| Date Initiated by Firm | November 18, 2025 |
| Date Posted | December 12, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0920-2026 |
| Recall Event ID |
97969 |
| 510(K)Number | K220058 |
| Product Classification |
System, network and communication, physiological monitors - Product Code MSX
|
| Product | Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1
HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1) |
| Code Information |
Audio Driver Version: 6.0.9484.1. UDI-DI: 06936415972417. HP Pro Mini 400 G9 Desktop PC. Realtek HD Audio Driver Released by HP.
1. WS host package (mini PC). Part Number 115-050935-00. Quantity 902
2. Exch, Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4
|
Recalling Firm/ Manufacturer |
Mindray DS USA, Inc. dba Mindray North America 800 Macarthur Blvd Mahwah NJ 07430-2001
|
| For Additional Information Contact | James Clifford +1-201-9958407 |
Manufacturer Reason for Recall | When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing. |
FDA Determined Cause 2 | Software design |
| Action | Mindray notified consignees of the recall on about 11/18/2025 via certified mail. Consignees were informed that they may continue to use the Central Monitoring System. If the HP Pro Mini 400 G9 Desktop PC used by the WS experiences the audio failure or screen freezes, restart the HP Pro Mini 400 G9 Desktop PC. Consignees were instructed to ensure applicable clinical staff are aware of the issue and letter and shared with any customer if further distributed. Consignees were also asked to complete and return the Acknowledgement and Receipt form provided. Mindray will provide a patch update for the audio driver of the affected HP Pro Mini 400 G9 Desktop PCs located at each facility and is anticipated to take approximately 10-15 minutes per HP Pro Mini 400 G9 Desktop PC. A representative from Mindray's Care team will contact consignees to schedule the update. |
| Quantity in Commerce | 906 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|