| Date Initiated by Firm | November 21, 2024 |
| Date Posted | December 27, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0774-2025 |
| Recall Event ID |
95845 |
| Product Classification |
Pump, infusion, implanted, programmable - Product Code LKK
|
| Product | SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900E CLINICIAN TABLET GLBL SMSNG and PROG CT900E CLINICIAN TABLET PAPER EN with the A810 CP app version 2.x, CFN CT900E |
| Code Information |
GTIN: 00763000447090
00763000729141
00763000737290
00763000447090;
Serial Numbers: NPL3020725
NPL3021958
NPL3021963
NPL3021966
NPL3021969
NPL3021970
NPL3021971
NPL3021974
NPL3021983
NPL3021984
NPL3021990
NPL3021997
NPL3022002
NPL3022005
NPL3022014
NPL3022022
NPL3022031
NPL3022032
NPL3022038
NPL3022044
NPL3022052
NPL3022062
NPL3022069
NPL3022070
NPL3022071
NPL3022072
NPL3022073
NPL3022074
NPL3022109
NPL3022114
NPL3022118
NPL3022124
NPL3022125
NPL3022131
NPL3022139
NPL3022187
NPL3022190
NPL3022193
NPL3022194
NPL3022195
NPL3022212
NPL3022213
NPL3022216
NPL3022218
NPL3022245
NPL3022257
NPL3022278
NPL3022280
NPL3022289
NPL3022291
NPL3022304
NPL3022311
NPL3022313
NPL3022314
NPL3022328
NPL3022352
NPL3022357
NPL3022364
NPL3022365
NPL3022711
NPL3022713
NPL3022714
NPL3022715
NPL3022717
NPL3022735
NPL3022761
NPL3022762
NPL3022765
NPL3022767
NPL3022768
NPL3022769
NPL3022776
NPL3022791
NPL3022793
NPL3022795
NPL3022810
NPL3022811
NPL3022815
NPL3022817
NPL3022818
NPL3022831
NPL3022836
NPL3022838
NPL3022841
NPL3022844
NPL3022856
NPL3022858
NPL3022869
NPL3022870
NPL3022873
NPL3022874
NPL3022876
NPL3022877
NPL3022879
NPL3022893
NPL3022899
NPL3022900
NPL3022912
NPL3022918
NPL3022922
NPL3022926
NPL3022939
NPL3022940
NPL3022942
NPL3022954
NPL3022957
NPL3022958
NPL3022959
NPL3022968
NPL3022969
NPL3022973
NPL3022974
NPL3022981
NPL3022999
NPL3023030
NPL3023032
NPL3023035
NPL3023038
NPL3023042
NPL3023043
NPL3023048
NPL3023050
NPL3023054
NPL3023059
NPL3023061
NPL3023065
NPL3023067
NPL3023076
NPL3023078
NPL3023086
NPL3023093
NPL3023097
NPL3023123
NPL3023131
NPL3023132
NPL3023133
NPL3023136
NPL3023146
NPL3023148
NPL3023149
NPL3023150
NPL3023153
NPL3023156
NPL3023167
NPL3023175
NPL3023180
NPL3023192
NPL3023194
NPL3023197
NPL3023205
NPL3023218
NPL3023229
NPL3023237
NPL3023239
NPL3023241
NPL3023242
NPL3023245
NPL3023254
NPL3023256
NPL3023257
NPL3023260
NPL3023262
NPL3023268
NPL3023272
NPL3023278
NPL3023288
NPL3023292
NPL3023315
NPL3023317
NPL3023319
NPL3023322
NPL3023323
NPL3023330
NPL3023331
NPL3023340
NPL3023347
NPL3023351
NPL3023352
NPL3023364
NPL3023367
NPL3023368
NPL3023373
NPL3023380
NPL3023388
NPL3023389
NPL3023394
NPL3023409
NPL3023411
NPL3023415
NPL3023422
NPL3023423
NPL3023424
NPL3023425
NPL3023429
NPL3023432
NPL3023435
NPL3023437
NPL3023439
NPL3023440
NPL3023443
NPL3023446
NPL3023447
NPL3023448
NPL3023449
NPL3023450
NPL3023459
NPL3023460
NPL3023463
NPL3023464
NPL3023466
NPL3023467
NPL3023474
NPL3023491
NPL3023494
NPL3023498
NPL3023504
NPL3023510
NPL3023511
NPL3023512
NPL3023520
NPL3023526
NPL3023541
NPL3023548
NPL3023550
NPL3023554
NPL3023555
NPL3023556
NPL3023559
NPL3023563
NPL3023573
NPL3023583
NPL3023593
NPL3023597
NPL3023598
NPL3023599
NPL3023600
NPL3023602
NPL3023607
NPL3023610
NPL3023615
NPL3023620
NPL3023635
NPL3023638
NPL3023640
NPL3023642
NPL3023646
NPL3023647
NPL3023649
NPL3023651
NPL3023659
NPL3023660
NPL3023661
NPL3023664
NPL3023665
NPL3023666
NPL3023689
NPL3023690
NPL3023692
NPL3023695
NPL3023704
NPL3023708
NPL3023711
NPL3023716
NPL3023718
NPL3023722
NPL3023723
NPL3023728
NPL3023729
NPL3023734
NPL3023744
NPL3023751
NPL3023752
NPL3023760
NPL3023762
NPL3023776
NPL3023778
NPL3023781
NPL3023782
NPL3023784
NPL3023785
NPL3023792
NPL3023803
NPL3023809
NPL3023829
NPL3023830
NPL3023831
NPL3023837
NPL3023841
NPL3023852
NPL3023854
NPL3023855
NPL3023857
NPL3023858
NPL3023860
NPL3023868
NPL3023873
NPL3023885
NPL3023902
NPL3023905
NPL3023915
NPL3023922
NPL3023930
NPL3023933
NPL3023936
NPL3023938
NPL3023948
NPL3023949
NPL3023955
NPL3023957
NPL3023959
NPL3023974
NPL3023977
NPL3023980
NPL3023983
NPL3023984
NPL3023990
NPL3023991
NPL3023998
NPL3024013
NPL3024015
NPL3024022
NPL3024023
NPL3024024
NPL3024025
NPL3024026
NPL3024028
NPL3024031
NPL3024033
NPL3024037
NPL3024040
NPL3024043
NPL3024044
NPL3024046
NPL3024054
NPL3024055
NPL3024056
NPL3024059
NPL3024072
NPL3024079
NPL3024082
NPL3024084
NPL3024087
NPL3024088
NPL3024092
NPL3024099
NPL3024100
NPL3024102
NPL3024103
NPL3024108
NPL3024111
NPL3024113
NPL3024117
NPL3024122
NPL3024123
NPL3024124
NPL3024127
NPL3024129
NPL3024139
NPL3024145
NPL3024148
NPL3024149
NPL3024150
NPL3024151
NPL3024152
NPL3024153
NPL3024155
NPL3024158
NPL3024161
NPL3024162
NPL3024166
NPL3024167
NPL3024170
NPL3024176
NPL3024179
NPL3024185
NPL3024188
NPL3024193
NPL3024196
NPL3024197
NPL3024198
NPL3024199
NPL3024200
NPL3024206
NPL3024207
NPL3024208
NPL3024209
NPL3024210
NPL3024212
NPL3024218
NPL3024220
NPL3024221
NPL3024225
NPL3024230
NPL3024233
NPL3024234
NPL3024235
NPL3024239
NPL3024240
NPL3024245
NPL3024246
NPL3024247
NPL3024256
NPL3024257
NPL3024258
NPL3024261
NPL3024266
NPL3024274
NPL3024276
NPL3024277
NPL3024278
NPL3024280
NPL3024284
NPL3024294
NPL3024296
NPL3024298
NPL3024299
NPL3024300
NPL3024303
NPL3024305
NPL3024307
NPL3024308
NPL3024309
NPL3024310
NPL3024311
NPL3024312
NPL3024313
NPL3024316
NPL3024321
NPL3024324
NPL3024327
NPL3024328
NPL3024329
NPL3024337
NPL3024341
NPL3024348
NPL3024350
NPL3024355
NPL3024356
NPL3024358
NPL3024360
NPL3024367
NPL3024370
NPL3024379
NPL3024381
NPL3024384
NPL3024393
NPL3024395
NPL3024399
NPL3024404
NPL3024405
NPL3024407
NPL3024408
NPL3024416
NPL3024417
NPL3024418
NPL3024419
NPL3024420
NPL3024424
NPL3024428
NPL3024430
NPL3024437
NPL3024439
NPL3024440
NPL3024442
NPL3024452
NPL3024455
NPL3024461
NPL3024462
NPL3024464
NPL3024472
NPL3024478
NPL3024479
NPL3024480
NPL3024482
NPL3024484
NPL3024485
NPL3024486
NPL3024487
NPL3024490
NPL3024491
NPL3024492
NPL3024493
NPL3024498
NPL3024499
NPL3024501
NPL3024502
NPL3024507
NPL3024509
NPL3024518
NPL3024519
NPL3024523
NPL3024531
NPL3024533
NPL3024534
NPL3024540
NPL3024541
NPL3024542
NPL3024544
NPL3024545
NPL3024549
NPL3024550
NPL3024551
NPL3024552
NPL3024554
NPL3024555
NPL3024556
NPL3024557
NPL3024558
NPL3024559
NPL3024563
NPL3024565
NPL3024567
NPL3024569
NPL3024572
NPL3024574
NPL3024576
NPL3024578
NPL3024582
NPL3024583
NPL3024585
NPL3024587
NPL3024589
NPL3024592
NPL3024594
NPL3024596
NPL3024598
NPL3024600
NPL3024602
NPL3024603
NPL3024604
NPL3024606
NPL3024608
NPL3024610
NPL3024615
NPL3024618
NPL3024619
NPL3024621
NPL3024625
NPL3024626
NPL3024631
NPL3024632
NPL3024635
NPL3024637
NPL3024639
NPL3024642
NPL3024643
NPL3024647
NPL3024649
NPL3024654
NPL3024656
NPL3024657
NPL3024658
NPL3024659
NPL3024660
NPL3024661
NPL3024662
NPL3024664
NPL3024665
NPL3024669
NPL3024671
NPL3024673
NPL3024676
NPL3024678
NPL3024680
NPL3024688
NPL3024694
NPL3024702
NPL3024709
NPL3024712
NPL3024717
NPL3024721
NPL3024722
NPL3024726
NPL3024728
NPL3024729
NPL3024730
NPL3024731
NPL3024734
NPL3024736
NPL3024737
NPL3024739
NPL3024740
NPL3024748
NPL3024750
NPL3024751
NPL3024755
NPL3024758
NPL3024759
NPL3024764
NPL3024765
NPL3024769
NPL3024771
NPL3024772
NPL3024774
NPL3024775
NPL3024777
NPL3024778
NPL3024780
NPL3024782
NPL3024784
NPL3024785
NPL3024786
NPL3024787
NPL3024794
NPL3024797
NPL3024799
NPL3024800
NPL3024803
NPL3024807
NPL3024808
NPL3024817
NPL3024819
NPL3024821
NPL3024823
NPL3024827
NPL3024834
NPL3024838
NPL3024843
NPL3024851
NPL3024857
NPL3024858
NPL3024859
NPL3024861
NPL3024867
NPL3024869
NPL3024870
NPL3024872
NPL3024875
NPL3024884
NPL3024890
NPL3024896
NPL3024897
NPL3024899
NPL3024903
NPL3024905
NPL3024906
NPL3024909
NPL3024919
NPL3024921
NPL3024929
NPL3024933
NPL3024934
NPL3024935
NPL3024936
NPL3024939
NPL3024940
NPL3024943
NPL3024945
NPL3024950
NPL3024951
NPL3024953
NPL3024954
NPL3024956
NPL3024964
NPL3024970
NPL3024971
NPL3024972
NPL3024975
NPL3024980
NPL3024982
NPL3024983
NPL3024985
NPL3024988
NPL3024993
NPL3024995
NPL3024997
NPL3024998
NPL3024999
NPL3025002
NPL3025011
NPL3025021
NPL3025024
NPL3025025
NPL3025026
NPL3025047
NPL3025048
NPL3025057
NPL3025058
NPL3025060
NPL3025062
NPL3025068
NPL3025070
NPL3025074
NPL3025084
NPL3025090
NPL3025097
NPL3025103
NPL3025105
NPL3025117
NPL3025119
NPL3025132
NPL3025133
NPL3025146
NPL3025153
NPL3025158
NPL3025160
NPL3025164
NPL3025167
NPL3025169
NPL3025179
NPL3025180
NPL3025185
NPL3025191
NPL3025197
NPL3025207
NPL3025217
NPL3025224
NPL3025227
NPL3025228
NPL3025231
NPL3025234
NPL3025237
NPL3025240
NPL3025241
NPL3025244
NPL3025249
NPL3025252
NPL3025253
NPL3025257
NPL3025263
NPL3025265
NPL3025270
NPL3025271
NPL3025275
NPL3025281
NPL3025289
NPL3025300
NPL3025302
NPL3025304
NPL3025305
NPL3025309
NPL3025312
NPL3025313
NPL3025320
NPL3025322
NPL3025327
NPL3025329
NPL3025331
NPL3025332
NPL3025335
NPL3025338
NPL3025339
NPL3025344
NPL3025354
NPL3025356
NPL3025370
NPL3025382
NPL3025385
NPL3025386
NPL3025388
NPL3025392
NPL3025394
NPL3025397
NPL3025409
NPL3025410
NPL3025419
NPL3025424
NPL3025433
NPL3025436
NPL3025439
NPL3025441
NPL3025446
NPL3025448
NPL3025449
NPL3025451
NPL3025452
NPL3025454
NPL3025458
NPL3025461
NPL3025462
NPL3025465
NPL3025466
NPL3025468
NPL3025470
NPL3025477
NPL3025479
NPL3025480
NPL3025482
NPL3025483
NPL3025490
NPL3025494
NPL3025502
NPL3025506
NPL3025508
NPL3025509
NPL3025513
NPL3025514
NPL3025515
NPL3025518
NPL3025526
NPL3025527
NPL3025530
NPL3025532
NPL3025533
NPL3025534
NPL3025535
NPL3025538
NPL3025542
NPL3025547
NPL3025549
NPL3025558
NPL3025562
NPL3025563
NPL3025572
NPL3025576
NPL3025577
NPL3025580
NPL3025584
NPL3025588
NPL3025589
NPL3025591
NPL3025595
NPL3025596
NPL3025598
NPL3025600
NPL3025601
NPL3025602
NPL3025606
NPL3025613
NPL3025617
NPL3025620
NPL3025621
NPL3025622
NPL3025624
NPL3025625
NPL3025629
NPL3025633
NPL3025638
NPL3025642
NPL3025644
NPL3025646
NPL3025649
NPL3025650
NPL3025652
NPL3025653
NPL3025661
NPL3025667
NPL3025669
NPL3025673
NPL3025674
NPL3025676
NPL3025677
NPL3025686
NPL3025689
NPL3025690
NPL3025691
NPL3025697
NPL3025703
NPL3025704
NPL3025707
NPL3025712
NPL3025713
NPL3025723
NPL3025725
NPL3025727
NPL3025739
NPL3025740
NPL3025741
NPL3025742
NPL3025745
NPL3025751
NPL3025754
NPL3025755
NPL3025757
NPL3025758
NPL3025775
NPL3025776
NPL3025777
NPL3025787
NPL3025795
NPL3025797
NPL3025799
NPL3025800
NPL3025807
NPL3025811
NPL3025813
NPL3025815
NPL3025816
NPL3025819
NPL3025820
NPL3025823
NPL3025825
NPL3025826
NPL3025827
NPL3025828
NPL3025832
NPL3025834
NPL3025837
NPL3025838
NPL3025839
NPL3025842
NPL3025845
NPL3025849
NPL3025850
NPL3025853
NPL3025855
NPL3025873
NPL3025879
NPL3025881
NPL3025882
NPL3025892
NPL3025893
NPL3025895
NPL3025900
NPL3025901
NPL3025908
NPL3025909
NPL3025916
NPL3025918
NPL3025921
NPL3025929
NPL3025932
NPL3025937
NPL3025938
NPL3025939
NPL3025946
NPL3025947
NPL3025949
NPL3025951
NPL3025953
NPL3025955
NPL3025963
NPL3025964
NPL3025965
NPL3025967
NPL3025975
NPL3025981
NPL3025998
NPL3025999
NPL3026005
NPL3026008
NPL3026009
NPL3026010
NPL3026012
NPL3026017
NPL3026019
NPL3026020
NPL3026021
NPL3026035
NPL3026038
NPL3026044
NPL3026050
NPL3026052
NPL3026053
NPL3026054
NPL3026057
NPL3026060
NPL3026061
NPL3026066
NPL3026068
NPL3026071
NPL3026076
NPL3026077
NPL3026081
NPL3026085
NPL3026088
NPL3026092
NPL3026094
NPL3026097
NPL3026099
NPL3026101
NPL3026102
NPL3026104
NPL3026108
NPL3026109
NPL3026111
NPL3026118
NPL3026120
NPL3026121
NPL3026123
NPL3026127
NPL3026134
NPL3026141
NPL3026142
NPL3026143
NPL3026144
NPL3026147
NPL3026148
NPL3026150
NPL3026151
NPL3026152
NPL3026153
NPL3026154
NPL3026155
NPL3026159
NPL3026161
NPL3026163
NPL3026165
NPL3026168
NPL3026169
NPL3026172
NPL3026173
NPL3026177
NPL3026178
NPL3026183
NPL3026184
NPL3026185
NPL3026187
NPL3026192
NPL3026193
NPL3026195
NPL3026196
NPL3026197
NPL3026198
NPL3026199
NPL3026201
NPL3026202
NPL3026203
NPL3026204
NPL3026206
NPL3026207
NPL3026208
NPL3026209
NPL3026210
NPL3026212
NPL3026215
NPL3026216
NPL3026217
NPL3026218
NPL3026221
NPL3026223
NPL3026224
NPL3026229
NPL3026233
NPL3026238
NPL3026241
NPL3026244
NPL3026245
NPL3026250
NPL3026258
NPL3026259
NPL3026261
NPL3026262
NPL3026263
NPL3026264
NPL3026267
NPL3026270
NPL3026271
NPL3026272
NPL3026273
NPL3026274
NPL3026275
NPL3026286
NPL3026289
NPL3026291
NPL3026292
NPL3026296
NPL3026300
NPL3026301
NPL3026303
NPL3026307
NPL3026313
NPL3026315
NPL3026317
NPL3026319
NPL3026321
NPL3026324
NPL3026325
NPL3026333
NPL3026334
NPL3026335
NPL3026337
NPL3026338
NPL3026343
NPL3026350
NPL3026351
NPL3026352
NPL3026356
NPL3026357
NPL3026358
NPL3026360
NPL3026362
NPL3026367
NPL3026374
NPL3026381
NPL3026382
NPL3026383
NPL3026384
NPL3026390
NPL3026391
NPL3026393
NPL3026396
NPL3026397
NPL3026398
NPL3026402
NPL3026405
NPL3026406
NPL3026407
NPL3026408
NPL3026414
NPL3026421
NPL3026422
NPL3026423
NPL3026425
NPL3026427
NPL3026428
NPL3026430
NPL3026432
NPL3026435
NPL3026438
NPL3026440
NPL3026441
NPL3026442
NPL3026444
NPL3026445
NPL3026448
NPL3026451
NPL3026452
NPL3027458
NPL3027462
NPL3027463
NPL3027464
NPL3027465
NPL3027467
NPL3027477
NPL3027480
NPL3027482
NPL3027487
NPL3027496
NPL3027516
NPL3027518
NPL3027520
NPL3027522
NPL3027526
NPL3027535
NPL3027538
NPL3027545
NPL3027549
NPL3027571
NPL3027592
NPL3027595
NPL3027601
NPL3027604
NPL3027606
NPL3027607
NPL3027610
NPL3027611
NPL3027614
NPL3027620
NPL3027627
NPL3027631
NPL3027634
NPL3027635
NPL3027636
NPL3027651
NPL3027652
NPL3027656
NPL3027658
NPL3027661
NPL3027664
NPL3027674
NPL3027676
NPL3027678
NPL3027680
NPL3027681
NPL3027682
NPL3027684
NPL3027685
NPL3027686
NPL3027693
NPL3027695
NPL3027710
NPL3027712
NPL3027714
NPL3027716
NPL3027726
NPL3027727
NPL3027731
NPL3027732
NPL3027733
NPL3027734
NPL3027736
NPL3027737
NPL3027745
NPL3027753
NPL3027760
NPL3027767
NPL3027770
NPL3027773
NPL3027774
NPL3027777
NPL3027779
NPL3027780
NPL3027787
NPL3027790
NPL3027791
NPL3027796
NPL3027801
NPL3027803
NPL3027810
NPL3027816
NPL3027820
NPL3027823
NPL3027829
NPL3027830
NPL3027832
NPL3027834
NPL3027836
NPL3027839
NPL3027840
NPL3027841
NPL3027843
NPL3027844
NPL3027846
NPL3027848
NPL3027850
NPL3027851
NPL3027852
NPL3027853
NPL3027856
NPL3027857
NPL3027863
NPL3027864
NPL3027865
NPL3030786
NPL3030794
NPL3030796
NPL3030799
NPL3030806
NPL3030807
NPL3030809
NPL3030811
NPL3030814
NPL3030817
NPL3030820
NPL3030824
NPL3030825
NPL3030836
NPL3030846
NPL3030848
NPL3030860
NPL3030865
NPL3030867
NPL3030869
NPL3030870
NPL3030873
NPL3030875
NPL3030879
NPL3030880
NPL3030882
NPL3030884
NPL3030889
NPL3030893
NPL3030900
NPL3030901
NPL3030902
NPL3030904
NPL3030905
NPL3030908
NPL3030913
NPL3030916
NPL3030917
NPL3030926
NPL3030929
NPL3030942
NPL3030943
NPL3030950
NPL3030951
NPL3030954
NPL3030955
NPL3030962
NPL3030964
NPL3030971
NPL3030978
NPL3030979
NPL3030983
NPL3030987
NPL3030989
NPL3030990
NPL3030992
NPL3030996
NPL3031005
NPL3031010
NPL3031011
NPL3031012
NPL3031013
NPL3031014
NPL3031016
NPL3031018
NPL3031019
NPL3031022
NPL3031024
NPL3031026
NPL3031027
NPL3031030
NPL3031031
NPL3031032
NPL3031035
NPL3031039
NPL3031041
NPL3031042
NPL3031043
NPL3031047
NPL3031053
NPL3031056
NPL3031058
NPL3031060
NPL3031061
NPL3031062
NPL3031063
NPL3031074
NPL3031075
NPL3031079
NPL3031080
NPL3031081
NPL3031082
NPL3031083
NPL3031084
NPL3031085
NPL3031087
NPL3031088
NPL3031089
NPL3031090
NPL3031091
NPL3031092
NPL3031095
NPL3031096
NPL3031097
NPL3031099
NPL3031100
NPL3031101
NPL3031102
NPL3031103
NPL3031109
NPL3031113
NPL3031117
NPL3031118
NPL3031122
NPL3031123
NPL3031124
NPL3031127
NPL3031128
NPL3031134
NPL3031135
NPL3031136
NPL3031137
NPL3031141
NPL3031144
NPL3031148
NPL3031150
NPL3031151
NPL3031152
NPL3031153
NPL3031154
NPL3031156
NPL3031158
NPL3031159
NPL3031160
NPL3031161
NPL3031165
NPL3031166
NPL3031171
NPL3031172
NPL3031177
NPL3031178
NPL3031180
NPL3031183
NPL3031187
NPL3031188
NPL3031190
NPL3031191
NPL3031193
NPL3031194
NPL3031197
NPL3031198
NPL3031200
NPL3031201
NPL3031209
NPL3031212
NPL3031215
NPL3031220
NPL3031222
NPL3031225
NPL3031227
NPL3031228
NPL3031231
NPL3031232
NPL3031233
NPL3031239
NPL3031240
NPL3031241
NPL3031242
NPL3031244
NPL3031245
NPL3031247
NPL3031248
NPL3031250
NPL3031251
NPL3031252
NPL3031257
NPL3031258
NPL3031259
NPL3031261
NPL3031265
NPL3031266
NPL3031267
NPL3031273
NPL3031277
NPL3031278
NPL3031279
NPL3031281
NPL3031282
NPL3031283
NPL3031284
NPL3031290
NPL3031292
NPL3031294
NPL3031296
NPL3031300
NPL3031305
NPL3031306
NPL3031307
NPL3031308
NPL3031309
NPL3031310
NPL3031311
NPL3031312
NPL3031313
NPL3031314
NPL3031315
NPL3031317
NPL3031320
NPL3031321
NPL3031323
NPL3031331
NPL3031334
NPL3031336
NPL3031339
NPL3031349
NPL3031352
NPL3031354
NPL3031357
NPL3031358
NPL3031359
NPL3031361
NPL3031362
NPL3031363
NPL3031366
NPL3031367
NPL3031368
NPL3031369
NPL3031371
NPL3031373
NPL3031374
NPL3031375
NPL3031377
NPL3031378
NPL3031379
NPL3031380
NPL3031383
NPL3031387
NPL3031388
NPL3031393
NPL3031395
NPL3031396
NPL3031397
NPL3031402
NPL3031403
NPL3031404
NPL3031405
NPL3031407
NPL3031408
NPL3031413
NPL3031416
NPL3031419
NPL3031420
NPL3031421
NPL3031423
NPL3031428
NPL3031429
NPL3031430
NPL3031431
NPL3031432
NPL3031433
NPL3031437
NPL3031444
NPL3031447
NPL3031448
NPL3031451
NPL3031452
NPL3031453
NPL3031456
NPL3031459
NPL3031461
NPL3031463
NPL3031466
NPL3031467
NPL3031471
NPL3031472
NPL3031473
NPL3031475
NPL3031477
NPL3031478
NPL3031482
NPL3031483
NPL3031487
NPL3031488
NPL3031491
NPL3031495
NPL3031496
NPL3031500
NPL3031501
NPL3031504
NPL3031508
NPL3031509
NPL3031511
NPL3031514
NPL3031515
NPL3031518
NPL3031521
NPL3031522
NPL3031525
NPL3031529
NPL3031531
NPL3031534
NPL3031537
NPL3031538
NPL3031540
NPL3031544
NPL3031546
NPL3031547
NPL3031548
NPL3031550
NPL3031551
NPL3031555
NPL3031558
NPL3031562
NPL3031563
NPL3031566
NPL3031569
NPL3031570
NPL3031575
NPL3031577
NPL3031582
NPL3031585
NPL3031587
NPL3031591
NPL3031592
NPL3031593
NPL3031594
NPL3031595
NPL3031596
NPL3031597
NPL3031600
NPL3031602
NPL3031603
NPL3031604
NPL3031606
NPL3031608
NPL3031609
NPL3031612
NPL3031614
NPL3031620
NPL3031624
NPL3031627
NPL3031632
NPL3031635
NPL3031638
NPL3031640
NPL3031641
NPL3031644
NPL3031646
NPL3031648
NPL3031649
NPL3031650
NPL3031651
NPL3031653
NPL3031654
NPL3031657
NPL3031658
NPL3031659
NPL3031661
NPL3031665
NPL3031670
NPL3031671
NPL3031672
NPL3031674
NPL3031676
NPL3031679
NPL3031680
NPL3031687
NPL3031690
NPL3031693
NPL3031694
NPL3031695
NPL3031696
NPL3031698
NPL3031700
NPL3031701
NPL3031702
NPL3031705
NPL3031708
NPL3031709
NPL3031710
NPL3031713
NPL3031718
NPL3031725
NPL3031727
NPL3031735
NPL3031737
NPL3031740
NPL3031742
NPL3031743
NPL3031744
NPL3031746
NPL3031750
NPL3031758
NPL3031759
NPL3031760
NPL3031765
NPL3031767
NPL3031773
NPL3031775
NPL3031777
NPL3031782
NPL3031786
NPL3031790
NPL3031792
NPL3031795
NPL3031796
NPL3031805
NPL3031807
NPL3031809
NPL3031815
NPL3031816
NPL3031817
NPL3031819
NPL3031823
NPL3031824
NPL3031826
NPL3031830
NPL3031832
NPL3031840
NPL3031848
NPL3031849
NPL3031852
NPL3031859
NPL3031863
NPL3031864
NPL3031865
NPL3031868
NPL3031870
NPL3031872
NPL3031873
NPL3031877
NPL3031880
NPL3031881
NPL3031883
NPL3031884
NPL3031895
NPL3031900
NPL3031902
NPL3031905
NPL3031907
NPL3031911
NPL3031912
NPL3031913
NPL3031914
NPL3031920
NPL3031923
NPL3031927
NPL3031930
NPL3031931
NPL3031932
NPL3031938
NPL3031947
NPL3031950
NPL3031951
NPL3031952
NPL3031953
NPL3031954
NPL3031957
NPL3031964
NPL3031968
NPL3031971
NPL3031975
NPL3031979
NPL3031986
NPL3031987
NPL3031988
NPL3031993
NPL3031994
NPL3031995
NPL3032001
NPL3032005
NPL3032007
NPL3032009
NPL3032015
NPL3032017
NPL3032018
NPL3032019
NPL3032021
NPL3032023
NPL3032024
NPL3032025
NPL3032028
NPL3032031
NPL3032033
NPL3032035
NPL3032036
NPL3032037
NPL3032038
NPL3032041
NPL3032046
NPL3032049
NPL3032051
NPL3032054
NPL3032056
NPL3032058
NPL3032061
NPL3032062
NPL3032067
NPL3032068
NPL3032069
NPL3032072
NPL3032074
NPL3032076
NPL3032077
NPL3032078
NPL3032079
NPL3032081
NPL3032085
NPL3032087
NPL3032088
NPL3032089
NPL3032094
NPL3032095
NPL3032104
NPL3032106
NPL3032107
NPL3032109
NPL3032110
NPL3032111
NPL3032112
NPL3032117
NPL3032118
NPL3032120
NPL3032122
NPL3032123
NPL3032124
NPL3032135
NPL3032136
NPL3032143
NPL3032144
NPL3032145
NPL3032146
NPL3032150
NPL3032151
NPL3032153
NPL3032154
NPL3032163
NPL3032165
NPL3032167
NPL3032168
NPL3032173
NPL3032180
NPL3032183
NPL3032187
NPL3032190
NPL3032192
NPL3032193
NPL3032195
NPL3032200
NPL3032201
NPL3032202
NPL3032206
NPL3032207
NPL3032210
NPL3032212
NPL3032214
NPL3032215
NPL3032217
NPL3032218
NPL3032219
NPL3032223
NPL3032227
NPL3032229
NPL3032231
NPL3032232
NPL3032236
NPL3032237
NPL3032238
NPL3032239
NPL3032241
NPL3032243
NPL3032244
NPL3032246
NPL3032250
NPL3032251
NPL3032256
NPL3032261
NPL3032262
NPL3032263
NPL3032266
NPL3032268
NPL3032273
NPL3032277
NPL3032281
NPL3032283
NPL3032286
NPL3032291
NPL3032296
NPL3032308
NPL3032310
NPL3032311
NPL3032320
NPL3032328
NPL3032330
NPL3032344
NPL3032346
NPL3032347
NPL3032348
NPL3032356
NPL3032360
NPL3032361
NPL3032366
NPL3032373
NPL3032375
NPL3032377
NPL3032378
NPL3032380
NPL3032382
NPL3032383
NPL3032384
NPL3032385
NPL3032387
NPL3032390
NPL3032399
NPL3032401
NPL3032403
NPL3032407
NPL3032408
NPL3032410
NPL3032416
NPL3032418
NPL3032420
NPL3032421
NPL3032424
NPL3032426
NPL3032428
NPL3032435
NPL3032436
NPL3032437
NPL3032440
NPL3032441
NPL3032442
NPL3032450
NPL3032451
NPL3032453
NPL3032454
NPL3032455
NPL3032456
NPL3032457
NPL3032458
NPL3032462
NPL3032463
NPL3032465
NPL3032467
NPL3032470
NPL3032471
NPL3032474
NPL3032482
NPL3032483
NPL3032485
NPL3032486
NPL3032488
NPL3032489
NPL3032491
NPL3032492
NPL3032493
NPL3032495
NPL3032496
NPL3032497
NPL3032500
NPL3032501
NPL3032505
NPL3032507
NPL3032510
NPL3032512
NPL3032516
NPL3032522
NPL3032523
NPL3032530
NPL3032531
NPL3032532
NPL3032536
NPL3032547
NPL3032548
NPL3032549
NPL3032550
NPL3032551
NPL3032556
NPL3032557
NPL3032560
NPL3032562
NPL3032566
NPL3032568
NPL3032572
NPL3032573
NPL3032574
NPL3032577
NPL3032579
NPL3032583
NPL3032587
NPL3032588
NPL3032591
NPL3032592
NPL3032594
NPL3032595
NPL3032597
NPL3032598
NPL3032599
NPL3032601
NPL3032602
NPL3032603
NPL3032609
NPL3032610
NPL3032611
NPL3032612
NPL3032613
NPL3032614
NPL3032616
NPL3032620
NPL3032621
NPL3032622
NPL3032623
NPL3032624
NPL3032626
NPL3032627
NPL3032628
NPL3032629
NPL3032630
NPL3032632
NPL3032633
NPL3032640
NPL3032641
NPL3032642
NPL3032648
NPL3032650
NPL3032651
NPL3032652
NPL3032656
NPL3032660
NPL3032664
NPL3032665
NPL3032666
NPL3032668
NPL3032672
NPL3032681
NPL3032683
NPL3032687
NPL3032688
NPL3032689
NPL3032690
NPL3032692
NPL3032693
NPL3032696
NPL3032697
NPL3032701
NPL3032702
NPL3032703
NPL3032706
NPL3032709
NPL3032714
NPL3032715
NPL3032717
NPL3032718
NPL3032719
NPL3032722
NPL3032724
NPL3032727
NPL3032728
NPL3032729
NPL3032735
NPL3032736
NPL3032739
NPL3032740
NPL3032741
NPL3032742
NPL3032745
NPL3032746
NPL3032747
NPL3032748
NPL3032750
NPL3032756
NPL3032759
NPL3032761
NPL3032762
NPL3032763
NPL3032764
NPL3032770
NPL3032772
NPL3032773
NPL3032774
NPL3032776
NPL3032777
NPL3032778
NPL3032779
NPL3032781
NPL3032782
NPL3032783
NPL3032784
NPL3032790
NPL3032791
NPL3032797
NPL3032798
NPL3032799
NPL3032803
NPL3032806
NPL3032807
NPL3032810
NPL3032814
NPL3032816
NPL3032818
NPL3032819
NPL3032821
NPL3032823
NPL3032824
NPL3032826
NPL3032828
NPL3032833
NPL3032835
NPL3032837
NPL3032838
NPL3032839
NPL3032848
NPL3032850
NPL3032856
NPL3032858
NPL3032860
NPL3032862
NPL3032863
NPL3032867
NPL3032869
NPL3032871
NPL3032872
NPL3032876
NPL3032877
NPL3032878
NPL3032879
NPL3032880
NPL3032882
NPL3032883
NPL3032885
NPL3032888
NPL3032891
NPL3032892
NPL3032893
NPL3032895
NPL3032896
NPL3032900
NPL3032904
NPL3032906
NPL3032907
NPL3032909
NPL3032911
NPL3032915
NPL3032916
NPL3032918
NPL3032920
NPL3032921
NPL3032922
NPL3032926
NPL3032927
NPL3032929
NPL3032931
NPL3032932
NPL3032934
NPL3032938
NPL3032939
NPL3032942
NPL3032944
NPL3032945
NPL3032949
NPL3032952
NPL3032953
NPL3032954
NPL3032959
NPL3032965
NPL3032968
NPL3032970
NPL3032973
NPL3032975
NPL3032978
NPL3032980
NPL3032982
NPL3032983
NPL3032985
NPL3032986
NPL3032989
NPL3032991
NPL3032995
NPL3032996
NPL3032997
NPL3032998
NPL3033001
NPL3033003
NPL3033007
NPL3033008
NPL3033011
NPL3033015
NPL3033017
NPL3033019
NPL3033020
NPL3033022
NPL3033023
NPL3033025
NPL3033026
NPL3033027
NPL3033029
NPL3033030
NPL3033032
NPL3033035
NPL3033036
NPL3033037
NPL3033038
NPL3033039
NPL3033040
NPL3033042
NPL3033043
NPL3033046
NPL3033047
NPL3033048
NPL3033049
NPL3033051
NPL3033052
NPL3033054
NPL3033057
NPL3033059
NPL3033062
NPL3033064
NPL3033066
NPL3033067
NPL3033068
NPL3033069
NPL3033070
NPL3033071
NPL3033072
NPL3033073
NPL3033077
NPL3033079
NPL3033080
NPL3033081
NPL3033087
NPL3033088
NPL3033089
NPL3033094
NPL3033095
NPL3033096
NPL3033100
NPL3033102
NPL3033105
NPL3033106
NPL3033107
NPL3033109
NPL3033111
NPL3033113
NPL3033114
NPL3033115
NPL3033116
NPL3033122
NPL3033123
NPL3033124
NPL3033127
NPL3033130
NPL3033132
NPL3033133
NPL3033136
NPL3033137
NPL3033138
NPL3033139
NPL3033140
NPL3033142
NPL3033144
NPL3033145
NPL3033148
NPL3033150
NPL3033151
NPL3033155
NPL3033157
NPL3033158
NPL3033164
NPL3033165
NPL3033166
NPL3033167
NPL3033169
NPL3033172
NPL3033173
NPL3033174
NPL3033182
NPL3033185
NPL3033186
NPL3033187
NPL3033190
NPL3033191
NPL3033192
NPL3033194
NPL3033196
NPL3033199
NPL3033200
NPL3033201
NPL3033203
NPL3033205
NPL3033206
NPL3033207
NPL3033208
NPL3033210
NPL3033211
NPL3033216
NPL3033217
NPL3033219
NPL3033220
NPL3033221
NPL3033222
NPL3033225
NPL3033227
NPL3033228
NPL3033230
NPL3033233
NPL3033235
NPL3033237
NPL3033238
NPL3033239
NPL3033241
NPL3033242
NPL3033243
NPL3033244
NPL3033245
NPL3033247
NPL3033248
NPL3033249
NPL3033250
NPL3033253
NPL3033254
NPL3033256
NPL3033258
NPL3033259
NPL3033260
NPL3033262
NPL3033265
NPL3033266
NPL3033268
NPL3033270
NPL3033275
NPL3033278
NPL3033279
NPL3033282
NPL3033283
NPL3033284
NPL3033285
NPL3033286
NPL3033289
NPL3033290
NPL3033292
NPL3033293
NPL3033295
NPL3033298
NPL3033300
NPL3033302
NPL3033305
NPL3033306
NPL3033307
NPL3033308
NPL3033311
NPL3033318
NPL3033322
NPL3033323
NPL3033327
NPL3033330
NPL3033334
NPL3033338
NPL3033342
NPL3033345
NPL3033347
NPL3033348
NPL3033349
NPL3033351
NPL3033354
NPL3033357
NPL3033358
NPL3033362
NPL3033363
NPL3033364
NPL3033365
NPL3033368
NPL3033370
NPL3033371
NPL3033372
NPL3033373
NPL3033375
NPL3033376
NPL3033382
NPL3033383
NPL3033384
NPL3033386
NPL3033388
NPL3033390
NPL3033391
NPL3033396
NPL3033401
NPL3033407
NPL3033411
NPL3033415
NPL3033418
NPL3033426
NPL3033428
NPL3033429
NPL3033430
NPL3033432
NPL3033434
NPL3033439
NPL3033440
NPL3033442
NPL3033451
NPL3033463
NPL3033465
NPL3033467
NPL3033469
NPL3033471
NPL3033484
NPL3033488
NPL3033490
NPL3033492
NPL3033493
NPL3033498
NPL3033499
NPL3033500
NPL3033502
NPL3033504
NPL3033506
NPL3033508
NPL3033512
NPL3033513
NPL3033514
NPL3033519
NPL3033521
NPL3033523
NPL3033526
NPL3033528
NPL3033529
NPL3033530
NPL3033531
NPL3033532
NPL3033533
NPL3033535
NPL3033536
NPL3033537
NPL3033542
NPL3033547
NPL3033554
NPL3033557
NPL3033559
NPL3033560
NPL3033566
NPL3033567
NPL3033568
NPL3033573
NPL3033581
NPL3033583
NPL3033585
NPL3033588
NPL3033594
NPL3033598
NPL3033602
NPL3033604
NPL3033606
NPL3033609
NPL3033610
NPL3033611
NPL3033614
NPL3033616
NPL3033619
NPL3033620
NPL3033622
NPL3033623
NPL3033624
NPL3033625
NPL3033628
NPL3033629
NPL3033635
NPL3033638
NPL3033640
NPL3033641
NPL3033644
NPL3033645
NPL3033651
NPL3033653
NPL3033656
NPL3033659
NPL3033661
NPL3033663
NPL3033667
NPL3033673
NPL3033678
NPL3033680
NPL3033681
NPL3033685
NPL3033687
NPL3033696
NPL3033698
NPL3033699
NPL3033700
NPL3033706
NPL3033708
NPL3033716
NPL3033717
NPL3033718
NPL3033719
NPL3033721
NPL3033725
NPL3033726
NPL3033729
NPL3033731
NPL3033732
NPL3033733
NPL3033734
NPL3033735
NPL3033736
NPL3033739
NPL3033741
NPL3033742
NPL3033743
NPL3033746
NPL3033749
NPL3033750
NPL3033751
NPL3033753
NPL3033755
NPL3033756
NPL3033758
NPL3033760
NPL3033761
NPL3033762
NPL3033764
NPL3033765
NPL3033767
NPL3033768
NPL3033772
NPL3033775
NPL3033776
NPL3033777
NPL3033778
NPL3033779
NPL3033785
NPL3033786
NPL3033787
NPL3033790
NPL3033791
NPL3033796
NPL3033800
NPL3033802
NPL3033803
NPL3033805
NPL3033809
NPL3033811
NPL3033812
NPL3033813
NPL3033815
NPL3033816
NPL3033818
NPL3033822
NPL3033823
NPL3033824
NPL3033825
NPL3033828
NPL3033829
NPL3033830
NPL3033834
NPL3033836
NPL3033838
NPL3033839
NPL3033840
NPL3033841
NPL3033843
NPL3033844
NPL3033846
NPL3033849
NPL3033854
NPL3033857
NPL3033858
NPL3033859
NPL3033861
NPL3033862
NPL3033865
NPL3033868
NPL3033869
NPL3033872
NPL3033873
NPL3033874
NPL3033878
NPL3033879
NPL3033882
NPL3033883
NPL3033887
NPL3033888
NPL3033891
NPL3033893
NPL3033899
NPL3033900
NPL3033901
NPL3033902
NPL3033903
NPL3033908
NPL3033909
NPL3033911
NPL3033912
NPL3033913
NPL3033914
NPL3033916
NPL3033917
NPL3033918
NPL3033919
NPL3033922
NPL3033923
NPL3033925
NPL3033926
NPL3033927
NPL3033928
NPL3033930
NPL3033934
NPL3033935
NPL3033937
NPL3033940
NPL3033943
NPL3033944
NPL3033945
NPL3033948
NPL3033949
NPL3033954
NPL3033955
NPL3033956
NPL3033958
NPL3033959
NPL3033960
NPL3033962
NPL3033964
NPL3033966
NPL3033968
NPL3033975
NPL3033977
NPL3033978
NPL3033979
NPL3033980
NPL3033981
NPL3033983
NPL3033984
NPL3033987
NPL3033990
NPL3033991
NPL3033992
NPL3033993
NPL3033994
NPL3033995
NPL3033996
NPL3034000
NPL3034002
NPL3034004
NPL3034005
NPL3034007
NPL3034008
NPL3034009
NPL3034011
NPL3034013
NPL3034017
NPL3034018
NPL3034021
NPL3034022
NPL3034024
NPL3034026
NPL3034030
NPL3034032
NPL3034034
NPL3034035
NPL3034036
NPL3034039
NPL3034040
NPL3034042
NPL3034043
NPL3034044
NPL3034045
NPL3034046
NPL3034047
NPL3034048
NPL3034049
NPL3034050
NPL3034052
NPL3034054
NPL3034055
NPL3034056
NPL3034057
NPL3034058
NPL3034060
NPL3034062
NPL3034064
NPL3034065
NPL3034066
NPL3034067
NPL3034068
NPL3034072
NPL3034074
NPL3034075
NPL3034076
NPL3034078
NPL3034079
NPL3034080
NPL3034081
NPL3034083
NPL3034084
NPL3034085
NPL3034086
NPL3034088
NPL3034089
NPL3034090
NPL3034091
NPL3034092
NPL3034093
NPL3034095
NPL3034096
NPL3034097
NPL3034098
NPL3034100
NPL3034102
NPL3034103
NPL3034104
NPL3034105
NPL3034107
NPL3034108
NPL3034109
NPL3034111
NPL3034112
NPL3034115
NPL3034116
NPL3034119
NPL3034120
NPL3034121
NPL3034123
NPL3034124
NPL3034125
NPL3034126
NPL3034127
NPL3034128
NPL3034130
NPL3034131
NPL3034133
NPL3034134
NPL3034135
NPL3034137
NPL3034138
NPL3034141
NPL3034142
NPL3034143
NPL3034145
NPL3034148
NPL3034151
NPL3034152
NPL3034157
NPL3034158
NPL3034159
NPL3034161
NPL3034163
NPL3034165
NPL3034168
NPL3034169
NPL3034171
NPL3034172
NPL3034173
NPL3034178
NPL3034179
NPL3034181
NPL3034182
NPL3034183
NPL3034185
NPL3034186
NPL3034187
NPL3034188
NPL3034189
NPL3034192
NPL3034194
NPL3034198
NPL3034199
NPL3034200
NPL3034201
NPL3034203
NPL3034205
NPL3034206
NPL3034209
NPL3034210
NPL3034211
NPL3034215
NPL3034216
NPL3034226
NPL3034227
NPL3034229
NPL3034231
NPL3034232
NPL3034233
NPL3034234
NPL3034236
NPL3034237
NPL3034238
NPL3034239
NPL3034240
NPL3034243
NPL3034244
NPL3034247
NPL3034250
NPL3034252
NPL3034253
NPL3034256
NPL3034259
NPL3034260
NPL3034262
NPL3034264
NPL3034266
NPL3034267
NPL3034269
NPL3034270
NPL3034271
NPL3034275
NPL3034276
NPL3034277
NPL3034278
NPL3034279
NPL3034280
NPL3034285
NPL3034286
NPL3034290
NPL3034291
NPL3034292
NPL3034294
NPL3034295
NPL3034296
NPL3034299
NPL3034301
NPL3034303
NPL3034306
NPL3034307
NPL3034308
NPL3034310
NPL3034316
NPL3034317
NPL3034319
NPL3034323
NPL3034324
NPL3034325
NPL3034326
NPL3034330
NPL3034331
NPL3034332
NPL3034334
NPL3034335
NPL3034339
NPL3034341
NPL3034342
NPL3034343
NPL3034345
NPL3034347
NPL3034348
NPL3034351
NPL3034352
NPL3034353
NPL3034354
NPL3034355
NPL3034357
NPL3034359
NPL3034360
NPL3034361
NPL3034363
NPL3034366
NPL3034367
NPL3034368
NPL3034370
NPL3034371
NPL3034372
NPL3034373
NPL3034374
NPL3034378
NPL3034380
NPL3034381
NPL3034382
NPL3034383
NPL3034384
NPL3034385
NPL3034388
NPL3034390
NPL3034391
NPL3034392
NPL3034397
NPL3034400
NPL3034402
NPL3034404
NPL3034405
NPL3034406
NPL3034407
NPL3034418
NPL3034421
NPL3034425
NPL3034427
NPL3034431
NPL3034437
NPL3034440
NPL3034441
NPL3034444
NPL3034449
NPL3034450
NPL3034459
NPL3034460
NPL3034461
NPL3034464
NPL3034465
NPL3034466
NPL3034467
NPL3034468
NPL3034470
NPL3034472
NPL3034474
NPL3034475
NPL3034477
NPL3034478
NPL3034479
NPL3034480
NPL3034483
NPL3034484
NPL3034490
NPL3034492
NPL3034493
NPL3034494
NPL3034495
NPL3034496
NPL3034498
NPL3034500
NPL3034502
NPL3034505
NPL3034506
NPL3034507
NPL3034513
NPL3034519
NPL3034520
NPL3034529
NPL3034534
NPL3034536
NPL3034539
NPL3034543
NPL3034544
NPL3034547
NPL3034548
NPL3034549
NPL3034550
NPL3034552
NPL3034553
NPL3034555
NPL3034556
NPL3034557
NPL3034560
NPL3034563
NPL3034571
NPL3034572
NPL3034574
NPL3034575
NPL3034592
NPL3034595
NPL3034597
NPL3034605
NPL3034608
NPL3034610
NPL3034615
NPL3034616
NPL3034618
NPL3034619
NPL3034634
NPL3034635
NPL3034636
NPL3034638
NPL3034641
NPL3034644
NPL3034648
NPL3034649
NPL3034651
NPL3034652
NPL3034658
NPL3034659
NPL3034661
NPL3034662
NPL3034665
NPL3034669
NPL3034670
NPL3034671
NPL3034673
NPL3034674
NPL3034676
NPL3034678
NPL3034679
NPL3034680
NPL3034681
NPL3034683
NPL3034685
NPL3034686
NPL3034688
NPL3034692
NPL3034693
NPL3034697
NPL3034702
NPL3034703
NPL3034706
NPL3034711
NPL3034714
NPL3034718
NPL3034720
NPL3034722
NPL3034723
NPL3034724
NPL3034728
NPL3034730
NPL3034732
NPL3034734
NPL3034737
NPL3034738
NPL3034740
NPL3034744
NPL3034747
NPL3034748
NPL3034752
NPL3034753
NPL3034757
NPL3034758
NPL3034759
NPL3034761
NPL3034762
NPL3034769
NPL3034772
NPL3034774
NPL3034776
NPL3034777
NPL3034778
NPL3034781
NPL3034785
NPL3036791
NPL3036795
NPL3036798
NPL3036802
NPL3036805
NPL3036806
NPL3036808
NPL3036810
NPL3036811
NPL3036817
NPL3036818
NPL3036819
NPL3036820
NPL3036821
NPL3036822
NPL3036824
NPL3036825
NPL3036827
NPL3036835
NPL3036842
NPL3036843
NPL3036859
NPL3036860
NPL3036863
NPL3036865
NPL3036866
NPL3036867
NPL3036871
NPL3036872
NPL3036875
NPL3036876
NPL3036877
NPL3036883
NPL3036885
NPL3036886
NPL3036889
NPL3036890
NPL3036902
NPL3036905
NPL3036908
NPL3036910
NPL3036916
NPL3036926
NPL3036927
NPL3036932
NPL3036934
NPL3036936
NPL3036940
NPL3036942
NPL3036944
NPL3036945
NPL3036946
NPL3036947
NPL3036948
NPL3036954
NPL3036956
NPL3036957
NPL3036959
NPL3036960
NPL3036962
NPL3036963
NPL3036967
NPL3036968
NPL3036969
NPL3036985
NPL3036991
NPL3036997
NPL3036999
NPL3037014
NPL3037022
NPL3037030
NPL3037038
NPL3037048
NPL3037052
NPL3037061
NPL3037062
NPL3037064
NPL3037066
NPL3037070
NPL3037072
NPL3037083
NPL3037085
NPL3037087
NPL3037088
NPL3037089
NPL3037094
NPL3037107
NPL3037108
NPL3037111
NPL3037112
NPL3037113
NPL3037115
NPL3037118
NPL3037119
NPL3037121
NPL3037123
NPL3037124
NPL3037126
NPL3037127
NPL3037131
NPL3037132
NPL3037136
NPL3037137
NPL3037139
NPL3037141
NPL3037143
NPL3037145
NPL3037146
NPL3037147
NPL3037148
NPL3037149
NPL3037150
NPL3037151
NPL3037152
NPL3037160
NPL3037163
NPL3037165
NPL3037167
NPL3037168
NPL3037171
NPL3037172
NPL3037173
NPL3037174
NPL3037176
NPL3037177
NPL3037178
NPL3037179
NPL3037180
NPL3037181
NPL3037183
NPL3037186
NPL3037188
NPL3037189
NPL3037191
NPL3037194
NPL3037198
NPL3037201
NPL3037203
NPL3037205
NPL3037208
NPL3037211
NPL3037212
NPL3037213
NPL3037215
NPL3037216
NPL3037217
NPL3037219
NPL3037221
NPL3037224
NPL3037225
NPL3037228
NPL3037229
NPL3037230
NPL3037231
NPL3037233
NPL3037237
NPL3037239
NPL3037240
NPL3037241
NPL3037243
NPL3037244
NPL3037246
NPL3037248
NPL3037250
NPL3037251
NPL3037252
NPL3037253
NPL3037254
NPL3037256
NPL3037260
NPL3037262
NPL3037264
NPL3037265
NPL3037266
NPL3037267
NPL3037269
NPL3037270
NPL3037271
NPL3037272
NPL3037277
NPL3037278
NPL3037280
NPL3037281
NPL3037283
NPL3037284
NPL3037285
NPL3037289
NPL3037292
NPL3037293
NPL3037294
NPL3037295
NPL3037297
NPL3037303
NPL3037305
NPL3037306
NPL3037308
NPL3037311
NPL3037319
NPL3037323
NPL3037324
NPL3037326
NPL3037327
NPL3037328
NPL3037329
NPL3037330
NPL3037331
NPL3037334
NPL3037336
NPL3037337
NPL3037340
NPL3037344
NPL3037346
NPL3037349
NPL3037352
NPL3037353
NPL3037355
NPL3037361
NPL3037362
NPL3037363
NPL3037364
NPL3037365
NPL3037366
NPL3037368
NPL3037369
NPL3037370
NPL3037371
NPL3037380
NPL3037381
NPL3037382
NPL3037383
NPL3037384
NPL3037387
NPL3037388
NPL3037389
NPL3037391
NPL3037396
NPL3037397
NPL3037399
NPL3037400
NPL3037402
NPL3037403
NPL3037404
NPL3037405
NPL3037406
NPL3037408
NPL3037411
NPL3037412
NPL3037414
NPL3037418
NPL3037420
NPL3037421
NPL3037425
NPL3037426
NPL3037429
NPL3037430
NPL3037431
NPL3037432
NPL3037433
NPL3037436
NPL3037440
NPL3037441
NPL3037444
NPL3037445
NPL3037446
NPL3037451
NPL3037456
NPL3037457
NPL3037459
NPL3037460
NPL3037462
NPL3037465
NPL3037466
NPL3037469
NPL3037476
NPL3037491
NPL3037493
NPL3037495
NPL3037499
NPL3037501
NPL3037503
NPL3037520
NPL3037526
NPL3037528
NPL3037530
NPL3037531
NPL3037542
NPL3037559
NPL3037563
NPL3037567
NPL3037568
NPL3037569
NPL3037571
NPL3037581
NPL3037586
NPL3037594
NPL3037596
NPL3037598
NPL3037601
NPL3037602
NPL3037603
NPL3037604
NPL3037607
NPL3037611
NPL3037614
NPL3037616
NPL3037617
NPL3037618
NPL3037620
NPL3037621
NPL3037624
NPL3037628
NPL3037629
NPL3037630
NPL3037631
NPL3037634
NPL3037635
NPL3037636
NPL3037637
NPL3037640
NPL3037641
NPL3037642
NPL3037645
NPL3037646
NPL3037647
NPL3037648
NPL3037650
NPL3037651
NPL3037652
NPL3037653
NPL3037654
NPL3037655
NPL3037657
NPL3037658
NPL3037659
NPL3037660
NPL3037661
NPL3037662
NPL3037664
NPL3037668
NPL3037669
NPL3037670
NPL3037672
NPL3037673
NPL3037674
NPL3037677
NPL3037678
NPL3037679
NPL3037681
NPL3037684
NPL3037688
NPL3037689
NPL3037690
NPL3037691
NPL3037695
NPL3037699
NPL3037700
NPL3037703
NPL3037704
NPL3037705
NPL3037706
NPL3037707
NPL3037708
NPL3037711
NPL3037713
NPL3037714
NPL3037715
NPL3037719
NPL3037720
NPL3037725
NPL3037728
NPL3037729
NPL3037732
NPL3037736
NPL3037737
NPL3037740
NPL3037741
NPL3037742
NPL3037745
NPL3037746
NPL3037747
NPL3037748
NPL3037750
NPL3037757
NPL3037759
NPL3037761
NPL3037762
NPL3037764
NPL3037765
NPL3037766
NPL3037767
NPL3037768
NPL3037771
NPL3037774
NPL3037776
NPL3037779
NPL3037780
NPL3037784
NPL3037785
NPL3037786
NPL3037789
NPL3037793
NPL3037796
NPL3037800
NPL3037802
NPL3037805
NPL3037811
NPL3037813
NPL3037814
NPL3037817
NPL3037818
NPL3037823
NPL3037826
NPL3037828
NPL3037830
NPL3037832
NPL3037833
NPL3037835
NPL3037837
NPL3037839
NPL3037842
NPL3037843
NPL3037844
NPL3037845
NPL3037847
NPL3037849
NPL3037850
NPL3037852
NPL3037855
NPL3037858
NPL3037860
NPL3037865
NPL3037870
NPL3037872
NPL3037879
NPL3037883
NPL3037884
NPL3037889
NPL3037899
NPL3037902
NPL3037907
NPL3037909
NPL3037912
NPL3037914
NPL3037915
NPL3037917
NPL3037922
NPL3037925
NPL3037928
NPL3037929
NPL3037930
NPL3037931
NPL3037932
NPL3037934
NPL3037937
NPL3037939
NPL3037944
NPL3037946
NPL3037948
NPL3037954
NPL3037956
NPL3037958
NPL3037960
NPL3037963
NPL3037965
NPL3037966
NPL3037973
NPL3037974
NPL3037975
NPL3037976
NPL3037977
NPL3037978
NPL3037979
NPL3037986
NPL3037990
NPL3037992
NPL3037996
NPL3037999
NPL3038000
NPL3038005
NPL3038012
NPL3038014
NPL3038015
NPL3038019
NPL3038020
NPL3038041
NPL3038042
NPL3038044
NPL3038045
NPL3038046
NPL3038048
NPL3038049
NPL3038051
NPL3038060
NPL3038062
NPL3038063
NPL3038064
NPL3038067
NPL3038069
NPL3038071
NPL3038072
NPL3038074
NPL3038075
NPL3038077
NPL3038078
NPL3038079
NPL3038080
NPL3038081
NPL3038091
NPL3038093
NPL3038097
NPL3038100
NPL3038107
NPL3038108
NPL3038109
NPL3038110
NPL3038113
NPL3038115
NPL3038119
NPL3038125
NPL3038126
NPL3038128
NPL3038129
NPL3038130
NPL3038133
NPL3038134
NPL3038136
NPL3038137
NPL3038138
NPL3038140
NPL3038142
NPL3038143
NPL3038151
NPL3038153
NPL3038154
NPL3038156
NPL3038160
NPL3038161
NPL3038162
NPL3038164
NPL3038167
NPL3038170
NPL3038173
NPL3038174
NPL3038175
NPL3038176
NPL3038177
NPL3038179
NPL3038180
NPL3038181
NPL3038184
NPL3038186
NPL3038187
NPL3038188
NPL3038189
NPL3038191
NPL3038193
NPL3038199
NPL3038205
NPL3038206
NPL3038214
NPL3038215
NPL3038217
NPL3038221
NPL3038222
NPL3038223
NPL3038224
NPL3038227
NPL3038229
NPL3038232
NPL3038233
NPL3038234
NPL3038239
NPL3038240
NPL3038243
NPL3038245
NPL3038246
NPL3038249
NPL3038251
NPL3038252
NPL3038253
NPL3038255
NPL3038256
NPL3038257
NPL3038260
NPL3038261
NPL3038263
NPL3038264
NPL3038265
NPL3038266
NPL3038269
NPL3038274
NPL3038276
NPL3038277
NPL3038280
NPL3038282
NPL3038283
NPL3038286
NPL3038287
NPL3038288
NPL3038291
NPL3038292
NPL3038293
NPL3038296
NPL3038297
NPL3038298
NPL3038300
NPL3038310
NPL3038316
NPL3038319
NPL3038320
NPL3038321
NPL3038323
NPL3038324
NPL3038325
NPL3038326
NPL3038327
NPL3038331
NPL3038334
NPL3038337
NPL3038343
NPL3038344
NPL3038346
NPL3038347
NPL3038348
NPL3038349
NPL3038350
NPL3038352
NPL3038355
NPL3038356
NPL3038361
NPL3038363
NPL3038365
NPL3038366
NPL3038370
NPL3038371
NPL3038372
NPL3038373
NPL3038374
NPL3038375
NPL3038376
NPL3038379
NPL3038380
NPL3038381
NPL3038385
NPL3038386
NPL3038388
NPL3038390
NPL3038395
NPL3038397
NPL3038399
NPL3038400
NPL3038401
NPL3038402
NPL3038403
NPL3038404
NPL3038405
NPL3038409
NPL3038412
NPL3038413
NPL3038414
NPL3038416
NPL3038418
NPL3038419
NPL3038421
NPL3038423
NPL3038427
NPL3038429
NPL3038430
NPL3038432
NPL3038434
NPL3038440
NPL3038443
NPL3038445
NPL3038446
NPL3038461
NPL3038462
NPL3038475
NPL3038494
NPL3038495
NPL3038496
NPL3038497
NPL3038509
NPL3038510
NPL3038512
NPL3038513
NPL3038514
NPL3038515
NPL3038516
NPL3038517
NPL3038518
NPL3038519
NPL3038520
NPL3038522
NPL3038524
NPL3038526
NPL3038527
NPL3038528
NPL3038530
NPL3038532
NPL3038534
NPL3038539
NPL3038545
NPL3038550
NPL3038553
NPL3038580
NPL3038581
NPL3038585
NPL3038588
NPL3038599
NPL3038603
NPL3038604
NPL3038614
NPL3038617
NPL3038630
NPL3038714
NPL3038718
NPL3038719
NPL3038736
NPL3038737
NPL3038738
NPL3038741
NPL3038742
NPL3038779
NPL3038780
NPL3038781
NPL3038802
NPL3038804
NPL3038827
NPL3038828
NPL3038829
NPL3038830
NPL3038831
NPL3038832
NPL3038833
NPL3038847
NPL3038849
NPL3038851
NPL3038852
NPL3038854
NPL3038856
NPL3038865
NPL3038873
NPL3038875
NPL3038876
NPL3038877
NPL3038885
NPL3038888
NPL3038891
NPL3038892
NPL3038895
NPL3038898
NPL3038899
NPL3038900
NPL3038905
NPL3038908
NPL3038909
NPL3038911
NPL3038916
NPL3038917
NPL3038918
NPL3038920
NPL3038921
NPL3038928
NPL3038946
NPL3038947
NPL3038951
NPL3038953
NPL3038954
NPL3038955
NPL3038956
NPL3038958
NPL3038959
NPL3038966
NPL3038972
NPL3038973
NPL3038974
NPL3038975
NPL3038976
NPL3038982
NPL3038984
NPL3038985
NPL3038987
NPL3038989
NPL3038990
NPL3038992
NPL3038996
NPL3039002
NPL3039003
NPL3039012
NPL3039013
NPL3039014
NPL3039015
NPL3039016
NPL3039019
NPL3039020
NPL3039022
NPL3039023
NPL3039024
NPL3039025
NPL3039028
NPL3039033
NPL3039034
NPL3039035
NPL3039036
NPL3039040
NPL3039043
NPL3039044
NPL3039046
NPL3039047
NPL3039055
NPL3039058
NPL3039059
NPL3039060
NPL3039061
NPL3039062
NPL3039063
NPL3039067
NPL3039071
NPL3039072
NPL3039073
NPL3039074
NPL3039076
NPL3039084
NPL3039085
NPL3039087
NPL3039088
NPL3039097
NPL3039119
NPL3039121
NPL3039123
NPL3039127
NPL3039131
NPL3039132
NPL3039134
NPL3039137
NPL3039138
NPL3039152
NPL3039154
NPL3039168
NPL3039179
NPL3039182
NPL3039183
NPL3039187
NPL3039188
NPL3039189
NPL3039190
NPL3039194
NPL3039195
NPL3039201
NPL3039209
NPL3039211
NPL3039212
NPL3039231
NPL3039232
NPL3039233
NPL3039234
NPL3039237
NPL3039239
NPL3039240
NPL3039241
NPL3039242
NPL3039244
NPL3039245
NPL3039246
NPL3039252
NPL3039254
NPL3039255
NPL3039257
NPL3039260
NPL3039262
NPL3039263
NPL3039265
NPL3039266
NPL3039267
NPL3039268
NPL3039269
NPL3039270
NPL3039271
NPL3039272
NPL3039273
NPL3039274
NPL3039278
NPL3039279
NPL3039280
NPL3039291
NPL3039292
NPL3039293
NPL3039296
NPL3039305
NPL3039308
NPL3039309
NPL3039310
NPL3039326
NPL3039327
NPL3039329
NPL3039330
NPL3039331
NPL3039332
NPL3039333
NPL3039334
NPL3039335
NPL3039340
NPL3039345
NPL3039346
NPL3039347
NPL3039348
NPL3039349
NPL3039353
NPL3039354
NPL3039356
NPL3039357
NPL3039359
NPL3039360
NPL3039364
NPL3039365
NPL3039374
NPL3039375
NPL3039376
NPL3039378
NPL3039382
NPL3039383
NPL3039384
NPL3039385
NPL3039386
NPL3039387
NPL3039388
NPL3039389
NPL3039390
NPL3039391
NPL3039392
NPL3039393
NPL3039394
NPL3039395
NPL3039396
NPL3039397
NPL3039398
NPL3039399
NPL3039400
NPL3039401
NPL3039402
NPL3039403
NPL3039404
NPL3039405
NPL3039406
NPL3039407
NPL3039408
NPL3039419
NPL3039420
NPL3039429
NPL3039430
NPL3039435
NPL3039436
NPL3039440
NPL3039456
NPL3039457
NPL3039460
NPL3039479
NPL3039484
NPL3039485
NPL3039486
NPL3039488
NPL3039489
NPL3039490
NPL3039495
NPL3039496
NPL3039497
NPL3039498
NPL3039501
NPL3039504
NPL3039509
NPL3039510
NPL3039511
NPL3039512
NPL3039513
NPL3039514
NPL3039515
NPL3039516
NPL3039517
NPL3039518
NPL3039519
NPL3039523
NPL3039525
NPL3039526
NPL3039528
NPL3039530
NPL3039537
NPL3039542
NPL3039543
NPL3039544
NPL3039545
NPL3039548
NPL3039555
NPL3039556
NPL3039557
NPL3039558
NPL3039560
NPL3039562
NPL3039611
NPL3039612
NPL3039613
NPL3039614
NPL3039615
NPL3039619
NPL3039622
NPL3039625
NPL3039626
NPL3039627
NPL3039640
NPL3039642
NPL3039647
NPL3039656
NPL3039657
NPL3039659
NPL3039660
NPL3039664
NPL3039672
NPL3039678
NPL3039679
NPL3039680
NPL3039681
NPL3039686
NPL3039688
NPL3039701
NPL3039703
NPL3039704
NPL3039705
NPL3039709
NPL3039713
NPL3039732
NPL3039735
NPL3039736
NPL3039776
NPL3039801
NPL3039806
NPL3039844
NPL3039850
NPL3039852
NPL3039862
NPL3039865
NPL3039866
NPL3026119
NPL3026156
NPL3038531
NPL3025012
NPL3026130
|
Recalling Firm/ Manufacturer |
Medtronic Neuromodulation 7000 Central Ave Ne Minneapolis MN 55432-3568
|
| For Additional Information Contact | Naomi Rodiles 763-514-2000 |
Manufacturer Reason for Recall | Software issues were identified in application version 2.x. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An URGENT: MEDICAL DEVICE CORRECTION notification letter dated November 2024 was sent to customers.
Actions
Share this notice with all those who need to be aware of this issue within your organization and maintain a copy of this notice in your records.
Please complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information. The form may be emailed to neuro.quality@medtronic.com.
Additional Information
Medtronic is actively working on a permanent solution and will inform you as soon as it becomes available.
Medtronic is communicating this information to the appropriate regulatory agencies. Adverse events or quality problems experienced with this product should be reported to Medtronic and the FDA MedWatch Adverse Event Reporting program via:
Medtronic email at RS.NeuAdverseEventReporting@Medtronic.com or Medtronic Technical Services at 1-800-707-0933 weekdays 7:00 AM 6:00 PM CT.
***Update January 2026***
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated January 2026 was sent to customers.
The purpose of this letter is to inform you that a new version of the Model A810 SynchroMed Clinician Programmer Application (CP App), v2.0.3320, is now available for installation. This update addresses previously communicated software issues associated with earlier versions used for the Model 8637 SynchroMed II and Model 8667 SynchroMed III infusion pumps.
These issues were initially communicated to you in the letter enclosed from November 2024. The new version of the A810 CP App resolves these issues.
Customer Actions:
" Download and install A810 CP App v2.0.3320 at your earliest convenience and contact your local Medtronic representative if you have any questions related to the update.
" Sign and return the Confirmation Form enclosed with this letter, acknowledging that you have received this information. The form may be emailed to neuro.quality@medtronic.com. Retain a copy of this lette |
| Quantity in Commerce | 3981 systems |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|