| | Class 1 Device Recall CADD Solis HSPCA |  |
| Date Initiated by Firm | April 10, 2025 |
| Date Posted | May 07, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1662-2025 |
| Recall Event ID |
96628 |
| 510(K)Number | K170982 |
| Product Classification |
Pump, infusion, pca - Product Code MEA
|
| Product | CADD Solis HSPCA Pump
***Updated 5/27/26***
Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope:
PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA
21-2111-0403-51
PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA
21-2111-0403-78
PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA
21-2112-0403-51
PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA
21-2112-0403-02
PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA
21-2111-0300-01
PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA
21-2111-0300-50
PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA
21-2112-0300-01
PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA
21-2112-0300-02
PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA
21-2112-0300-50
PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA
21-2101-51
PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA
21-2111-0401-78
PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA
21-2111-0402-78
PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA
21-2111-0400-51
PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA
21-2112-0400-51
PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA
21-2111-0401-51
PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA
21-2112-0401-51
PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA
21-2111-0402-51
PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA
21-2112-0402-78
PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA
21-2112-0402-51
PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA
21-2111-0100-51
PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA
21-2112-0100-51 |
| Code Information |
Affected CADD-Solis Software Versions: 4.0, 4.1, 4.2, 4.2.1, 4.3
UDI-DI:
4546027095675
4546027095682
10610586034787
10610586038778
10610586038785
10610586038808
10610586038822
10610586039232
10610586041037
10610586041051
15019517099690
15019517101003
15019517101010
15019517101034
15019517101041
15019517128697
15019517135343
15019517154290
15019517154887
15019517154924
15019517154979
15019517154986
15019517154993
15019517155006
15019517155013
15019517155044
15019517246735
15019517247008
15019517247169
15019517247176
15019517247183
15019517247206
00610586038900
00610586039310
04546027095620
04546027095637
04546027095644
04546027095668
04546027095699
04546027095705
04546027095712
04546027095729
04546027095736
04546027095897
04546027095903
04546027095910
04546027422983
04546027424116
04546027424130
04546027424147
04546027424178
04546027424185
04546027424192
04546027424208
04546027424598
05019517154934
05019517155108
10610586031854
10610586032950
10610586033933
10610586033940
10610586035357
10610586035395
10610586035401
10610586035418
10610586035432
10610586035456
10610586035494
10610586035548
10610586038792
10610586038839
10610586038860
10610586038891
10610586038914
10610586038921
10610586038938
10610586038952
10610586038969
10610586038983
10610586038990
10610586039010
10610586039041
10610586039065
10610586039072
10610586039089
10610586039195
10610586039256
10610586039270
10610586039409
10610586039416
10610586039430
10610586039447
10610586039454
10610586039461
10610586039478
10610586039485
10610586039492
10610586041822
30610586035337
30610586035344
30610586035368
30610586035511
30610586039205
30610586039397
35019517128677
35019517128684
35019517128691
35019517128721
35019517128738
35019517128745
35019517128769
35019517128783
35019517128837
35019517135316
35019517135323
35019517135330
35019517135354
35019517135361
35019517154850
35019517154867
35019517154874
35019517154904
35019517154911
35019517154942
35019517154959
35019517154966
35019517155055
35019517155062
35019517155079
35019517155086
35019517155130
|
| FEI Number |
3012307300
|
Recalling Firm/ Manufacturer |
Smiths Medical ASD, Inc.
6000 Nathan Ln N
Minneapolis
MN
55442-1690
|
| For Additional Information Contact | Technical Support 800-241-4002 |
Manufacturer Reason for Recall | Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death. |
FDA Determined Cause 2 | Software design |
| Action | Firm began notifying customers on April 10, 2025 via letters titled "Urgent Medical Device Correction."
Customers were instructed to validate any wireless network changes or updates to hospital network settings with the pump/communication module to ensure compatibility of the wireless connection before deploying network setting changes or updates to the production environment.
If the alarm recurs or is unable to be cleared by power cycling the pump, the customer should revert to prior compatible network settings or turn off the wireless function of the communication module.
The firm will also address the issue with a software update. Smiths Medical will contact customers to schedule implementation of the software update once it is available.
***Updated 5/28/2026***
On 5/27/26, firm revised the scope of the action to apply specifically to CADD customers using the wireless system. All current wireless customers, as well as any customers who upgrade to wireless, will receive the updated CADD firmware connection. All future implementations - both wireless and non-wireless - will include the corrected CADD firmware. |
| Quantity in Commerce | US: 83,833 items ; OUS: 71,843 items |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MEA
|
|
|
|