| | Class 2 Device Recall ICU Medical Plum Duo |  |
| Date Initiated by Firm | June 10, 2026 |
| Date Posted | July 10, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2724-2026 |
| Recall Event ID |
99192 |
| 510(K)Number | K242114 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | ICU Medical Plum Solo Precision IV Pump, v1.0.3, Item Number 400010401; infusion pump |
| Code Information |
UDI/DI +M335400011, Serial Numbers: 60000316, 60000363, 60000375, 60000388, 60000428, 60000451, 60000469, 60000495, 60000529, 60000569, 60000580, 60000637, 60000643, 60000647, 60000652, 60000668, 60000670, 60000689, 60000692, 60000695, 60000703, 60000706, 60000710, 60000712, 60000716, 60000718, 60000724, 60000726, 60000754, 60000757, 60000759, 60000765, 60000775, 60000776, 60000781, 60000790, 60000793, 60000801, 60000805, 60000816, 60000830, 60000835, 60000850, 60000857, 60000860, 60000879, 60000883, 60000886, 60000899, 60000903, 60000909, 60000916, 60000917, 60000920, 60000927, 60000933, 60000938, 60000939, 60000942, 60000944, 60000948, 60000951, 60000954, 60000959, 60000960, 60000964, 60000966, 60000967, 60000980, 60000985, 60000987, 60000988, 60000990, 60001007, 60001008, 60001013, 60001028, 60001030, 60001031, 60001034, 60001035, 60001036, 60001037, 60001038, 60001039, 60001040, 60001041, 60001042, 60001043, 60001044, 60001045, 60001047, 60001048, 60001049, 60001050, 60001051, 60001052, 60001053, 60001054, 60001055, 60001056, 60001057, 60001058, 60001060, 60001061, 60001062, 60001063, 60001064, 60001065, 60001066, 60001067, 60001068, 60001069, 60001070, 60001071, 60001072, 60001073, 60001074, 60001075, 60001076, 60001077, 60001078, 60001079, 60001080, 60001081, 60001082, 60001083, 60001084, 60001085, 60001086, 60001087, 60001088, 60001089, 60001090, 60001091, 60001092, 60001093, 60001094, 60001095, 60001096, 60001097, 60001098, 60001099, 60001100, 60001101, 60001102, 60001103, 60001104, 60001105, 60001106, 60001107, 60001108, 60001109, 60001110, 60001111, 60001112, 60001113, 60001114, 60001115, 60001116, 60001117, 60001118, 60001119, 60001120, 60001121, 60001122, 60001123, 60001124, 60001125, 60001126, 60001127, 60001128, 60001129, 60001130, 60001131, 60001132, 60001133, 60001134, 60001135, 60001136, 60001138, 60001139, 60001140, 60001142, 60001143, 60001144, 60001145, 60001146, 60001147, 60001148, 60001149, 60001150, 60001151, 60001152, 60001153, 60001154, 60001155, 60001157, 60001158, 60001159, 60001160, 60001161, 60001162, 60001163, 60001164, 60001165, 60001166, 60001167, 60001168, 60001169, 60001170, 60001171, 60001172, 60001173, 60001174, 60001175, 60001176, 60001177, 60001178, 60001179, 60001180, 60001181, 60001182, 60001183, 60001184, 60001185, 60001186, 60001187, 60001188, 60001189, 60001190, 60001191, 60001192, 60001193, 60001194, 60001195, 60001196, 60001197, 60001198, 60001199, 60001200, 60001201, 60001202, 60001203, 60001204, 60001205, 60001206, 60001207, 60001208, 60001209, 60001210, 60001211, 60001212, 60001213, 60001214, 60001215, 60001216, 60001217, 60001218, 60001219, 60001220, 60001221, 60001222, 60001223, 60001224, 60001225, 60001226, 60001227, 60001228, 60001229, 60001230, 60001231, 60001232, 60001233, 60001236, 60001237, 60001238, 60001240, 60001243, 60001245, 60001252, 60001253, 60001255, 60001262, 60001266, 60001271, 60001280, 60001285, 60001286, 60001291, 60001293, 60001295, 60001296, 60001297, 60001300, 60001302, 60001303, 60001304, 60001305, 60001306, 60001307, 60001308, 60001309, 60001310, 60001311, 60001312, 60001313, 60001314, 60001315, 60001316, 60001317, 60001318, 60001319, 60001320, 60001321, 60001322, 60001323, 60001324, 60001325, 60001326, 60001327, 60001328, 60001329, 60001330, 60001331, 60001332, 60001333, 60001334, 60001335, 60001336, 60001337, 60001338, 60001339, 60001340, 60001341, 60001342, 60001343, 60001344, 60001345, 60001346, 60001347, 60001348, 60001349, 60001350, 60001351, 60001352, 60001353, 60001354, 60001355, 60001357, 60001358, 60001359, 60001360, 60001361, 60001362, 60001363, 60001364, 60001365, 60001366, 60001367, 60001368, 60001370, 60001371, 60001373, 60001374, 60001375, 60001376, 60001377, 60001378, 60001380, 60001381, 60001382, 60001383, 60001385, 60001386, 60001387, 60001388, 60001390, 60001391, 60001392, 60001393, 60001394, 60001395, 60001396, 60001397, 60001398, 60001400, 60001401, 60001402, 60001403, 60001404, 60001405, 60001406, 60001407, 60001408, 60001409, 60001410, 60001411, 60001413, 60001414, 60001416, 60001417, 60001418, 60001420, 60001421, 60001422, 60001423, 60001424, 60001425, 60001426, 60001427, 60001428, 60001429, 60001430, 60001431, 60001432, 60001433, 60001434, 60001435, 60001436, 60001438, 60001439, 60001443, 60001444, 60001445, 60001446, 60001447, 60001448, 60001449, 60001451, 60001452, 60001453, 60001454, 60001455, 60001456, 60001457, 60001458, 60001459, 60001461, 60001462, 60001463, 60001464, 60001465, 60001466, 60001467, 60001468, 60001469, 60001470, 60001472, 60001473, 60001475, 60001476, 60001477, 60001478, 60001480, 60001481, 60001482, 60001483, 60001484, 60001485, 60001486, 60001487, 60001488, 60001489, 60001490, 60001492, 60001494, 60001495, 60001496, 60001497, 60001498, 60001499, 60001500, 60001501, 60001502, 60001503, 60001504, 60001506, 60001507, 60001508, 60001509, 60001510, 60001511, 60001512, 60001513, 60001514, 60001516, 60001518, 60001519, 60001520, 60001521, 60001522, 60001523, 60001525, 60001526, 60001527, 60001528, 60001529, 60001530, 60001531, 60001532, 60001533, 60001535, 60001537, 60001538, 60001539, 60001540, 60001542, 60001543, 60001544, 60001545, 60001546, 60001547, 60001549, 60001550, 60001551, 60001552, 60001553, 60001554, 60001555, 60001556, 60001558, 60001559, 60001560, 60001561, 60001562, 60001563, 60001564, 60001566, 60001567, 60001568, 60001569, 60001570, 60001571, 60001573, 60001574, 60001575, 60001577, 60001578, 60001579, 60001580, 60001581, 60001582, 60001583, 60001584, 60001585, 60001586, 60001587, 60001588, 60001589, 60001590, 60001591, 60001592, 60001593, 60001594, 60001595, 60001596, 60001597, 60001598, 60001599, 60001600, 60001601, 60001602, 60001603, 60001604, 60001605, 60001606, 60001607, 60001608, 60001609, 60001610, 60001611, 60001612, 60001613, 60001614, 60001615, 60001616, 60001617, 60001618, 60001619, 60001620, 60001621, 60001622, 60001623, 60001624, 60001625, 60001626, 60001627, 60001628, 60001629, 60001630, 60001631, 60001632, 60001633, 60001634, 60001635, 60001636, 60001637, 60001638, 60001639, 60001640, 60001641, 60001642, 60001643, 60001644, 60001645, 60001646, 60001647, 60001648, 60001649, 60001650, 60001651, 60001652, 60001653, 60001654, 60001655, 60001656, 60001657, 60001658, 60001659, 60001660, 60001661, 60001662, 60001663, 60001665, 60001666, 60001667, 60001668, 60001670, 60001671, 60001672, 60001673, 60001674, 60001675, 60001676, 60001677, 60001678, 60001679, 60001680, 60001681, 60001683, 60001684, 60001685, 60001686, 60001687, 60001688, 60001690, 60001691, 60001692, 60001693, 60001694, 60001695, 60001696, 60001697, 60001699, 60001700, 60001701, 60001702, 60001703, 60001704, 60001705, 60001706, 60001707, 60001708, 60001709, 60001710, 60001711, 60001712, 60001713, 60001715, 60001716, 60001718, 60001719, 60001720, 60001721, 60001722, 60001723, 60001724, 60001725, 60001726, 60001727, 60001729, 60001731, 60001732, 60001733, 60001734, 60001735, 60001736, 60001737, 60001739, 60001741, 60001743, 60001744, 60001745, 60001746, 60001747, 60001748, 60001749, 60001750, 60001751, 60001752, 60001753, 60001754, 60001755, 60001756, 60001757, 60001758, 60001759, 60001760, 60001761, 60001763, 60001765, 60001767, 60001770, 60001771, 60001772, 60001773, 60001774, 60001775, 60001776, 60001778, 60001779, 60001780, 60001781, 60001782, 60001783, 60001784, 60001785, 60001786, 60001788, 60001789, 60001790, 60001791, 60001792, 60001793, 60001794, 60001795, 60001796, 60001797, 60001799, 60001800, 60001801, 60001802, 60001803, 60001804, 60001805, 60001806, 60001807, 60001808, 60001809, 60001810, 60001811, 60001813, 60001814, 60001815, 60001816, 60001817, 60001818, 60001819, 60001821, 60001824, 60001825, 60001828, 60001829, 60001830, 60001831, 60001832, 60001833, 60001836, 60001837, 60001838, 60001839, 60001840, 60001842, 60001843, 60001844, 60001845, 60001847, 60001848, 60001852, 60001853, 60001854, 60001856, 60001860, 60001861, 60001862, 60001863, 60001864, 60001865, 60001866, 60001867, 60001869, 60001870, 60001871, 60001872, 60001873, 60001874, 60001875, 60001876, 60001877, 60001878, 60001879, 60001880, 60001881, 60001883, 60001885, 60001887, 60001888, 60001889, 60001890, 60001891, 60001892, 60001893, 60001894, 60001897, 60001899, 60001900, 60001901, 60001902, 60001905, 60001906, 60001908, 60001911, 60001912, 60001913, 60001914, 60001915, 60001916, 60001917, 60001919, 60001920, 60001921, 60001922, 60001923, 60001924, 60001925, 60001926, 60001927, 60001929, 60001930, 60001931, 60001932, 60001935, 60001936, 60001937, 60001938, 60001939, 60001940, 60001941, 60001942, 60001945, 60001947, 60001948, 60001949, 60001950, 60001951, 60001952, 60001953, 60001955, 60001956, 60001957, 60001958, 60001959, 60001960, 60001961, 60001962, 60001964, 60001965, 60001966, 60001967, 60001968, 60001969, 60001970, 60001971, 60001972, 60001973, 60001974, 60001976, 60001978, 60001979, 60001980, 60001981, 60001983, 60001984, 60001985, 60001986, 60001987, 60001988, 60001989, 60001990, 60001991, 60001992, 60001993, 60001994, 60001995, 60001996, 60001998, 60001999, 60002000, 60002001, 60002002, 60002003, 60002004, 60002005, 60002006, 60002007, 60002008, 60002009, 60002010, 60002011, 60002012, 60002013, 60002014, 60002015, 60002016, 60002017, 60002018, 60002019, 60002020, 60002021, 60002022, 60002023, 60002024, 60002025, 60002026, 60002027, 60002028, 60002029, 60002030, 60002031, 60002032, 60002033, 60002034, 60002035, 60002036, 60002037, 60002038, 60002039, 60002040, 60002041, 60002042, 60002043, 60002044, 60002045, 60002046, 60002047, 60002048, 60002049, 60002050, 60002051, 60002052, 60002053, 60002054, 60002055, 60002056, 60002057, 60002058, 60002059, 60002060, 60002061, 60002062, 60002063, 60002064, 60002065, 60002066, 60002067, 60002068, 60002069, 60002070, 60002071, 60002072, 60002073, 60002074, 60002075, 60002076, 60002077, 60002078, 60002079, 60002080, 60002081, 60002082, 60002083, 60002084, 60002085, 60002086, 60002087, 60002088, 60002089, 60002090, 60002091, 60002092, 60002094, 60002095, 60002096, 60002097, 60002098, 60002099, 60002100, 60002101, 60002102, 60002103, 60002104, 60002105, 60002106, 60002107, 60002108, 60002109, 60002110, 60002111, 60002112, 60002113, 60002114, 60002115, 60002116, 60002117, 60002118, 60002119, 60002120, 60002121, 60002122, 60002123, 60002124, 60002125, 60002126, 60002127, 60002128, 60002129, 60002130, 60002131, 60002132, 60002133, 60002134, 60002135, 60002136, 60002137, 60002138, 60002139, 60002140, 60002141, 60002142, 60002143, 60002144, 60002145, 60002146, 60002147, 60002148, 60002149, 60002150, 60002151, 60002152, 60002153, 60002154, 60002155, 60002156, 60002157, 60002158, 60002159, 60002160, 60002161, 60002162, 60002163, 60002164, 60002165, 60002166, 60002167, 60002168, 60002169, 60002170, 60002171, 60002172, 60002173, 60002174, 60002176, 60002177, 60002178, 60002179, 60002180, 60002181, 60002182, 60002183, 60002184, 60002185, 60002186, 60002187, 60002188, 60002189, 60002190, 60002191, 60002192, 60002193, 60002194, 60002195, 60002196, 60002197, 60002198, 60002199, 60002200, 60002201, 60002202, 60002203, 60002204, 60002205, 60002206, 60002207, 60002208, 60002209, 60002210, 60002211, 60002212, 60002213, 60002214, 60002215, 60002216, 60002217, 60002218, 60002219, 60002220, 60002221, 60002222, 60002223, 60002224, 60002225, 60002226, 60002227, 60002228, 60002661, 60002663, 60002664, 60002665, 60002666, 60002667, 60002668, 60002669, 60002671, 60002672, 60002673, 60002674, 60002675, 60002676, 60002677, 60002678, 60002679, 60002680, 60002681, 60002682, 60002683, 60002684, 60002685, 60002686, 60002687, 60002688, 60002689, 60002690, 60002691, 60002692, 60002693, 60002694, 60002695, 60002696, 60002697, 60002698, 60002699, 60002700, 60002701, 60002702, 60002704, 60002705, 60002706, 60002707, 60002708, 60002709, 60002710, 60002711, 60002712, 60002713, 60002714, 60002715, 60002716, 60002717, 60002718, 60002719, 60002720, 60002721, 60002722, 60002723, 60002724, 60002725, 60002726, 60002727, 60002728, 60002729, 60002731, 60002732, 60002733, 60002734, 60002735, 60002736, 60002737, 60002739, 60002740, 60002741, 60002743, 60002744, 60002745, 60002746, 60002747, 60002748, 60002749, 60002750, 60002751, 60002752, 60002754, 60002755, 60002756, 60002757, 60002760, 60002761, 60002762, 60002763, 60002764, 60002765, 60002766, 60002767, 60002768, 60002769, 60002770, 60002771, 60002772, 60002773, 60002774, 60002775, 60002776, 60002777, 60002778, 60002779, 60002780, 60002781, 60002782, 60002783, 60002784, 60002785, 60002786, 60002787, 60002788, 60002789, 60002791, 60002793, 60002794, 60002795, 60002797, 60002798, 60002799, 60002800, 60002801, 60002803, 60002804, 60002805, 60002806, 60002808, 60002809, 60002810, 60002811, 60002812, 60002814, 60002816, 60002817, 60002818, 60002819, 60002820, 60002821, 60002822, 60002823, 60002825, 60002826, 60002827, 60002828, 60002830, 60002831, 60002832, 60002833, 60002834, 60002835, 60002836, 60002837, 60002838, 60002839, 60002841, 60002842, 60002843, 60002844, 60002845, 60002847, 60002850, 60002851, 60002852, 60002853, 60002854, 60002856, 60002857, 60002859, 60002860, 60002861, 60002862, 60002864, 60002865, 60002866, 60002867, 60002868, 60002869, 60002870, 60002871, 60002872, 60002873, 60002874, 60002875, 60002876, 60002877, 60002878, 60002879, 60002880, 60002881, 60002882, 60002884, 60002885, 60002886, 60002888, 60002889, 60002890, 60002891, 60002892, 60002893, 60002894, 60002895, 60002896, 60002897, 60002898, 60002899, 60002900, 60002901, 60002902, 60002903, 60002905, 60002906, 60002907, 60002908, 60002910, 60002912, 60002913, 60002915, 60002917, 60002918, 60002919, 60002920, 60002921, 60002922, 60002923, 60002924, 60002925, 60002926, 60002928, 60002929, 60002930, 60002931, 60002932, 60002933, 60002934, 60002935, 60002936, 60002937, 60002938, 60002939, 60002941, 60002942, 60002944, 60002945, 60002946, 60002947, 60002948, 60002949, 60002950, 60002951, 60002952, 60002953, 60002954, 60002955, 60002956, 60002957, 60002958, 60002959, 60002960, 60002961, 60002962, 60002963, 60002964, 60002965, 60002966, 60002967, 60002968, 60002969, 60002970, 60002971, 60002973, 60002974, 60002975, 60002976, 60002977, 60002979, 60002980, 60002981, 60002982, 60002983, 60002984, 60002985, 60002986, 60002987, 60002988, 60002989, 60002990, 60002991, 60002992, 60002993, 60002994, 60002995, 60002996, 60002997, 60002998, 60002999, 60003000, 60003002, 60003003, 60003004, 60003005, 60003006, 60003008, 60003009, 60003010, 60003011, 60003012, 60003013, 60003014, 60003015, 60003016, 60003017, 60003018, 60003019, 60003020, 60003021, 60003022, 60003023, 60003025, 60003026, 60003028, 60003029, 60003030, 60003031, 60003032, 60003034, 60003035, 60003036, 60003037, 60003038, 60003039, 60003040, 60003041, 60003042, 60003043, 60003044, 60003045, 60003046, 60003047, 60003048, 60003049, 60003050, 60003051, 60003053, 60003054, 60003055, 60003056, 60003057, 60003058, 60003059, 60003060, 60003061, 60003062, 60003064, 60003065, 60003066, 60003067, 60003068, 60003069, 60003070, 60003071, 60003072, 60003073, 60003074, 60003075, 60003076, 60003077, 60003078, 60003079, 60003080, 60003081, 60003082, 60003083, 60003084, 60003085, 60003086, 60003087, 60003088, 60003090, 60003091, 60003092, 60003093, 60003094, 60003096, 60003097, 60003098, 60003099, 60003100, 60003102, 60003103, 60003104, 60003105, 60003106, 60003107, 60003108, 60003109, 60003110, 60003111, 60003112, 60003113, 60003114, 60003115, 60003116, 60003117, 60003118, 60003119, 60003120, 60003121, 60003122, 60003123, 60003124, 60003125, 60003126, 60003127, 60003128, 60003129, 60003131, 60003132, 60003133, 60003134, 60003135, 60003136, 60003137, 60003138, 60003139, 60003140, 60003141, 60003142, 60003143, 60003144, 60003145, 60003146, 60003147, 60003148, 60003150, 60003152, 60003153, 60003154, 60003155, 60003157, 60003158, 60003159, 60003160, 60003161, 60003162, 60003163, 60003164, 60003165, 60003167, 60003168, 60003169, 60003170, 60003171, 60003173, 60003174, 60003175, 60003177, 60003179, 60003180, 60003181, 60003182, 60003183, 60003184, 60003185, 60003186, 60003188, 60003190, 60003191, 60003192, 60003193, 60003197, 60003200, 60003201, 60003203, 60003204, 60003206, 60003207, 60003208, 60003209, 60003210, 60003211, 60003212, 60003213, 60003214, 60003215, 60003216, 60003217, 60003218, 60003220, 60003221, 60003222, 60003223, 60003224, 60003225, 60003226, 60003227, 60003229, 60003230, 60003231, 60003232, 60003233, 60003234, 60003235, 60003236, 60003238, 60003239, 60003240, 60003241, 60003242, 60003244, 60003245, 60003246, 60003247, 60003248, 60003250, 60003251, 60003253, 60003254, 60003256, 60003257, 60003258, 60003259, 60003260, 60003261, 60003262, 60003263, 60003264, 60003265, 60003266, 60003267, 60003270, 60003271, 60003272, 60003273, 60003274, 60003275, 60003276, 60003277, 60003278, 60003279, 60003280, 60003281, 60003283, 60003284, 60003285, 60003286, 60003287, 60003289, 60003290, 60003291, 60003292, 60003293, 60003294, 60003295, 60003296, 60003297, 60003298, 60003299, 60003300, 60003301, 60003302, 60003303, 60003304, 60003305, 60003306, 60003307, 60003308, 60003309, 60003310, 60003311, 60003312, 60003313, 60003314, 60003315, 60003316, 60003317, 60003318, 60003319, 60003320, 60003321, 60003322, 60003323, 60003324, 60003326, 60003327, 60003328, 60003329, 60003330, 60003331, 60003332, 60003333, 60003334, 60003335, 60003336, 60003337, 60003338, 60003339, 60003340, 60003341, 60003342, 60003344, 60003345, 60003346, 60003347, 60003348, 60003349, 60003350, 60003351, 60003352, 60003353, 60003354, 60003355, 60003356, 60003357, 60003358, 60003359, 60003360, 60003362, 60003363, 60003364, 60003365, 60003366, 60003367, 60003368, 60003369, 60003370, 60003371, 60003372, 60003373, 60003374, 60003375, 60003376, 60003378, 60003379, 60003380, 60003381, 60003382, 60003383, 60003384, 60003385, 60003386, 60003387, 60003388, 60003389, 60003390, 60003391, 60003392, 60003393, 60003394, 60003395, 60003396, 60003397, 60003398, 60003399, 60003400, 60003401, 60003402, 60003403, 60003404, 60003405, 60003406, 60003407, 60003409, 60003410, 60003411, 60003412, 60003413, 60003414, 60003415, 60003416, 60003417, 60003418, 60003419, 60003420, 60003421, 60003422, 60003423, 60003424, 60003425, 60003426, 60003427, 60003428, 60003429, 60003430, 60003431, 60003432, 60003433, 60003434, 60003435, 60003436, 60003437, 60003438, 60003439, 60003440, 60003441, 60003443, 60003444, 60003445, 60003446, 60003447, 60003448, 60003449, 60003450, 60003451, 60003452, 60003453, 60003454, 60003455, 60003456, 60003457, 60003458, 60003459, 60003460, 60003461, 60003462, 60003463, 60003464, 60003465, 60003467, 60003468, 60003469, 60003470, 60003471, 60003472, 60003474, 60003475, 60003476, 60003477, 60003478, 60003479, 60003480, 60003481, 60003483, 60003484, 60003485, 60003486, 60003487, 60003488, 60003489, 60003490, 60003491, 60003492, 60003493, 60003494, 60003495, 60003496, 60003497, 60003498, 60003499, 60003500, 60003501, 60003502, 60003503, 60003504, 60003505, 60003506, 60003507, 60003508, 60003509, 60003510, 60003511, 60003512, 60003513, 60003514, 60003515, 60003516, 60003517, 60003518, 60003519, 60003520, 60003521, 60003522, 60003523, 60003525, 60003526, 60003527, 60003528, 60003529, 60003530, 60003532, 60003533, 60003534, 60003535, 60003536, 60003537, 60003538, 60003539, 60003540, 60003541, 60003542, 60003543, 60003544, 60003545, 60003546, 60003547, 60003548, 60003549, 60003550, 60003551, 60003552, 60003553, 60003554, 60003555, 60003556, 60003557, 60003559, 60003560, 60003561, 60003562, 60003563, 60003564, 60003565, 60003566, 60003567, 60003568, 60003569, 60003570, 60003571, 60003572, 60003574, 60003575, 60003576, 60003577, 60003578, 60003579, 60003580, 60003581, 60003582, 60003583, 60003584, 60003585, 60003586, 60003587, 60003588, 60003589, 60003590, 60003591, 60003592, 60003593, 60003594, 60003595, 60003596, 60003597, 60003598, 60003599, 60003600, 60003601, 60003602, 60003603, 60003604, 60003605, 60003606, 60003607, 60003608, 60003609, 60003610, 60003611, 60003612, 60003613, 60003614, 60003615, 60003616, 60003617, 60003618, 60003619, 60003620, 60003621, 60003622, 60003623, 60003624, 60003625, 60003626, 60003627, 60003628, 60003629, 60003630, 60003632, 60003633, 60003634, 60003635, 60003636, 60003637, 60003638, 60003640, 60003641, 60003642, 60003643, 60003644, 60003645, 60003646, 60003647, 60003648, 60003649, 60003650, 60003651, 60003652, 60003653, 60003654, 60003655, 60003657, 60003658, 60003659, 60003661, 60003662, 60003663, 60003664, 60003665, 60003666, 60003667, 60003668, 60003669, 60003670, 60003671, 60003672, 60003673, 60003674, 60003675, 60003676, 60003677, 60003678, 60003679, 60003680, 60003681, 60003682, 60003683, 60003684, 60003685, 60003686, 60003687, 60003688, 60003689, 60003690, 60003691, 60003692, 60003693, 60003694, 60003696, 60003697, 60003698, 60003699, 60003700, 60003701, 60003702, 60003703, 60003704, 60003705, 60003706, 60003707, 60003708, 60003709, 60003710, 60003711, 60003712, 60003714, 60003715, 60003716, 60003717, 60003718, 60003719, 60003720, 60003723, 60003724, 60003725, 60003726, 60003728, 60003729, 60003730, 60003731, 60003732, 60003733, 60003734, 60003735, 60003736, 60003737, 60003738, 60003739, 60003740, 60003741, 60003742, 60003743, 60003744, 60003745, 60003746, 60003747, 60003748, 60003749, 60003750, 60003751, 60003752, 60003753, 60003754, 60003755, 60003756, 60003757, 60003758, 60003759, 60003760, 60003761, 60003762, 60003764, 60003765, 60003766, 60003767, 60003768, 60003769, 60003770, 60003771, 60003772, 60003773, 60003774, 60003775, 60003776, 60003778, 60003779, 60003780, 60003781, 60003782, 60003783, 60003784, 60003785, 60003786, 60003787, 60003788, 60003789, 60003790, 60003791, 60003792, 60003793, 60003794, 60003795, 60003796, 60003797, 60003798, 60003799, 60003800, 60003801, 60003802, 60003804, 60003805, 60003806, 60003807, 60003808, 60003809, 60003810, 60003811, 60003812, 60003813, 60003814, 60003815, 60003816, 60003817, 60003818, 60003819, 60003820, 60003821, 60003822, 60003823, 60003824, 60003825, 60003826, 60003827, 60003828, 60003829, 60003830, 60003831, 60003832, 60003833, 60003834, 60003836, 60003837, 60003838, 60003839, 60003840, 60003841, 60003842, 60003843, 60003844, 60003845, 60003846, 60003848, 60003849, 60003850, 60003851, 60003852, 60003853, 60003854, 60003855, 60003856, 60003857, 60003858, 60003859, 60003860, 60003861, 60003862, 60003863, 60003864, 60003865, 60003866, 60003868, 60003869, 60003870, 60003871, 60003872, 60003873, 60003874, 60003875, 60003876, 60003877, 60003878, 60003879, 60003880, 60003881, 60003882, 60003883, 60003884, 60003885, 60003886, 60003887, 60003888, 60003889, 60003890, 60003891, 60003892, 60003893, 60003894, 60003895, 60003896, 60003897, 60003898, 60003899, 60003900, 60003901, 60003902, 60003903, 60003904, 60003905, 60003906, 60003907, 60003908, 60003909, 60003910, 60003911, 60003912, 60003913, 60003914, 60003915, 60003916, 60003917, 60003918, 60003919, 60003920, 60003921, 60003922, 60003923, 60003924, 60003925, 60003926, 60003927, 60003928, 60003929, 60003930, 60003931, 60003932, 60003933, 60003934, 60003936, 60003937, 60003938, 60003939, 60003940, 60003941, 60003942, 60003943, 60003944, 60003945, 60003946, 60003947, 60003948, 60003949, 60003950, 60003951, 60003952, 60003953, 60003954, 60003955, 60003956, 60003957, 60003958, 60003959, 60003960, 60003961, 60003962, 60003963, 60003964, 60003965, 60003966, 60003968, 60003969, 60003970, 60003971, 60003972, 60003973, 60003974, 60003975, 60003976, 60003977, 60003978, 60003979, 60003980, 60003981, 60003982, 60003983, 60003984, 60003985, 60003986, 60003987, 60003988, 60003989, 60003990, 60003991, 60003992, 60003993, 60003994, 60003996, 60003997, 60003998, 60004000, 60004001, 60004002, 60004003, 60004004, 60004005, 60004006, 60004007, 60004008, 60004009, 60004010, 60004011, 60004012, 60004013, 60004014, 60004015, 60004016, 60004017, 60004018, 60004019, 60004020, 60004021, 60004022, 60004023, 60004024, 60004025, 60004026, 60004028, 60004029, 60004030, 60004031, 60004032, 60004034, 60004035, 60004036, 60004037, 60004038, 60004039, 60004040, 60004041, 60004042, 60004043, 60004044, 60004045, 60004046, 60004047, 60004048, 60004049, 60004050, 60004052, 60004053, 60004055, 60004056, 60004057, 60004058, 60004059, 60004060, 60004061, 60004062, 60004063, 60004064, 60004065, 60004066, 60004067, 60004068, 60004069, 60004070, 60004071, 60004072, 60004073, 60004074, 60004075, 60004076, 60004077, 60004078, 60004079, 60004080, 60004081, 60004082, 60004083, 60004084, 60004085, 60004086, 60004087, 60004088, 60004089, 60004090, 60004091, 60004092, 60004093, 60004094, 60004095, 60004096, 60004097, 60004098, 60004099, 60004100, 60004101, 60004102, 60004103, 60004104, 60004105, 60004106, 60004107, 60004108, 60004109, 60004110, 60004111, 60004112, 60004113, 60004114, 60004115, 60004116, 60004117, 60004118, 60004119, 60004120, 60004121, 60004122, 60004123, 60004124, 60004125, 60004126, 60004127, 60004128, 60004129, 60004130, 60004131, 60004132, 60004133, 60004134, 60004135, 60004136, 60004137, 60004138, 60004139, 60004140, 60004141, 60004142, 60004144, 60004145, 60004146, 60004147, 60004148, 60004149, 60004151, 60004152, 60004153, 60004154, 60004155, 60004156, 60004157, 60004158, 60004159, 60004160, 60004161, 60004162, 60004163, 60004164, 60004165, 60004166, 60004167, 60004168, 60004169, 60004170, 60004172, 60004173, 60004174, 60004175, 60004176, 60004177, 60004178, 60004179, 60004180, 60004181, 60004182, 60004183, 60004184, 60004185, 60004186, 60004187, 60004188, 60004189, 60004190, 60004191, 60004192, 60004193, 60004194, 60004195, 60004196, 60004197, 60004198, 60004199, 60004200, 60004201, 60004202, 60004203, 60004204, 60004205, 60004207, 60004208, 60004209, 60004210, 60004211, 60004212, 60004214, 60004215, 60004216, 60004217, 60004219, 60004222, 60004224, 60004225, 60004226, 60004228, 60004229, 60004230, 60004231, 60004232, 60004233, 60004234, 60004236, 60004237, 60004238, 60004239, 60004240, 60004241, 60004242, 60004243, 60004244, 60004245, 60004246, 60004247, 60004248, 60004249, 60004250, 60004251, 60004252, 60004253, 60004254, 60004255, 60004256, 60004257, 60004258, 60004259, 60004260, 60004261, 60004262, 60004263, 60004264, 60004266, 60004267, 60004268, 60004269, 60004270, 60004271, 60004272, 60004273, 60004274, 60004275, 60004276, 60004277, 60004278, 60004279, 60004280, 60004281, 60004282, 60004283, 60004284, 60004285, 60004286, 60004287, 60004288, 60004289, 60004290, 60004291, 60004292, 60004293, 60004294, 60004295, 60004296, 60004297, 60004298, 60004299, 60004300, 60004301, 60004302, 60004303, 60004304, 60004305, 60004306, 60004307, 60004308, 60004309, 60004310, 60004312, 60004313, 60004314, 60004315, 60004316, 60004317, 60004318, 60004319, 60004320, 60004321, 60004322, 60004323, 60004324, 60004325, 60004326, 60004327, 60004328, 60004329, 60004330, 60004331, 60004332, 60004333, 60004334, 60004335, 60004336, 60004337, 60004338, 60004339, 60004340, 60004341, 60004342, 60004343, 60004344, 60004345, 60004346, 60004347, 60004348, 60004349, 60004351, 60004352, 60004353, 60004355, 60004356, 60004357, 60004358, 60004359, 60004360, 60004361, 60004362, 60004363, 60004364, 60004365, 60004366, 60004367, 60004368, 60004370, 60004374, 60004375, 60004376, 60004377, 60004379, 60004380, 60004381, 60004382, 60004383, 60004385, 60004386, 60004387, 60004390, 60004391, 60004392, 60004394, 60004395, 60004396, 60004397, 60004398, 60004399, 60004400, 60004401, 60004402, 60004403, 60004404, 60004405, 60004406, 60004407, 60004408, 60004409, 60004410, 60004411, 60004412, 60004413, 60004414, 60004415, 60004416, 60004417, 60004418, 60004419, 60004420, 60004422, 60004423, 60004426, 60004427, 60004428, 60004429, 60004431, 60004432, 60004435, 60004436, 60004437, 60004439, 60004440, 60004441, 60004443, 60004444, 60004445, 60004447, 60004448, 60004449, 60004450, 60004451, 60004452, 60004453, 60004454, 60004455, 60004456, 60004457, 60004458, 60004460, 60004461, 60004462, 60004463, 60004464, 60004465, 60004466, 60004468, 60004469, 60004470, 60004471, 60004472, 60004473, 60004474, 60004475, 60004476, 60004477, 60004478, 60004479, 60004480, 60004483, 60004485, 60004486, 60004487, 60004489, 60004490, 60004491, 60004492, 60004493, 60004494, 60004496, 60004497, 60004498, 60004499, 60004501, 60004502, 60004503, 60004506, 60004507, 60004508, 60004509, 60004512, 60004513, 60004514, 60004515, 60004516, 60004518, 60004519, 60004520, 60004521, 60004522, 60004523, 60004524, 60004525, 60004526, 60004527, 60004528, 60004529, 60004530, 60004531, 60004532, 60004533, 60004534, 60004535, 60004536, 60004537, 60004538, 60004539, 60004540, 60004541, 60004542, 60004543, 60004544, 60004545, 60004547, 60004549, 60004550, 60004551, 60004552, 60004553, 60004555, 60004556, 60004557, 60004558, 60004559, 60004560, 60004561, 60004562, 60004563, 60004564, 60004565, 60004566, 60004569, 60004570, 60004572, 60004573, 60004574, 60004575, 60004576, 60004577, 60004578, 60004579, 60004582, 60004583, 60004584, 60004585, 60004586, 60004587, 60004588, 60004590, 60004591, 60004593, 60004594, 60004596, 60004597, 60004598, 60004599, 60004600, 60004601, 60004602, 60004603, 60004604, 60004605, 60004607, 60004609, 60004610, 60004611, 60004612, 60004613, 60004615, 60004616, 60004617, 60004618, 60004619, 60004620, 60004621, 60004622, 60004623, 60004624, 60004625, 60004626, 60004627, 60004628, 60004629, 60004630, 60004631, 60004632, 60004633, 60004634, 60004635, 60004636, 60004637, 60004638, 60004639, 60004640, 60004641, 60004642, 60004643, 60004644, 60004645, 60004646, 60004647, 60004648, 60004649, 60004650, 60004651, 60004652, 60004653, 60004654, 60004655, 60004656, 60004657, 60004658, 60004659, 60004660, 60004661, 60004662, 60004665, 60004667, 60004668, 60004670, 60004683, 60004686, 60004695, 60004699, 60004700, 60004703, 60004704, 60004714, 60004722, 60004723, 60004724, 60004725, 60004726, 60004727, 60004728, 60004729, 60004730, 60004731, 60004733, 60004734, 60004736, 60004737, 60004738, 60004739, 60004740, 60004741, 60004742, 60004743, 60004744, 60004745, 60004746, 60004747, 60004748, 60004749, 60004750, 60004751, 60004752, 60004753, 60004754, 60004755, 60004756, 60004757, 60004758, 60004759, 60004760, 60004761, 60004762, 60004763, 60004764, 60004765, 60004766, 60004767, 60004768, 60004769, 60004770, 60004771, 60004772, 60004773, 60004774, 60004775, 60004776, 60004777, 60004778, 60004779, 60004780, 60004781, 60004782, 60004783, 60004784, 60004785, 60004787, 60004788, 60004789, 60004790, 60004791, 60004792, 60004793, 60004794, 60004795, 60004796, 60004797, 60004798, 60004799, 60004800, 60004801, 60004802, 60004803, 60004804, 60004805, 60004806, 60004808, 60004809, 60004810, 60004811, 60004812, 60004813, 60004814, 60004815, 60004816, 60004817, 60004819, 60004821, 60004822, 60004823, 60004824, 60004825, 60004826, 60004827, 60004828, 60004829, 60004830, 60004831, 60004832, 60004833, 60004834, 60004835, 60004836, 60004837, 60004838, 60004839, 60004840, 60004841, 60004842, 60004843, 60004844, 60004845, 60004846, 60004847, 60004848, 60004849, 60004850, 60004852, 60004853, 60004854, 60004855, 60004856, 60004857, 60004858, 60004860, 60004862, 60004863, 60004864, 60004865, 60004866, 60004867, 60004868, 60004869, 60004870, 60004871, 60004872, 60004873, 60004874, 60004875, 60004876, 60004878, 60004881, 60004882, 60004883, 60004884, 60004885, 60004886, 60004888, 60004889, 60004890, 60004891, 60004892, 60004893, 60004894, 60004895, 60004896, 60004897, 60004899, 60004902, 60004903, 60004904, 60004905, 60004906, 60004907, 60004908, 60004909, 60004910, 60004911, 60004912, 60004913, 60004914, 60004915, 60004916, 60004917, 60004918, 60004919, 60004920, 60004922, 60004923, 60004924, 60004925, 60004928, 60004929, 60004930, 60004931, 60004932, 60004933, 60004935, 60004936, 60004937, 60004938, 60004939, 60004940, 60004942, 60004943, 60004944, 60004945, 60004946, 60004947, 60004948, 60004949, 60004950, 60004951, 60004952, 60004953, 60004954, 60004955, 60004957, 60004958, 60004959, 60004960, 60004961, 60004962, 60004964, 60004965, 60004967, 60004968, 60004969, 60004970, 60004971, 60004972, 60004973, 60004974, 60004975, 60004977, 60004978, 60004979, 60004980, 60004981, 60004982, 60004983, 60004985, 60004986, 60004987, 60004988, 60004989, 60004990, 60004991, 60004992, 60004993, 60004994, 60004995, 60004996, 60004998, 60004999, 60005000, 60005001, 60005002, 60005003, 60005004, 60005005, 60005006, 60005007, 60005008, 60005009, 60005010, 60005011, 60005012, 60005013, 60005014, 60005015, 60005016, 60005017, 60005018, 60005019, 60005020, 60005021, 60005022, 60005023, 60005024, 60005025, 60005026, 60005027, 60005028, 60005029, 60005030, 60005031, 60005032, 60005033, 60005034, 60005035, 60005036, 60005037, 60005038, 60005039, 60005040, 60005041, 60005043, 60005044, 60005045, 60005046, 60005047, 60005048, 60005049, 60005050, 60005051, 60005053, 60005054, 60005055, 60005056, 60005057, 60005058, 60005059, 60005060, 60005061, 60005062, 60005063, 60005064, 60005065, 60005066, 60005067, 60005069, 60005070, 60005071, 60005072, 60005073, 60005074, 60005075, 60005077, 60005078, 60005079, 60005080, 60005081, 60005082, 60005083, 60005084, 60005085, 60005086, 60005087, 60005088, 60005089, 60005091, 60005093, 60005094, 60005095, 60005096, 60005097, 60005098, 60005099, 60005100, 60005106, 60005107, 60005108, 60005110, 60005112, 60005113, 60005116, 60005117, 60005127, 60005131, 60005132, 60005133, 60005134, 60005135, 60005136, 60005138, 60005139, 60005140, 60005141, 60005143, 60005144, 60005145, 60005146, 60005147, 60005149, 60005153, 60005154, 60005156, 60005157, 60005159, 60005160, 60005161, 60005163, 60005164, 60005165, 60005166, 60005167, 60005168, 60005169, 60005170, 60005172, 60005173, 60005175, 60005176, 60005177, 60005178, 60005179, 60005180, 60005182, 60005183, 60005184, 60005185, 60005186, 60005187, 60005188, 60005189, 60005190, 60005191, 60005192, 60005193, 60005194, 60005195, 60005196, 60005197, 60005198, 60005199, 60005200, 60005201, 60005203, 60005204, 60005205, 60005206, 60005207, 60005209, 60005210, 60005212, 60005213, 60005214, 60005215, 60005216, 60005217, 60005218, 60005219, 60005221, 60005222, 60005223, 60005224, 60005225, 60005226, 60005227, 60005228, 60005230, 60005233, 60005236, 60005237, 60005239, 60005240, 60005241, 60005242, 60005243, 60005244, 60005245, 60005247, 60005248, 60005249, 60005250, 60005251, 60005252, 60005253, 60005254, 60005255, 60005256, 60005257, 60005258, 60005259, 60005260, 60005261, 60005262, 60005263, 60005264, 60005265, 60005266, 60005267, 60005268, 60005269, 60005270, 60005271, 60005272, 60005273, 60005274, 60005275, 60005276, 60005277, 60005278, 60005281, 60005283, 60005284, 60005285, 60005286, 60005294, 60005299, 60005301, 60005302, 60005303, 60005304, 60005305, 60005311, 60005313, 60005318, 60005325, 60005327, 60005340, 60005342, 60005365, 60005371, 60005375, 60005376, 60005378, 60005379, 60005382, 60005383, 60005384, 60005385, 60005388, 60005390, 60005391, 60005392, 60005393, 60005394, 60005395, 60005396, 60005397, 60005398, 60005399, 60005400, 60005401, 60005403, 60005404, 60005405, 60005406, 60005407, 60005408, 60005409, 60005410, 60005411, 60005412, 60005413, 60005414, 60005415, 60005416, 60005417, 60005418, 60005419, 60005420, 60005421, 60005422, 60005423, 60005424, 60005425, 60005426, 60005427, 60005428, 60005429, 60005430, 60005431, 60005432, 60005436, 60005437, 60005438, 60005439, 60005440, 60005441, 60005442, 60005444, 60005445, 60005446, 60005447, 60005448, 60005449, 60005450, 60005451, 60005452, 60005453, 60005454, 60005455, 60005456, 60005457, 60005458, 60005459, 60005460, 60005461, 60005462, 60005465, 60005466, 60005467, 60005468, 60005469, 60005470, 60005471, 60005472, 60005473, 60005474, 60005475, 60005477, 60005478, 60005479, 60005480, 60005481, 60005482, 60005483, 60005484, 60005485, 60005487, 60005489, 60005491, 60005493, 60005494, 60005495, 60005496, 60005497, 60005498, 60005499, 60005503, 60005504, 60005505, 60005506, 60005507, 60005508, 60005509, 60005510, 60005511, 60005512, 60005513, 60005514, 60005515, 60005516, 60005517, 60005518, 60005519, 60005520, 60005521, 60005522, 60005523, 60005524, 60005525, 60005526, 60005527, 60005528, 60005529, 60005530, 60005534, 60005535, 60005537, 60005538, 60005539, 60005540, 60005541, 60005542, 60005545, 60005546, 60005547, 60005548, 60005549, 60005550, 60005552, 60005553, 60005554, 60005555, 60005557, 60005558, 60005559, 60005560, 60005561, 60005562, 60005563, 60005564, 60005565, 60005566, 60005567, 60005568, 60005569, 60005570, 60005571, 60005572, 60005573, 60005574, 60005576, 60005577, 60005578, 60005579, 60005580, 60005581, 60005582, 60005583, 60005584, 60005586, 60005587, 60005588, 60005589, 60005590, 60005591, 60005592, 60005593, 60005595, 60005596, 60005597, 60005598, 60005600, 60005601, 60005602, 60005603, 60005604, 60005605, 60005606, 60005607, 60005609, 60005610, 60005611, 60005612, 60005613, 60005614, 60005615, 60005616, 60005617, 60005618, 60005619, 60005620, 60005621, 60005622, 60005624, 60005625, 60005626, 60005627, 60005628, 60005629, 60005630, 60005631, 60005632, 60005633, 60005634, 60005635, 60005637, 60005638, 60005639, 60005640, 60005641, 60005642, 60005643, 60005644, 60005645, 60005646, 60005647, 60005648, 60005649, 60005650, 60005651, 60005652, 60005653, 60005654, 60005655, 60005656, 60005657, 60005658, 60005659, 60005660, 60005661, 60005662, 60005663, 60005664, 60005665, 60005666, 60005667, 60005668, 60005669, 60005670, 60005671, 60005672, 60005673, 60005674, 60005675, 60005676, 60005677, 60005678, 60005679, 60005681, 60005682, 60005683, 60005684, 60005685, 60005686, 60005687, 60005688, 60005689, 60005690, 60005691, 60005692, 60005693, 60005694, 60005695, 60005696, 60005697, 60005698, 60005699, 60005700, 60005701, 60005702, 60005703, 60005704, 60005705, 60005706, 60005707, 60005708, 60005709, 60005710, 60005711, 60005712, 60005713, 60005714, 60005716, 60005717, 60005718, 60005719, 60005720, 60005721, 60005722, 60005723, 60005724, 60005725, 60005726, 60005727, 60005728, 60005729, 60005730, 60005731, 60005732, 60005734, 60005736, 60005737, 60005739, 60005741, 60005742, 60005743, 60005744, 60005745, 60005746, 60005747, 60005748, 60005749, 60005750, 60005751, 60005752, 60005753, 60005754, 60005756, 60005757, 60005758, 60005759, 60005760, 60005761, 60005762, 60005763, 60005764, 60005765, 60005766, 60005767, 60005768, 60005769, 60005770, 60005771, 60005774, 60005775, 60005776, 60005777, 60005778, 60005779, 60005780, 60005781, 60005783, 60005786, 60005790, 60005791, 60005792, 60005793, 60005794, 60005795, 60005796, 60005797, 60005798, 60005800, 60005801, 60005802, 60005803, 60005804, 60005805, 60005806, 60005807, 60005808, 60005809, 60005810, 60005811, 60005812, 60005814, 60005815, 60005816, 60005817, 60005819, 60005820, 60005821, 60005822, 60005823, 60005824, 60005825, 60005827, 60005828, 60005829, 60005830, 60005833, 60005834, 60005835, 60005836, 60005837, 60005838, 60005839, 60005840, 60005841, 60005842, 60005843, 60005844, 60005845, 60005846, 60005847, 60005849, 60005850, 60005851, 60005853, 60005854, 60005855, 60005856, 60005857, 60005858, 60005859, 60005860, 60005861, 60005862, 60005863, 60005864, 60005865, 60005866, 60005867, 60005868, 60005869, 60005870, 60005871, 60005873, 60005874, 60005875, 60005876, 60005877, 60005878, 60005879, 60005880, 60005881, 60005882, 60005883, 60005884, 60005885, 60005886, 60005887, 60005888, 60005889, 60005890, 60005891, 60005892, 60005893, 60005894, 60005895, 60005896, 60005897, 60005898, 60005899, 60005900, 60005901, 60005902, 60005903, 60005904, 60005905, 60005906, 60005907, 60005909, 60005910, 60005911, 60005912, 60005913, 60005914, 60005915, 60005916, 60005917, 60005918, 60005919, 60005920, 60005921, 60005922, 60005923, 60005924, 60005925, 60005926, 60005927, 60005928, 60005929, 60005930, 60005931, 60005932, 60005933, 60005934, 60005935, 60005936, 60005939, 60005941, 60005942, 60005943, 60005944, 60005945, 60005946, 60005948, 60005949, 60005950, 60005951, 60005952, 60005953, 60005954, 60005955, 60005956, 60005957, 60005958, 60005959, 60005960, 60005961, 60005962, 60005964, 60005965, 60005966, 60005967, 60005968, 60005969, 60005970, 60005971, 60005972, 60005973, 60005975, 60005976, 60005977, 60005978, 60005979, 60005980, 60005981, 60005982, 60005983, 60005984, 60005985, 60005986, 60005987, 60005988, 60005989, 60005991, 60005992, 60005993, 60005994, 60005995, 60005996, 60005997, 60005998, 60005999, 60006000, 60006001, 60006002, 60006003, 60006006, 60006007, 60006008, 60006009, 60006010, 60006011, 60006012, 60006013, 60006014, 60006016, 60006017, 60006018, 60006019, 60006021, 60006022, 60006025, 60006026, 60006027, 60006028, 60006029, 60006030, 60006032, 60006034, 60006035, 60006036, 60006037, 60006038, 60006039, 60006040, 60006041, 60006042, 60006043, 60006044, 60006045, 60006047, 60006048, 60006049, 60006050, 60006051, 60006052, 60006053, 60006054, 60006055, 60006056, 60006057, 60006059, 60006060, 60006061, 60006062, 60006063, 60006064, 60006065, 60006066, 60006067, 60006068, 60006069, 60006070, 60006071, 60006072, 60006073, 60006074, 60006075, 60006076, 60006077, 60006078, 60006079, 60006080, 60006081, 60006082, 60006083, 60006084, 60006085, 60006086, 60006087, 60006088, 60006090, 60006091, 60006092, 60006094, 60006095, 60006096, 60006097, 60006099, 60006101, 60006102, 60006103, 60006104, 60006105, 60006106, 60006107, 60006108, 60006109, 60006110, 60006112, 60006113, 60006114, 60006116, 60006117, 60006118, 60006120, 60006121, 60006122, 60006123, 60006124, 60006125, 60006126, 60006127, 60006128, 60006129, 60006131, 60006133, 60006136, 60006137, 60006138, 60006139, 60006140, 60006141, 60006143, 60006144, 60006145, 60006146, 60006147, 60006148, 60006150, 60006151, 60006152, 60006153, 60006155, 60006156, 60006157, 60006158, 60006159, 60006161, 60006163, 60006164, 60006165, 60006166, 60006168, 60006169, 60006170, 60006171, 60006172, 60006173, 60006174, 60006175, 60006177, 60006178, 60006179, 60006182, 60006183, 60006184, 60006185, 60006186, 60006187, 60006188, 60006189, 60006190, 60006191, 60006192, 60006193, 60006195, 60006196, 60006197, 60006198, 60006199, 60006200, 60006201, 60006202, 60006203, 60006204, 60006206, 60006207, 60006208, 60006209, 60006210, 60006211, 60006212, 60006213, 60006214, 60006215, 60006216, 60006218, 60006219, 60006220, 60006221, 60006222, 60006223, 60006224, 60006225, 60006226, 60006227, 60006228, 60006229, 60006230, 60006231, 60006232, 60006233, 60006234, 60006235, 60006236, 60006237, 60006238, 60006239, 60006240, 60006241, 60006242, 60006243, 60006244, 60006245, 60006246, 60006247, 60006249, 60006250, 60006251, 60006252, 60006253, 60006254, 60006255, 60006256, 60006257, 60006258, 60006259, 60006260, 60006261, 60006263, 60006264, 60006265, 60006266, 60006267, 60006268, 60006269, 60006270, 60006271, 60006272, 60006274, 60006276, 60006277, 60006278, 60006279, 60006280, 60006281, 60006282, 60006283, 60006286, 60006287, 60006288, 60006289, 60006290, 60006291, 60006292, 60006293, 60006295, 60006296, 60006297, 60006298, 60006299, 60006300, 60006301, 60006302, 60006303, 60006304, 60006305, 60006306, 60006307, 60006308, 60006309, 60006310, 60006311, 60006312, 60006313, 60006314, 60006315, 60006316, 60006317, 60006318, 60006319, 60006320, 60006321, 60006322, 60006323, 60006324, 60006325, 60006326, 60006327, 60006328, 60006329, 60006330, 60006331, 60006332, 60006333, 60006334, 60006335, 60006336, 60006337, 60006338, 60006339, 60006340, 60006341, 60006342, 60006343, 60006344, 60006345, 60006347, 60006348, 60006349, 60006350, 60006351, 60006352, 60006353, 60006354, 60006355, 60006356, 60006357, 60006358, 60006359, 60006360, 60006361, 60006362, 60006363, 60006364, 60006365, 60006366, 60006367, 60006368, 60006369, 60006370, 60006371, 60006372, 60006373, 60006375, 60006376, 60006377, 60006378, 60006379, 60006380, 60006381, 60006382, 60006383, 60006384, 60006385, 60006386, 60006387, 60006388, 60006389, 60006390, 60006391, 60006392, 60006393, 60006394, 60006395, 60006396, 60006398, 60006399, 60006400, 60006401, 60006402, 60006403, 60006404, 60006405, 60006406, 60006407, 60006409, 60006410, 60006411, 60006412, 60006414, 60006415, 60006416, 60006417, 60006418, 60006419, 60006420, 60006421, 60006422, 60006423, 60006424, 60006425, 60006426, 60006428, 60006429, 60006430, 60006431, 60006432, 60006433, 60006434, 60006435, 60006436, 60006437, 60006438, 60006439, 60006440, 60006441, 60006442, 60006443, 60006444, 60006445, 60006446, 60006447, 60006448, 60006449, 60006450, 60006451, 60006452, 60006453, 60006454, 60006455, 60006456, 60006457, 60006458, 60006459, 60006460, 60006461, 60006462, 60006463, 60006464, 60006465, 60006466, 60006467, 60006468, 60006469, 60006470, 60006471, 60006472, 60006473, 60006474, 60006475, 60006476, 60006477, 60006478, 60006480, 60006481, 60006482, 60006483, 60006484, 60006485, 60006486, 60006487, 60006489, 60006490, 60006493, 60006495, 60006497, 60006498, 60006499, 60006500, 60006501, 60006502, 60006503, 60006504, 60006505, 60006506, 60006507, 60006508, 60006509, 60006510, 60006513, 60006515, 60006516, 60006517, 60006518, 60006519, 60006520, 60006522, 60006523, 60006524, 60006525, 60006526, 60006527, 60006528, 60006529, 60006530, 60006532, 60006533, 60006534, 60006537, 60006538, 60006539, 60006541, 60006543, 60006544, 60006545, 60006546, 60006548, 60006549, 60006550, 60006551, 60006552, 60006553, 60006554, 60006555, 60006557, 60006558, 60006559, 60006561, 60006562, 60006564, 60006565, 60006567, 60006568, 60006571, 60006573, 60006575, 60006576, 60006582, 60006583, 60006586, 60006587, 60006589, 60006590, 60006591, 60006593, 60006595, 60006596, 60006599, 60006602, 60006603, 60006606, 60006607, 60006610, 60006615, 60006616, 60006617, 60006619, 60006622, 60006626, 60006630, 60006632, 60006633, 60006635, 60006637, 60006639, 60006640, 60006643, 60006644, 60006646, 60006649, 60006651, 60006656, 60006659, 60007216, 60007239, 60007268, 60007297, 60007461, 60007482, 60007525, 60007536, 60007544, 60007561, 60007573, 60007577, 60007579, 60007587, 60007590, 60007607, 60007637, 60007651, 60007653, 60007654, 60007656, 60007659, 60007690, 60007692, 60007693, 60007699, 60007700, 60007704, 60007708, 60007714, 60007717, 60007718, 60007719, 60008664, 60008666, 60008748, 60008753, 60008803, 60008836, 60008846, 60008865, 60008866, 60008869, 60008870, 60008872, 60008882, 60008884, 60008887, 60008888, 60008892, 60008895, 60008896, 60008897, 60008899, 60008900, 60008904, 60008907, 60008908, 60008909, 60008910, 60008911, 60008913, 60008917, 60008921, 60008923, 60008941, 60008944, 60008948, 60008951, 60008955, 60008964, 60008965, 60008969, 60008971, 60008974, 60008977, 60008978, 60008980, 60008982, 60008990, 60008991, 60008999, 60009001, 60009014, 60009017, 60009020, 60009023, 60009028, 60009035, 60009036, 60009041, 60009042, 60009044, 60009046, 60009053, 60009055, 60009056, 60009058, 60009062, 60009063, 60009068, 60009069, 60009073, 60009076, 60009078, 60009081, 60009083, 60009084, 60009086, 60009089, 60009090, 60009093, 60009098, 60009101, 60009103, 60009104, 60009106, 60009108, 60009109, 60009110, 60009113, 60009115, 60009116, 60009117, 60009118, 60009119, 60009122, 60009123, 60009124, 60009125, 60009126, 60009128, 60009129, 60009134, 60009135, 60009138, 60009139, 60009140, 60009142, 60009144, 60009145, 60009146, 60009148, 60009149, 60009151, 60009153, 60009155, 60009156, 60009157, 60009159, 60009160, 60009161, 60009162, 60009169, 60009170, 60009171, 60009174, 60009175, 60009176, 60009177, 60009178, 60009179, 60009180, 60009182, 60009184, 60009186, 60009188, 60009190, 60009192, 60009193, 60009196, 60009199, 60009200, 60009202, 60009206, 60009207, 60009211, 60009213, 60009214, 60009216, 60009218, 60009220, 60009224, 60009225, 60009226, 60009229, 60009230, 60009231, 60009232, 60009233, 60009234, 60009235, 60009237, 60009240, 60009241, 60009243, 60009245, 60009246, 60009249, 60009253, 60009254, 60009257, 60009258, 60009259, 60009261, 60009262, 60009264, 60009265, 60009266, 60009271, 60009272, 60009273, 60009274, 60009276, 60009279, 60009280, 60009281, 60009282, 60009283, 60009284, 60009285, 60009287, 60009289, 60009290, 60009294, 60009295, 60009296, 60009297, 60009299, 60009302, 60009303, 60009304, 60009306, 60009307, 60009308, 60009309, 60009311, 60009313, 60009316, 60009317, 60009319, 60009320, 60009322, 60009323, 60009324, 60009326, 60009327, 60009329, 60009331, 60009332, 60009333, 60009334, 60009336, 60009337, 60009338, 60009339, 60009342, 60009344, 60009346, 60009349, 60009351, 60009352, 60009353, 60009357, 60009358, 60009361, 60009364, 60009365, 60009370, 60009372, 60009376, 60009378, 60009384, 60009387, 60009389, 60009390, 60009392, 60009394, 60009395, 60009398, 60009402, 60009403, 60009404, 60009405, 60009407, 60009411, 60009413, 60009414, 60009415, 60009416, 60009417, 60009418, 60009420, 60009422, 60009423, 60009424, 60009425, 60009427, 60009428, 60009430, 60009431, 60009432, 60009433, 60009434, 60009435, 60009436, 60009437, 60009439, 60009440, 60009441, 60009442, 60009443, 60009444, 60009445, 60009446, 60009447, 60009448, 60009449, 60009450, 60009451, 60009452, 60009453, 60009455, 60009456, 60009457, 60009458, 60009459, 60009462, 60009465, 60009466, 60009467, 60009469, 60009470, 60009471, 60009472, 60009473, 60009474, 60009475, 60009476, 60009477, 60009478, 60009479, 60009480, 60009481, 60009482, 60009483, 60009484, 60009488, 60009489, 60009490, 60009491, 60009492, 60009494, 60009495, 60009496, 60009497, 60009498, 60009499, 60009501, 60009502, 60009504, 60009505, 60009508, 60009511, 60009513, 60009514, 60009515, 60009516, 60009517, 60009518, 60009519, 60009520, 60009522, 60009523, 60009524, 60009525, 60009526, 60009529, 60009530, 60009531, 60009532, 60009533, 60009534, 60009535, 60009536, 60009537, 60009538, 60009539, 60009540, 60009541, 60009543, 60009544, 60009545, 60009546, 60009548, 60009550, 60009551, 60009553, 60009554, 60009556, 60009557, 60009567, 60009574, 60009577, 60009578, 60009580, 60009581, 60009584, 60009588, 60009589, 60009590, 60009593, 60009594, 60009596, 60009599, 60009600, 60009602, 60009605, 60009610, 60009651, 60009655, 60009675, 60009686, 60009693, 60009696, 60009707, 60009711, 60009715, 60009718, 60009723, 60009724, 60009727, 60009728, 60009729, 60009737, 60009740, 60009745, 60009758, 60000312, 60000318, 60000323, 60000338, 60000349, 60000362, 60000365, 60000387, 60000410, 60000429, 60000436, 60000482, 60000610, 60000618, 60000659, 60000677, 60000732, 60000748, 60000756, 60000762, 60000772, 60000782, 60000815, 60000823, 60000836, 60000846, 60000847, 60000849, 60000853, 60000861, 60000868, 60000912, 60000914, 60000926, 60000935, 60000955, 60000956, 60000961, 60000971, 60000973, 60000975, 60000977, 60001000, 60001003, 60001015, 60001017, 60001018, 60001259, 60001282, 60001299, 60000306, 60001505. |
| FEI Number |
3013319212
|
Recalling Firm/ Manufacturer |
ICU Medical, Inc. 600 N Field Dr Lake Forest IL 60045-4835
|
| For Additional Information Contact | Mrs. Mary J. Ferreira 845-907-1429 |
Manufacturer Reason for Recall | In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | ICU Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 06/10/2026 via hand delivered letter. The notice explained the issue, potential risk, and requested the following:
"Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification.
If the pump is in the state shown in Figures 1 and 2, do not press the Start button. Wait for the pump to complete the Cassette Test before taking further action. It is important to note that if the clinician waits for the Cassette Test to complete (and the START button is only pressed once), this potential scenario does not occur.
If the pump displays a green background on the line with active animation indicating delivery is occurring and the VTBI is not decrementing, stop delivery and restart the line to resolve this issue.
Complete and return the attached Customer Response Form to ICUMedMarketActionsTeam@icumed.com within 10 days of receipt to acknowledge your understanding of this notification."
For further inquiries:
For any questions regarding this action: https://icumed.custhelp.com/app/market-action.
Additional technical information or assistance: Technical Support, 1-(800)-241-4002, option 3 or tsc.support@icumed.com.
To report adverse events or product complaints: Global Complaint Management: 1-844-654-7780 (M-F, 8:00 am 5:00 pm CT) or ProductComplaintsPP@icumed.com. |
| Quantity in Commerce | 5059 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|