| | Class 2 Device Recall Philips |  |
| Date Initiated by Firm | September 29, 2025 |
| Date Posted | October 23, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0241-2026 |
| Recall Event ID |
97682 |
| 510(K)Number | K183101 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Zenition 50;
Product Number (REF): 718096; Interventional fluoroscopic x-ray system |
| Code Information |
Product Number (REF): 718096;
UDI-DI: 00884838091535;
All serial numbers;
|
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
|
Manufacturer Reason for Recall | Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board. |
FDA Determined Cause 2 | Process control |
| Action | On September 29, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this notice to all users and follow the instructions in the letter. 2. Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). 3. Ensure the letter is in a place likely to be seen/viewed. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Starting in October 2025, Philips will replace the 11W control board of the affected Zenition 50 systems.
Your local Philips representative will contact you to schedule a visit to perform this activity in your system. |
| Quantity in Commerce | 29 units |
| Distribution | Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|