| | Class 2 Device Recall Inflation Syringe |  |
| Date Initiated by Firm | December 04, 2025 |
| Date Posted | January 16, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1123-2026 |
| Recall Event ID |
98139 |
| Product Classification |
Syringe, Balloon Inflation, Exempt - Product Code PTM
|
| Product | Merit Medical Inflation Device.basixCOMPAK.30 atm/bar.20 mL
Reference Numbers: ENDO-AN2030/B IN4130/CNK IN4130/H IN4130/JPH IN4130/K IN4130/KRK IN4130/L IN4130/T IN4230/JPH IN4230/K IN4330/CNK IN4330/K IN4352/K IN4430/K IN4530/CNK IN4530/K IN4530/T IN4802/K IN4802/T IN4852/K IN4901/A IN4902/A IN4903/A IN4904/A IN4905/A IN4906/A IN4907/A IN4908 IN4909 IN4910 IN4911/A IN4912 IN4913/A IN4914/A IN4915 IN4916/A IN4917/A IN4918/A IN4919/A |
| Code Information |
Reference Number/UDI Number/Lot Number(s)
ENDO-AN2030/B 00884450645161 I2559664
00884450645161 I2617758
00884450645161 I2622046
00884450645161 I2657805
00884450645161 I2937082
00884450645161 I3056430
00884450645161 I3085664
00884450645161 I3118225
00884450645161 I3251180
00884450645161 I3282751
IN4130/CNK 00884450636695 I2559550
00884450636695 I2559551
00884450636695 I2563338
00884450636695 I2563339
00884450636695 I2563340
00884450636695 I2564655
00884450636695 I2564665
00884450636695 I2567481
00884450636695 I2567482
00884450636695 I2567483
00884450636695 I2578099
00884450636695 I2578100
00884450636695 I2584021
00884450636695 I2584024
00884450636695 I2584025
00884450636695 I2584026
00884450636695 I2607299
00884450636695 I2607301
00884450636695 I2607302
00884450636695 I2617759
00884450636695 I2617760
00884450636695 I2628998
00884450636695 I2629000
00884450636695 I2629001
00884450636695 I2629002
00884450636695 I2630702
00884450636695 I2648658
00884450636695 I2648659
00884450636695 I2648660
00884450636695 I2648661
00884450636695 I2657828
00884450636695 I2657833
00884450636695 I2657884
00884450636695 I2657889
00884450636695 I2663025
00884450636695 I2663026
00884450636695 I2663027
00884450636695 I2669899
00884450636695 I2669900
00884450636695 I2674531
00884450636695 I2674532
00884450636695 I2682701
00884450636695 I2682702
00884450636695 I2684463
00884450636695 I2684522
00884450636695 I2684523
00884450636695 I2699591
00884450636695 I2699592
00884450636695 I2703002
00884450636695 I2703003
00884450636695 I2703004
00884450636695 I2703006
00884450636695 I2703081
00884450636695 I2703082
00884450636695 I2703083
00884450636695 I2707735
00884450636695 I2707736
00884450636695 I2715012
00884450636695 I2715013
00884450636695 I2719928
00884450636695 I2719929
00884450636695 I2719930
00884450636695 I2719931
00884450636695 I2719933
00884450636695 I2752077
00884450636695 I2752078
00884450636695 I2755218
00884450636695 I2755219
00884450636695 I2755496
00884450636695 I2760565
00884450636695 I2773971
00884450636695 I2773972
00884450636695 I2784600
00884450636695 I2784603
00884450636695 I2844301
00884450636695 I2853612
00884450636695 I2853613
00884450636695 I2853615
00884450636695 I2858649
00884450636695 I2858650
00884450636695 I2858651
00884450636695 I2869290
00884450636695 I2886413
00884450636695 I2886423
00884450636695 I2886429
00884450636695 I2886432
00884450636695 I2887831
00884450636695 I2887832
00884450636695 I2887836
00884450636695 I2888430
00884450636695 I2892032
00884450636695 I2892033
00884450636695 I2894577
00884450636695 I2902444
00884450636695 I2902445
00884450636695 I2902446
00884450636695 I2902447
00884450636695 I2903533
00884450636695 I2903535
00884450636695 I2917165
00884450636695 I2917167
00884450636695 I2921904
00884450636695 I2921905
00884450636695 I2927911
00884450636695 I2933517
00884450636695 I2937693
00884450636695 I2937694
00884450636695 I2971987
00884450636695 I2973163
00884450636695 I2973184
00884450636695 I2976519
00884450636695 I2976520
00884450636695 I2976521
00884450636695 I2983704
00884450636695 I2986048
00884450636695 I2986049
00884450636695 I2989136
00884450636695 I2989137
00884450636695 I3001128
00884450636695 I3001129
00884450636695 I3001130
00884450636695 I3001131
00884450636695 I3013083
00884450636695 I3013085
00884450636695 I3014308
00884450636695 I3014309
00884450636695 I3014311
00884450636695 I3014312
00884450636695 I3044881
00884450636695 I3048387
00884450636695 I3056432
00884450636695 I3056433
00884450636695 I3067548
00884450636695 I3067549
00884450636695 I3072605
00884450636695 I3074753
00884450636695 I3074754
00884450636695 I3074755
00884450636695 I3074756
00884450636695 I3079091
00884450636695 I3079092
00884450636695 I3079093
00884450636695 I3079094
00884450636695 I3079095
00884450636695 I3079096
00884450636695 I3085666
00884450636695 I3085667
00884450636695 I3085668
00884450636695 I3097690
00884450636695 I3097691
00884450636695 I3098952
00884450636695 I3098953
00884450636695 I3098955
00884450636695 I3118226
00884450636695 I3118227
00884450636695 I3118228
00884450636695 I3118230
00884450636695 I3129660
00884450636695 I3129661
00884450636695 I3129662
00884450636695 I3129663
00884450636695 I3129664
00884450636695 I3146015
00884450636695 I3146016
00884450636695 I3150024
00884450636695 I3150025
00884450636695 I3150028
00884450636695 I3150029
00884450636695 I3161167
00884450636695 I3161168
00884450636695 I3161169
00884450636695 I3161170
00884450636695 I3186043
00884450636695 I3186044
00884450636695 I3186047
00884450636695 I3186048
00884450636695 I3186064
00884450636695 I3186065
00884450636695 I3204095
00884450636695 I3204096
00884450636695 I3204119
00884450636695 I3204120
00884450636695 I3204124
00884450636695 I3213949
00884450636695 I3232111
00884450636695 I3232118
00884450636695 I3232119
00884450636695 I3237621
00884450636695 I3237646
00884450636695 I3251181
00884450636695 I3251182
00884450636695 I3251183
00884450636695 I3251184
00884450636695 I3251185
00884450636695 I3251186
IN4130/H 00884450636695 I2623447
00884450636695 I2639668
00884450636695 I2657810
00884450636695 I2663024
00884450636695 I2976522
00884450636695 I2989138
00884450636695 I3014317
00884450636695 I3014321
00884450636695 I3023697
00884450636695 I3098957
00884450636695 I3118231
00884450636695 I3161171
00884450636695 I3186059
00884450636695 I3204098
00884450636695 I3213970
00884450636695 I3213991
00884450636695 I3282764
IN4130/JPH 00884450636695 I2563341
00884450636695 I2584111
00884450636695 I2593621
00884450636695 I2600945
00884450636695 I2617761
00884450636695 I2622049
00884450636695 I2648662
00884450636695 I2648663
00884450636695 I2812806
00884450636695 I2813127
00884450636695 I2819767
00884450636695 I2868572
00884450636695 I2887840
00884450636695 I2892034
00884450636695 I2917168
00884450636695 I2921906
00884450636695 I2927912
00884450636695 I2976523
00884450636695 I3001132
00884450636695 I3014322
00884450636695 I3023700
00884450636695 I3037358
00884450636695 I3074757
00884450636695 I3085670
00884450636695 I3098958
00884450636695 I3110563
00884450636695 I3146019
00884450636695 I3200761
00884450636695 I3213973
00884450636695 I3232041
00884450636695 I3262820
00884450636695 I3282766
IN4130/K 00884450637524 I2719948
00884450637524 I2559561
00884450637524 I2572851
00884450637524 I2572852
00884450637524 I2617762
00884450637524 I2649063
00884450637524 I2649065
00884450637524 I2707737
00884450637524 I2714726
00884450637524 I2719962
00884450637524 I2748232
00884450637524 I2748233
00884450637524 I2748234
00884450637524 I2755246
00884450637524 I2759867
00884450637524 I2759868
00884450637524 I2759871
00884450637524 I2760566
00884450637524 I2770156
00884450637524 I2773967
00884450637524 I2773969
00884450637524 I2773970
00884450637524 I2784607
00884450637524 I2784608
00884450637524 I2853617
00884450637524 I2854185
00884450637524 I2858648
00884450637524 I2868573
00884450637524 I2868574
00884450637524 I2887838
00884450637524 I2887839
00884450637524 I2892035
00884450637524 I2903541
00884450637524 I2903542
00884450637524 I2917171
00884450637524 I2917172
00884450637524 I2918172S1
00884450637524 I2921907
00884450637524 I2921908
00884450637524 I2927913
00884450637524 I2933518
00884450637524 I2937696
00884450637524 I2937697
00884450637524 I2976524
00884450637524 I2988260S1
00884450637524 I2989139
00884450637524 I2989140
00884450637524 I2989141
00884450637524 I3001133
00884450637524 I3014325
00884450637524 I3014326
00884450637524 I3014327
00884450637524 I3014328
00884450637524 I3031136
00884450637524 I3031137
00884450637524 I3047976
00884450637524 I3048199
00884450637524 I3059081
00884450637524 I3074758
00884450637524 I3074759
00884450637524 I3084132
00884450637524 I3085671
00884450637524 I3097692
00884450637524 I3098960
00884450637524 I3118232
00884450637524 I3129665
00884450637524 I3129666
00884450637524 I3129667
00884450637524 I3150026
00884450637524 I3161172
00884450637524 I3161173
00884450637524 I3161174
00884450637524 I3182211
00884450637524 I3199105
00884450637524 I3199127
00884450637524 I3199131
00884450637524 I3204113
00884450637524 I3204114
00884450637524 I3204117
00884450637524 I3204121
00884450637524 I3204122
00884450637524 I3204123
00884450637524 I3232112
00884450637524 I3248089
00884450637524 I3262821
00884450637524 I3290570
00884450637524 I3290571
00884450637524 I3310655
00884450637524 I3310659
00884450637524 I3359468
IN4130/KRK 00884450751121 I2917719
00884450751121 I2933519
00884450751121 I2976526
00884450751121 I3014329
00884450751121 I3056436
00884450751121 I3074760
00884450751121 I3119323
00884450751121 I3232075
IN4130/L 00884450638101 I2593642
00884450638101 I2593643
00884450638101 I2593648
00884450638101 I2593649
00884450638101 I2628991
00884450638101 I2628993
00884450638101 I2639673
00884450638101 I2648667
00884450638101 I2657830
00884450638101 I2657831
00884450638101 I2669897
00884450638101 I2693439
00884450638101 I2693440
00884450638101 I2693441
00884450638101 I2703009
00884450638101 I2714727
00884450638101 I2755249
00884450638101 I2759872
00884450638101 I2759873
00884450638101 I2784624
00884450638101 I2812807
00884450638101 I2834503
00884450638101 I2839309
00884450638101 I2844308
00884450638101 I2853620
00884450638101 I2858643
00884450638101 I2868575
00884450638101 I2887843
00884450638101 I2887844
00884450638101 I2903545
00884450638101 I2921909
00884450638101 I2921910
00884450638101 I2927914
00884450638101 I2933520
00884450638101 I2933521
00884450638101 I2976527
00884450638101 I2989152
00884450638101 I2989153
00884450638101 I3001134
00884450638101 I3001135
00884450638101 I3014332
00884450638101 I3014333
00884450638101 I3014334
00884450638101 I3023701
00884450638101 I3023702
00884450638101 I3031138
00884450638101 I3048389
00884450638101 I3056437
00884450638101 I3056438
00884450638101 I3074761
00884450638101 I3074762
00884450638101 I3079097
00884450638101 I3079098
00884450638101 I3085672
00884450638101 I3098961
00884450638101 I3098962
00884450638101 I3098963
00884450638101 I3129669
00884450638101 I3200762
IN4130/T 00884450860953 I3134851
00884450860953 I3161176
00884450860953 I3161177
00884450860953 I3186046
00884450860953 I3186049
00884450860953 I3204102
00884450860953 I3204110
00884450860953 I3213985
00884450860953 I3232042
00884450860953 I3248073
00884450860953 I3282771
00884450860953 I3282772
00884450860953 I3282773
00884450860953 I3290574
IN4230/JPH 00884450658963 I2629020
00884450658963 I2657814
IN4230/K 00884450638316 I2567486
00884450638316 I2593651
00884450638316 I2607305
00884450638316 I2617770
00884450638316 I2617771
00884450638316 I2639677
00884450638316 I2639678
00884450638316 I2648669
00884450638316 I2648670
00884450638316 I2648672
00884450638316 I2657809
00884450638316 I2657822
00884450638316 I2663022
00884450638316 I2669893
00884450638316 I2678047
00884450638316 I2684456
00884450638316 I2689145
00884450638316 I2714730
00884450638316 I2773990
00884450638316 I2784605
IN4330/CNK 00884450675120 I2701204
00884450675120 I2752079
00884450675120 I2759877
00884450675120 I2989154
00884450675120 I3014337
00884450675120 I3014338
00884450675120 I3023703
00884450675120 I3031139
00884450675120 I3048391
00884450675120 I3186062
00884450675120 I3200763
00884450675120 I3204109
00884450675120 I3213942
IN4330/K 00884450638323 I2535054
00884450638323 I2578105
00884450638323 I2593652
00884450638323 I2607445
00884450638323 I2617772
00884450638323 I2617773
00884450638323 I2617774
00884450638323 I2622051
00884450638323 I2622052
00884450638323 I2639680
00884450638323 I2794464
00884450638323 I2815171
00884450638323 I2827040
00884450638323 I2844309
00884450638323 I2866343
00884450638323 I2868576
00884450638323 I2868577
00884450638323 I2887845
00884450638323 I2887846
00884450638323 I2917174
00884450638323 I2921911
00884450638323 I2921912
00884450638323 I2933522
00884450638323 I2937698
00884450638323 I2937699
00884450638323 I2976528
00884450638323 I2989155
00884450638323 I2989156
00884450638323 I3014341
00884450638323 I3023704
00884450638323 I3023705
00884450638323 I3031141
00884450638323 I3031142
00884450638323 I3085673
00884450638323 I3098964
00884450638323 I3098965
00884450638323 I3098966
00884450638323 I3118234
00884450638323 I3118235
00884450638323 I3146023
00884450638323 I3161189
00884450638323 I3186045
00884450638323 I3186063
00884450638323 I3186066
00884450638323 I3213982
00884450638323 I3232104
00884450638323 I3232114
00884450638323 I3245232
00884450638323 I3251192
00884450638323 I3262822
00884450638323 I3262823
00884450638323 I3280486
IN4352/K 00884450638347 I2519350
00884450638347 I2578108
00884450638347 I2584168
00884450638347 I2584169
00884450638347 I2593666
00884450638347 I2607447
00884450638347 I2617775
00884450638347 I2622053
00884450638347 I2622054
00884450638347 I2639669
00884450638347 I2639670
00884450638347 I2648675
00884450638347 I2657807
00884450638347 I2657808
00884450638347 I2663020
00884450638347 I2678046
00884450638347 I2684457
00884450638347 I2839305
00884450638347 I2853623
00884450638347 I2865636
00884450638347 I2868578
00884450638347 I2868579
00884450638347 I2868580
00884450638347 I2892036
00884450638347 I2903547
00884450638347 I2903549
00884450638347 I2917176
00884450638347 I2937701
00884450638347 I2937702
00884450638347 I2976530
00884450638347 I2976531
00884450638347 I3001136
00884450638347 I3007188
00884450638347 I3014346
00884450638347 I3014347
00884450638347 I3014348
00884450638347 I3023706
00884450638347 I3023707
00884450638347 I3085674
00884450638347 I3110564
00884450638347 I3118236
00884450638347 I3129671
00884450638347 I3129673
00884450638347 I3146025
00884450638347 I3161198
00884450638347 I3161199
00884450638347 I3172026
00884450638347 I3186056
00884450638347 I3186057
00884450638347 I3195191
00884450638347 I3232079
IN4430/K 00884450638354 I2584171
00884450638354 I2593668
00884450638354 I2607449
00884450638354 I2607450
00884450638354 I2617777
00884450638354 I2629013
00884450638354 I2648680
00884450638354 I2657827
00884450638354 I2684458
00884450638354 I2773991
00884450638354 I2808576
00884450638354 I2830967
00884450638354 I2839286
00884450638354 I2844324
00884450638354 I2868581
00884450638354 I2892037
00884450638354 I2917177
00884450638354 I2921913
00884450638354 I2921914
00884450638354 I2937703
00884450638354 I2989157
00884450638354 I2989158
00884450638354 I3014350
00884450638354 I3074770
00884450638354 I3074771
00884450638354 I3079100
00884450638354 I3085675
00884450638354 I3110565
00884450638354 I3161204
00884450638354 I3161205
00884450638354 I3204099
00884450638354 I3213943
00884450638354 I3245237
00884450638354 I3262824
00884450638354 I3262825
00884450638354 I3310663
IN4530/CNK 00884450638361 I2811612
00884450638361 I2976582
00884450638361 I2989159
00884450638361 I3001137
00884450638361 I3001138
00884450638361 I3014351
00884450638361 I3023708
00884450638361 I3048392
00884450638361 I3056440
00884450638361 I3110566
00884450638361 I3118237
00884450638361 I3146027
00884450638361 I3204097
00884450638361 I3213947
00884450638361 I3245235
IN4530/K 00884450638378 I2563342
00884450638378 I2564657
00884450638378 I2579800
00884450638378 I2584172
00884450638378 I2593670
00884450638378 I2593671
00884450638378 I2607461
00884450638378 I2617778
00884450638378 I2618399S1
00884450638378 I2622055
00884450638378 I2648681
00884450638378 I2648682
00884450638378 I2657799
00884450638378 I2663018
00884450638378 I2669891
00884450638378 I2678045
00884450638378 I2684459
00884450638378 I2693445
00884450638378 I2693477
00884450638378 I2714734
00884450638378 I2759878
00884450638378 I2794461
00884450638378 I2812803
00884450638378 I2812804
00884450638378 I2812805
00884450638378 I2819769
00884450638378 I2819770
00884450638378 I2827041
00884450638378 I2834500
00884450638378 I2839288
00884450638378 I2839295
00884450638378 I2839297
00884450638378 I2868589
00884450638378 I2868593
00884450638378 I2887847
00884450638378 I2887849
00884450638378 I2892038
00884450638378 I2903553
00884450638378 I2903554
00884450638378 I2921915
00884450638378 I2921916
00884450638378 I2932642
00884450638378 I2933523
00884450638378 I2937704
00884450638378 I2976557
00884450638378 I2976558
00884450638378 I2989160
00884450638378 I2989161
00884450638378 I3001139
00884450638378 I3001140
00884450638378 I3014353
00884450638378 I3014354
00884450638378 I3023709
00884450638378 I3023710
00884450638378 I3031143
00884450638378 I3031144
00884450638378 I3037362
00884450638378 I3048393
00884450638378 I3056441
00884450638378 I3065899
00884450638378 I3074773
00884450638378 I3074774
00884450638378 I3074775
00884450638378 I3079101
00884450638378 I3079102
00884450638378 I3085676
00884450638378 I3085678
00884450638378 I3110115
00884450638378 I3118238
00884450638378 I3118239
00884450638378 I3129675
00884450638378 I3161424
00884450638378 I3198556
00884450638378 I3213987
00884450638378 I3232117
00884450638378 I3262828
00884450638378 I3262829
00884450638378 I3298678
00884450638378 I3310665
00884450638378 I3333085
IN4530/T 00884450860106 I3135192
00884450860106 I3162534
00884450860106 I3162535
00884450860106 I3186050
00884450860106 I3186055
00884450860106 I3204103
00884450860106 I3232088
00884450860106 I3245228
00884450860106 I3248056
00884450860106 I3250060
00884450860106 I3282780
00884450860106 I3290580
00884450860106 I3310666
IN4802/K 00884450638385 I2563343
00884450638385 I2593673
00884450638385 I2593675
00884450638385 I2596323
00884450638385 I2596324
00884450638385 I2596325
00884450638385 I2617779
00884450638385 I2629004
00884450638385 I2657832
00884450638385 I2669888
00884450638385 I2678044
00884450638385 I2684460
00884450638385 I2689143
00884450638385 I2693447
00884450638385 I2693448
00884450638385 I2720068
00884450638385 I2738932
00884450638385 I2738933
00884450638385 I2747737
00884450638385 I2794459
00884450638385 I2794460
00884450638385 I2827043
00884450638385 I2834501
00884450638385 I2839289
00884450638385 I2839293
00884450638385 I2873736
00884450638385 I2887850
00884450638385 I2887851
00884450638385 I2892039
00884450638385 I2921917
00884450638385 I2921918
00884450638385 I2927915
00884450638385 I2927916
00884450638385 I2933524
00884450638385 I2937705
00884450638385 I2976560
00884450638385 I2989162
00884450638385 I3001141
00884450638385 I3014357
00884450638385 I3014358
00884450638385 I3014359
00884450638385 I3031145
00884450638385 I3031146
00884450638385 I3037363
00884450638385 I3079103
00884450638385 I3085679
00884450638385 I3098968
00884450638385 I3098970
00884450638385 I3098971
00884450638385 I3110567
00884450638385 I3118240
00884450638385 I3129676
00884450638385 I3138883
00884450638385 I3150031
00884450638385 I3161209
00884450638385 I3161210
00884450638385 I3161211
00884450638385 I3172451
00884450638385 I3234856
00884450638385 I3251194
00884450638385 I3275873
00884450638385 I3275874
00884450638385 I3275889
00884450638385 I3277190
00884450638385 I3290585
00884450638385 I3298679
00884450638385 I3298680
00884450638385 I3310667
00884450638385 I3310668
00884450638385 I3333086
IN4802/T 00884450860113 I3135188
00884450860113 I3170171
IN4852/K 00884450638392 I2794462
00884450638392 I2887852
00884450638392 I2917180
00884450638392 I2933525
00884450638392 I2976562
00884450638392 I3001143
00884450638392 I3023717
00884450638392 I3031148
00884450638392 I3048394
00884450638392 I3056442
00884450638392 I3097696
00884450638392 I3097697
00884450638392 I3098972
00884450638392 I3110570
00884450638392 I3130278
00884450638392 I3161212
00884450638392 I3161213
00884450638392 I3204108
00884450638392 I3262831
00884450638392 I3310671
IN4901/A 00884450638408 I2564660
00884450638408 I2607467
00884450638408 I2617781
00884450638408 I2629015
00884450638408 I2648684
00884450638408 I2648685
00884450638408 I2663014
00884450638408 I2684461
00884450638408 I2693456
00884450638408 I2714736
00884450638408 I2719978
00884450638408 I2747555
00884450638408 I2822684
00884450638408 I2844330
00884450638408 I2868607
00884450638408 I2887841
00884450638408 I2887842
00884450638408 I2917181
00884450638408 I2921919
00884450638408 I2976563
00884450638408 I2989168
00884450638408 I3014360
00884450638408 I3037365
00884450638408 I3048395
00884450638408 I3056443
00884450638408 I3079104
00884450638408 I3097698
00884450638408 I3098974
00884450638408 I3118241
00884450638408 I3141085
00884450638408 I3146032
00884450638408 I3186061
00884450638408 I3200766
00884450638408 I3206777
00884450638408 I3213971
00884450638408 I3232096
00884450638408 I3242604
00884450638408 I3251195
00884450638408 I3279228
00884450638408 I3290595
00884450638408 I3298681
00884450638408 I3333091
IN4902/A 00884450638415 I2607468
00884450638415 I2684465
00884450638415 I2738937
00884450638415 I2844331
00884450638415 I2868608
00884450638415 I2976564
00884450638415 I2989174
00884450638415 I3098975
00884450638415 I3118242
00884450638415 I3146037
00884450638415 I3247347
00884450638415 I3262832
IN4903/A 00884450638422 I2648686
00884450638422 I2663012
00884450638422 I2669883
00884450638422 I2674448
00884450638422 I2839311
00884450638422 I2887853
00884450638422 I2927917
00884450638422 I2937706
00884450638422 I2976565
00884450638422 I3014361
00884450638422 I3023711
00884450638422 I3031156
00884450638422 I3074776
00884450638422 I3079112
00884450638422 I3098976
00884450638422 I3184893
00884450638422 I3199765
00884450638422 I3200769
00884450638422 I3213959
00884450638422 I3234851
00884450638422 I3298682
00884450638422 I3310672
00884450638422 I3333092
IN4904/A 00884450638439 I2794465
00884450638439 I2827044
00884450638439 I2903555
00884450638439 I2917182
00884450638439 I2921920
00884450638439 I2976566
00884450638439 I3001144
00884450638439 I3014363
00884450638439 I3023712
00884450638439 I3074778
00884450638439 I3079113
00884450638439 I3098978
00884450638439 I3118245
00884450638439 I3146044
00884450638439 I3186053
00884450638439 I3195193
00884450638439 I3200770
00884450638439 I3232085
00884450638439 I3282785
00884450638439 I3298683
IN4905/A 00884450638446 I2622058
00884450638446 I2639679
00884450638446 I2648688
00884450638446 I2693458
00884450638446 I2719968
00884450638446 I2738935
00884450638446 I2755250
00884450638446 I2755251
00884450638446 I2770154
00884450638446 I2834504
00884450638446 I2853629
00884450638446 I2868611
00884450638446 I2903558
00884450638446 I2915017
00884450638446 I2933526
00884450638446 I2937707
00884450638446 I2989176
00884450638446 I3001145
00884450638446 I3031157
00884450638446 I3056444
00884450638446 I3079114
00884450638446 I3085682
00884450638446 I3129679
00884450638446 I3146045
00884450638446 I3161252
00884450638446 I3185199
00884450638446 I3204106
00884450638446 I3245241
00884450638446 I3262834
00884450638446 I3333094
IN4906/A 00884450638453 I2607469
00884450638453 I2617783
00884450638453 I2639671
00884450638453 I2648689
00884450638453 I2657813
00884450638453 I2669871
00884450638453 I2684462
00884450638453 I2693459
00884450638453 I2808575
00884450638453 I2834502
00884450638453 I2839306
00884450638453 I2903559
00884450638453 I2933527
00884450638453 I2989177
00884450638453 I3014364
00884450638453 I3023714
00884450638453 I3056449
00884450638453 I3065880
00884450638453 I3085683
00884450638453 I3097700
00884450638453 I3146046
00884450638453 I3204104
00884450638453 I3213967
00884450638453 I3282786
IN4907/A 00884450638460 I2684550
00884450638460 I2707738
00884450638460 I2707741
00884450638460 I2747546
00884450638460 I2770153
00884450638460 I2773980
00884450638460 I2812829
00884450638460 I2819771
00884450638460 I2839313
00884450638460 I2868612
00884450638460 I2887855
00884450638460 I2903561
00884450638460 I2917184
00884450638460 I2933528
00884450638460 I2976568
00884450638460 I2976569
00884450638460 I3001146
00884450638460 I3014365
00884450638460 I3014366
00884450638460 I3065280
00884450638460 I3074779
00884450638460 I3098979
00884450638460 I3129680
00884450638460 I3146047
00884450638460 I3213979
00884450638460 I3248082
00884450638460 I3282787
IN4908 00884450544570 I2623336
00884450544570 I2654504
00884450544570 I2684552
00884450544570 I2794463
00884450544570 I2839307
00884450544570 I2892042
00884450544570 I2903564
00884450544570 I2988261S1
00884450544570 I2989180
00884450544570 I3043208
00884450544570 I3056451
00884450544570 I3098980
00884450544570 I3129681
00884450544570 I3204100
00884450544570 I3232095
00884450544570 I3262837
IN4909 00884450544587 I2639691
00884450544587 I2684549
00884450544587 I2693461
00884450544587 I2853630
00884450544587 I2858652
00884450544587 I2891843
00884450544587 I2915034
00884450544587 I2921921
00884450544587 I2927918
00884450544587 I2989178
00884450544587 I3014368
00884450544587 I3023715
00884450544587 I3031158
00884450544587 I3056452
00884450544587 I3098981
00884450544587 I3110572
00884450544587 I3146048
00884450544587 I3161258
00884450544587 I3204105
00884450544587 I3226196
00884450544587 I3232116
00884450544587 I3248083
00884450544587 I3282789
00884450544587 I3298686
IN4910 00884450544594 I3152059
00884450544594 I3333097
IN4911/A 00884450638477 I2629018
00884450638477 I2678041
00884450638477 I2693462
00884450638477 I2719972
00884450638477 I2738936
00884450638477 I2773993
00884450638477 I2839312
00884450638477 I2868614
00884450638477 I2903566
00884450638477 I2903567
00884450638477 I2917187
00884450638477 I2921923
00884450638477 I2927919
00884450638477 I2927921
00884450638477 I2933529
00884450638477 I2976570
00884450638477 I2989184
00884450638477 I3014369
00884450638477 I3014370
00884450638477 I3023716
00884450638477 I3031159
00884450638477 I3031160
00884450638477 I3053785
00884450638477 I3056454
00884450638477 I3057825
00884450638477 I3097703
00884450638477 I3118246
00884450638477 I3129682
00884450638477 I3186060
00884450638477 I3204112
00884450638477 I3213984
00884450638477 I3248093
00884450638477 I3262841
00884450638477 I3290597
00884450638477 I3298687
00884450638477 I3333099
IN4912 00884450544617 I2593678
00884450544617 I2617785
00884450544617 I2639688
00884450544617 I2812812
00884450544617 I2892043
00884450544617 I2921924
00884450544617 I2927922
00884450544617 I2976571
00884450544617 I3079115
00884450544617 I3110574
00884450544617 I3129684
00884450544617 I3200774
00884450544617 I3213980
00884450544617 I3251196
00884450544617 I3282790
00884450544617 I3298688
IN4913/A 00884450638484 I2684555
00884450638484 I2693463
00884450638484 I2759884
00884450638484 I2834505
00884450638484 I2853632
00884450638484 I2868615
00884450638484 I2933532
00884450638484 I2976572
00884450638484 I3031161
00884450638484 I3048396
00884450638484 I3074780
00884450638484 I3119310
00884450638484 I3200775
00884450638484 I3213972
00884450638484 I3232073
IN4914/A 00884450638491 I2684547
00884450638491 I2707743
00884450638491 I2827045
00884450638491 I2976573
00884450638491 I3014371
00884450638491 I3074782
00884450638491 I3079116
00884450638491 I3146049
00884450638491 I3204116
00884450638491 I3298689
IN4915 00884450544648 I2607298
00884450544648 I2617786
00884450544648 I2629010
00884450544648 I2648691
00884450544648 I2657803
00884450544648 I2684554
00884450544648 I2693465
00884450544648 I2738938
00884450544648 I2827046
00884450544648 I2839316
00884450544648 I2858644
00884450544648 I2885752
00884450544648 I2892044
00884450544648 I2917188
00884450544648 I2921925
00884450544648 I2927923
00884450544648 I2976574
00884450544648 I3001147
00884450544648 I3085684
00884450544648 I3097705
00884450544648 I3110575
00884450544648 I3129222
00884450544648 I3161288
00884450544648 I3204111
00884450544648 I3226333
00884450544648 I3232109
00884450544648 I3262842
IN4916/A 00884450544648 I2629021
00884450544648 I2639674
00884450544648 I2892045
00884450544648 I2921926
00884450544648 I2976575
00884450544648 I3022141
00884450544648 I3119234
00884450544648 I3130590
00884450544648 I3204101
00884450544648 I3251197
00884450544648 I3279574
IN4917/A 00884450638514 I2559659
00884450638514 I2578110
00884450638514 I2593680
00884450638514 I2648692
00884450638514 I2648693
00884450638514 I2678040
00884450638514 I2903572
00884450638514 I2917190
00884450638514 I2921927
00884450638514 I2976576
00884450638514 I3037366
00884450638514 I3056455
00884450638514 I3110576
00884450638514 I3200777
00884450638514 I3204107
00884450638514 I3213956
00884450638514 I3231624
IN4918/A 00884450638521 I2693466
00884450638521 I2868616
00884450638521 I2903573
00884450638521 I2933533
00884450638521 I2976578
00884450638521 I3056456
00884450638521 I3074783
00884450638521 I3098982
00884450638521 I3146050
00884450638521 I3200779
00884450638521 I3275917
00884450638521 I3277193
IN4919/A 00884450638538 I2559662
00884450638538 I2584179
00884450638538 I2669854
00884450638538 I2759886
00884450638538 I2903576
00884450638538 I2915043
00884450638538 I2920389
00884450638538 I2933534
00884450638538 I2976579
00884450638538 I3056457
00884450638538 I3079117
00884450638538 I3085685
00884450638538 I3097706
00884450638538 I3275918
00884450638538 I3310676
|
Recalling Firm/ Manufacturer |
Merit Medical Systems, Inc. 1600 W Merit Pkwy South Jordan UT 84095-2416
|
| For Additional Information Contact | Merit Customer Service 1800-356-3748 |
Manufacturer Reason for Recall | Inflation device handle may detach from the syringe during procedure. |
FDA Determined Cause 2 | Process control |
| Action | On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions:
1. Please read and understand the recall notification.
2. Should the handle detach during use and the balloon not completely deflate, pull the
handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast
3. Ensure that applicable personnel within your organization are made aware of this field action.
3. If the product has been further distributed to other facilities, institutions, or manufacturers, please
ensure this notice is immediately shared with them.
4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within seven (7) calendar days |
| Quantity in Commerce | 4,163,123 units |
| Distribution | Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Korea, Republic of Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Madagascar, Mexico, Monaco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Trinidad and Tobago, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian, Republic of Viet Nam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|