• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline Convenience Kits

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medline Convenience Kitssee related information
Date Initiated by FirmMay 21, 2026
Date PostedJune 29, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2537-2026
Recall Event ID 99162
Product Classification Cardiovascular procedure kit - Product Code OEZ
ProductMedline Convenience Kits: 1) DYNJ0621189X, Model Number: SM RADIOLOGY-ANGIO PACK-LF; 2) PAIN1470A, Model Number: CONT. 20G EPI MRI COMPATIBLE; 3) TVS4000L, Model Number: TVS4000 ENDOVENOUS PACK
Code Information 1) DYNJ0621189X, UDI-DI: 10195327638566(each), 40195327638567(case), Lot Number: 25ILA890; 2) PAIN1470A, UDI-DI: 10193489957365(each), 40193489957366(case), Lot Number: 25JLA514; 3) TVS4000L, UDI-DI: 10195327223540(each), 40195327223541(case), Lot Number: 25JDB583; 4) TVS4000L, UDI-DI: 10195327223540(each), 40195327223541(case), Lot Number: 25KDA180; 5) TVS4000L, UDI-DI: 10195327223540(each), 40195327223541(case), Lot Number: 25KDB274
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 5/21/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-082-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce333 kits
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-