| Date Initiated by Firm | December 22, 2025 |
| Date Posted | February 06, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1325-2026 |
| Recall Event ID |
98277 |
| 510(K)Number | K234064 |
| Product Classification |
Reprocessed intravascular ultrasound catheter - Product Code OWQ
|
| Product | Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters:
ACUSON AcuNav Ultrasound Catheter, Use on GE Systems
10135910RH; 10043342RH
ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems
10135936RH, 08255790RH |
| Code Information |
10135910RH
UDI-DI 10197344026500
Lots
EP250324
EP250429
EP250521
EP250609
EP250630
EP250710
EP250905;
10135936RH
UDI-DI 10197344026517
Lots
EP250324
EP250429
EP250521
EP250609
EP250630
EP250710
EP250905
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barlay 866-359-1704 |
Manufacturer Reason for Recall | These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025 |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. |
| Quantity in Commerce | 511 |
| Distribution | Distribution US nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWQ
|