| | Class 2 Device Recall Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 |  |
| Date Initiated by Firm | July 15, 2025 |
| Date Posted | September 24, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2620-2025 |
| Recall Event ID |
97285 |
| Product Classification |
Light, surgical, ceiling mounted - Product Code FSY
|
| Product | Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 |
| Code Information |
UDI-DI: 07613327296167 and 07613327651317
REF/Serial Numbers:
CH00000001/7000010116, 7000010117, 7000010118, 7000010119, 7000010120, 7000010121, 7000010122, 7000010123, 7000010124, 7000010125, 7000010141, 7000010142, 7000010177, 7000010178, 7000010148, 7000010149, 7000010157, 7000010137, 7000010175, 7000010176, 7000010145, 7000010146, 7000010147, 7000010309, 7000010305, 7000010307, 7000010296, 7000010143, 7000010150, 7000010151, 7000010155.
P60034/1000022, 1000023, 1000024, 1000016, 1000017, 1000018, 1000019, 1000020, 1000021 |
| FEI Number |
3002975342
|
Recalling Firm/ Manufacturer |
Stryker Communications 571 Silveron Blvd Flower Mound TX 75028
|
| For Additional Information Contact | 972-410-7100 |
Manufacturer Reason for Recall | Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
|
FDA Determined Cause 2 | Component change control |
| Action | On 7/15/2025, recall notices were mailed or hand delivered to customers who were asked to do the following:
Inform individuals within your organization who need to be aware of this action.
2. A Stryker representative will be in contact with your account to schedule a field service visit at your earliest convenience to complete the following:
a. Deinstall impacted Oculan light(s) and reinstall with SLX lights to ensure your operating room stays in use.
b. Upon product remediation and availability, Stryker will return to reinstall Oculan within your operating room(s). Your local Stryker representative will provide timing updates as they are available.
3. Complete and return the Business Reply Form to CommunicationsRA1@stryker.com
4. Maintain awareness of this communication internally until all required actions have been completed within your facility. |
| Quantity in Commerce | 40 |
| Distribution | US Nationwide distribution in the states of TX, IL, NY, UT, OR, AZ, VT, NY, IL, MN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|