| Date Initiated by Firm | May 08, 2026 |
| Date Posted | June 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2359-2026 |
| Recall Event ID |
99036 |
| 510(K)Number | K123309 |
| Product Classification |
Warmer, Infant Radiant - Product Code FMT
|
| Product | Panda Freestanding Warmer, Infant Radiant Warmer
Product Identifiers: 2063568-001
Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance. |
| Code Information |
UDI-DI of service blender component: 00813841020010
Affected devices were installed with the following serial numbers of affected service blender components:
002I08B
002X08B
008S10B
008Y09B
013S09B
016A09B
016R09B
018P10B
019J09B
028C08B
02X08B
033N09B
037E08B
038D10B
038N08B
042F10B
042S10B
042T10B
045808B
045M08B
047I08B
047J08B
054C10B
054I10B
054L10B
054N09B
054O09B
054U10B
055D09B
055V09B
057908B
061K09B
062C10B
068W10B
077M09B
078509B
078M09B
078S09B
084I10B
084J10B
092G08B
099F08B
1 B1405629
100E08B
100R08B
104A09B
12105790
133D08B
136708B
136W08B
141J09B
142009B
142W09B
144J08B
144N08B
149Q09B
163608B
169D08B
169G08B
169H08B
172308B
175P08B
185I08B
185Q08B
186208B
189808B
189A09B
189C08B
189F09B
193L08B
193N08B
195D09B
195E09B
195J09B
195K09B
198H08B
210K09B
210U09B
218N09B
219C09B
219K09B
220E08B
226F08B
244L09B
260X09B
266M08B
267808B
280F09B
281H09B
285J07B
295H09B
295J09B
295P08B
310I09B
338707B
338E07B
338K07B
352Y09B
355P09B
7979700385
B1003093
B1003112
B1003124
B1003140
B1003143
B1003147
B1003153
B1003343
B1003452
B1003550
B1003867
B1003944
B1004844
B1004861
B1004924
B1005029
B1005036
B1005044
B1005422
B1005534
B1005560
B1005571
B1005839
B1006102
B1006113
B1006304
B1006858
B1006935
B1006971
B1007078
B1007220
B1007262
B1007355
B1007360
B1007376
B1007383
B1007528
B1007605
B1101215
B1101224
B1101241
B1101445
B1101446
B1101547
B1101862
B1101869
B1101998
B1102011
B1102015
B1102153
B1102384
B1102405
B1102571
B1103145
B1103159
B1103251
B1103256
B1103422
B1103434
B1103576
B1103710
B1103844
B1103846
B1103850
B1103856
B1103968
B1104440
B1104441
B1104450
B1104651
B1104658
B1104764
B1105179
B1105180
B1105232
B1105244
B1105418
B1105548
B1105806
B1106117
B1106789
B1106804
B1106812
B1106823
B1107074
B1107357
B1107360
B1107581
B1107658
B1201350
B1201361
B1201601
B1201990
B1202187
B1202199
B1202740
B1203408
B1203414
B1203520
B1203557
B1203643
B1203892
B1203905
B1203909
B1203913
B1203917
B1204710
B1205518
B1205537
B1205804
B1205820
B1206201
B1206202
B1206224
B1206632
B1206909
B1206948
B1207178
B1207180
B1207181
B1207189
B1207201
B1207202
B1207215
B1207240
B1207308
B1207331
B1207515
B1207555
B1208045
B1208066
B1208067
B1208068
B1208069
B1208102
B1208104
B1208141
B1208296
B1208702
B1208729
B1208736
B1208739
B1208759
B1208979
B1208989
B1208991
B1209004
B1209089
B1209090
B1209099
B1209175
B1209181
B1209188
B1209197
B1209199
B1209432
B1209463
B1209467
B1209641
B1209642
B1209647
B1209796
B1209805
B1209956
B1211743
B1211784
B1211907
B1211913
B1212670
B1212683
B1212706
B1212817
B1300102
B1300110
B1300126
B1300416
B1300430
B1300445
B1300805
B1300812
B1300830
B1300834
B1300858
B1300919
B1302107
B1302132
B1302133
B1302180
B1302527
B1302530
B1302583
B1302593
B1302945
B1302982
B1302983
B1303084
B1303098
B1303452
B1303453
B1303454
B1303492
B1303581
B1303854
B1303868
B1303910
B1304069
B1304357
B1304374
B1304376
B1304450
B1304518
B1304564
B1304570
B1304574
B1304751
B1305163
B1305175
B1305213
B1305215
B1305337
B1305363
B1305523
B1306031
B1306066
B1306068
B1306078
B1306902
B1307502
B1307507
B1307514
B1307526
B1307553
B1307814
B1307820
B1307878
B1308177
B1308332
B1308340
B1308362
B1308367
B1308874
B1309123
B1309124
B1309135
B1309139
B1309150
B1309158
B1309683
B1309705
B1309706
B1309962
B1310226
B1310237
B1400022
B1400237
B1400243
B1400264
B1400278
B1400292
B1400688
B1400713
B1400760
B1400771
B1400946
B1401337
B1401342
B1401343
B1401757
B1401785
B1401893
B1402142
B1402162
B1402168
B1402196
B1402198
B1402224
B1402504
B1402565
B1402854
B1403053
B1403079
B1403128
B1403140
B1403155
B1403196
B1403216
B1403230
B1403366
B1403732
B1403769
B1403778
B1403880
B1403893
B1403911
B1403914
B1403934
B1403952
B1404215
B1404267
B1404291
B1404430
B1404469
B1404473
B1404480
B1404481
B1404495
B1404723
B1404738
B1404754
B1404760
B1404786
B1404925
B1404981
B1405020
B1405134
B1405213
B1405218
B1405542
B1405544
B1405553
B1405558
B1405574
B1405576
B1405577
B1405589
B1405601
B1405608
B1405621
B1405622
B1405629
B1405630
B1405649
B1405688
B1405909
B1405923
B1405998
B1406104
B1406186
B1406190
B1406191
B1406196
B1406197
B1406432
B1406441
B1406479
B1406482
B1406506
B1406507
B1406518
B1406518-2
B1406528
B1406537
B1406599
B1407065
B1407092
B1407755
B1407776
B1407777
B1407786
B1408337
B1408379
B1408525
B1408533
B1409022
B1409048
B1409051
B1409052
B1409053
B1409324
B1409328
B1409329
B1409366
B1409509
B1409533
B1409546
B1409569
B1409588
B1409590
B1409651
B1409722
B1409731
B1409759
B1501314
B1501349
B1501652
B1501657
B1501675
B1501686
B1501687
B1501693
B1502040
B1502054
B1502157
B1502195
B1502200
B1502616
B1502655
B1502656
B1502659
B1502668
B1502675
B1502822
B1502887
B1502888
B1502892
B1502894
B1502899
B1503153
B1503178
B1503210
B1503224
B1503229
B1503240
B1503274
B1503472
B1503476
B1503500
B1503608
B1503611
B1503644
B1503665
B1503667
B1503673
B1503687
B1503698
B1504073
B1504076
B1504078
B1504339
B1504538
B1504754
B1504755
B1504761
B1504765
B1504800
B1504919
B1504951
B1504999
B1505000
B1505085
B1505090
B1505091
B1505092
B1505093
B1505096
B1505276
B1505301
B1505304
B1505527
B1505529
B1505536
B1505558
B1505570
B1505573
B1505575
B1505576
B1505624
B1505636
B1505637
B1505638
B1505655
B1505656
B1505825
B1505839
B1505840
B1505841
B1505842
B1505843
B1505854
B1505879
B1505884
B1505884 (2)
B1505897
B1505899
B1505954
B1505981
B1505990
B1505992
B1505994
B1505997
B1506198
B1506203
B1506208
B1506209
B1506209-2
B1506215
B1506216
B1506220
B1506237
B1506240
B1801608
B2001326
B2260943
BC2260311
BX1101219
BX1103422
BX1206948
BX1209647
BX1302527
BX1409022
BX1501179
BX1501323
BX1501350
BX1501429
BX1501514
BX1501668
BX1501686
BX1501827
BX1501849
BX1501857
BX1501865
BX1501894
BX1501895
BX1501897
BX1501899
BX1501900
BX1501903
BX1501911
BX1502336
BX1502340
BX1502392
BX1502444
BX1502705
BX1502719
BX1502739
BX1502748
BX1502751
BX1502753
BX1502759
BX1502764
BX1502766
BX1505884
BX1601010
BX1601032
BX1601044
BX1601050
BX1601062
BX1601063
BX1601066
BX1601268
BX1601277
BX1601286
BX1601307
BX1601311
BX1601322
BX1601392
BX1601394
BX1601397
BX1601432
BX1601467
BX1601516
BX1601518
BX1601519
BX1601521
BX1601523
BX1601524
BX1601525
BX1601526
BX1601527
BX1601528
BX1601537
BX1601538
BX1601541
BX1601556
BX1601797
BX1601861
BX1601881
BX1601941
BX1601946
BX1602072
BX1602083
BX1602195
BX1602315
BX1602418
BX1602465
BX1602497
BX1602527
BX1602586
BX1602614
BX1602621
BX1602650
BX1602653
BX1602746
BX1602764
BX1602779
BX1602818
BX1602823
BX1602825
BX1602830
BX1602839
BX1602850
BX1602856
BX1602857
BX1602860
BX1602862
BX1602864
BX1602887
BX1602908
BX1602910
BX1602915
BX1602925
BX1602932
BX1602934
BX1602935
BX1602936
BX1602950
BX1602969
BX1602986
BX1602989
BX1603007
BX1603014
BX1603033
BX1603090
BX1603111
BX1603115
BX1603269
BX1603279
BX1603285
BX1603356
BX1603377
BX1603553
BX1603594
BX1603595
BX1603671
BX1603677
BX1603680
BX1603687
BX1603689
BX1603694
BX1604080
BX1604340
BX1604353
BX1604421
BX1604532
BX1604659
BX1604754
BX1604819
BX1605003
BX1605014
BX1605016
BX1605119
BX1605122
BX1605127
BX1605190
BX1605252
BX1605255
BX1605294
BX1605318
BX1605538
BX1605699
BX1605701
BX1605835
BX1606122
BX1606145
BX1606148
BX1606163
BX1606177
BX1701008
BX1701117
BX1701123
BX1701415
BX1701451
BX1701460
BX1701463
BX1701471
BX1701474
BX1701477
BX1701484
BX1701597
BX1701598
BX1701600
BX1701648
BX1701666
BX1701674
BX1701739
BX1701752
BX1701779
BX1701808
BX1701817
BX1701821
BX1701864
BX1701990
BX1702032
BX1702056
BX1702057
BX1702067
BX1702070
BX1702321
BX1702357
BX1702372
BX1702375
BX1702461
BX1702467
BX1702557
BX1702568
BX1702573
BX1702602
BX1702646
BX1702652
BX1702665
BX1702726
BX1702831
BX1702874
BX1702894
BX1703020
BX1703023
BX1703078
BX1703155
BX1703163
BX1703173
BX1703184
BX1703194
BX1703244
BX1703266
BX1703267
BX1703284
BX1703285
BX1703310
BX1703369
BX1703504
BX1703506
BX1703573
BX1703681
BX1703682
BX1703694
BX1703696
BX1703703
BX1703715
BX1703725
BX1703730
BX1703856
BX1703949
BX1704125
BX1704128
BX1704132
BX1704183
BX1704220
BX1704279
BX1704481
BX1704518
BX1704529
BX1704599
BX1704692
BX1704835
BX1704848
BX1704854
BX1704859
BX1704860
BX1704896
BX1704962
BX1704965
BX1704971
BX1705011
BX1705014
BX1705061
BX1705067
BX1705098
BX1705155
BX1705232
BX1705323
BX1705376
BX1705390
BX1705454
BX1705468
BX1705484
BX1705498
BX1705525
BX1705532
BX1705567
BX1705671
BX1705687
BX1705769
BX1705783
BX1705820
BX1705823
BX1705838
BX1705871
BX1705872
BX1705876
BX1705913
BX1705920
BX1705932
BX1705941
BX1706088
BX1706123
BX1706139
BX1801007
BX1801011
BX1801026
BX1801102
BX1801158
BX1801245
BX1801354
BX1801438
BX1801446
BX1801456
BX1801491
BX1801537
BX1801546
BX1801586
BX1801589
BX1801590
BX1801591
BX1801593
BX1801598
BX1801600
BX1801603
BX1801604
BX1801606
BX1801608
BX1801610
BX1801611
BX1801663
BX1801722
BX1801732
BX1802029
BX1802091
BX1802207
BX1802218
BX1802250
BX1802299
BX1802305
BX1802314
BX1802316
BX1802404
BX1802490
BX1802613
BX1802733
BX1802781
BX1802895
BX1802902
BX1802934
BX1802972
BX1803013
BX1803015
BX1803043
BX1803046
BX1803049
BX1803058
BX1803061
BX1803065
BX1803097
BX1803107
BX1803122
BX1803127
BX1803153
BX1803204
BX1803206
BX1803207
BX1803220
BX1803228
BX1803279
BX1803316
BX1803331
BX1803334
BX1803346
BX1803352
BX1803354
BX1803368
BX1803369
BX1803371
BX1803373
BX1803381
BX1803440
BX1803458
BX1803517
BX1803518
BX1803528
BX1803558
BX1803635
BX1803654
BX1803661
BX1803663
BX1803690
BX1803709
BX1803741
BX1803759
BX1803769
BX1803771
BX1803837
BX1803839
BX1803840
BX1803845
BX1803857
BX1803881
BX1803901
BX1803905
BX1803949
BX1803950
BX1803962
BX1804015
BX1804032
BX1804040
BX1804136
BX1804163
BX1804206
BX1804219
BX1804220
BX1804265
BX1804289
BX1804290
BX1804293
BX1804294
BX1804323
BX1804325
BX1804349
BX1804366
BX1804372
BX1804389
BX1804426
BX1804439
BX1804443
BX1804500
BX1804508
BX1804510
BX1804548
BX1804553
BX1804586
BX1804655
BX1804660
BX1804761
BX1804778
BX1804836
BX1804870
BX1804950
BX1804960
BX1805024
BX1805032
BX1805035
BX1805142
BX1805196
BX1805259
BX1805260
BX1805322
BX1805404
BX1805411
BX1805418
BX1805423
BX1805445
BX1805449
BX1805486
BX1805487
BX1805648
BX1805701
BX1805743
BX1805745
BX1805768
BX1805770
BX1805794
BX1805809
BX1805871
BX1805876
BX1805914
BX1805918
BX1805923
BX1805936
BX1805937
BX1805941
BX1806054
BX1806154
BX1806187
BX1806195
bx1806197
BX1806251
BX1806254
BX1806302
BX1806303
BX1806445
BX1806636
BX1901006
BX1901142
BX1901149
BX1901170
BX1901173
BX1901176
BX1901178
BX1901206
BX1901227
BX1901229
BX1901233
BX1901254
BX1901299
BX1901323
BX1901429
BX1901432
BX1901555
BX1901614
BX1901719
BX1901842
BX1901881
BX1901889
BX1901892
BX1901901
BX1901938
BX1901938 (2)
BX1901960
BX1901969
BX1901978
BX1901981
BX1901992
BX1901998
BX1902014
BX1902036
BX1902040
BX1902054
BX1902056
BX1902066
BX1902076
BX1902125
BX1902125-2
BX1902133
BX1902135
BX1902137
BX1902163
BX1902166
BX1902170
BX1902193
BX1902194
BX1902199
BX1902211
BX1902235
BX1902297
BX1902300
BX1902302
BX1902367
BX1902426
BX1902448
BX1902475
BX1902557
BX1902562
BX1902572
BX1902576
BX1902579
BX1902582
BX1902605
BX1902679
BX1902692
BX1902746
BX1902749
BX1902809
BX1902828
BX1902847
BX1902874
BX1902894
BX1902907
BX1902920
BX1902925
BX1902927
BX1902933
BX1902941
BX1902960
BX1902986
BX1902989
BX1903022
BX1903072
BX1903081
BX1903142
BX1903156
BX1903183
BX1903200
BX1903204
BX1903235
BX1903247
BX1903255
BX1903357
BX1903365
BX1903367
BX1903370
BX1903386
BX1903411
BX1903417
BX1903424
BX1903427
BX1903434
BX1903447
BX1903525
BX1903602
BX1903624
BX1903650
BX1903688
BX1903722
BX1903819
BX1903845
BX1903851
BX1903875
BX1903878
BX1903947
BX1903952
BX1903971
BX1903982
BX1904005
BX1904081
BX1904093
BX1904124
BX1904148
BX1904170
BX1904224
BX1904255
BX1904290
BX1904302
BX1904355
BX1904357
BX1904374
BX1904385
BX1904478
BX1904555
BX1904569
BX1904570
BX1904571
BX1904573
BX1904667
BX1904749
BX1904771
BX1904812
BX1904826
BX1904846
BX1904927
BX1904930
BX1904935
BX1905048
BX1905103
BX1905106
BX1905112
BX1905116
BX1905124
BX1905130
BX1905145
BX1905155
BX1905168
BX1905174
BX1905175
BX1905176
BX1905220
BX1905238
BX1905245
BX1905247
BX1905252
bx1905253
BX1905254
BX1905258
BX1905308
BX1905313
BX1905390
BX1905449
BX1905497
BX1905553
BX1905589
BX1905733
BX1905758
BX1905859
BX1905866
BX1905871
BX1905954
BX1905976
BX1905989
BX1906004
BX1906024
BX1906025
BX1906034
BX2001023
BX2001024
BX2001030
BX2001070
BX2001080
BX2001083
BX2001085
BX2001089
BX2001091
BX2001095
BX2001101
BX2001113
BX2001140
BX2001142
BX2001146
BX2001168
BX2001177
BX2001208
BX2001237
BX2001304
BX2001313
BX2001320
BX2001326
BX2001334
BX2001335
BX2001342
BX2001347
BX2001355
BX2001395
BX2001414
BX2001415
BX2001454
BX2001458
BX2001485
BX2001495
BX2001513
BX2001520
BX2001528
BX2001605
BX2001606
BX2001608
BX2001617
BX2001620
BX2001625
BX2001632
BX2001634
BX2001639
BX2001641
BX2001646
bx2001664
BX2001687
BX2001692
BX2001759
BX2001762
BX2001764
BX2001768
BX2001789
BX2001891
BX2001912
BX2001917
BX2001953
BX2001956
BX2001962
BX2001964
BX2001974
BX2001984
BX2001986
BX2001996
BX2002003
BX2002036
BX2002041
BX2002048
BX2002052
BX2002062
BX2002063
BX2002064
BX2002077
BX2002081
BX2002089
BX2002100
BX2002101
BX2002103
BX2002104
BX2002109
BX2002116
BX2002119
BX2002151
BX2002178
BX2002182
BX2002197
BX2002198
BX2002208
BX2002209
BX2002210
BX2002216
BX2002217
BX2002267
BX2002268
BX2002282
BX2002286
BX2002290
BX2002296
BX2002323
BX2002327
BX2002346
BX2002377
BX2002381
BX2002387
BX2002393
BX2002439
BX2002442
BX2002476
BX2002546
BX2002550
BX2002557
BX2002558
BX2002565
BX2002566
BX2002574
BX2002581
BX2002595
BX2002637
BX2002709
BX2002737
BX2002747
BX2002787
BX2002794
BX2002802
BX2002812
BX2002824
BX2002835
BX2002838
BX2002842
BX2002847
BX2002857
BX2002892
BX2002925
BX2002927
BX2002962
BX2002967
BX2002987
BX2003019
BX2003042
BX2003067
BX2003140
BX2003151
BX2003155
BX2003161
BX2003269
BX2003280
BX2003403
BX2003405
BX2003456
BX2003473
BX2003479
BX2003591
BX2003621
BX2003641
BX2003647
BX2003653
BX2003665
BX2003666
BX2003669
BX2003675
BX2003687
BX2003688
BX2003692
BX2003695
BX2003697
BX2003716
BX2003717
BX2003807
BX2003808
BX2003862
BX2003871
BX2003881
BX2003912
BX2003915
BX2003922
BX2003950
BX2003953
BX2003960
BX2003975
BX2003989
BX2003995
BX2004004
BX2004009
BX2004018
BX2004020
BX2004024
BX2004030
BX2004031
BX2004032
BX2004043
BX2004049
BX2004051
BX2004053
BX2004059
BX2004074
BX2004076
BX2004077
BX2004080
BX2004085
BX2004088
BX2004091
BX2004094
BX2004100
BX2004108
BX2004121
BX2004132
BX2004133
BX2004138
BX2004141
BX2004144
BX2004152
BX2004154
BX2004155
BX2004157
BX2004165
BX2004176
BX2004177
BX2004205
BX2004207
BX2004235
BX2004251
BX2004255
BX2004258
BX2004258-2
BX2004260
BX2004283
BX2004289
BX2004305
BX2004309
BX2004314
BX2004322
BX2004332
BX2004338
BX2004371
BX2004386
BX2004388
BX2004413
BX2004426
BX2004443
BX2004446
BX2004454
BX2004455
BX2004463
BX2004465
BX2004473
BX2004476
BX2004478
BX2004479
BX2004480
BX2004482
BX2004497
BX2004500
BX2004509
BX2004525
BX2004554
BX2004604
BX2004618
BX2004624
BX2004625
BX2004636
BX2004643
BX2004653
BX2004699
BX2004708
BX2004717
BX2004779
BX2004788
BX2004790
BX2004796
BX2004800
BX2004805
BX2004806
BX2004810
BX2004869
BX2004870
BX2004892
BX2004918
BX2004938
BX2004990
BX2004991
BX2004999
BX2005003
BX2005014
BX2005018
BX2005036
BX2005042
BX2005069
BX2005072
BX2005084
BX2005093
BX2005097
BX2005100
BX2005115
BX2005117
BX2005125
BX2005130
BX2005131
BX2005133
BX2005136
BX2005145
BX2005146
BX2005151
BX2005152
BX2005153
BX2005154
BX2005158
BX2005175
BX2005179
BX2005181
BX2005193
BX2005199
BX2005203
BX2005223
BX2005224
BX2005226
BX2005233
BX2005234
BX2005240
BX2005246
BX2005250
BX2005255
BX2005271
BX2005278
BX2005280
BX2005286
BX2005289
BX2005295
BX2005314
BX2005315
BX2005317
BX2005321
BX2005324
BX2005328
BX2005329
BX2005348
BX2005410
BX2005424
BX2005450
BX2005452
BX2005518
BX2005528
BX2005540
BX2005543
BX2005547
BX2005548
BX2005550
BX2005565
BX2005573
BX2005576
BX2005578
BX2005579
BX2005585
BX2005587
BX2005588
BX2005594
BX2005601
BX2005627
BX2005635
BX2005680
BX2005681
BX2005690
BX2005691
BX2005692
BX2005693
BX2005696
BX2005697
BX2005700
BX2005884
BX2101019
BX2101023
BX2101027
BX2101032
BX2101068
BX2101080
BX2101092
BX2101095
BX2101147
BX2101152
BX2101156
BX2101166
BX2101170
BX2101200
BX2101250
BX2101294
BX2101297
BX2101385
BX2101386
BX2101389
BX2101392
BX2101393
BX2101394
BX2101451
BX2101463
BX2101468
BX2101469
BX2101478
BX2101479
BX2101480
BX2101499
BX2101512
BX2101514
bx2101519
BX2101528
BX2101549
BX2101550
BX2101553
BX2101554
BX2101557
BX2101567
BX2101581
BX2101583
BX2101597
BX2101600
BX2101601
BX2101635
BX2101642
BX2101652
BX2101657
BX2101659
BX2101678
BX2101685
BX2101686
BX2101690
BX2101694
BX2101695
BX2101696
BX2101699
BX2101700
BX2101702
BX2101707
BX2101716
BX2101720
BX2101721
BX2101723
BX2101727
BX2101744
BX2101753
BX2101756
BX2101760
BX2101767
BX2101782
BX2101788
BX2101815
BX2101830
BX2101841
BX2101853
BX2101856
BX2101859
BX2101861
BX2101909
BX2101913
BX2101930
BX2101931
BX2101938
BX2101959
BX2101970
BX2101990
BX2102037
BX2102051
BX2102056
BX2102063
BX2102066
BX2102074
BX2102090
BX2102091
BX2102104
BX2102109
BX2102123
BX2102126
BX2102132
BX2102150
BX2102153
BX2102155
BX2102168
BX2102173
BX2102177
BX2102178
BX2102188
BX2102192
BX2102331
BX2102348
BX2102375
BX2102394
BX2102396
BX2102437
BX2102451
BX2102459
BX2102460
BX2102464
BX2102466
BX2102476
BX2102497
BX2102503
BX2102606
BX2102662
BX2102686
BX2102697
BX2102703
BX2102715
BX2102716
BX2102753
BX2102780
BX2102799
BX2102816
BX2102821
BX2102822
BX2102853
BX2102868
BX2102876
BX2102881
BX2102893
BX2102919
BX2102920
BX2102922
BX2102923
BX2102931
BX2102947
BX2102954
BX2103002
BX2103003
BX2103012
BX2103028
BX2103059
BX2103078
BX2103183
BX2103186
BX2103191
BX2103203
BX2103210
BX2103221
BX2103224
BX2115131
BX2154512
BX2154520
BX2154521
BX2154526
BX2154653
BX2154660
BX2154718
BX2154719
BX2154731
BX2154738
BX2154740
BX2154741
BX2154742
BX2154743
BX2154745
BX2154751
BX2154757
BX2154769
BX2154770
BX2154780
BX2154792
BX2154793
BX2154796
BX2154799
BX2155056
BX2155083
BX2155087
BX2155203
BX2155210
BX2155211
BX2155212
BX2155218
BX2155220
BX2155242
BX2155394
BX2155409
BX2155411
BX2155611
BX2155619
BX2155624
BX2155685
BX2155687
BX2155701
BX2156131
BX2156141
BX2156237
BX2156326
BX2156328
BX2156618
BX2156623
BX2156763
BX2156923
BX2156927
BX2156943
BX2156978
BX2156980
BX2157060
BX2157150
BX2157151
BX2157168
BX2157193
BX2157307
BX2157319
BX2157325
BX2157329
BX2157351
BX2157352
BX2157358
BX2157612
BX2157636
BX2157646
BX2157879
BX2157891
BX2157897
BX2158012
BX2158111
BX2158115
BX2158356
BX2158369
BX2158387
BX2158568
BX2158648
BX2158670
BX2158827
BX2158986
BX2159265
BX2159280
BX2159289
BX2159596
BX2159708
BX2159713
BX2159754
BX2169840
BX217188
BX2251008
BX2251206
BX2251210
BX2251211
BX2251223
BX2251227
BX2251229
BX2251756
BX2251877
BX2251878
BX2251885
BX2251887
BX2251891
BX2251896
BX2251904
BX2251912
BX2251913
BX2251920
BX2252118
BX2252124
BX2252352
BX2252365
BX2252390
BX2252393
BX2252439
BX2252444
BX2252454
BX2252458
BX2252912
BX2252934
BX2253368
BX2253369
BX2253372
BX2253378
BX2253382
BX2253383
BX2253384
BX2253438
BX2253453
BX2253583
BX2253607
BX2253610
BX2253621
BX2253937
BX2254048
BX2254049
BX2254051
BX2254054
BX2254055
BX2254056
BX2254066
BX2254069
BX2254072
BX2254075
BX2254078
BX2254118
BX2254303
BX2254409
BX2254410
BX2254411
BX2254413
BX2254416
BX2254428
BX2254430
BX2254441
BX2254447
BX2254452
BX2254598
BX2254602
BX2254605
BX2254611
BX2254619
BX2254621
BX2254629
BX2254720
BX2254780
BX2255019
BX2255519
BX2255521
BX2255523
BX2255524
BX2255531
BX2255534
BX2255546
BX2255581
BX2255599
BX2255604
BX2255620
BX2255635
BX2255636
BX2255644
BX2255646
BX2255649
BX2255651
BX2255652
BX2255680
BX2255681
BX2255685
BX2255686
BX2255707
BX2255709
BX2255710
BX2255711
BX2255712
BX2255713
BX2255716
BX2255722
BX2255725
BX2255731
BX2255734
BX2255736
BX2255744
BX2255902
BX2255903
BX2255907
BX2255910
BX2255966
BX2256247
BX2256251
BX2256253
BX2256254
BX2256281
BX2256288
BX2256292
BX2256299
BX2256307
BX2256312
BX2256318
BX2256401
BX2256542
BX2256545
BX2256546
BX2256554
BX2256556
BX2256563
BX2256567
BX2256568
BX2256569
BX2256572
BX2256575
BX2256578
BX2256579
BX2256581
BX2256587
BX2256593
BX2256595
BX2256606
BX2256608
BX2256613
BX2256614
BX2256642
BX2256825
BX2256827
BX2256829
BX2256838
BX2256847
BX2257009
BX2257016
BX2257035
BX2257038
BX2257039
BX2257044
BX2257048
BX2257051
BX2257052
BX2257053
BX2257054
BX2257055
BX2257058
BX2257059
BX2257123
BX2257124
BX2257125
BX2257126
BX2257127
BX2257130
BX2257137
BX2257138
BX2257139
BX2257140
BX2257185
BX2257190
BX2257194
BX2257195
BX2257196
BX2257197
BX2257202
BX2257205
BX2257215
BX2257218
BX2257219
BX2257226
BX2257241
BX2257242
BX2257244
BX2257247
BX2257249
BX2257253
BX2257256
BX2257257
BX2257377
BX2257381
BX2257382
BX2257383
BX2257384
BX2257387
BX2257389
BX2257395
BX2257398
BX2257400
BX2257408
BX2257410
BX2257413
BX2257437
BX2257449
BX2257625
BX2257630
BX2257636
BX2257639
BX2257731
BX2257874
BX2257876
BX2257882
BX2257897
BX2257903
BX2257914
BX2257917
BX2257924
BX2257928
BX2258080
bx2258083
BX2258099
BX2258256
BX2258337
BX2258463
BX2258464
BX2258465
BX2258470
BX2258471
BX2258474
BX2258478
BX2258479
BX2258486
BX2258487
BX2258492
BX2258494
BX2258495
BX2258498
BX2258501
BX2258502
BX2258503
BX2258507
BX2258510
BX2258511
BX2258514
BX2258515
BX2258538
BX2258542
BX2258543
BX2258544
BX2258546
BX2258551
BX2258570
BX2258571
BX2258734
BX2258748
BX2258753
BX2258757
BX2258759
BX2258761
BX2258763
BX2258764
BX2258767
BX2258770
BX2258845
BX2258851
BX2258911
BX2258912
BX2258914
BX2258918
BX2258924
BX2258928
BX2258942
BX2258955
BX2259032
BX2259035
BX2259041
BX2259046
BX2259064
BX2259068
BX2259070
BX2259220
BX2259230
BX2259251
BX2259377
BX2259380
BX2259392
BX2259400
BX2259460
BX2259462
BX2259699
BX2259723
BX2259727
BX2259728
BX2259730
BX2259735
BX2259737
BX2259739
BX2259742
BX2259745
BX2259746
BX2259747
BX2259749
BX2259753
BX2259763
BX2259764
BX2259825
BX2259826
BX2259853
BX2259860
BX2259869
BX2259873
BX2259875
BX2259876
BX2259888
BX2259889
BX2259893
BX2259968
BX2260007
BX2260303
BX2260305
BX2260306
BX2260311
BX2260312
BX2260313
BX2260316
BX2260319
BX2260323
BX2260331
BX2260332
BX2260337
BX2260345
BX2260347
BX2260348
BX2260352
BX2260354
BX2260359
BX2260360
BX2260361
BX2260555
BX2260560
BX2260603
BX2260622
BX2260643
BX2260667
BX2260777
BX2260778
BX2260865
BX2260867
BX2260922
BX2260943
BX2260952
BX2260955
BX2260960
BX2260965
BX2260966
BX2261007
BX2261008
BX2261138
BX2261181
BX2261201
BX2261204
BX2261209
BX2261211
BX2261214
BX2261215
BX2261216
BX2261217
BX2261254
BX2261303
BX2261537
BX2261538
BX2261557
BX2261610
BX2261624
BX2261628
BX2261629
BX2261631
BX2261632
BX2261644
BX2261645
BX2261647
BX2261659
BX2261688
BX2261689
BX2261963
BX2262017
BX2262018
BX2262020
BX2262021
BX2262022
BX2262025
BX2262026
BX2262028
BX2262029
BX2262030
BX2262034
BX2262035
BX2262048
BX2262233
BX2262325
BX2262363
BX2351222
BX2351263
BX2351265
BX2351267
BX2351306
BX2351309
bx2351311
BX2351315
BX2351322
BX2351324
BX2351329
BX2351347
BX2351371
BX2351376
BX2351543
BX2351544
BX2351551
BX2351553
bx2351555
BX2351562
BX2351564
BX2351636
BX2351645
BX2351646
BX2351647
BX2351892
BX2351898
BX2351899
BX2351908
BX2351910
BX2351913
BX2351920
BX2351924
BX2352160
BX2352184
BX2352202
BX2352257
BX2352267
BX2352268
BX2352270
BX2352437
BX2352467
BX2352468
BX2352477
BX2352490
BX2352495
BX2352499
BX2352500
BX2352513
BX2352518
BX2352522
BX2352528
BX2352538
BX2352551
BX2352560
BX2352572
BX2352579
BX2352608
BX2352904
BX2352905
BX2352906
BX2352908
BX2352918
BX2352940
BX2352958
BX2352964
BX2352966
BX2352977
BX2352980
BX2352985
BX2352986
BX2353112
BX2353118
BX2353168
BX2353171
BX2353195
BX2353561
BX2353585
BX2353604
BX2353615
BX2353621
BX2353656
BX2353659
BX2353660
BX2353668
BX2353670
BX2353671
BX2353672
BX2353675
BX2353677
BX2353678
BX2353682
BX2353685
BX2353764
BX2353771
BX2353774
BX2353782
BX2353985
BX2354039
BX2354051
BX2354057
BX2354063
BX2354069
BX2354072
BX2354073
BX2354075
BX2354133
BX2354135
BX2354298
BX2354299
BX2354300
BX2354324
BX2354407
BX2354426
BX2354427
BX2354589
BX2354591
BX2354599
BX2354748
BX2354761
BX2354763
BX2354775
BX2354780
BX2354782
BX2354786
BX2354787
BX2354812
BX2354940
BX2355010
BX2355012
BX2355352
BX2355357
BX2355363
BX2355370
BX2355376
BX2355385
BX2355386
BX2355388
BX2355391
BX2355809
BX2355829
BX2355859
BX2356461
BX2356469
BX2356470
BX2356634
BX2356807
BX2356810
BX2356814
BX2356817
BX2356823
BX2356964
BX2356968
BX2357427
BX2357432
BX2357456
BX2357460
BX2357525
BX2357540
BX2357551
BX2358305
BX2358306
BX2358746
BX2358864
BX2358879
BX2358926
BX2359465
BX2359467
BX2359484
BX2359493
BX2359718
BX2359758
BX2359765
BX2359771
BX2359775
BX2360544
BX2360565
BX2360567
BX2360568
BX2360738
BX2360747
BX2360748
BX2360752
BX2361021
BX2361124
BX2361164
BX2361175
BX2361180
BX2361249
BX2361255
BX2361311
BX2361361
BX2361362
BX2361376
BX2361382
BX2361436
BX2361544
BX2361546
BX2361547
BX2361548
BX2361553
BX2361555
BX2361677
BX2361687
BX2361688
BX2361689
BX2361690
BX2361695
BX2361704
BX2361710
BX2361712
BX2362100
BX2362771
BX2363124
BX2363387
BX2363619
BX2363861
BX2363955
BX2451081
BX2451636
BX2451640
BX2451704
BX2451911
BX2451940
BX2451956
BX2451960
BX2452030
BX2452060
BX2452264
BX2452335
BX2452346
BX2452363
BX2452433
BX2452446
BX2452460
BX2452466
BX2452605
BX2452693
BX2452695
BX2452714
BX2452866
BX2452872
BX2453163
BX2453449
BX2453472
BX2453816
BX2453839
BX2454335
BX2454344
BX2454366
BX2454963
BX2455414
BX2455497
BX2455552
BX2455583
BX2455638
BX2455641
BX2456393
BX2457402
BX2457414
BX2457489
BX2457630
BX2457688
BX2458083
BX2458244
BX2458486
BX2458691
BX2458702
BX2458705
BX2459415
BX2459916
BX2460840
BX2461267
BX2501899
BX2551067
BX2551103
BX2551707
BX2553937
BX2553940
BX2557105
BX2557114
BX2557218
BX2557224
BX2560529
|
| FEI Number |
3036470679
|
Recalling Firm/ Manufacturer |
Datex-Ohmeda Inc 3114 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient. |
FDA Determined Cause 2 | Process control |
| Action | Firm began notifying customers on 5/8/2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION."
Customers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative. |
| Quantity in Commerce | 550 (367 US; 183 OUS) |
| Distribution | Pending |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FMT
|