| | Class 2 Device Recall SafeControl Handset |  |
| Date Initiated by Firm | July 11, 2022 |
| Date Posted | November 08, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0329-2025 |
| Recall Event ID |
95490 |
| Product Classification |
Bed, therapeutic, ac-powered, adjustable home-use - Product Code LLI
|
| Product | SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198 |
| Code Information |
UDI/DI 4050621100401, 19-0990
UDI/DI 4050621100012, 19-1062
UDI/DI 4050621100432, 19-1152
UDI/DI 4050621100449, 19-1119
UDI/DI 4050621100470, 19-1155
UDI/DI 4050621100487, 19-1156
|
| FEI Number |
3015979371
|
Recalling Firm/ Manufacturer |
WISSNER-BOSSERHOFF GMBH
Hauptstr. 4-6
Wickede (Ruhr)
Germany
|
Manufacturer Reason for Recall | When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved. |
FDA Determined Cause 2 | Device Design |
| Action | Wissner-bosserhoff issued an Urgent safety information notice along with supplement use instructions notice to its sole US consignee on 07/11/2022. The notice explained the problem, risk to patient, and requested the user update the IFU for the device. Distributors were directed to notify their customers. |
| Quantity in Commerce | 77 units |
| Distribution | US Nationwide distribution in the states of NC. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|