• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall pro mattress

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall pro mattresssee related information
Date Initiated by FirmMarch 05, 2024
Date PostedMarch 29, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1400-2024
Recall Event ID 94192
Product Classification Bed, flotation therapy, powered - Product Code IOQ
ProductHill-Rom PRO+ 36" MRS Surface, REF P7924A03, The pro+ mattress is intended for patient support and for the prevention and/or treatment of pressure injuries.
Code Information UDI/DI 00887761977884, Serial Numbers: W330BP6363, W333BP6439, W334BP6477, W334BP6535, W337BP6631, W341BP6675, X047BP7353, X049BP7431, W330BP6365, W333BP6440, W334BP6478, W334BP6536, W337BP6632, W341BP6676, X047BP7354, X049BP7432, W330BP6366, W333BP6442, W334BP6480, W334BP6537, W337BP6633, W341BP6678, X047BP7355, X049BP7433, W330BP6367, W333BP6445, W334BP6481, W337BP6565, W337BP6634, W341BP6679, X047BP7356, X063BP7505, W330BP6368, W333BP6446, W334BP6485, W337BP6566, W337BP6635, W341BP6681, X047BP7357, X063BP7506, W330BP6382, W333BP6447, W334BP6486, W337BP6567, W337BP6636, W341BP6683, X047BP7358, X063BP7507, W330BP6383, W333BP6448, W334BP6487, W337BP6568, W337BP6637, W341BP6684, X047BP7359, X063BP7508, W330BP6385, W333BP6450, W334BP6489, W337BP6569, W337BP6638, W341BP6687, X047BP7360, X063BP7509, W330BP6387, W333BP6451, W334BP6492, W337BP6570, W337BP6641, X046BP7330, X047BP7361, X063BP7510, W330BP6397, W333BP6452, W334BP6493, W337BP6571, W337BP6642, X046BP7331, X047BP7362, X063BP7511, W333BP6399, W333BP6453, W334BP6496, W337BP6572, W337BP6643, X046BP7332, X047BP7363, X063BP7512, W333BP6400, W333BP6455, W334BP6500, W337BP6573, W337BP6644, X046BP7333, X047BP7364, X063BP7513, W333BP6401, W333BP6456, W334BP6501, W337BP6574, W337BP6645, X046BP7334, X047BP7365, X063BP7514, W333BP6404, W333BP6457, W334BP6502, W337BP6575, W341BP6646, X046BP7335, X047BP7366, X066BP7516, W333BP6405, W333BP6458, W334BP6504, W337BP6577, W341BP6647, X046BP7336, X047BP7367, X066BP7517, W333BP6406, W333BP6459, W334BP6505, W337BP6578, W341BP6648, X046BP7337, X047BP7368, X066BP7518, W333BP6415, W333BP6460, W334BP6511, W337BP6579, W341BP6655, X046BP7338, X047BP7369, X066BP7519, W333BP6418, W333BP6461, W334BP6513, W337BP6583, W341BP6658, X046BP7339, X047BP7370, X066BP7520, W333BP6420, W333BP6463, W334BP6514, W337BP6614, W341BP6659, X046BP7340, X047BP7371, X066BP7521, W333BP6421, W333BP6465, W334BP6525, W337BP6615, W341BP6660, X047BP7342, X047BP7372, X066BP7522, W333BP6422, W333BP6466, W334BP6526, W337BP6621, W341BP6661, X047BP7343, X047BP7373, X066BP7523, W333BP6423, W333BP6467, W334BP6527, W337BP6622, W341BP6663, X047BP7344, X047BP7374, X066BP7524, W333BP6424, W334BP6468, W334BP6528, W337BP6623, W341BP6666, X047BP7345, X047BP7375, X066BP7525, W333BP6425, W334BP6469, W334BP6529, W337BP6624, W341BP6667, X047BP7346, X047BP7376, X066BP7526, W333BP6426, W334BP6470, W334BP6530, W337BP6625, W341BP6668, X047BP7347, X047BP7378, X066BP7527, W333BP6427, W334BP6471, W334BP6531, W337BP6626, W341BP6669, X047BP7348, X047BP7379, X066BP7531, W333BP6431, W334BP6473, W334BP6532, W337BP6628, W341BP6670, X047BP7350, X047BP7381, X066BP7532, W333BP6432, W334BP6475, W334BP6533, W337BP6629, W341BP6671, X047BP7351, X049BP7429, X066BP7533, W333BP6435, W334BP6476, W334BP6534, W337BP6630, W341BP6674, X047BP7352, X049BP7430, X066BP7540
FEI Number 1417572
Recalling Firm/
Manufacturer
Baxter Healthcare Corporation
1 Baxter Pkwy
Deerfield IL 60015-4625
For Additional Information ContactCenter for One Baxter
800-422-9837
Manufacturer Reason
for Recall
Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as potentially impacted by these service inconsistencies and therefore, Baxter will need to reassess the impacted mattresses to confirm they have been corrected as identified by the service order, and to confirm the correct serial numbers have been identified.
FDA Determined
Cause 2
Under Investigation by firm
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/05/2024 via USPS First Class Mail. The notice explained the issue, hazard involved, and requested the consignee inspect the device to check for any indication of error present. If an error is detected, contact Baxter Technical Support by phone at 800-445-3720, available Monday through Thursday, between 8:00 am and 6:30 pm Eastern Time, and Friday, between 8:00 am and 6:00 pm Eastern Time, or by email at HRC_Technical_Support@baxter.com to have the pro+ mattress corrected. A Baxter representative will contact your facility to arrange for an inspection of your device and correction, if required.
Quantity in Commerce232 units
DistributionUS Nationwide distribution in the state of Colorado.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-