| | Class 1 Device Recall Arrow Kits and Sets that contain Lidocaine and Bupivacaine |  |
| Date Initiated by Firm | April 21, 2026 |
| Date Posted | June 13, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2377-2026 |
| Recall Event ID |
98793 |
| 510(K)Number | K171146 K810962 K820009 K862056 K021120 |
| Product Classification |
Catheter, percutaneous - Product Code DQY
|
| Product | Description/REF:
ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6
RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP
ACCESS TRAY/ASK-04001-MS
RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1
VESSEL CATH KIT: 20GA X 5IN/AK-04550-S
CVC KIT: 1-LUMEN 5 FR X 20 CM/AK-04301-SP
CVC KIT: 16 GA X 6IN (16 CM)/AK-04306-SP
CVC KIT: 16GA X 6 IN/AK-04800-S
CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-12402
CVC KIT: 3-LUMEN 7 FR X 16 CM/AK-12703-CDC
CVC KIT: 2L 4 FR X 13 CM/AK-14402
CVC KIT: 2-LUMEN 5 FR X 13 CM/AK-14502
CVC KIT: 3-LUMEN 5.5 FR X 30 CM/AK-14553
CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-14703-SP
CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-15402
CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-15553
CVC KIT: 2-LUMEN 4 FR X 30 CM/AK-16402
CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-16553
CVC KIT: 5 FR X 20 CM/ASK-04301-HH2
CVC KIT 3L: 7 FR X 6 IN (16 CM)/ASK-12703-MGH3 |
| Code Information |
REF(Material)/UDI-DI/Lot(Batch):
ASK-04510-MM6/10801902193275, 10801902221909/33F25E0590, 33F25G0196, 33F25M0005, 33F25M0149, 33F26A0302, 33F26C0723, 33F24K0212;
NA-04020-X1ASP/10801902193275/33F25J0575, 33F25J0756, 33F26B0065, 33F26B0027;
ASK-04001-MS/10801902193275/33F24K0706, 33F25F0375, 33F25F0005;
ASK-00820-UM1/10801902193275/33F24G0150;
AK-04550-S/10801902196719, 10801902193275, 60801902121709/33F24A0029, 33F24A0355, 33F24A1014, 33F24A1619, 33F24B0112, 33F24C0548, 33F24D0017, 33F24E0174, 33F24G0328, 33F24G0576, 33F24J0276, 33F24J0494, 33F24K0092, 33F24L0720, 33F25A0458, 33F25C0950 33F25E0041, 33F25E0911, 33F25G0080, 33F25H0013, 33F25J0820, 33F25L0402, 33F26B0550, 33F26C0267, 33F23M0175;
AK-04301-SP/10801902121650, 10801902193275, 10801902121674/33F24A1010, 33F24C0093, 33F24D0068, 33F24E0325, 33F24G0325, 33F24J0274, 33F24K0090, 33F24L0809, 33F25C0757, 33F25E0038, 33F25G0760, 33F25H0721, 33F25J0462, 33F25L0239, 33F25M0888, 33F24A0146;
AK-04306-SP/10801902121674, 10801902193275, 10801902192230/33F24A0846, 33F24B0174, 33F24C0094, 33F24E0327, 33F24E0329, 33F24J0275, 33F24J0492, 33F24L0717, 33F25B0249, 33F25C0783, 33F25E0039, 33F25J0463, 33F25L0054, 33F25L0345, 33F26B0684, 33F23J0175;
AK-04800-S/60801902121709, 10801902193275, 10801902209501/33F24B0437, 33F24C0735, 33F24E0708, 33F24G0120, 33F24G0647, 33F24H0415, 33F24J0448, 33F24L0037, 33F24M0128, 33F25A0396, 33F25B0111, 33F25D0028, 33F25G0509, 33F25H0326, 33F25J0997, 33F26C0274, 33F24C0013;
AK-12402/10801902168471, 10801902193275, 10801902168488/33F24E0342, 33F24F0325, 33F24G0347, 33F24J0512, 33F24K0113, 33F24L0750, 33F25C0109, 33F25C0964, 33F25E0063, 33F25E0926, 33F25J0474, 33F24A0408;
AK-12703-CDC/10801902168488, 10801902193275, 10801902168495/33F24A1020, 33F24B0187, 33F24C0104, 33F24D0078, 33F24G0348, 33F24J0513, 33F24K0114, 33F24L0596, 33F25C0110, 33F25C0759, 33F25J0834, 33F25M0948, 33F24A0401;
AK-14402/10801902168495, 10801902193275, 10801902168495, 10801902168518/33F24E0177, 33F24E0343, 33F24F0328, 33F24G0349, 33F24J0514, 33F24K0115, 33F25B0369, 33F25C0875, 33F25E0064, 33F25E0928, 33F25H0029, 33F25J0475, 33F26B0564, 33F24A0813;
AK-14502/10801902168518, 10801902193275, 30801902122040/33F24E0344, 33F24G0350, 33F24K0116, 33F24L0656, 33F25C0878, 33F25E0065, 33F25G0089, 33F25J0476, 33F26B0568, 33F24A0849;
AK-14553/30801902122040, 10801902193275, 10801902122077/33F24B0190, 33F24C0107, 33F24E0345, 33F24E0346, 33F24F0332, 33F24G0351, 33F24K0117, 33F24L0689, 33F25C0112, 33F25C0848, 33F25G0090, 33F25J0835, 33F24B0191;
AK-14703-SP/10801902122077, 10801902193275, 10801902197136/33F24E0347, 33F24G0352, 33F24J0516, 33F24K0118, 33F24L0737, 33F25A0476, 33F25C0887, 33F25E0067, 33F25E0929, 33F25J0477, 33F25L0399, 33F25M0905, 33F24D0079;
AK-15402/10801902168532, 10801902193275, 10801902168532, 10801902168549/33F24B0194, 33F24E0350, 33F24G0354, 33F24H0034, 33F24J0518, 33F24K0120, 33F24L0751, 33F25A0479, 33F25C0799, 33F25E0069, 33F25E0932, 33F25H0723, 33F25J0836, 33F25L0348, 33F26C0317, 33F24A1268;
AK-15553/10801902168549, 10801902193275, 10801902168549, 10801902168563/33F24E0714, 33F24G0226, 33F24G0687, 33F24H0249, 33F24J0353, 33F25B0332, 33F25D0195, 33F25H0796, 33F25J0081, 33F26B0287, 33F24A1269;
AK-16402/10801902168563, 10801902193275, 10801902168563, 10801902168570/33F24C0738, 33F24E0715, 33F24H0065, 33F24H0416, 33F24J0212, 33F24J0426, 33F25A0046, 33F25C0087, 33F25C0604, 33F25G0458, 33F25J0082, 33F26A1658, 33F26B0069, 33F26C0446, 33F24A0796;
AK-16553/10801902168570, 10801902193275, 30801902122248/33F24E0178, 33F24E0351, 33F24G0355, 33F24J0519, 33F24K0121, 33F24L0790, 33F25C0980, 33F25E0070, 33F25J0478, 33F24C0025;
ASK-04301-HH2/10801902193275, 20801902114987/33F25H0640, 33F25K0070, 33F25K0090, 33F24D0733;
ASK-12703-MGH3/10801902193275, 10801902115260/33F25D0263, 33F24F0838
|
| FEI Number |
3013162291
|
Recalling Firm/ Manufacturer |
ARROW INTERNATIONAL, LLC
3015 Carrington Mill Blvd
Morrisville
NC
27560-5437
|
| For Additional Information Contact | Customer Service 866-396-2111 |
Manufacturer Reason for Recall | Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following:
Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets.
Medical facilities
1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred.
2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product.
3. Complete and return the response form virtually via:
https://fca.myteleflex.com/en/recall/000608
Distributors:
1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor.
2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product.
3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608
4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email.
If you have questions contact firm's Customer Service:
Link: https://fca.myteleflex.com/en/recall/000608,
Customer Service: Telephone: 1-866-396-2111,
FAX: 1-855-419-8507 Email: Recalls@teleflex.com
On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices.
On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets. |
| Quantity in Commerce | 126,530 |
| Distribution | US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DQY 510(K)s with Product Code = DQY
|
|
|
|