| | Class 2 Device Recall FRED |  |
| Date Initiated by Firm | September 05, 2025 |
| Date Posted | October 10, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0108-2026 |
| Recall Event ID |
97580 |
| PMA Number | P180027 P180027S002 |
| Product Classification |
Intracranial aneurysm flow diverter - Product Code OUT
|
| Product | Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827;
FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X |
| Code Information |
REF/UDI-DI(Lot):
FRED5019/00811425020388(0000235839), 00842429117163(0000544498);
FRED5029/00811425020395(0000256343), 00842429117170(0000507209);
FRED4528/00811425020425(0000265666);
FRED4026/00811425021033(0000333125, 0000576973);
FRED5526/00842429117194(0000361117);
FRED4508/00842429117095(0000372088);
FRED5514/00842429117187(0000429786, 0000478385);
FRED3536/00842429117033(0000460268, 0000445555);
FRED3511/00842429117002(0000452923);
FRED5014/00842429117156(0000563591, 0000464401);
FRED3516/00842429117019(0000513146, 0000531733, 0000536217);
MV-F451827/04987892121906(0000565094);
XFRED4017-MVE/00842429107508(0000478838);
XFRED4528-MVE/00842429107560(0000518323, 0000663751);
MV-F501427X/04987892151231(0000579031);
XFRED4518-MVE/00842429107553(0000663074, 0000663750);
XFRED5526-MVE/00842429107638(0000704203);
XFRED4539-MVE/00842429107577(00007311620);
FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155);
FREDX3522-PMA/00842429114315(0000469598);
FREDX4525-PMA/00842429114414(0000469603);
FREDX5015-PMA/00842429114445(0000469606);
FREDX5532-PMA/00842429114490(0000469610);
FREDX4013-PMA/00842429114346(0000469613, 0000663758, 0000668189);
FREDX3513-PMA/00842429114292(0000478834);
FREDX4515-PMA/00842429114391(0000478840);
FREDX3517-PMA/00842429114308(0000731304);
MV-F352427X/04987892151101(0000760963)
|
Recalling Firm/ Manufacturer |
MICROVENTION INC. 35 Enterprise Aliso Viejo CA 92656
|
| For Additional Information Contact | Terrence Callahan 949-899-7194 |
Manufacturer Reason for Recall | Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences. |
FDA Determined Cause 2 | Device Design |
| Action | On 9/5/2025, recall notices were mailed to customers who were asked to do the following:
1) Stop using and quarantine affected devices.
2) Provide the notice to medical facilities or users to whom you have distributed recalled devices.
3) Return effected devices.
4) Complete and return the customer acknowledgement form via email to Terumo8943@sedgwick.com
Direct questions to the firm at terrence.callahan@microvention.com
On 10/29/2025, additional recall notices were distributed identifying additional distributed devices. |
| Quantity in Commerce | 467 |
| Distribution | Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|