 Class 2 Device Recall Centurionsee related information
Date Initiated by FirmFebruary 23, 2024
Date PostedMarch 29, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1439-2024
Recall Event ID 94136
Product Classification Curette - Product Code HTF
ProductCenturion Curette kits labeled as: a) STERILE BUCK EAR CURETTE "00" (25697), Product Code 66745; b) STERILE "1" EAR CURETTE (505629), Product Code 67050; c) STERILE FOX DERMAL CURETTE, 5MM(32923), Product Code 67430; d) STERILE EXCAVATOR CURETTE (P/S), Product Code I67970; e) ST BUCK EAR CURETTE BLNT STR SZ 00, Product Code I68040; f) STERILE BRUNS CURETTE #4 (BC400S), Product Code I68340; g) 6MM FOX DERMAL CURETTE STERILE, Product Code I68380; h) STERILE KEVORKIAN CURETTE 4MM (KEC4S), Product Code I68470; i) ST ENDOCERVICAL BIOPSY CURETTE 2MM, Product Code I68480
Code Information a) 66745, UDI/DI 70653160060330 (case) 00653160060331 (each), Lot Numbers: 2023103090; b) 67050, UDI/DI 00653160357509 (case) 10653160357506 (each), Lot Numbers: 2023103090; c) 67430, UDI/DI 10653160123514 (case) 00653160123517 (each), Lot Numbers: 2023022190; d) I67970, UDI/DI 20653160224744 (case) 00653160224740 (each), Lot Numbers: 2023041390, 2023041490, 2023061990, 2023101290; e) I68040, UDI/DI 20653160228605 (case) 00653160228601 (each), Lot Numbers: 2023022190, 2023102090; f) I68340, UDI/DI 00653160279863 (case) 10653160279860 (each), Lot Numbers: 2023042490; g) I68380, UDI/DI 00653160284553 (case) 10653160284550 (each), Lot Numbers: 2023032790, 2023101790; h) I68470, UDI/DI 00653160290721 (case) 10653160290728 (each), Lot Numbers: 2023032190, 2023032390; i) I68480, UDI/DI 00653160292640 (case) 10653160292647 (each), Lot Numbers: 2023013190
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
FDA Determined
Cause 2
Process control
ActionMedline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.
Quantity in Commerce20614 units
DistributionWorldwide distribution - US Nationwide and the countries of Panama, Canada.