| Date Initiated by Firm | March 04, 2025 |
| Date Posted | April 15, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1571-2025 |
| Recall Event ID |
96519 |
| 510(K)Number | K221925 |
| Product Classification |
Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets From Clinical Specimens In Near-Patient Settings - Product Code QWR
|
| Product | Brand Name: ID NOW COVID-19 2.0 24T
Product Name: ID NOW COVID-19 2.0 24T
Model/Catalog Number: 192-000
Software Version: N/A
Product Description: Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets
Component: No |
| Code Information |
Model No 192-000; GTIN/DI 10811877011351; Lot 000M884910
Model No 192-000; GTIN/DI 10811877011351; Lot 000M885121
Model No 192-000; GTIN/DI 10811877011351; Lot 000M911157
Model No 192-000; GTIN/DI 10811877011351; Lot 000M913227
Model No 192-000; GTIN/DI 10811877011351; Lot 000M913286
Model No 192-000; GTIN/DI 10811877011351; Lot 000M913303
Model No 192-000; GTIN/DI 10811877011351; Lot 000M913641
Model No 192-000; GTIN/DI 10811877011351; Lot 000M914188
Model No 192-000; GTIN/DI 10811877011351; Lot 000M914221
Model No 192-000; GTIN/DI 10811877011351; Lot 000M914852
Model No 192-000; GTIN/DI 10811877011351; Lot 000M915373
Model No 192-000; GTIN/DI 10811877011351; Lot 000M916267
Model No 192-000; GTIN/DI 10811877011351; Lot 000M921404
Model No 192-000; GTIN/DI 10811877011351; Lot 000M922204
Model No 192-000; GTIN/DI 10811877011351; Lot 000M922578
Model No 192-000; GTIN/DI 10811877011351; Lot 000M922601
Model No 192-000; GTIN/DI 10811877011351; Lot 000M922634
Model No 192-000; GTIN/DI 10811877011351; Lot 000M924048
Model No 192-000; GTIN/DI 10811877011351; Lot 000M925615
Model No 192-000; GTIN/DI 10811877011351; Lot 000M938692
Model No 192-000; GTIN/DI 10811877011351; Lot 00M922634A
Model No 192-000; GTIN/DI 10811877011351; Lot 00M924048A
****Updated 5/8/2025****
04/29/2025 (Adding Additional Lots):
Model No 192-000; GTIN/DI 10811877011351; Lot 000M913147
Model No 192-000; GTIN/DI 10811877011351; Lot 000M913159
Model No 192-000; GTIN/DI 10811877011351; Lot 000M922950
Model No 192-000; GTIN/DI 10811877011351; Lot 000M922988
Model No 192-000; GTIN/DI 10811877011351; Lot 000M926188 |
| FEI Number |
1221359
|
Recalling Firm/ Manufacturer |
Abbott Diagnostics Scarborough, Inc.
10 Southgate Rd
Scarborough
ME
04074-8303
|
| For Additional Information Contact | Vicki Assardo 224-6681505 |
Manufacturer Reason for Recall | The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use. |
FDA Determined Cause 2 | Process change control |
| Action | On March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers.
Please complete the following actions, as applicable.
1. If you have impacted inventory in stock, then:
" Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures.
" Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot).
" Please retain this letter for your records.
2. If you have forwarded the product listed above to others in your network, then:
- Inform them of this Field Correction Notice, provide to them a copy of this notice and request they take the necessary action.
3. If you do not have impacted inventory in stock, then:
- All product lots not identified in Appendix I can continue to be used.
If you have questions regarding this information, please contact Sedgwick by phone at 888 671 8858 (available 8am 5pm EST, Monday - Friday).
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178).
If you have experienced any patient or user injury associated with this Field Correction Notice, please immediately report the event to your local area Customer Service.
****Update added 5/8/2025****
On April 29, 2025 FIELD CORRECTION NOTICE letters were sent to customers informing them of additional lots included in this recall. |
| Quantity in Commerce | 339456 eaches (337560 US, 1896 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom; Uruguay. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = QWR
|