| | Class 2 Device Recall ID NOW Influenza A/B 2 24T |  |
| Date Initiated by Firm | March 04, 2025 |
| Date Posted | April 15, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1575-2025 |
| Recall Event ID |
96500 |
| 510(K)Number | K232775 |
| Product Classification |
Influenza A And Influenza B Multiplex Nucleic Acid Assay - Product Code OZE
|
| Product | Brand Name: ID NOW Influenza A/B 2 24T
Product Name: ID NOW Influenza A/B 2 24T
Model/Catalog Number: 427-000
Software Version: N/A
Product Description: Influenza A And Influenza B Multiplex Nucleic Acid Assay; Real Time Nucleic Acid
Component: No |
| Code Information |
Model 427-000; GTIN/DI 10811877010422; Lot 000M889669
Model 427-000; GTIN/DI 10811877010422; Lot 000M890889
Model 427-000; GTIN/DI 10811877010422; Lot 000M916327
Model 427-000; GTIN/DI 10811877010422; Lot 000M916649
Model 427-000; GTIN/DI 10811877010422; Lot 000M916672
Model 427-000; GTIN/DI 10811877010422; Lot 000M916724
Model 427-000; GTIN/DI 10811877010422; Lot 000M917131
Model 427-000; GTIN/DI 10811877010422; Lot 000M917164
Model 427-000; GTIN/DI 10811877010422; Lot 000M917180
Model 427-000; GTIN/DI 10811877010422; Lot 000M917508
Model 427-000; GTIN/DI 10811877010422; Lot 000M917581
Model 427-000; GTIN/DI 10811877010422; Lot 000M919907
Model 427-000; GTIN/DI 10811877010422; Lot 000M921626
Model 427-000; GTIN/DI 10811877010422; Lot 000M921636
Model 427-000; GTIN/DI 10811877010422; Lot 000M921728
Model 427-000; GTIN/DI 10811877010422; Lot 000M923485
Model 427-000; GTIN/DI 10811877010422; Lot 000M926234
Model 427-000; GTIN/DI 10811877010422; Lot 000M926370
Model 427-000; GTIN/DI 10811877010422; Lot 000M927991
Model 427-000; GTIN/DI 10811877010422; Lot 000M928421
Model 427-000; GTIN/DI 10811877010422; Lot 000M929248
Model 427-000; GTIN/DI 10811877010422; Lot 000M931072
Model 427-000; GTIN/DI 10811877010422; Lot 000M934181
Model 427-000; GTIN/DI 10811877010422; Lot 000M939117
Model 427-000; GTIN/DI 10811877010422; Lot 000M939160
Model 427-000; GTIN/DI 10811877010422; Lot 000M939186
Model 427-000; GTIN/DI 10811877010422; Lot 000M939468
Model 427-000; GTIN/DI 10811877010422; Lot 000M939480
Model 427-000; GTIN/DI 10811877010422; Lot 000M952883
|
| FEI Number |
1221359
|
Recalling Firm/ Manufacturer |
Abbott Diagnostics Scarborough, Inc.
10 Southgate Rd
Scarborough
ME
04074-8303
|
| For Additional Information Contact | Vicki Assardo 224-6681505 |
Manufacturer Reason for Recall | the impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use. |
FDA Determined Cause 2 | Process change control |
| Action | On March 4, 2025 PRODUCT REMOVAL NOTICE letters were sent to distributors.
Necessary Actions to be Taken by Distributor:
1. If you have impacted inventory held in stock, then:
"Immediately quarantine impacted product per your local procedure.
"Complete and return the Distributor Acknowledgment Form (Form must be completed, signed and returned to receive replacement product).
"Contact Customer Service to arrange return of your product by Abbott.
2. If you have forwarded the product listed above to others within your distribution network, then:
"Inform them of this Product Removal Notice and provide to them a copy of this notice and request they take the necessary action.
3. If you have forwarded the impacted product to your customers, then:
"Provide customer list to Abbott based on the product information in Appendix II as outlined in the distribution agreement within 48 hours of receipt of this Product Removal Notice.
"Abbott will notify customers of this Product Removal Notice.
4. If you do not have any inventory of the impacted lot(s), then:
"All product lots not identified in Appendix I can continue to be used.
If you have questions regarding this information, U.S. Customers please contact Sedgwick at 1 888 671 8858 (available 8am 6pm EST, Monday - Friday).
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178).
If you have experienced any patient or user injury associated with this Removal Notice, please immediately report the event to your local area Customer Service.
****Update 05/08/2025****
On April 29, 2025 Updated PRODUCT REMOVAL NOTICE letters were sent to distributors notifying of additional lots that were included in the recall. |
| Quantity in Commerce | 2,776,416 eaches |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OZE
|
|
|
|