| | Class 2 Device Recall Exactech Equinoxe |  |
| Date Initiated by Firm | March 06, 2024 |
| Date Posted | March 29, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1405-2024 |
| Recall Event ID |
94092 |
| 510(K)Number | K042021 K093430 |
| Product Classification |
Prosthesis, shoulder, semi-constrained, metal/polymer cemented - Product Code KWS
|
| Product | Exactech Equinoxe GLENOID,KEELED, CEMENTED,47, 50, 53 head, beta curvature, Sizes:
a) Small, Item Number 314-02-12, b) Medium, Item Number 314-02-13, c) Large, Item Number 314-02-14, d) Extra Large, Item Number 314-02-15, Shoulder Arthroplasty |
| Code Information |
a) Item Number 314-02-12, UDI/DI 10885862084507, Serial Numbers: 0790902, 0790901, 0790903, 0790904, 0790907, 0790908, 0790910, 0790919, 0790899, 0790912, 0797553, 0790917, 0797597, 0797605, 0797554, 0797582, 0790914, 0797567, 0797583, 0790900, 0797537, 0797606, 0790918, 0797559, 1340014, 1356689, 1356703, 1361002, 1361004, 1388246, 0797538, 0797563, 0797603, 0797573, 0797560, 0797572, 0797578, 0790898, 0797550, 0797574, 0790893, 1340017, 1340012, 1425621, 1388234, 1388236, 1425335, 1425344, 1360995, 1535521, 1356704, 1425620, 1340013, 1361005, 1425347, 1425626, 1425636, 1425338, 1425629, 1591628, 1591621, 1425334, 1360998, 1388247, 1425619, 1425634, 0797602, 0797552, 0797540, 0797587, 0790897, 1594675, 1356696, 1535516, 1591609, 1594683, 1591634, 1535523, 1594687, 1361008, 1340015, 1591616, 1425341, 1594692, 1425627, 1594681, 1522746, 1591610, 1425349, 1425351, 1388245, 1594693, 1591627, 1340010, 1704508, 1667046, 1704498, 1704531, 1802791, 1704533, 1802807, 1893661, 1704491, 1704530, 1704506, 1704526, 1864986, 1802815, 1704488, 1783187, 1667062, 1864969, 1864971, 1704500, 1704504, 1667056, 1704484, 1704501, 1704535, 1704494, 1704502, 1704536, 1667036, 1704521, 1704519, 1704524, 1802802, 1704522, 1802816, 1802818, 1704532, 1802795, 1802825, 1802810, 1802801, 1802804, 1704513, 1704517, 1888844, 1704514, 1704485, 1704486, 1704487, 1704493, 1704497, 1783178, 1802812, 1802821, 1864974, 1535512, 1425336, 1356706, 1591613, 1594678, 1425330, 1340011, 1356698, 1356690, 1594696, 1388239, 1425622, 1425329, 2048566, 1909798, 1802823, 1909794, 1802806, 1909801, 1667052, 1955133, 1704509, 1926351, 1667041, 1864990, 1802800, 1704489, 1888845, 2073046, 1783186, 1888841, 1802796, 1955140, 1783180, 1802799, 1704529, 1909797, 1783190, 1667054, 1783189, 1783184, 1909813, 1888849, 1864993, 1667057, 1909808, 2073051, 1888843, 1893653, 2073035, 2073043, 1909803, 2048551, 2048535, 1667033, 2048568, 2118375, 2118380, 2118381, 2118387, 1425631, 1361000, 1591626, 1591614, 1425331, 1535524, 1594686, 1425633, 1388238, 1425352, 1425353, 1591606, 1425350, 1356688, 1667060, 1704525, 1864998, 1926348, 1802797, 1802792, 1704511, 2048556, 2048532, 1783181, 1704503, 1802803, 2048549, 1704528, 1667053, 1893651, 2073044, 1909809, 1955137, 1704515, 2048538, 1864964, 2048537, 2048555, 1909802, 2048561, 2073054, 1864961, 1864963, 1864981, 1888848, 1864973, 1864999, 1864989, 2048550, 2284433, 2118379, 2284420, 2284446, 2153156, 2118386, 2284461, 2153167, 2366829, 2366813, 2118382, 2118374, 2366828, 2366822, 2284445, 2284460, 2284437, 2284439, 2284450, 2284436, 2284448, 2284454, 2284466, 2284467, 2284423, 2393757, 2393737, 2393736, 2393744, 2366837, 2153161, 2284440, 2393735, 2393756, 2366823, 2393738, 2469344, 2366826, 2366839, 2366836, 2153169, 2153176, 2153162, 1388244, 1591620, 1591623, 1388235, 1594672, 1909795, 1704483, 1802805, 1783182, 1802820, 1909799, 1926350, 1893657, 1864970, 1667037, 1926344, 1864968, 1893655, 1893652, 1893644, 1864988, 1893646, 1667039, 1783185, 2073052, 2284431, 2284419, 2118378, 2284438, 2284432, 2393745, 2522168, 2284441, 2536939, 2393739, 2153160, 2284427, 2366817, 2366814, 2284449, 2554489, 2554462, 2638558, 2638560, 2638563, 2638568, 2638577, 2638585, 2638587, 2554469, 2554474, 2638572, 2638589, 2638540, 2638545, 2638555, 2638562, 2638566, 2638581, 2638541, 2638552, 2777932, 2777934, 2777935, 2777936, 2777939, 2777941, 2777944, 2777933, 2777937, 2777938, 2777940, 2777942, 2777943, 2790565, 2790568, 2807383, 2807384, 2807386, 2807387, 2807389, 2790566, 2790570, 2638559, 2638570, 2790569, 2790571, 2844275, 2844281, 2844287, 2844288, 2844290, 2844292, 2854373, 2854380, 2638544, 2638547, 2638571, 2854376, 2854377, 2854381, 2854385, 2854387, 2854389, 2854395, 2854396, 2854397, 2638567, 2638576, 2638584, 2638578, 2638551, 2638565, 2638557, 2638546, 2554475, 1594673, 1594691, 1926345, 1864996, 1955139, 1667050, 1864976, 2073050, 2048563, 1802793, 1667035, 2393732, 2153164, 2536934, 2284455, 2366820, 2469339, 2284451, 2284458, 2393753, 2284457, 2118388, 2366841, 2536935, 2469337, 2366831, 2844274, 2638573, 3565464, 2554465, 2844273, 2898565, 2907026, 2907029, 2898566, 2854378, 2978107, 2969650, 3565478, 3565479, 2854393, 3565473, 3565474, 2854379, 2854394, 2854372, 2854375, 2638542, 3655853, 2554490, 2898567, 1802811, 1704520, 1802809, 2073049, 2048553, 2536936, 2153179, 2393746, 2366838, 2284462, 2284453, 3646045, 2844277, 2978101, 2978095, 2638549, 2854383, 2854392, 2969649, 2978093, 2978100, 2978102, 2854382, 3655862, 3655858, 2638564, 3655849, 3655863, 3655841, 3655845, 3655854, 2907021, 2898560, 3655861, 3963288, 4124447, 4124446, 4141698, 4141704, 4141710, 3963282, 3963298, 4045754, 3931359, 3931368, 3963293, 3931366, 2153166, 2536921, 2469330, 2536923, 2536929, 2469348, 2522171, 2393761, 2284434, 2554464, 2807391, 2638583, 2638543, 3565492, 2790564, 2790572, 3655852, 4221518, 4221517, 4087747, 4221531, 4087755, 4141720, 4221530, 4221511, 4221516, 4221527, 4468804, 4468823, 4468806, 4468819, 3963297, 4221532, 4141705, 3931379, 4141707, 4141713, 2469345, 3565489, 2898558, 2554473, 2554468, 2907023, 3655847, 4221520, 4045733, 4087753, 4045748, 4468802, 4045745, 3963281, 3963290, 4221509, 3963303, 4045738, 4221534, 3963295, 4221523, 3963300, 4221524, 4468821, 4468811, 4141718, 4932814, 4695178, 4695177, 4989465, 5064824, 4695176, 4903105, 4932816, 3565500, 3565496, 3646039, 4087751, 3931371, 4212549, 4212569, 4468817, 4141714, 4045752, 4221515, 3963287, 4967434, 5074761, 4903092, 5074751, 4837236, 5178263, 5178267, 5178255, 5178253, 4695191, 5178282, 5074747, 4903102, 4932815, 5074750, 4903093, 5074756, 5178259, 5074748, 4903107, 3963286, 4860545, 4903111, 4903103, 5178271, 5178256, 4695180, 5178258, 5925073, 5925082, 5925083, 5925084, 5925085, 6060038, 6060040, 6060042, 6060045, 6060047, 6060056, 6060057, 6060062, 6060039, 5925069, 6060249, 6060251, 6060255, 6060259, 5963038, 5963039, 5963048, 5963052, 6060079, 6060097, 5963047, 5074742, 6060265, 6060058, 6060044, 6060096, 6060078, 6060108, 6060254, 4903108, 5963037, 6060048, 6751144, 6751149, 6751150, 6751152, 6931677, 6931679, 6931683, 6931684, 6931685, 6751139, 6773175, 6751136, 6751155, 6060088, 6060017, 6773170, 6880994, 6880996, 6881000, 6751113, 6060252, 6060086, 7023040, 7023036, 7023042, 7023050, 7023066, 7023031, 7023033, 7023061, 7023062, 7023065, 7023069, 7023054, 6883218, 6773158;
b) Item Number 314-02-13, UDI/DI 10885862084514, Serial Numbers: 0767969, 0767996, 0746108, 0767974, 0767959, 0767995, 0746113, 0767961, 0746119, 0767956, 0746120, 0746111, 0767976, 0767960, 0746122, 0767935, 0767983, 0746121, 0767993, 0767937, 0767955, 0746128, 0767977, 0767951, 0767944, 0767924, 0775207, 0774857, 0767923, 0774855, 0746125, 0768254, 0768168, 0767972, 0768202, 0767934, 0768169, 0767952, 0767967, 0768218, 0767982, 0768167, 0768160, 0768258, 0767925, 0746109, 0767985, 0768166, 0768233, 0768203, 0768207, 0775205, 0767968, 0768175, 0746117, 0767986, 0767988, 0767932, 0767931, 0768193, 0768199, 0775188, 0767957, 0775192, 0746106, 0767954, 0746118, 0767947, 0768248, 0768243, 0768171, 0775183, 0768184, 0768178, 0767979, 0768200, 0768247, 0768238, 0768179, 0768181, 0767950, 0768189, 0768226, 0768223, 0746112, 0775200, 0767933, 0767939, 0767940, 0767981, 0767989, 0768242, 0767936, 0768217, 0768161, 0768214, 0768249, 0768259, 0768370, 0768358, 0768349, 0775204, 0768241, 0768158, 0768363, 0768206, 0768346, 0775206, 0768196, 0768260, 0768234, 0768195, 0768186, 0768235, 0768192, 0768194, 0767975, 0768230, 0767941, 0768366, 0767994, 0768237, 0768159, 0768255, 0768154, 0768163, 0768227, 0767938, 0768357, 0768231, 0775194, 0768198, 0768204, 0768373, 0768352, 0768155, 0768190, 0768173, 0768213, 0768361, 0767973, 0768365, 0775186, 0768157, 0768362, 0768172, 0768224, 0768364, 0774853, 0767943, 0768351, 0768191, 0768359, 0768221, 0768209, 0768182, 0768251, 0768188, 0768345, 0746107, 0775201, 0767922, 0768369, 0767945, 0768180, 0768215, 0775184, 0768229, 0768355, 0768183, 0768354, 0768205, 0768240, 0767958, 0768228, 0746110, 1021962, 1021967, 1021891, 1021952, 1021940, 1021960, 1021903, 1021900, 1021968, 1021937, 1021907, 1021956, 1021933, 1021901, 1021892, 1021928, 1021935, 1021949, 1021908, 1021904, 1021911, 1021915, 1021921, 1021926, 1021975, 1021890, 1021958, 1021961, 0768348, 0775191, 0768225, 0768246, 0768164, 0767992, 0768156, 0767990, 0768353, 0768176, 0775202, 0775190, 0767926, 0768245, 0775193, 0768371, 0767953, 0768244, 0768211, 0767978, 0774854, 0768222, 0775189, 0768220, 0768261, 0767980, 0768174, 0768170, 1146852, 1146809, 1146827, 1146792, 1146817, 1146793, 1146851, 1146847, 1146842, 1146831, 1146825, 1146800, 1146795, 1146822, 1146834, 1146798, 1146803, 1146833, 1146801, 1146838, 1146850, 1146815, 1146808, 1146810, 1146846, 1146839, 1146836, 1146845, 1146844, 1146797, 1146829, 1146813, 1146807, 1146841, 1146796, 1146814, 1021965, 1021948, 1021922, 1021920, 1021955, 1021924, 1021959, 1021914, 1021905, 1021977, 1021944, 1021888, 1021916, 1021893, 1021912, 1021919, 1021972, 1021981, 1021950, 1021918, 1021934, 1021978, 1021963, 1021929, 1021974, 1021932, 1021970, 1021923, 1021946, 1021941, 1021945, 1021899, 1021947, 1021931, 1021976, 1021966, 1165246, 1165234, 1165233, 1165241, 1165244, 1165239, 1165237, 1165238, 1165243, 1165242, 1165235, 1246456, 1285452, 1285406, 1285443, 1246485, 1246529, 1285465, 1285393, 1246471, 1285424, 1362322, 1285421, 1285418, 1246483, 1246533, 1246487, 1246522, 1285466, 1246507, 1285431, 1246486, 1246528, 1246514, 1362333, 1246530, 1285426, 1285414, 1285434, 1246520, 1246498, 1285441, 1246465, 1285364, 1246512, 1285395, 1246474, 1362339, 1246459, 1246532, 1285461, 1285397, 1285413, 1285451, 1362320, 1362336, 1246463, 1246504, 1285417, 1362326, 1246453, 1246458, 1246460, 1246472, 1246473, 1246475, 1246476, 1246481, 1246482, 1246488, 1246493, 1246495, 1246499, 1246500, 1246513, 1246521, 1285373, 1285374, 1285462, 1246484, 1246477, 1285371, 0775208, 0768356, 0768256, 0768368, 0768252, 0768350, 0768239, 0767930, 0768177, 0767991, 0768236, 0768210, 0768212, 1146805, 1146818, 1146799, 1146821, 1146820, 1146843, 1146806, 1146819, 1146824, 1146811, 1146791, 1146826, 1146837, 1146832, 1146835, 1146848, 1146823, 1146828, 1021925, 1021969, 1021979, 1021953, 1021917, 1021894, 1021938, 1021939, 1021973, 1021980, 1021909, 1021895, 1021927, 1021913, 1021889, 1021957, 1021954, 1021902, 1165247, 1165240, 1246534, 1424852, 1424876, 1457324, 1424860, 1477973, 1285407, 1487822, 1285447, 1285450, 1285382, 1285448, 1246506, 1362347, 1487819, 1487801, 1285384, 1450604, 1408317, 1362330, 1285453, 1551782, 1424839, 1408320, 1487824, 1487803, 1551785, 1246517, 1246503, 1285396, 1362321, 1362334, 1362327, 1433369, 1433365, 1487815, 1450600, 1487808, 1477976, 1487813, 1457308, 1285389, 1496960, 1496972, 1477964, 1246469, 1487812, 1362329, 1408322, 1285402, 1424844, 1457313, 1487821, 1285457, 1246501, 1362346, 1457310, 1487804, 1424864, 1457320, 1285409, 1551806, 1285440, 1362308, 1408312, 1285437, 1285423, 1408328, 1285459, 1477969, 1433371, 1362305, 1424851, 1487807, 1450603, 1285392, 1285455, 1487825, 1246510, 1285442, 1433370, 1424850, 1285381, 1285370, 1433367, 1424846, 1246478, 1246457, 1551805, 1285363, 1487816, 1285410, 1433360, 1285445, 1424871, 1285360, 1285438, 1424840, 1362315, 1285460, 1285401, 1285365, 1408318, 1408315, 1285362, 1450601, 1433359, 1487828, 1496964, 1551814, 1285436, 1477966, 1362310, 1433364, 1362335, 1433358, 1408329, 1285458, 1424873, 1522730, 1362309, 1424837, 1496969, 1362332, 1362307, 1285446, 1424841, 1433361, 1424858, 1285376, 1424848, 1424836, 1362328, 1246524, 1362342, 1285416, 1285391, 1424866, 1246468, 1477975, 1285380, 1362306, 1408314, 1424847, 1362337, 1477968, 1551780, 1408319, 1487820, 1424845, 1457318, 1408321, 1457311, 1496982, 1496980, 1551791, 1285366, 1457306, 1496971, 1285435, 1285425, 1285375, 1496963, 1424838, 1362314, 1285411, 1285422, 1424863, 1457309, 1477972, 1496975, 1362311, 1246508, 1424865, 1362331, 1487805, 1285449, 1285432, 1285403, 1362325, 1285387, 1362316, 1522731, 1496947, 1246461, 1362313, 1285388, 1285427, 1362318, 1424867, 1285378, 1424859, 1285367, 1285372, 1285390, 1424882, 1465105, 1477958, 1487814, 1246480, 1424857, 1433357, 1477961, 1433354, 1551787, 1496984, 1424878, 1362344, 1408326, 1408313, 1246489, 1246515, 1285419, 1551793, 1246455, 1433355, 1424855, 1424862, 1408316, 1424842, 1477962, 1551788, 1487826, 0768250, 0775187, 0768372, 1146802, 1146816, 1021930, 1021936, 1021898, 1021896, 1021910, 1246505, 1285464, 1246523, 1424868, 1285398, 1551810, 1496983, 1551781, 1285467, 1424875, 1424849, 1362340, 1408327, 1457321, 1496961, 1551792, 1433368, 1551804, 1551817, 1551801, 1551789, 1551809, 1285429, 1465104, 1457317, 1450602, 1477967, 1465103, 1496970, 1246535, 1285412, 1408324, 1487809, 1487800, 1522729, 1551779, 1487810, 1408323, 1457305, 1285383, 1551786, 1496962, 1246526, 1496978, 1246509, 1487811, 1285394, 1424874, 1424869, 1424872, 1477965, 1362319, 1424856, 1285379, 1362345, 1477960, 1285433, 1285428, 1457323, 1496986, 1477971, 1496952, 1246516, 1362338, 1285369, 1465106, 1496956, 1285420, 1362303, 1551812, 1246467, 1551813, 1424870, 1551784, 1551790, 1424853, 1551799, 1465107, 1362312, 1496957, 1551796, 1433356, 1496950, 1551794, 1496967, 1424843, 1551797, 1487818, 1477963, 1362341, 1457312, 1246464, 1246490, 1285368, 1285386, 1457319, 1496946, 1496954, 1551798, 1285405, 1496981, 1551808, 1496976, 1496955, 1457307, 1457315, 1424854, 1825668, 1733271, 1803096, 1733269, 1668203, 1668230, 1668228, 1768279, 1768127, 1668233, 1733279, 1668213, 1619944, 1625186, 1625168, 1668200, 1668222, 1668219, 1768281, 1668201, 1803087, 1825677, 1733260, 1825663, 1803110, 1825670, 1803086, 1733280, 1768262, 1733284, 1896077, 1625194, 1920011, 1889473, 1668224, 1768265, 1768294, 1825672, 1619960, 1803120, 1920015, 1625201, 1768272, 1768282, 1919987, 1733276, 1619949, 1619953, 1803122, 1768124, 1619956, 1625171, 1803099, 1625199, 1719772, 1625177, 1733256, 1668207, 1889487, 1625176, 1768134, 1668225, 1733277, 1768121, 1889472, 1625169, 1668232, 1719775, 1803119, 1668208, 1625198, 1825660, 1896076, 1619943, 1625192, 1768291, 1803112, 1803114, 1768111, 1803088, 1768117, 1768118, 1668217, 1625196, 1625173, 1625195, 1668220, 1889489, 1768112, 1768123, 1896107, 1768114, 1619945, 1625179, 1825674, 1825662, 1768138, 1625200, 1619942, 1889485, 1768135, 1919992, 1768113, 1619951, 1668214, 1733288, 1733259, 1768263, 1625185, 1668239, 1803092, 1803076, 1625180, 1619947, 1668240, 1733267, 1733281, 1803123, 1619957, 1733251, 1803108, 1668243, 1825675, 1625181, 1733255, 1668241, 1825659, 1625188, 1668223, 1625167, 1668237, 1896088, 1625184, 1619958, 1619950, 1668212, 1625187, 1896109, 1619952, 1625197, 1733263, 1768139, 1825666, 1768128, 1896090, 1733278, 1625189, 1733247, 1668227, 1733253, 1668249, 1889477, 1733270, 1733258, 1920021, 1803113, 1668221, 1668216, 1668247, 1803104, 1668215, 1668236, 1668242, 1733261, 1625182, 1668202, 1733250, 1768284, 1768283, 1803089, 1803103, 1803085, 1803098, 1803109, 1768280, 1828738, 1828760, 1828732, 1828740, 1768293, 1733266, 1733282, 1919985, 1920017, 1803095, 1803115, 1768269, 1803074, 1803090, 1146794, 1146812, 1146804, 1146849, 1021897, 1021964, 1246492, 1496985, 1285439, 1477970, 1246470, 1496973, 1285430, 1424835, 1285377, 1457316, 1487802, 1496966, 1433366, 1285456, 1551819, 1551818, 1477959, 1362317, 1522732, 1487823, 1433362, 1551783, 1246462, 1551795, 1496949, 1285454, 1246519, 1285444, 1424834, 1424881, 1487827, 1246497, 1433363, 1496968, 1496977, 1285361, 1496965, 1408330, 1246466, 1246496, 1246511, 1246531, 1551811, 1424877, 1424879, 1433372, 1496948, 1496953, 1496979, 1920012, 1896091, 1989668, 1828753, 2070908, 1828751, 1625178, 1828737, 1962002, 1828731, 1803094, 1989654, 1889483, 1668204, 1803117, 2016222, 1803106, 1989666, 1896075, 1828759, 1984340, 2056942, 1919996, 1733272, 2016239, 1989655, 1828750, 1961996, 2056987, 1989651, 1896074, 1989673, 1828747, 1733248, 2016218, 1989667, 2070888, 1625202, 1668248, 1961986, 1985467, 1719771, 1957586, 1768120, 1668206, 1825664, 1920007, 1733283, 1896101, 2070903, 1896103, 2016221, 1768119, 1733285, 1961991, 1828746, 1896110, 1733264, 1896106, 1985474, 1961994, 1896081, 1828745, 1768290, 1668238, 1803101, 1828736, 1825669, 1825671, 1985472, 1889476, 1668205, 1920000, 1803093, 1619941, 1803116, 1668235, 1961988, 1896108, 1803097, 1889488, 1984344, 2056991, 2016230, 1768276, 1919998, 1985476, 1985475, 1989653, 1733287, 1668229, 1768268, 1768142, 1985470, 1985471, 1733262, 1889478, 1768277, 2070904, 1889470, 1768130, 1919989, 1668209, 1768116, 1989671, 1733268, 1803082, 1828741, 1896083, 1828734, 1768273, 1668245, 1961993, 2016223, 1989663, 1919993, 2056963, 1989660, 2070896, 1768141, 2016226, 1984343, 1984338, 1961985, 1768274, 2056972, 1889480, 1619955, 1989652, 1896095, 1920010, 1920002, 1828733, 1919991, 1961987, 1825673, 1961984, 2016234, 2016240, 1957584, 2070885, 2070905, 2056966, 2016220, 1896096, 1896093, 1896094, 1985478, 1984339, 1919994, 1896078, 1619954, 1733252, 1803100, 2056957, 1768132, 1896084, 1961978, 1896112, 1768292, 2016236, 1920023, 1768125, 1989659, 2056953, 1920013, 1989665, 1668234, 1961981, 1828742, 1962000, 2056959, 2056947, 1919997, 1625191, 2056980, 1668244, 2070880, 1920016, 1828758, 1896080, 2016224, 1768285, 2070883, 1896089, 1668211, 1803107, 2056974, 1803080, 1889486, 1989669, 1828755, 1989670, 1768131, 2056945, 1961995, 1828757, 1961990, 1896086, 1733289, 1989658, 1625174, 1668210, 1825661, 1768275, 1619959, 1619948, 1896079, 1619946, 1668218, 1920009, 1889482, 1768270, 1961998, 1803091, 1985465, 1919986, 2070887, 2056976, 1896104, 1768144, 2016238, 1896092, 2070897, 2070894, 1919984, 2056967, 1733274, 1733273, 1889481, 2070879, 1896082, 1889475, 1896098, 2056955, 1920006, 1985477, 1803081, 1919982, 2056962, 1733286, 2070902, 1961980, 2056983, 1668231, 2016232, 2016237, 2070895, 1989664, 1828743, 1919981, 1920005, 1985468, 1828744, 2056956, 1625175, 1768140, 1828756, 1919988, 2056943, 2070882, 1889474, 2016228, 1803121, 1768266, 1961989, 1896111, 1896087, 2056950, 1984347, 1889479, 1961992, 1896100, 1984341, 2178935, 2084755, 2178762, 2178764, 2178936, 2178757, 2132497, 2121704, 2132508, 2118363, 2132510, 2121710, 2132517, 2206224, 2178944, 2121716, 2132487, 2132492, 2178934, 2178750, 2084752, 2132515, 2132489, 2121698, 2132472, 2084754, 2121713, 2132501, 2118347, 2084753, 2118369, 2132491, 2118352, 2132484, 2178922, 2121719, 2132521, 2178930, 2121701, 2132520, 2084750, 2132474, 2132502, 2084740, 2118341, 2118362, 2121692, 2121711, 2118353, 2118354, 2118355, 2121697, 2121715, 2121720, 2132485, 2132499, 2121699, 2132476, 1146840, 1165236, 1246454, 1408325, 1285400, 1496958, 1362343, 1424880, 1285385, 1496974, 1424861, 1522733, 1551807, 1551815, 1551816, 1477974, 1457314, 1989661, 1768286, 1825665, 1733257, 2070892, 2056965, 2056975, 2070881, 1768122, 1920018, 2070906, 1825676, 2056970, 1825667, 2056982, 1803084, 2070893, 2056988, 2056977, 1919995, 1920020, 1733265, 1896102, 1828748, 2056968, 2070891, 2056958, 2056961, 1989649, 1957583, 1989650, 2070899, 2056984, 1828754, 1803105, 2056990, 1985473, 1985466, 1668246, 1768289, 1961979, 2056989, 2070909, 2056979, 1896105, 2070884, 1961997, 1768133, 1733254, 1989656, 1768129, 2070898, 1625170, 1920008, 1768267, 1919999, 1920004, 2056949, 2070900, 1989672, 1989662, 1919983, 2056946, 2016233, 2016235, 2056944, 2056948, 1961983, 2206213, 2224647, 2324787, 2315320, 2473897, 2178923, 2315301, 2206220, 2206210, 2276412, 2250246, 2359720, 2359710, 2250239, 2359815, 2430607, 2526613, 2224643, 2132475, 2178938, 2359705, 2178942, 2359691, 2206216, 2132478, 2084746, 2250262, 2084745, 2221742, 2250274, 2224618, 2118364, 2498435, 2118367, 2224648, 2224622, 2206225, 2250260, 2250266, 2250261, 2178767, 2132506, 2132498, 2505944, 2121709, 2121708, 2359731, 2178785, 2178755, 2178916, 2440686, 2178928, 2178782, 2132495, 2121718, 2324772, 2178780, 2250244, 2224627, 2315305, 2132516, 2473899, 2430602, 2121691, 2250276, 2178753, 2359736, 2430588, 2451390, 2430597, 2118365, 2250241, 2324773, 2250245, 2430610, 2473892, 2132519, 2132481, 2359699, 2178940, 2221739, 2315298, 2359702, 2489201, 2276410, 2324771, 2276391, 2132507, 2178937, 2224621, 2276388, 2315300, 2276433, 2276428, 2206221, 2359714, 2430587, 2118356, 2451397, 2359808, 2451402, 2276430, 2359692, 2324798, 2276405, 2178771, 2276427, 2178763, 2250272, 2359802, 2221749, 2324775, 2178777, 2276409, 2359729, 2178783, 2224645, 2132477, 2315309, 2498436, 2359733, 2178765, 2178943, 2224654, 2178924, 2276387, 2276389, 2359728, 2221743, 2359698, 2315308, 2132486, 2276431, 2178945, 2324786, 2178927, 2315311, 2276432, 2359812, 2250258, 2324782, 2206214, 2224641, 2121707, 2315307, 2359713, 2498427, 2221736, 2250256, 2276386, 2324780, 2250254, 2276384, 2315310, 2178769, 2084751, 2178779, 2221750, 2221741, 2118345, 2276419, 2132488, 2315302, 2132505, 2178774, 2359690, 2084749, 2276401, 2206211, 2118368, 2224629, 2178778, 2132503, 2224623, 2250236, 2250270, 2178939, 2121714, 2359683, 2250257, 2276424, 2178926, 2178941, 2121717, 2250248, 2276390, 2359716, 2221740, 2118342, 2132482, 2451385, 2359686, 2224614, 2206204, 2359813, 2132504, 2430592, 2132493, 2250232, 2224619, 2359704, 2178761, 2224630, 2324796, 2359735, 2315312, 2359697, 2359706, 2276395, 2430600, 2178766, 2359821, 2206218, 2250230, 2178770, 2121694, 2084741, 2121703, 2178775, 2276398, 2315303, 2132490, 2276397, 2315318, 2359797, 2359817, 2359693, 2359810, 2359844, 2221737, 2224613, 2121696, 2250263, 2276394, 2276406, 2276407, 2276411, 2276393, 2276423, 2276404, 2276426, 2315299, 2315317, 2324779, 2359688, 2359811, 2359827, 2359830, 2178768, 2359685, 2324789, 2359689, 2359814, 2359816, 2440675, 2440687, 2440689, 2440679, 2440668, 2440669, 2440688, 2451404, 2505925, 2505946, 2526619, 2132514, 2224652, 2451391, 2221747, 2221748, 2224620, 2359708, 2359719, 2359732, 2359687, 2359721, 2359725, 2250264, 2359803, 2315319, 2178917, 2132479, 2178931, 2315315, 2250268, 2324777, 2430598, 2451401, 1362304, 1487806, 1551803, 1362302, 1496959, 1828739, 2070907, 1625193, 1896097, 2056981, 1985464, 2016231, 1768136, 1803078, 1961999, 1768143, 2016217, 1957585, 1984345, 2070890, 1719773, 2056986, 2056952, 2118349, 2498432, 2526593, 2440676, 2315306, 2440673, 2587540, 2587554, 2498434, 2430593, 2587548, 2224625, 2359846, 2359712, 2430604, 2178921, 2084747, 2526618, 2132483, 2121712, 2118340, 2178754, 2250273, 2250255, 2276396, 2132500, 2526587, 2084748, 2526591, 2587466, 2505933, 2498433, 2505932, 2324776, 2250231, 2451400, 2587552, 2178760, 2324784, 2324781, 2526616, 2224650, 2359727, 2359726, 2250265, 2178920, 2132512, 2359737, 2359715, 2489194, 2224616, 2206205, 2526604, 2178925, 2224617, 2505948, 2473891, 2359800, 2498431, 2276421, 2451394, 2359730, 2276385, 2473898, 2526594, 2121695, 2178758, 2430609, 2489196, 2587465, 2498426, 2324792, 2587533, 2587480, 2359823, 2587577, 2498437, 2359834, 2359804, 2489202, 2587551, 2276425, 2587546, 2587565, 2221738, 2178759, 2276399, 2359838, 2587481, 2451405, 2505930, 2359707, 2359825, 2587542, 2473895, 2526601, 2359805, 2359818, 2315316, 2587476, 2587494, 2250243, 2359723, 2451407, 2587583, 2505943, 2118350, 2206212, 2498423, 2276400, 2250249, 2221744, 2526623, 2587477, 2587483, 2084744, 2206226, 2526597, 2526611, 2359807, 2224655, 2587534, 2250271, 2359832, 2206227, 2505959, 2526588, 2473900, 2430590, 2526602, 2451406, 2587475, 2440674, 2178756, 2587490, 2526589, 2359701, 2473901, 2359819, 2473894, 2315322, 2451388, 2224626, 2473902, 2587463, 2359839, 2526605, 2359722, 2324790, 2206215, 2505939, 2587532, 2489208, 2526598, 2178784, 2324778, 2276415, 2359822, 2587496, 2587489, 2587530, 2359847, 2430589, 2315297, 2587488, 2430601, 2132509, 2473896, 2526599, 2250267, 2324794, 2451403, 2498425, 2587572, 2587512, 2587524, 2587528, 2587537, 2587543, 2587544, 2587545, 2359836, 2505957, 2587567, 2505954, 2526614, 2587575, 2587566, 2505936, 2587561, 2132518, 2587518, 2505941, 2587576, 2587510, 2587513, 2587521, 2587574, 2526592, 2505956, 2526615, 2430605, 2451384, 2359841, 2587547, 2324793, 2440690, 2250229, 2587535, 2118357, 2276422, 2250250, 2178933, 2587553, 2587549, 2587522, 2178751, 2359806, 2224646, 2526590, 2587526, 2121705, 2505960, 2746735, 2746725, 2690668, 2626746, 2795450, 2746732, 2626744, 2746752, 2690662, 2651484, 2651489, 2690677, 2844310, 2626729, 2736282, 2626747, 2626751, 2795448, 2690694, 2690682, 2690693, 2854976, 2690667, 2736290, 2626756, 2626719, 2651494, 2690687, 2651500, 2626750, 2651499, 2626726, 2690663, 2626722, 2626732, 2626757, 2795455, 2626745, 2854985, 2690673, 2626753, 2690685, 2795438, 2746738, 2626723, 2651503, 2626717, 2626733, 2626738, 2626761, 2651492, 2651493, 2651498, 2746734, 2746736, 2746742, 2626762, 2651495, 2746714, 2746754, 2626764, 2795437, 2795442, 2795445, 2795451, 2795452, 2795453, 2626724, 2626766, 2795440, 2795443, 2795444, 2626742, 2853412, 2844296, 2844297, 2844298, 2844299, 2844311, 2844312, 2853403, 2853416, 2853420, 2853421, 2853424, 2853425, 2853406, 2853411, 2853413, 2853426, 2844300, 2844302, 2853408, 2853414, 2854981, 2844294, 2844295, 2844301, 2844307, 2853404, 2853405, 2853409, 2853410, 2853422, 2853423, 2853427, 2854977, 2746710, 2746715, 2690690, 2651490, 2626739, 2746737, 2746753, 2746723, 2746739, 2746721, 2746727, 2746747, 2626749, 2651491, 2626752, 2690659, 2795458, 1457322, 1803079, 1668226, 2016216, 1803083, 1803102, 1920014, 1920003, 1896113, 1803118, 2016227, 1768126, 1920001, 2526610, 2315314, 2359694, 2505947, 2440672, 2359700, 2178918, 2440671, 2526606, 2206209, 2359820, 2440681, 2587484, 2276402, 2505926, 2359799, 2451389, 2451396, 2587467, 2224615, 2250247, 2587516, 2587559, 2505927, 2587493, 2359842, 2526600, 2084742, 2498438, 2440670, 2451398, 2250238, 2587536, 2430595, 2587538, 2473893, 2505931, 2440685, 2526585, 2587473, 2315304, 2118359, 2359696, 2451392, 2178919, 2587464, 2526608, 2132496, 2587491, 2359845, 2587495, 2250237, 2587573, 2587557, 2359703, 2489199, 2324785, 2526620, 2489198, 2526596, 2587555, 2587531, 2221746, 2505955, 2489205, 2489206, 2473890, 2178932, 2587520, 2587584, 2587511, 2224636, 2489197, 2084738, 2587560, 2587578, 2440677, 2250275, 2505949, 2746751, 2980584, 2626763, 2746741, 2690689, 2746709, 2927875, 2889194, 2844314, 2746743, 2927852, 2690688, 2746730, 2854980, 2946779, 2736283, 3617039, 2690678, 2746712, 2946775, 2690661, 2889191, 2690681, 2854962, 2889183, 2626758, 2651487, 2795439, 2626728, 2927876, 2889174, 2626765, 2690660, 2690665, 2927870, 3563558, 2854966, 2626736, 2980566, 2626743, 2795447, 3591925, 2889178, 3583203, 2927861, 2651485, 3583210, 2626718, 2736281, 2690674, 2651482, 2736286, 2651496, 2906927, 2906929, 2927874, 2690670, 2736284, 2626759, 2746750, 2844306, 2906920, 2746726, 2746711, 2946781, 2946758, 2690664, 2626754, 2906917, 2626727, 2854968, 2927862, 3563574, 2980574, 2854975, 2854969, 2854978, 2946777, 2853417, 2980563, 2853407, 2853418, 2853419, 2854982, 3583202, 2853415, 2854984, 2690684, 2746744, 2844313, 2889171, 2889190, 2906926, 2746731, 2980581, 2946759, 3617033, 3656788, 2906939, 3601474, 3563563, 2906918, 2889185, 2889173, 2889187, 2980571, 3601482, 3591916, 3591911, 3641057, 3641065, 3641066, 3617047, 3641067, 3641073, 3617048, 2946760, 2690692, 2736285, 2690671, 2651501, 2854973, 3563562, 2795457, 2889176, 2626730, 2651488, 2746722, 3591931, 3617044, 2651486, 3583217, 3617031, 2889188, 2056954, 2056978, 1989657, 1768115, 1803111, 2056969, 1889484, 2016219, 1768288, 1803077, 2070889, 2056964, 1984342, 2587579, 2505945, 2587514, 2505935, 2250234, 2359801, 2587563, 2224624, 2121693, 2430591, 2224653, 2206222, 2118346, 2505953, 2430586, 2451383, 2440684, 2587564, 2451399, 2440666, 2276416, 2498429, 2587562, 2587581, 2440678, 2587479, 2132511, 2206207, 2587523, 2587525, 2250242, 2118351, 2587487, 2587582, 2224657, 2084739, 2498428, 2206206, 3617050, 3617056, 3641069, 2626735, 2980575, 3656802, 2927866, 2946782, 3617029, 3641063, 3601475, 3563560, 2927853, 2946764, 2844309, 3641064, 2889172, 3641058, 2946771, 2746740, 2889177, 2690672, 2946778, 2906923, 2854964, 2906931, 3563565, 2690669, 3591909, 2980573, 3641068, 3591913, 2746756, 2927873, 3617026, 2854963, 3563554, 2906933, 3641049, 2889192, 2854974, 3563573, 2690675, 2746745, 3601483, 2746720, 2889193, 2927872, 2626721, 3656780, 3563571, 3563553, 3583199, 3583201, 2946766, 2980579, 2946761, 2946762, 2946763, 2946765, 2946767, 2946768, 2946772, 2980583, 2980567, 3656785, 3617038, 2980565, 3617041, 3641062, 3656784, 3656786, 3656797, 3656799, 3656800, 3656803, 3656804, 3617042, 3617043, 2795454, 3563552, 3641071, 2626734, 2889180, 3656782, 3563561, 2980586, 3641070, 3583214, 2746724, 3617054, 3583197, 3656795, 3583212, 3656794, 3641053, 3617037, 3583196, 3656781, 3656798, 2980582, 3898842, 3898894, 3927939, 3898880, 3915953, 3915966, 3915957, 3915971, 3927958, 3927966, 3927960, 3927978, 3898832, 3898841, 3898810, 3898820, 3898889, 3927934, 3927940, 3959663, 3959664, 3927974, 3959656, 3959665, 3959673, 3959675, 3898897, 3898895, 3898896, 3927954, 3898883, 3898877, 3898817, 2250235, 2587571, 2526595, 2505951, 2178929, 2276403, 2359829, 2587468, 2430608, 2324774, 2587558, 2505958, 2526609, 2359798, 2526586, 2084743, 2587568, 2118361, 2489204, 2587556, 3656790, 2927869, 3563557, 2927871, 2690691, 2626725, 2854970, 2946774, 3583209, 3601478, 3601481, 3583206, 2946780, 3563551, 3617028, 2651502, 2746733, 3563569, 3601486, 2626741, 3656787, 3583215, 3591928, 3641054, 3583211, 3591918, 3641056, 2746718, 2795446, 3927973, 3927970, 3898799, 4384657, 3927967, 3898845, 3927952, 3898838, 3927969, 4095144, 3898831, 3898803, 3915950, 4444453, 3898835, 3898890, 4384655, 3898891, 4095128, 4368680, 4095126, 3959669, 3898812, 3898834, 4267058, 4384648, 4368665, 3898802, 4095134, 3915970, 4095130, 3959654, 3927950, 3898823, 3898825, 3959655, 3927947, 4095143, 4095142, 3915963, 4095135, 4095132, 4095133, 4368676, 4368666, 4384660, 4384664, 4384652, 4444438, 4444439, 4444446, 4444436, 4444451, 4444447, 4444435, 4444437, 4368678, 4444456, 3959661, 3927956, 4444445, 3927957, 3927968, 4095125, 4095124, 4095120, 4095121, 4095138, 3898819, 3915962, 3898878, 3898801, 3915948, 3898886, 3915959, 4384646, 3915965, 4368679, 3898808, 3898893, 3927943, 4267045, 3898873, 3959672, 4368682, 4368670, 3927948, 3959659, 3898874, 4384658, 3959676, 3927963, 3898876, 3927977, 3927938, 3898892, 2498430, 2587471, 2505942, 2587474, 2587550, 2587529, 2526624, 2526607, 2587492, 2587527, 3563570, 3591914, 3601494, 2927858, 2854979, 2927855, 2906928, 2927854, 3591923, 3563577, 3563559, 2746755, 2927859, 2690676, 2906924, 2626748, 2946770, 2736287, 3583204, 3641061, 3601490, 3617049, 3591910, 2980578, 4267050, 3927946, 4474067, 3915967, 4095136, 3898875, 3959660, 4474057, 4474069, 4267060, 4474079, 3898837, 3915947, 4474077, 4474068, 3959668, 3898829, 4444441, 4444442, 4444448, 4267054, 3927949, 3898816, 3927933, 3927941, 4384659, 3959677, 4444450, 3898814, 4368674, 4384666, 4095137, 4444454, 4368672, 4474073, 4384651, 4384653, 3927962, 4474063, 3898821, 5015196, 4776112, 4776111, 4851689, 4875840, 5015165, 4851677, 4851674, 4810717, 4836951, 5015150, 4776105, 4776109, 4776113, 4903160, 4851681, 4851684, 4903142, 4903143, 4903157, 4903141, 4903155, 4903158, 5015155, 5015164, 5015169, 5015171, 5075638, 5075651, 4875848, 4921387, 4776100, 4776091, 4875853, 4875849, 4921380, 5075643, 4776098, 4776104, 4776094, 4810714, 4836950, 4836966, 4836952, 4903144, 4903150, 4903151, 4921385, 5060146, 5075649, 5060145, 5075644, 5015160, 4776095, 4921379, 5015185, 4875858, 4875852, 4875851, 4776110, 5075639, 4776099, 4810711, 2854983, 2746749, 3591927, 2906941, 3641051, 2746748, 3617040, 3591921, 3583218, 2746729, 4474081, 4384644, 4474059, 4444457, 3898887, 4368677, 4095122, 3915954, 3898798, 3898840, 4267055, 4384645, 4474070, 4384643, 4267052, 4095139, 3927945, 5015177, 5015200, 5237105, 4836953, 5015189, 5015174, 5247543, 4851692, 5247551, 5015173, 4851675, 5015175, 4776092, 5015191, 4875847, 5060151, 4836959, 5060152, 5237123, 5247554, 5247528, 5154864, 5154866, 4836949, 5154874, 4810718, 5154872, 5247531, 4810721, 5247525, 5237122, 5237124, 5154882, 5060134, 5154863, 5015187, 5015176, 4836960, 5154881, 4776103, 5075647, 5154878, 5237098, 4903154, 4903159, 5154877, 4776097, 4903149, 4875856, 4921377, 5237108, 5015163, 4875857, 5060140, 4875842, 4875855, 5154871, 5247532, 5247548, 5247530, 4875854, 5075642, 5060138, 5060143, 4875859, 4810706, 5015193, 4921384, 3617027, 3583198, 3601480, 3641050, 4474060, 3898815, 4474065, 4444455, 3898839, 4444459, 4384650, 4474078, 3959666, 4267066, 4474076, 3915951, 3927965, 4384665, 3927961, 4836965, 5015149, 5237121, 5060148, 5015152, 4836948, 5015179, 4776089, 5237109, 4851687, 5154867, 5247546, 5015204, 5237104, 4903152, 5015182, 5247542, 5923651, 5923664, 5923668, 5966284, 5966285, 5966286, 5966293, 5969831, 5969840, 5969841, 5969844, 6005842, 6005764, 6005766, 6005788, 6005797, 6005799, 6005804, 6005808, 6005809, 6004744, 6004733, 6004751, 6004752, 6005846, 6005849, 5924997, 5966280, 6005780, 6005789, 5966296, 5966288, 5969832, 5925005, 6005790, 6005793, 6005263, 6005264, 6004747, 5966282, 5966295, 5924998, 6005761, 5966290, 5923652, 5923659, 4384656, 4474071, 4267048, 5237112, 4776090, 4810719, 5237117, 4851676, 5060153, 5247553, 5015202, 5015195, 4776101, 5015153, 5154865, 5154868, 5015192, 4903145, 4921382, 4810705, 6004748, 6005767, 6005284, 5925002, 6005276, 5925008, 6005272, 6004739, 6005750, 6005751, 6004738, 6480498, 6480506, 6004749, 6506394, 6004750, 6005290, 6480504, 5969834, 6005291, 5925004, 5925001, 5923654, 6421007, 6421014, 6421000, 6421013, 5237097, 5247545, 5015206, 5237113, 4810703, 4851682, 5237118, 5015170, 4836962, 5154875, 5075634, 5154873, 6506404, 6005770, 6506398, 6480503, 6480500, 6480511, 6480502, 6480512, 6480518, 5969826, 6480507, 6506392, 6506407, 6506416, 6480517, 6506419, 6005787, 6506409, 6004743, 6004758, 6005753, 6005796, 6005807, 6506401, 6506402, 6480515, 5924995, 6506413, 6005292, 6004730, 5969833, 5923657, 6506427, 6819510, 6819067, 6819511, 6819512, 6819523, 7032907, 7032922, 7032926, 6819633, 6819639, 6819058, 6819061, 6819064, 6819066, 6819069, 6819070, 6819074, 6819078, 6819091, 6819095, 6421002, 7032920, 7032924, 6819077, 6421016, 6819094, 4921378, 5075652, 5015151, 6005779, 6506396, 6506410, 5923656, 6506403, 6005268, 6005847, 6005782, 5924993, 6004755, 6004740, 6506408, 6506415, 6004746, 5924994, 6004736, 5969829, 6005848, 5925000, 6506417, 6005801, 6480494, 6506405, 6421006, 6819518, 7032913, 7032866, 7032870, 7032837, 7032898, 7032942, 7032843, 7032854, 7032861, 7032899, 7032890, 6819093, 7032871, 7041714, 7041715, 6819087, 7032848, 7032873, 7032892, 7032928, 7032869, 7032882, 7032880, 6819072, 6421017, 6819084, 7032938, 6819082, 7032885, 7032886, 7032921, 6819529, 7041725, 6420991, 5924999, 6005776, 6480514, 6005768, 6480495, 6005777, 6005773, 6506418, 6480499, 7032932, 7032872, 7032931, 7041713, 7032856, 7041731, 7032896, 7032918, 6819076, 7032909, 6420996, 6819065, 6819085, 6819643, 7032895, 7041732, 7032891;
c) Item Number 314-02-14, UDI/DI 10885862084521, Serial Numbers: 0817079, 0855052, 0855033, 0855041, 0855013, 0855017, 0855032, 0817078, 0855026, 0855020, 0855000, 0855007, 0855009, 0855019, 0855062, 0855008, 0854999, 0817071, 0855055, 0855001, 0855004, 0855046, 0855057, 0855027, 0817084, 0855023, 0855066, 0817082, 0817070, 0817072, 0817074, 0817075, 0817076, 0817080, 0817073, 0855014, 0855038, 0855012, 0855058, 0855031, 0855010, 0855035, 0855045, 0855003, 0817083, 0855018, 0855029, 0855011, 0855039, 0855053, 0855044, 0855060, 0855024, 0855002, 0855028, 0855030, 0855021, 0855054, 0855016, 0855005, 0855036, 0855037, 0855051, 0855050, 0855043, 0817081, 0855061, 0855006, 1116033, 1118444, 1118499, 0957760, 0957651, 0957758, 0957747, 0957769, 0957746, 0957628, 0957742, 0957649, 0957648, 0957622, 0957620, 0948359, 0957630, 0957640, 0957766, 0957768, 0957767, 0957644, 0957735, 0957637, 0957763, 0957737, 0957762, 0948361, 0957779, 0957775, 0957777, 0957638, 0957761, 0948363, 0957617, 0957772, 0957736, 0957656, 0957754, 0957625, 0957652, 0957752, 0957635, 0957615, 0957614, 0957624, 0957741, 0957647, 0957658, 0948362, 0957744, 0957748, 0957783, 0957641, 0957745, 0957757, 0957753, 0957764, 0957632, 0957765, 0957627, 0855047, 0855040, 0855042, 0855059, 0855067, 0855049, 0855015, 1116034, 1118448, 1118505, 1118486, 1116037, 1116022, 1118514, 1116031, 1118498, 1118481, 1116013, 1118492, 1116014, 1118506, 1116016, 1116036, 1118469, 1118491, 1118468, 1118482, 1118485, 1118493, 1118502, 1118455, 1118459, 1116023, 1118479, 1118447, 1118445, 1118512, 1118484, 1118465, 1118478, 1118466, 1118488, 1118460, 1116020, 1118449, 1118507, 1118476, 1116025, 1118504, 1118451, 1116029, 1116030, 1116026, 1118472, 1116032, 1118516, 1118474, 1118471, 1116018, 1118515, 1118450, 1116017, 1118489, 1118496, 1118470, 1118494, 1118518, 1118511, 1118457, 1118456, 1118495, 1116019, 1118446, 1118509, 0957773, 0957634, 0957781, 0957642, 0957743, 0957616, 0957629, 0957626, 0957770, 0957618, 0948360, 0957621, 0957739, 0957631, 0957643, 0957740, 0957774, 0957646, 0957659, 0957780, 0957660, 0957751, 0957654, 0957650, 0957619, 0957653, 0957633, 0957738, 0957636, 0957755, 0957655, 0957782, 0957750, 1228763, 1228758, 1228760, 1228677, 1228717, 1228752, 1279896, 1228709, 1228740, 1279938, 1228694, 1279894, 1228676, 1228666, 1279943, 1361419, 1228671, 1279917, 1279937, 1279880, 1228714, 1228702, 1228733, 1279893, 1228686, 1228756, 1279887, 1228718, 1228727, 1279874, 1228747, 1279922, 1228690, 1228662, 1228757, 1228746, 1279947, 1228697, 1279928, 1279911, 1279952, 1228739, 1279907, 1228726, 1228710, 1361452, 1228687, 1228715, 1228673, 1279910, 1228729, 1279930, 1228688, 1279942, 1279903, 1228664, 1228701, 1361407, 1279879, 1361463, 1228665, 1228680, 1279921, 1228742, 1361406, 1361409, 1361416, 1361420, 1361451, 1279914, 1279935, 1228716, 1228695, 1361423, 1361434, 1279912, 1228658, 1228678, 1228699, 1228700, 1228703, 1228711, 1228737, 1279873, 1279890, 1279891, 1279898, 1279899, 1279919, 1279920, 1279940, 1228748, 1279900, 0855022, 0855063, 0817077, 1118501, 1118508, 1116035, 1118475, 1118497, 1116021, 1118490, 1116028, 1118463, 1118462, 1118480, 1116024, 1118467, 1118461, 1118458, 1118454, 1118453, 0957623, 0957776, 0957645, 0957771, 1361437, 1404960, 1279881, 1433245, 1433241, 1279936, 1361431, 1433248, 1503516, 1549089, 1279949, 1361410, 1404984, 1404986, 1361433, 1423439, 1503524, 1361453, 1404946, 1404972, 1423434, 1279897, 1361440, 1361418, 1228685, 1279888, 1228668, 1361430, 1503519, 1404985, 1361464, 1279906, 1535555, 1279876, 1549095, 1503511, 1589065, 1279934, 1361445, 1228736, 1279877, 1404978, 1279968, 1228674, 1228661, 1228669, 1589069, 1228712, 1279882, 1549083, 1535543, 1228741, 1279883, 1361456, 1404944, 1279913, 1361424, 1361413, 1404942, 1433243, 1433246, 1433238, 1423442, 1404981, 1404958, 1279905, 1279901, 1228683, 1423446, 1423422, 1228724, 1279895, 1404971, 1423440, 1228725, 1404948, 1503515, 1423436, 1404943, 1535566, 1503518, 1589072, 1361411, 1503523, 1361462, 1361417, 1361427, 1423443, 1404950, 1361449, 1404955, 1535565, 1279957, 1361428, 1279958, 1228753, 1228679, 1404951, 1433234, 1279961, 1423447, 1535549, 1404989, 1279875, 1404983, 1228659, 1228675, 1423428, 1279953, 1522718, 1404962, 1361444, 1589079, 1361412, 1279944, 1404952, 1433247, 1228723, 1503513, 1404974, 1361422, 1535554, 1535552, 1535563, 1535551, 1549099, 1589043, 1503521, 1279970, 1228720, 1361429, 1361421, 1279885, 1228764, 1404954, 1404969, 1361458, 1228692, 1423427, 1279909, 1279948, 1522719, 1423441, 1535568, 1404949, 1404956, 1535559, 1279884, 1522721, 1549101, 1589076, 1549092, 1279904, 1279966, 1279927, 1361442, 1503504, 1503526, 1279929, 1522728, 1535542, 1228738, 1228722, 1279941, 1361426, 1279918, 1279878, 1279939, 1228670, 1535557, 1522720, 1228754, 1279923, 1423420, 1404964, 1361435, 1433252, 1228657, 1361447, 1279931, 1361439, 1361446, 1404967, 1361438, 1433240, 1503512, 1589080, 1503514, 1279951, 1279945, 1228755, 1404945, 1404953, 1423431, 1423435, 1423430, 1503509, 1279925, 0855048, 0855056, 1116027, 1118464, 1116015, 1118483, 1118487, 1118513, 0957756, 0957778, 0957639, 0957749, 1361450, 1589071, 1589086, 1228732, 1589081, 1404961, 1549084, 1535545, 1404980, 1279916, 1535567, 1404966, 1361454, 1361425, 1522727, 1589089, 1404973, 1589088, 1433237, 1279964, 1522722, 1423445, 1549098, 1549103, 1549090, 1535564, 1589049, 1589083, 1589061, 1589051, 1549097, 1589058, 1228705, 1549100, 1361415, 1589064, 1404982, 1589045, 1404977, 1279962, 1423438, 1589087, 1503507, 1589082, 1589054, 1589059, 1589066, 1404988, 1361408, 1433239, 1535546, 1279971, 1228660, 1228682, 1503525, 1423421, 1279956, 1404970, 1228689, 1228750, 1549087, 1589053, 1549104, 1589084, 1589090, 1549093, 1279902, 1423433, 1361432, 1589048, 1279963, 1404976, 1404957, 1433235, 1535556, 1535550, 1535553, 1433250, 1522717, 1361455, 1423419, 1589077, 1589044, 1535544, 1404947, 1535562, 1423437, 1589070, 1228719, 1589057, 1503510, 1535561, 1589062, 1279967, 1589055, 1535547, 1279926, 1279932, 1589047, 1589042, 1522724, 1279950, 1361448, 1279933, 1535558, 1279908, 1423424, 1549088, 1228761, 1589067, 1423444, 1228681, 1361441, 1361460, 1589063, 1423432, 1589046, 1279960, 1404965, 1589041, 1503508, 1279954, 1503505, 1626867, 1727222, 1827405, 1677004, 1727220, 1804634, 1728500, 1626828, 1762473, 1727237, 1762480, 1676987, 1626859, 1676983, 1626827, 1786810, 1728509, 1722261, 1727226, 1819189, 1786792, 1827422, 1626852, 1819207, 1774789, 1626855, 1626844, 1786813, 1626865, 1677014, 1626838, 1626831, 1804628, 1827428, 1762464, 1827411, 1819197, 1819193, 1728493, 1728494, 1676999, 1727223, 1804648, 1819181, 1676984, 1762479, 1762474, 1626856, 1774788, 1677009, 1804616, 1676990, 1804636, 1827421, 1626864, 1722257, 1728497, 1804622, 1786812, 1626854, 1728518, 1722259, 1626850, 1727224, 1762466, 1774805, 1762467, 1774797, 1676991, 1626851, 1626875, 1677005, 1819204, 1626845, 1728506, 1626862, 1677006, 1727213, 1727235, 1728501, 1804632, 1774791, 1762455, 1786806, 1819208, 1728515, 1804629, 1774792, 1804614, 1626841, 1727225, 1762475, 1626874, 1676982, 1626840, 1774777, 1819176, 1774782, 1676985, 1827419, 1626877, 1762460, 1786808, 1819179, 1804612, 1774807, 1804624, 1819200, 1819186, 1626820, 1626857, 1819205, 1626833, 1626821, 1626869, 1677010, 1728498, 1762463, 1626832, 1786793, 1722256, 1786804, 1819203, 1626870, 1626825, 1626835, 1626836, 1677003, 1626834, 1626847, 1626823, 1727233, 1626830, 1626848, 1722263, 1728504, 1727234, 1728496, 1774802, 1727238, 1786800, 1774794, 1727232, 1774785, 1804635, 1804618, 1819206, 1804620, 1804631, 1762458, 1786798, 1827409, 1819191, 1804638, 1774796, 1626868, 1827418, 1626873, 1676993, 1677011, 1804621, 1762481, 1676994, 1677013, 1774799, 1819177, 1827403, 1827432, 1727230, 1827404, 1722258, 1804623, 1827431, 1626853, 1774800, 1728516, 1774806, 1762459, 1676996, 1819183, 1804643, 1677015, 1676997, 1677012, 1727215, 1728499, 1728513, 1626846, 1626829, 1727221, 1727231, 1728512, 1727236, 1804627, 1804615, 1804644, 1804646, 1804647, 1804650, 1827406, 1827407, 1827427, 1804639, 1762469, 1762471, 1819209, 1786796, 1774798, 1774801, 1626822, 1722260, 1762472, 1118500, 1279889, 1228745, 1503506, 1549085, 1228704, 1433242, 1228735, 1522723, 1549105, 1228706, 1361461, 1404959, 1522725, 1423423, 1549086, 1228696, 1228759, 1361443, 1549094, 1589078, 1433249, 1228751, 1228731, 1503503, 1228734, 1279924, 1228667, 1279892, 1589052, 1361414, 1228707, 1503520, 1404990, 1589085, 1279946, 1361457, 1228749, 1279969, 1404987, 1503517, 1535560, 1228743, 1228663, 1228728, 1549091, 1404963, 1626872, 1762462, 1819201, 1774778, 1926114, 1774790, 1936462, 1774804, 1988328, 1925796, 1988311, 1988315, 2012655, 1925786, 1925797, 2002088, 1925768, 1988292, 2002100, 2002097, 1786803, 2070783, 2002101, 1936461, 1926104, 2002079, 2002090, 1727218, 1936482, 1936476, 1728514, 1762456, 2070794, 1988320, 1925763, 2055998, 1762470, 1936475, 2012637, 1827430, 1925765, 1936443, 1925790, 1936467, 1925766, 1936460, 1988304, 1988318, 2056003, 2002075, 1626837, 1936455, 1925770, 1936446, 1827402, 1827413, 2012636, 1827398, 1936454, 1988313, 1819185, 1626876, 1936444, 1988312, 1925785, 1626826, 1925798, 1988298, 1988299, 2012649, 1988329, 2002103, 1936477, 2056010, 2012635, 2070802, 1988330, 2070804, 2070811, 2055996, 1819190, 1786794, 1926100, 1774779, 1819180, 1819192, 1762478, 1676992, 1819184, 1925788, 1988290, 1925779, 1762461, 1936481, 1988324, 2002077, 2012630, 1786797, 2012653, 1728503, 1677008, 1827417, 1988308, 1988307, 2070790, 1762457, 1926102, 1626861, 1936473, 1786807, 1827420, 1827400, 1936463, 1827429, 2056035, 2012652, 2002099, 2070786, 1819195, 1827424, 2056034, 2056022, 1988331, 1926120, 1926116, 1804633, 1926105, 1786814, 1786802, 1925771, 1925767, 2012660, 2070796, 2070808, 1936445, 1804613, 1925781, 1936457, 1926092, 1936471, 1988305, 1988310, 1936456, 1925775, 2002089, 1727219, 1626843, 1819196, 2012641, 2070810, 2012643, 2056027, 1926098, 1819198, 2070797, 1676995, 2012644, 1925764, 1988302, 2002083, 2070812, 1988294, 2056038, 2002078, 2002091, 2056030, 1762465, 2012659, 1804637, 1677000, 1774803, 2012657, 2070799, 1786816, 1804619, 1926118, 1804645, 1728507, 1936474, 1926121, 1925795, 1925791, 1722262, 1925778, 1988306, 1988296, 1626863, 2056012, 1728519, 1827401, 2056033, 1827414, 1988317, 1774784, 1926097, 2002098, 1925772, 1936452, 1804640, 2070792, 1925794, 1936451, 1926099, 1925780, 2056020, 2070805, 2056025, 1925783, 1936448, 2056009, 1925784, 1827415, 1786795, 1925789, 1819178, 1936469, 1925793, 1925776, 1926101, 1988323, 1774787, 1988326, 1988293, 1827425, 2012642, 1988333, 2002087, 1827399, 1925769, 2002086, 1936466, 2002095, 2070801, 2070784, 1926094, 1925800, 1728510, 1819202, 1774786, 1819199, 1926109, 1827416, 1926096, 1819194, 1936447, 1626824, 1936459, 1727216, 1936453, 1988314, 1988291, 1786817, 1827423, 1774781, 2012632, 1936479, 1626842, 2056008, 1727227, 1988309, 1819210, 1926119, 1936470, 1804641, 1804649, 1926107, 1925777, 2012646, 1988316, 1676989, 2012634, 2012640, 2012648, 2012656, 2056013, 2070800, 1786805, 2002081, 2183075, 2116973, 2082886, 2082878, 2082875, 2132650, 2132668, 2132670, 2082884, 2082889, 2082881, 2116943, 2082879, 2145730, 2116965, 2116970, 2132673, 2145710, 2145713, 2132665, 2116982, 2116963, 2116947, 2082876, 2132649, 2116944, 2116983, 2116985, 2145726, 2145721, 2169999, 2183072, 2116957, 2116988, 2132674, 2082882, 2132678, 2132659, 2132660, 2170025, 2170016, 2170017, 2132656, 2082885, 2132652, 2132663, 2132667, 2170021, 2145715, 2082898, 2145705, 2082880, 2116942, 2116951, 2132676, 2183069, 2116956, 2116986, 2116949, 2170008, 2170006, 2170003, 2183081, 2170018, 2116959, 2132657, 2132661, 2145722, 2170002, 2082877, 2170000, 2116958, 2116961, 2116966, 2116967, 1118477, 1228693, 1228672, 1228708, 1535548, 1549102, 1589050, 1228691, 1549096, 1433236, 1228762, 1361459, 1423426, 2002084, 1827410, 2056000, 1926103, 1676998, 2056018, 2002094, 1626866, 1936458, 1728495, 1774795, 2002096, 1988334, 1936478, 2056014, 2070806, 1925799, 1804630, 2056001, 1626849, 1926113, 1677007, 1727228, 2012658, 1925773, 2012627, 2070809, 1774793, 1676986, 1626871, 1728511, 2056011, 2056017, 2012638, 2002102, 2012633, 2056029, 1926095, 1926117, 1936472, 1988319, 2012628, 1926111, 2002092, 2012629, 2055997, 2056040, 2056006, 1804626, 2012626, 2056032, 2012631, 1819188, 2070793, 2012639, 1804642, 2056026, 1926093, 1804617, 2055995, 1936450, 2012650, 1988321, 2012651, 2070788, 2070791, 1926106, 2056024, 2056036, 1988301, 2070795, 1988325, 1936449, 2056002, 2070785, 1988300, 1727214, 1936465, 2002082, 1936468, 1676988, 2056019, 2440715, 2381828, 2296603, 2451419, 2451430, 2280116, 2253043, 2183064, 2204662, 2280112, 2322465, 2322485, 2440711, 2451449, 2132655, 2204669, 2204659, 2238803, 2204661, 2204629, 2280138, 2204660, 2296623, 2322460, 2280131, 2393477, 2322456, 2253033, 2323741, 2145731, 2204664, 2280139, 2204639, 2116968, 2238778, 2522821, 2253046, 2381641, 2238798, 2116952, 2323738, 2145732, 2204671, 2322455, 2238781, 2145711, 2170014, 2451472, 2204668, 2238788, 2204646, 2082891, 2381824, 2451423, 2253048, 2322457, 2381835, 2204670, 2204627, 2280109, 2204633, 2116940, 2145717, 2280122, 2204677, 2238795, 2170019, 2280108, 2132664, 2253024, 2323729, 2393459, 2451463, 2116989, 2280127, 2296615, 2451411, 2451436, 2204636, 2116964, 2204651, 2393469, 2253028, 2381838, 2132651, 2296607, 2204630, 2170010, 2238797, 2253053, 2323711, 2322483, 2204626, 2204640, 2253045, 2322492, 2322470, 2253049, 2253042, 2204667, 2116974, 2238789, 2253032, 2280136, 2132666, 2280105, 2204672, 2170029, 2451499, 2280129, 2253027, 2204650, 2170026, 2204680, 2204635, 2253044, 2170015, 2082872, 2280132, 2322491, 2323714, 2238780, 2381638, 2323716, 2170011, 2323734, 2280134, 2381842, 2451414, 2296598, 2238776, 2296618, 2170001, 2322474, 2204638, 2204649, 2204666, 2280120, 2204631, 2296617, 2323733, 2322484, 2296606, 2170032, 2253036, 2451454, 2440717, 2183079, 2204637, 2393455, 2238802, 2280135, 2505910, 2451481, 2451479, 2381823, 2145706, 2381825, 2323707, 2204632, 2296610, 2145734, 2253056, 2381632, 2451413, 2116969, 2280106, 2296619, 2296600, 2204655, 2183067, 2183062, 2253021, 2296616, 2204674, 2253058, 2145729, 2183071, 2238777, 2296601, 2238794, 2170027, 2082888, 2280103, 2253030, 2170024, 2170012, 2253052, 2116972, 2253029, 2451474, 2381623, 2238786, 2451447, 2132658, 2170005, 2204679, 2145714, 2204663, 2451492, 2204642, 2204628, 2145733, 2238785, 2145728, 2253047, 2238796, 2183080, 2280118, 2451429, 2145719, 2381811, 2132654, 2381819, 2253031, 2381814, 2322490, 2204658, 2183063, 2238799, 2170031, 2393454, 2082892, 2170013, 2116978, 2322454, 2238784, 2082887, 2145707, 2238783, 2170009, 2322469, 2280128, 2132677, 2323737, 2145724, 2381822, 2116962, 2082895, 2253022, 2381834, 2132671, 2296609, 2381628, 2505915, 2543283, 2381836, 2296602, 2296605, 2296613, 2296614, 2322463, 2322471, 2253034, 2322468, 2322481, 2322482, 2322489, 2280102, 2296608, 2393462, 2393476, 2381626, 2381642, 2381625, 2393467, 2170020, 2381816, 2440708, 2440712, 2440710, 2451424, 2451438, 2451444, 2451470, 2451503, 2296604, 2451448, 2451458, 2451465, 2451475, 2451477, 2505903, 2505908, 2280130, 2393453, 2451453, 2238790, 2238791, 2238800, 2505916, 2322479, 2204656, 2322493, 2132669, 2280125, 1361436, 1589075, 1279872, 1589068, 1404975, 1926108, 1786799, 2002085, 1728517, 2012645, 2055999, 1926110, 2070789, 2002093, 2056015, 1819182, 2056021, 2056037, 2056007, 1786811, 1925774, 2056004, 1827426, 1925787, 1786815, 2323724, 2253041, 2381639, 2451437, 2381833, 2183070, 2522809, 2451468, 2082874, 2451445, 2204643, 2522808, 2451432, 2280133, 2322461, 2393465, 2543307, 2253050, 2543313, 2451441, 2505888, 2393481, 2440714, 2204653, 2393464, 2543285, 2505907, 2543300, 2505912, 2451421, 2505904, 2381840, 2322459, 2145725, 2543297, 2522806, 2323709, 2451409, 2393450, 2505887, 2132662, 2204645, 2543293, 2543323, 2451500, 2323718, 2451446, 2522803, 2522801, 2451408, 2543309, 2543281, 2543301, 2323722, 2522843, 2381630, 2522816, 2082871, 2280107, 2505918, 2451422, 2145708, 2522824, 2522826, 2322487, 2322458, 2522815, 2116960, 2116953, 2280113, 2204647, 2522822, 2451410, 2322488, 2451461, 2381839, 2280141, 2451460, 2116976, 2451493, 2440709, 2170030, 2522811, 2183068, 2253037, 2280104, 2451507, 2296621, 2323715, 2296624, 2381817, 2522802, 2322472, 2204641, 2505913, 2451431, 2522797, 2522842, 2522832, 2522827, 2183076, 2381635, 2451490, 2381831, 2505890, 2451484, 2253060, 2183077, 2440713, 2522805, 2543287, 2505894, 2323736, 2145709, 2393474, 2323719, 2145718, 2296622, 2543280, 2393475, 2451434, 2522807, 2323710, 2451443, 2522813, 2543275, 2543319, 2543308, 2451420, 2543316, 2451426, 2082873, 2522836, 2451459, 2522833, 2322464, 2451471, 2280124, 2183065, 2393460, 2451417, 2381818, 2204652, 2543320, 2543322, 2323725, 2543317, 2638953, 2648870, 2560855, 2560851, 2560844, 2690755, 2690771, 2560850, 2648884, 2560833, 2648896, 2638958, 2844489, 2790595, 2638955, 2560849, 2690757, 2560842, 2648905, 2648901, 2560831, 2648874, 2560836, 2560854, 2648873, 2648900, 2560843, 2648899, 2648903, 2560837, 2690786, 2560835, 2648908, 2690783, 2690788, 2690791, 2690760, 2648893, 2690764, 2629824, 2560853, 2648876, 2648877, 2648885, 2648886, 2648891, 2648892, 2648906, 2629811, 2629821, 2629823, 2629832, 2629838, 2629839, 2629814, 2758994, 2758996, 2758997, 2758998, 2758999, 2759001, 2758983, 2629816, 2629837, 2758986, 2758988, 2758991, 2758995, 2790579, 2790574, 2790575, 2790577, 2790586, 2790589, 2790592, 2629822, 2790597, 2790587, 2790591, 2790602, 2853431, 2853432, 2853433, 2853434, 2853439, 2853440, 2853442, 2853443, 2853444, 2853447, 2853449, 2853451, 2629836, 2629827, 2853428, 2853429, 2853430, 2853435, 2853436, 2853438, 2853441, 2853445, 2853446, 2853448, 2853450, 2648871, 2648882, 2638944, 2638956, 2648883, 2638949, 2629825, 2758984, 2758989, 2758990, 2758993, 2758992, 2790590, 2790598, 2790603, 2638947, 2758985, 2758987, 2759000, 2560848, 2629831, 2638952, 2070798, 2012654, 1774780, 1804625, 1936480, 2002074, 2505899, 2505897, 2170023, 2381627, 2393457, 2505920, 2381636, 2082896, 2204673, 2323723, 2322475, 2145720, 2522837, 2381634, 2253039, 2505911, 2543282, 2381633, 2543296, 2381631, 2280110, 2451456, 2505891, 2522840, 2381624, 2543277, 2543324, 2204678, 2322466, 2451495, 2381820, 2543302, 2253054, 2381622, 2543276, 2170007, 2082893, 2451467, 2451415, 2543318, 2323717, 2323732, 2451455, 2543278, 2183074, 2323739, 2253059, 2522831, 2323740, 2253025, 2381841, 2323726, 2505896, 2522810, 2204657, 2296599, 2522845, 2451489, 2204676, 2451466, 2451498, 2381815, 2116955, 2323721, 2393463, 2322462, 2116980, 2116984, 2116987, 2204644, 2322486, 2629826, 2790606, 2961268, 2844491, 2648897, 2886761, 2648904, 2844485, 2690772, 2961277, 3565522, 2690781, 2961265, 2886756, 2844482, 3565517, 2954050, 2690780, 2922156, 2629813, 2690765, 2844496, 2922152, 3565503, 2961263, 2648898, 2790600, 2690763, 2560840, 2560841, 2690769, 2648895, 2560846, 2961275, 2690779, 2844499, 2648907, 2690754, 2844498, 2922172, 3565527, 2638945, 2629842, 2922157, 2560839, 2690793, 3565513, 3565509, 2690750, 2648872, 2844476, 2629828, 2690778, 2690782, 2844481, 2886770, 2638954, 2954065, 3599827, 2629830, 2954053, 2886762, 2886766, 2886763, 2922174, 2954066, 2954054, 2886773, 3592009, 2629818, 2886774, 2922160, 2961283, 3565519, 3565535, 2790607, 2922165, 2844486, 2790583, 2886769, 2844474, 2790596, 1988303, 1988295, 1926115, 2002080, 2505895, 2393472, 2116977, 2145716, 2082897, 2505922, 2393451, 2296620, 2505892, 2451485, 2451480, 2322476, 2322480, 2116971, 2253057, 2116954, 2451452, 2522830, 2543321, 2322477, 2451482, 2451496, 2451425, 2322467, 2543284, 2381645, 2451462, 2522798, 2451488, 2440716, 2323720, 2451433, 2393471, 2323727, 2296612, 2451504, 2505898, 2505923, 2451457, 2116981, 2381827, 2393473, 2253040, 2543288, 2543312, 2648878, 2922173, 2886758, 2844477, 2954044, 2954047, 3592007, 3592004, 2922164, 2560845, 3599826, 2638960, 3591999, 3660746, 3592021, 2690756, 2648902, 2844494, 2961271, 2844483, 2961276, 3599815, 3565533, 3632628, 3632609, 2844490, 2690767, 3592001, 2690768, 3565508, 3565512, 3565518, 2690753, 3599837, 2844479, 2690751, 2690794, 3565505, 3599817, 2648875, 2629834, 2844475, 3599836, 3599814, 3632620, 2690762, 3660725, 2922161, 3632608, 3632631, 3599823, 3660748, 3592005, 2954048, 2954057, 2954068, 2954059, 2954062, 2954063, 3599830, 3599834, 3632610, 2922166, 2922175, 2961270, 3632606, 3632612, 3632619, 3632625, 3660743, 3660747, 3660752, 3660753, 3660732, 3660738, 3660749, 3565514, 3599821, 2629833, 2629835, 3599813, 2790601, 2954067, 3565525, 2961274, 3632621, 3915973, 3927921, 3915994, 3927922, 3927881, 3915977, 3899875, 3927894, 4053246, 3927891, 3927899, 3915982, 3915988, 3927885, 4086012, 3927918, 3915979, 4086019, 3915992, 4086020, 3927887, 3927925, 3927926, 3899872, 3927928, 2056028, 2543299, 2505889, 2543289, 2381829, 2451483, 2451412, 2238787, 2543292, 2145727, 2132672, 2505901, 2451487, 2451494, 2451501, 2505921, 2522834, 2543290, 2522820, 2522818, 2543310, 2393470, 2204648, 2322473, 2844480, 3632607, 2844488, 2844493, 2560847, 2844501, 2690758, 2790581, 3565524, 2690775, 2560832, 2961267, 2954052, 2922155, 2638946, 2648894, 3592011, 2922167, 3599820, 2638959, 3592008, 3660750, 3592014, 2886757, 3599816, 2690790, 3660751, 2638957, 2629840, 3660735, 3915996, 4221373, 4221378, 4221393, 3927927, 4053245, 3899885, 4221396, 3927896, 3927906, 4475483, 4053261, 3899892, 4221425, 3899876, 3915981, 3927904, 3927915, 3927923, 3927917, 4053264, 4053266, 3915991, 4221405, 4053256, 4221400, 4221414, 4221371, 4414841, 4414840, 4414845, 4414846, 4475481, 4475461, 4475466, 4475477, 4475494, 4475485, 4086023, 4086009, 4221382, 4221389, 4221370, 4086010, 4475468, 4221421, 4414848, 3927919, 4221376, 4414839, 3899888, 4086030, 3915980, 3915987, 3899879, 2505914, 2543298, 2451439, 2381810, 2543315, 2954049, 2844502, 2961280, 3632622, 2961264, 2690774, 3632611, 2629812, 2954064, 2690746, 3565531, 2922171, 3592010, 3599831, 2922154, 3660728, 2922159, 2629841, 3660733, 2961269, 2629817, 3599825, 2790588, 2922158, 4221383, 4221395, 3927901, 4053255, 4221417, 3927907, 3927892, 4221368, 4053253, 4053265, 4221422, 4475478, 4221403, 4086028, 3927888, 4221391, 4414824, 4053244, 4221390, 4221428, 4221369, 4414826, 4414836, 3927913, 4475482, 3927924, 4475484, 3899890, 4221385, 4414829, 4086027, 3915972, 3927929, 4837204, 4982773, 4884317, 4795878, 4884312, 4837205, 4884313, 5085526, 4884302, 4825436, 4903216, 4837197, 4884311, 4982787, 4933006, 4933018, 4982774, 4982777, 4982779, 4982768, 4982772, 4825435, 4933007, 4837206, 5085533, 4933005, 5085531, 4825434, 5060125, 5173052, 4982771, 1626839, 3660742, 3565516, 3599833, 3592018, 3632614, 3599828, 2886771, 3660727, 2790576, 2790594, 2790599, 2844497, 2844478, 3915983, 4053254, 4475475, 3899877, 4053258, 3927910, 4086014, 4414844, 4086022, 3927882, 4475486, 4086025, 4221406, 3915990, 3915985, 4086032, 4414827, 4903199, 4825433, 4837199, 4933002, 5085534, 4982786, 4932999, 5237126, 5173055, 5060132, 4825442, 5060130, 4884301, 5060129, 5173053, 5173076, 5237130, 5173061, 5173074, 5237144, 5237153, 5173077, 4795883, 4933009, 5237127, 4903209, 5237140, 4825449, 5060116, 5085520, 4825437, 5085523, 4884300, 3592003, 3660737, 3899874, 4221381, 3927912, 4086024, 3927903, 4221433, 3899889, 4414833, 4086033, 4903203, 4982770, 4795875, 4837202, 4837196, 5060122, 4903198, 5237135, 5237133, 4837198, 5173064, 4933008, 4933017, 5085519, 4884308, 5060124, 4982783, 5924675, 5925139, 5925133, 5925142, 5925092, 5925100, 5925101, 5925103, 6105923, 6105933, 6105939, 6105946, 5925066, 5925164, 6105896, 6105907, 6105918, 6105929, 6105936, 6105941, 6105942, 5924659, 5925067, 5925165, 5925056, 5925157, 5925095, 5925155, 5925156, 5924660, 5925131, 5925058, 5925049, 5924669, 5925096, 5925102, 4475465, 3927902, 4221408, 4475491, 4221418, 5173066, 5237136, 5237152, 5237128, 4837201, 4903211, 5237147, 5237142, 5060123, 5060115, 4795868, 5924673, 5925055, 6105906, 5925068, 5925130, 5925136, 5925106, 5925062, 5925107, 6571579, 6571591, 6571597, 6571603, 6571610, 6571611, 6571617, 6571538, 6571616, 6571580, 6571582, 6571587, 6571605, 6571491, 6571499, 6571508, 6571509, 4414838, 4903205, 5060118, 5060121, 5237132, 5173056, 5060114, 5085525, 5173058, 5085528, 5925154, 6105964, 5924676, 6105902, 5925162, 6105950, 6105914, 6105932, 6105960, 6105904, 5924664, 5925137, 6571507, 6573020, 6573036, 6571607, 6571493, 6571494, 6571484, 6571506, 6571593, 6573003, 6573013, 6571529, 6645467, 6571504, 6573034, 6571606, 6571503, 6573007, 6573027, 6573022, 6571601, 7031762, 4903206, 5085536, 4982769, 4933001, 6105975, 6105928, 5925134, 5925064, 6105901, 5925160, 6105927, 5925147, 6105940, 6645502, 6645519, 7031770, 7031794, 7031671, 7031676, 7031682, 7031695, 7031702, 7031674, 7031718, 6571581, 6573033, 7031761, 7031786, 7031672, 7031673, 7031678, 7031679, 6571525, 6571613, 6573008, 6571590, 7031757, 6573005, 5060126, 6105962, 6105898, 5925050, 7040400, 7031765, 6571488, 7031771, 6571483, 7031677, 7031690, 7031712.7031714, 7031703, 7040396, 7031764, 7031778, 7040437, 7040440, 7040422, 7031785, 7040435, 7040406, 7031716, 7031689, 7031709, 7031800, 7040407, 7040409, 7040430, 6645510, 7031706, 6571490, 7040426, 7031779, 7031685, 6571518, 6571527, 7031782, 7040402, 6571481, 6571514, 6645475, 6645477, 6645498, 7031759, 6645491, 7031758, 7031790, 6573030, 7031799, 7031783;
d) Item Number 314-02-15, UDI/DI, 10885862084538, Serial Numbers: 1936710, 1936726, 1936722, 2010829, 2103110, 2140614, 2010824, 2010799, 2010830, 2010801, 2010827, 2010797, 1965997, 1966019, 2010809, 2010812, 2010814, 1936724, 2140615, 2016861, 1936713, 1966014, 1965995, 1936704, 1936742, 1965987, 1965991, 2016853, 1966022, 1965692, 1965686, 2103122, 1966021, 1965988, 1936705, 2103124, 1936707, 1936739, 2103130, 2140612, 2239792, 2010819, 1965992, 2140602, 2010798, 2239775, 2140600, 2239791, 1936709, 2103104, 2016850, 1966008, 1965985, 2016856, 1965999, 2140606, 2194933, 2016870, 2016867, 1936735, 1936733, 2512944, 2654840, 2010831, 2016872, 1966011, 2239794, 2239803, 2538037, 2103123, 2140607, 1936725, 2194949, 2140603, 2194924, 1965687, 2654878, 2239789, 2512963, 1965670, 2194955, 1966013, 2239802, 2654837, 2103117, 2239812, 2103129, 2194938, 1966005, 2016849, 2016866, 2140619, 3591032, 3591049, 2907652, 2654880, 1936720, 1965701, 2512945, 2654877, 2103127, 2239811, 2194926, 2654841, 2654848, 2654857, 2654865, 2654870, 2654881, 2654859, 2654867, 3591030, 3591029, 2239779, 2239808, 2538025, 4141882, 3609986, 3610005, 2907637, 2538017, 2654876, 4209701, 3591022, 3591050, 4222532, 5177380, 4595390, 4913656, 4627746, 5163375, 4535575, 4535578, 5163390, 5163389, 5177356, 5327684, 5327693, 5971607, 5971610, 4559166, 4222496, 4695071, 4695076, 6059455, 6883333, 6883336, 6883338, 6059411, 6059492, 6059409, 6059482, 6883014, 6882987, 6059423, 6059471
|
Recalling Firm/ Manufacturer |
Exactech, Inc. 2320 Nw 66th Ct Gainesville FL 32653-1630
|
| For Additional Information Contact | Thien Doan 352-377-1140 |
Manufacturer Reason for Recall | The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken:
"" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required.
" IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing.
" EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession.
" COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com.
The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device.
Contact your local Exactech Representative if you have any questions regarding this communication. |
| Quantity in Commerce | 6840 units |
| Distribution | Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KWS
|
|
|
|