| | Class 2 Device Recall STA Satellite |  |
| Date Initiated by Firm | July 21, 2025 |
| Date Posted | August 28, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2473-2025 |
| Recall Event ID |
97287 |
| 510(K)Number | K082248 |
| Product Classification |
System, multipurpose for in vitro coagulation studies - Product Code JPA
|
| Product | STA Satellite. Product Code: All references. All software versions.
The STA Satellite is a fully automated benchtop analyzer capable of simultaneously performing clotting, chromogenic, and immunologic assays. |
| Code Information |
Product Code: All references. UDI-DI: 03607450581041, 03607450581027. All software versions.
|
| FEI Number |
2245451
|
Recalling Firm/ Manufacturer |
Diagnostica Stago, Inc.
5 Century Dr
Parsippany
NJ
07054-4607
|
| For Additional Information Contact | Technical Hotline 800-7250607 |
Manufacturer Reason for Recall | Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications. |
FDA Determined Cause 2 | Mixed-up of materials/components |
| Action | Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed. |
| Quantity in Commerce | 3,650 units (538 US) |
| Distribution | Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JPA
|
|
|
|