| | Class 2 Device Recall Introducer Kit Coaxial Dilator |  |
| Date Initiated by Firm | March 27, 2025 |
| Date Posted | May 07, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1735-2025 |
| Recall Event ID |
96622 |
| Product Classification |
Percutaneous sheath introducer kit - Product Code OFL
|
| Product | Introducer Kit- Coaxial Dilator
REF:
KIT-002-28, 00841268104730
KIT-002-34, 00841268104792
KIT-002-35, 00841268104808
KIT-011-40, 00841268105553
KIT-011-62, 00841268105614
KIT-038-04, 00841268104556
KIT-039-03, 00841268104587
KIT-039-13, 00841268106055
KIT-081-02 / PS2-38, 00841268108394
These Micro-Introducer Kits are intended to introduce up to a .038 inch guidewire or catheter into the peripheral vascular system following a small gauge needle stick. |
| Code Information |
Catalog Number:
KIT-002-28
UDI-DI code: 00841268104730
Lot Number:
G25038156
KIT-002-34,
UDI-DI code: 00841268104792
Lot Number:
G25064259
KIT-002-35,
UDI-DI code: 00841268104808
Lot Number:
G25029077
KIT-011-40,
UDI-DI code: 00841268105553
Lot Number:
S25031227
KIT-011-62,
UDI-DI code: 00841268105614
Lot Numbers:
G25052163
G25080165
KIT-038-04,
UDI-DI code: 00841268104556
Lot Number:
G25092186
KIT-039-03,
UDI-DI code: 00841268104587
Lot Numbers:
G25050190
G25078192
KIT-039-13,
UDI-DI code: 00841268106055
Lot Number:
G25050202
KIT-081-02 / PS2-38,
UDI-DI code: 00841268108394
Lot Number:
G25056231
|
Recalling Firm/ Manufacturer |
Galt Medical Corporation 2220 Merritt Dr Garland TX 75041-6137
|
Manufacturer Reason for Recall | Due to a potential open seal in the sterile barrier packaging. |
FDA Determined Cause 2 | Process control |
| Action | On 03/27/2025, the firm emailed customers an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that affected products may have an open seal in the sterile barrier packaging that will render the product non-sterile and will increase the risk of patient harm due to microbial and environmental contamination.
Customers are instructed to:
1.DO NOT USE OR DISTRIBUTE THE DEVICEs LISTED IN THIS RECALL. Identify and segregate the affected lot(s) that are in their possession as well as those in the possession of their end-users.
2. Complete the enclosed Field Correction Reply Forms and email or fax them to the attention of the Corrections Coordinator at quality@galtmedical.com or 214-778-1433. We have enclosed two forms: one to account for inventory at their facility and the other for capturing these exact details from their end users. The form lists the product number, lot number, and quantity that the firm's records indicate their facility has received.
It is essential that, even if they do not have any product remaining in their possession, they complete the attached forms (Attachment A & B), indicating zero quantity to be returned, and email or fax the forms to GALT MEDICAL CORP.
3. Ship the returned product to GALT MEDICAL CORP. using Galt s carrier account information listed on the forms.
4. Reference Return Authorization Number RMA# 1200 on the outside of the shipping box and include a copy of the Field Correction Reply Forms with their shipment.
5. Once the completed Field Correction Reply Forms have been received and processed, Galt will issue a credit to customers for the returned product. Please place a new purchase order (PO), noting Replacement Parts for new products.
For questions, contact Quality Director at 1-214-778-1306 |
| Quantity in Commerce | 4,280 kits |
| Distribution | Worldwide - U.S. Nationwide distribution in the states of AL, FL, GA, IL, MA, OH, PA, and UT. The countries of Canada, Ireland, Israel, Korea, Qatar, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|