| | Class 2 Device Recall ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS icono (VE40A) (ceiling); ARTIS phe |  |
| Date Initiated by Firm | May 15, 2025 |
| Date Posted | June 27, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2022-2025 |
| Recall Event ID |
97135 |
| 510(K)Number | K241572 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | interventional fluoroscopic x-ray system |
| Code Information |
4056869149325 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
| For Additional Information Contact | 610-219-4834 |
Manufacturer Reason for Recall | In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose. |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | Siemens Healthineers sent customers a Customer Advisory Notice dated May 15, 2025 informing customers of a potential issue with ARTIS pheno and ARTIS icono systems with the following model numbers:
ARTIS pheno 10849000,
ARTIS icono biplane 11327600,
ARTIS icono ceiling 11328100,
ARTIS icono floor 11327700; due to in very rare situations, communication between the sensor measuring dose applied to the area and the corresponding
software fails. As a result of the failure, the system may incorrectly report an application of a high dose and appropriate software update will be implemented, based on the current software version of the affected system. |
| Distribution | Worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|