| | Class 2 Device Recall VITROS XT 7600 Integrated System |  |
| Date Initiated by Firm | December 19, 2024 |
| Date Posted | January 21, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1006-2025 |
| Recall Event ID |
96031 |
| 510(K)Number | K182063 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Brand Name: VITROS
Product Name: VITROS XT 7600 Integrated System
Model/Catalog Number: 6844461 (New), 6272222 (Certified), 6904003 (China)
Software Version: version 3.8.0 or higher
Product Description: In vitro diagnostic chemistry analyzer.
FDA 3 letter product code: JJE
Component: No, not a component |
| Code Information |
Lot Code: VITROS XT 7600 Integrated System
Product Code: 6844461 (New)
UDI: 10758750012343
VITROS XT 7600 System (Certified)
Product Code: 6272222
UDI: 10758750012343
Lot Code: VITROS XT 7600 Integrated System
Product Code: 6904003 (China)
UDI: 10758750035809
Range of installed serial numbers:76000001-76002199
Manufacture date range: July 2018 to present
Distribution date range; July 2018 to present
Total systems currently installed in the US: 994
Total systems currently installed outside the US: 899
Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined.
Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today.
Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
|
| FEI Number |
1000305840
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, INc.
513 Technology Blvd
Rochester
NY
14626-3601
|
| For Additional Information Contact | Joe Falvo 585-4533452 |
Manufacturer Reason for Recall | The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.
|
FDA Determined Cause 2 | Software Design Change |
| Action | On December 19, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION
" Do not use expired Signal Reagent packs. See the Questions & Answers section for more details.
" Complete the enclosed Confirmation of Receipt form no later than December 31, 2024.
" Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System in your laboratory until the issue has been resolved.
Contact Information
We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088. |
| Quantity in Commerce | 1,893 units (994 US, 899 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|