• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmNovember 21, 2024
Date PostedDecember 30, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0779-2025
Recall Event ID 95852
Product Classification Central venous catheter tray - Product Code OFF
ProductMedline custom medical procedure kits labeled as: 1) VP SHUNT CDS, REF CDS983470D; 2) 1/2X1/2 NEURO SPONGE PACK-LF, REF DYNJ0713177; 3) BURR HOLE SHUNT PACK, REF DYNJ56819D; 4) BURR HOLE SHUNT PACK, REF DYNJ56819F; 5) BURR HOLE SHUNT PACK, REF DYNJ56819G; 6) RO VP SHUNT PACK, REF PHS392834010D; 7) VP SHUNT, REF SYNJ10307A.
Code Information 1) REF CDS983470D, UDI/DI 10193489275339 (EA) 40193489275330(CS), Lot numbers: 20BDB519, 20EDA360, 20EDC040, 20GDA474, 20JDA107, 20KDA855, 21CDA110, 21EDA674, 21FDA748; 2) REF DYNJ0713177, UDI/DI 10889942085088 (EA) 40889942085089(CS), Lot numbers: 19AKC072, 19HKC645, 19KKA304, 20FKA539; 3) REF DYNJ56819D, UDI/DI 10193489254273 (EA) 40193489254274(CS), Lot numbers: 20AKA332, 20BKA860; 4) REF DYNJ56819F, UDI/DI 10193489353532 (EA) 40193489353533(CS), Lot numbers: 20EKB648, 20HKA426, 20JKA217, 20LKB026, 21FBT545, 21HBL828, 21KBY910, 21KBY914; 5) REF DYNJ56819G, UDI/DI 10195327073749 (EA) 40195327073740(CS), Lot numbers: 22CBJ951, 22HBJ403, 22LBQ584, 23BBH164, 23CBG816, 23FBM524; 6) REF PHS392834010D, UDI/DI 10193489222432 (EA) 40193489222433(CS), Lot numbers: 21LDA084; 7) REF SYNJ10307A, UDI/DI 10193489377026 (EA) 40193489377027(CS), Lot numbers: 20HBU444, 20HBW312, 21EBD895, 21FBE508, 21HBS523.
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMedline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Quantity in Commerce1402 units
DistributionWorldwide distribution - US Nationwide and the countries of UAE, and Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-