• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Erbe ERBEFLO 2 and ERBEFLO CleverCap Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Erbe ERBEFLO 2 and ERBEFLO CleverCap Devicessee related information
Date Initiated by FirmMarch 31, 2026
Date PostedMay 01, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2062-2026
Recall Event ID 98674
510(K)NumberK132340 K162152 
Product Classification Pump, air, non-manual, for endoscope - Product Code FEQ
ProductERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208
Code Information REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249; 20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243; 20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367; 20325-207/04065655000689/W4469939, W4469940, W4469941; 20325-248/04065655001006/W4467050, W4467051, W4467055, W4467056, W4467682, WO470898, WO470899, WO470900, WO470901, WO470902, WO470903, WO471252, WO471253, WO471254, WO471255, WO471256; 20325-208/04065655001099/W4467040, W4468271, WO468760
FEI Number 3020712480
Recalling Firm/
Manufacturer
Erbe Medical, LLC
405 W Geneva Dr
Tempe AZ 85282-2003
Manufacturer Reason
for Recall
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
FDA Determined
Cause 2
Component change control
ActionOn 3/31/2026, correction notices were emailed to Medical Directors, Risk Managers, Medical Device Safety Officers, or Healthcare Professionals informing them to take the following actions: 1) Discontinue use of unsupported third-party Scope Port Connectors. 2) Provide this notice to all personnel within your organization who use, service, purchase, stock, or otherwise manage. 3) If you further distributed any affected devices, promptly forward this notice to those accounts and request that they take the same actions. 4) Continue following the instructions for use, including verifying connector compatibility, ensuring that all connectors are fully seated, confirming that all joints and interfaces are secure, and verifying correct placement and alignment of the tubing within the pump head. 5) Utilize the compatibility table in the notice until updated Notes on Use is received. 6) Complete and return the response form via email to fieldaction.medical.us@erbegroup.com. Firm contacts: Correction: fieldaction.medical.us@erbegroup.com, 480-564-1899. Routine Customer Service and Complaints: techservices@erbegroup.com, 1-800-778-3723
Quantity in Commerce260,962
DistributionUS Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = FEQ
-
-