2035607, 2035608, 2035609, 2035610, 2035611, 2035612, 2035613, 2035614, 2035615, 2035616, 2035617, 2035618, 2035619, 2035620, 2035621, 2035622, 2035623, 2035624, 2035625, 2035626, 2035627, 2035628, 2035629, 2035630, 2035631, 2035632, 2035633, 2035634, 2035635, 2035637, 2035638, 2035639, 2035640, 2035641, 2035642, 2035643, 2035644, 2035645, 2035646, 2035647, 2035648, 2035649, 2035650, 2035651, 2035652, 2035653, 2035654, 2035655, 2035656, 2035657, 2035658, 2035659, 2035660, 2035661, 2035662, 2035663, 2035664, 2035665, 2035666, 2035667, 2035668, 2035692, 2035693, 2035694, 2035695, 2035696, 2035697, 2035698, 2035699, 2035700, 2035701, 2035702, 2035703, 2035704, 2035705, 2035706, 2035707, 2035708, 2035709, 2035710, 2035711, 2035712, 2035713, 2035714, 2035715, 2035716, 2035717, 2035718, 2035719, 2035720, 2035721, 2035722, 2035723, 2035724, 2035725, 2035726, 2035727, 2035728, 2035729, 2035730, 2035731, 2035732, 2035733, 2035734, 2035735, 2035736, 2035737, 2035738, 2035739, 2035740, 2035741, 2035742, 2035743, 2035744, 2035745, 2035746, 2035747, 2035748, 2035749, 2035750, 2035751, 2035752, 2035753, 2035754, 2035755, 2035756, 2035757, 2035758, 2035759, 2035760, 2035761, 2035762, 2035763, 2035764, 2035765, 2035766, 2035767, 2035768, 2035769, 2035770, 2035771, 2035772, 2035773, 2035774, 2035775, 2035776, 2035777, 2035778, 2035779, 2035780, 2035781, 2035782, 2035783, 2035784, 2035785, 2035786, 2035787, 2035788, 2035789, 2035790, 2035791, 2035792, 2035793, 2035794, 2035795, 2035796, 2035797, 2035798, 2035799, 2035800, 2035801, 2035802, 2035803, 2035804, 2035805, 2035806, 2035807, 2035808, 2035809, 2035810, 2035811, 2035812, 2035813, 2035814, 2035815, 2035816, 2035817, 2035818, 2035819, 2035820, 2035821, 2035822, 2035823, 2035824, 2035825, 2035826, 2035827, 2035828, 2035829, 2035830, 2035831, 2035832, 2035833, 2035834, 2035835, 2035836, 2035837, 2035838, 2035839, 2035840, 2035841, 2035842, 2035843, 2035844, 2035845, 2035846, 2035847, 2035848, 2035849, 2035850, 2035851, 2035852, 2035853, 2035854, 2035855, 2035856, 2035857, 2035858, 2035859, 2035860, 2035861, 2035862, 2035863, 2035864, 2035865, 2035866, 2035867, 2035868, 2035869, 2035870, 2035871, 2035872, 2035873, 2035874, 2035875, 2035876, 2035877, 2035878, 2035879, 2035880, 2035881, 2035882, 2035883, 2035884, 2035885, 2035886, 2035887, 2035888, 2035889, 2035890, 2035891, 2035892, 2035893, 2035894, 2035895, 2035896, 2035897, 2035898, 2035899, 2035900, 2035901, 2035902, 2035903, 2035904, 2035905, 2035906, 2035907, 2035908, 2035909, 2035910, 2035911, 2035912, 2035913, 2035914, 2035915, 2035916, 2035917, 2035918, 2035919, 2035920, 2035921, 2035922, 2035923, 2035924, 2035925, 2035926, 2035927, 2035928, 2035929, 2035930, 2035931, 2035932, 2035933, 2035934, 2035935, 2035936, 2035937, 2035938, 2035939, 2035940, 2035941, 2035942, 2035943, 2035944, 2035945, 2035946, 2035947, 2035948, 2035949, 2035950, 2035951, 2035952, 2035953, 2035954, 2035955, 2035956, 2035957, 2035958, 2035959, 2035960, 2035961, 2035962, 2035963, 2035964, 2035965, 2035966, 2035967, 2035968, 2035969, 2035970, 2035971, 2035972, 2035973, 2035974, 2035975, 2035976, 2035977, 2035978, 2035979, 2035980, 2035981, 2035982, 2035983, 2035984, 2035985, 2035986, 2035987, 2035988, 2035989, 2035990, 2035991, 2035992, 2035993, 2035994, 2035995, 2035996, 2035997, 2035998, 2035999, 2036000, 2036001, 2036002, 2036003, 2036004, 2036005, 2036006, 2036007, 2036008, 2036009, 2036010, 2036011, 2036012, 2036013, 2036014, 2036015, 2036016, 2036017, 2036018, 2036019, 2036020, 2036021, 2036022, 2036023, 2036024, 2036025, 2036027, 2036028, 2036029, 2036030, 2036031, 2036032, 2036033, 2036034, 2036035, 2036036, 2036037, 2036038, 2036039, 2036040, 2036041, 2036042, 2036043, 2036044, 2036045, 2036046, 2036047, 2036048, 2036049, 2036050, 2036051, 2036052, 2036053, 2036054, 2036055, 2036056, 2036057, 2036058, 2036059, 2036060, 2036061, 2036062, 2036063, 2036064, 2036065, 2036066, 2036067, 2036068, 2036069, 2036070, 2036071, 2036072, 2036073, 2036074, 2036075, 2036076, 2036077, 2036078, 2036079, 2036080, 2036081, 2036082, 2036083, 2036084, 2036085, 2036086, 2036087, 2036088, 2036089, 2036090, 2036091, 2036092, 2036093, 2036094, 2036095, 2036096, 2036097, 2036098, 2036099, 2036100, 2036101, 2036102, 2036103, 2036104, 2036105, 2036106, 2036107, 2036108, 2036109, 2036110, 2036111, 2036112, 2036113, 2036114, 2036115, 2036116, 2036117, 2036118, 2036119, 2036120, 2036121, 2036122, 2036123, 2036124, 2036125, 2036126, 2036127, 2036128, 2036129, 2036130, 2036131, 2036132, 2036133, 2036134, 2036135, 2036136, 2036137, 2036138, 2036139, 2036140, 2036141, 2036142, 2036143, 2036144, 2036145, 2036146, 2036147, 2036148, 2036149, 2036150, 2036151, 2036152, 2036153, 2036154, 2036155, 2036156, 2036157, 2036158, 2036159, 2036160, 2036161, 2036162, 2036163, 2036164, 2036165, 2036166, 2036167, 2036168, 2036169, 2036170, 2036171, 2036172, 2036173, 2036174, 2036175, 2036176, 2036177, 2036178, 2036179, 2036180, 2036181, 2036182, 2036183, 2036184, 2036185, 2036186, 2036187, 2036188, 2036189, 2036190, 2036191, 2036192, 2036193, 2036194, 2036195, 2036196, 2036197, 2036198, 2036199, 2036200, 2036201, 2036202, 2036203, 2036204, 2036205, 2036206, 2036207, 2036208, 2036209, 2036210, 2036211, 2036212, 2036213, 2036214, 2036215, 2036216, 2036217, 2036218, 2036219, 2036220, 2036221, 2036222, 2036223, 2036224, 2036225, 2036226, 2036227, 2036228, 2036229, 2036230, 2036231, 2036232, 2036233, 2036234, 2036235, 2036236, 2036237, 2036238, 2036240, 2036241, 2036242, 2036243, 2036244, 2036245, 2036246, 2036247, 2036248, 2036249, 2036250, 2036251, 2036254, 2036255, 2036256, 2036257, 2036258, 2036259, 2036260, 2036261, 2036262, 2036263, 2036264, 2036265, 2036266, 2036267, 2036268, 2036269, 2036270, 2036271, 2036272, 2036273, 2036274, 2036275, 2036276, 2036277, 2036278, 2036279, 2036280, 2036281, 2036282, 2036283, 2036284, 2036285, 2036286, 2036287, 2036288, 2036289, 2036290, 2036291, 2036292, 2036293, 2036294, 2036295, 2036296, 2036297, 2036298, 2036299, 2036300, 2036301, 2036302, 2036303, 2036304, 2036305, 2036306, 2036307, 2036308, 2036309, 2036310, 2036311, 2036312, 2036313, 2036314, 2036315, 2036316, 2036317, 2036318, 2036319, 2036320, 2036321, 2036322, 2036323, 2036324, 2036325, 2036326, 2036327, 2036328, 2036329, 2036331, 2036332, 2036333, 2036334, 2036335, 2036336, 2036337, 2036338, 2036339, 2036340, 2036341, 2036342, 2036343, 2036344, 2036345, 2036346, 2036347, 2036348, 2036349, 2036350, 2036351, 2036352, 2036353, 2036354, 2036355, 2036356, 2036357, 2036358, 2036359, 2036360, 2036361, 2036362, 2036363, 2036364, 2036365, 2036366, 2036367, 2036368, 2036369, 2036370, 2036371, 2036372, 2036373, 2036374, 2036375, 2036376, 2036377, 2036379, 2036380, 2036381, 2036382, 2036383, 2036384, 2036385, 2036386, 2036387, 2036388, 2036389, 2036390, 2036391, 2036392, 2036393, 2036394, 2036395, 2036396, 2036397, 2036398, 2036399, 2036400, 2036401, 2036403, 2036404, 2036405, 2036406, 2036407, 2036408, 2036409, 2036410, 2036411, 2036412, 2036413, 2036414, 2036415, 2036416, 2036417, 2036418, 2036419, 2036420, 2036421, 2036422, 2036423, 2036424, 2036425, 2036426, 2036427, 2036428, 2036429, 2036430, 2036431, 2036432, 2036433, 2036434, 2036435, 2036436, 2036437, 2036438, 2036439, 2036440, 2036441, 2036442, 2036443, 2036444, 2036445, 2036446, 2036447, 2036448, 2036449, 2036450, 2036451, 2036452, 2036453, 2036454, 2036455, 2036456, 2036457, 2036458, 2036459, 2036460, 2036461, 2036462, 2036463, 2036464, 2036465, 2036466, 2036467, 2036468, 2036469, 2036470, 2036471, 2036472, 2036473, 2036474, 2036475, 2036476, 2036477, 2036479, 2036480, 2036481, 2036482, 2036484, 2036485, 2036486, 2036487, 2036488, 2036489, 2036490, 2036491, 2036492, 2036493, 2036495, 2036496, 2036497, 2036498, 2036499, 2036500, 2036501, 2036502, 2036503, 2036504, 2036505, 2036506, 2036507, 2036508, 2036509, 2036510, 2036511, 2036512, 2036513, 2036514, 2036515, 2036516, 2036517, 2036518, 2036519, 2036520, 2036521, 2036522, 2036523, 2036524, 2036525, 2036526, 2036527, 2036528, 2036529, 2036530, 2036531, 2036532, 2036533, 2036534, 2036535, 2036536, 2036537, 2036538, 2036539, 2036540, 2036541, 2036542, 2036543, 2036544, 2036545, 2036546, 2036547, 2036548, 2036549, 2036550, 2036551, 2036552, 2036553, 2036554, 2036555, 2036556, 2036557, 2036558, 2036559, 2036560, 2036561, 2036562, 2036563, 2036564, 2036565, 2036566, 2036567, 2036568, 2036569, 2036570, 2036571, 2036572, 2036573, 2036574, 2036575, 2036576, 2036577, 2036578, 2036579, 2036580, 2036581, 2036582, 2036583, 2036584, 2036585, 2036586, 2036587, 2036588, 2036589, 2036590, 2036591, 2036592, 2036593, 2036594, 2036595, 2036596, 2036597, 2036598, 2036599, 2036600, 2036601, 2036602, 2036603, 2036604, 2036605, 2036606, 2036607, 2036608, 2036609, 2036610, 2036611, 2036612, 2036613, 2036614, 2036615, 2036616, 2036617, 2036618, 2036619, 2036620, 2036621, 2036622, 2036623, 2036624, 2036625, 2036627, 2036628, 2036629, 2036630, 2036631, 2036632, 2036633, 2036634, 2036635, 2036636, 2036637, 2036638, 2036639, 2036640, 2036642, 2036644, 2036645, 2036646, 2036647, 2036648, 2036649, 2036650, 2036651, 2036652, 2036653, 2036654, 2036655, 2036656, 2036657, 2036658, 2036659, 2036660, 2036661, 2036662, 2036663, 2036664, 2036665, 2036666, 2036667, 2036668, 2036669, 2036670, 2036672, 2036673, 2036674, 2036675, 2036676, 2036677, 2036678, 2036679, 2036680, 2036681, 2036682, 2036683, 2036684, 2036685, 2036686, 2036687, 2036688, 2036689, 2036690, 2036691, 2036692, 2036693, 2036694, 2036695, 2036696, 2036697, 2036698, 2036699, 2036700, 2036701, 2036702, 2036703, 2036704, 2036705, 2036706, 2036707, 2036708, 2036709, 2036710, 2036711, 2036712, 2036713, 2036714, 2036715, 2036716, 2036717, 2036718, 2036719, 2036720, 2036721, 2036722, 2036723, 2036724, 2036725, 2036726, 2036727, 2036728, 2036729, 2036730, 2036731, 2036732, 2036733, 2036734, 2036735, 2036736, 2036737, 2036738, 2036739, 2036740, 2036741, 2036742, 2036743, 2036744, 2036745, 2036746, 2036747, 2036748, 2036749, 2036750, 2036751, 2036752, 2036753, 2036754, 2036755, 2036756, 2036757, 2036758, 2036759, 2036760, 2036761, 2036762, 2036763, 2036764, 2036765, 2036766, 2036767, 2036768, 2036769, 2036770, 2036771, 2036772, 2036773, 2036774, 2036775, 2036776, 2036777, 2036778, 2036779, 2036780, 2036781, 2036782, 2036783, 2036784, 2036785, 2036786, 2036787, 2036788, 2036789, 2036790, 2036791, 2036792, 2036793, 2036794, 2036795, 2036796, 2036797, 2036798, 2036799, 2036800, 2036801, 2036802, 2036803, 2036804, 2036805, 2036806, 2036807, 2036808, 2036809, 2036810, 2036811, 2036812, 2036813, 2036814, 2036815, 2036816, 2036817, 2036818, 2036819, 2036820, 2036821, 2036822, 2036823, 2036824, 2036825, 2036826, 2036827, 2036828, 2036829, 2036830, 2036831, 2036832, 2036833, 2036834, 2036835, 2036836, 2036837, 2036838, 2036839, 2036840, 2036841, 2036842, 2036843, 2036844, 2036845, 2036846, 2036847, 2036848, 2036849, 2036850, 2036851, 2036852, 2036853, 2036854, 2036855, 2036856, 2036857, 2036858, 2036859, 2036860, 2036861, 2036862, 2036863, 2036864, 2036865, 2036866, 2036867, 2036868, 2036869, 2036870, 2036871, 2036872, 2036873, 2036874, 2036875, 2036876, 2036877, 2036878, 2036879, 2036880, 2036881, 2036882, 2036883, 2036884, 2036885, 2036886, 2036887, 2036888, 2036889, 2036890, 2036891, 2036892, 2036893, 2036894, 2036895, 2036896, 2036897, 2036898, 2036899, 2036900, 2036901, 2036902, 2036903, 2036904, 2036905, 2036906, 2036907, 2036908, 2036909, 2036910, 2036911, 2036912, 2036913, 2036914, 2036915, 2036916, 2036917, 2036918, 2036919, 2036920, 2036921, 2036922, 2036923, 2036924, 2036925, 2036926, 2036927, 2036928, 2036929, 2036930, 2036931, 2036932, 2036933, 2036934, 2036935, 2036936, 2036937, 2036938, 2036939, 2036941, 2036942, 2036943, 2036944, 2036945, 2036946, 2036947, 2036948, 2036949, 2036950, 2036951, 2036952, 2036953, 2036954, 2036955, 2036956, 2036957, 2036958, 2036959, 2036960, 2036961, 2036962, 2036963, 2036964, 2036965, 2036966, 2036967, 2036968, 2036969, 2036970, 2036971, 2036972, 2036973, 2036974, 2036975, 2036976, 2036977, 2036978, 2036979, 2036980, 2036981, 2036982, 2036983, 2036984, 2036985, 2036986, 2036987, 2036988, 2036989, 2036990, 2036991, 2036992, 2036993, 2036994, 2036995, 2036996, 2036997, 2036998, 2036999, 2037000, 2037001, 2037002, 2037003, 2037004, 2037005, 2037006, 2037007, 2037008, 2037009, 2037010, 2037011, 2037012, 2037013, 2037014, 2037015, 2037016, 2037017, 2037018, 2037019, 2037020, 2037021, 2037022, 2037023, 2037024, 2037025, 2037026, 2037027, 2037028, 2037029, 2037030, 2037031, 2037032, 2037033, 2037034, 2037035, 2037036, 2037037, 2037038, 2037039, 2037040, 2037041, 2037042, 2037043, 2037044, 2037045, 2037046, 2037047, 2037048, 2037049, 2037050, 2037051, 2037052, 2037053, 2037054, 2037055, 2037056, 2037057, 2037058, 2037059, 2037060, 2037061, 2037062, 2037063, 2037064, 2037065, 2037066, 2037067, 2037068, 2037069, 2037070, 2037071, 2037072, 2037073, 2037074, 2037075, 2037076, 2037077, 2037078, 2037079, 2037080, 2037081, 2037082, 2037083, 2037085, 2037086, 2037087, 2037088, 2037089, 2037090, 2037091, 2037092, 2037093, 2037094, 2037095, 2037096, 2037097, 2037098, 2037099, 2037100, 2037101, 2037102, 2037103, 2037104, 2037105, 2037106, 2037107, 2037108, 2037109, 2037110, 2037111, 2037112, 2037113, 2037114, 2037115, 2037116, 2037117, 2037118, 2037119, 2037120, 2037121, 2037122, 2037123, 2037124, 2037125, 2037126, 2037127, 2037128, 2037129, 2037130, 2037131, 2037132, 2037133, 2037134, 2037135, 2037136, 2037137, 2037138, 2037139, 2037140, 2037141, 2037142, 2037143, 2037144, 2037145, 2037146, 2037147, 2037148, 2037149, 2037150, 2037151, 2037152, 2037153, 2037154, 2037155, 2037156, 2037157, 2037158, 2037159, 2037160, 2037161, 2037162, 2037163, 2037164, 2037165, 2037166, 2037167, 2037168, 2037169, 2037170, 2037171, 2037172, 2037173, 2037174, 2037175, 2037176, 2037177, 2037178, 2037179, 2037181, 2037182, 2037183, 2037184, 2037185, 2037186, 2037187, 2037188, 2037189, 2037190, 2037191, 2037192, 2037193, 2037194, 2037195, 2037196, 2037197, 2037198, 2037199, 2037200, 2037201, 2037202, 2037203, 2037204, 2037205, 2037206, 2037207, 2037208, 2037209, 2037210, 2037211, 2037212, 2037213, 2037214, 2037215, 2037216, 2037217, 2037218, 2037219, 2037220, 2037221, 2037222, 2037223, 2037224, 2037225, 2037226, 2037227, 2037228, 2037229, 2037230, 2037231, 2037232, 2037233, 2037235, 2037236, 2037237, 2037238, 2037239, 2037240, 2037241, 2037242, 2037243, 2037244, 2037245, 2037246, 2037247, 2037248, 2037249, 2037250, 2037251, 2037252, 2037253, 2037254, 2037255, 2037256, 2037257, 2037258, 2037259, 2037260, 2037261, 2037262, 2037263, 2037264, 2037265, 2037266, 2037267, 2037268, 2037269, 2037270, 2037271, 2037272, 2037273, 2037274, 2037275, 2037276, 2037277, 2037278, 2037279, 2037280, 2037281, 2037282, 2037283, 2037284, 2037285, 2037286, 2037287, 2037288, 2037289, 2037290, 2037291, 2037292, 2037293, 2037294, 2037295, 2037296, 2037297, 2037298, 2037299, 2037300, 2037301, 2037302, 2037303, 2037304, 2037305, 2037306, 2037307, 2037308, 2037309, 2037310, 2037311, 2037312, 2037313, 2037314, 2037315, 2037316, 2037317, 2037318, 2037319, 2037320, 2037321, 2037322, 2037323, 2037324, 2037325, 2037326, 2037327, 2037328, 2037329, 2037330, 2037331, 2037332, 2037333, 2037334, 2037335, 2037336, 2037337, 2037338, 2037339, 2037340, 2037341, 2037342, 2037343, 2037344, 2037345, 2037346, 2037347, 2037348, 2037349, 2037350, 2037351, 2037352, 2037353, 2037354, 2037355, 2037356, 2037357, 2037358, 2037359, 2037360, 2037362, 2037363, 2037364, 2037365, 2037366, 2037367, 2037368, 2037369, 2037370, 2037371, 2037372, 2037373, 2037374, 2037375, 2037376, 2037377, 2037378, 2037379, 2037380, 2037381, 2037382, 2037383, 2037384, 2037385, 2037386, 2037387, 2037388, 2037389, 2037390, 2037391, 2037392, 2037393, 2037394, 2037395, 2037396, 2037397, 2037398, 2037399, 2037400, 2037401, 2037402, 2037403, 2037404, 2037405, 2037406, 2037407, 2037408, 2037409, 2037410, 2037411, 2037412, 2037413, 2037414, 2037415, 2037416, 2037417, 2037418, 2037419, 2037420, 2037421, 2037422, 2037423, 2037424, 2037425, 2037426, 2037427, 2037428, 2037429, 2037430, 2037431, 2037432, 2037433, 2037434, 2037435, 2037436, 2037437, 2037438, 2037439, 2037440, 2037441, 2037442, 2037443, 2037444, 2037445, 2037446, 2037447, 2037448, 2037449, 2037450, 2037451, 2037452, 2037453, 2037454, 2037455, 2037456, 2037457, 2037458, 2037459, 2037460, 2037461, 2037462, 2037463, 2037464, 2037465, 2037466, 2037467, 2037468, 2037469, 2037470, 2037471, 2037473, 2037474, 2037475, 2037476, 2037478, 2037479, 2037480, 2037481, 2037482, 2037483, 2037484, 2037485, 2037486, 2037487, 2037488, 2037489, 2037490, 2037491, 2037492, 2037493, 2037494, 2037495, 2037496, 2037497, 2037498, 2037499, 2037500, 2037501, 2037502, 2037503, 2037504, 2037505, 2037506, 2037507, 2037508, 2037509, 2037510, 2037511, 2037512, 2037513, 2037514, 2037515, 2037516, 2037517, 2037518, 2037519, 2037520, 2037521, 2037522, 2037524, 2037525, 2037526, 2037527, 2037528, 2037529, 2037530, 2037531, 2037532, 2037533, 2037534, 2037535, 2037536, 2037537, 2037538, 2037539, 2037540, 2037541, 2037542, 2037543, 2037544, 2037545, 2037546, 2037547, 2037548, 2037549, 2037550, 2037551, 2037552, 2037553, 2037554, 2037555, 2037556, 2037557, 2037558, 2037559, 2037560, 2037561, 2037562, 2037563, 2037564, 2037565, 2037566, 2037567, 2037568, 2037569, 2037570, 2037571, 2037572, 2037573, 2037574, 2037575, 2037576, 2037577, 2037578, 2037579, 2037580, 2037581, 2037582, 2037583, 2037584, 2037585, 2037586, 2037587, 2037588, 2037589, 2037590, 2037591, 2037592, 2037593, 2037594, 2037595, 2037596, 2037597, 2037598, 2037599, 2037600, 2037601, 2037602, 2037603, 2037604, 2037605, 2037606, 2037607, 2037608, 2037609, 2037610, 2037611, 2037612, 2037613, 2037614, 2037615, 2037616, 2037617, 2037618, 2037619, 2037620, 2037621, 2037622, 2037623, 2037624, 2037625, 2037626, 2037627, 2037628, 2037629, 2037630, 2037631, 2037632, 2037633, 2037634, 2037635, 2037636, 2037637, 2037638, 2037639, 2037640, 2037641, 2037642, 2037643, 2037644, 2037645, 2037646, 2037647, 2037648, 2037649, 2037650, 2037651, 2037652, 2037653, 2037654, 2037655, 2037656, 2037657, 2037658, 2037659, 2037660, 2037661, 2037662, 2037663, 2037664, 2037665, 2037666, 2037667, 2037668, 2037669, 2037670, 2037671, 2037672, 2037673, 2037674, 2037675, 2037676, 2037677, 2037678, 2037679, 2037680, 2037681, 2037682, 2037683, 2037684, 2037685, 2037686, 2037687, 2037688, 2037689, 2037690, 2037691, 2037692, 2037693, 2037694, 2037695, 2037696, 2037697, 2037698, 2037699, 2037700, 2037701, 2037702, 2037703, 2037704, 2037705, 2037706, 2037707, 2037708, 2037709, 2037711, 2037712, 2037713, 2037714, 2037715, 2037716, 2037717, 2037718, 2037719, 2037720, 2037721, 2037722, 2037723, 2037724, 2037725, 2037726, 2037727, 2037728, 2037729, 2037730, 2037731, 2037732, 2037733, 2037734, 2037735, 2037736, 2037737, 2037738, 2037739, 2037740, 2037741, 2037742, 2037743, 2037744, 2037745, 2037746, 2037747, 2037748, 2037749, 2037750, 2037751, 2037752, 2037753, 2037754, 2037755, 2037756, 2037757, 2037758, 2037759, 2037760, 2037761, 2037762, 2037763, 2037764, 2037765, 2037766, 2037767, 2037768, 2037769, 2037770, 2037771, 2037772, 2037773, 2037774, 2037775, 2037776, 2037777, 2037778, 2037779, 2037780, 2037781, 2037782, 2037783, 2037784, 2037785, 2037786, 2037787, 2037788, 2037789, 2037790, 2037791, 2037792, 2037793, 2037794, 2037795, 2037796, 2037797, 2037798, 2037799, 2037800, 2037801, 2037802, 2037803, 2037804, 2037805, 2037806, 2037807, 2037808, 2037809, 2037810, 2037811, 2037812, 2037813, 2037814, 2037815, 2037816, 2037817, 2037818, 2037819, 2037820, 2037821, 2037822, 2037823, 2037824, 2037825, 2037826, 2037827, 2037829, 2037830, 2037831, 2037832, 2037833, 2037834, 2037835, 2037836, 2037837, 2037838, 2037839, 2037840, 2037841, 2037842, 2037843, 2037844, 2037845, 2037846, 2037847, 2037848, 2037849, 2037850, 2037851, 2037852, 2037853, 2037854, 2037855, 2037856, 2037857, 2037858, 2037859, 2037860, 2037861, 2037862, 2037863, 2037864, 2037865, 2037866, 2037867, 2037868, 2037869, 2037870, 2037871, 2037872, 2037873, 2037874, 2037875, 2037876, 2037877, 2037878, 2037879, 2037880, 2037881, 2037882, 2037883, 2037884, 2037885, 2037886, 2037887, 2037888, 2037889, 2037890, 2037891, 2037892, 2037893, 2037894, 2037895, 2037896, 2037897, 2037898, 2037899, 2037900, 2037901, 2037902, 2037903, 2037904, 2037905, 2037906, 2037907, 2037908, 2037909, 2037910, 2037911, 2037912, 2037913, 2037914, 2037915, 2037916, 2037917, 2037918, 2037919, 2037920, 2037921, 2037922, 2037923, 2037924, 2037925, 2037926, 2037927, 2037928, 2037929, 2037930, 2037931, 2037932, 2037933, 2037934, 2037935, 2037936, 2037937, 2037938, 2037939, 2037940, 2037941, 2037942, 2037943, 2037944, 2037945, 2037946, 2037947, 2037948, 2037949, 2037950, 2037951, 2037952, 2037953, 2037954, 2037955, 2037956, 2037957, 2037958, 2037959, 2037960, 2037961, 2037962, 2037963, 2037964, 2037965, 2037966, 2037967, 2037968, 2037969, 2037970, 2037971, 2037972, 2037973, 2037974, 2037975, 2037976, 2037977, 2037978, 2037979, 2037980, 2037981, 2037982, 2037983, 2037984, 2037985, 2037986, 2037988, 2037989, 2037990, 2037991, 2037992, 2037993, 2037994, 2037995, 2037996, 2037997, 2037998, 2037999, 2038000, 2038001, 2038002, 2038003, 2038004, 2038005, 2038006, 2038007, 2038008, 2038009, 2038010, 2038011, 2038012, 2038013, 2038014, 2038015, 2038017, 2038018, 2038019, 2038020, 2038021, 2038022, 2038023, 2038024, 2038025, 2038026, 2038027, 2038028, 2038029, 2038030, 2038031, 2038032, 2038033, 2038034, 2038035, 2038036, 2038037, 2038039;
j) 4000-0106-231, UDI/DI 15019517070996, Serial Numbers: 2026690, 2035023, 2035024, 2035025, 2035026, 2035031, 2035033, 2035035, 2035036, 2036643
|
| FEI Number |
3012307300
|
Recalling Firm/ Manufacturer |
Smiths Medical ASD Inc.
6000 Nathan Ln N
Minneapolis
MN
55442-1690
|
| For Additional Information Contact | icumed.custhelp.com/app/market¿ action 763-383-3072 |
Manufacturer Reason for Recall | Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded. |
FDA Determined Cause 2 | Process control |
| Action | Smiths Medical issued an Updated UREGNT MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2023 via US mail and email. The notice explained the issue, potential risk, and requested the following actions be taken:
"Customer Required Actions - Medfusion Infusion Pumps:
1. Locate any affected Medfusion Syringe Pumps that may be in your possession by referring to the list of affected devices included with the Response Form. This list includes any specific pump model/serial number(s) your organization purchased that were manufactured or serviced with potentially affected Barrel Clamp Guides. Each pump has a unique serial number found on the label on the bottom of the pump.
2. You may continue to use the pumps but utilize the Syringe Verification Reference Tool originally included with the 2018 Notice (also provided as Attachment 2 of this communication) until pumps containing potentially affected Barrel Clamp Guides are repaired.
Customer Required Actions - Individually Sold Barrel Clamp Guides:
1. Locate any affected Barrel Clamp Guide (Part Number G6000716) lots in your parts inventory or within the Medfusion Syringe Pump(s) you have repaired at your facility. NOTE - If you have purchased affected individually sold Barrel Clamp Guides, the Response Form included with this notice will indicate the order numbers shipped to your organization and a Return Label will be provided.
2. If any potentially affected Barrel Clamp Guides have already been installed in pumps, you may continue to use the pumps, but utilize the Syringe Verification Reference Tool originally included with the 2018 Notice (also provided as Attachment 2 of this communication) until the pumps installed with potentially affected Barrel Clamp Guides are repaired. If you have affected devices, please contact Smiths Medical for repair of the pump."
"DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Ask them to c |
| Quantity in Commerce | 7160 pumps |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|