• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall BD Pyxis " Main, BD Pyxis " MedBank " Mini Main, BD Pyxis " MedBank " Tower Main, and BD Pyxis " MedFlex

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall BD Pyxis " Main, BD Pyxis " MedBank " Mini Main, BD Pyxis " MedBank " Tower Main, and BD Pyxis " MedFlexsee related information
Date Initiated by FirmJuly 10, 2025
Date PostedAugust 13, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2321-2025
Recall Event ID 97245
Product Classification Cabinet, table and tray, anesthesia - Product Code BRY
ProductDescription/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/169-102, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/169-101 , BD PYXIS MEDBANK MINI 1FH-2HH/155311-01, BD PYXIS MEDBANK MINI 4HH/155310-01, BD PYXIS MEDBANK MINI 1FH-1FM/155288-01, BD PYXIS MEDBANK MINI 4HM/155171-01, BD PYXIS MEDBANK MINI 3HH-1HM/155169-01, BD PYXIS MEDBANK MINI 3HH-1FG/155168-01, BD PYXIS MEDBANK MINI 2HM-1FM/155167-01, BD PYXIS MEDBANK MINI 2HH-2HM/155166-01, BD PYXIS MEDBANK MINI 2HH-1FM/155165-01, BD PYXIS MEDBANK MINI 2FM/155163-01, BD PYXIS MEDBANK MINI 1HH-3HM/155161-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/155160-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/155159-01, BD PYXIS MEDBANK MINI 2HM-1FM/139131-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/139122-01, BD PYXIS MEDBANK MINI 3HH-1FG/139091-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/138921-01, BD PYXIS MEDBANK MINI 4HM/138915-01, BD PYXIS MEDBANK MINI 1HH-3HM/138912-01, BD PYXIS MEDBANK MINI 4HH/138910-01, BD PYXIS MEDBANK MINI 3HH-1HM/138909-01, BD PYXIS MEDBANK MINI 2HH-1FH/138908-01, BD PYXIS MEDBANK MINI 2HH-2HM/138907-01, BD PYXIS MEDBANK MINI 2HH-1FM/138906-01, BD PYXIS MEDBANK MINI 2FM/138903-01, BD PYXIS MEDBANK MINI 2FH/138902-01, MEDBANK MINI MAIN/1147-00, MEDBANK MAIN/1137-00, BD PYXIS MEDBANK TWR MN 8HH-3FM/169-81, BD PYXIS MEDBANK TWR MN 8HH-3FH/169-80, BD PYXIS MEDBANK TWR MN 8HH-3FF/169-79, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/169-78, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/169-77, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/169-76, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/169-75, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/169-74, BD PYXIS MEDBANK TWR MN 7FM/169-73, BD PYXIS MEDBANK TWR MN 7FH/169-72, BD PYXIS MEDBANK TWR MN 6HH-8HM/169-71, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/169-70, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/169-69, BD PYXIS MEDBANK TWR MN 6HH-4FM/169-68, BD PYXIS MEDBANK TWR MN 6HH-4FH /169-67, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/169-66, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/169-65, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/169-64, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/169-63, BD PYXIS MEDBANK TWR MN 6FH-1FM/169-62, BD PYXIS MEDBANK TWR MN 5HH-9HM/169-61, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/169-60, BD PYXIS MEDBANK TWR MN 4HM-5FM/169-59, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/169-58, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/169-57, BD PYXIS MEDBANK TWR MN 4HH-5FM/169-56, BD PYXIS MEDBANK TWR MN 4HH-5FH /169-55, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/169-54, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/169-53, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/169-52, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/169-51, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/169-50, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/169-49, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/169-48, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/169-45, BD PYXIS MEDBANK TWR MN 4HH-10HM/169-44, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/169-43, BD PYXIS MEDBANK TWR MN 3FH-4FM/169-42, BD PYXIS MEDBANK TWR MN 2HM-6FM/169-41, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/169-40, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/169-39, BD PYXIS MEDBANK TWR MN 2HH-6FM/169-38, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/169-37, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/169-36, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/169-35, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/169-34, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/169-33, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/169-32, BD PY
Code Information UDI-DI: 10885403512544, 10885403517181, 10885403512568, 10885403517167, 10885403512704. Serial Numbers: 17458898 17449648 17449652 17443178 17443179 17443177 17438233 17422325 17422322 17422331 17412631 17405923 17405924 17404980 17404945 17404960 17404988 17404991 17404947 17404992 17404985 17404944 17402533 17402530 17398216 17394106 17394107 17394118 17394144 17394142 17394149 17394152 17394124 17394139 17394137 17394130 17394125 17391193 17391434 17391189 17389466 17389463 17389464 17379928 17379924 17379926 17379917 17379919 17379921 17379923 17379901 17379907 17379912 17380888 17380893 17380890 17380885 17380886 17377201 17377216 17377218 17377217 17376116 17377219 17376131 17377173 17376124 17376132 17376118 17377208 17377212 17376130 17376133 17376115 17376128 17376125 17376111 17376117 17376122 17376121 17376113 17377210 17376114 17377214 17377211 17377204 17377195 17373685 17364636 17359323 17355092 17355088 17355107 17353367 17355093 17353839 17353837 17348448 17349084 17349073 17348455 17349088 17349103 17349083 17349106 17348457 17344748 17344782 17343434 17344760 17344783 17344781 17344780 17344745 17342689 17343421 17342683 17344768 17342686 17343423 17342702 17342621 17344788 17339144 17339145 17339146 17339150 17337848 17337907 17339300 17337908 17337910 17337845 17337878 17339295 17337866 17337876 17337843 17337894 17339152 17337917 17337913 17337858 17337039 17337040 17337041 17337042 17337043 17337044 17337045 17337046 17337047 17337048 17337049 17337050 17337051 17337052 17337053 17337054 17337055 17337056 17337057 17337058 17285941 17285942 17278814 17278822 17275121 17266351 17266357 17266314 17266308 17264519 17264521 17264524 17264527 17266319 17264540 17264544 17253267 17247731 17247737 17247740 17247745 17247782 17247720 17247812 17247804 17247811 17247803 17247789 17247795 17247802 17247763 17247758 17237343 17237373 17237347 17231090 17231070 17231068 17231065 17231072 17231083 17231087 17231077 17231093 17230261 17230019 17230020 17230116 17230017 17230018 17230117 17230119 17226644 17226038 17226638 17222014 17221891 17221892 17222019 17222017 17222021 17222022 17222023 17221896 17222026 17222024 17222025 17222009 17221897 17221898 17222006 17222008 17222016 17221890 17221895 17221893 17214916 17214918 17214923 17214917 17214919 17212547 17212546 17212545 17212544 17212541 17212539 17212538 17212540 17212537 17212533 17212534 17212535 17186603 17187560 17181676 17181672 17181670 17181673 17170832 17170835 17170837 17170838 17170839 17170842 17170783 17170915 17170779 17170780 17170782 17170936 17170931 17170939 17160759 17160756 17160767 17160769 17160764 17160760 17160763 17160765 17160039 17160040 17160017 17160037 17160018 17160041 17160022 17160019 17160021 17160023 17160038 17150146 17146445 17146444 17146449 17146512 17146515 17146518 17146520 17146439 17146467 17146555 17146553 17146469 17146551 17146438 17146456 17146543 17146547 17146461 17146433 17146452 17146530 17146527 17146532 17146526 17146546 17146509 17140533 17140986 17140988 17140976 17140973 17140987 17140982 17140972 17140974 17140978 17140955 17140965 17124145 17124146 17117177 17117179 17118826 17118818 17118817 17118819 17118827 17118816 17118825 17118821 17113391 17113397 17113399 17113858 17113400 17113880 17113398 17113860 17113844 17113395 17113856 17113878 17113877 17113857 17113866 17113863 17113868 17103252 17103287 17103244 17103254 17103242 17103289 17103281 17103241 17103267 17102985 17103257 17090997 17091000 17090993 17090990 17088230 17088233 17088238 17088235 17088231 17088237 17088232 17088226 17088227 17088229 17088234 17088236 17088228 17085484 17076844 17076874 17076878 17076882 17076885 17076887 17076890 17076892 17076894 17076897 17076872 17076900 17076922 17076924 17076926 17076841 17076928 17077313 17076836 17076833 17076865 17076849 17075823 17076808 17076807 17076930 17076932 17076933 17076857 17076859 17076832 17070480 17070479 17070473 17070477 17070472 17070470 17070462 17070469 17070459 17070465 17070475 17070466 17070461 17070454 17070457 17070476 17070455 17070452 17070467 17070450 17070441 17070642 17070446 17070445 17070640 17070444 17070447 17070443 17068658 17065970 17065969 17064235 17064233 17064240 17055660 17055658 17055666 17055665 17055659 17055654 17055656 17055667 17055663 17055649 17055657 17027301 17027219 17027303 17027241 17027295 17027261 17027201 17027281 17027289 17027268 17027175 17027255 17023869 17023870 17023871 17023868 17024246 17023498 17024245 17024244 17023510 17012287 17012241 17012252 17012239 17012237 17012240 17012236 17012265 17012251 17012293 17012294 17012290 17012274 17012269 17012271 17012273 17012272 17012268 16998700 17001338 17001320 17001341 17001340 17001339 17001321 17001331 17001327 17001329 17001328 16995873 16995447 16995453 16995395 16995390 16995451 16995388 16995384 16995469 16994460 16994461 16995449 16995450 16992605 16992607 16992799 16992797 16992798 16992350 16992349 16992337 16970798 16970804 16970801 16970809 16935674 16935657 16935647 16913476 16913472 16913487 16913483 16910103 16910106 16910112 16910096 16910108 16910098 16891632 16891639 16891635 16891640 16891637 16885231 16885056 16879334 16879330 16879323 16879329 16879326 16879332 16876539 16876542 16876537 16876544 16876541 16876538 16876543 16876530 16876553 16876642 16876529 16859280 16859279 16859282 16859283 16859278 16845151 16845154 16845153 16839260 16839263 16839262 16835645 16835627 16835635 16835639 16828263 16828262 16827800 16827801 16827790 16827775 16827776 16828271 16827774 16827779 16828269 16828270 16827805 16828209 16827770 16819236 16814469 16809178 16809171 16809173 16809157 16809172 16809153 16809156 16809154 16809177 16809175 16809176 16809168 16790085 16790043 16790091 16790041 16790060 16790065 16790066 16790067 16790069 16790072 16790052 16790051 16790050 16785705 16785703 16785700 16785701 16778147 16778149 16777334 16778155 16778158 16778164 16777447 16778005 16778151 16777301 16777434 16777438 16778171 16778177 16778178 16778183 16778166 16778182 16778179 16778169 16777329 16778181 16777319 16778153 16778165 16764209 16764213 16765178 16765205 16765206 16765176 16765199 16765200 16765179 16765195 16765173 16757991 16757816 16757828 16757830 16757833 16757835 16757825 16757827 16757823 16757826 16757834 16757829 16748233 16748232 16747878 16748234 16748235 16748236 16748231 16747877 16747879 16747883 16747881 16747882 16747880 16747873 16747875 16735600 16727771 16727778 16727776 16721475 16721479 16721483 16721482 16719626 16719595 16719596 16719599 16719598 16719593 16717428 16717424 16712903 16712889 16714006 16714012 16714011 16714004 16714003 16714010 16714002 16714009 16714016 16714005 16714013 16714015 16706796 16706798 16706797 16701103 16693084 16691046 16691047 16691048 16691053 16691050 16691044 16691049 16685050 16687034 16686999 16687007 16687030 16687031 16687012 16687009 16687013 16687028 16687003 16687018 16675215 16675168 16675865 16675809 16675810 16675812 16675815 16675816 16675806 16675871 16675807 16675861 16675859 16675880 16675878 16661054 16661052 16661053 16661317 16661312 16661055 16661068 16661066 16661069 16661306 16661059 16661070 16661057 16661065 16661067 16661071 16661072 16661174 16661056 16661058 16650692 16647712 16647718 16647715 16647720 16647710 16641391 16643140 16626428 16626427 16628039 16628037 16628038 16623551 16623549 16623560 16623576 16623580 16623591 16623574 16623565 16623584 16623567 16623590 16623589 16623558 16623586 16623561 16623581 16623563 16623554 16620623 16620622 16620654 16620696 16620694 16620692 16620697 16620693 16620695 16619487 16619506 16619490 16619491 16619492 16619508 16619507 16612169 16612167 16604645 16601178 16598591 16598590 16598754 16598748 16598747 16594681 16596312 16596355 16596358 16596298 16596302 16596353 16596369 16596367 16596352 16596361 16596364 16596372 16596371 16596291 16596290 16587457 16587543 16587545 16587531 16587522 16587454 16587458 16587465 16587474 16587468 16587470 16587473 16587467 16587471 16587469 16587462 16587460 16587510 16587500 16586539 16587488 16587490 16587491 16587495 16579074 16577899 16577882 16579156 16577898 16577897 16579071 16579138 16579150 16577892 16577888 16577890 16577889 16579151 16579152 16579142 16579147 16577887 16579076 16577900 16577886 16577895 16579079 16579135 16579140 16579145 16579161 16579094 16574848 16574504 16575830 16575876 16575873 16575886 16574502 16574539 16575878 16575881 16575868 16574505 16575874 16575875 16574497 16574465 16574483 16575883 16574501 16569428 16570197 16570171 16570208 16570206 16570173 16570135 16570172 16569289 16569290 16570132 16570257 16570215 16565598 16565583 16565590 16566155 16565593 16566151 16565842 16562377 16558936 16558933 16558929 16558938 16558942 16550170 16550109 16550070 16550059 16550092 16550104 16550054 16550091 16550111 16550089 16550097 16550065 16550068 16550113 16550051 16550101 16550099 16550096 16550057 16550073 16550061 16550105 16550094 16550107 16550124 16543333 16543327 16543329 16543110 16543106 16543331 16543112 16543115 16543121 16543103 16543134 16543137 16543139 16543118 16538799 16538803 16538800 16538797 16534640 16534642 16531501 16531500 16531503 16531506 16527419 16528927 16528926 16528924 16528929 16525037 16521603 16521274 16521606 16516780 16516782 16516779 16516771 16516768 16516775 16510272 16510269 16510270 16506554 16503897 16502516 16503896 16502330 16502329 16502257 16491396 16490496 16491408 16491384 16491407 16491392 16491388 16491409 16491400 16491403 16491379 16488020 16487428 16487425 16479191 16478387 16474392 16472326 16469897 16468081 16468197 16468199 16468196 16468198 16468079 16468200 16468082 16468080 16468103 16468104 16468101 16468105 16461396 16461398 16461401 16461411 16460718 44658907 16455056 16455110 16455468 16455755 16450291 16450313 16450314 16450315 16450309 16450316 16446397 16437155 16434526 16434523 16430664 16427306 16427307 16424562 16424717 16423194 16423191 16423192 16423188 16423195 16423051 16422217 16422218 16422219 16417406 16417403 16416523 16416516 16416515 16416513 16416514 16416524 16416521 16416522 16416525 16416517 16416112 16416190 16416191 16416189 16415214 16415210 16415212 16415211 16415209 16415213 16414139 16414133 16414134 16414135 16414132 16413726 16413723 16413721 16413725 16413722 16413701 16413724 16413720 16413409 16413417 16413414 16413419 16413415 16413416 16413420 16413413 16413418 16413421 16413365 16413370 16413361 16413362 16413363 16413366 16413359 16413369 16413367 16413371 16413364 16413368 16413360 16413372 16413357 16413358 16412200 16412201 16412204 16412202 16412199 16412203 16412169 16412167 16412171 16412114 16412173 16412168 16412172 16412113 16412166 16411894 16411893 16411895 16411896 16411897 16411898 16411892 16410148 16409382 16409385 16409381 16409386 16409388 16409383 16409387 16407123 16407705 16407703 16407698 16407700 16407702 16407598 16407122 16406085 16406056 16406057 16405232 16406061 16406059 16406060 16406058 16403805 16404334 16403809 16404332 16404330 16403807 16403806 16403808 16403810 16404356 16404354 16404350 16399720 16399721 16399725 16399723 16398277 16398853 16398851 16398854 16398278 16398861 16398862 16398850 16398279 16397522 16397530 16397529 16396551 16396552 16396553 16397531 16396550 16396548 16396549 16396554 16397528 16395522 16395527 16395523 16395524 16395525 16394968 16394967 16394970 16394976 16394977 16394974 16394969 16394971 16394973 16390540 16390541 16390534 16389774 16390530 16390531 16389914 16390532 16390533 16390529 16387858 16387868 16388244 16388837 16388252 16388835 16388840 16388838 16388834 16388249 16388845 16388844 16388836 16388843 16388839 16388215 16388251 16388219 16388233 16388221 16388230 16388225 16388231 16388223 16388227 16388237 16383211 16383212 16383203 16383206 16383210 16383209 16383207 16383204 16383205 16381296 16381297 16381284 16381291 16381292 16381298 16381294 16381282 16381293 16381295 16381283 16381300 16381301 16375388 16374659 16374257 16374087 16374261 16374258 16374260 16374259 16374657 16374090 16374083 16373119 16372466 16373059 16372443 16372445 16373055 16372452 16372446 16372444 16372437 16372440 16372435 16372450 16366692 16366693 16366691 16366694 16365317 16365321 16365319 16365318 16365316 16362405 16362407 16362404 16362406 16360888 16361332 16358680 16356145 16356143 16356144 16355644 16354646 16354647 16354648 16354505 16354507 16354503 16352787 16349691 16346261 16346262 16346259 16346260 16344721 16344719 16340968 16340991 16340990 16340995 16340967 16340994 16340993 16340992 16339701 16337394 16337040 16337050 16337034 16337048 16337036 16337035 16337039 16337037 16337038 16335926 16335925 16335406 16335408 16335407 16335374 16335376 16331376 16331477 16331371 16331475 16331372 16331474 16331237 16331373 16331476 16331374 16331473 16331235 16331239 16331243 16331241 16331234 16329842 16329843 16329841 16329844 16329840 16329845 16330464 16330466 16330461 16330456 16330459 16329984 16328672 16328663 16328664 16328661 16328667 16328668 16328662 16328776 16328777 16328781 16328778 16328779 16328780 16328783 16328771 16327197 16327189 16327194 16327195 16327190 16327193 16327183 16327196 16327188 16327191 16327187 16327192 16327200 16327209 16327210 16327205 16327208 16327202 16327211 16327213 16327214 16324881 16324883 16324880 16324858 16324856 16324882 16324949 16324943 16324940 16324939 16324947 16324945 16324946 16324942 16324950 16323807 16323805 16322982 16322984 16322983 16321804 16321806 16321809 16321812 16321814 16321788 16321799 16321815 16321794 16321792 16319674 16320192 16320186 16320189 16320188 16320193 16320184 16320187 16320194 16320190 16320191 16317878 16316472 16316474 16316476 16316478 16316470 16316468 16316484 16314884 16314269 16314272 16314270 16314264 16314271 16314267 16314268 16314265 16314266 16314273 16313000 16313001 16312997 16310002 16310003 16310008 16310011 16310004 16310010 16310006 16310013 16310012 16310007 16310005 16310001 16310009 16308993 16308990 16308989 16308977 16308979 16308973 16308987 16308998 16308971 16308975 16308992 16308991 16306024 16306411 16306401 16306402 16306413 16306393 16306395 16306408 16306419 16306403 16306397 16306404 16306399 16305129 16305136 16305134 16305130 16305133 16303465 16303467 16303469 16303463 16303468 16303466 16302339 16302328 16302327 16302337 16302331 16302343 16302338 16302342 16302332 16302333 16302330 16302329 16302340 16302341 16301869 16301974 16301023 16301024 16301018 16301021 16301015 16301020 16301016 16300643 16300641 16300639 16300640 16300642 16297183 16297187 16297697 16297696 16297666 16296379 16296380 16296360 16296384 16296362 16296378 16295928 16295929 16295930 16295931 16295917 16295915 16295933 16295919 16295920 16294039 16294040 16290131 16290130 16285719 16285712 16283985 16283988 16283982 16284016 16284007 16284015 16284011 16284014 16283992 16282653 16282655 16282654 16282657 16282658 16282716 16282717 16282709 16282713 16282712 16282714 16282710 16282715 16282711 16282656 16282651 16282191 16281132 16281131 16281129 16281130 16278997 16277309 16277298 16277307 16277303 16277301 16277299 16277305 16277306 16277296 16277295 16276630 16276629 16276634 16276633 16276631 16276632 16274621 16274623 16274823 16274023 16274218 16274219 16274217 16274214 16274216 16274215 16274024 16273979 16273982 16273994 16273995 16273993 16273978 16273998 16273991 16273992 16273984 16273980 16273981 16273983 16274012 16273996 16274013 16274005 16274008 16273970 16274022 16274019 16274004 16273999 16274011 16274016 16274009 16274015 16274014 16274010 16274002 16274001 16274017 16274018 16274220 16273971 16274003 16274007 16273523 16273525 16272999 16273001 16273000 16273341 16273002 16273332 16273337 16273326 16273328 16271697 16271723 16271722 16271205 16271198 16271203 16271201 16271195 16271200 16271202 16271207 16271204 16271199 16271197 16271196 16271206 16271173 16271168 16271172 16271164 16271165 16271162 16271163 16271167 16271160 16271166 16271161 16271171 16271169 16271170 16271070 16270827 16270829 16270830 16270834 16270836 16270236 16270237 16270241 16270246 16270239 16270235 16270238 16270247 16270240 16269765 16269757 16269756 16269758 16269760 16269763 16269762 16269761 16269759 16269098 16269099 16269097 16269096 16269095 16269094 16269112 16269114 16269110 16269111 16269113 16269273 16268999 16268998 16268996 16263866 16263864 16262553 16262214 16262208 16262209 16262206 16262213 16262207 16262204 16262200 16262212 16262201 16262203 16262211 16262210 16262202 16262215 16261298 16261299 16260884 16260054 16259235 16259147 16259148 16259145 16259146 16258341 16258339 16258336 16258459 16257751 16257783 16257780 16257784 16257782 16257779 16257468 16257460 16257462 16257465 16257461 16257457 16257464 16257466 16257459 16257467 16257469 16257470 16256356 16256358 16256359 16255649 16255652 16255651 16255648 16255650 16255572 16255575 16255574 16255087 16255088 16255094 16255091 16255093 16255092 16255090 16255089 16254403 16254401 16254402 16254400 16254405 16254404 16254361 16254375 16254366 16254373 16254370 16254368 16254365 16254376 16254364 16254374 16254363 16254369 16254377 16254367 16254372 16254371 16254362 16254384 16254360 16253979 16253978 16253977 16253421 16253423 16253383 16253382 16253258 16253260 16253259 16253264 16253261 16252908 16252916 16252917 16252912 16252913 16252915 16252911 16252914 16252910 16252909 16252357 16252358 16252354 16252356 16252353 16252355 16250856 16250858 16250860 16250859 16250857 16250861 16250218 16250220 16250221 16250222 16250217 16250216 16249865 16249864 16249866 16249390 16249393 16249392 16249394 16249389 16249395 16249300 16249301 16248579 16248217 16248200 16248197 16248207 16248209 16248211 16248208 16248203 16248214 16248216 16247093 16247462 16247096 16247098 16247095 16247097 16247232 16247233 16247231 16247230 16247461 16247464 16246162 16246160 16246163 16246166 16246164 16246170 16246169 16246167 16246165 16246168 16246161 16245711 16245713 16245712 16245718 16245686 16245246 16245245 16245244 16245243 16245242 16245198 16245121 16245120 16245119 16244879 16244880 16244881 16244808 16244807 16244197 16243953 16243544 16243897 16243893 16243894 16243555 16243895 16243896 16243554 16243891 16243890 16243892 16243552 16243549 16243551 16243196 16243195 16242834 16242807 16242806 16242840 16242842 16242373 16242378 16242369 16242368 16242370 16242372 16242374 16242375 16242371 16242377 16242376 16242366 16242365 16242380 16242381 16242364 16242367 16242032 16242036 16242035 16242030 16242033 16242029 16242037 16242034 16242031 16242040 16242038 16242028 16242039 16240300 16240310 16240306 16240298 16240301 16240302 16240307 16240304 16240296 16240297 16240303 16240308 16240309 16240305 16240299 16240280 16240289 16240287 16240291 16240294 16240292 16240295 16240288 16240293 16240290 16239870 16239866 16239862 16239864 16239865 16239867 16239868 16239200 16239201 16239203 16239197 16239198 16239199 16239414 16238932 16237810 16237809 16237812 16237813 16237807 16237811 16237852 16237831 16237833 16237834 16237832 16237830 16237838 16237840 16237843 16237842 16237846 16237844 16237841 16237848 16237845 16237850 16237837 16237466 16237465 16236598 16236599 16236600 16236601 16236602 16236297 16236298 16236299 16236303 16236306 16236304 16236300 16236305 16236307 16236308 16236296 16236301 16236302 16235776 16235771 16235780 16235779 16235773 16235778 16235777 16235774 16235579 16235577 16235578 16235580 16235041 16235040 16235044 16235049 16235042 16235048 16235045 16235043 16235047 16234920 16234919 16234918 16234871 16234815 16234816 16234814 16234810 16234812 16234809 16234811 16234813 16234817 16234386 16234385 16234387 16234389 16234388 16233314 16233175 16233228 16233313 16233310 16233311 16233133 16233312 16233309 16232563 16232562 16232316 16232268 16232271 16232270 16231933 16232116 16232044 16232047 16232046 16231480 16231481 16231482 16231087 16231477 16231478 16231483 16231476 16231479 16230235 16228958 16228959 16228623 16228645 16228637 16228639 16228489 16228602 16228632 16228633 16228634 16228635 16228636 16228125 16227643 16227641 16227634 16227635 16227642 16227645 16227646 16227644 16227636 16227637 16227640 16227638 16227555 16227552 16227553 16227554 16227551 16227550 16227549 16227548 16227389 16227401 16227399 16227393 16227397 16227394 16227398 16227396 16227400 16227395 16227392 16227390 16226698 16226697 16226700 16226699 16226704 16226705 16226701 16226706 16226702 16226622 16226623 16226159 16225844 16225354 16225353 16225356 16225357 16225355 16225063 16225064 16225065 16225068 16225062 16225066 16223966 16223714 16223709 16223710 16223713 16223712 16223711 16223708 16223715 16223719 16223716 16223722 16223717 16223718 16223706 16223721 16223720 16223723 16223380 16223379 16223378 16223377 16221614 16221619 16221611 16221620 16221610 16221613 16221621 16221612 16221609 16221608 16221607 16221615 16221618 16221616 16221617 16221558 16221559 16221561 16221560 16221563 16221562 16221550 16221552 16221553 16221551 16221554 16221555 16221565 16221445 16221447 16221444 16221448 16221449 16221450 16220530 16220635 16220532 16220634 16220636 16220529 16220637 16220631 16220626 16220630 16220629 16220624 16220625 16220623 16220616 16220617 16220615 16220229 16220228 16220235 16220238 16220239 16220237 16220234 16220232 16220231 16220242 16220227 16220236 16220240 16220233 16220241 16220230 16219658 16219653 16219665 16219652 16219663 16219666 16219664 16219662 16219657 16219640 16219654 16219643 16219659 16219642 16219661 16219634 16219639 16219641 16219632 16219660 16220056 16220057 16219621 16219620 16219247 16219248 16217681 16217687 16217691 16217684 16217694 16217692 16217682 16217689 16217690 16217688 16217686 16217693 16217683 16217685 16217695 16217696 16216632 16216081 16216105 16216104 16216103 16216102 16216080 16215111 16215103 16215110 16215109 16215107 16215108 16215104 16215105 16215106 16214567 16214275 16214272 16214273 16214274 16214269 16214268 16214270 16214271 16212665 16212667 16212782 16212666 16212785 16212786 16212784 16212017 16212021 16212022 16212020 16212019 16212018 16211927 16211930 16211573 16211589 16211592 16211590 16211591 16211574 16211579 16211578 16211577 16211575 16211576 16211580 16211572 16210992 16210993 16210994 16210910 16210911 16210470 16210472 16210474 16210464 16210463 16210461 16210462 16210465 16210480 16210479 16210466 16210467 16210459 16210915 16210914 16210481 16210469 16210477 16210335 16210334 16210333 16209956 16209852 16209854 16209850 16209289 16209288 16209287 16209074 16208658 16208667 16208521 16208520 16208516 16208518 16208517 16208519 16208515 16208514 16208513 16208512 16208510 16208511 16207669 16207668 16207670 16207707 16207658 16207657 16207747 16207712 16207709 16207711 16207659 16207738 16207728 16207726 16206142 16206144 16206143 16206145 16206121 16206120 16205732 16205729 16205730 16205726 16205723 16205720 16205721 16205722 16205725 16205719 16205724 16205016 16204872 16204868 16204871 16204867 16204870 16204869 16204682 16204683 16203712 16203699 16203714 16203709 16203144 16203145 16203142 16202459 16202458 16202462 16203036 16202461 16203032 16203034 16203039 16202460 16203037 16202463 16201161 16201162 16201160 16200466 16200464 16200465 16200467 16200472 16200469 16200468 16200470 16200481 16200068 16200073 16200066 16199481 16199483 16199482 16198303 16195042 16195054 16195029 16195048 16195050 16195033 16195038 16194942 16194944 16194943 16194403 16194405 16194394 16194393 16194391 16194388 16194389 16194390 16193269 16192178 16192176 16191236 16191232 16191231 16191235 16191234 16191237 16191230 16191233 16191220 16191222 16191219 16190207 16190429 16190431 16190426 16190432 16190430 16190428 16190427 16189541 16189536 16189542 16189540 16189538 16188931 16188924 16188930 16188926 16188927 16188678 16188674 16188675 16188673 16188676 16188677 16188346 16188345 16188344 16188343 16188081 16188083 16188084 16188082 16187584 16187382 16186998 16186997 16186999 16187001 16186501 16186252 16186240 16186236 16186243 16186237 16186242 16186239 16186238 16186235 16185367 16184668 16184667 16184666 16184665 16184762 16182535 16182061 16182324 16182062 16182329 16182327 16182325 16182326 16181895 16181893 16181892 16182330 16182059 16182322 16182054 16182042 16182057 16182056 16180995 16180325 16180516 16179611 16179852 16179371 16179370 16179367 16179368 16178602 16178404 16177061 16177060 16175309 16175310 16175311 16175312 16175313 16175314 16175316 16175317 16175318 16175320 16175306 16175308 16174520 16174398 16174521 16174081 16173244 16173243 16173246 16173349 16173247 16173242 16170660 16170659 16170538 16170539 16170657 16169588 16169580 16169584 16169587 16169582 16169586 16169583 16169592 16169591 16169594 16169595 16169590 16169589 16169597 16169596 16169599 16169600 16168841 16168377 16168375 16168376 16168373 16167620 16167617 16166277 16166281 16166283 16166285 16166278 16166280 16165553 16165554 16165552 16165551 16165547 16165548 16165550 16165555 16165549 16165544 16164289 16164287 16164261 16164260 16164257 16164259 16164258 16162537 16162544 16162547 16162536 16162541 16161539 16161551 16161552 16161553 16161540 16161555 16161557 16161558 16161556 16161541 16161576 16161577 16160275 16160283 16160284 16160289 16160287 16160286 16160285 16160288 16160274 16160108 16160246 16160247 16160248 16159989 16159987 16159988 16159758 16159759 16159757 16159781 16159780 16159789 16159783 16159556 16159137 16159138 16158753 16158752 16158519 16158512 16158517 16158518 16158515 16158513 16158514 16158464 16158465 16158463 16157021 16157016 16157023 16157018 16156626 16155522 16155520 16155518 16155517 16155521 16154519 16154518 16154249 16154250 16154252 16154254 16154253 16153564 16152697 16151701 16151699 16151700 16151584 16151585 16151583 16151468 16151381 16151382 16151378 16151379 16151380 16151043 16151040 16151042 16151041 16148982 16148978 16148979 16148959 16149476 16149477 16148945 16148946 16147788 16147608 16147609 16147499 16147319 16146405 16146408 16146406 16145924 16145923 16145922 16145921 16144807 16144811 16144810 16144331 16144339 16144336 16144333 16144335 16142010 16141588 16141585 16141549 16141547 16141550 16141551 16141542 16141543 16141545 16140468 16140524 16139797 16139686 16139836 16139676 16139678 16139679 16139672 16139684 16139153 16139154 16138837 16138836 16138831 16138827 16138825 16138835 16138829 16138828 16137673 16137674 16137675 16137617 16137618 16136731 16136732 16136734 16136730 16136728 16136729 16136727 16136726 16136733 16136378 16136377 16136376 16136375 16136374 16136418 16136419 16136421 16136422 16136379 16136380 16135189 16135191 16135187 16135186 16135190 16135192 16135188 16134693 16132831 16132438 16132248 16131741 16130129 16130311 16130312 16130314 16130309 16130308 16130310 16129507 16129770 16128854 16128845 16129009 16129013 16128859 16128855 16128848 16129017 16128847 16128853 16128851 16129016 16129014 16128856 16128861 16129015 16129012 16128850 16128844 16128849 16128843 16128394 16127653 16127655 16127654 16127652 16127651 16126673 16126158 16126161 16125701 16125697 16125678 16125679 16125699 16125676 16125698 16125674 16124909 16124910 16124905 16124907 16124605 16124603 16124604 16124086 16124076 16124081 16124082 16124083 16124093 16124084 16124077 16124085 16124087 16124080 16124078 16124079 16123677 16122575 16122576 16121796 16120931 16120938 16120948 16120950 16120952 16120708 16120707 16120704 16120070 16119930 16119877 16119931 16119932 16119933 16119929 16119934 16119405 16119404 16119407 16119408 16119406 16119409 16118353 16118352 16118366 16118368 16118364 16118367 16118369 16118372 16118373 16118370 16118371 16115895 16116015 16115959 16115958 16115957 16116027 16113462 16113461 16113241 16113239 16113240 16112808 16112853 16112812 16112398 16112397 16112384 16112385 16112396 16112388 16112393 16112392 16112386 16112382 16112381 16112379 16112380 16112378 16112377 16112376 16112390 16112391 16112178 16112179 16112187 16112184 16112185 16112186 16112040 16112175 16112176 16112174 16112177 16112047 16112048 16112044 16112045 16112046 16112173 16112171 16112042 16111850 16111853 16111847 16111849 16111852 16111848 16111846 16110888 16110889 16110532 16110145 16110135 16110144 16110138 16110143 16109984 16109890 16109988 16109891 16109892 16110003 16109993 16109886 16109995 16109889 16110002 16109997 16109998 16109994 16109999 16109885 16109996 16109888 16110001 16109989 16109991 16109992 16109319 16109077 16109067 16109320 16109322 16109323 16109318 16109317 16109075 16109076 16109073 16109227 16109066 16109068 16109228 16109069 16109070 16109072 16109071 16109224 16109225 16108913 16108917 16108830 16108832 16108831 16108828 16108833 16108827 16108914 16108824 16108543 16108544 16108545 16108537 16108539 16108527 16108530 16108531 16108536 16108529 16108532 16108533 16108535 16108528 16108526 16108534 16108541 16108132 16108134 16108133 16107569 16107567 16107568 16107091 16107092 16107015 16107016 16107017 16107020 16107018 16107019 16107287 16107021 16107289 16107290 16107014 16107288 16107022 16107286 16107291 16106589 16106595 16106636 16106631 16106594 16106591 16106592 16106587 16106202 16106201 16106203 16106204 16106205 16106206 16105996 16105995 16105997 16105993 16105992 16105991 16105080 16105087 16105083 16105084 16105092 16105088 16105090 16105091 16105086 16105085 16105081 16097437 16097441 16097442 16095340 16095337 16095336 16095333 16095335 16095334 16095332 16095338 16095339 16093717 16093716 16093640 16093639 16093533 16092801 16092803 16092802 16092805 16092804 16092800 16092775 16092773 16091885 16091820 16091818 16091227 16091258 16090687 16090406 16090401 16090325 16090308 16090399 16090202 16090204 16089912 16089562 16089561 16088912 16088913 16088914 16088911 16088909 16088910 16089000 16089001 16088999 16088915 16088917 16088919 16088920 16089022 16088921 16089021 16088905 16088369 16088371 16088370 16088365 16088366 16088364 16088367 16088362 16088363 16088361 16088368 16087688 16087687 16087692 16087694 16087811 16087808 16087809 16087813 16087691 16087812 16087689 16087696 16087695 16087690 16087693 16087009 16087006 16087002 16086998 16087001 16087007 16087008 16087003 16087004 16086485 16086487 16086488 16086489 16086486 16086058 16086062 16086059 16086067 16086061 16086072 16086075 16086068 16086074 16086063 16086079 16086069 16086077 16086066 16086070 16086073 16086065 16086071 16086078 16086055 16085447 16085444 16085440 16085675 16085674 16085446 16085677 16085676 16085443 16085441 16085442 16084959 16084954 16084962 16084960 16084955 16084961 16085412 16084963 16084502 16084466 16084465 16084492 16084469 16084464 16084471 16084470 16084463 16084028 16084024 16084025 16084026 16084020 16084023 16084022 16083755 16083757 16083747 16083752 16083751 16083748 16083756 16083753 16083749 16083754 16083489 16083483 16083497 16083493 16083498 16083495 16083496 16083494 16083492 16083501 16083499 16083490 16083500 16082788 16082802 16082800 16082797 16082798 16082799 16082189 16082239 16082187 16082213 16082214 16082216 16082217 16081624 16081522 16081525 16081312 16081478 16081524 16081052 16077584 16077585 16075071 16074261 16073427 16073423 16073425 16073529 16073530 16073454 16073456 16073518 16073458 16073462 16073514 16073494 16073522 16073523 16073515 16073520 16073516 16073419 16073418 16073486 16073420 16073468 16073491 16073525 16073490 16073432 16073448 16073443 16073437 16073488 16073430 16073435 16073436 16073440 16073464 16073466 16073467 16073481 16073526 16073528 16072260 16070529 16069111 16069143 16069123 16069124 16069118 16069117 16069119 16069125 16069114 16069116 16069121 16069115 16069113 16069120 16069122 16069112 16069126 16069142 16068553 16068557 16068550 16068551 16068552 16068554 16068556 16068555 16068039 16066740 16066060 16066061 16066049 16066062 16066051 16066050 16066058 16066055 16065033 16065031 16065032 16065246 16065250 16065034 16065035 16064701 16064702 16064700 16064699 16064703 16064705 16063911 16063826 16063250 16063244 16062430 16062431 16062432 16060924 16060910 16060914 16060913 16060923 16060922 16060921 16060918 16060915 16060920 16060919 16058636 16058635 16058110 16058111 16058116 16058205 16058114 16056921 16056926 16056934 16056920 16056923 16056933 16056935 16056929 16056930 16056922 16056928 16056927 16056467 16056463 16056395 16056466 16056464 16056394 16056400 16056399 16056398 16056397 16056401 16056465 16056396 16056470 16056469 16056471 16055967 16055302 16055213 16055121 16055230 16055229 16055231 16055243 16055234 16055123 16055124 16054986 16054923 16054917 16054916 16054919 16054921 16054983 16054985 16054920 16054922 16054987 16054918 16054663 16054661 16054660 16053361 16052084 16051741 16051737 16051734 16051736 16051731 16051738 16051732 16051733 16049572 16049602 16049573 16049596 16049575 16049574 16049576 16049577 16049580 16049607 16049606 16049609 16049592 16048771 16048764 16048765 16048769 16048767 16048768 16048636 16048634 16048637 16047561 16047564 16047565 16047566 16047562 16046165 16046169 16046170 16046162 16046167 16046163 16046164 16046168 16045065 16045112 16045109 16045107 16045113 16044053 16044075 16044082 16044091 16044100 16044090 16044086 16044093 16044094 16044084 16044097 16044098 16044081 16044087 16044088 16044095 16044096 16043229 16043221 16043226 16042994 44019580 16042131 16042046 16042047 16042123 16040911 16040919 16040917 16040913 16040914 16040915 16040918 16040729 16040728 16040732 16040730 16040731 16040194 16040233 16040234 16040231 16040193 16040154 16040065 16040063 16040064 16040066 16040067 16040068 16039769 16039767 16039772 16039768 16039774 16039771 16039770 16039776 16039729 16039556 16039561 16039565 16039553 16039555 16039315 16039316 16039317 16038622 16038620 16038624 16038621 16038623 16038618 16038626 16037627 16037628 16037631 16037630 16037632 16037633 16037634 16037635 16036885 16037352 16037350 16037351 16037349 16036880 16036875 16036872 16036884 16036390 16036389 16035949 16034791 16035157 16035155 16035156 16034341 16034342 16034344 16034345 16033122 16033124 16033123 16033125 16032398 16032399 16032397 16030464 16029669 43964026 43958073 43958061 16023039 16022057 16019470 16019472 16016146 16015777 16015806 16015795 16015778 16015771 16015770 16015802 16015804 16015768 16015803 16015792 16015791 16015796 16015782 16015801 16015776 16015789 16015766 16015774 16015779 16015780 16015784 16015786 16015785 16015767 16015772 16015797 16015799 16015800 16015775 16015783 16015793 16015788 16015773 16015790 16015798 16015794 16015781 16015805 16015769 16015787 16012871 16012866 16012864 16012873 16012869 16012865 16012867 16012860 16012872 16012164 43914334 43914378 43915150 43915157 43915128 43914388 43914382 43914390 43910326 43910385 16008004 16008006 43903214 43903210 16006036 16006044 16006039 16006034 16006045 16006041 16006035 16006043 16006042 16006030 16006029 16006028 16006027 16006046 16006037 16006038 16006040 16006033 16006032 16006031 43899554 16005392 16005391 16005136 16005132 16005133 43895678 43895682 16003931 16003516 16003484 16003518 16003495 16003482 16003501 16003510 16003503 16003493 16003508 16003514 16003491 16003507 16003505 16003512 16002712 16002713 16002030 16002120 16002124 16002112 16002109 16002110 16002111 16002123 16002107 16002122 16002108 16002036 16002034 16002125 16001055 16001045 16001048 16001039 16001049 16001042 16001040 16001041 16001070 16001043 16001038 16001069 16001046 16001071 16001054 16001044 16000751 16000752 16000754 16000750 16000755 16000753 16000745 16000759 16000746 16000748 16000756 16000760 16000749 16000758 16000747 16000762 16000761 16000481 16000234 43880149 43880150 43880151 43879115 43879258 43879259 15999945 15999946 15999947 15999943 15999942 15999948 15999944 15999425 15999440 15999443 15999428 15999429 15999426 15999421 15999418 15999420 15999413 15999414 15999422 15999412 15999442 15999411 15999423 15999424 15999416 15999419 15999415 15999444 15997860 15997859 15997871 15997861 15997870 15997862 15997873 15997863 15997869 15997878 15997867 15997866 15997877 15997875 15997858 15997864 15997874 15997868 15996672 15996667 15996659 15996664 15996663 15996655 15996665 15996668 15996662 15996661 15996656 15996657 15996658 43867610 43867690 43867691 15994366 15994368 15994364 15994371 15994370 15994365 15994367 15994369 15993972 15993958 15993966 15994033 15993963 15993969 15993965 15993968 15993967 15994032 15993962 15993964 15993959 15993971 15993961 15993970 15993957 15993960 15992165 15992178 15992169 15992177 15992187 15992176 15992174 15992171 15992175 15992181 15992172 15992170 15992179 15992173 15992182 15992162 15992188 15992167 15992186 15992168 15992184 15992183 15992185 15992164 15992180 15991738 15991741 15991785 15991784 15991736 15991737 15991739 15991742 15991744 15991783 15991749 15991731 15991732 15991786 15991795 15991733 15991734 15991728 15991798 15991730 15991729 15991726 15991725 15991797 15991724 15991790 15991791 15991793 15991792 15991794 15991796 15991789 15991727 15991735 15991787 15991788 43859910 15991650 15991649 15991651 15991648 43858201 43858192 43858180 43858186 43857553 43857558 43857897 43857563 43857568 43858197 43857583 43857573 43857542 43857543 43857578 43854813 43854814 15989190 15989189 15989194 15989196 15989197 15989193 15989195 15989200 15989201 15989191 15989192 15989198 15989199 15989202 43853028 43852984 43852969 43853015 43852927 43852974 43852979 43853037 43853020 43852892 43852932 43852907 15988587 15988586 15988639 15988581 15988580 15988589 15988574 15988575 15988582 15988584 15988583 15988579 15988577 15988592 15988901 15988898 15988588 15988573 15988899 15988591 15988576 15988900 15988902 15988590 15987755 15987757 15987756 15987751 43847963 43847675 43847655 43847630 43847635 43848507 43847660 43849070 43848516 43847686 43848510 43847646 43847665 43848519 43848504 43848495 43846437 43846895 43846824 43846903 43846837 43846889 43846926 43846809 43846918 43846852 43846911 43846845 43846817 43846884 43846840 43846929 43846872 43846832 43844688 43844266 43844555 43844575 43844546 43844262 43844572 43844566 43844549 15984461 15984460 15984464 15984462 15984465 15983616 15983618 15983617 15983615 15983619 15980602 15980600 15980604 15980589 15980588 15980599 15980591 15980590 15980609 15980595 15980593 15980606 15980608 15980605 15980601 15980597 15980607 15980592 15980598 15980610 15980596 15980594 15979513 15979512 15979509 15979508 15979510 15979518 15979516 15979511 15979524 15979517 15979522 15979520 15979521 15979523 15979519 15979514 15979515 15978031 15967376 15967374 15967375 15967386 15967390 15967414 15967408 15967388 15967383 15967381 15967380 15967415 15967362 15967385 15967402 15967403 15967399 15967377 15967401 15967398 15967393 15967378 15967379 15967396 15967406 15967407 15967394 15967384 15967389 15967404 15967391 15967405 15961979 15962018 15961986 15961996 15962002 15962003 15962006 15962008 15962016 15961998 15962019 15962000 15962021 15962013 15962001 15961989 15961987 15962011 15962009 15962014 15962004 15961999 15962005 15962007 15962020 15962010 15961994 15961993 15961990 15961981 15961992 15961991 15961995 15961982 15961984 15961985 15961997 15962012 15960099 15958444 15958445 15958434 15958435 15958436 15958463 15958452 15958440 15958453 15958450 15958428 15958456 15958448 15958447 15958430 15958431 15958446 15958437 15958421 15958426 15958422 15958443 15958427 15958442 15958449 15958423 15958429 15958457 15958462 15958424 15958420 15958438 15958451 15958458 15958461 15958433 15958432 15958459 15957628 15957624 15957607 15957618 15957608 15957603 15957605 15957606 15957602 15957598 15957617 15957615 15957600 15957614 15957610 15957612 15957599 15957609 15957627 15957611 15957596 15957601 15957613 15957622 15957616 15957626 15957597 15957625 15957002 15956999 15952364 15952378 15952361 15952366 15952362 15952367 15952363 15952352 15952354 15952356 15952369 15952376 15952353 15952371 15952368 15952358 15952360 15952359 15952355 15952365 15952357 15952374 15952375 15952373 15952372 15952351 15952370 15950396 15950398 15949568 15949569 15949576 15949577 15949578 15949558 15949553 15949559 15949580 15949560 15949555 15949550 15949595 15949552 15949575 15949579 15949556 15949594 15949557 15949567 15949561 15949566 15949573 15949571 15949572 15949563 15949562 15949565 15949564 15943766 15943774 15943759 15943758 15943752 15943761 15943760 15943755 15943778 15943771 15943751 15943765 15943777 15943753 15943754 15943773 15942157 15942158 15942159 15941662 15941663 15941677 15941682 15941696 15941685 15941660 15941669 15941658 15941664 15941659 15941679 15941681 15941675 15941687 15941701 15941666 15941661 15941671 15941692 15941690 15941311 15940620 15940593 15940368 15940371 15940387 15940380 15940394 15940405 15940378 15940384 15940382 15940408 15940398 15940385 15940372 15940393 15940410 15940401 15940377 15940369 15940370 15940407 15940412 15940396 15940391 15940114 15940124 15940125 15940138 15940139 15940141 15940142 15940143 15940144 15940145 15940118 15940112 15940116 15940099 15940119 15940105 15940086 15940096 15940085 15940097 15940130 15940090 15940103 15940126 15940101 15940094 15940129 15940123 15940134 15940092 15940083 15940107 15940106 15940110 15940108 15940109 15940135 15940136 15940137 15940146 15940147 15939081 15938055 15938054 15937957 15937578 15937582 15937577 15937574 15937576 15937575 15937581 15937587 15937586 15937585 15935612 15935615 15935596 15935629 15935613 15935621 15935618 15935620 15935619 15935622 15935586 15935626 15933737 15933712 15933706 15933727 15933726 15933738 15933728 15933725 15933721 15933746 15933750 15933748 15933723 15933717 15933724 15933735 15933744 15933743 15933719 15933057 15933056 15932837 15932835 15932841 15932832 15932823 15932833 15932831 15932839 15932829 15932827 15932828 15932836 15932821 15932822 15932834 15931343 15931008 15931016 15931019 15930653 15930748 15930758 15930652 15930745 15930796 15930754 15930750 15930795 15930794 15930746 15929989 15929559 15929558 15928765 15928764 15928762 15928763 15928168 15928169 15927991 15927984 15927980 15927971 15927973 15927974 15927986 15927985 15927982 15927983 15927981 15927992 15927989 15927990 15925590 15925593 15925592 15925589 15925595 15925594 15925599 15925598 15925602 15925601 15924566 15924567 15924535 15924534 15924541 15924533 15924542 15924539 15924538 15923966 15923969 15923716 15922819 15922821 15922823 15922810 15922808 15922813 15922814 15922812 15922809 15921441 15921430 15921439 15921466 15921431 15921456 15921457 15921459 15921450 15921452 15921454 15920018 15920016 15920000 15920020 15919982 15919983 15919998 15920004 15920006 15919999 15920005 15919997 15920002 15919981 15920007 15920011 15920003 15920010 15914133 15914130 15914125 15914132 15914131 15914136 15914113 15914122 15914121 15914114 15914116 15914120 15914115 15914119 15914117 15914123 15912620 15912621 15910808 15910807 15910849 15910848 15910837 15910834 15910829 15910831 15910845 15910828 15910846 15910833 15910169 15906182 15906065 15906051 15906069 15906046 15906062 15906088 15906078 15906076 15906047 15906082 15906049 15906060 15906059 15906084 15906087 15906055 15906058 15906080 15906053 15906071 15906077 15906073 15906075 15906056 15906079 15906052 15906066 15902416 15902424 15902415 15902423 15902420 15901343 15895496 15895423 15895427 15895425 15895424 15895420 15895422 15888951 15888952 15888954 15888965 15888959 15888957 15888961 15888962 15879989 15879984 15879985 15879991 15879987 15879998 15879976 15879982 15872415 15872422 15872430 15872419 15872421 15872427 15872418 15872411 15872414 15872412 15867348 15867344 15867349 15867347 15867341 15867343 15867337 15867340 15867338 15867336 15867342 15867335 15855651 15855650 15855649 15855652 15855653 15855489 15855492 15855496 15855495 15855493 15855494 15855491 15855198 15855085 15855080 15854989 15854748 15854735 15854729 15854734 15854737 15854732 15854736 15854740 15854741 15854745 15854739 15854743 15854742 15854744 15854738 15854592 15854198 15854201 15854203 15854204 15854192 15854193 15854202 15854217 15854196 15854206 15854194 15853794 15853295 15853299 15853300 15853298 15853303 15853296 15853043 15853045 15853056 15853057 15853053 15853051 15853046 15853048 15853055 15853041 15853042 15853059 15853049 15853044 15853058 15853052 15852900 15835058 15835059 15834992 15832664 15832666 15832657 15832668 15832670 15832663 15832655 15832667 15832651 15832671 15832662 15832654 15832656 15832665 15832649 15832660 15832652 15832676 15830892 15824208 15824231 15824236 15824211 15824212 15824224 15824229 15824235 15824230 15824222 15824220 15815829 15815821 15815834 15815827 15815832 15815831 15815818 15815833 15815824 15815820 15815828 15815822 15815817 15815840 15806707 15806718 15806711 15806706 15806720 15806694 15806691 15806689 15806699 15806713 15806698 15806692 15806717 15806687 15806695 15806701 15806708 15806716 15803799 15803598 15803606 15803611 15803621 15803617 15803614 15803616 15803615 15803609 15803612 15803605 15803620 15803603 15803610 15803602 15803601 15788872 15788870 15788868 15786152 15786168 15786147 15786161 15786155 15786160 15786149 15786156 15786146 15786166 15786151 15786163 15786169 15786164 15786171 15786157 15786158 15786148 15786167 15786172 15783807 15783805 15783781 15783791 15783793 15783798 15783790 15783785 15783786 15783800 15783801 15783808 15783802 15783787 15783788 15783778 15783780 15783794 15783796 15783804 15783789 15783806 15779891 15775769 15775767 15775750 15775780 15775774 15775759 15775764 15775772 15775776 15775751 15775755 15775756 15775765 15775775 15775781 15774188 15770172 15770179 15770181 15770187 15770207 15770215 15770217 15770221 15770194 15770197 15770205 15770210 15770219 15770202 15770177 15770223 15770186 15761382 15761397 15761388 15761386 15761389 15748541 15748543 15748544 15747072 15747087 15747056 15747062 15743375 15743374 15743384 15743378 15743382 15743386 15743368 15743369 15743388 15743380 15743370 15743379 15743385 15743371 15743387 15743391 15743392 15736824 15736832 15736823 15736438 15736430 15736424 15736427 15736440 15736437 15736441 15728657 15728652 15728656 15728641 15728648 15728642 15728666 15723806 15723817 15723814 15723800 15723809 15723821 15723816 15723797 15723818 15723804 15723808 15715742 15715748 15715761 15715745 15715760 15715735 15715739 15715741 15715746 15715764 15715759 15715766 15713952 15713950 15709847 15709875 15709865 15709862 15693279 15691543 15684799 15684812 15684808 15684803 15684807 15684816 15684806 15684813 15684815 15684802 15684809 15684817 15684810 15684811 15684801 15684814 15682324 15679247 15679245 15679269 15679254 15679243 15679263 15679237 15679417 15671102 15666771 15666762 15666764 15662821 15662824 15662830 15662831 15662814 15662813 15662829 15662827 15662816 15662804 15662807 15662833 15662832 15662818 15662820 15662835 15662811 15661682 15661681 15654701 15654698 15654695 15654709 15654718 15654702 15654721 15654711 15654722 15654717 15654720 15654704 15654697 15654710 15654712 15654714 15653519 15652145 15652147 15652138 15652143 15652059 15650666 15647364 15645826 15645828 15645830 15645825 15643124 15643125 15643128 15639702 15639704 15639708 15639699 15639676 15639721 15639695 15639700 15639697 15639684 15639694 15639679 15639715 15639693 15639707 15639687 15639724 15639718 15639720 15639729 15639713 15639674 15639683 15639705 15639714 15639690 15639698 15639688 15638496 15638495 15633002 15632966 15632987 15632988 15632982 15632975 15632967 15632979 15632973 15632983 15632990 15632970 15632978 15632971 15632984 15632976 15632969 15632965 15632974 15632991 15632989 15632980 15631112 15631098 15631093 15631107 15631125 15631116 15631119 15631109 15631122 15631104 15631099 15631126 15631108 15631094 15631097 15631114 15631101 15631106 15631117 15622613 15622611 15622635 15617717 15618006 15614585 15613096 15606160 15606155 15606156 15606149 15606144 15606152 15606143 15606154 15606170 15606150 15606167 15606151 15606161 15606164 15606159 15606165 15606168 15606163 15600761 15600778 15600763 15600770 15600762 15600767 15600777 15600775 15600768 15600783 15600782 15600784 15600757 15600756 15600771 15600776 15600766 15600786 15600773 15600781 15598106 15598091 15598092 15598095 15598109 15598104 15598083 15598093 15598105 15598088 15590119 15590120 15588927 15587203 15587204 15587205 15584199 15583141 15581964 15581965 15575589 15575994 15575976 15575974 15575978 15575979 15575983 15575995 15575990 15575984 15575991 15575977 15569626 15569607 15569629 15569608 15569630 15569612 15569624 15569616 15569627 15569625 15569613 15569621 15569604 15569606 15569605 15569623 15569618 15569610 15564894 15564893 15564896 15564895 15564897 15562706 15561697 15561264 15561265 15561268 15561275 15561251 15561254 15561273 15561272 15561255 15561259 15561252 15561258 15561262 15561253 15561261 15561256 15561270 15557611 15554707 15550857 15550844 15550852 15550848 15550853 15550858 15550828 15550861 15550839 15550855 15550834 15550829 15550838 15550842 15550846 15550850 15550847 15550835 15550860 15550843 15547821 15546742 15537380 15521201 15521200 15521199 15521203 15519704 15519700 15518256 15518261 15518267 15518262 15518259 15518260 15518266 15518270 15518275 15518276 15518271 15518268 15518269 15518273 15518272 15518265 15518258 15518257 15518274 15517161 15513835 15513831 15513825 15513826 15513833 15513834 15513830 15509359 15509361 15509364 15509365 15509366 15506570 15504839 15504833 15504835 15504837 15504836 15504832 15504822 15504819 15504820 15504830 15504825 15504829 15504834 15504840 15504838 15503514 15503510 15503515 15503511 15503518 15503519 15503509 15503516 15500854 15500700 15500693 15500698 15500704 15500696 15500689 15500695 15500697 15500685 15500702 15500691 15500706 15498908 15499064 15492248 15492211 15492649 15492260 15492266 15492267 15492271 15492259 15492265 15492261 15492270 15492278 15492273 15492268 15492272 15492275 15492258 15492274 15492276 15492263 15492257 15492264 15492262 15490744 15490718 15490719 15490742 15490720 15490721 15490734 15490740 15490722 15490723 15490724 15490725 15490726 15490727 15490728 15490731 15490729 15490747 15490748 15490751 15490732 15490735 15490733 15490743 15490736 15490737 15490739 15490741 15490730 15490745 15490753 15490752 15490754 15489245 15489243 15489244 15483513 15475980 15474596 15474551 15473692 15473693 15471183 15470033 15470035 15470034 15467518 15464898 15463097 15463095 15463096 15463216 15463220 15463217 15463219 15463218 15461858 15460996 15460997 15460705 15460706 15460707 15452296 15452358 15452359 15442786 15442794 15442792 15442787 15442793 15442791 15442788 15437809 15437811 15437831 15437814 15437820 15437825 15437828 15437816 15437824 15437812 15437819 15437826 15437830 15437827 15437833 15437815 15437817 15437821 15437823 15430952 15430964 15430948 15430963 15430966 15430951 15430946 15430965 15430947 15430958 15430959 15430944 15430961 15430955 15430956 15430960 15430954 15422138 15422139 15422141 15422129 15422127 15422131 15422133 15422154 15422147 15422152 15422148 15422153 15422142 15422132 15422128 15422137 15422140 15422136 15422145 15422134 15422143 15422130 15410884 15410881 15410886 15410891 15410887 15410882 15410892 15410883 15410885 15410894 15410900 15410902 15410895 15410893 15410896 15410897 15410898 15398897 15398911 15398901 15398909 15398899 15396726 15393558 15392443 15392451 15392449 15392447 15391311 15391644 15389013 15389673 15389562 15389564 15389566 15389567 15389559 15389565 15389570 15389561 15389563 15383483 15383487 15383471 15383475 15383486 15383480 15383481 15383478 15383474 15383473 15383470 15383484 15383485 15383477 15383469 15383476 15383488 15383472 15376276 15376146 15371603 15371600 15371599 15369939 15366948 15366949 15366950 15366947 15366326 15366328 15366316 15366325 15366329 15366309 15366310 15366317 15366322 15366323 15366318 15366324 15366315 15366319 15366313 15366314 15366311 15366327 15366321 15361559 15361555 15361553 15361554 15356965 15356968 15356977 15355354 15353919 15353870 15353872 15353889 15353886 15353888 15353878 15353883 15353884 15353876 15353891 15353874 15353879 15353890 15353905 15353894 15353906 15353898 15353897 15353896 15353893 15353895 15353902 15353892 15353899 15353903 15353901 15353904 15353868 15340242 15340240 15340245 15340241 15340239 15340243 15340251 15337472 15331136 15328507 15328493 15328499 15328491 15328514 15328498 15328510 15328496 15328502 15328492 15328511 15328509 15328495 15328494 15328512 15328503 15328500 15328501 15326804 15322152 15322148 15322145 15322318 15322321 15322319 15322325 15322322 15322320 15322317 15322316 15322323 15322324 15322326 15318393 15318400 15318395 15317099 15317097 15317104 15317101 15317102 15317100 15315572 15315526 15315573 15314017 15312346 15312347 15312344 15312343 15309654 15308357 15308353 15308354 15305242 15305249 15305090 15305089 15303223 15303224 15303222 15302367 15299942 15299938 15299940 15298294 15298296 15298295 15295794 15295766 15295775 15295774 15295777 15295781 15295765 15295787 15295784 15295809 15295778 15295773 15295805 15295797 15295768 15295783 15295790 15295802 15295782 15295786 15295785 15295815 15295811 15295772 15295788 15295800 15295780 15295771 15295767 15295776 15295769 15295799 15294475 15294474 15290669 15290668 15290673 15290675 15279452 15279453 15279455 15279454 15279442 15279443 15279444 15279458 15276088 15276086 15276090 15274578 15274579 15274576 15268831 15268840 15268836 15268835 15268851 15268838 15268856 15268850 15267546 15267502 15267544 15267501 15267549 15267500 15263414 15260921 15260923 15257954 15257622 15254778 15254776 15254774 15254775 15254772 15253442 15245775 15245774 15245776 15245777 15242596 15242606 15242603 15242602 15242605 15242599 15240793 15236627 15229142 15229109 15229116 15229115 15229107 15229108 15229119 15229118 15223562 15223570 15223560 15223568 15223571 15222620 15216161 15216160 15213239 15210095 15210091 15210094 15210089 15210104 15210088 15210090 15210106 15206270 15206267 15206269 15201877 15201878 15192525 15190476 15188733 15182585 15182137 15176580 15176582 15175656 15080056 15041225 14813912 14787922 14750087 14750089 14737824 14737827 14723237 14699415 14697051 14697048 14697049 14649204 14635120 14607363 14603162 14603159 14603157 14533097 14473362 14473367 14473370 14473365 14473360 14473361 14473366 14473372 14472490 14443845 14443835 14443832 14419505 14414503 14404307 14395945 14364547 14364542 14347748 14338618 14314757 14309974 14297227 14254813 14177214 14159916 14101916 14036852 14017694 14011500 14006853 13985115 13948724 13939451 13909109 13884848 13885104 13883357 13873024 13853918 13832440 13826187 13727050
FEI Number 2016493
Recalling Firm/
Manufacturer
CareFusion 303, Inc.
10020 Pacific Mesa Blvd
San Diego CA 92121-4386
Manufacturer Reason
for Recall
Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
FDA Determined
Cause 2
Use error
ActionOn 7/10/2025, correction notices were mailed and emailed to customers who were informed of the following: Firm strengthened its user guide content regarding manual access to medications. To access the updated content for the user guide and nursing training material, visit the MedBank Customer Portal at https://bd.com/medbankportal When selecting a medication from override/add item function, it is important for users to use extra precautions and verify correct medications are selected and pulled as undesired medications could be displayed on the override list. Firm recommends the following actions to prevent delays in access to medication: 1) Make sure staff know how to use the override/add-item workflow. 2) Have device keys available on-site in a secure location known to designated staff. 3) Train staff on how to manually access the device, see User Guide. 4) Ensure downtime policies and procedures are implemented and staff are appropriately trained Enable Over-ride permission. 5) Do not use the station as a crash cart or code box. Medication or items may not be immediately accessible for an emergent situation causing a delay in treatment which could lead to serious injury or death. 6) For pharmacies providing products to clinical users that are not physically located at the pharmacy site, ensure clinical users are aware of the risk of delay in access and provide access to keys. 7) Protocols for retrieving medications from alternative inventory locations 8) Guidance on storing time-sensitive or critical medications outside of automated dispensing devices to prevent access delays. 9) Complete and return the response form via email to BDRC9@bd.com Distributors: provide a copy of this customer letter to any customers who may have received impacted devices. Firm is evaluating design enhancements to improve usability of the override/Add Item workflow. Report any device issues via phone: 866-930-9251
Quantity in Commerce6076
DistributionUS Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH, FL, WA, MS, OK, TN, PA, RI, TX, SC, AR, MI, NJ, CT, MD, AZ, CA, WV, NC, UT, WY, CO, SD, ID, IL, IN, NV, NM, NE, KS, GA, WI, MO, IA, ME, LA, MN, AL, VT, NH, DE, MT, ND, AK, HI, DC.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-