Class 2 Device Recall Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034see related information
Date Initiated by FirmJuly 15, 2025
Date PostedSeptember 24, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2620-2025
Recall Event ID 97285
Product Classification Light, surgical, ceiling mounted - Product Code FSY
ProductChromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
Code Information UDI-DI: 07613327296167 and 07613327651317 REF/Serial Numbers: CH00000001/7000010116, 7000010117, 7000010118, 7000010119, 7000010120, 7000010121, 7000010122, 7000010123, 7000010124, 7000010125, 7000010141, 7000010142, 7000010177, 7000010178, 7000010148, 7000010149, 7000010157, 7000010137, 7000010175, 7000010176, 7000010145, 7000010146, 7000010147, 7000010309, 7000010305, 7000010307, 7000010296, 7000010143, 7000010150, 7000010151, 7000010155. P60034/1000022, 1000023, 1000024, 1000016, 1000017, 1000018, 1000019, 1000020, 1000021
FEI Number 3002975342
Recalling Firm/
Manufacturer
Stryker Communications
571 Silveron Blvd
Flower Mound TX 75028
For Additional Information Contact
972-410-7100
Manufacturer Reason
for Recall
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
FDA Determined
Cause 2
Component change control
ActionOn 7/15/2025, recall notices were mailed or hand delivered to customers who were asked to do the following: Inform individuals within your organization who need to be aware of this action. 2. A Stryker representative will be in contact with your account to schedule a field service visit at your earliest convenience to complete the following: a. Deinstall impacted Oculan light(s) and reinstall with SLX lights to ensure your operating room stays in use. b. Upon product remediation and availability, Stryker will return to reinstall Oculan within your operating room(s). Your local Stryker representative will provide timing updates as they are available. 3. Complete and return the Business Reply Form to CommunicationsRA1@stryker.com 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Quantity in Commerce40
DistributionUS Nationwide distribution in the states of TX, IL, NY, UT, OR, AZ, VT, NY, IL, MN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-