| | Class 2 Device Recall MONOCRYL (poliglecaprone 25) Suture |  |
| Date Initiated by Firm | May 10, 2024 |
| Date Posted | June 14, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2148-2024 |
| Recall Event ID |
94643 |
| 510(K)Number | K960653 K964072 |
| Product Classification |
Suture, absorbable, synthetic - Product Code GAN
|
| Product | MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated
PRODUCT CODE:
Y315H
Y345H
Y359H
Y398H
Y416H
Y417H
Y426H
Y427H
Y935H
Y936H
Y945H |
| Code Information |
GTIN:
GTIN
10705031058729
10705031058897
10705031059047
10705031059092
10705031059122
30705031059133
10705031112674
10705031059146
10705031059153
10705031059245
10705031059894
10705031059917
10705031059962
PRODUCT CODE PRODUCT LOTS
Y315H UAMQCX
Y345H UBMCAJ
Y359H UBMHQS
Y398H UAMLMU UAMPJB UAMMXQ
UBMJPB UBMQQR
Y416H UAMLQM UAMLKM UAMRMZ
Y417H UCMEDU
Y426H UAMKCL UAMKHE UAMPQS
UBMJTE
Y427H UAMLXX UAMMJR UAMPSC
Y935H UBMCAA UBMDKJ UBMDKC
UBMHDD
Y936H UBMJQM
Y945H UAMKCD
|
Recalling Firm/ Manufacturer |
Ethicon, Inc. 1000 Route 202 Raritan NJ 08869-1425
|
Manufacturer Reason for Recall | Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection . |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Erhicon issued Urgent: Medical Device Recall (Removal) letter on May 10, 2024 to US Distributors, primary and secondary consignees via UPS second day.
Ethicon Johnson & Johnson affiliates in EMEA, LATAM, APAC and Canada will be contacting all their impacted primary and secondary consignees and/or end customers.
Letter states reason for recall, health risk and action to take:
1.
Determine whether you have inventory of the lots listed in Attachment 1. If so, you may
"
Option 1: Remove the first package of suture from a complete box,
i.
Quarantine and return it for credit. You may use the remaining 35 sutures in accordance with the IFU. (See Figure 2 below for location of first package)
OR
ii.
Examine the first package of a complete box prior to use. Quarantine and return this unit if a packaging defect is identified. (See Figure 2 below for location of first package). If no defect is found, the product and remaining 35 sutures in the box may be used accordance with the IFU. OR
"Option 2: Quarantine all product in scope and return all inventory for credit.
Please maintain a copy of this notice with the quarantined product and keep a copy for your records.
2.
Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating.
3.
Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 800-901-0335 or ethicon8329@sedgwick.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall.
4.
Customers are required to return unused sutures with holes in the primary packaging subject to this recall that are in inventory immediately. To receive credit for items returned, customers must return product no |
| Quantity in Commerce | 306000 units |
| Distribution | Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = GAN
|
|
|
|