| | Class 2 Device Recall ARTIS pheno (VE30A) |  |
| Date Initiated by Firm | May 15, 2025 |
| Date Posted | June 27, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2019-2025 |
| Recall Event ID |
97135 |
| 510(K)Number | K230949 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | ARTIS is a family of dedicated angiography systems developed for single and
biplane diagnostic imaging and interventional procedures including, but not
limited to, pediatric and obese patients. Procedures that can be performed
with the ARTIS family include cardiac angiography, neuro-angiography, general
angiography, rotational angiography, multipurpose angiography and whole
body radiographic/fluoroscopic procedures as well as procedures next to the
table for i.e., patient extremities. This does not include projection radiography.
Additional procedures that can be performed include angiography in the
operating room, image guided surgery by X-ray, by image fusion, and by
navigation systems. The examination table as an integrated part of the system
can be used for X-ray imaging, surgery and interventions. ARTIS can also
support the acquisition of position triggered imaging for spatial data synthesis. |
| Code Information |
4056869046877 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
| For Additional Information Contact | 610-219-4834 |
Manufacturer Reason for Recall | In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose. |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | Siemens Healthineers sent customers a Customer Advisory Notice dated May 15, 2025 informing customers of a potential issue with ARTIS pheno and ARTIS icono systems with the following model numbers:
ARTIS pheno 10849000,
ARTIS icono biplane 11327600,
ARTIS icono ceiling 11328100,
ARTIS icono floor 11327700; due to in very rare situations, communication between the sensor measuring dose applied to the area and the corresponding
software fails. As a result of the failure, the system may incorrectly report an application of a high dose and appropriate software update will be implemented, based on the current software version of the affected system. |
| Distribution | Worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|