• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Abbott Medical TactiFlex

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Abbott Medical TactiFlexsee related information
Date Initiated by FirmSeptember 10, 2025
Date PostedOctober 10, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0112-2026
Recall Event ID 97595
PMA NumberP220013 
Product Classification Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation - Product Code OAE
ProductTactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter
Code Information GTIN 05415067034618, Batch Numbers: 10093213, 10093588, 10095366, 10096269, 10118676, 10131389, 10151333, 10180969, 10286877, 10315661, 10317649, 10329929, 10332297, 10344177, 10350400, 10360030, 10374171, 10375844, 10418954, 10431765, 10433617, 10450462, 10492093, 10494298, 10537000, 10537747, 10538321, 10557710, 10557878, 10558450, 10559655, 10569703, 10570507, 10571902, 10571907, 10574230, 10575746, 10575752, 10577567, 10578953, 10579724, 10580699, 10581612, 10581681, 10583203, 10583793, 10587033, 10587043, 10587653, 10589068, 10589921, 10591178, 10591238, 10592227, 10606628, 10606720, 10608438, 10608708, 10611469, 10611683, 10612686, 10614973, 10616329, 10618038, 10619154, 10621180, 10622172, 10622962, 10622967, 10625101, 10630682, 10630687, 10632744, 10633145, 10634830, 10636059, 10637023, 10639373, 10639374, 10639376, 10639377, 10641739, 10641741, 10642345, 10643887, 10644681, 10645838, 10646429, 10648576, 10662844, 10665077, 10666137, 10667210, 10667217, 10668065, 10668185, 10670353, 10670360, 10671604, 10671663, 10672934, 10674544, 10675329, 10675541, 10677719, 10677722, 10678784, 10680052, 10680962, 10684449, 10691502, 10691504, 10691507, 10692805, 10696082, 10696083, 10696084, 10705825, 10708749, 10708800, 10709735, 10711022, 10711803, 10712639, 10712641, 10714607, 10714611, 10715850, 10717049, 10717050, 10719520, 10721238, 10721239, 10725441, 10725442, 10747196, 10749817, 10749820, 10749823, 10751484, 10762186, 10762464, 10764517, 10764523, 10765657, 10766897, 10767820, 10769643, 10770889, 10772623, 10772625, 10773775, 10773811, 10776251, 10777548, 10778157, 10779451, 10808389, 10808390, 10808393, 10811066, 10811067, 10811262, 10821034, 10827717, 10829276, 10829990, 10830744, 10832015, 10832019, 10832727, 10834411, 10834412, 10835187, 10836468, 10836482, 10836911, 10837558, 10837559, 10839959, 10839960, 10841679, 10841690, 10842962, 10845198, 10845199, 10846600, 10847688, 10847695, 10850757, 10850761, 10856004, 10856065, 10858260, 10867490, 10867495, 10868706, 10873737, 10873924, 10875772, 10875773, 10877194, 10897765, 10899500, 10899504, 10902557, 10902569, 10905178, 10905907, 10906368, 10907259, 10909995, 10910050, 10911332, 10911333, 10913977, 10915227, 10915230, 10920111, 10922315, 10922528, 10924643, 10924802, 10927218, 10954504, 10954507, 10955716, 10956520, 10956521, 10958150, 10959364, 10959373, 10960184, 10962699, 10965714, 10970014, 10970054, 10985513, 10989401, 10990017, 10990759, 10992074, 10992078, 10987781.
FEI Number 3005334138
Recalling Firm/
Manufacturer
Abbott
5050 Nathan Ln N
Plymouth MN 55442-3209
For Additional Information ContactShelley Lange
651-756-4091
Manufacturer Reason
for Recall
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
FDA Determined
Cause 2
Package design/selection
ActionAbbott issued a Medical Device Advisory to its consignees on 09/10/2025 by letter delivered via Federal Express. The notice explained the issue, impact and associated risk to the patient and requested the following: "Action Requested/Recommendations/Next Steps . Continue following the product IFU by inspecting the packaging and catheter for damage and/or kinking prior to use. . Review the guidance in Appendix A information on proper removal technique of the TactiFlex" Ablation Catheter, Sensor Enabled" from the packaging tray, ensuring to inspect the electrodes and catheter tip carefully for integrity and overall condition prior to use. . Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal from the packaging. The deployment of the new packaging has begun. Please complete the provided Acknowledgement Form with the assistance of your Abbott representative. This form ensures you are aware of the advisory notice and have been given an opportunity to ask questions. Please forward this notice to anyone in your organization who may need to be notified and maintain this notice for your records to ensure effectiveness of this communication. Should you have questions about this issue, please contact your local Abbott representative or Abbott Support at 1-855-478-5833 (U.S), 8:30 a.m. - 5:30 p.m. Central Standard Time, Monday thru Friday."
Quantity in Commerce11810 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = OAE
-
-