• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BD Pyxis

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BD Pyxissee related information
Date Initiated by FirmFebruary 14, 2023
Date PostedJanuary 17, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0974-2025
Recall Event ID 96053
Product Classification Cabinet, table and tray, anesthesia - Product Code BRY
ProductBD Pyxis MedStation 4000 AUX, REF 306
Code Information UDI/DI 10885403512636, Serial Numbers: 9304700, 9305723, 9309302, 9413001, 9413014, 9415602, 9415665, 9417311, 9417312, 9417335, 9417340, 9417428, 9417430, 9417433, 9417437, 9418065, 9418068, 9418100, 9418105, 9418111, 9418112, 9418113, 9418114, 9418115, 9418129, 9418568, 9523679, 9525535, 9525558, 9525560, 9525561, 9525562, 9525582, 9525590, 9525600, 9525605, 9525606, 9525613, 9525624, 9525630, 9525638, 9525656, 9525657, 9526867, 9528530, 9528836, 9528859, 9528890, 9531253, 9531257, 12439518, 12439520, 12439521, 12439907, 12440208, 12440217, 12440218, 12440219, 12440220, 12440227, 12440339, 12440984, 12440985, 12443922, 12443923, 12445433, 12447385, 12447387, 12447388, 12447389, 12447394, 12447395, 12447396, 12448203, 12448204, 12452249, 12452265, 12452646, 12452649, 12453000, 12453215, 12453217, 12453238, 12453242, 12460386, 12460388, 12461366, 12461367, 12461927, 12462829, 12462831, 12462833, 12462835, 12463516, 12463517, 12463518, 12464145, 12464146, 12465451, 12465452, 12465453, 12465454, 12465455, 12465456, 12465457, 12465458, 12465459, 12466105, 12466106, 12470354, 12470398, 12470399, 12470890, 12472293, 12474035, 12474844, 12474845, 12476540, 12476541, 12476694, 12482559, 12482560, 12482561, 12482562, 12487398, 12487399, 12487400, 12487401, 12487402, 12487403, 12487404, 12487405, 12487407, 12488733, 12488734, 12490126, 12490127, 12494564, 12494565, 12494769, 12494770, 12494771, 12495340, 12495341, 12495344, 12495345, 12498244, 12498247, 12500586, 12500587, 12500588, 12500589, 12500591, 12502193, 12502865, 12502887, 12507268, 12507287, 12507288, 12511509, 12511516, 12511936, 12511937, 12512135, 12515208, 12515688, 12517918, 12517919, 12517922, 12517929, 12517932, 12517940, 12524687, 12525675, 12530269, 12536035, 12539925, 12549048, 12549049, 12549050, 12549064, 12549093, 12549212, 12550827, 12550828, 12553557, 12555134, 12561011, 12561891, 12564119, 12567515, 12568591, 12570581, 12570583, 12570585, 12570588, 12570591, 12571514, 12571516, 12573404, 12573411, 12574278, 12575617, 12575618, 12577371, 12577374, 12579121, 12579122, 12579124, 12579334, 12580215, 12580216, 12581915, 12581966, 12581967, 12581968, 12581969, 12581993, 12581995, 12581996, 12583672, 12583674, 12586354, 12588479, 12595252, 12599174, 12599393, 12602410, 12605978, 12605979, 12608119, 12608120, 12613253, 12618230, 12618231, 12618234, 12618235, 12623557, 12626539, 12627331, 12627333, 12629812, 12629813, 12629814, 12629819, 12629922, 12629923, 12629928, 12634465, 12635146, 12635147, 12635148, 12635369, 12635370, 12635638, 12658974, 12659199, 12659210, 12659212, 12659665, 12659669, 12662147, 12668807, 12680615, 12680616, 12681705, 12681762, 12682952, 12685164, 12686454, 12688387, 12688392, 12688393, 12690138, 12690333, 12693393, 12693395, 12693958, 12695306, 12695309, 12695699, 12696372, 12697189, 12697190, 12697591, 12697593, 12697594, 12697630, 12697633, 12697999, 12698369, 12698376, 12699437, 12699438, 12699439, 12699440, 12700049, 12700050, 12700051, 12700052, 12700053, 12700054, 12700064, 12700065, 12700066, 12700067, 12700068, 12700069, 12700070, 12700076, 12700077, 12700078, 12700079, 12700080, 12700081, 12700082, 12700083, 12700084, 12700085, 12700297, 12700617, 12701912, 12702061, 12702062, 12702081, 12702110, 12702311, 12702312, 12703855, 12703939, 12704168, 12705208, 12705739, 12705742, 12706339, 12706341, 12706342, 12706360, 12706884, 12709776, 12710726, 12710727, 12710728, 12712274, 12712277, 12712830, 12714589, 12716740, 12716742, 12716744, 12719022, 12719029, 12720148, 12720149, 12721805, 12721806, 12721953, 12722836, 12725536, 12726415, 12726416, 12726417, 12729974, 12731258, 12734620, 12734762, 12737718, 12737721, 12737722, 12746950, 12746951, 12747608, 12747727, 12751610, 12752806, 12752807, 12752808, 12758039, 12758697, 12765458, 12771652, 12771653, 12771654, 12771655, 12772979, 12773813, 12773814, 12773818, 12774834, 12774836, 12774945, 12777930, 12777931, 12784101, 12784102, 12784103, 12786113, 12786114, 12786115, 12786116, 12786117.12786118, 12786826, 12786827, 12787395, 12788886, 12788887, 12788888, 12788890, 12790295, 12794336, 12794338, 12794469, 12794937, 12797101, 12797102, 12797103, 12797195, 12797844, 12797845, 12797846, 12797847, 12797848, 12798057, 12799772, 12800460, 12802678, 12803691, 12803692, 12803697, 12803698, 12803928, 12804358, 12804359, 12804360, 12804767, 12804768, 12804783, 12804873, 12804875, 12804876, 12805526, 12805527, 12805923, 12805925, 12805927, 12805928, 12805929, 12805932, 12805933, 12805934, 12805935, 12805936, 12805937, 12805939, 12805941, 12805944, 12806570, 12806571, 12807448, 12808059, 12808060, 12808131, 12808132, 12808134, 12809490, 12809491, 12809492, 12809493, 12809494, 12809496, 12809497, 12809498, 12810330, 12811945, 12811947, 12811949, 12811950, 12811955, 12813228, 12813245, 12814191, 12814192, 12815157, 12815831, 12815832, 12815834, 12815835, 12815840, 12815841, 12815842, 12815843, 12815844, 12815846, 12815847, 12815849, 12815850, 12815892, 12815893, 12815895, 12817760, 12817761, 12817763, 12817766, 12817767, 12817770, 12817771, 12817814, 12820030, 12822654, 12822819, 12822821, 12822829, 12822830, 12823390, 12823393, 12823759, 12824110, 12824112, 12824113, 12824116, 12824117, 12824118, 12824119, 12824120, 12824121, 12824213, 12824214, 12824302, 12824303, 12825225, 12826707, 12826709, 12826797, 12828139, 12828922, 12829784, 12829785, 12831634, 12831635, 12833821, 12833823, 12833825, 12833875, 12833876, 12833877, 12833878, 12833879, 12833884, 12833885, 12836252, 12836254, 12836255, 12837502, 12837503, 12837504, 12837541, 12838402, 12838404, 12838405, 12838406, 12838407, 12838408, 12840394, 12840431, 12840432, 12840433, 12840434, 12843821, 12846195, 12851244, 12851245, 12851246, 12851247, 12851248, 12851249, 12851250, 12852330, 12852336, 12852338, 12852351, 12853614, 12853615, 12853617, 12853618, 12853624, 12855845, 12858808, 12858810, 12858825, 12863663, 12868847, 12868864, 12869725, 12869726, 12869727, 12870855, 12872360, 12872873, 12878432, 12880471, 12881195, 12881196, 12881197, 12881198, 12881201, 12881202, 12881203, 12881566, 12881567, 12881568, 12881569, 12881570, 12881582, 12881593, 12881595, 12881598, 12882011, 12882012, 12884989, 12884992, 12884997, 12884998, 12885006, 12886585, 12886587, 12886986, 12888141, 12888319, 12888320, 12888321, 12888322, 12888323, 12888324, 12888325, 12888599, 12888656, 12890308, 12890309, 12890334, 12890335, 12893880, 12893883, 12893884, 12894386, 12894387, 12894388, 12894389, 12894390, 12894391, 12894975, 12895333, 12895380, 12895382, 12895927, 12895929, 12895931, 12896408, 12896409, 12896410, 12896411, 12896412, 12896413, 12896416, 12896417, 12897588, 12898019, 12899000, 12899001, 12899022, 12899420, 12899421, 12899422, 12901425, 12901426, 12901427, 12901428, 12901429, 12901430, 12901431, 12901432, 12901433, 12901434, 12901435, 12901436, 12901437, 12901578, 12901579, 12901984, 12902487, 12902725, 12903356, 12903357, 12903367, 12903368, 12903369, 12904052, 12904053, 12904054, 12904180, 12904295, 12904296, 12904297, 12904299, 12904300, 12904327, 12904328, 12904906, 12905668, 12906165, 12906166, 12906181, 12906358, 12906359, 12906360, 12906387, 12907111, 12907112, 12907113, 12907114, 12907115, 12907116, 12907117, 12907118, 12907221, 12907902, 12908359, 12908360, 12908370, 12908371, 12908379, 12908555, 12908556, 12908557, 12908558, 12909323, 12909324, 12909871, 12909879, 12909955, 12910308, 12910309, 12910592, 12910593, 12910890, 12910891, 12910894, 12911069, 12911846, 12912033, 12912214, 12912798, 12913322, 12913837, 12913971, 12913972, 12914033, 12914429, 12914433, 12914434, 12914488, 12914695, 12915288, 12915316, 12915904, 12916415, 12917019, 12917692, 12918868, 12922022, 12923024, 12923025, 12923583, 12924959, 12927034, 12927770, 12928119, 12928120, 12928189, 12931929, 12931930, 12931931, 12931932, 12931933, 12931935, 12931937, 12931938, 12931939, 12933996, 12933997, 12933998, 12933999, 12938494, 12939335, 12939336, 12939608, 12948908, 12948909, 12948935, 12950084, 12950090.12950100, 12951318, 12952556, 12952557, 12952558, 12952559, 12952577, 12952578, 12952598, 12954031, 12954032, 12954703, 12954704, 12954705, 12954708, 12954709, 12954833, 12955409, 12957461, 12957462, 12957463, 12957464, 12957465, 12957501, 12957502, 12959528, 12959536, 12959538, 12959562, 12959563, 12959564, 12959565, 12960382, 12960384, 12961112, 12963362, 12963818, 12963846, 12964165, 12965353, 12966355, 12966356, 12967099, 12967867, 12972165, 12972357, 12975310, 12976612, 12976613, 12977813, 12977816, 12977837, 12978788, 12978789, 12978790, 12978834, 12978841, 12981303, 12983206, 12983207, 12983222, 12983224, 12983225, 12983226, 12984623, 12985653, 12985654, 12985656, 12985657, 12985678, 12986463, 12986473, 12986574, 12986575, 12989845, 12989846, 12990238, 12990239, 12990240, 12991593, 12992890, 12992891, 12993632, 12996830, 12996831, 12996832, 12996833, 12996834, 12996835, 12996836, 12996853, 12996854, 12996876, 12996880, 12996881, 12996882, 12997803, 12998735, 12998736, 12998737, 12998738, 12998739, 12998740, 12998762, 12998763, 12998795, 12998796, 12999547, 12999548, 12999549, 12999550, 12999551, 13000371, 13000866, 13001801, 13001802, 13003677, 13003796, 13003797, 13004707, 13004708, 13008201, 13008205, 13009132, 13009166, 13015141, 13020050, 13021126, 13021127, 13022161, 13022162, 13022189, 13022191, 13022216, 13022467, 13022468, 13022469, 13022470, 13022473, 13022474, 13022475, 13022476, 13022477, 13022478, 13022479, 13023150, 13023904, 13023905, 13023906, 13024394, 13024395, 13024401, 13024402, 13024432, 13024668, 13024670, 13024986, 13024987, 13024989, 13025097, 13025496, 13025936, 13025937, 13025964, 13025967, 13026395, 13026485, 13026769, 13026783, 13026784, 13026785, 13026810, 13026811, 13026815, 13026816, 13027077, 13028154, 13031714, 13031715, 13031716, 13031718, 13031719, 13031720, 13032525, 13032526, 13033187, 13034495, 13034496, 13036048, 13036049, 13036050, 13036051, 13036052, 13036880, 13036908, 13036909, 13036921, 13039023, 13039601, 13039602, 13040143, 13040154, 13040155, 13040156, 13040157, 13040158, 13040159, 13040177, 13041003, 13041777, 13042603, 13043071, 13043074, 13043076, 13043077, 13043116, 13043175, 13043176, 13043178, 13044228, 13047033, 13047438, 13047497, 13047657, 13049698, 13049704, 13051044, 13051093, 13052852, 13052853, 13052854, 13052855, 13052895, 13054053, 13055118, 13057651, 13058397, 13058398, 13058928, 13058942, 13058943, 13059630, 13059631, 13060872, 13060904, 13062872, 13064171, 13064172, 13064173, 13064219, 13064504, 13064505, 13064506, 13064507, 13064508, 13064691, 13065460, 13067109, 13067110, 13069693, 13069694, 13069695, 13069696, 13069697, 13069698, 13070505, 13072207, 13072208, 13072209, 13072210, 13072230, 13072231, 13072239, 13072240, 13073038, 13073040, 13073041, 13073072, 13073073, 13073089, 13073090, 13073091, 13073092, 13073094, 13074510, 13074569, 13074570, 13074579, 13074998, 13074999, 13076423, 13076424, 13076425, 13076440, 13076441, 13076443, 13076467, 13077014, 13077015, 13077659, 13077752, 13078270, 13078271, 13078272, 13078273, 13078274, 13078672, 13078675, 13079051, 13079237, 13079238, 13079240, 13080366, 13080367, 13080373, 13080374, 13080888, 13081172, 13081174, 13081744, 13083264, 13083265, 13083957, 13083958, 13084015, 13084036, 13084553, 13084554, 13084555, 13084556, 13084557, 13085978, 13085979, 13087606, 13087647, 13088316, 13088317, 13088320, 13088442, 13088443, 13088444, 13088445, 13088446, 13088447, 13089187, 13089188, 13089189, 13089226, 13091169, 13091676, 13091872, 13091873, 13091874, 13092537, 13092538, 13092539, 13092540, 13092559, 13092620, 13092626, 13096575, 13096576, 13096577, 13096578, 13096579, 13096582, 13096583, 13096585, 13096632, 13098696, 13098697, 13098698, 13098699, 13098700, 13098701, 13098702, 13098703, 13098705, 13100286, 13101554, 13103221, 13103222, 13103223, 13103840, 13103841, 13103970, 13105090, 13105093, 13107007, 13107098, 13107106, 13107108, 13107109, 13109980, 13109981, 13109983, 13109985, 13109986, 13112666, 13112669.13112710, 13112711, 13112712, 13114103, 13114694, 13114695, 13114696, 13115829, 13115830, 13118427, 13118689, 13118690, 13119516, 13120302, 13120304, 13120314, 13120315, 13120397, 13120398, 13120399, 13121374, 13121895, 13121962, 13124117, 13124118, 13124119, 13124120, 13124269, 13124270, 13124271, 13125049, 13125051, 13125052, 13125129, 13125130, 13125354, 13125376, 13125410, 13126746, 13126747, 13126748, 13126749, 13126750, 13126792, 13128449, 13128450, 13128451, 13128458, 13128471, 13128473, 13128785, 13128787, 13128789, 13129667, 13131362, 13131363, 13132165, 13132626, 13132634, 13132635, 13132636, 13132637, 13132679, 13132768, 13132775, 13134527, 13135399, 13136244, 13136245, 13137828, 13138519, 13138538, 13139320, 13139323, 13139799, 13139801, 13139802, 13139803, 13139806, 13139807, 13139808, 13139809, 13139810, 13139994, 13139995, 13139998, 13140031, 13140722, 13140735, 13141508, 13141509, 13141510, 13141511, 13141512, 13141526, 13141674, 13141675, 13142154, 13142157, 13142167, 13142342, 13142343, 13142571, 13142592, 13143979, 13143980, 13143981, 13145425, 13145426, 13145427, 13145428, 13145447, 13146796, 13146798, 13146799, 13147621, 13147622, 13147623, 13147624, 13147625, 13147626, 13147810, 13148050, 13148051, 13148086, 13149243, 13149244, 13149248, 13149249, 13150846, 13151658, 13151698, 13152765, 13153207, 13153254, 13153256, 13153277, 13153278, 13154286, 13154287, 13154288, 13154289, 13154290, 13154291, 13154293, 13154294, 13154296, 13154297, 13154298, 13154299, 13154300, 13154301, 13155237, 13155279, 13155922, 13156267, 13156268, 13156269, 13156270, 13156271, 13156272, 13156348, 13156832, 13156859, 13158568, 13158569, 13158570, 13158571, 13158572, 13158605, 13158606, 13159258, 13159259, 13161900, 13162933, 13165018, 13165019, 13165033, 13165034, 13168313, 13169406, 13169407, 13169409, 13169411, 13172601, 13172602, 13172603, 13172604, 13173220, 13174001, 13174002, 13174003, 13174004, 13174718, 13174719, 13175428, 13176356, 13178987, 13179313, 13179329, 13179331, 13179867, 13179868, 13179869, 13180066, 13181837, 13181838, 13181839, 13181840, 13181841, 13181891, 13181892, 13181894, 13181895, 13181896, 13181897, 13181938, 13182477, 13185800, 13192436, 13192437, 13192438, 13192439, 13192440, 13192441, 13193100, 13193101, 13193589, 13193590, 13193591, 13193592, 13193593, 13193594, 13193595, 13193596, 13193990, 13193991, 13194413, 13194421, 13195102, 13195103, 13195113, 13198469, 13198470, 13198613, 13198624, 13200244, 13200245, 13200253, 13200409, 13200410, 13200411, 13200412, 13200413, 13200461, 13201324, 13201325, 13201327, 13201346, 13201347, 13201348, 13201349, 13201352, 13201425, 13201426, 13201427, 13201429, 13201430, 13201451, 13202179, 13202180, 13202181, 13202182, 13202183, 13202184, 13202185, 13202186, 13202200, 13202201, 13202227, 13204213, 13204362, 13204925, 13204926, 13205468, 13206692, 13206722, 13206725, 13207459, 13207499, 13207976, 13208149, 13208150, 13208680, 13211148, 13211657, 13211693, 13211707, 13214385, 13214386, 13214387, 13214388, 13214461, 13214463, 13215907, 13215909, 13215915, 13216869, 13217950, 13217951, 13217952, 13217976, 13217977, 13217980, 13217981, 13218023, 13218024, 13218028, 13218029, 13219794, 13219795, 13219796, 13219797, 13321179, 13323820, 13324892, 13324937, 13325160, 13325161, 13325162, 13325163, 13325164, 13325165, 13333136, 13336000, 13343952, 13343956, 13343960, 13347238, 13347239, 13347409, 13347537, 13347541, 13350354, 13350355, 13350357, 13350458, 13350459, 13350460, 13350461, 13350462, 13350463, 13350464, 13350465, 13350466, 13350467, 13350468, 13350469, 13350470, 13350471, 13350472, 13350473, 13350796, 13350797, 13350798, 13350799, 13350800, 13350802, 13350804, 13350806, 13350809, 13350811, 13350813, 13350816, 13350818, 13350821, 13350823, 13350824, 13351252, 13351253, 13351254, 13351274, 13351276, 13351821, 13351845, 13352742, 13352743, 13352744, 13358594, 13358595, 13358596, 13360774, 13362749, 13363046, 13363054, 13364060, 13364065, 13364512, 13366335, 13366341.13368306, 13372368, 13372369, 13373614, 13373796, 13373797, 13373869, 13373871, 13373873, 13373875, 13373882, 13377674, 13377679, 13378119, 13378140, 13378141, 13378142, 13378143, 13381825, 13382002, 13382003, 13382004, 13382005, 13382006, 13383694, 13384048, 13384049, 13384050, 13384051, 13384053, 13384054, 13384232, 13384233, 13385354, 13385367, 13385368, 13387331, 13387332, 13387488, 13387489, 13387497, 13387498, 13387499, 13387510, 13387511, 13387515, 13388048, 13388170, 13388179, 13388182, 13390975, 13390980, 13390987, 13390990, 13391997, 13392651, 13392652, 13392685, 13392749, 13392818, 13392881, 13392882, 13392883, 13392884, 13392895, 13392923, 13392926, 13393048, 13393049, 13393065, 13393066, 13393067, 13393080, 13393081, 13393082, 13393084, 13393085, 13393086, 13395233, 13395236, 13395247, 13395396, 13395418, 13395419, 13395420, 13395421, 13395422, 13398383, 13398384, 13398385, 13398386, 13399456, 13399457, 13399458, 13399462, 13406264, 13407100, 13407292, 13407293, 13407294, 13407295, 13407296, 13407297, 13407298, 13407299, 13408613, 13408928, 13409189, 13409191, 13409193, 13410004, 13410399, 13410402, 13410404, 13410406, 13410409, 13410410, 13411422, 13414953, 13414954, 13414956, 13414957, 13414958, 13416345, 13416348, 13416361, 13416364, 13416365, 13416366, 13416367, 13416368, 13416369, 13416370, 13416371, 13419887, 13419888, 13419892, 13420570, 13420676, 13420736, 13420737, 13420791, 13423529, 13423530, 13425681, 13425770, 13425772, 13426591, 13426592, 13426619, 13426860, 13426861, 13426864, 13426865, 13426867, 13429369, 13429370, 13429377, 13430474, 13430477, 13430478, 13430479, 13430480, 13430481, 13430482, 13430483, 13431256, 13431258, 13431267, 13431268, 13433343, 13433344, 13433345, 13433346, 13434867, 13434868, 13434869, 13434870, 13434871, 13434872, 13435067, 13435071, 13435072, 13435073, 13436448, 13436453, 13436454, 13436464, 13437505, 13441981, 13441982, 13441984, 13443070, 13443152, 13443153, 13443154, 13445289, 13445292, 13445293, 13445294, 13445295, 13445296, 13445297, 13445298, 13445304, 13447453, 13447454, 13449409, 13449410, 13449422, 13449425, 13449429, 13449431, 13449432, 13449435, 13449438, 13449470, 13449472, 13450449, 13450450, 13450465, 13450466, 13451184, 13451185, 13451186, 13451188, 13451537, 13452349, 13452353, 13452357, 13452358, 13452359, 13452360, 13452361, 13452362, 13453628, 13455532, 13455533, 13458488, 13458493, 13458496, 13458498, 13458500, 13458501, 13461261, 13461262, 13461267, 13461268, 13461271, 13461272, 13461273, 13461289, 13461291, 13461293, 13461311, 13461312, 13461313, 13461314, 13461315, 13461317, 13461318, 13461319, 13462251, 13462253, 13462254, 13462255, 13462256, 13462257, 13462258, 13462259, 13462280, 13462281, 13462282, 13463262, 13463264, 13463273, 13463274, 13463275, 13463276, 13463277, 13463281, 13463302, 13463979, 13463980, 13464849, 13465021, 13465024, 13467782, 13467798, 13467799, 13467801, 13467815, 13469640, 13469641, 13469644, 13469645, 13469647, 13469649, 13469691, 13469725, 13471678, 13471679, 13472044, 13473245, 13473246, 13473353, 13476105, 13476836, 13476838, 13476839, 13476840, 13476841, 13476874, 13476875, 13476876, 13483280, 13484588, 13487024, 13491312, 13491313, 13491316, 13491909, 13495129, 13495130, 13495131, 13495132, 13495252, 13495253, 13495255, 13495257, 13495258, 13495259, 13495260, 13495261, 13495263, 13495266, 13495267, 13495268, 13495269, 13495271, 13495272, 13495275, 13496557, 13496558, 13496559, 13496560, 13496561, 13496562, 13496564, 13496565, 13496566, 13496567, 13499061, 13499062, 13501404, 13501417, 13501437, 13501440, 13501441, 13501446, 13501449, 13505648, 13505649, 13505650, 13505651, 13505652, 13505653, 13506361, 13506362, 13506374, 13506375, 13506377, 13506631, 13509369, 13509370, 13509371, 13509372, 13509373, 13509549, 13510117, 13510126, 13510161, 13510162, 13510163, 13510664, 13510699, 13510701, 13511540, 13512230, 13512232, 13512234, 13512235, 13512246, 13512247, 13512325, 13512326, 13515952, 13517656, 13517699, 13519761.13520988, 13521085, 13521086, 13523120, 13524198, 13524199, 13524247, 13524257, 13524488, 13526148, 13526149, 13530515, 13530517, 13530628, 13530629, 13530765, 13531064, 13533119, 13533124, 13533129, 13533132, 13533182, 13533183, 13533184, 13533185, 13534278, 13534383, 13534384, 13536875, 13542667, 13543504, 13543506, 13545759, 13545761, 13545890, 13548996, 13549036, 13551138, 13551227, 13555445, 13557692, 13558536, 13559017, 13559865, 13559866, 13559867, 13559868, 13562669, 13562751, 13563602, 13565951, 13568179, 13573178, 13573179, 13573881, 13574512, 13574513, 13574603, 13575571, 13575576, 13575578, 13575588, 13575589, 13575600, 13575603, 13575626, 13575633, 13575637, 13575638, 13575642, 13577652, 13577693, 13577795, 13577796, 13577797, 13579014, 13581712, 13582635, 13585820, 13585821, 13585822, 13585823, 13585825, 13585826, 13585827, 13585828, 13588075, 13588079, 13588095, 13591587, 13591588, 13591590, 13591667, 13593094, 13593095, 13593096, 13594247, 13594248, 13594249, 13594250, 13594251, 13595307, 13595901, 13596999, 13597001, 13597003, 13598061, 13598163, 13598652, 13598792, 13600383, 13600395, 13600596, 13604607, 13604617, 13604619, 13604782, 13605987, 13605994, 13606005, 13606009, 13607312, 13607313, 13607314, 13607315, 13607316, 13607317, 13607318, 13607319, 13607320, 13607321, 13609965, 13610013, 13611195, 13611268, 13611271, 13611573, 13612361, 13612362, 13612365, 13612368, 13612369, 13612374, 13612375, 13612376, 13612404, 13612405, 13614802, 13614803, 13614891, 13614893, 13614895, 13614900, 13614904, 13614917, 13617213, 13619249, 13619250, 13619256, 13621387, 13621388, 13621409, 13622795, 13622797, 13622799, 13622800, 13622838, 13624302, 13624304, 13624308, 13624315, 13624424, 13624441, 13624442, 13624443, 13624445, 13624448, 13625671, 13625675, 13625676, 13625677, 13625678, 13625698, 13625710, 13628001, 13628006, 13628009, 13628024, 13630267, 13630273, 13630274, 13630275, 13630276, 13631565, 13632671, 13633809, 13633810, 13633899, 13635880, 13635881, 13635883, 13635884, 13635885, 13635886, 13635887, 13635888, 13637557, 13638313, 13638314, 13638315, 13640593, 13640595, 13640596, 13640598, 13640978, 13640985, 13642263, 13642264, 13643687, 13643695, 13643696, 13644774, 13645006, 13647249, 13647263, 13647264, 13647269, 13647270, 13647271, 13647272, 13650654, 13654676, 13654683, 13654706, 13655649, 13656526, 13657555, 13657556, 13657567, 13658710, 13660423, 13660427, 13662199, 13666380, 13666381, 13666382, 13667279, 13667280, 13667281, 13668693, 13668699, 13668700, 13668701, 13670994, 13672195, 13672197, 13672198, 13674805, 13674844, 13674846, 13676174, 13676175, 13676590, 13676591, 13676592, 13676593, 13676594, 13676595, 13676596, 13676597, 13677337, 13677779, 13677780, 13677781, 13677782, 13677786, 13677788, 13678697, 13678698, 13678699, 13678709, 13678711, 13678713, 13678714, 13678717, 13678720, 13678728, 13678732, 13685191, 13685197, 13685212, 13686256, 13686453, 13686462, 13686464, 13686466, 13686467, 13686473, 13688069, 13688134, 13688135, 13688136, 13688137, 13688138, 13688140, 13688141, 13688142, 13688143, 13688144, 13688146, 13688148, 13688149, 13688187, 13688247, 13692496, 13693748, 13693753, 13693755, 13694081, 13694180, 13694181, 13694182, 13694183, 13694184, 13694185, 13694186, 13694187, 13694188, 13694189, 13694190, 13694200, 13694862, 13698186, 13698499, 13698502, 13698506, 13698507, 13698508, 13698509, 13698510, 13699886, 13700113, 13700115, 13700118, 13700120, 13700996, 13701497, 13701499, 13701520, 13701579, 13701653, 13701806, 13701811, 13701813, 13701946, 13701956, 13702117, 13702128, 13702129, 13702130, 13702131, 13702132, 13702134, 13702135, 13702136, 13702605, 13702606, 13702608, 13702609, 13703209, 13703211, 13703214, 13703217, 13703220, 13703222, 13703223, 13703230, 13703594, 13703595, 13703596, 13703920, 13703958, 13703960, 13703961, 13703963, 13703965, 13704442, 13704449, 13705396, 13705397, 13705398, 13705399, 13705400, 13705451, 13705456, 13705457, 13705465, 13705466, 13705467, 13705469.13705523, 13705526, 13705527, 13707408, 13707409, 13708267, 13708318, 13708319, 13708320, 13708868, 13710443, 13710449, 13710452, 13710525, 13710526, 13710949, 13710968, 13710980, 13710981, 13710986, 13711004, 13711035, 13711386, 13711387, 13711389, 13712791, 13712793, 13712796, 13713331, 13714429, 13714430, 13714713, 13714714, 13714715, 13715997, 13715998, 13718236, 13718239, 13718432, 13718433, 13720049, 13720050, 13720057, 13720058, 13720688, 13720689, 13720690, 13720691, 13720692, 13720693, 13720694, 13720768, 13720770, 13720772, 13720774, 13720779, 13720780, 13720815, 13720817, 13720850, 13720854, 13720880, 13720882, 13720884, 13720888, 13720899, 13722704, 13724491, 13725424, 13725476, 13725477, 13725486, 13725488, 13725490, 13725492, 13727061, 13727064, 13727065, 13728287, 13728289, 13728407, 13728410, 13728411, 13728737, 13728738, 13728739, 13730426, 13730428, 13730430, 13731594, 13734869, 13736216, 13736221, 13736222, 13736817, 13736826, 13738077, 13738078, 13738359, 13738360, 13738365, 13738366, 13744377, 13745847, 13746788, 13747825, 13749167, 13749169, 13749170, 13749171, 13749758, 13750847, 13750848, 13750855, 13750858, 13750861, 13750862, 13750864, 13750972, 13750973, 13750974, 13750976, 13750977, 13750984, 13750985, 13751024, 13754080, 13754081, 13754082, 13754083, 13754084, 13754085, 13754086, 13754087, 13754088, 13754089, 13754090, 13754091, 13754092, 13754093, 13754094, 13754095, 13754102, 13754103, 13754104, 13754105, 13754128, 13754129, 13754136, 13754137, 13754138, 13754139, 13754140, 13754141, 13754142, 13754143, 13754144, 13754145, 13754146, 13754169, 13754181, 13754188, 13754190, 13754191, 13755046, 13755049, 13755051, 13757435, 13757436, 13757437, 13757438, 13757445, 13757446, 13757447, 13757448, 13757449, 13757450, 13757493, 13757495, 13757497, 13757500, 13757503, 13757505, 13758099, 13758100, 13758116, 13758499, 13758502, 13758556, 13758557, 13758558, 13758560, 13758562, 13758589, 13758599, 13758600, 13758603, 13758604, 13758606, 13759343, 13759345, 13759348, 13759349, 13759361, 13759362, 13759363, 13759364, 13759365, 13759367, 13759370, 13759374, 13759376, 13759431, 13759432, 13759433, 13759440, 13760593, 13760594, 13760636, 13760637, 13760638, 13760640, 13760641, 13760642, 13760643, 13761761, 13762727, 13762729, 13763173, 13763178, 13763179, 13763181, 13763821, 13764660, 13764661, 13764662, 13765557, 13765559, 13766063, 13766064, 13766065, 13766732, 13766733, 13766734, 13766735, 13766736, 13766743, 13766746, 13766748, 13766755, 13767785, 13767786, 13767882, 13772515, 13778225, 13778509, 13778510, 13779649, 13783399, 13783400, 13783401, 13783402, 13783403, 13783404, 13783405, 13783406, 13783408, 13783409, 13783410, 13783413, 13783418, 13783424, 13783425, 13784528, 13784615, 13784620, 13784621, 13784622, 13790775, 13790776, 13790777, 13790780, 13790781, 13790782, 13790806, 13796585, 13796586, 13796587, 13797280, 13798047, 13804084, 13805270, 13805271, 13805272, 13810954, 13810955, 13813199, 13813207, 13813239, 13813240, 13816724, 13816740, 13816742, 13816744, 13816746, 13816748, 13816754, 13816760, 13816763, 13816765, 13816768, 13817517, 13819652, 13825296, 13825298, 13826519, 13826520, 13828100, 13828101, 13828103, 13829200, 13829201, 13829202, 13829203, 13829204, 13829205, 13829207, 13829209, 13829211, 13829215, 13829221, 13829223, 13829226, 13829351, 13829352, 13829353, 13829367, 13829368, 13829371, 13829372, 13829374, 13834405, 13834407, 13836030, 13836518, 13836906, 13836908, 13836910, 13837619, 13838564, 13838583, 13840721, 13840758, 13843260, 13843264, 13843266, 13843308, 13844076, 13846556, 13846557, 13846559, 13846560, 13847266, 13847267, 13848153, 13848216, 13848218, 13848309, 13848310, 13850957, 13850958, 13852482, 13852485, 13852495, 13852498, 13852499, 13854722, 13854723, 13855411, 13856636, 13857747, 13857748, 13859236, 13859237, 13860463, 13860469, 13860474, 13861590, 13861593, 13861596, 13861646, 13863550, 13863553, 13863554, 13863555, 13863556, 13864731, 13864750, 13865651, 13865654.13866620, 13868034, 13868064, 13868069, 13868071, 13868072, 13868073, 13868702, 13868707, 13869427, 13869428, 13869429, 13870357, 13871476, 13872210, 13872571, 13872652, 13872654, 13872657, 13872791, 13874162, 13874163, 13874176, 13874183, 13874185, 13875098, 13875104, 13875105, 13875723, 13875724, 13875725, 13875727, 13875728, 13875729, 13875731, 13875733, 13875736, 13875737, 13875747, 13875748, 13875749, 13875750, 13876612, 13876629, 13876630, 13879050, 13879907, 13880431, 13881070, 13881173, 13882164, 13882509, 13883128, 13883129, 13883968, 13885025, 13885102, 13885103, 13890200, 13890201, 13890837, 13891832, 13892268, 13892269, 13892270, 13892271, 13892272, 13892273, 13892274, 13892275, 13892284, 13892285, 13892286, 13892287, 13894426, 13894996, 13896068, 13896070, 13896768, 13902707, 13902708, 13902778, 13902799, 13902801, 13902802, 13912829, 13912830, 13912844, 13912907, 13912908, 13912910, 13914685, 13914686, 13914687, 13914688, 13914689, 13914690, 13914705, 13919634, 13919636, 13919637, 13919645, 13919651, 13919653, 13919684, 13919685, 13919715, 13919716, 13919717, 13919718, 13920177, 13920181, 13920184, 13920186, 13920190, 13920931, 13927118, 13927999, 13928000, 13928002, 13930177, 13930178, 13930179, 13930184, 13930185, 13932106, 13932109, 13933755, 13933796, 13933797, 13933798, 13933799, 13933824, 13933829, 13935093, 13937581, 13941416, 13943095, 13943096, 13943126, 13943127, 13943549, 13943595, 13944020, 13944021, 13944095, 13944163, 13944169, 13944170, 13946361, 13947174, 13947176, 13947439, 13947440, 13948941, 13948943, 13948944, 13949824, 13949827, 13949828, 13951409, 13951410, 13951411, 13951412, 13951413, 13951878, 13951925, 13951938, 13951939, 13951940, 13951941, 13951956, 13951957, 13951958, 13951963, 13951964, 13951965, 13952018, 13952019, 13952020, 13958616, 13960823, 13961657, 13961658, 13965072, 13965073, 13966270, 13971997, 13973533, 13973534, 13973536, 13985499, 13985500, 13987461, 13992443, 13992498, 13993281, 13993299, 13997722, 13998149, 13998150, 13998505, 13998506, 13998507, 13998524, 13998525, 13998551, 13998553, 13998554, 14000714, 14000715, 14005889, 14006846, 14006847, 14006848, 14006901, 14006911, 14006912, 14006914, 14006916, 14006918, 14006922, 14006925, 14006930, 14006931, 14010149, 14010150, 14010158, 14010160, 14010162, 14010163, 14010167, 14010176, 14010177, 14010179, 14010181, 14010190, 14010192, 14010193, 14010196, 14010197, 14010201, 14010205, 14014112, 14016102, 14019211, 14019221, 14019223, 14024288, 14027067, 14027069, 14027071, 14027074, 14029475, 14029476, 14029480, 14030744, 14030747, 14030751, 14030754, 14030769, 14033927, 14033943, 14034050, 14034052, 14034556, 14034584, 14034585, 14034586, 14036676, 14036678, 14036853, 14042864, 14042868, 14043989, 14043990, 14054009, 14062188, 14065578, 14065580, 14065587, 14066296, 14072940, 14072941, 14072942, 14072943, 14072944, 14072945, 14072946, 14072947, 14072948, 14072958, 14072994, 14072995, 14072996, 14072997, 14076125, 14076127, 14076129, 14076131, 14076133, 14080218, 14080221, 14080222, 14080223, 14080224, 14080226, 14080234, 14080235, 14080237, 14080238, 14080245, 14080246, 14080254, 14080262, 14080263, 14080267, 14080270, 14080271, 14080272, 14080279, 14080282, 14080286, 14081825, 14085341, 14085410, 14085411, 14085412, 14085414, 14085415, 14085418, 14085419, 14085427, 14085429, 14085436, 14085437, 14085438, 14085439, 14085442, 14085445, 14085446, 14085447, 14085457, 14085458, 14085459, 14085460, 14085461, 14085462, 14085463, 14089260, 14092620, 14092621, 14092622, 14092623, 14095370, 14101525, 14101632, 14101633, 14101661, 14101879, 14101880, 14101881, 14101882, 14101883, 14101884, 14103722, 14105036, 14105037, 14112256, 14119132, 14119133, 14133109, 14136383, 14136465, 14136467, 14143871, 14145717, 14149623, 14150632, 14150633, 14150634, 14150641, 14150650, 14152452, 14152453, 14152454, 14152455, 14152456, 14153243, 14153248, 14153249, 14153251, 14153252, 14153259, 14153878, 14154094, 14154095, 14154096, 14154146, 14154148.14154150, 14154151, 14154153, 14154154, 14155758, 14155759, 14155760, 14155768, 14155769, 14155770, 14155771, 14155772, 14155773, 14155774, 14155780, 14155781, 14155782, 14155783, 14156779, 14157187, 14157529, 14158296, 14159152, 14159153, 14159154, 14159155, 14159424, 14159425, 14160240, 14160241, 14160242, 14160248, 14160249, 14164432, 14166690, 14167865, 14172687, 14172691, 14173733, 14174834, 14174836, 14174838, 14177117, 14177220, 14183582, 14184102, 14184104, 14185636, 14186458, 14188214, 14188215, 14188216, 14191550, 14191552, 14192864, 14192865, 14192939, 14192941, 14192943, 14192945, 14192947, 14192949, 14192956, 14194869, 14194870, 14194871, 14194882, 14197746, 14202655, 14202828, 14210783, 14210850, 14210851, 14214126, 14214128, 14215599, 14219806, 14225880, 14225885, 14225886, 14225960, 14225962, 14225973, 14228622, 14231577, 14232957, 14232958, 14234340, 14242418, 14243712, 14243797, 14249065, 14250533, 14252203, 14256158, 14256160, 14256162, 14257049, 14257051, 14258497, 14258498, 14258508, 14258509, 14258510, 14258512, 14258514, 14258516, 14258518, 14258650, 14265207, 14265209, 14265354, 14265357, 14267723, 14269817, 14269875, 14269896, 14270078, 14270079, 14270080, 14270252, 14270253, 14275239, 14281396, 14281397, 14281398, 14281402, 14281404, 14281406, 14281408, 14281410, 14281412, 14281414, 14281417, 14281426, 14281428, 14281430, 14281432, 14281434, 14281436, 14284202, 14284203, 14284205, 14284207, 14290737, 14292901, 14292902, 14298754, 14298758, 14298760, 14298812, 14298877, 14308120, 14308121, 14308136, 14308137, 14308138, 14308139, 14308140, 14308141, 14313131, 14313205, 14313206, 14313605, 14313606, 14313607, 14313608, 14313609, 14313612, 14313613, 14313615, 14313617, 14313618, 14313619, 14324046, 14324047, 14325471, 14325474, 14325477, 14325485, 14329293, 14329418, 14334602, 14338624, 14339779, 14339780, 14339781, 14339794, 14339795, 14342953, 14350107, 14350109, 14350112, 14350113, 14350114, 14350175, 14350176, 14350177, 14350178, 14350179, 14350180, 14350181, 14350183, 14350184, 14350186, 14350471, 14350472, 14350473, 14350475, 14350476, 14354943, 14354944, 14354945, 14355750, 14355751, 14355752, 14373486, 14373488, 14378336, 14378337, 14386457, 14386458, 14386459, 14386460, 14386461, 14386462, 14386463, 14386464, 14386465, 14386466, 14386467, 14386468, 14386469, 14386470, 14386471, 14386472, 14386473, 14386474, 14386475, 14386476, 14392725, 14394972, 14395958, 14400966, 14400968, 14406548, 14410198, 14412233, 14412234, 14412644, 14425437, 14425438, 14425439, 14425445, 14425449, 14425451, 14425453, 14425462, 14425463, 14425464, 14425465, 14425466, 14425473, 14425475, 14425476, 14425477, 14425478, 14425479, 14425549, 14425554, 14425555, 14425556, 14425558, 14427002, 14427004, 14427005, 14427007, 14428961, 14432114, 14436316, 14437874, 14439035, 14442247, 14446215, 14446280, 14446673, 14447422, 14447441, 14447442, 14447443, 14448356, 14448357, 14448358, 14448359, 14448360, 14448361, 14456049, 14456058, 14456059, 14456060, 14457920, 14460711, 14460712, 14462580, 14462582, 14463766, 14463767, 14463768, 14463769, 14463770, 14463771, 14463772, 14463773, 14466739, 14467794, 14467803, 14470299, 14470604, 14470606, 14470608, 14470610, 14471052, 14471053, 14471054, 14472451, 14472452, 14472453, 14472477, 14472478, 14472479, 14472480, 14472484, 14472485, 14472486, 14473340, 14474505, 14475087, 14477670, 14479627, 14479629, 14479630, 14479631, 14479781, 14479782, 14479783, 14479784, 14479785, 14479965, 14479966, 14483192, 14483595, 14483596, 14483597, 14483598, 14483601, 14483602, 14483603, 14484473, 14485771, 14490416, 14494587, 14494589, 14494593, 14494597, 14496613, 14496618, 14499906, 14501162, 14501164, 14503130, 14503131, 14503457, 14503459, 14505043, 14507877, 14507879, 14507881, 14511343, 14511792, 14525068, 14527540, 14528361, 14529067, 14529068, 14529097, 14529107, 14533576, 14536107, 14538218, 14541553, 14546230, 14547484, 14552394, 14557543, 14557568, 14557569, 14558485, 14558486, 14558487, 14558488.14558490, 14558491, 14558534, 14559087, 14560643, 14561459, 14561460, 14561461, 14561502, 14565985, 14568021, 14571262, 14572199, 14572200, 14575680, 14577392, 14577394, 14583203, 14583205, 14586237, 14586238, 14589356, 14589357, 14589358, 14589359, 14589360, 14589363, 14591863, 14593356, 14593358, 14594579, 14594580, 14594581, 14599882, 14605415, 14605416, 14606232, 14606233, 14608047, 14608048, 14608049, 14610039, 14610197, 14610198, 14610203, 14617429, 14617430, 14617433, 14617434, 14630104, 14630105, 14630107, 14633851, 14634580, 14634581, 14635629, 14635630, 14636807, 14640604, 14643172, 14643710, 14643711, 14643712, 14643713, 14648231, 14648459, 14648462, 14649038, 14649039, 14649040, 14649049, 14653331, 14653426, 14653427, 14653428, 14655017, 14655018, 14655019, 14655020, 14655022, 14657398, 14658759, 14658760, 14658761, 14659661, 14662133, 14662135, 14662136, 14662137, 14662140, 14662144, 14662145, 14662148, 14662149, 14662150, 14662190, 14666874, 14668951, 14668952, 14668953, 14668954, 14668957, 14671345, 14672086, 14672087, 14677081, 14686708, 14686709, 14687299, 14689052, 14689054, 14689056, 14689058, 14689848, 14689849, 14690745, 14690746, 14695679, 14697338, 14699772, 14712597, 14713428, 14713429, 14713430, 14718694, 14724741, 14724742, 14728013, 14728015, 14728144, 14729448, 14730317, 14730319, 14730321, 14730324, 14735045, 14735047, 14735052, 14739114, 14739140, 14740320, 14740323, 14740936, 14740937, 14742672, 14743700, 14743701, 14743702, 14743705, 14745421, 14745422, 14745423, 14745424, 14745425, 14745426, 14753670, 14753671, 14753672, 14753673, 14753674, 14753675, 14753676, 14753775, 14753777, 14753779, 14754563, 14755790, 14756802, 14770827, 14779713, 14793732, 14795111, 14797740, 14797934, 14800123, 14801280, 14801286, 14801787, 14802401, 14802402, 14802403, 14802404, 14802405, 14807724, 14808743, 14817149, 14823190, 14823192, 14823700, 14823702, 14827223, 14827931, 14831139, 14831141, 14832125, 14832127, 14832128, 14832130, 14832131, 14832135, 14832136, 14832138, 14832139, 14832141, 14832142, 14832145, 14832148, 14832151, 14832153, 14832155, 14832156, 14832158, 14832159, 14832161, 14832162, 14832164, 14832166, 14832167, 14832169, 14832170, 14832172, 14832173, 14832181, 14832188, 14832190, 14832193, 14832599, 14832600, 14836423, 14836425, 14836427, 14841010, 14841011, 14841118, 14841120, 14849245, 14849246, 14849247, 14849248, 14855393, 14860499, 14860501, 14860504, 14863277, 14877901, 14878636, 14878638, 14878640, 14878643, 14879832, 14879834, 14879837, 14879841, 14879844, 14879846, 14879859, 14879860, 14879861, 14879862, 14879863, 14879864, 14879865, 14886461, 14886462, 14886464, 14886465, 14886466, 14887229, 14891491, 14897829, 14897834, 14898339, 14898343, 14898344, 14898345, 14898995, 14898996, 14900323, 14900327, 14900328, 14900329, 14902385, 14902387, 14902389, 14908405, 14908406, 14908843, 14908844, 14914163, 14914164, 14923143, 14923145, 14923147, 14923150, 14923159, 14923301, 14923305, 14923310, 14923327, 14927755, 14927756, 14927758, 14927760, 14927761, 14937423, 14938025, 14938029, 14938031, 14938035, 14938036, 14938037, 14938039, 14938040, 14938041, 14947849, 14947850, 14947851, 14947852, 14951339, 14951341, 14951343, 14951346, 14951355, 14951356, 14951358, 14951361, 14951363, 14951365, 14951368, 14951370, 14960370, 14964153, 14964156, 14964157, 14964341, 14970760, 14970762, 14970765, 14972128, 14972133, 14972134, 14972135, 14972139, 14972146, 14972149, 14972151, 14975833, 14975837, 14975838, 14975839, 14975844, 14975845, 14975847, 14975848, 14990207, 14990220, 14992709, 14992710, 14995980, 15003457, 15014108, 15018442, 15028923, 15033045, 15033067, 15033068, 15034878, 15036015, 15036016, 15036017, 15036018, 15036080, 15036081, 15036082, 15036083, 15036084, 15042046, 15042049, 15046773, 15046775, 15047888, 15047890, 15047892, 15047896, 15047901, 15047904, 15051002, 15051003, 15051004, 15057103, 15063920, 15064917, 15064938, 15064939, 15064940, 15064941, 15079066, 15079067, 15079069, 15079078.15091874, 15096845, 15096860, 15101505, 15101507, 15101508, 15101509, 15101510, 15101511, 15101611, 15103481, 15103482, 15103483, 15103484, 15105048, 15107218, 15107219, 15111539, 15122077, 15139294, 15147745, 15147746, 15149713, 15154981, 15154983, 15154986, 15154988, 15157910, 15157911, 15165044, 15165056, 15167672, 15169934, 15169936, 15169937, 15174680, 15185453, 15196206, 15198028, 15199756, 15202573, 15205253, 15206076, 15210294, 15211172, 15211173, 15212980, 15212982, 15212983, 15213022, 15213024, 15213026, 15213032, 15213610, 15213618, 15222366, 15222367, 15222371, 15222373, 15223200, 15223206, 15223420, 15223422, 15223427, 15228571, 15229859, 15229860, 15230031, 15248356, 15253485, 15254530, 15254531, 15258736, 15271248, 15271250, 15275951, 15275952, 15290920, 15292305, 15295606, 15298205, 15318477, 15318478, 15318480, 15318481, 15318482, 15318483, 15318491, 15321561, 15321563, 15324655, 15324656, 15324657, 15324658, 15324659, 15324660, 15324668, 15324670, 15324671, 15328692, 15331797, 15331799, 15331801, 15341065, 15341205, 15341207, 15341209, 15341212, 15341215, 15344341, 15344342, 15344343, 15344344, 15344345, 15344362, 15351759, 15356810, 15356812, 15356814, 15356816, 15356818, 15358155, 15366775, 15366776, 15366779, 15366780, 15366782, 15366783, 15366784, 15371702, 15371703, 15376140, 15380785, 15380787, 15380788, 15381473, 15381582, 15382380, 15382381, 15382804, 15382809, 15393446, 15393514, 15393521, 15393915, 15395452, 15395453, 15395473, 15396688, 15396689, 15396690, 15396691, 15396692, 15396693, 15396694, 15396696, 15396697, 15396698, 15396700, 15397006, 15397007, 15397009, 15397011, 15397014, 15397015, 15397016, 15397017, 15397018, 15397077, 15398646, 15398647, 15402312, 15402313, 15405926, 15405931, 15405934, 15405937, 15406215, 15408957, 15422984, 15426474, 15426475, 15426478, 15426479, 15429500, 15429504, 15429833, 15435510, 15435532, 15435533, 15435534, 15435536, 15435538, 15435540, 15435542, 15440148, 15440149, 15440150, 15441301, 15441302, 15442057, 15442058, 15442059, 15447907, 15447908, 15447909, 15448269, 15454464, 15454465, 15455679, 15455681, 15455684, 15455692, 15459248, 15461851, 15467554, 15467555, 15472530, 15472532, 15473649, 15473651, 15473664, 15473666, 15473669, 15473671, 15473673, 15473686, 15473688, 15477283, 15480766, 15481789, 15481794, 15481796, 15481798, 15481801, 15481803, 15492609, 15492613, 15492617, 15492618, 15492634, 15492638, 15493878, 15500445, 15513457, 15513458, 15513459, 15513460, 15513461, 15513462, 15513464, 15513465, 15513466, 15513467, 15513468, 15513469, 15513471, 15513479, 15513488, 15513489, 15516319, 15528163, 15534111, 15538927, 15538929, 15549057, 15549060, 15550368, 15550370, 15554800, 15554801, 15554802, 15554803, 15554804, 15554808, 15563534, 15563915, 15568554, 15568917, 15569594, 15570622, 15572940, 15575710, 15575711, 15575790, 15581099, 15581100, 15582853, 15591365, 15593200, 15604510, 15622739, 15622749, 15622751, 15622753, 15622755, 15622763, 15622767, 15622770, 15622772, 15622775, 15622806, 15622810, 15622812, 15622815, 15622817, 15622819, 15622822, 15622825, 15622827, 15622829, 15622831, 15622834, 15622838, 15622840, 15622844, 15622848, 15622850, 15622852, 15622856, 15622858, 15622860, 15622862, 15622865, 15622867, 15622869, 15627475, 15630844, 15630845, 15630849, 15668480, 15683478, 15686336, 15702122, 15702123, 15714232, 15720455, 15720457, 15720459, 15720461, 15720464, 15720466, 15720469, 15720471, 15720474, 15720476, 15720479, 15720481, 15737978, 15737979, 15770266, 15770281, 15772857, 15788716, 15788766, 15789401, 15789402, 15791695, 15792207, 15792210, 15798754, 15813793, 15813794, 15831050, 15835152, 15835154, 15835156, 15845636, 15852321, 15854270, 15854271, 15854272, 15854273, 15854846, 15854847, 15854851, 15854852, 15854853, 15854859, 15854882, 15855486, 15855995, 15856109, 15856110, 15856180, 15856933, 15856934, 15856939, 15856941, 15856953, 15856955, 15856957, 15857842, 15857846, 15857847, 15857848, 15857849, 15857850, 15857851, 15857852.15858073, 15858075, 15860312, 15860313, 15860422, 15860424, 15860425, 15860427, 15860932, 15862134, 15862135, 15863158, 15863161, 15863163, 15863184, 15863185, 15863186, 15863188, 15863189, 15864816, 15866021, 15868996, 15869138, 15869444, 15869445, 15869446, 15869448, 15869449, 15869451, 15869454, 15869457, 15869460, 15895368, 15895399, 15895400, 15905075, 15907775, 15907776, 15910452, 15910453, 15910454, 15910455, 15910456, 15910457, 15918235, 15918236, 15920351, 15921886, 15924489, 15924490, 15924491, 15925085, 15925086, 15927750, 15928628, 15928629, 15930186, 15930187, 15932256, 15932257, 15934607, 15934608, 15934656, 15937821, 15938634, 15942624, 15945070, 15945071, 15945745, 15949851, 15954768, 15958166, 15960803, 15976015, 15984871, 15984872, 15984874, 16001210, 16001212, 16001214, 16001216, 16001218, 16001220, 16012509, 16012511, 16052917, 16069684, 16079413, 16083959, 16083960, 16173446, 16187325, 16187326, 16187327, 710008244, 710012860, 710012861, 710013384, 710013385, 710013387, 710019634, 710023893, 710024516, 710024678, 710024679, 710025019, 710039922, 710051515, 710080317, 710080318, 710080604, 710080607, 710084186, 710085009, 710099037, 710100576, 710101030, 710110161, 710121947, 710137725, 710151434, 710151435, 710151436, 710153312, 710153664, 710153666, 710155295, 710155386, 710155545, 710155830, 710156019, 710156103, 710156104, 710156275, 710156276, 710156327, 710156397, 710156472, 710156539, 710156585, 710156586, 710166514, 710166694, 710171427, 710173990, 8000001026, 8000002255, 8000002449, 8000002450, 8000002451, 8000002491, 8000002492, 8000004555, 8000004704, 8000004936, 8000006756, 8000006864, 8000006866, 8000006867, 8000007042, 8000007223, 8000007387, 8000007388, 8000007389, 8000007390, 8000009968, 8000009969, 8000010270, 8000010271, 8000010272, 8000011487, 8000011785, 8000017421, 8000017422, 8000017488, 8000017489, 8000017865, 8000017866, 8000018004, 8000018005, 8000018006, 8000018007, 8000018008, 8000018009, 8000018010, 8000018011, 8000018012, 8000018013, 8000018014, 8000018625, 8000023138, 8000024064, 8000024065, 8000024201, 8000024460, 8000024579, 8000024580, 8000025074, 8000025222, 8000025223, 8000025224, 8000027247, 8000027248, 8000027477, 8000027478, 8000028937, 8000028938, 8000028939, 8000028940, 8000029924, 8000030028, 8000030433, 8000030434, 8000030978, 8000030979, 8000030980, 8000031228, 8000031229, 8000031230, 8000031231, 8000031340, 8000031341, 8000031342, 8000031343, 8000031344, 8000031363, 8000031590, 8000033140, 8000033141, 8000033706, 8000035114, 8000035115, 8000035750, 8000035833, 8000037326, 8000037327, 8000037328, 8000037329, 8000037478, 8000039131, 8000039133, 8000039134, 8000039135, 8000040554, 8000040815, 8000040816, 8000040817, 8000040818, 8000040819, 8000041146, 8000041147, 8000042850, 8000042953, 8000042954, 8000043711, 8000044291, 8000044517, 8000044809, 8000047011, 8000047012, 8000047013, 8000047014, 8000047015, 8000047016, 8000047521, 8000047776, 8000049081, 8000050319, 8000050675, 8000050676, 8000050677, 8000050678, 8000050679, 8000050680, 8000052091, 8000052092, 8000052093, 8000052865, 8000053409, 8000053901, 8000054554, 8000054557, 8000055548, 8000055549, 8000055550, 8000055551, 8000055960, 8000057082, 8000057083, 8000058226, 8000058234, 8000059217, 8000059218, 8000060479, 8000060726, 8000061481, 8000062360, 8000065365, 8000065439, 8000066845, 8000066846, 8000066847, 8000066848, 8000066849, 8000067786, 8000067787, 8000067871, 8000068382, 8000068458, 8000069448, 8000070244, 8000070245, 8000070246, 8000070247, 8000070248, 8000070538, 8000071575, 8000071576, 8000071577, 8000071578, 8000071693, 8000071694, 8000071695, 8000072880, 8000073428, 8000073429, 8000073430, 8000074804, 8000075227, 8000077162, 8000077163, 8000077164, 8000077165, 8000077178, 8000077179, 8000077180, 8000077181, 8000077300, 8000077301, 8000077302, 8000077303, 8000077304, 8000077305, 8000077306, 8000077307, 8000077473, 8000078086, 8000078089, 8000078090, 8000078093, 8000078094, 8000080311, 8000080312, 8000080313, 8000080314, 8000080315, 8000080316, 8000080317, 8000080319, 8000080392, 8000080505, 8000080702, 8000081426, 8000081759, 8000082404, 8000082893, 8000083582, 8000083583, 8000083944, 8000084392, 8000084393, 8000084394, 8000086095, 8000086342, 8000086343, 8000087010, 8000087456, 8000087457, 8000087458, 8000087459, 8000087501, 8000088188, 8000088438, 8000088578, 8000089042, 8000089729, 8000090022, 8000090023, 8000091120, 8000091121, 8000091122, 8000091123, 8000091124, 8000091125, 8000091126, 8000091793, 8000091794, 8000092926, 8000092928, 8000093218, 8000093219, 8000093220, 8000093221, 8000093222, 8000093409, 8000093410, 8000094256, 8000094257, 8000094985, 8000094987, 8000094988, 8000095112, 8000095113, 8000095114, 8000095115, 8000095116, 8000095117, 8000095777, 8000096107, 8000096369, 8000096370, 8000096371, 8000096372, 8000096373, 8000096374, 8000096481, 8000096944, 8000098329, 8000098330, 8000098702, 8000099694, 8000099695, 8000100037, 8000100038, 8000100039, 8000100040, 8000100425, 8000100654, 8000100750, 8000100858, 8000100859, 8000100860, 8000100862, 8000100864, 8000101404, 8000101642, 8000101754, 8000102347, 8000102348, 8000103937, 8000104029, 8000104030, 8000104266, 8000104267, 8000104268, 8000105112, 8000105113, 8000105114, 8000105115, 8000106736, 8000107229, 8000107537, 8000107538, 8000107569, 8000107575, 8000107725, 8000107726, 8000107797, 8000107798, 8000107799, 8000107800, 8000107801, 8000108400, 8000108406, 8000108845, 8000108846, 8000108847, 8000108848, 8000109325, 8000109326, 8000109327, 8000109328, 8000109825, 8000109826, 8000109827, 8000110985, 8000110987, 8000110989, 8000110990, 8000110991, 8000110992, 8000110993, 8000110995, 8000111585, 8000111796, 8000111797, 8000111980, 8000112223, 8000112477, 8000112478, 8000112722, 8000113415, 8000113416, 8000113417, 8000114001, 8000114347, 8000114348, 8000115246, 8000115337, 8000115339, 8000115340, 8000115341, 8000115688, 8000115792, 8000115925, 8000115926, 8000115927, 8000116166, 8000116167, 8000116198, 8000116771, 8000116772, 8000116773, 8000116778, 8000116779, 8000116780, 8000116781, 8000116782, 8000116783, 8000116853, 8000116931, 8000116932, 8000117110, 8000117415, 8000117416, 8000117417, 8000117418, 8000117419, 8000117420, 8000117421, 8000117678, 8000117679, 8000117978, 8000117979, 8000117980, 8000117981, 8000117982, 8000118210, 8000118211, 8000118212, 8000118979, 8000118980, 8000118993, 8000118994, 8000119117, 8000119199, 8000119353, 8000119354, 8000119357, 8000119742, 8000119743, 8000119744, 8000119745, 8000119746, 8000119747, 8000119748, 8000119749, 8000119750, 8000119751, 8000119927, 8000119928, 8000120102, 8000120104, 8000121072, 8000121175, 8000121176, 8000121385, 8000121797, 8000121798, 8000122405, 8000122549, 8000122599, 8000123097, 8000123098, 8000123099, 8000123100, 8000123101, 8000123102, 8000123103, 8000124202, 8000124203, 8000124204, 8000124205, 8000124469, 8000124470, 8000124476, 8000124489, 8000124804, 8000124873, 8000125072, 8000126290, 8000126291, 8000126292, 8000126330, 8000126331, 8000126332, 8000126333, 8000126334, 8000126335, 8000126336, 8000126337, 8000129060, 8000129061, 8000129570, 8000129582, 8000129583, 8000129584, 8000129718, 8000130123, 8000130124, 8000130572, 8000131140, 8000131141, 8000131142, 8000131143, 8000131144, 8000131410, 8000131411, 8000131412, 8000131571, 8000132198, 8000132317, 8000132318, 8000132319, 8000132320, 8000132638, 8000132639, 8000132667, 8000132668, 8000132669, 8000132670, 8000132671, 8000132672, 8000132673, 8000132675, 8000132676, 8000132677, 8000132678, 8000132680, 8000132721, 8000132748, 8000132780, 8000132781, 8000133328, 8000133329.
FEI Number 2016493
Recalling Firm/
Manufacturer
CareFusion 303, Inc.
10020 Pacific Mesa Blvd
San Diego CA 92121-4386
Manufacturer Reason
for Recall
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
FDA Determined
Cause 2
Nonconforming Material/Component
ActionBD issued an Urgent Medical Device Product Advisory to its consignees on 02/14/2023 via letter. The notice explained the problems with the device, health hazards, and requested the following actions be taken: "1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 4. If you have specific questions about this letter, please contact 1-866-583-8783." The firm issued a second notice on 01/09/2024 with the same information.
Quantity in Commerce5056 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-