• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall AMS ALIGNED MEDICAL SOLUTIONS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall AMS ALIGNED MEDICAL SOLUTIONSsee related information
Date Initiated by FirmJune 17, 2026
Date PostedJuly 27, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2818-2026
Recall Event ID 99347
Product Classification General surgery tray - Product Code LRO
ProductAMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Loop Recorder Pack 2 -RX, Part Number AMS10164; 2) NY Tray - RX, Part Number AMS10276A; 3) NYC Tray - RX, Part Number AMS10281B; 4) Vein Pack - RX, Part Number AMS11398C; 5) Radiology Tray-RX, Part Number AMS1217A; 6) Custom Pack - RX, Part Number AMS12757B; 7) Biopsy Tray w/Lidocaine - RX, Part Number AMS13700; 8) Arthrogram Tray - RX, Part Number AMS13755A; 9) Myelogram Tray - RX, Part Number AMS14805A; 10) Arthrogram Tray - RX, Part Number AMS15071A; 11) Vein Pack - RX, Part Number AMS15270; 12) Myelogram Tray - RX, Part Number AMS15298; 13) Loop Pack - RX, Part Number AMS15643; 14) Basic Biopsy Tray - RX, Part Number AMS15662A; 15) Pain Pack -RX, Part Number AMS17021; 16) Pain Tray -RX, Part Number AMS17817; 17) Vasectomy Pack - RX, Part Number AMS18226; 18) Arthrogram Tray- RX, Part Number AMS2886(A; 19) Custom Pack - RX, Part Number AMS3081C; 20) Basic Procedure Pack - RX, Part Number AMS5497; 21) Basic Pack - RX, Part Number AMS8130C; 22) Supply Kit - RX, Part Number AMS9521A; 23) Maximal Sterile Barrier Vascular Access Procedure Kit - RX, Part Number VK0003MB.
Code Information Part Number AMS10164, UDI/DI B098AMS101640, Lot Numbers: 230816; Part Number AMS10276A, UDI/DI B098AMS10276A0, Lot Numbers: 230588, 230589, 230590, 230893, 241165; Part Number AMS10281B, UDI/DI B098AMS10281B0, Lot Numbers: 215784, 231259, 241171; Part Number AMS11398C, UDI/DI B098AMS11398C0, Lot Numbers: 230680; Part Number AMS1217A, UDI/DI B098AMS1217A0, Lot Numbers: 217313; Part Number AMS12757B, UDI/DI B098AMS12757B0, Lot Numbers: 240360; Part Number AMS13700, UDI/DI B098AMS137000, Lot Numbers: 217431; Part Number AMS13755A, UDI/DI B098AMS13755A0, Lot Numbers: 230376, 241103; Part Number AMS14805A, UDI/DI B098AMS14805A0, Lot Numbers: 217790; Part Number AMS15071A, UDI/DI B098AMS15071A0, Lot Numbers: 230689; Part Number AMS15270, UDI/DI B098AMS152700, Lot Numbers: 230667; Part Number AMS15298, UDI/DI B098AMS152980, Lot Numbers: 216398; Part Number AMS15643, UDI/DI B098AMS156430, Lot Numbers: 241219; Part Number AMS15662A, UDI/DI B098AMS15662A0, Lot Numbers: 241307; Part Number AMS17021, UDI/DI B098AMS170210, Lot Numbers: 230737, 231023; Part Number AMS17817, UDI/DI B098AMS178170, Lot Numbers: 241136; Part Number AMS18226, UDI/DI B098AMS182260, Lot Numbers: 241031; Part Number AMS2886(A, UDI/DI B098AMS2886(A0, Lot Numbers: 230745; Part Number AMS3081C, UDI/DI B098AMS3081C0, Lot Numbers: 230716; Part Number AMS5497, UDI/DI B098AMS54970, Lot Numbers: 230851; Part Number AMS8130C, UDI/DI B098AMS8130C0, Lot Numbers: 217634, 217635, 231235, 231236, 240963; Part Number AMS9521A, UDI/DI B098AMS9521A0, Lot Numbers: 217055; Part Number VK0003MB, UDI/DI 00813623010666, Lot Numbers: 217429.
FEI Number 1000125955
Recalling Firm/
Manufacturer
Windstone Medical Packaging, Inc.
1602 4th Ave N
Billings MT 59101-1521
For Additional Information Contact
406-259-6387
Manufacturer Reason
for Recall
Windstone Medical Packaging issued a recall for specific item(s) and lot(s) of Aligned Medical Solutions (AMS) convenience kits containing Elevaris Lidocaine HCl injection USP. Elevaris Medical Devices Inc. recalled the Lidocaine HCl injection USP due to supplier quality issues.
FDA Determined
Cause 2
Material/Component Contamination
ActionAligned Medical Solutions (AMS) issues an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 6/17/2026 via email. The notice explained the issue, potential risk, and requested the following: "AMS requests that you undertake the following activities: 1. Immediately check your inventory for the recalled products listed above and place them under quarantine. 2. Due to the manufacturing process any finished sterilized packs remaining in inventory, or at your facility will require over labeling for all products affected by this recall. Packs will be labeled with a sticker attached to each pack. This sticker will read: 'RECALL NOTICE ELEVARIS announced the recall of Lidocaine 5mL Ampules contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. " At the time the kit is opened for use any Elevaris Lidocaine 5mL Ampule(s) should be identified and set aside. " The recalled Elevaris Lidocaine should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution.' 3. Complete the attached Recall Reply Form listing all inventory of affected product by lot number and quantity. 4. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product recall. Your notification to your customers should include a copy of this notification letter. 5. Please complete the Recall Reply Form within 5 business days and return via email to fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand. 6. Indicate on the Recall Reply Form if a replacement Lidocaine Ampules are needed." for questions or concerns, please do not hesitate to contact us at fieldcorrectiveaction@alignedmedicalsolutions.com.
Quantity in Commerce3378 kits
DistributionUS Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IL, KS, LA, MA, MD, MN, MO, NH, NJ, NV, NY, OR, PA, TX, VA, WA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-